key: cord-0073561-efo9wpoy authors: Casalino, Gabriella; Castellano, Giovanna; Zaza, Gianluca title: Evaluating the robustness of a contact-less mHealth solution for personal and remote monitoring of blood oxygen saturation date: 2022-01-14 journal: J Ambient Intell Humaniz Comput DOI: 10.1007/s12652-021-03635-6 sha: a94831d060e4e17c0ca746dc66087c34635e1fa7 doc_id: 73561 cord_uid: efo9wpoy MHealth technologies play a fundamental role in epidemiological situations such as the ongoing outbreak of COVID-19 because they allow people to self-monitor their health status (e.g. vital parameters) at any time and place, without necessarily having to physically go to a medical clinic. Among vital parameters, special care should be given to monitor blood oxygen saturation (SpO(2)), whose abnormal values are a warning sign for potential COVID-19 infection. SpO(2) is commonly measured through the pulse oximeter that requires skin contact and hence could be a potential way of spreading contagious infections. To overcome this problem, we have recently developed a contact-less mHealth solution that can measure blood oxygen saturation without any contact device but simply processing short facial videos acquired by any common mobile device equipped with a camera. Facial video frames are processed in real-time to extract the remote photoplethysmographic signal useful to estimate the SpO(2) value. Such a solution promises to be an easy-to-use tool for both personal and remote monitoring of SpO(2). However, the use of mobile devices in daily situations holds some challenges in comparison to the controlled laboratory scenarios. One main issue is the frequent change of perspective viewpoint due to head movements, which makes it more difficult to identify the face and measure SpO(2). The focus of this work is to assess the robustness of our mHealth solution to head movements. To this aim, we carry out a pilot study on the benchmark PURE dataset that takes into account different head movements during the measurement. Experimental results show that the SpO(2) values obtained by our solution are not only reliable, since they are comparable with those obtained with a pulse oximeter, but are also insensitive to head motion, thus allowing a natural interaction with the mobile acquisition device. The current COVID-19 (Coronavirus Disease 2019) pandemic has changed our daily lives, by imposing some strict behavioural uses, such as wearing mask, distancing, and in most severe conditions lockdown. At the time of writing this paper, many countries worldwide are still trying to limit the spread of the virus by requiring people to remain in self-isolation at home and limit human contacts as much as possible. On the other side, hospital facilities have been taken by storm by the huge inflow of Covid patients, also causing slowdowns in treatment of other diseases. Despite the arrival of the first vaccines against Corona virus, these precautionary measures will have to be observed for a long time to come. In this context, mHealth, which is a new frontier of telehealth encompassing the use of mobile and wireless technologies to support digital health services, assumes crucial importance (Lupton 2013; Swan 2012; Ardito et al. 2020; Tripathi et al. 2020) . Thanks to mHealth solutions, that allow people to self-monitor their vital parameters in any place and at any time through the use of mobile devices, patients can have their vital signs continuously monitored without having to physically go to a medical clinic, thus reducing the risk of virus transmission (Kim and Choi 2020) . Moreover, the monitoring of vital signs undertaken by mobile devices allow immediate medical intervention when there is a deterioration in the health status, and reduces unnecessary hospital admissions, time and costs spent on transport, and healthrelated costs. Actually, the potential for using mHealth and in general telehealth solutions during emergencies has been emphasized in Lurie and Carr (2018) and even WHO (World Health Organization) has given specific highlight of telehealth as a powerful tool to curb the spread of the Corona virus and for early detection of the disease at the very onset. In particular, monitoring blood oxygen saturation (SpO 2 ) is of fundamental importance for early detection of COVID-19 since abnormal values of this vital parameter are a warning sign for potential infection even before patients start having short breath. Indeed, it has been observed 1 . that the SpO 2 values may fall below 90% much more before the first symptoms of Covid pneumonia appear. SpO 2 monitoring can be easily done at home through the pulse oximeter that however requires skin contact and hence could be a potential way of spreading contagious infections. To overcome this problem, in Casalino et al. (2020) we have recently developed a mHealth solution for contact-less estimation of blood oxygen saturation. Its basic idea is based on extracting the photoplethysmographic signal from facial video frames, hence it does not require any skin contact for measurement of the vital sign. The first experimental evidences reported in Casalino et al. (2020) have shown that our mHealth solution works well in controlled environments and provides several advantages. It is cheap, since it just requires a mobile device equipped with a camera (smartphone, laptop, tablet, ...), and does not need specific medical devices. It is contact-less, thus avoiding potential infections. It is easyto-use, since the patient experiences the self-monitoring of vital parameters through a simple and natural action that can be part of his daily routine. Moreover, it can be easily integrated within a patient-centered telehealth platform for both personal and remote health monitoring. Finally, it is fast, since it provides reliable SpO 2 measurements in a couple of seconds. Although the proposed solution has been calibrated to perform the measurement of blood oxygen saturation aligned with a common pulse oximeter, there is still a lack of validation studies concerning its applications in real world scenarios. Indeed, measuring the SpO 2 using mobile devices in real-world daily situations holds some challenges in comparison to the controlled laboratory scenarios where the first experiments of our solution were conducted. One main issue is the frequent change of perspective viewpoint due to head movements, which leads to challenges in detecting the human's face and computing an estimate of SpO 2 . Hence, to better assess the validity of our solution for in-thewild scenarios, it is fundamental to assess its robustness to head movements. This is the goal of the present work. To understand to what extent our mHealth solution is robust to head movements, in this paper we carry out a pilot study on the PURE benchmark dataset that takes into account different head movements during the SpO 2 measurement. Experimental results have shown that measurements obtained by our contact-less solution are reliable since they are comparable with those obtained using a commercial pulse oximeter, and are insensitive to head movements. The paper is organized as follows. In Sect. 2 we overview some related works. Section 3 describes the problem of SpO 2 measurement via photopletismography. Section 4 summarises the proposed solution. In Sect. 5 we discuss experimental results on a benchmark dataset to evaluate how head motion affects the SpO 2 measurements. Conclusions and future works are outlined in Sect. 6. Several mHealth solutions have been recently proposed that allow patients to self-monitor their health status and stay remotely in touch with physicians, thus avoiding them to go physically to clinical centers. Most of these solutions rely on the well established technology known as remote photopletismography (rPPG) that has been largely used to develop contact-less solutions for monitoring vital parameters including SpO 2 . In the pioneering work by Verkruysse et al. (2008) , a digital camera in ambient light is used to estimate heart rate and breath rate values based on rPPG. Poh et al. (2010) proposed the use of the blind source separation of colour channels into independent components to process facial videos for heart rate estimation. Hassan et al. (2017) and Rouast et al. (2018) present an overview of rPPG techniques based on images of faces for heart rate estimation. Since vital parameters must be correctly estimated by means of trustable devices, several works studied the limits of these approaches, due to their sensitivity to external factors, such as light in Yang et al. (2021) and Song et al. (2021) , variations in environmental conditions and subject demographics in Dasari et al. (2021) , presence of face masks in Speth et al. (2021) , and motion in Liu et al. (2021) . However, none of these works studied the influence of those factors on the measurement of blood oxygen saturation. Only Allado et al. (2021) have recently designed a clinical trial to evaluate the effectiveness of rPPG-based measurements of blood oxygen saturation (as well as heart rate and respiratory rate) compared to those obtained with contact devices. First attempts to use PPG-based measurements employed specialized and expensive devices that are difficult to retrieve and use, thus largely excluding a portion of the potential target users. More recently, low cost devices or common devices such as smartphones have been employed for largescale adoption in daily life. Many free or low-cost software applications specifically designed for self-measuring of vital parameters through smartphones have been recently developed. For example Tayfur and Afacan (2019) Besides smartphones, other cheap and ergonomic mobile devices have been proposed for personal and remote measurement of vital parameters. An example is the smart bracelet developed within the Telemachus (Telemedicine Monitoring and Collaborative Hub-and-Spoke System) project 2 . that allows to remotely monitor Covid patients in all phases of the disease by detecting the vital parameters of the wearer and communicating via geolocalization or via Bluetooth. Another device for data collection and remote monitoring of COVID-19 patients in home isolation is described in the cosinuss • project 3 . It consists of a device in the wearable in-ear using photoplethysmography to monitor vital parameters, a data gateway and a cloud database server (Baldinger et al. 2021 ). In Sabbadini et al. (2021) heart rate and breathing rate are measured through the use of a face mask using the photoplethysmography technique to detect the presence of environmental and physical stress factors in workers. Despite the current proliferation of personal mobile/ wearable devices for vital parameter self-monitoring, the state-of-art literature shows that the most suitable sensor for contact-less measurement of SpO 2 is the video camera that satisfies both easy-of-use and contact-less requirements. The present work represents an advancement of the state of the art by introducing a general camera-based mhealth solution for the reliable measurement of blood oxygen saturation. Given the absence of studies on the influence of external factors on rPPG measurement of blood oxygen saturation, in this study we investigate the influence of head movements to better assess the robustness of our rPPG-based solution. The problem we face in this work is to estimate the value of blood oxygen saturation through a contact-less solution exploiting remote photopletismography. According to Challoner (1979) , the photoplethysmography (PPG) signal represents the variation of lights due to cardiovascular activities in some regions of the face tissue. These variations can be used to estimate the hearth rate and the blood oxygen saturation (Allen 2007) . Specifically, according to Kong et al. (2013) , an estimation of SpO 2 can be obtained by considering the cardiovascular pulse wave signal at two different wavelengths, namely the red and blue bands, being the blue band representative of the infrared wavelength used in the traditional pulse oximeter. The underlying principle is that the impulse of cardiovascular wave that flows through the body periodically, causes stretch in the vessel walls, with consequent fluctuations in blood volume. These fluctuations modulate the absorbency of light passing through a given volume of tissue, so it is possible to evaluate the variation of light during a normal cardiac cycle. There are two methods to implement PPG measurement of SpO 2 . The first method is contact PPG, in which the user places his/her finger on the sensor and the flash light provides the light source required for the blood cells to become visible (Kim et al. 1986 ). The second method is contact-less or remote PPG, in which the camera is held in front of the user's face and the measurement is undertaken without the need for skin contact. For the motivations described in section I, we chose to implement a contact-less PPG estimation of SpO 2 . The following sections outlines the methodological pipeline underlying out contact-less solution. Our approach for contact-less measurement of blood oxygen saturation is based on processing a short video frame sequence that capture the face of the user in order to extract the photopletismographic signal and derive an estimate of SpO 2 . The detailed description of the methods underlying our solution are described in Casalino et al. (2020) . Here we briefly describe the main phases of the methodological pipeline (Fig. 1) . In the first phase, a short video capturing the user's face is acquired through the video camera and a sequence of video frames 2 seconds long is processed. Resolution of the video frames depend on the characteristics of the video camera integrated in the mobile device. Each frame is processed for face detection and three regions of interest (ROIs) are identified, specifically the two cheeks and the forehead, which represent the areas where the plethysmographic signal is most evident (Challoner 1979) . We used three different regions in order to make the acquisition more robust to occlusion, such as glasses, beard and so on, as it has been proven in our previous work by Casalino et al. (2020) . To cope with head movements, a fourth ROI is located in the middle of the face and used for face tracking. This further ROI is connected with the other three ROIs (cheeks and forehead), so that together they can indicate the movement of the face. This enables face tracking that makes the system less sensitive to movements and thus avoids the user having to stand still in front of the camera. The second phase is devoted to extract the photopletismographic signal from the identified ROIs. Signal processing techniques are applied to the three different ROIs for noise reduction, and the most informative ROI is selected for SpO 2 measurement, as suggested in Solomon (1991) . Particularly, for each of the three ROI, the RGB signals are extracted. Then for each of the three channels (red, blue and green) and for each frame in the video, the average value over all the pixels is computed. Thus a three dimensional signal is derived for each frame. High-frequency noise in signals are filtered (Speake and Mersereau (1981) ) and the the chrominance method (De Haan and Jeanne (2013)) is used to limit the influence of motion and illumination variations on the signal. Finally linear interpolation is applied to uniform the signal. After signal cleaning, red and blue components of the signal are used to compute SpO 2 value according to the following formula that was suggested by Kong et al. (2013) : where the AC RED and AC BLUE , DC RED and DC BLUE , are respectively the standard deviations and the mean of the red and blue signals computed from the sequence frames. A and B are calibration parameters. We used A = 125 and B = 26 as in Kong et al. (2013) . Using the above described process, an estimation of blood oxygen saturation is obtained in real-time. Indeed, the whole process requires only a couple of seconds to obtain the SpO 2 value. The methodology was implemented as a web-based application, hence after acquiring the short video, all the processing steps run on a server, and web-services are used for communication. Being web-based, our solution can be employed on different mobile devices (smartphone, tablet, laptop, mobile robots,....), as long as they are equipped with a camera and are able to connect to internet. The described pipeline was implemented as a prototype in Casalino et al. (2020) . Preliminary results under controlled conditions of in-vitro experiments have shown that the methodology is reliable and comparable to a classical pulse (1) Fig. 1 Methodological pipeline for contact-less SpO 2 measurement oximeter under controlled conditions, i.e. when the users were required to stay still in front of the camera. However, to definitively assert the robustness of the method, we need to test it in more real scenarios. Actually, in real situations of use of our method on mobile devices, it is essential to determine if the measurement is susceptible to changes in case of head movements. The aim of this work is indeed to evaluate how robust the developed method is to the movements that the user may naturally make during the interaction with the mobile device. To this aim, we have conducted a pilot study to test the method on a dataset including facial videos of users characterized by different head movements. The following section shows the results of this study. The goal of this work is to examine the influence of artifacts introduced by head motion on the SpO 2 measurements made by our contact-less solution originally proposed in Casalino et al. (2020) . To this aim we used the PURE dataset (Stricker et al. 2014) . It is composed of videos of 10 subjects (8 males, 2 females) standing in front of a camera at a distance of about 1.1 meters and performing 6 different head motions, for a total of 60 videos. The videos were recorded with an RGB digital camera. Each video has a length of about 1 minute and was captured at a frame rate of 30 Hz with a cropped of resolution of 640 × 480 pixels. The dataset provides Ground Truth values of heart rate and SpO 2 that were measured with a pulse oximeter at the same time as the videos were recorded. The adopted pulse oximeter is a CMS50E with a sampling rate of 60 Hz. Natural lighting condition were used with frontal daylight on the subjects, and changing illumination conditions due to clouds. In all controlled setups the user was sitting in front of the camera. The following six different configurations are considered: -Steady (S). This setup is used as baseline for measurements. The subject looked at the camera while avoiding head motion. Figure 2 shows frames of a single subject in the six different poses. 4 We processed each video using our method by extracting a number of measurements that is proportional to the length of the video. We obtained an average number of 20 measures for each video. Due to the higher sampling rate of the pulse oximeter, there are more ground truth values than the values obtained by our method. To enable comparison between our measurements and the pulse oximeter measurements, for each video we calculated the mean value of the SpO 2 values computed by our system and the mean value of the corresponding SpO 2 ground truth values. We quantitatively evaluated the accordance between the Figure 5 shows Bland-Altman graphs that visualise the differences between the two measurements and define the Level of Agreement (LoA). For reliable measurements the 95% of the differences should be within the LoA, that are the dotted lines in the graph (lower and upper limits LoA). The central line represents the mean bias. Values lower than zero indicates that the proposed approach under-estimate the ground truth. On the contrary, values higher than zero suggest that the proposed approach returns values that are higher than the real ones. For a good agreement mean bias should be zero or close to it. We highlight that due to measurements errors related to the pulse oximeter, SpO 2 is considered valid if the mean bias is within ±2% (Chan et al. 2019) . In detail, Fig. 5 shows the B&A graph with all mean values of the measures without considering the movement of the subjects. We can observe that there is agreement between the two measurements since all the points except two are within the LoA. In addition, we computed the B&A graph based on subject motion (Fig. 3) . It can be seen that the movements slightly affect the agreement but there are no significant differences among the graphs of different movements. All the points are between the LoA, except for one measurement in settings S, ST and MR, that are still very close to the LoA bands. Figure 6 shows the average SpO 2 values obtained from ground truth and our system for each subject without accounting for motion. Overall, it can be seen that the values in the plot are in agreement with each other and there are no significant differences between the ground truth values and the values in our system. This is also supported by the total MAE value (1.332) in Table 1 . In addition, Fig. 4 shows the average values of ground truth and our system, for each subject, in relation to motion. Again, there are no significant differences between the values. Table 1 shows that comparable results of MAE and RMSE are obtained for the six movements. Indeed if we consider the steady case as a baseline, low differences are observed in the other cases. Even with abrupt changes, as fast translation and medium rotation, we obtain low error values. Pearson analysis returns comparable results for the setups S, T, FT and MR. Surprisingly the worst results are obtained with ST and SR setups. We observed that these results depend on significant errors in rPPG acquisitions that were due to bad light conditions. In particular, three out ten subject videos had low lights in the Slow Translations setting. On the contrary, in the Fast Translation setting the scene was well lite, and consequently lower errors were observed. Thus we conclude that low illumination may affect the rPPG-based measurement. However, it is worth noting that the proposed methodology acquires the rPPG signal from a sequence of frames that together compose the whole movement. Thus, during real-time acquisition in complex setups such as ST or FT, the user will still be stationary in some frames, as in Steady setup. In fact, in Fig. 2 , frames corresponding to ST and FT setups show the user's head in front of the camera. Moreover, even though the setups were controlled, the users might perform slightly different movements. Leveraging these results, we can conclude that SpO 2 measurements obtained by the proposed contact-less solution approximate quite well measurements given by the pulse oximeter, and the movements do not affect significantly the estimate of SpO 2 . This is also obtained thanks to the tracking module that allows the system to cope with head movements such as slow rotations. Indeed in case of head movements, the ROIs are still tracked by the system (as shown in Fig. 2) , thus the rPPG signal is still correctly extracted from the ROIs. In this paper a mHealth solution for personal and remote measurement of blood oxygen saturation has been presented. The key feature of our contact-less solution is the use of a video camera to acquire video frames of the face from which the plethysmographic signal is extracted and finally an estimate of the SpO 2 value is derived. The proposed solution allows a continuous personal control of the blood oxygen saturation in any place, avoiding the use of dedicated measuring devices (e.g. saturometers, pulse oximeters), but simply using a mobile device equipped with a camera. Since it is designed to be as portable as possible, our solution can run on different mobile devices of daily use, thanks to a web-based system, thus making the measurement easy, noninvasive and pervasive. The study in this work was mainly aimed to evaluate the robustness of the proposed approach to different motion setups. Experimental results showed no significant differences among the SpO 2 measurements in presence of different slight head movements. We believe the robustness of the method is achieved also thanks to the tracking mechanism that has been implemented as a main step of the methodological pipeline. On the overall, we can conclude that the proposed mHealth solution is robust enough to be used in real scenarios where the user should be free to move and not be forced to be still in front of the camera. This enhances the user experience, as well as the acceptability, leading the user needs to trust the technology and feel safe while using it. However, experimental results have shown a certain influence of the illumination conditions on the acquisition. Future works will be addressed to better study light conditions that could affect the real-time measurement in indoor environments. Moreover, the dataset considered in this work mainly encompasses videos of healthy people. We plan to test the robustness of the method on data coming from sick patients, who have critical values of SpO 2 . Unfortunately such cases are difficult to collect at the moment, due to the pandemic situation, but we plan to collect more data in the future. With the growing use of mobile technology in the health field, we foresee that patients will increasingly use mobile devices to monitor daily their vital parameters or for personal triage assessment before going to the hospital. For this reason, we are currently extending our solution so as to 6 Comparing the SpO 2 values measured by our system with the ground truth enable contact-less measurement of other vital parameters besides SpO 2 , such as temperature, heart rate, and breathing rate. Also, we are working to integrate it into a telehealth platform so as the acquired data is sent in real time to physicians who can review it from mobile through an app and from desktop through a web dashboard. Innovative measurement of routine physiological variables (heart rate, respiratory rate and oxygen saturation) using a remote photoplethysmography imaging system: a prospective comparative trial protocol Photoplethysmography and its application in clinical physiological measurement Management at the edge of situation awareness during patient telemonitoring. International conference of the italian association for artificial intelligence TELECOVID: remote vital signs monitoring of COVID-19 risk patients in home isolation with an in-ear wearable A mhealth solution for contact-less self-monitoring of blood oxygen saturation Intelligent analysis of data streams about phone calls for bipolar disorder monitoring Photoelectric plethysmography for estimating cutaneous blood flow. Non-invasive physiological measurements A smartphone oximeter with a fingertip probe for use during exercise training: usability, validity and reliability in individuals with chronic lung disease and healthy controls Evaluation of biases in remote photoplethysmography methods Robust pulse rate from chrominancebased rppg End-to-end deep learning framework for coronavirus (covid-19) detection and monitoring Mobile health in remote patient monitoring for chronic diseases: principles, trends, and challenges Heart rate estimation using facial video: a review Considerations on the implementation of the telemedicine system encountered with stakeholders' resistance in Covid-19 pandemic Pulse oximetry and circulatory kinetics associated with pulse volume amplitude measured by photoelectric plethysmography Covid-19 mobile apps: a systematic review of the literature Non-contact detection of oxygen saturation based on visible light imaging device using ambient light Motion-robust multimodal heart rate estimation using bcg fused remote-ppg with deep facial roi tracker and pose constrained Kalman filter The digitally engaged patient: Self-monitoring and self-care in the digital health era The role of telehealth in the medical response to disasters Application of smartphone technologies in disease monitoring: a systematic review Mobile application for monitoring patients under home oxygen therapy: a protocol for a randomized controlled trial Non-contact, automated cardiac pulse measurements using video imaging and blind source separation Remote heart rate measurement using low-cost rgb face video: a technical literature review A smart face mask based on photoplethysmography for cardiorespiratory monitoring in occupational settings A cardiac early warning system with multi channel SCG and ECG monitoring for mobile health Psd computations using Welch's method Remote photoplethysmography with an eemd-mcca method robust against spatially uneven illuminations A note on the use of windows for twodimensional fir filter design Remote pulse estimation in the presence of face masks 2021) bwhealthapp: A software system to support personalized medicine by individual monitoring of vital parameters of outpatients Non-contact video-based pulse rate measurement on a mobile service robot Health 2050: the realization of personalized medicine through crowdsourcing, the quantified self, and the participatory biocitizen Reliability of smartphone measurements of vital parameters: a prospective study using a reference method Accuracy of smartphone-based pulse oximetry compared with hospital-grade pulse oximetry in healthy children Sms: a secure healthcare model for smart cities Remote plethysmographic imaging using ambient light Non-contact vital-sign monitoring of patients undergoing haemodialysis treatment Assessment of deep learning-based heart rate estimation using remote photoplethysmography under different illuminations Acknowledgements All the authors are members of the GNCS-INDAM Gruppo Nazionale per il Calcolo Scientifico of Istituto Nazionale di Alta Matematica, Francesco Severi, P.le Aldo Moro, Roma, Italy. This work was partially supported by INdAM GNCS within the research project "Computational Intelligence methods for Digital Health". The authors G. Casalino and G. Castellano are with the CITEL -Centro Interdipartimentale di Telemedicina, Università degli studi di Bari Aldo Moro.