key: cord-0155865-ut8janvm authors: Estevez, Dimitri; Mours, Benoit; Pradier, Thierry title: Newtonian calibrator tests during the Virgo O3 data taking date: 2020-11-07 journal: nan DOI: nan sha: 2cba59239e6822530eeafe6de9320742fcda216e doc_id: 155865 cord_uid: ut8janvm The gravitational-wave detectors outputs from the LIGO and Virgo collaborations have been a source of scientific results of prime importance in various domains such as astrophysics, cosmology or fundamental physics. With the upgrades of the detectors and their improved sensitivities, new challenges are set for these instruments calibration. A calibration method based on the local variations of the Newtonian gravitational field could be the next absolute reference of calibration for the interferometers network. We report new tests of Newtonian calibrators (NCal) on the Advanced Virgo detector performed during the LIGO-Virgo observing run O3. The NCal-induced strain on a mirror of the interferometer has been computed both using analytical calculations and numerical simulations with results in very good agreement. The strains given by the numerical model have been used to analyse the data of the NCals and have been compared to the reference method of calibration using photon radiation pressure. New methods to measure the NCal to mirror distance and the NCal offset with respect to the plane of the interferometer using two NCals are also presented and used to correct the NCal data which improves the agreement with the current Advanced Virgo reference of calibration. Accurate calibration of gravitational-wave detectors is critical for proper sky localization, testing waveforms and General Relativity, or extracting measurements like the Hubble constant [1] [2] [3] . The requirements will become even more severe with the planned sensitivity improvements [4] . Calibration using a locally induced variation of the gravity field, using rotating objects is known to be an option. This kind of system, named Newtonian Calibrator or NCal (or sometime GCal), has been first tested on a large interferometer (Virgo) during the so called O2 data taking [6] . Since then, new prototypes have been built and tested at LIGO [5] , KAGRA [7] and Virgo [8] . The calibration of a gravitational-wave detector requires the injection of a well known signal [9] [10] . This means that we need to build a device which will induce a known displacement of a test mass of the interferometer. We also need to accurately predict the injected signal. Following the first tests made on Virgo during O2, new improved NCals have been built for the O3 data taking (April 2019, to March 2020). The modeling of the expected signal has also been developed, both on the analytical side, with a more complete formalism, and on the numerical side, with a new detailed finite element analysis program named FROMAGE. This paper reports these new developments and the achieved results. It starts by the presentation of the Virgo NCal layout used during O3. Next, the analytical and numerical modeling are detailed. The description of the collected data and first checks follow. They are used to derive the NCals positions relative to the mirror location. Using this corrected positions, a check of the h(t) reconstruction is performed and compared to the check made with the reference calibration technique for Virgo, the so called photon calibrator. Finally, the future evolution of the Virgo NCal system is briefly discussed. Following the NCal tests made during the Virgo Observing run O2, a new NCal system was built for O3. The goal was to achieve a higher frequency signal, with a quieter system and reduced systematic uncertainties. A more compact rotor was designed, as described on Figure 1 . It is still made of aluminum for easiness of machining, but with two 90 degrees sectors for optimal strength. The external rotor diameter is 205 mm. Two sets of two cavities are made on both sides of the rotor, with a nominal inner radius of 64 mm, outer radius of 95 mm and 37 mm depth. One of the rotor was measured with a 3d system, and a systematic deviation of the cavities depth of about 0.1 mm was observed on both sides of the rotor. Therefore, when computing the expected displacement, the thickness of the "active" part of the rotor is taken as b = 73.8 mm. The uncertainty on the rotor thickness is set to this correction as a conservative approach. The outer radius of the cavities was found to be systematically larger by 0.045 mm. Again, when computing the radius of the active part of the rotor we used this measure value of r max = 95.045 mm. The uncertainty on r max is this correction. The rotors are made of aluminum 7075, whose expected density is ρ = 2805 kg/m 3 with a typical uncertainty of 5 kg/m 3 , larger than the air density that is neglected. A global test was made by measuring the weight of the rotor: 4.837 kg, which compares very well with the expected CAD value of 4.838 kg, when using the measured rotor parameters. This agreement is an indication that the uncertainties on the rotor parameters are not underestimated. In first approximation, the amplitude of the calibration signal injected by the NCal is proportional to the rotor thickness, density and to the fourth power of the rotor external radius (see equation 8) . Therefore, we can derive the uncertainty on this amplitude from the uncertainties on the main rotor parameters (see Table 1 ). Both sides of the rotor are covered by plastic covers to avoid air motion. The rotor is enclosed in an aluminum housing for safety reason. Instead of driving the rotor with pulleys and belt like for the first prototype, a frameless motor is directly installed on the rotor axis, reducing significantly the vibrations generated by the rotor rotation. The size of the bearing used was reduce to minimize the losses, but their choice and mounting turned out to be difficult, with sometimes not fully reproducible result. The frequency of the rotor as well as the phase information is retrieved using a photodiode sensing and a LED located on the other side of the rotor through a hole made in the rotor. Instead of having the NCal just sitting on the base of the vacuum chamber like during the O2 test, we suspended it from the structure surrounding the vacuum chamber. This locates the NCal on the plane of the interferometer, which significantly reduces the effect of the vertical positioning uncertainty. The knowledge of the mirror to NCal distance is the most difficult part. A direct measurement is impossible, since the mirror is inside the vacuum chamber, without direct view from the NCal position. Furthermore, the suspended solution used to filter the NCal residual vibrations makes this distance not well known. Therefore, we built two NCal devices that, after some tests, were installed on the same side of the vacuum chamber. They are mounted on a rigid rod, with a relative distance of 680 mm. The expected mirror to NCal distance was 1.267 m for the NCal closer to the mirror, or near NCal, and 1.947 m for the far NCal. The angle between the beam axis and the NCals to mirror axis was the same for both NCal: 34.71 degrees. The vertical positions of the two NCals have been measured to be the same within 0.5 mm. The uncertainty on the difference of the phase origins of the two NCals is less than 0.5 mrad, and 1 mrad for each of the individual absolute phase origin. The far NCal was flipped compared to the near NCal, meaning that they were rotating in opposite directions, a feature that allow us to extract the vertical position as described in a following section. The point-mass analytical calculation of the quadrupole force of the Virgo NCal has been refined by including an extended rotor and an extended mirror. Figure 2 (a) and 2(b) show respectively a top view and a side view of the considered NCal configuration. The reference frame is (O, x, y, z) with the center of the mirror in O. The mirror is assumed to be a full cylinder of radius r mir and thickness x mir . The coordinates of a point of the mirror in the reference frame are: (x ; r cos(β); r sin(β)) Here we assume for simplicity that the rotor axis is in the plane of the interferometer (z = 0) but a detailed calculation when z = 0 is given in Appendix A. The coordinates of a point of the rotor sector masses (+, −) in the reference frame are: Defining the small quantities = r d , = r d , = x d and = b d , the longitudinal force induced by the rotor masses (+, −) is: where G is the gravitational constant, m is the mass of the masses (+, −), M is the mass of the mirror, d is the distance from the center of the rotor to the center of the mirror and: The expression from Eq. 2 can be expanded at the fourth order in X ± using: Keeping only the time dependent terms at twice the rotor frequency up to the fourth order and summing the longitudinal force of both rotor sectors, the total longitudinal force F x is: where the odd terms in cos(β), sin(β), and have been omitted for clarity reasons since they cancel when integrating over a small volume element of the mirror in next equations. Considering a small element of the rotor and of the mirror respectively with a mass dm rot = ρ rot rdr dψ db and dm mir = ρ mir r dr dβ dx , the mass of a rotor sector and of the mirror are: It is then possible to obtain the total longitudinal force for an extended rotor and an extended mirror by integrating Eq. 4 using Eq. 5 and Eq. 6. Since the amplitude of an Advanced Virgo mirror motion follows a free test mass response above the resonance frequency of the suspension (f 0.6 Hz), it is expressed as: where f 2rot = 2f rot with f rot the rotor frequency. The resulting amplitude is thus: We can also define a strain which is the amplitude normalized by the interferometer arm length L = 3 km for Virgo: For instance, computing the amplitude of the mirror motion using the Virgo O3 NCal near parameters one can find 1.007136e-14 f −2 2rot m or in strain unit 3.35712e-18 f −2 2rot . The analytical formula for the NCal works well when the rotor and mirror geometry can be described in simple volume elements such as cylinders, sectors, cuboids etc... However, in reality the objects we consider have some imperfections that cannot be modeled in a simple analytical manner. A straightforward way to compute the Newtonian gravitational force for the NCal with an arbitrary geometry is to use a finite element analysis by discretizing the rotor and the mirror into small 3D elements with simple geometrical forms as shown in Figure 3 . This is the aim of the simulation tool, called FROMAGE ‡ [11] , we developed. Computing and adding the gravitational force of each rotor element applied to each mirror element, the total force on the mirror can be derived: with I = (i, j, k) and J = (l, p, q) the indexes of the mirror and the rotor elements respectively, d I,J the distance between the mirror element I and the rotor element J and u I,J a unit vector along d I,J . Eq. 10 can then be projected onto the x axis to obtain the NCal-induced longitudinal force on the mirror: In the case of the Virgo NCal this force is expected to be π−periodic since the two sectors should only induce a quadrupole component (i.e at 2f rot ). However, in order to properly extract any contribution of harmonics in the force, we expand the numerical force in Fourier series such that: with N the maximum order to which the signal is expanded, C k the Fourier coefficients of the force and Φ k the phase of the Fourier coefficients. The results reported further in this paper for the Virgo NCal are indeed the |C 2 | values extracted from Eq. 12. It is also important to define the grid size adapted to our simulations so that it does not induce a significant numerical error on the results. The grid size used to compute the gravitational force with the Virgo NCal has thus been optimized by doing a convergence test. We call (n mir,x , n mir,α , n mir,r ) the mirror grid and (n rot1/2,y , n rot1/2,α , n rot1/2,r ) the grid of a rotor sector. Setting the numerical error on the force at 0.005% per grid parameter we found the following optimized grid for the mirror (12, 30, 8) and for a rotor sector (8, 65 , 40) as shown in Figure 4 . The layout of the Virgo NCal during O3 was made of two rotors respectively at d N = 1.2666 m and d F = 1.9466 m. We first computed the expected amplitude of the mirror motion with FROMAGE for both rotors using simple geometry approximations. The mirror has been set as a full cylinder of radius 175 mm and thickness 200 mm and the rotor has been considered as two opposite cylindrical sectors with the parameters given in Sec. 2. Then we made refined simulations adding details on the geometry such as the flats, the ears and the anchors which are used to suspend the mirror (see Table 2 ). The holes, the screws, the central aluminum plate and the mechanical fillets for the rotor have also been taken into account according to the rotor plans. Table 2 : Parameters to define the flats on the mirror and the cuboids to model the ears and anchors. Table 3 gives the numerical amplitude of the quadrupole contribution for both rotors using a simple and a more realistic geometry. For the simple geometry case, the amplitude given by the analytical formula from Eq. 9 is compared to the numerical results. For the more realistic geometry case, the numerical amplitude is compared to the simple geometry case numerical amplitude. The systematic uncertainty on the computed strain with FROMAGE due to the modeling method could come from three terms: Simple geometry More realistic geometry NCal position Analytical relative error −0.013% −0.004% --Numerical relative error --−0.010% +0.015% Table 3 : Comparison of the computed strain between the numerical and analytical models for the NCal near and NCal far located at a distance d N = 1.2666 m (near) and d F = 1.9466 m (far) respectively at the real angle φ = 0.6058 rad. The relative errors given for the analytical models are computed as ( analytical numerical simple − 1) × 100 and for the numerical models ( numerical real numerical simple − 1) × 100. • The uncertainty coming from the finite resolution of the grid, which is set to 0.005% per grid parameter which results in 0.007% summing all the contributions. This value is an upper limit of the numerical error since it has been computed for the near NCal. • The error coming from the modeling method. This is conservatively taken as the difference between the analytical model and the numerical model (next to last line of Table 3 ). • The error coming from the possible small details of the mirror or rotor geometry that have been ignored. An upper value for this error is the difference between the FROMAGE simulation with the simple geometry and with the more detailed geometry (last line of Table 3 ). Taking the quadratic sum of three terms, we get the following relative systematic uncertainties from the modeling method on the expected injected strain amplitude: 0.018% for the near NCal and 0.017% for the far NCal. One of the benefit of the finite element analysis is to make possible the computation of the torque induced by the NCal on the mirror, even with a complex geometry. The contribution of this torque to the longitudinal mirror motion in the gravitational-wave interferometers appears when the main interferometer beam is not centered on the mirrors. During O3, in Advanced Virgo the position of the beam was controlled to be centered on the mirrors better than ±0.5 mm. Using the maximum values of ±0.5 mm in the y and z directions the contribution to the longitudinal mirror motion is respectively bounded to ±0.05% and ±0.03% for the near and far NCals. From those values, it is then possible to predict the contribution of the torques for any beam offsets since they are linearly dependent. Several data set were collected during O3, that allow us to adjust the NCal calibration procedure. We describe here the last and longer test, six hours of data collected on March 24, 2020 with the Virgo interferometer running at its nominal sensitivity (around 55 Mpc), just before the premature end of the run triggered by the Covid-19 pandemic. During this calibration period, both NCals were used, with their rotation often set to close values in order to extract the mirror to NCal distances, as it will be described later. Figure 5 presents the typical h(t) spectrum observed during the NCals operation, compared to the spectrum taken an hour later, when the NCals were switched off. The differences are marginal, besides the two NCal lines at 55 and 119 Hz. The NCal line at 119 Hz is not a perfect monochromatic, due to small fluctuations of the NCal rotation since the NCal motor was a standalone device with just a simple commercial driver. It is worse noting that despite this time was tagged as calibration time, and therefore not used for regular analysis, the NCals operation is not very invasive with usually just two additional stable calibration lines that could be subtracted. Therefore, if an exceptional event would have been detected during this time, the data could probably have been used. The NCal motor used during O3 was a standalone system, or in other words, the NCal rotation was independent of the Virgo timing system. The phase and rotation speed was therefore extracted from the data, using the pulses of the LED + rotor hole + photodiode system, recorded by the Virgo data acquisition. The time of each individual pulse was computed by taking the average value of the time of the rising and falling fronts. Each of these times is computed by doing a linear interpolation of the time of the samples around the crossing of 50 % of the pulse height, going up or down. This method gives the central value of the pulse and is independent of the rotor speed. Once the pulse times are known, the rotation period is just the time difference between two pulses and the instantaneous frequency the invert of this period. The NCal lines amplitude is extracted by doing the transfer function between the h(t) channel and a signal describing the effective NCal rotation to be immune to small variations of the rotor rotation. For this purpose, a set of sine-waves of unit amplitude is built with the same sampling rate as the h(t) channel. Their frequencies are N times the rotor instantaneous frequency (typical value: N=2). The NCal h(t) amplitude for the harmonic N of the rotor is then computed by reading, at the harmonic frequency, the transfer function value between the reconstructed h(t) and the corresponding sine wave. This transfer function is computed over 20 seconds, meaning that a new value of the NCal h(t) recovered amplitude is available with the same period. The phase is extracted in the same way. Given the position of the hole in the rotor as well as the LED/photodiode position, when the two massive sectors are vertical, the phase is zero. Once the reconstructed NCal line amplitudes are computed and averaged over some time, they can be compared to the expected values or "injected" signal, computed using the FROMAGE model. The result is presented in Figure 6 , where only statistical errors are shown. Deviations from the expected values of 1 for the ratio, and 0 for the phase difference are observed and will be addressed later in the paper. But more interesting at this stage of the study is to notice the differences between the near and far NCals, that are supposed to be identical, coming from the same production batch. Such difference could easily be explained by slight deviations of the NCal position relative to the true mirror location and will be discussed in the following sections. The main discrepancy between the near and far NCal amplitude points may be explained by an offset d 0 in the expected distance between both rotors and the mirror. Since the distance between both NCals is fixed and known, d N F = 680 mm, it is possible to extract a measured distance by comparing the NCal amplitudes and correct the predicted NCal signal accordingly. Let's call A i the amplitude of the mirror motion induced by the NCal i with i = {N, F } at the distance d i + d 0 . The amplitude of the time dependent gravitational force at twice the rotor frequency can be expressed as: with C i the gravitational coupling factor. From Eq. 13 and assuming d 0 d i we can write: with K i the amplitude of the mirror motion computed with a rotor at the estimated distance d i from the mirror. The difference of ratios between the real amplitude A and the estimated amplitude K of both NCals is: One can check that this approximation holds for small offsets d 0 , typically smaller than 1 cm, using FROMAGE. Figure 7 shows the comparison between the numerical results and the approximation of Eq. 15. The linear approximation is in agreement with the numerical results within better than 0.05% for any offset d 0 ≤ 1 cm. Table 4 gathers the far and near NCals h rec /h inj ratios used to compute distance offset d 0 . For each measurement with the far NCal, a couple of measurement at frequencies that bracket the far NCal frequency were performed with the near NCal, to minimize some possible frequency dependent response. Then the average value of the two near NCal measurements at the "low" and "high" frequencies is made and Eq. 15 is used to compute the corresponding d 0 values reported in this Table 4 : Values of the h rec /h inj ratios for the far and near NCals at different frequencies and estimated distance offset d 0 . The three d 0 values report a systematic offset, with an average value of 7.6 mm computed with the exact d 0 values (the table reports rounded values). Removing part of the measurement time do not change significantly each of the d 0 values. The statistical uncertainty on each d 0 , derived from the h rec /h inj errors, is of the order of 1 mm, well below the observed dispersion of the three values. In a normal situation, this would call for extra measurements, that were expected to be made during the end of run period. But due to the Covid-19 pandemic, the end of run was anticipated and the interferometer was shutdown a few days after the measurements reported in this paper. Therefore, the measurement uncertainty associated to d 0 is set to the maximum difference between the mean value and the farthest value: 4.1 mm (here again, the difference has been computed with the exact d 0 values and not the rounded ones from the table). This method of extracting d 0 assumes identical NCals. This could have be checked by swapping the two NCals and repeating the d 0 measurement procedure. But again, it was not possible due to the Covid-19 pandemic. Therefore we use the 0.53% amplitude uncertainty from the rotor geometry as the maximum difference between the signals produced by the two rotors and derive the corresponding error on the measured distance: 4.8 mm. Finally, we do the quadratic sum of both uncertainties and obtain: d 0 = 7.6 ± 6.4 mm 7. Extracting the NCal offset with respect to the plane of the interferometer The phase difference between the near and far NCal points is mainly driven by the offset z of the rotor with respect to the plane of the interferometer. From the calculations in Appendix A, the real analytical phase offset for the NCal i at the first order in d −1 can be approximated by: The rotors have been set on Advanced Virgo such that they spin in opposite direction. They thus have an opposite sign on their phase and it is possible to extract the offset using this phase difference. Let's call ξ i the phase of the mirror motion with the NCal i whose rotor is in the plane of the interferometer (z = 0). In this case, we expect ξ i = 0. Assuming the offset z to be the same for both NCals the difference between the phase points can be computed as: Assuming Ψ N < 0 and Ψ F > 0 without any loss of generality, we can verify that this approximation holds using FROMAGE for small offsets z with typical values smaller than 1 cm. Figure 8 shows the comparison between the numerical results and the approximation of Eq. 17. The linear approximation differs by a quasi-constant value of 0.81% for any offset z ≤ 1 cm. This difference mainly comes from the fact that we did not take into account the second order terms in d −2 in Eq. 16. We can thus normalize the phase for a better approximation by redefining Ψ i as: Table 5 gives the values of the phase at different frequencies for the near and far NCals with the estimated offset z. As for the distance measurements, we compute the average value of these offsets and report it in the first line of Table 6 . This line presents also the phase shifts computed using Eq. 18. The quoted z uncertainty on the second line, that is also translated to phase uncertainties, is the linear sum of two terms. The first one, 1.0 mm, is the maximum difference between the mean value and the farthest z value. The second term is from the mechanical uncertainty in the relative phase origin, 0.5 mrad, which translates to twice this value at the NCal signal frequency and finally to a z systematic error of 0.3 mm. Finally, the last line of the table is the overall systematic phase uncertainty that is the quadratic sum of the second line numbers plus 2 mrad coming for the 1 mrad mechanical absolute phase origin uncertainty, which again should be doubled at the NCal signal frequency. Using the measured distance and vertical offset for both NCals given in previous sections, we recomputed the new expected injected amplitudes and phases of the mirror motion with FROMAGE. These new values have thus been used to correct the data points of the h rec to h inj comparison and are shown in Figure 9 . The error bars reported on the figures are just statistical. Several contributions to the systematic uncertainties on the expected amplitude were discussed in previous sections. The NCal to mirror angle uncertainty coming from the NCal positioning is expected Table 6 : Value of the measured z and its systematic uncertainty. The change in the phase difference h rec − h inj for both NCals is also reported with the associated systematic uncertainties as well as the overall uncertainties. to be small. But since there is a large measured offset in the NCal to mirror distance, we assumed the same offset as maximum possible lateral shift of the NCals and derived the corresponding angles for the two NCals. Table 7 summarizes them, and presents the overall values. The systematic uncertainty of the NCal phase is expected to primarily come from the uncertainty on the NCal signals readout timing of 10 µs, which translates to a phase uncertainty of 6 mrad at 100 Hz to be added with the uncertainty given in Table 6 . The NCal points are compared to the reference method of calibration for Advanced Virgo during O3 using Photon Calibrators (PCal) [12] . The principle of PCals is to induce a known mirror motion using radiation pressure of auxiliary lasers. The systematic uncertainty associated to the NE PCal points used in the figure is 1.4%. The frequency dependent shape of the NCal amplitude points is similar to the PCal ones but a difference of about 3% is observed between both techniques. This difference, is not incompatible with the sum of the systematic uncertainties of the two methods, that are expected to be frequency independent. On the phase, the NCal and PCal points are in good agreement, both regarding the shape and the absolute values. A new version of NCal was successfully tested on the Virgo detector during O3. It was possible to check the h(t) reconstruction up to 120 Hz, and to confirm the shape observed with the reference method that uses a photon calibrator. A difference of about 3% was observed, that is of the order of the sum of the systematic uncertainties of both methods. The anticipated end of the O3 data taking due to the Covid-19 pandemic prevented us to perform Table 7 : Uncertainties on calibration signal amplitude all the needed measurements, like determining the relative strain induced by the two NCals, the dominant term in our amplitude systematic uncertainty. Nevertheless, the achieved amplitude systematic uncertainty for the far NCal was the same as for the PCal, confirming the potential of the NCal technique. We are planning to build additional and slightly upgraded NCals for the upcoming O4 data run, scheduled to start in 2022. Having more of them, especially on both sides of the mirror, will allow us to have a better measurement of the mirror to NCal distance and bring the systematic error well below 1%. where: C = 9G 8d 4 ρ rot b(r 4 max − r 4 min ) sin(α)ρ mir πr 2 mir x mir cos(φ)(A 2 + B 2 ) 1/2 (39) The amplitude of the supended mirror motion at twice the rotor frequency follows a free test mass response above the resonant frequency of the suspension (f 0.6 Hz): One can notice that when the center of the rotor is in the plane of the interferometer (z = 0), the phase shift of the force is Ψ = 0 and the amplitude of the force is C = A. The amplitude of the mirror motion at twice the rotor frequency is eventually: GWTC-1: A Gravitational-Wave Transient Catalog of Compact Binary Mergers Observed by LIGO and Virgo during the First and Second Observing Runs Tests of general relativity with the binary black hole signals from the LIGO-Virgo catalog GWTC-1 A gravitational-wave standard siren measurement of the Hubble constant Prospects for Observing and Localizing Gravitational-Wave Transients with Advanced LIGO, Advanced Virgo and KAGRA First Results from the LIGO Gravitational Calibrator First Tests of a Newtonian Calibrator on an Interferometric Gravitational Wave Detector Improving the absolute accuracy of the gravitational wave detectors by combining the photon pressure and gravity field calibrators Analysis of the NCal data collected during O3B, Virgo TDS Calibration of Advanced Virgo and Reconstruction of the Gravitational Wave Signal h(t) during the observing run O2 Characterization of systematic error in Advanced LIGO calibration Numerical simulations for Virgo O3 NCal using FROMAGE, Virgo TDS, VIR-0759B-20 The Advanced Virgo Photon Calibrators We are indebted to the Virgo Collaboration for allowing us to use the data collected during the tests reported here, and are grateful for support provided by the Virgo Collaboration and the European Gravitational Observatory during those tests. We thank our colleagues in the Virgo Collaboration and in the LIGO Scientific Collaboration for useful discussions. We thank the technical staff at LAPP for their help in building the O3 NCals, especially B. Lieunard for the mechanical design and installation, L. Journet for the assembly, G Deleglise for the metrology and S. Petit for the motor control. We also thanks D. Dangelser from IPHC for improving the NCal support system. We provide here a detailed calculation of the Virgo NCal-induced mirror amplitude in an off-axis configuration and with an offset z with respect to the plane of the interferometer. The coordinates of a point of the rotor masses (+, −) are:and the coordinates of a point of the mirror are:The gravitational force induced by a point of a rotor sector on a point of the mirror along the x axis is given by:where G is the gravitational constant, m is the mass of the masses (+, −), M is the mass of the mirror, d is the distance from the center of the rotor to the center of the mirror in the plane of the interferometer and D is the distance between a point of a rotor sector and a point of the mirror:We define the small quantities = r d , = r d , = x d , = b d and z = z d , the longitudinal force then becomes:where:The expression from Eq. 21 can be expanded at the fourth order in X ± using:Let's define: and since they further cancel when integrating over the mirror and the rotor and defining Θ = θ + ψ we get:Y 2 ≈ cos(φ) 8 2 cos 2 (Θ) + 8 4 cos 2 (Θ) + 8 2 2 cos 2 (Θ) + 24 2 2 cos 2 (Θ) + 8 2 2 cos 2 (Θ) + 24 2 2 cos 2 (φ) cos 2 (Θ) + 16 2 z cos(Θ) sin(Θ) + 24 2 2 sin 2 (φ) cos 2 (β) cos 2 (Θ) + 8 2 2 sin 2 (β) sin 2 (Θ) (29) Y 3 ≈ cos(φ) 24 4 cos 2 (Θ) + 24 2 2 cos 2 (Θ) + 72 2 2 cos 2 (Θ) + 24 2 2 cos 2 (Θ) + 16 4 cos 4 (Θ) + 144 2 2 cos 2 (φ) cos 2 (Θ) + 24 2 z 2 cos 2 (Θ) + 144 2 2 sin 2 (φ) cos 2 (β) cos 2 (Θ) (30) Y 4 ≈ cos(φ) 32 4 cos 4 (Θ) + 192 2 2 sin 2 (φ) cos 2 (β) cos 2 (Θ) + 192 2 2 cos 2 (φ) cos 2 (Θ)The total time dependent longitudinal force can then be expressed as:The mass of a rotor sector and of the mirror are: with ρ rot and ρ mir respectively the density of the rotor material and the mirror material. Integrating Eq. 32 using Eq. 33 and Eq. 34 the NCal longitudinal force becomes: The force can thus be expressed as: