key: cord-0062496-hncqs1cs authors: Kumar, Rahul; Dubey, P. K.; Zafer, Afaqul; Kumar, Ashok; Yadav, Sanjay title: Design and Development of a Temperature-Compensated Body Mass Index Measuring System date: 2021-04-30 journal: MAPAN DOI: 10.1007/s12647-021-00448-y sha: b37ce9b676d0b2788c0f699c1e4d1b99a12ca9ad doc_id: 62496 cord_uid: hncqs1cs The measurement of body mass index (BMI) is an important indicator of preliminary assessment of human health. A higher BMI value increases the risk factor for critical diseases, e.g., high blood pressure (hypertension), cardiovascular disease, gallbladder disease, type 2 diabetes, etc. Obesity also increases the risk factor of COVID-19 in patients. BMI, the ratio of mass (in kilogram) to the square of height (in meter), is an estimation of an individual's physique. It depends upon the adult’s height and mass and is intended to evaluate the tissue mass. In this article, we have described the design and developmental aspects of a temperature-compensated BMI measuring system. The developed BMI measuring system is comprised of an ultrasonic sensor for height measurement, strain gauge-based load cell for precise and accurate mass measurement, a temperature sensor to eliminate the effect of temperature, and a microcontroller with related electronics, software, and display system. The developed BMI system is capable to measure mass precisely in the range (0–200) kg with a resolution of 0.01 kg, and height in the range of (0–4) m with resolution of 0.01 m. The expanded uncertainty in height measurement is estimated as 2.8 mm. The BMI value is determined with the help of these parameters and displayed with precision of 0.01. The novelty of the developed system is temperature compensation in the height measurement using an ultrasonic sensor, which has been introduced first time. Also, system is improved for its highly precise and accurate measurements at a comparatively lower cost than the commercially available existing systems. BMI is used to calculate body fat, and its value provides information about patients' risk for diseases due to over mass and obesity. Obesity is identified as a cause of risk factor for many non-communicable diseases, such as diabetes, blood pressure, hypertension, cholesterol, triglycerides, and insulin resistance [1] [2] [3] [4] [5] [6] [7] [8] [9] [10] [11] [12] . The higher BMI value also increases the risk of cancer of the breast, colon, prostate, endometrium, kidney, and gall bladder [13] . Mortality rate increases with increasing degree of over mass, as the measured value of BMI. All over the world, as per World Health Organization (WHO) report, around 2.8 million people die each year due to over mass or obesity. An estimated 35.8 million people (2.3%) of global disability-adjusted life years (DALYs) are caused by over mass or obesity. Apart from the over mass, under mass is also the reason for malnutrition, vitamin deficiencies, anaemia, osteoporosis, a decrease in immune function, growth, and development issues particularly in children and teenagers. In India, children under the age of 5 years, more than 30%, are suffering from malnutrition which is very high [14] [15] [16] [17] . In most of the countries, BMI measuring system is used to identify a person having under mass or over mass. It is widely used due to its fast response and ease of operations. As per WHO recommendation, based on BMI value nutritional status of adults is categorized in Table 1 . Recently, it is reported that the obese person has a higher risk factor of COVID-19, bacterial and viral pneumonia, acute respiratory distress syndrome (ARDS), and acute respiratory failure after lung transplantation. In some of the studies, it is reported that obesity is more common among patients who require hospital admission for COVID-19, that it may be associated with an increased risk *Corresponding author, E-mail: rprahulprajapati2@gmail.com for death. Early reports also suggest that diabetes, hypertension, and cardiovascular disease are associated with invasive mechanical ventilation and death in COVID-19 [19] [20] [21] [22] [23] [24] [25] . In the USA, around 42.4% adult population is obese, and out of that 9.2% are severely obese. The Centers for Disease Control and Prevention lists severe obesity at any age that means a patient having BMI C 40 kg/m 2 has a high risk of the condition being affected by COVID-19. Given the high prevalence of obesity, the potential effect of COVID-19 in the US population is higher [26] [27] [28] . Therefore, it is essential to monitor the BMI of every person and those who are suffering from obesity. It is advisable for these people to take extra precautions and prevention from the COVID-19 in this pandemic time [29] [30] [31] [32] [33] [34] [35] [36] . The developed system measures BMI with high precision and accuracy as compared to existing systems [37] [38] [39] [40] . In case of India, the under mass issue is normally applicable for the people below the poverty line where the nutrient and healthy food availability is not always adequate or enough. The main objective of the present study was to work toward finding the suitable mean for precise, accurate, cheaper, ease in operation and 'Made in India' medical device to regularly monitor the vital health parameters of the patients, the high-risk users, and even for the healthier people as a preventive measure. This has more relevance in the current humongous COVID-19 situations. This has motivated us to design and develop an improved, indigenous, and low-cost BMI measuring system. It also provides a simple, fast, and low-cost health assessment of the vital parameters to the users and stakeholders. In the present investigation, the height of an object/person is measured using an ultrasonic sensor (40 kHz). To measure the height, a transmitter is used to transmit ultrasonic waves, which are reflected from the object. The reflected waves are received using a receiver. The time delay between the transmitted and received waves is measured using a microcontroller (ATmega 328). The ultrasonic sensor (HC-SR04) is used in the development of the BMI measuring system which works on the principle of the pulse-echo method of ultrasonic. In the pulse-echo method, ultrasonic pulse is transmitted and when an object comes into its path echo is generated. This echo is received by the receiver of the ultrasonic sensor. So, the distance travel by the ultrasound pulse becomes 2d. Temperature-compensated ultrasonic velocity is determined using Eq. (3). Then, temperature-compensated distance is determined as: where d is the height between the platform and the sensor, c(T) is the temperature-compensated ultrasonic velocity in air, and t d is the time delay between transmitted and received waves. The working principle of the ultrasonic distance measurement is graphically depicted in Fig. 1 . The ultrasonic velocity in air is found to vary with changes in air temperature. Therefore, any change in the prevailing environment temperature around the place of use of the device will have a significant effect in the measured height. To compensate the effect of temperature in height measurement, a temperature sensor (LM-35) is used. The four load cell sensors, commercially available, have been used to measure the mass precisely. The load cell converts the mass into the form of the electrical signal which is appropriately amplified and displayed. In the developed BMI system, four load cells are mounted beneath a flat platform. The object/person under test is placed /used to stand on the platform. Normally, the placing or standing positions of the object/person are not uniform for providing the equal distribution on load cells. Therefore, the use of four strain gauge-based load cells addresses this issue amicably. The stress or mass generates an electrical output of the load cell is interfaced with a 24-bit analog-to-digital converter (ADC) to achieve higher resolution, and this ADC module is interfaced with a microcontroller which provides the precise value of the mass on a large liquid crystal display (LCD). A simplified block diagram and the photograph of the developed BMI measuring system are shown in Fig. 2a and b, respectively. The BMI measuring system is comprised of an ultrasonic sensor (HC-SR04), four strain gauge-based load cell sensors, an Arduino development board, a temperature sensor (LM-35), a 24-bit analog-to-digital converter module, and a large (2 9 16 character) (LCD). The ultrasonic sensor has been used to measure the height. The ultrasonic sensor has four pins for interfacing with the microcontroller. Accordingly, the four pins, namely Vcc, Trigger, Echo, and Ground, were interfaced with the microcontroller. The ultrasonic sensor module triggers (connected at digital pin-8 of the microcontroller) the 10 ls pulse and then it transmits the eight cycles of sonic burst which is reflected back when any object comes in the range of its path as shown in the timing diagram in Fig. 1 . The echo of the ultrasonic wave is received on the echo pin which is interfaced on the digital pin-9 of the microcontroller. The time delay between the transmitted and received waves is measured using microcontroller. The Fig. 3 Flowchart of the developed algorithm for BMI measuring system R. Kumar et al. distance between the ultrasonic sensor and object is determined using Eq. (1). The height of the person (h in m), the difference of the distance between the ultrasonic sensor and platform (d in m), and the distance between sensor and person (y in m). Then, the height of the person h is determined using The ultrasonic sensor is operated at ? 5 V. For the mass measurement, four strain gauge-based load cells have been used. As load cell generates the electrical output in the order of mV, its analog electrical output is converted into digital by using a 24-bit ADC module (HX-711) which finally provides measured values of mass in kg. As mentioned earlier, a temperature sensor (LM-35) is used to compensate the temperature effect in height measurement. The liquid crystal display (LCD) is interfaced with the microcontroller using a four-wire data communication protocol which is used to display the height, mass, and calculated BMI of an person. The programming of the BMI device is done in the Arduino Integrated Development Environment (IDE). The flowchart of the developed BMI measuring system program code is shown in Fig. 3 . For the temperature compensation in the height measurement, the value of ultrasonic velocity as 347.1 m/s (at 25°C and 50% relative humidity) measured experimentally as well as reported in the literature [41] has been programmed for height measurement [42] . A linear relationship between ultrasonic velocity and temperature is shown in Eq. 3. where c T ð Þ is the temperature-compensated ultrasonic velocity in m/s and T is temperature in°C. A linear relationship is also obtained between temperature-compensated ultrasonic velocity and the temperature at 50% RH as shown in Fig. 4 . Figure 4 clearly shows the evidence for dependence of ultrasonic velocity on environmental conditions. The uncertainty in the height measurement without temperature correction is 2.8 mm (at k = 2) while measuring the height of 738 mm with a temperature variation well within 0.5°C. The experimental data obtained on the height measurement as a function of temperature as well as without temperature are shown in Table 2 . The linear relationship between height measurement with temperature correction and temperature is as follows: In the linear fitted Eq. (4), the slope is 0.2 which shows that as temperature increases the measurement of height is changed, while in the linear fitted Eq. (5), the value of the slope is significantly higher as compared to Eq. (4), e.g., 0.88 which significantly shows that variation in height measurement is more as compared to Eq. (4) as shown in Fig. 5 . This variation in the height measurement comes due to temperature. Therefore, the developed BMI measuring system compensates the temperature effect in height measurement and measures accurate BMI value. For the testing and performance evaluation of the device, the BMI is measured for a sample set of six persons in the laboratory, and measurements were repeated at least 5 times on each person to calculate the BMI manually. The BMI values obtained using the present system are in close agreement with the values calculated manually as shown in Fig. 6 . The maximum variation in the measured BMI value is found to be 0.5 kg/m 2 (Table 3) . A simple, low cost and indigenous digital BMI measuring system is designed and developed. The developed system is capable to measure and altogether the values of vital health parameters of a person under test related to height with a resolution of 0.01 m in the range (0-4) m; mass with a resolution of 0.01 kg in the range (0-200) kg; and BMI value with a precision of 0.01. The unique design provides the temperature-compensated height measurements which are first time made available in the present system. The expanded measurement uncertainty associated with height measurement is estimated as 2.8 mm which improves the overall precision of the BMI value to 0.01. The developed system would be highly useful for the hospitals, private medical practitioners, sports associations, and other stakeholders as a low cost and quick solution for assessment of vital health parameters of individuals or community purpose and accordingly planning for the next course of action. Visceral adiposity and its anatomical distribution as predictors of the metabolic syndrome and cardiometabolic risk factor levels Impact of abdominal visceral and subcutaneous adipose tissue on cardiometabolic risk factors: the Jackson Heart Study Associations of visceral and abdominal subcutaneous adipose tissue with markers of cardiac and metabolic risk in obese adults Visceral fat volume predicts new onset type 2 diabetes in patients with chronic hepatitis C Relationship between visceral fat accumulation and anti-lipolytic action of insulin in patients with type 2 diabetes mellitus Visceral fat: a key mediator of steatohepatitis in metabolic liver disease Body fat distribution, incident cardiovascular disease, cancer, and all-cause mortality Visceral obesity, metabolic syndrome, insulin resistance and cancer Past, present and future of blood pressure measuring instruments and their calibration Long-term follow-up of patients with NAFLD and elevated liver enzymes Skeletal muscle attenuation determined by computed tomography is associated with skeletal muscle lipid content Obesity and cancer risk: recent review and evidence Expert panel on integrated guidelines for cardiovascular health and risk reduction in children and adolescents Suboptimal feeding and caring practices among young Indian children ages 12 to 24 mo living in the slums of New Delhi Populations at risk for severe or complicated influenza illness: systematic review and meta analysis Obesity is associated with higher risk of intensive care unit admission and death in influenza A (H1N1) patients: a systematic review and meta-analysis Body mass index is associated with the development of acute respiratory distress syndrome Obesity paradox'' in acute respiratory distress syndrome: a systematic review and metaanalysis Obesity and primary graft dysfunction after lung transplantation: the Lung Transplant Outcomes Group Obesity Study Clinical characteristics of coronavirus disease 2019 in China Clinical characteristics and outcomes of 112 cardiovascular disease patients infected by 2019-nCoV Obesity could shift severe COVID-19 disease to younger ages Prevalence of Obesity and Severe Obesity Among Adults: United States Body mass index and risk for intubation or death in SARS-CoV-2 infection: a retrospective cohort study Overmass, obesity, and risk of hospitalization for COVID-19: a community-based cohort study of adults in the United Kingdom Prevalence of malnutrition and analysis of related factors in elderly patients with COVID-19 in Wuhan, China COVID-19 and severe obesity: a big problem Advanced body composition assessment: from body mass index to body composition profiling Lifestyle risk factors for cardiovascular disease in relation to COVID-19 hospitalization: a community-based cohort study of 387,109 adults in UK. medRxiv Measuring weight with e-scales in clinical and research settings during the COVID-19 pandemic Obesity and mortality among patients diagnosed with COVID-19: results from an integrated health care organization BMI assessment machine with recommended ideal weight Body mass index measurement system as a desktop-based nutrition monitor Digital and analog body mass index calculating device: a comparative study A Framework for Internet of Things-Based Body Mass Index Estimation and Obesity Prediction Development of remote wireless environmental conditions measurement, monitoring and recording device for metrological and other scientific applications Acknowledgements The authors would like to thank the Director, CSIR-National Physical Laboratory, New Delhi for providing the necessary facilities to carry out the above work. The authors are also thankful to their colleagues Mr. Raman Kumar Sharma and Mr. Lalit Kumar for their constant support. One of the authors, Mr. Rahul Kumar (CSIR-SRF), is thankful to the Council of Scientific and Industrial Research (CSIR), New Delhi, India, for providing financial support.