key: cord-0066071-y8ex6vxu authors: Chen, Xin-Fei; Hou, Xin; Xiao, Meng; Zhang, Li; Cheng, Jing-Wei; Zhou, Meng-Lan; Huang, Jing-Jing; Zhang, Jing-Jia; Xu, Ying-Chun; Hsueh, Po-Ren title: Matrix-Assisted Laser Desorption/Ionization Time of Flight Mass Spectrometry (MALDI-TOF MS) Analysis for the Identification of Pathogenic Microorganisms: A Review date: 2021-07-19 journal: Microorganisms DOI: 10.3390/microorganisms9071536 sha: 517649a217a1bf4ea1b72ad97290532f2b346004 doc_id: 66071 cord_uid: y8ex6vxu Matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) has been used in the field of clinical microbiology since 2010. Compared with the traditional technique of biochemical identification, MALDI-TOF MS has many advantages, including convenience, speed, accuracy, and low cost. The accuracy and speed of identification using MALDI-TOF MS have been increasing with the development of sample preparation, database enrichment, and algorithm optimization. MALDI-TOF MS has shown promising results in identifying cultured colonies and rapidly detecting samples. MALDI-TOF MS has critical research applications for the rapid detection of highly virulent and drug-resistant pathogens. Here we present a scientific review that evaluates the performance of MALDI-TOF MS in identifying clinical pathogenic microorganisms. MALDI-TOF MS is a promising tool in identifying clinical microorganisms, although some aspects still require improvement. Matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) has been applied in the field of clinical microbiology for a dozen years. With the development of new technology and optimization of methods, novel rapid and accurate approaches have been developed to improve identification accuracy. Clinical microbiology laboratories are constantly expanding the application of MALDI-TOF MS in new areas of infectious disease diagnostics. Accurate identification of microbes at the species level in clinical situations is imperative in most cases. Therefore, this review focuses on the efficacy of MALDI-TOF MS in identifying pathogens at the species level in a clinical setting. MALDI-TOF MS is not limited to identifying strains cultured on solid media but can also directly identify them from the blood culture, cerebrospinal fluid, and urine [1] [2] [3] [4] [5] . MALDI-TOF MS identification has been used to identify gram-negative and gram-positive bacteria, aerobes, anaerobes, mycobacteria, Nocardia, yeasts, filamentous fungi and viruses [6] [7] [8] [9] [10] [11] [12] [13] . MALDI-TOF MS has good identification outcomes for common, clinically isolated bacteria [14, 15] . In fungi, the identification of Candida spp. and filamentous fungi has made good progress [4, 11] . MALDI-TOF MS plays an important role in the identification of viruses, including influenza viruses, hepatitis viruses, and herpesviruses [16] [17] [18] [19] . New studies have shown remarkable progress in the detection of microorganisms by MALDI-TOF MS and the identification of drug-resistant strains [20] [21] [22] [23] [24] [25] . The present review aimed to summarize the efficacy of MALDI-TOF MS for pathogen identification in clinical microbiology laboratories, including sample preparation and identification performance of microorganisms. In addition, the current problems associated with MALDI-TOF MS in clinical laboratories are also discussed. During MALDI-TOF MS detection, the microbial samples, with or without a preceding extraction step, and the matrix are added to the target and dry to form crystals. The sample and matrix absorb energy and vaporize into a positively charged cloud when the plate is exposed to laser irradiation. These ionized proteins are accelerated in the flight tube under an electrostatic field. The speed at which ions fly to the ion detector depends on the mass-to-charge ratio (m/z). The value of m/z is shown on the x-axis, while the signal strength is shown on the y-axis of the mass spectrum. The height of the peak in the mass spectrum is related to the abundance of a protein. MALDI-TOF MS can analyze the positively charged proteins of 2000-30,000 Da for the identification of microorganisms. Different MALDI-TOF MS systems can detect ionized proteins in different ranges: the Shimadzu system can detect ionized proteins smaller than 30 kD, while the Bruker system can detect ionized proteins smaller than 20 kD. Therefore, the spectra (in clinical settings) only included a small part of the entire spectra, composed of the proteins of the microorganisms, and subsequently were compared with the database [26] . The majority of bacterial and yeast samples in clinical laboratories are separated on a solid agar medium, following which, an appropriate colony is selected and smeared directly on a target plate as a thin film. However, Mycobacterium spp., Nocardia spp., and filamentous fungi have complex wall structures, resulting in poor protein extraction with the matrix process. The formic acid extraction method (on-plate) can improve extraction efficiency. Unlike bacteria and fungi, which can be cultured on solid media, viruses can only be cultured in living cells. In addition, due to different detection principles, the preparation for MALDI-TOF MS-based virus identification is nucleic acid extraction. With the recent development of MALDI-TOF MS, there has been an improved study of direct sample identification [27] [28] [29] [30] . Positive blood bottles and urine samples may require further processing to retain pathogens. In addition, workflow and preparation process for different samples need to be established to improve MALDI-TOF MS identification. Identification of culture strains is a routine clinical laboratory technique. For most bacterial and yeast strains, direct deposition is an easy and rapid method [31] . This method includes two steps: first, the deposition of a single colony on the target and fixing it with a matrix. On-target extraction has a good effect on yeast and some difficult-to-extract bacteria [32] [33] [34] . The smeared samples were first treated with formic acid. After air-drying, the samples were dropped onto the matrix. If the direct deposit or on-target extraction methods cannot identify the microorganisms in strains with a thick wall or capsule layer, tube extraction is recommended [35] . This method uses ethanol treatment followed by formic acid and acetonitrile to extract the proteins. An easy and effective preparation method is preferable for routine identification. Protein extraction from single colonies on a solid medium must be optimized [36] . To date, two primary methods have been reported for virus detection by MALDI-TOF MS. One is to detect virus proteins via the analysis of protein peaks by MALDI-TOF MS [16] . This process is similar to the preparation of bacteria and fungi. The second is to detect amplified viral nucleic acids. General preparation procedures can be roughly divided into two stages: (1) nucleic acid extraction and PCR amplification, and (2) single base extension of quality probes [19, 37, 38 ]. MALDI-TOF MS has been applied in clinical liquid samples, including blood culture, urine, cerebrospinal fluid (CSF), and various other body fluids; however, it is limited by the low number of pathogens and abundance of human cells [39] . Bacteria with a load of 10 5 -10 7 colony-forming units (CFUs) can be identified by MALDI-TOF MS [40] . For positive blood cultures, the sample needs to be concentrated and purified before identification. Various protocols have been reported to identify microorganisms present in positive blood culture broth accurately. Home-brew protocols were developed as follows: 1. Stepwise differential sedimentation of blood cells and microorganisms; 2. Centrifugation and lysis procedures; 3. Lysis-vacuum filtration; and 4. Centrifugation and membrane filtration techniques [20] . These processes may be performed by differential centrifugation and washing, selective lysis of blood cells, serum separator tubes, or filtration. Recently, a short incubation period based on positive blood culture has been used to improve the identification rate by using MALDI-TOF MS. These techniques measure the positive blood culture broth on a solid medium incubated for approximately 4-6 h and identify the colonies by using MALDI-TOF MS [27, 41] . Although this short incubation method is tedious, it can benefit the diagnosis and treatment of patients with bloodstream infections. However, the lack of standardized protocols, the use of different software for mass analysis, and different blood culture bottles make it difficult to compare the performances of the different methods. In addition to the home-brew methods, there are three major commercial kits: the Sepsityper ® kit (Bruker Daltonics GmbH, Bremen, Germany), the VITEK MS blood culture kit (bioMerieux, Marcy-l'Étoile, France), and the BACpro ® II kit (Nittobo Medical Coo, Tokyo, Japan) [20, 42, 43] . Currently, the cost of commercial kits is high, and it must be reduced. Pathogens in samples of positive urine routine tests detected by flow cytometry can be directly identified using MALDI-TOF MS [44] . Pre-treatment of the urine sample aims to remove human cells via centrifugation. The bacterial pellet is treated with formic acid, followed by MALDI-TOF MS. This method is of great significance for the rapid identification of pathogens in CSF samples. Previous studies have shown that positive smear microscopic examination samples treated after pathogen enrichment and centrifugal stratification to remove human cells can be used for MALDI-TOF MS identification [45] . The cell walls of Mycobacterium spp. and Nocardia spp. are thick. As a result, formic acid and matrix have limited effects on their cell walls. Therefore, the method of on-plate extraction does not perform well in identifying Mycobacterium spp. and Nocardia spp. Hence, the tube extraction method can effectively enhance the identification of the Mycobacterium spp. and Nocardia spp. The key to in-tube extraction is to use physical or chemical methods to destroy the cell wall and then use formic acid and acetonitrile to extract the proteins. The extraction methods for mycobacteria mainly include commercial-kit based extraction, with VITEK MS Mycobacterium/Nocardia kit being the commonly used commercial kit, and some protocols including sonication, bead beating, and vortexing in the presence of silica beads [46] [47] [48] . The sample preparation of Nocardia spp. by MALDI-TOF MS is similar to that of Mycobacterium spp. The identification efficacy of the extraction method has been reported to be better than that of the direct smear method [49] . The mycobacterium genus contains various species, some of which are highly pathogenic and must be inactivated before identification. The inactivation methods are different in the routine operation guidelines of Bruker Biotyper and VITEK MS. Bruker Biotyper uses the thermal inactivation method, whereas VITEK MS uses mechanical crushing with alcohol as the inactivation method. The samples were then treated with formic acid and acetonitrile, the extract was dripped onto the target plate, and the matrix was added after drying [50] . Currently, the identification protocol for filamentous fungi uses the home-brew method in the laboratory [11, 51] . The conventional matrix-assisted dissociation effect of formic acid and acetonitrile and a matrix in the laboratory is imperfect, and the extraction effect must be improved by physical or enzymatic lysis of the microbial cell wall. The mycelium is extracted by a modified formic acid extraction method. The ethanoltreated mycelium extract was harvested by centrifugation and air-dried. Then the extract was incubated with formic acid and acetonitrile, and finally dropped onto a target [11, 52] . MALDI-TOF MS technology shows advantages over traditional laboratory diagnostic technologies. The traditional identification of routine clinical culture bacteria relies on biochemical and microscopic examinations and other methods [53] . MALDI-TOF MS shortens the time required for traditional biochemical identification and improves the accuracy of traditional microscopic examinations [39, 54] . During the evaluation and analysis of common clinical isolates, the accuracy of determining the species level of common clinical pathogens was found to be close to 90% (Table 1 ). In 2009, a study comparing different assays for the identification of 1116 clinically encountered bacteria by using MALDI-TOF MS and biochemical testing systems showed correct species identification in 95.2% of cases [14] . Another study obtained the correct identification rate of 1660 bacterial isolates from 109 different species in 84.1% of cases [55] . In a study by Carbonnelle et al., 94 .9% and 83.4% of 296 isolates involving gram-positive and gram-negative bacteria, respectively, were correctly identified at the genus and species levels by MALDI-TOF MS [56] . In 2010, a study demonstrated a high rate of correct identification by MALDI-TOF MS compared to traditional biochemical tests in 99.1% of cases [5] . Several studies have been performed to evaluate the ability of MALDI-TOF MS to identify gram-negative bacterial strains [57, 58] and have identified it as a suitable and reliable tool for the identification of gram-negative rods [6, 59] . In a more recent study, Faron et al. evaluated the Bruker MALDI Biotyper system to identify gram-negative bacteria in a study including 2263 isolates and correctly identified 98.2% of the species [60] . In another study, correct identification was achieved in 87.83% of cases at the species level and 95.9% at the genus level [61] . Rychert et al. reported a multi-center evaluation of MALDI-TOF MS that correctly identified 92.8% grampositive pathogens at the species level [62] . In addition, Kassim et al. identified anaerobic gram-positive strains [61] . In conclusion, MALDI-TOF MS has a good identification rate in case of common clinical non-fermentative bacteria, Enterobacteriaceae, Staphylococcus, Streptococcus, Enterococcus, Corynebacterium, and HACEK strains [63] . Anaerobic bacteria are a major contributor to the human microbiota, with some colonizing the human body. Several species of anaerobic bacteria are the leading causes of clinical infections leading to death. Thus, the accurate identification of anaerobic bacteria plays an essential role in clinical diagnosis and treatment while avoiding unnecessary medication. MALDI-TOF MS has overcome some of the limitations of biochemical tests, greatly reducing the turnaround time and providing faster results [53, 64] . Several studies have evaluated the ability of MALDI-TOF MS in identifying anaerobic bacteria. These studies have shown that the identification rate of MALDI-TOF MS is above 80% (Table 1) . MALDI-TOF MS can identify common anaerobes in the clinic, and the accuracy of identification is higher than that of traditional biochemical test methods. Garner et al. showed that MALDI-TOF MS correctly identified 92.5% (602/651) anaerobic bacteria at the genus level and 91.2% (591/651) at the species level [65] . Another study reported that the identification rates of MALDI-TOF MS were 99.3% and 89.1% at the genus and species levels, respectively [7] . Rodríguez-Sánchez et al. showed 97.0% correct identification at the genus level and 85.8% at the species level [66] . In a meta-analysis, Li et al. evaluated the efficacy of MALDI-TOF MS in identification of clinical pathogenic anaerobes, including Bacteroides spp., Lactobacillus spp., Parabacteroides spp., Clostridium spp., Propionibacterium spp., Prevotella spp., Veillonella spp., and Peptostreptococcus spp. The authors reported that MALDI-TOF MS correctly identified the anaerobes at the genus and species levels at 92% and 84%, respectively. The identification accuracy of Bacteroides spp. was the highest, whereas the highest identification errors were for rare species [67] . Wang [3] . Therefore, MALDI-TOF MS is a proven useful method, although some anaerobes whose profiles are similar cannot be identified to the species level [68] . Clinically, Mycobacterium is the most important genus of bacteria. The Mycobacterium tuberculosis complex causes tuberculosis in humans. Traditionally, the biochemical phenotypic identification of mycobacteria lacks accuracy [69] . Molecular methods are widely used in clinical laboratories because of their high identification accuracy; however, these methods are time-consuming [70] . MALDI-TOF MS overcomes the shortcomings of phenotypic and molecular techniques. Recently, several studies have shown that MALDI-TOF MS is a valuable tool for identifying Mycobacterium isolates ( [71] . At present, MALDI-TOF MS identification of M. tuberculosis can only identify the M. tuberculosis complex [8] ; for nontuberculous mycobacterium, the detection of Mycobacterium avium complex is not sufficient to distinguish between Mycobacterium intracellulare and Mycobacterium chimera [72] . However, Epperson et al. reported that a novel MALDI Biotyper algorithm might accurately identify Mycobacterium intracellulare and Mycobacterium chimera [73] . For the identification of Nocardia spp., the mass spectra data available for Nocardia spp. are limited; thus, a self-built database can improve the identification rate. Blosser et al. conducted a multi-center study using three Nocardia libraries (Bruker, National Institutes of Health, and Ohio State University Libraries) to evaluate the efficacy of the Bruker MS method [74] . The identification efficacy upon using the three databases was significantly better than using a single database in all centers. Therefore, constantly updating the database is an indispensable factor for improving the MALDI-TOF MS identification accuracy rate [74] . A commercial database combined with a self-built database was used, and it resulted in an identification accuracy of 90% [75] . Durand et al. evaluated the performance of VITEK ® MS (V3.0) for the identification of Nocardia spp. The identification rates at the species level can be improved by using repeat or new extracts [9] . Most yeast species spectra that cause infections in humans have been included in the commercial database, and evaluations of the identification of yeasts have shown good performance [71, 76, 77] . In a study by Zhang et al., the performance of the VITEK MS v2.0 system for the identification of yeast isolates collected from patients with invasive fungal infections in the National China Hospital Invasive Fungal Surveillance Net 2011 program was evaluated. In that study, a total of 1243 isolates representing 31 yeast species were analyzed, and 97.3% of the isolates were correctly identified to the species level. Based on these results, a testing algorithm combining the VITEK MS system with selected supplementary ribosomal DNA sequencing was developed, which could be practically implemented in strategic programs for fungal infection surveillance [10] . Based on this program, Wang et al. conducted a comprehensive evaluation of the Bruker MS (v3.1) and VITEK MS (v2.0) systems, with the latter system accurately identifying 95.4% of the isolates (2559/2683) and the former system accurately identified 98.8% (2651/2683) of the isolates [78] . In that study, neither MALDI-TOF MS systems differentiated between Meyerozyma caribbica and M. guilliermondii among the M. guilliermondii complex isolates [78] . Autof MS 1000 (Autobio Diagnostics, Zhengzhou, China), a commercial MALDI-TOF MS system, has been available for routine pathogen identification in many clinical laboratories in China since 2018, and its accuracy (v2.0.18) was evaluated for the identification of 1228 isolates representing 14 different yeast species within five closely related species complexes compared to the VITEK MS (v3.0). For Candida albicans complex, C. glabrata complex isolates, C. parapsilosis complex, and C. neoformans complex, the accuracy of the identification rate to the species level was 99.4% vs. 96.3%, 98.9% vs. 94.7%, 99.0% vs. 79.1%, and 99.4% vs. 95.2% when using the Autof MS 1000 and VITEK MS, respectively. In addition, the Autof MS 1000 and VITEK MS systems both performed well for the identification of C. auris [4] . Therefore, MALDI-TOF MS is valuable for the routine identification of yeast species. Wang et al. reported that different culture media affected the identification accuracy of the Bruker Biotyper MALDI-TOF MS and VITEK MS systems for identifying C. tropicalis [79] . Further research is required to determine the effect of different culture media on microbial identification. For a long time, identifying filamentous fungi has been a challenge in clinical microbiology laboratories. The traditional identification methods of filamentous fungi, which rely primarily on morphology and require experienced staff, are time-consuming [80, 81] . Molecular technology, which is the gold standard for identification, has solved most of these problems. However, for filamentous fungi, molecular identification often requires sequencing multiple gene loci, which is expensive [81] . Fortunately, with the application of MALDI-TOF MS in clinical laboratories, the identification accuracy of filamentous fungi has improved (Table 1) , shortening the identification cycle. A recent study found that the correct identification rate of MALDI-TOF MS for Aspergillus filamentous fungi was more than 95% [82] . Another study showed that the MALDI-TOF MS system correctly identified 98.8% (1094/1107) to the species level [83] . In addition, for non-Aspergillus filamentous fungi, Li [88] . Furthermore, MALDI-TOF MS was applied for the identification of influenza virus types. The results showed 100% (29/29) concordance with results of molecular methods [17] . Bloodstream infection is a life-threatening disease that requires timely diagnosis and precise treatment. Compared to traditional methods, direct identification of pathogenic microorganisms from positive blood culture broth by using MALDI-TOF MS is quicker and less labor-intensive. The key to this process is to differentiate microorganisms from blood broth. Several commercial and laboratory-developed protocols have been reported to date. Owing to the lack of standardization techniques, the overall performance of these methods varies greatly, ranging from 60% to 100% at the species level [20, 44] . However, direct comparisons between these studies are difficult, because the identification rates can be affected by factors such as the pre-treatment method, microorganism distribution, and the cut-off value for species identification. In a previous study, Zhou et al. developed an in-house protocol for direct MALDI-TOF MS-based identification of organisms in positive blood cultures [89] . The data showed that the in-house protocol exhibited higher performance for gram-negative bacteria than for gram-positive bacteria, shown by Zhou et al. [89] . However, Zhou et al. also developed an in-house saponin-based extraction method for the direct identification of organisms in VersaTREK-positive blood cultures by using a Bruker Biotyper MALDI-TOF MS system (MALDI Biotyper). The identification rate of these protocols was 90.6% (184/203) and 82.3% (167/203) at the genus and species levels, respectively [21] . In addition, using the MIXED method, which is recommended by the manufacturer for suspected mixed organisms, two species were identified in more than half of the polymicrobial blood cultures [89] . Apart from direct pathogen identification, the potential of MALDI-TOF MS for antimicrobial susceptibility testing has also been explored [23] . Johansson et al. applied MALDI-TOF MS to screen for cfiA-positive B. fragilis strains directly from blood culture bottles with various ertapenem minimum inhibitory concentrations, yielding results within 3 h [90] . Oviano et al. expanded their use to detect carbapenem-resistant gram-negative bacteria causing bloodstream infections [91] . A MALDI-TOF MS-based assay was established by measuring the hydrolysis of imipenem in blood cultures spiked with Pseudomonas aeruginosa, Acinetobacter baumannii, or Enterobacteriaceae producing different carbapenemases [92] . The analysis was performed using MBT Compass STAR-BL module software (Bruker Daltonics), automatically providing a result (sensitivity or resistance) based on the degree of hydrolysis of the antibiotic. The assay achieved 98% sensitivity and 100% specificity after 30 min of incubation, and both values increased to 100% after 60 min of incubation [92] . However, performing antimicrobial susceptibility testing directly from blood cultures with MALDI-TOF MS is impossible for most laboratories. Further research is required to achieve this goal; however, the potential applications of MALDI-TOF MS are promising for use in all aspects of blood culture testing. For the identification of urine samples, positive samples were measured using flow cytometry of routine urine tests, and many bacteria were suitable for direct identification. Ilki et al. reported a microorganism identification rate of 91.8% (484/538) for MALDI-TOF MS, directly from urine samples [22] . Other studies revealed that approximately 10 4 -10 6 CFU/mL of bacteria in urine could be detected by MALDI-TOF MS [93] [94] [95] . Li et al. reported that MALDI-TOF MS combined with UF1000i could be used for rapid identification within 1 h. The accuracy of direct identification was shown to be 86.42% (229/265) of single-microorganism samples [28] . However, similar to the identification of positive blood samples, identification directly from urine by MALDI-TOF MS is more effective for identifying single bacterial samples, whereas the effect on > 1 bacterial species is poor. Although the rapid identification of urine samples by MALDI-TOF MS is of great significance to patients, several associated problems remain. The identification rate of gram-positive bacteria is higher than that of gram-negative bacteria [96] . Several studies have shown that large amounts of urine are required, with several tests requiring 10 mL, which may be difficult to obtain from children [91, 97, 98] . Compared with urethral infection, CSF infection has a greater impact on patients, and rapid diagnosis and treatment are of great significance. Currently, there are few studies describing the use of MALDI-TOF MS to detect microorganisms in the CSF. One case of bacterial meningitis caused by Klebsiella pneumoniae was directly identified using MALDI-TOF MS [99] . Similarly, Hartmeyer et al. directly identified S. pneumoniae by using MALDI-TOF MS in CSF samples [100] . Torres et al. showed the direct identification of biomarkers in CSF samples from patients with enteroviral meningitis by using MALDI-TOF MS [29] . Furthermore, Bishop et al. reported that MALDI-TOF MS effectively detected gramnegative rods in smear-positive CSF samples; however, it was not suitable for identifying gram-positive bacteria [45] . In conclusion, with further study of the role of MALDI-TOF MS in CSF identification in the future, MALDI-TOF MS may be used more extensively. If a specific protein in pathogenic bacteria is related to antimicrobial resistance, the quality peak of the protein can be detected by MALDI-TOF MS to distinguish sensitive and drug-resistant pathogens. MALDI-TOF MS can identify methicillin-resistant Staphylococcus aureus (MRSA) and methicillin-susceptible S. aureus (MSSA) by detecting small peptides of phenol-soluble modulin (PSM)-mec. The presence of the 2415 ± 2.00 m/z peak can predict mecA carriage in S. aureus. However, because of its low sensitivity, it is easily affected by strain differences and other factors. At present, the strains identified by MS with small peptides of PSM-mec are MRSA; however, strains that cannot be identified by MS as MSSA [101, 102] . An extended-spectrum class C beta-lactamase from A. baumannii was recently identified (m/z~40,279) as a biomarker for carbapenem resistance. The sensitivity and specificity of the method were 96% and 73%, respectively, compared with the microdilution imipenem susceptibility testing method [103] . An 11,109-Da peak was identified by MALDI-TOF MS to distinguish the protein patterns of Klebsiella pneumoniae carbapenemase (KPC)-producing and non-KPC-producing K. pneumoniae strains [104] . A previous study revealed that the 11,109-Da peak is a cleavage product of a hypothetical protein named p019 [105] . However, as p019 polypeptide was absent in a subset of bla KPC -harboring plasmids, negative results should also be confirmed by other tests. The identification of possible markers associated with drug resistance by MALDI-TOF MS has also been investigated in anaerobes. cfiApositive B. fragilis can be distinguished from cfiA-negative strains by a set of peak shifts in the range of 4000-5500 Da [106] . The screening of characteristic peaks related to drug resistance can detect the pathogen associated with drug resistance; however, the methods described above are specific for one resistance mechanism. Owing to its low sensitivity, the method must be optimized in the future to improve identification accuracy further. Serotype VI Group B Streptococcus (GBS) is an important invasive pathogen, and Li et al. found that the protein peak of 6251 Da appeared in most serotype VI GBS (20/24, 92%), and the protein peak of 6891 DA was observed in most serotype III (15/18, 83%) and Ib (19/23, 83%) strains. Latex agglutination methods combined with MALDI-TOF MS may quickly identify GBS serotypes [107] . Among the Clostridioides difficile isolates, the hypervirulent RT027 isolates caused largescale outbreaks in North America and European countries. Detection of the hypervirulent RT027 strains based on the absence of a peak at 6654 m/z and the presence of a peak at 6712 m/z by MALDI-TOF MS presented a sensitivity of 100%, specificity of 91.7%, and positive predicted value of 95% [24] . However, the major clone spreading in Asian countries was the hypovirulence strain, mainly ST37 and the recently identified ST81 strains. Our previous study identified five peaks, which revealed that five markers (2691. 43 At present, MALDI-TOF MS can correctly identify most pathogens; however, the identification accuracy of some pathogens with similar mass spectra by MALDI-TOF MS is low. These are some strains of complex bacterial groups, including A. calcoaceticus-A. baumannii complex, Enterobacter cloacae complex, Burkholderia cepacia complex, and S. mitis complex [15, [108] [109] [110] . In addition, Shigella and Escherichia coli cannot be accurately identified by MALDI-TOF MS because they have similar mass spectra [111] . For the M. tuberculosis complex too, one of the reasons that affect the effectiveness of MALDI-TOF MS is the similarity of the mass spectra. The identification accuracy of H. pylori is low because there are currently fewer mass spectra for this species in the database. With the continuous enrichment of spectral databases and improvement of spectral quality, MALDI-TOF MS will be more widely used in clinical laboratories [80, 83, 112] . MALDI-TOF MS has been widely used in clinical microbiology for a dozen years. Nowadays, MALDI-TOF MS can identify common microorganisms to the species level. Moreover, MALDI-TOF MS systems can technically identify some bacterial subspecies [114] [115] [116] [117] ; however, with optimizing methods and technological innovation, MALDI-TOF MS still has more application growth in this field. The reliability and accuracy of MALDI-TOF MS have been analyzed in many studies, and the spectrum database is vital for identification. In the future, MALDI-TOF MS is expected to play a more critical role in promoting the development of clinical microbiology because it is fast, accurate, sensitive, and has a high throughput. Informed Consent Statement: Not applicable. The data presented in this study are available on request from the corresponding author. Prod'hom, G. Performance of matrix-assisted laser desorption ionization-time of flight mass spectrometry for identification of bacterial strains routinely isolated in a clinical microbiology laboratory Matrix-assisted laser desorption ionization-time of flight mass spectrometry in clinical microbiology Evaluation of VITEK MS, Clin-ToF-II MS, Autof MS 1000 and VITEK 2 ANC card for identification of Bacteroides fragilis group isolates and antimicrobial susceptibilities of these isolates in a Chinese university hospital Evaluation of Autof MS 1000 and Vitek MS MALDI-TOF MS System in Identification of Closely-Related Yeasts Causing Invasive Fungal Diseases. Front. Cell Infect Comparison of two matrix-assisted laser desorption ionization-time of flight mass spectrometry methods with conventional phenotypic identification for routine identification of bacteria to the species level Rapid and reliable identification of Gram-negative bacteria and Gram-positive cocci by deposition of bacteria harvested from blood cultures onto the MALDI-TOF plate Comparison of two matrix-assisted laser desorption/ionization-time of flight (MALDI-TOF) mass spectrometry methods and API 20AN for identification of clinically relevant anaerobic bacteria Performance of Vitek MS v3.0 for Identification of Mycobacterium Species from Patient Samples by Use of Automated Liquid Medium Systems Assessment of VITEK(R) MS IVD database V3.0 for identification of Nocardia spp. using two culture media and comparing direct smear and protein extraction procedures Yeast identification algorithm based on use of the Vitek MS system selectively supplemented with ribosomal DNA sequencing: Proposal of a reference assay for invasive fungal surveillance programs in China Identification by Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry and Antifungal Susceptibility Testing of Non-Aspergillus Molds Identification of different respiratory viruses, after a cell culture step, by matrix assisted laser desorption/ionization time of flight mass spectrometry (MALDI-TOF MS) COVID-19, Chikungunya, Dengue and Zika Diseases: An Analytical Platform Based on MALDI-TOF MS, IR Spectroscopy and RT-qPCR for Accurate Diagnosis and Accelerate Epidemics Control. Microorganisms Performance of a matrix-assisted laser desorption ionization-time-of-flight mass spectrometry system for the identification of bacterial isolates in the clinical routine laboratory Factors Associated With MALDI-TOF Mass Spectral Quality of Species Identification in Clinical Routine Diagnostics Detection of SARS-CoV-2 in nasal swabs using MALDI-MS Establishment and Application of a Universal Coronavirus Screening Method Using MALDI-TOF Mass Spectrometry. Front Multiplex detection of 60 hepatitis B virus variants by maldi-tof mass spectrometry Multiplex detection of human herpesviruses from archival specimens by using matrix-assisted laser desorption ionization-time of flight mass spectrometry Mass spectrometry-based microbiological testing for blood stream infection Applicability of an in-house saponin-based extraction method in Bruker Biotyper matrix-assisted laser desorption/ionization time-of-flight mass spectrometry system for identifying bacterial and fungal species in positively flagged pediatric VersaTREK blood cultures An alternative for urine cultures: Direct identification of uropathogens from urine by MALDI-TOF MS Detection of Antibiotic-Resistance by MALDI-TOF Mass Spectrometry: An Expanding Area Bocanegra-Ibarias, P. Screening of biomarkers of drug resistance or virulence in ESCAPE pathogens by MALDI-TOF mass spectrometry Use of matrix-assisted laser desorption ionization-time of flight mass spectrometry to identify MLST clade 4 Clostridium difficile isolates Methods for the Identification of Cultured Microorganisms Using Matrix-Assisted Laser Desorptio Rapid identification of bacteria from positive blood culture bottles by MALDI-TOF MS following short-term incubation on solid media Rapid Identification and Antimicrobial Susceptibility Testing for Urinary Tract Pathogens by Direct Analysis of Urine Samples Using a MALDI-TOF MS-Based Combined Protocol Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF-MS) proteomic profiling of cerebrospinal fluid in the diagnosis of enteroviral meningitis: A proof-of-principle study A combined approach of MALDI-TOF mass spectrometry and multivariate analysis as a potential tool for the detection of SARS-CoV-2 virus in nasopharyngeal swabs The rapid identification of intact microorganisms using mass spectrometry Formic acid-based direct, on-plate testing of yeast and Corynebacterium species by Bruker Biotyper matrix-assisted laser desorption ionization-time of flight mass spectrometry Identification of anaerobic bacteria by Bruker Biotyper matrix-assisted laser desorption ionization-time of flight mass spectrometry with on-plate formic acid preparation Identification of Gram-positive cocci by use of matrix-assisted laser desorption ionization-time of flight mass spectrometry: Comparison of different preparation methods and implementation of a practical algorithm for routine diagnostics Matrix-assisted laser desorption ionization-time of flight mass spectrometry for fast and reliable identification of clinical yeast isolates Optimization of sample preparation for intact cell mass spectrometry (matrix-assisted laser desorption/ionization linear time-of-flight mass spectrometry) of endophytic Xylaria A new PCR-based mass spectrometry system for high-risk HPV, part I: Methods. Am Simultaneous detection and identification of enteric viruses by PCR-mass assay Current Status of Matrix-Assisted Laser Desorption/Ionization-Time-of-Flight Mass Spectrometry (MALDI-TOF MS) in Clinical Diagnostic Microbiology Applications of MALDI-TOF mass spectrometry in clinical diagnostic microbiology Rapid Identification of Microorganisms from Positive Blood Culture by MALDI-TOF MS After Short-Term Incubation on Solid Medium Applications of copolymer for rapid identification of bacteria in blood culture broths using matrix-assisted laser desorption ionization time-of-flight mass spectrometry Development of an improved rapid BACpro(R) protocol and a method for direct identification from blood-culture-positive bottles using matrix-assisted laser desorption ionization time-of-flight mass spectrometry Current Status of Proteomic Technologies for Discovering and Identifying Gingival Crevicular Fluid Biomarkers for Periodontal Disease The use of matrix-assisted laser desorption/ionization time-of-flight mass spectrometry for rapid bacterial identification in patients with smear-positive bacterial meningitis Evaluation of Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry for Identification of Nontuberculous Mycobacteria from Clinical Isolates Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry for Combined Species Identification and Drug Sensitivity Testing in Mycobacteria Improved Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry (MALDI-TOF MS)-Based Identification of Mycobacterium spp. by Use of a Novel Two-Step Cell Disruption Preparatory Technique Routine identification of Nocardia species by MALDI-TOF mass spectrometry Comparison of Sample Preparation Methods, Instrumentation Platforms, and Contemporary Commercial Databases for Identification of Clinically Relevant Mycobacteria by Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry Development and Validation of an In-House Library for Filamentous Fungi Identification by MALDI-TOF MS in a Clinical Laboratory in Medellin (Colombia) Species identification and antifungal susceptibility testing of Aspergillus strains isolated from patients with otomycosis in northern Matrix-assisted laser desorption ionization-time of flight mass spectrometry: A fundamental shift in the routine practice of clinical microbiology Prospective evaluation of a matrix-assisted laser desorption ionization-time of flight mass spectrometry system in a hospital clinical microbiology laboratory for identification of bacteria and yeasts: A bench-by-bench study for assessing the impact on time to identification and cost-effectiveness Ongoing revolution in bacteriology: Routine identification of bacteria by matrix-assisted laser desorption ionization time-of-flight mass spectrometry Robustness of two MALDI-TOF mass spectrometry systems for bacterial identification Comparison of Bruker Biotyper matrix-assisted laser desorption ionization-time of flight mass spectrometer to BD Phoenix automated microbiology system for identification of gram-negative bacilli Optimization of routine identification of clinically relevant Gram-negative bacteria by use of matrixassisted laser desorption ionization-time of flight mass spectrometry and the Bruker Biotyper MALDI-TOF identification of Gram-negative bacteria directly from blood culture bottles containing charcoal: Sepsityper(R) kits versus centrifugation-filtration method Multicenter Evaluation of the Bruker MALDI Biotyper CA System for the Identification of Clinical Aerobic Gram-Negative Bacterial Isolates Comparison of biomarker based Matrix Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry (MALDI-TOF MS) and conventional methods in the identification of clinically relevant bacteria and yeast Multicenter evaluation of the Vitek MS matrix-assisted laser desorption ionization-time of flight mass spectrometry system for identification of Gram-positive aerobic bacteria Performances and Reliability of Bruker Microflex LT and VITEK MS MALDI-TOF Mass Spectrometry Systems for the Identification of Clinical Microorganisms Emerging and Future Applications of Matrix-Assisted Laser Desorption Ionization Time-of-Flight (MALDI-TOF) Mass Spectrometry in the Clinical Microbiology Laboratory: A Report of the Association for Molecular Pathology Multi-centre evaluation of mass spectrometric identification of anaerobic bacteria using the VITEK(R) MS system Evaluation of MALDI-TOF MS (Matrix-Assisted Laser Desorption-Ionization Time-of-Flight Mass Spectrometry) for routine identification of anaerobic bacteria Application of MALDI-TOF MS to rapid identification of anaerobic bacteria Performance of mass spectrometric identification of clinical Prevotella species using the VITEK MS system: A prospective multi-center study Routine identification of microorganisms by matrix-assisted laser desorption ionization time-of-flight mass spectrometry: Success rate, economic analysis, and clinical outcome An improved simple method for the identification of Mycobacteria by MALDI-TOF MS (Matrix-Assisted Laser Desorption-Ionization mass spectrometry) Advantages of using matrix-assisted laser desorption ionization-time of flight mass spectrometry as a rapid diagnostic tool for identification of yeasts and mycobacteria in the clinical microbiological laboratory Comparison of methods available for identification of Mycobacterium chimaera Evaluation of a Novel MALDI Biotyper Algorithm to Distinguish Mycobacterium intracellulare From Mycobacterium chimaera Multicenter Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry Study for Identification of Clinically Relevant Nocardia spp Comparison of Mechanical Disruption Techniques for Rapid Inactivation of Mycobacterium and Nocardia Species before Identification Using Matrix-Assisted Laser Desorption Ionization-Time of Flight (MALDI-TOF) Mass Spectrometry Improved clinical laboratory identification of human pathogenic yeasts by matrix-assisted laser desorption ionization time-of-flight mass spectrometry Comparison of the accuracy of two conventional phenotypic methods and two MALDI-TOF MS systems with that of DNA sequencing analysis for correctly identifying clinically encountered yeasts A Comprehensive Evaluation of the Bruker Biotyper MS and Vitek MS Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry Systems for Identification of Yeasts, Part of the National China Hospital Invasive Fungal Surveillance Net (CHIF-NET) Study, 2012 to 2013 Evaluation of Bruker Biotyper and Vitek MS for the identification of Candida tropicalis on different solid culture media Identification of filamentous fungi isolates by MALDI-TOF mass spectrometry: Clinical evaluation of an extended reference spectra library R)) MS v3.0 System in the Identification of Filamentous Fungi Identification of clinical isolates of Aspergillus, including cryptic species, by matrix assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) Matrixassisted laser desorption ionization time-of-flight mass spectrometry: Revolutionizing clinical laboratory diagnosis of mould infections MS for the rapid identification and drug susceptibility testing of filamentous fungi Development of a clinically comprehensive database and a simple procedure for identification of molds from solid media by matrix-assisted laser desorption ionization-time of flight mass spectrometry Pseudallescheria/Scedosporium complex species identification by Matrix-Assisted Laser Desorption Ionization Time-Of-Flight Mass Spectrometry Development of high-throughput genotyping method of all 18 HR HPV based on the MALDI-TOF MS platform and compared with the Roche Cobas 4800 HPV assay using clinical specimens Sensitive and rapid detection of viruses associated with hand foot and mouth disease using multiplexed MALDI-TOF analysis An Improved In-house MALDI-TOF MS Protocol for Direct Cost-Effective Identification of Pathogens from Blood Cultures Instant screening and verification of carbapenemase activity in Bacteroides fragilis in positive blood culture, using matrix-assisted laser desorption ionization-time of flight mass spectrometry Rapid direct detection of carbapenemase-producing Enterobacteriaceae in clinical urine samples by MALDI-TOF MS analysis Universal protocol for the rapid automated detection of carbapenemresistant Gram-negative bacilli directly from blood cultures by matrix-assisted laser desorption/ionisation time-of-flight mass spectrometry (MALDI-TOF/MS) Direct identification of urinary tract pathogens from urine samples by matrix-assisted laser desorption ionization-time of flight mass spectrometry Direct Identification of Urinary Tract Pathogens From Urine Samples Using the Vitek MS System Based on Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry Rapid urine preparation prior to identification of uropathogens by MALDI-TOF MS Matrix-Assisted Laser Desorption/Ionization Time-of-Flight: The Revolution in Progress Diafiltration MALDI-TOF mass spectrometry method for culture-independent detection and identification of pathogens directly from urine specimens A multicentre study investigating parameters which influence direct bacterial identification from urine Direct application of MALDI-TOF mass spectrometry to cerebrospinal fluid for rapid pathogen identification in a patient with bacterial meningitis Mass spectrometry: Pneumococcal meningitis verified and Brucella species identified in less than half an hour. Scand Detection of methicillin-resistant coagulase-negative staphylococci harboring the class A mec complex by MALDI-TOF mass spectrometry Evaluation of Staphylococcus aureus Subtyping Module for Methicillin-Resistant Staphylococcus aureus Detection Based on Matrix-Assisted Laser Desorption Ionization Time-of-Flight Mass Spectrometry Direct detection of carbapenemase-associated proteins of Acinetobacter baumannii using nanodiamonds coupled with matrix-assisted laser desorption/ionization time-of-flight mass spectrometry Evaluation of Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry for Identification of KPC-Producing Klebsiella pneumoniae A rapid matrix-assisted laser desorption ionization-time of flight mass spectrometry-based method for single-plasmid tracking in an outbreak of carbapenem-resistant Enterobacteriaceae Differentiation of division I (cfiA-negative) and division II (cfiA-positive) Bacteroides fragilis strains by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry Identification of a proteomic biomarker associated with invasive ST1, serotype VI Group B Streptococcus by MALDI-TOF MS MALDI-TOF MS and chemometric based identification of the Acinetobacter calcoaceticus-Acinetobacter baumannii complex species Matrix-assisted laser desorption/ionization time-of-flight MS for the accurate identification of Burkholderia cepacia complex and Burkholderia gladioli in the clinical microbiology laboratory MALDI-TOF mass spectrometry for differentiation between Streptococcus pneumoniae and Streptococcus pseudopneumoniae Multiple independent origins of Shigella clones of Escherichia coli and convergent evolution of many of their characteristics Differentiation of Gastric Helicobacter Species Using MALDI-TOF Mass Spectrometry. Pathogens 2021 Identification of Candida yeasts: Conventional methods and MALDI-TOF MS Rapid discrimination of Bifidobacterium animalis subspecies by matrix-assisted laser desorption ionization-time of flight mass spectrometry Classification algorithm for subspecies identification within the Mycobacterium abscessus species, based on matrix-assisted laser desorption ionization-time of flight mass spectrometry Dissecting the taxonomic heterogeneity within Propionibacterium acnes: Proposal for Propionibacterium acnes subsp. acnes subsp. nov. and Propionibacterium acnes subsp. elongatum subsp. nov Fusobacterium nucleatum subspecies identification by matrix-assisted laser desorption ionization-time of flight mass spectrometry The authors declare no conflict of interest.