key: cord-0067586-qzurzhp8 authors: Stoecker, Allison; Lear, William; Johnson, Karsten; Bahm, Jared; Kruzic, Jamie J. title: Enhanced perfusion of elliptical wound closures using a novel adhesive suture retention device date: 2021-09-14 journal: Health Sci Rep DOI: 10.1002/hsr2.364 sha: 38e7db9d0d02401e1d953869d2903c352d4b7d5b doc_id: 67586 cord_uid: qzurzhp8 BACKGROUND AND AIMS: The purpose of this investigation was to test the hypothesis that a novel adhesive retention suture device (ARSD) can increase perfusion at elliptical wound closures by distributing stress away from the suture site. METHODS: Stress in the skin around a suture both with and without support from an ARSD was evaluated using a finite element model. A single‐center, randomized split‐scar comparison trial using laser speckle contrast analysis was used to quantify the perfusion at elliptical wound closures in human patients both with and without an ARSD. RESULTS: The finite element model revealed that the ARSD promoted load transfer to the skin over a larger area, thus reducing the local stress and deformation in the skin around the suture site. Results from the split‐scar study showed a mean improvement of 25% perfusion units with the ARSD, and the improvement was statistically significant (p = 0.002). CONCLUSION: The reduction in local stress and enhanced perfusion around the suture site reveals the potential benefit of using an ARSD to enable more efficient healing by avoiding complications associated with both low perfusion and skin tearing, such as dehiscence, infection, and cheese wiring. Adequate perfusion is required to supply the cellular and metabolic demands required for the wound healing phases. 4 Poor perfusion is associated with complications including wound dehiscence, infections, and delayed healing. [5] [6] [7] Measurements of perfusion have proven to be important prognostic factors in the clinical management of wound healing, with one example being the ankle-brachial index for the treatment of chronic lower extremity wounds. 8 Laser speckle contrast analysis (LASCA) is another effective method to measure perfusion by real-time camera imaging of a defined area over time. LASCA has been successfully used to monitor healing after burns and radiation therapy and is able to predict the need for excision and grafting. 9, 10 Furthermore, LASCA measurements of perfusion have proven to correlate well with laser Doppler measurements while enabling faster measurements over larger areas. 9, 10 F I G U R E 1 Solid model used for the finite element analysis with the x-axis defined as parallel to the wound, the y-axis defined along the suture loading direction, and the z-axis defined as perpendicular to the skin surface. (a) and (b) show the configuration of the simulated skin samples with and without the ARSD, respectively. (c) shows the fixed displacement boundary condition and (d) depicts a cross-sectional view of the y-z plane where displacement was constrained to 1 mm in the z-direction. Color codes for the layers of the ARSD are white = spunlace, blue = PETG, and purple = polyethylene Recently, an adhesive retention suture device (ARSD) has been reported to aid in the closure of fragile and high-tension wounds. 11 The ARSD has been shown to bolster skin strength and allow for closure of wounds even in situations where normal sutures had already torn the skin. 11 Additionally, clinical observations have shown less skin blanching of the skin with the use of the ARSD in high-tension wound closure. Here we hypothesize that the efficacy of the device is related to how stress is redistributed at the wound edge to create enhanced perfusion relative to a suture without the device. To test this hypothesis, we have developed a finite element analysis (FEA) model to assess the stress redistribution caused by the ARSD when it is added to a suture. We then examined the effect of the ARSD on perfusion in human patients when used with a nylon-interrupted suturing pattern. A quasi-static FEA model was created to compare the stress distribution and relative deformation around suture penetration sites in simulated skin samples both with and without the use of an ARSD. FEA was selected for this study because it is the method of choice for calculating the stresses involved in engineering designs and has had great impact in various biomedical applications, [12] [13] [14] including the design of medical devices. [15] [16] [17] [18] We refer the reader to various excellent sources for more information on FEA and its applications in medicine and medical devices. [15] [16] [17] [18] A solid model of one case that included an ARSD and another that did not ( Figure 1 ) was created using Onshape software (Onshape, Boston, Massachusetts), and the FEA was carried out using SimScale (SimScale GmbH, Munich, Germany). The finite element mesh was comprised of tetrahedral elements with refinement around the stress concentration locations (Figure 2 ) to give a total of 32,745 nodes. We refined the mesh until the stress values converged within~0.6%. The FEA model included the point loading induced by the suture; however, the deformation of the suture was not modeled in the present study. For both samples, a total tensile wound closure force of 25 N was applied for each case (Figure 1a ,b) while the top edge of each skin sample was fixed (Figure 1c ). The value of 25 N represents the upper end of typical wound closure forces based on in vivo human data. [19] [20] [21] [22] Additionally, the skin displacement out of the plane was constrained to be 1 mm in the z-direction to represent the realistic situation of the skin minimally pulling away from the site of a clinical wound closure ( Figure 1d ). The ARSD was modeled based on the same commercially available device that is used in our perfusion study (HEMIGARD ARS device, SUTUREGARD Medical Inc., Portland, Oregon). The ARSD consists of a layered polymer structure containing three materials: spunlace, polyethylene, and glycol-modified polyethylene terephthalate (PETG). The design of the ARSD includes spunlace that runs the whole length of the device, a layer of PETG covering the spunlace at the end with the two holes, and a polyethylene top layer that covers the PETG and extends over some of the spunlace (see Figure 1 for a color-coded depiction of the layers). All of the materials were considered to be homogeneous, isotropic, and linear elastic, and the material properties used for each layer and the skin are given in Table 1 . Thickness values for the ARSD layers were taken to exactly match the device. All values for skin were estimated based on literature values. 23, 24 While it is recognized the polymer layers and skin behave viscoelastically and viscoplastically under stress, 21,25,26 using simple elastic properties is considered a suitable methodology to compare the worst-case scenario that occurs upon initial loading before any stress relaxation can occur. Finally, the model assumed good adhesion between each material layer of the ARSD and also between the ARSD and the skin layer such than no separation between layers, or between the ARSD and the skin, occurred. We conducted a single-center, randomized split-scar comparison trial between August and October 2020. The split-scar model was selected to reduce uncontrollable variables and has been used previously to compare suturing techniques. 27 The Samaritan Regional Medical Centre Institutional Review Board in Corvallis, Oregon, approved this registered study (SHS IRB20-054). Seventeen patients were enrolled in the study, and all patients gave written informed consent prior to enrollment. Inclusion criteria were male and female patients over Study participant characteristics are summarized in Table 2 , and wound dimensions are summarized in Table 3 . All procedures were performed by the same surgeon. Each surgical ellipse was marked with a long axis parallel to relaxed skin tension lines. Each fusiform specimen was removed with a number 15 scalpel blade, and minimal electrosurgery was performed. Wounds were divided in half, with the midline marked with a sterile surgical marking F I G U R E 2 Schematic showing the mesh used for the finite element analysis and the mesh refinement employed around the stress concentration locations pen. Then, marks were made to indicate 1 cm from the midline on each side of the wound. On each of these lines, a marking was made 1 cm from the wound edge on each side of wound as depicted schematically in Figure 3 . The side to receive the ARSD was randomized using a random number generator. A transparent ARSD variant of the commercially available HEMIGARD was used to enable visualization of the entire wound. This ARSD differed from the commercial ARSD in that the spunlace layer However, the bonding of the ARSD to the skin allows for the load transfer to the skin to occur over a larger area, thus reducing the local stress and deformation in the skin around the suture site. We performed our subsequent in vivo perfusion tests based on the hypothesis that by having less skin under high stress (ie, less red in Figure 4 ), we would find enhanced perfusion for the ARSD case. A representative LASCA image is shown in Figure 5 . In Figure 5 The results of this study indicate a statistically significant 25% increase in perfusion at the time of wound closure with the use of an ARSD. Such results are consistent with prior studies that have shown how various suturing patterns can affect perfusion at the wound edge. 28 Furthermore, the FEA results suggest that the lower perfusion for the bare suture is correlated with the higher concentration of local stress around the suture site. Thus, our hypothesis that having less skin under high stress (red in Figure 4 ) would give enhanced perfusion for the ARSD case was not rejected. Wound dehiscence and delayed healing remain problems in cutaneous wound repair and are associated with increased costs and morbidity. 29, 30 Inadequate closure of some lower extremity wounds, including pretibial lacerations in the elderly, is associated with increased mortality. 31 Surgical procedures below the knee, such as saphenous vein graft harvesting and limb amputation, have an elevated risk of wound healing complications including dehiscence and edge necrosis. 30, 32 In foot and ankle surgery, wound dehiscence is the most common postoperative complication, affecting approximately 5% of all cases. 33 In orthopedic surgery, dehiscence occurrence over internal hardware results in further procedures, morbidity, and cost. Considering the importance of adequate perfusion in avoiding wound healing complications, 5-7 it should be expected that the 25% improvement in perfusion found in this study could reduce postoperative complications such as edge necrosis and dehiscence, especially in zones of the closure where tension is highest. The ARSD also has the advantages of bolstering skin that is fragile to prevent "cheese wiring" of the skin and elevates the suture above the skin surface to prevent suture ingrowth. 11, 34 Compared with other skin support methods such as using medical tape, 35 Steri-Strips, 36 tearing. Meanwhile, the trailing edge of spunlace is stretchy to prevent shear force blistering on frail skin. One limitation of this study is that the average age of patients in this study was slightly over 75 years. However, the data are highly relevant since the elderly are at higher risk of wound complications after lower limb procedures and worsened morbidity in the event of complications. 31 Jamie Kruzic had full access to all of the data in this study and takes complete responsibility for the integrity of the data and the accuracy of the data analysis. Jamie Kruzic affirms that this manuscript is an honest, accurate, and transparent account of the study being reported; that no important aspects of the study have been omitted; and that any discrepancies from the study as planned (and, if relevant, registered) have been explained. The data that support the findings of this study are available from the corresponding author upon reasonable request. ORCID Jamie J. Kruzic https://orcid.org/0000-0002-9695-1921 Cutaneous wound healing The healing of rat skin wounds annals of surgery General principles of wound healing Wound healing: part I. basic science Wound healing essentials: let there be oxygen Tissue oxygenation, anemia, and perfusion in relation to wound healing in surgical patients Using physiology to improve surgical wound outcomes Chronic wounds: evaluation and management Microvascular blood flow in scalds in children and its relation to duration of wound healing: a study using laser speckle contrast imaging Changes in skin microcirculation during radiation therapy for breast cancer A novel adhesive suture retention device for the closure of fragile skin under tension Assessment of fracture risk in proximal tibia with tumorous bone defects by a finite element method Biomechanical comparison of different implant inclinations and cantilever lengths in all-on-4 treatment concept by three-dimensional finite element analysis Finite element analysis of dental implants with and without microthreads under static and dynamic loading The impact of the finite element method on medical device design Application of finite element analysis in medicine Application of the finite element method in orthopedic implant design Finite Element Analysis for Biomedical Engineering Applications Measurements of closing force of surgical wounds and relation to the appearance of resultant scars The analysis of forces needed for the suturing of elliptical skin wounds In vivo stress relaxation of human scalp Forces on sutures when closing excisional wounds using the rule of halves Comparison of the mechanical properties of different skin sites for auricular and nasal reconstruction Poisson's ratio in skin Mechanical properties of the skin Viscoelasticity of Polymers: Theory and Numerical Algorithms. Springer Series in Materials Science 241 Effect of simple interrupted suture spacing on aesthetic and functional outcomes of skin closures The effect of suture pattern and tension on cutaneous blood flow as assessed by laser Doppler flowmetry in a pig model Postoperative wound dehiscence: predictors and associations Determining risk factors for surgical wound dehiscence: a literature review The management of pretibial lacerations Trends in lower limb amputation in the veterans health administration, 1989-1998 The most Common Patient Safety Problems in Elective Foot and Ankle Orthopedic Surgery Use of a novel adhesive suture retention wound closure device to prevent patient follow-up visits during the COVID-19 pandemic The Fixomull skin support method for wound closure in patients with fragile skin Suture/Steri-strip combination for the management of lacerations in thin-skinned individuals A novel method of skin closure for aging or fragile skin Enhanced perfusion of elliptical wound closures using a novel adhesive suture retention device