key: cord-0077049-1fwqaovu authors: Sengupta, Rudrarup; Khand, Heena; Sarusi, Gabby title: Terahertz Impedance Spectroscopy of Biological Nanoparticles by a Resonant Metamaterial Chip for Breathalyzer-Based COVID-19 Prompt Tests date: 2022-04-04 journal: ACS Appl Nano Mater DOI: 10.1021/acsanm.2c00954 sha: 1f8f6310a76f99e4466a782afb2ecfa41404029f doc_id: 77049 cord_uid: 1fwqaovu [Image: see text] We propose a tested, sensitive, and prompt COVID-19 breath screening method that takes less than 1 min. The method is nonbiological and is based on the detection of a shift in the resonance frequency of a nanoengineered inductor–capacitor (LC) resonant metamaterial chip, caused by viruses and mainly related exhaled particles, when performing terahertz spectroscopy. The chip consists of thousands of microantennas arranged in an array and enclosed in a plastic breathalyzer-like disposable capsule kit. After an appreciable agreement between numerical simulations (COMSOL and CST) and experimental results was reached using our metamaterial design, low-scale clinical trials were conducted with asymptomatic and symptomatic coronavirus patients and healthy individuals. It is shown that coronavirus-positive individuals are effectively screened upon observation of a shift in the transmission resonance frequency of about 1.5–9 GHz, which is diagnostically different from the resonance shift of healthy individuals who display a 0–1.5 GHz shift. The initial results of screening coronavirus patients yielded 88% agreement with the real-time quantitative polymerase chain reaction (RT-qPCR) results (performed concurrently with the breath test) with an outcome of a positive predicted value of 87% and a negative predicted value of 88%. During the last 2 years, the coronavirus and its variants have caused a global upsurge in the number of people being infected with COVID-19. The disease courses of COVID-19 range from asymptomatic cases to mild (e.g., dry cough and fever) and severe cases (e.g., pneumonia) and even to acute respiratory distress syndrome. In a metaanalysis of 95 studies covering 29,776,306 tested individuals, 0.25% (95% CI, 0.23− 0.27%) had asymptomatic infection and 40.50% (95% CI, 33.50−47.50%) of 19,884 confirmed individuals in this study were asymptomatic. 1 The high percentage of asymptomatic infections highlights the potential transmission risk of asymptomatic infections in communities. Furthermore, asymptomatic patients are considered a major source of COVID-19 spread. In the wake of this development, there is a tremendous need for prompt, large-scale, accurate, and costeffective screening detection of COVID-19 asymptomatic as well as symptomatic patients. As life gradually returns to normal, there is a great need for monitoring asymptomatic cases in order to decrease or prevent cluster outbreaks and transmission. The availability of a rapid, noninvasive, largescale, accurate, and cost-effective COVID-19 screening test is paramount. Current biological-based methods, such as polymerase chain reaction (PCR) tests, require more than 4 h to obtain the result. In parallel, a plethora of new COVID-19 rapid test techniques with varying efficiency have emerged, but these are not always commensurate with the real-time polymerase chain reaction (RT-PCR) standard. 2−4 In this work, we propose an innovative, prompt, noninvasive, nonbiological COVID-19 screening test. This test will be used at entry points, such as airports, stadiums, and shopping malls, and will enable quick differentiation between COVID-19 positive and negative cases. In general, spectroscopy across the electromagnetic frequencies is a powerful tool for analyzing chemical and biological materials. The most commonly used spectral ranges for spectroscopy are ultraviolet, visible, and infrared radiation. Terahertz (THz) spectroscopy is a relatively recent implementation toward analyzing chemical and biological materials, apart from nanoparticles. 5, 6 Using metamaterials in conjunction with THz spectroscopy is an important technique toward bridging the THz gap by providing a strong response to THz radiation of the inspected material and for lowconcentration nanoparticles, where most natural materials exhibit only weak responses. 7−11 With recent advances in THz spectroscopy, the interaction of THz radiation with metallic micro-and nanostructures has provided insights regarding the properties of nanoparticles, in general, and viruses, in particular. 7, 12 Engineered metamaterial nanostructures have been used to determine the dielectric properties of chemical and biological materials when introduced into metamaterial metallic structures, which has led to various innovative techniques for virus, bacteria, and fungus detection. 13−18 These methods constitute the fundamental concept of inductor−capacitor (LC) resonance-based micro-and nanoparticle testing. The metallic nanostructure is physically designed as an inductor and capacitor element, where its fundamental resonant frequency is F L C Here, the factor L (the inductance) is dictated by the geometric parameters of the fabricated inductor's nanostructure, while C (the capacitance) is highly dependent on the plate area (length and thickness), capacitive gap (W), and effective dielectric constant (ε eff ) of the material introduced inside the capacitor gap. A change in the resonant frequency of the metamaterial structure can be brought about by any foreign substance deposited in the capacitive gap region, thereby changing the ε eff and thus the capacitance, resulting in a red shift in the resonance frequency (ΔF) with respect to the pristine LC circuit in the array. Because these microstructures have an effective inductance in the nanohenry range and a capacitance in the femtofarad range, they are extremely sensitive to any decimal-point change in ε eff , thereby giving an indicative ΔF with the deposition of any nanoparticle or biological particle inside the capacitor gap. 17, 19 Moreover, the effectiveness and sensitivity of the metamaterial structure in detecting a specific type of particle can be enhanced by engineering the outer dimensions, capacitive area, and gap of the metamaterial, which makes the metamaterial structure more physically sensitive to a specific genre of foreign substances of a specific dimension. This phenomenon has been demonstrated multiple times, where the F 0 value of the metamaterial-based sensor shows a strong dependence on the capacitive gap as well as on any foreign particle present there. 19 Park et al. 19 demonstrated a sensitive bacteriophage detection using a THz-sensitive metamaterial made of classical, linear polarization, split-ring resonators (SRRs) with various capacitive gap widths. This THz SRR is actually a microantenna array of specific geometry-dependent inductance and capacitance, giving a discrete resonance frequency in the THz range. Upon deposition of the bacteriophages on the metamaterial sensors' surface, the collective resonance frequency of the microantenna array is red-shifted because of an increase in ε eff inside the capacitor gap. However, in order to adequately obtain a substantial ΔF, as required for highsensitivity detection of the bacteriophages, the proposed method 19 requires a very high concentration of bacteriophages (about 10 9 copies/mL), a concentration that can only be obtained in a laboratory experiment. If we consider using the above method to detect infected individuals having viruses related to a respiratory system infection, such an extremely high viral density cannot be obtained by any practical methodneither by a swab nor by a breath test. A recent study revealed that the viral load present in a breath or cough sample is 3000−15000 copies/mL of exhaled air for a low-tomoderate viral load and up to 2.6 × 10 6 ± 1.7 × 10 6 copies/ mL for a very high viral load. 19 For comparison, even for a very high viral load, 20,21 a nasopharyngeal swab sample provides a maximum of 16000−80000 copies/mL. Therefore, to attain rapid testing of coronavirus carriers, out of the two methods, the most convenient and noninvasive method would undoubtedly be a breath analysis, where the exhaled breath of a tested individual is used for detection. In addition, detection of only the viruses may provide a very small spectral shift because of their low density, which can be obtained inside the capacitor gap. Further, a variety of THz metamaterial resonator structures, including SRRs, 13−19 are polarization-dependent (linear polarization nanoantenna), which necessitates careful orientation of the metastructure when using a linear-polarized spectrometer and adds to the complexity of rapid testing because of the alignment requirements. Nevertheless, because these engineered metamaterials have been extensively used for a range of bacteria and fungus detection, their use could be a promising step toward the detection of coronavirus as well, provided that a new sensitive, resonant, polarization-independent metamaterial structure design is adopted. Our work presents the detection of a variety of nanometer-to-submicron biological particles and viruses at low concentrations, with sensitivity enhancement provided by a specific polarization-independent metamaterial structure. More particularly, this paper aims to demonstrate prompt, noninvasive screening of non-COVID-19 carriers and other respiratory-related viral diseases through a breath test with a newly engineered and sensitive LC resonant metamaterial structure in considerably lesser time than that of currently available tests. 22 Our hypothesis is that any respiratory-infected subject (viral or bacterial) exhales a vast amount of biological micro-and nanoparticle-based ingredients associated with his/ her disease during a breath test; these particles can be detected and analyzed using a metamaterial-based chip and THz spectroscopy. These particles could be cell debris, virus debris, cytokines, and other related proteins, fat molecules, and clusters associated with virus proliferation inside the respiratory system. Therefore, we hypothesize that this method will be sufficiently efficient to detect a combination of viruses and exhaled biological particles related to respiratory diseases that will give a specific ΔF signature, even at concentrations as low as those found in reported exhaled breath statistics. 19 ACS Applied Nano Materials www.acsanm.org Article ■ METAMATERIAL DESIGN CONSIDERATIONS, For the detection of COVID-19 carriers using metamaterialbased structures, we physically redesign the THz LC resonant geometry with the intention of maximizing the capacitor gap area. The capacitor gap (width) can be reduced in order to obtain a plasmonic enhancement that is associated with the nanometric scale of such a capacitor gap. 7 Placing the capacitor gaps at both geometric diagonals of a square inductor enables the exhaled viruses and particles to be detected in both the S and P polarization states (when using a single polarization spectrometer) with a rectangular all-around singular inductor structure (rectangular side arm), as shown in Figure 1a ,b, for its resonance detection. The initial concept of this "crossarrowhead" structure was inspired by a single-symmetry SRR unit cell metamaterial and based on a study that dealt with various structures of THz antennas, in general. 23−25 This improvement enables breathalyzer-based coronavirus and related particle detection by increasing the capacitive gap lengths compared to those of the basic SRR; 19 this effectively increases the sensitivity by increasing the probability of viruses/particles falling inside the capacitive gap and yet maintaining a submicron capacitor gap, thereby enabling more pronounced resonance frequency shifts for breath samples even with a lower virus load. In typical conditions, once an infected patient exhales onto the chip, the coronavirus and related particles will spread all over the metamaterial chip surface (which contains a plurality of nanoantenna metamaterial structures), but the virus/ particles that are deposited on the capacitive gap areas will only change the resonant frequency due to the change in ε eff . Moreover, the thousands of such nanoantenna structures in a few-millimeter-sized chip enables combined field enhancement, giving rise to a considerable ΔF value for COVID-19 carrier detection. The concentration of viruses only will be very low in each breath or cough test. Therefore, according to our hypothesis, the effective dielectric constant that will be measured by the metamaterial structures is a combination of all of the exhaled particles from an infected patient, which will be considerably different between an infected and a healthy person, yielding a ΔF range for coronavirus-infected patients that can be easily distinguished from that of healthy persons. Because this is a measurement of the electrical parameters of the metamaterial that is changed after the breath test, we do not aim to distinguish between any specific genetic material (variant) related to the coronavirus using this screening method. This technology proposition is expected to detect the entire combination of exhaled viruses and particles of all virus variants to screen healthy and infected individuals. If we can effectively distinguish coronavirus-infected patients from healthy ones, we can hypothesize that the collateral damage in the respiratory system due to different coronavirus variants is the same and that all infected patients exhale similar biological ingredients and particles because of the viruses' population multiplication inside the respiratory system regardless of the variant type. Therefore, the method will be equally efficient for any mutated variation of the coronavirus. In addition, this method is currently targeted to serve as a Conceptual design of our COVID-19 breathalyzer test, indicating a step-bystep process of the test: (c) the cap is unscrewed from the capsule, which reveals (d) our designed LC resonant metamaterial chip, containing thousands of nanoantennas; (e) the capsule is then screwed onto the mouthpiece, (f) facilitating the individual to blow into the capsule on the metamaterial chip surface; (g) finally, the mouthpiece is thrown away, and the capsule cap is screwed back in its place. The capsule, as shown in part g, can be placed in a THz spectrometer for testing after external sterilization. ACS Applied Nano Materials www.acsanm.org Article screening test at point of entries and not a diagnostic test. A screening test identifies "who is healthy" with a high accuracy, unlike a diagnostic test, which diagnoses "who is sick" along with the "type of sickness". Our method will likely identify any respiratory-infected individual as positive regardless of whether they have COVID-19, influenza, or another viral respiratory disease. At this moment, we do not differentiate between different types of respiratory diseases, because we want to be more specific about SARS-CoV-2, and obtain a specific threshold for "infected" versus "healthy" patients, so that we can separate healthy subjects from infected ones by a prompt screening test. In the case where the individual is found to be positive, he/she may then undergo a diagnostic biologicalbased test such as antigen or PCR. For the breath test, we designed a plastic breathalyzer-like kit, semitransparent to THz frequencies, that consists of a capsule housing for the nanoantenna chip and a mouthpiece, which directs the air blown by an individual onto the nanoantenna chip surface. In our prompt coronavirus breathalyzer test shown schematically in Figure 1c −g, after the individual blows onto the chip surface enclosed in a capsule, the capsule is placed in a THz spectrometer for scanning in transmission mode within a predefined frequency range around the expected resonance frequency. The resonance region of the LC resonant nanoantenna structure is characterized by a definitive dip in the transmission amplitude at the resonance frequency, which is compared with the reference resonance frequency of the pristine LC resonant chip to obtain the ΔF value. We aim to design our entire breathalyzer test to take less than 1 min to predict the results from the moment an individual blows air into the breathalyzer kit and to employ it as a screening tool at points of entry. Finite-element COMSOL multiphysics simulations were initially performed to evaluate and predict the resonance frequency in transmission mode of a pristine single-element LC resonator. To incorporate the dielectric environment of the nanocircuit, an effective refractive index, ε eff , was introduced in the simulation. Here, the initial ε eff during the reference measurement was a linear combination of the dielectric constant of the substrate (silicon, silica, glass, etc.) and air. To emulate the sample measurement in the COMSOL frequency domain electromagnetic solver, ε eff was varied by depositing carbon quantum dots (QDs) of a size similar to that of the coronavirus (∼100 nm diameter) in the capacitive gap (W), as shown in Figure 1a , in order to predict ΔF in the THz transmission function. All simulations were done with Si as the substrate, and mostly aluminum (and gold to compare the sensitivity of the nanoantennas) as the metallic metamaterial structure, where the deposition thicknesses of Al and Au are higher than their skin depths (at 0.7 THz, the skin depth of Au is around 80 nm, and for Al, it is around 98 nm 7, 26 ). Hence, the metal deposition thickness for Au was kept at 100 nm and for Al at 200 nm to avoid any penetration of THz fields into Si through the metamaterial structures and to ensure proper resonating interaction. Because the mass production of chips at CMOS technology-based fabs can be done with Al-metal only (because Au is known as a charge-carrier-lifetime killer in semiconductors), simulations were done to predict the stable For further modeling and fabrication, we chose a capacitor gap of W = 1.5 μm, with a resonance frequency at 710 GHz, which fits well with the trade-off between an increased plasmonic effect 7 (which demands a lower capacitive gap) and wider capacitive gaps that increase the probability of coronavirus/exhaled particles covering the capacitive gap. Figure 2b shows ΔF for varying surface densities of carbon QDs (100-nm-diameter particles per square micron), where we simulated ultralow QD concentrations, addressing the possibility of a low virus load in a breathalyzer-based test and taking advantage of the increased capacitive region of our cross-arrowhead structure. Now, according to the simulated results, we see an inverse relationship between ε eff and the resonant frequency, thereby leading to small red shifts (positive ΔF), which linearly increase with a small increase in the QD concentration compared to the pristine sample (reference). In addition, the fact that we have around 15000 nanoantennas on a 6 × 6 mm sized chip enables us to measure the overall effect of all antennas where the exhaled particles and viruses are spread randomly. Interestingly, according to this theory, a particle of a definitive size and shape should also generate a particular change in the resultant dielectric constant in the capacitive region, leading to a deterministic change in the resonant frequency because the nanometamaterial structures are sensitive enough to capture changes in the dielectric media due to their ultralow capacitance (femtofarad range). Hence, any specific particle can be expected to be detected by a deterministic shift of ΔF, which will be unique to only that particular substance. This notion can be well extended to different genres of viruses as well. We took advantage of the finite-element COMSOL machine to simulate such a condition. In Figure 2c , we varied the QD size while keeping the surface density constant, and we see a very gradual red shift of 10 GHz with increasing QD size. This shows the possibility of virus-specific testing with our cross-arrowhead LC resonant metamaterial. However, one can expect an ambiguity issue at this point if ε eff related to the specific density of particles of a particular size approaches that of a different particle size with different density. Viruses/particles of different kinds (both from the respiratory system tract) might show similar ΔF values, which might increase the number of false positives related to coronavirus patients. Nevertheless, at this point of the development, we aim to achieve a screening test at strategic entry points and not a diagnostic test. The decision regarding positive/negative for a specific individual will be based on a range of ΔF values, as discussed in the following, that is based on the particle density on the metamaterial chip surface. Further simulations graphically show the saturation of ΔF with an increase in the carbon QD concentration above a surface density of 2 particles per square micron (high surface density), whereas for lower surface densities, its relationship with ΔF is close to linear for different capacitive gaps (W), as shown in Figure 2d . When these simulations are extended to a Au metamaterial cross-arrowhead structure with a 100 nm deposition thickness (yet thicker than the skin depth), we observe higher sensitivity of the Au nanoantennas in Figure 2e compared to Al, which was also reported for SRR metamaterial structures. 27 However, it is worth noting that the overall simulated sensitivity of our geometrically optimized Al crossarrowhead structure is higher than that for Au-based linear SRRs as depicted by S. J. Park et. al. 19 To calculate and predict the collective effect of a plurality of such arrowhead nanostructures fabricated on a 6 × 6 mm chip with a 725-μm-thick Si substrate (the typical thickness of an 8in. wafer), in addition to the plastic capsule (wherein the chip is mounted during the spectral measurement), which serves as a radome to the microantenna array from the electromagnetic point of view, we performed further simulation with the CST Studio Suite. The enclosure of the chip is a THz semitransparent capsule with precisely calculated front-and backwall thicknesses and air gaps for optimal impedance matching of THz radiation with the nanoantenna array chip. Because of the thickness of the Si substrate, these simulations accurately predict the resonance, as well as Fabry−Perot oscillations originating by the Si substrate in transmission mode, such that there is zero loss of the transmitted signal when the capsule (radome) encloses the metamaterial chip, compared to the naked chip. This is achieved by designing suitable capsule architecture, thicknesses, and material (for impedance matching in the CST simulation, see Supporting Information section 2). Figure 3a gives the optimized dimensions of a 3D simulation, which was taken further for fabrication. Figure 3b shows the THz transmission spectrum of an Al nanometamaterial four-arrowhead chip in the resonance region enclosed in a capsule compared to an experimental spectrum in decibels. A good agreement is noticed at the resonance frequency of 794 GHz and in the amplitude between the experimental results and simulations, where both results are calculated in decibels. The resonance frequency obtained from the CST simulations is not exactly the same as that obtained from the COMSOL simulations because the latter deals with only a single element of the metamaterial, while the CST simulations account for the overall antenna array effect, taking into account the 725 μm Si substrate and the capsule that serves as a radome; this upgrades the CST simulations to a system-level simulation tool in our work. Apart from the Fabry−Perot oscillations, the resonance region can be clearly seen with a considerably lower decibel dip, indicating blocking of the transmitted THz waves by the resonant metamaterial. Because most biological entities are filled with cytoplasmic materials, in order to emulate coronaviruses and their related nanoparticles exhaled at a biological level, we used 100-nm-diameter-sized particles with relative permittivity identical with that of a biological cell as the dielectric. 28, 29 In Figure 3c , we studied the effect of ΔF with varying concentrations of such nanoentities, where we maintained a very low surface density in order to mimic a breathalyzer-like test. The precise numbers of ΔF for Al metamaterial chips are given in Figure 3d and range from 3 to 10.5 GHz, varying linearly with the nanoparticle surface density. The results are further compared with Au metamaterial chips with 100 nm deposition thickness, where ΔF varies linearly from 4.5 to 21.5 GHz as well. This quantifies the effect of the higher sensitivity of Au metamaterial structures at the system level, indicating that Au-based four- To verify our hypothesis, we conducted low-scale verification clinical trials of our breathalyzer-based coronavirus carrier screening kit at the Soroka Medical Center in the year 2021. In the initial clinical trials, each patient simultaneously gave samples for the swab-PCR test along with our breathalyzer test. The patient was asked to deeply exhale three times into the mouthpiece (for the air-flow design of our breathalyzer kit, see Supporting Information section 4). The sealed capsule samples from the patients were externally sterilized by dipping into alcohol for a few minutes. After the alcohol dried, the samples were scanned with a frequency domain THz spectrometer (see Supporting Information section 5) within 4 h from the sample collection. All measurements undertaken by us were completely "blind", and we correlated our predictions with the RT-PCR tests that were delivered to us 1 day after the breath trial; our results were provided within a few minutes to the trial principal investigator from the hospital. In order to effectively differentiate between "healthy" and "SARS-CoV-2infected" patients in terms of ΔF, we included only "completely healthy" and "sick with SARS-CoV-2" individuals in our low-scale verification clinical trials. Because we intend to design a screening test and not a diagnostic test, we have not included samples of other respiratory diseases, in order to obtain a strict threshold line between "sick" and "healthy". Figure 4 depicts the raw THz spectra (in photocurrent) of patients infected with coronavirus, compared with healthy ones. The black spectrum is a THz scan done on pristine samples, which acts as a reference, and the red line indicates the measurement samples (after the patient exhaled). Both the reference and measurement spectra are Gaussian-fitted at the resonance region, and the minima of individual fittings are extracted to compute ΔF. According to our simulations, we anticipated a transmittance dip around 800 GHz, signifying the resonance frequency for our four-arrowhead metamaterial structure. It is interesting to note that, for a coronaviruspositive patient with a CT value of 21 (high viral load), we observed ΔF = 9.2 GHz (Figure 4a ), whereas for a patient with a CT value of 30 (moderate viral load), we observed ΔF = 4.1 GHz (Figure 4b) . Hence, there is a direct correlation between the viral load (CT value of the PCR) and ΔF; the CT value indicates the viral infection level in the patients, which is strongly linked to the viral load. On the contrary, for healthy individuals, ΔF ∼ 0−1.5 GHz, where the reference and measurement spectra almost completely overlap, as observed in Figure 4c . This experiment supports our initial hypothesis that ΔF due to the coronavirus-related particles exhaled by an infected patient is considerably different from that of a healthy individual. By statistically conducting this trial for 40 individuals in hospitals, we reached the conclusion that, for screening purposes using an Al-based metamaterial, any ΔF in the range of 1.5−10 GHz will signify coronavirus-infected individuals with varying viral loads. ΔF values below 1.5 GHz can certify a coronavirus-negative individual. Small ΔF values up to the range of 1.5 GHz can be attributed to numerous other micro-to nanoparticles present in the breath. Interestingly, if we study the relationship between the viral load and ΔF for infected patients, we understand that roughly a higher viral load will lead to higher ε eff , leading to a decreased resonance frequency and an enlarged ΔF. The CT values are strongly linked to the viral load, which can be subsequently divided into zones of high, moderate, and low viral loads, depending on the marker used for the fluorescence of the RT-PCR process. 30, 31 The results are observed in Figure 4d . Here, we have assumed that the infected individual exhales properly ACS Applied Nano Materials www.acsanm.org Article into the breathalyzer, such that the higher viral load is properly correlated with the CT values. The three false-negative samples were taken from severely ill patients in critical condition who were unable to breathe properly into the breathalyzer (they all had very high CT values), and hence a sufficient ΔF value was not observed for these patients. However, because this test is intended to be used as a screening test, most of the individuals are assumed to be asymptomatic and would be able to give a proper breath test. Moreover, all of the samples with CT values higher than 29 were taken from walk-in patients, who took our breathalyzer-based test along with the RT-qPCR, which proves that our screening method works for lower viral loads. Among the remaining 24 healthy individuals, we had two false positives, as shown in Figure 4e . On the other hand, we also had one subject with a ΔF of 2.3 GHz, and with a CT value of 37, who tested negative at the time that we performed the breathalyzer test but was retested, upon showing symptoms, and found to be positive the next day. We should note at this point that the linear dependence of ΔF on the CT may serve in the future to be a tool for identifying "just infected" or "in recovery" phases of the disease, but this will require more intensive clinical trials with hundreds to thousands of tested subjects. Hence, our hypothesis seems to be confirmed and may pave the way for this technology to be used for the screening of symptomatic and asymptomatic patients with significant accuracy. An interesting correlation is also drawn between the ΔF data obtained in the clinical trials and the simulation results (see Supporting Information section 6). Upon evaluating the time taken to perform the entire breathalyzer test from the moment an individual exhales onto the chip, we could consistently deliver results within 50−55 s, which includes scanning of the sample and ΔF extraction with an automated algorithm inlayed into the software (see Supporting Information section 7). We have also included a detailed list of all patients recruited, mentioning their RT-qPCR test results along with the CT values and our predictions based on ΔF (see Supporting Information section 8). Figure 5 shows a scanning electron microscopy (SEM) image of an Au metamaterial chip collected from a coronavirus-infected patient during the initial proof-of-concept tests (in the year 2020), with coronaviruses and carriers located in the capacitive gap region, which gives a ΔF of 19.8 GHz (this ΔF is typical for a Au-based metamaterial for COVID-positive subjects). A fixation of the viruses with formaldehyde was done before scanning with the electron microscope. The coronavirus measured about 100 nm in diameter, which is well correlated with the literature. 31 We can also see various biological particles exhaled that are spread all over the metamaterial chip surface, but these are too microscopic to be resolved even at 2 μm resolution. For Au metamaterial chips, we had found that the maximum ΔF value that we can reach is up to 20 GHz for a similar viral load, which also strongly correlates with the simulation results shown in Figure 3d . Figure 5 helps to visualize our entire breathalyzer-based coronavirus carrier detection principle and mechanism with our arrowhead LC resonant metamaterial nanostructure chips. A total of 40 randomly selected individuals were recruited into this study and tested simultaneously using the SARS-CoV-2 rapid breathalyzer test and the gold standard RT-qPCR method. A total of 16 (40%) tested positive by RT-qPCR, while 24 tested negative. Considering the PCR results, we found 13 tests to be true positive and 2 to be false positive, whereas 22 tests were true negative and 3 tests were false negative ( Table 1) . The test reached a sensitivity of 81% and a specificity of 92%, a positive predictive value (PPV) of 87% and a negative predictive value (NPV) of 88.0%. In our proposed coronavirus screening test with crossarrowhead LC resonant metamaterial chips, we detected a combination of viruses and related biological particles (virus debris, cytokines, cell debris, and related proteins and fat molecules) that produce an effective change in the dielectric constant of the resonant metamaterial in the capacitive-gap regions. This phenomenon red shifts the resonance frequency (ΔF), which becomes the determining factor used to effectively screen infected patients and distinguish them from healthy individuals. The simulations correlate well with the clinical trial results, which give a distinct band of ΔF = 1.5−10 GHz for infected individuals, with a linear relationship of increasing ΔF with an increase in the viral load (and thus the amount of particles exhaled). The attractiveness of this noninvasive, rapid, and cost-effective breathalyzer test lies in its ease of handling, which does not require a complex setup procedure. The entire testing and analysis procedure is performed within 50−55 s, with 88% agreement with the RT-PCR analysis, based on the low-scale verification clinical trials conducted at the Soroka Medical Center. This method of prompt coronavirus screening will help to effectively distinguish healthy from coronavirus-infected individuals, both symptomatic and asymptomatic and with low viral loads. Because we aim to conduct this screening test at strategic entry points like airports, shopping malls, ports, etc., any individual taking the breathalyzer test may have other related respiratory and chronic diseases. This might identify as positive any individual having influenza, etc., and not just COVID-19. If the individual is screened as positive, he/she will ACS Applied Nano Materials www.acsanm.org Article have to undergo a biological-based test such as antigen or PCR diagnostic tests. Nevertheless, the infected person will be screened at entry points with PPV = 87% and NPV = 88%, using our prompt breathalyzer test. ■ ASSOCIATED CONTENT The Supporting Information is available free of charge at https://pubs.acs.org/doi/10.1021/acsanm.2c00954. Dependence of the metal thickness in the material on ΔF (S1), minimizing Fabry−Perot oscillations using the capsule (radome) (S2), fabrication methods and a working prototype (S3), air-flow design of our breathalyzer (S4), information about the THz spectrometer used (S5), correlation of the ΔF data of the clinical trials with the simulation results (S6), automation procedure for ΔF extraction (S7), and low-scale verification of the clinical trial dataset (S8) (PDF) Global Percentage of Asymptomatic SARS-CoV-2 Infections Among the Tested Population and Individuals With Confirmed COVID-19 Diagnosis: A Systematic Review and Meta-Analysis Conventional and Nanotechnology-Based Sensing Methods for SARS Coronavirus (2019-NCoV) COVID-19 Impact on Diagnostic Innovations: Emerging Trends and Implications The Use of Breath Analysis for Diagnosing COVID-19: Opportunities, Challenges, and Considerations for Future Pandemic Responses Ultrafast Multi-Terahertz Nano-Spectroscopy with Sub-Cycle Temporal Resolution Resonant Transmission of Terahertz Waves through Metallic Slot Antennas on Various Dielectric Substrates Graphene Plasmonics for Tunable Terahertz Metamaterials Reconfigurable Nanomechanical Photonic Metamaterials An Electromechanically Reconfigurable Plasmonic Metamaterial Operating in the Near-Infrared Nanoparticle-Enabled Terahertz Imaging for Cancer Diagnosis Highly Sensitive Detection of Staphylococcus Aureus by a THz Metamaterial Biosensor Based on Gold Nanoparticles and Rolling Circle Amplification Design and Study of a Coronavirus-Shaped Metamaterial Sensor Stimulated by Electromagnetic Waves for Terahertz Biosensing Metamaterial Absorber for Virus Detection Based on Spoof Surface Plasmon Polaritons Sensitive Detection of Yeast Using Terahertz Slot Antennas Sensing Viruses Using Terahertz Nano-Gap Metamaterials Analysis of Viral Load in Different Specimen Types and Serum Antibody Levels of COVID-19 Patients Colorimetric Test for Fast Detection of SARS-COV-2 in Nasal and Throat Swabs Prompt Viruses Infection Detection Using THz Spectroscopy in a Breathalyzer-Like Configuration 1084. (25) Yudistira, H. T.; Asril, M. The Study of the Altering Substrate Refractive Index on Single Symmetry Split-Ring Resonator Unit Cell Metamaterial Reduced Conductivity in the Terahertz Skin-Depth Layer of Metals Silver and Aluminum Nanoantennas to Enhance Spontaneous Emission Cells Electrical Characterization: Dielectric Properties, Mixture, and Modeling Theories Determination of Dielectric Properties of Cells Using Ac Electrokinetic-Based Microfluidic Platform: A Review of Recent Advances Real-Time RT-PCR Normalisation; Strategies and Considerations Dynamic Change Process of Target Genes by RT-PCR Testing of SARS-Cov-2 during the Course of a Coronavirus Disease The authors declare no competing financial interest. The authors acknowledge the contribution of Ben-Gurion University of the Negev in supporting this research. We also acknowledge support of the Israeli Ministry of Defense (Mafa'at) during the initial phase of this research. The help of Dr. Ifat Abadi-Korek to write the clinical results section in the proper medical terms is highly appreciated.