key: cord-0055945-1yptptgb authors: Alaifan, Taghreed; Altamimi, Asmaa; Obeid, Dalia; Alshehri, Turki; Almatrrouk, Shaihana; Albarrag, Ahmed title: SARS-CoV-2 direct real-time polymerase chain reaction testing in laboratories with shortage challenges date: 2021-02-03 journal: nan DOI: 10.2217/fvl-2020-0187 sha: d1b549ea3b73d431a5cb536bee6ca18e29c72ffb doc_id: 55945 cord_uid: 1yptptgb Aim: In our study, we propose the use of direct real-time polymerase chain reaction (RT-PCR), this test does not require extraction or a preheating step, which saves a lot of cost, labor, processing time and provides a solution for supply shortage. Materials & methods: We assayed 185 nasopharyngeal samples stored in viral transport media. The indirect method was done with RNA extraction step, and the direct RT-PCR was done without an extraction step, both assays were evaluated on a commercially validated severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) kit targeting E gene. Results: Our assay showed a sensitivity of 79%, a specificity of 100% and the agreement between methods was 72%. Conclusion: Overall, this simple direct RT-PCR approach can be utilized as a qualitative diagnostic tool for emergency SARS-CoV-2 surveillance in countries with limited resources and may help laboratories to continue testing and at greater frequency despite supply shortages, although with delay in cycle threshold value in comparison with indirect RT-PCR. time has been proposed the use of the loop-mediated isothermal amplification method [11] . Thus, focusing on improving the workflow of RT-PCR is very essential as it ensures high sensitivity and specificity in clinical settings. One of the most successful attempts in improving diagnostic testing was reported in South Korea. South Korea increased its testing capacity by recruiting domestic companies to provide testing kits, which have shortened the supply leading time drastically [12] . This approach is applicable in countries with biomedical resources and specialized factories, however, many countries either don't have these advancements in the biomedical field or cannot afford the cost. Direct PCR amplification from samples is a technique that could be useful in the current phase of the pandemic. Direct amplification can tremendously shorten (TAT) time and reduce testing costs by eliminating the extraction phase of the samples handling workflow. Moreover, this kind of method reduces the sample handling time which will decrease the probability of contamination across the samples. This method has shown previously promising results in bacterial culture, human cells and plants [13, 14] . Considering the current pandemic, the use of direct RT-PCR offers an interesting solution for laboratories with limited supplies and reagents. In our current study, we provide an alternative method that provides diagnostic results without the need for the RNA extraction step or any additives. This protocol is useful in laboratories with limited resources and infrequent supplies. The direct RT-PCR uses samples stored in VTM directly into the RT-PCR reaction. In our study we evaluated our assay, by comparing it to the nationally validated method in Saudi Arabia, to determine its sensitivity and specificity. Overall, we aim to deliver a cost-effective method for laboratories with supply challenges and that are overwhelmed with mass population samples. Specimens were selected from the diagnostic laboratories at the Saudi Center for Diseases Control and Prevention (SCDC). 185 clinical nasopharyngeal swabs (124 COVID-19 positives and 64 negatives) stored at -80 • C in 2 ml of VTM (Vicell, Spain) were tested by both the indirect RT-PCR method and the direct RT-PCR method. Samples included in our study were collected from regional laboratories across Saudi Arabia. To draw a fair evaluation for our methodology, we choose samples with variation in their cycle threshold (Ct) values. Initially, we aimed to evaluate the appropriate sample volume input in the RT-PCR and the effect of the heating step on the direct technique. Nine samples with Ct values ranged from 18 to 28 selected randomly. A total of 3, 5 and 10 μl of each sample were added respectively to 10 μl of master mix. Also, 40 previously positive samples with different Ct values ranging from 18 to 36 were selected randomly. Aliquots of 100 μl from each VTM were heated in a water bath at 60 • C for 15 min. Then a 3 μl of heat-treated samples and a 3 μl without heating were subjected directly to RT-PCR. For the indirect RT-PCR test, RNA was extracted on the 96 MagNA Pure instrument (Roche, Mannheim, Germany) using the MagNA Pure 96 DNA and Viral NA Small volume isolation Kit. The extraction protocol was done following the manufacture protocol using 200-μl sample input and producing 50-μl of RNA output. The RNA was then used in one-step RT-PCR, using LightMix R Modular RNA master mix (Roche, Cat. No. 06754155001) with E gene LightMix R as a target (Roche, Cat:53077696) to detect SARS-CoV-2 virus according to the manufacturer instructions. Briefly, a final reaction volume of 15 μl containing; 5.5 μl of DNAse/RNAse free Water, 0.5 μl of Reagent mix including probes and primers, 4.0 μl of Roche Master Mix, 0.1 μl of RT enzymes and 10 μl of RNA template (samples or control). The RT-PCR reaction was performed on the Light Cycler 480 II instrument (Roche) following the recommended cycling conditions: reverse transcription at 55 • C for 5 min and denatured at 95 • C for 5 min, followed by 45 cycles of PCR at 95 • C for 5 s, 60 • C for 15 s and 72 • C for 15 s, and then cooled at for 40 • C for 30 s. Samples and control templates (positive and negative) were run in triplicate wells, the Ct value reported was the average of the three wells. A Ct value above 40 was considered a negative result. Direct RT-PCR For our developed protocol, samples that were stored in VTM were vortexed for a minute and 3 μl were then directly used into the RT-PCR LightMix R reaction at the same conditions stated in the indirect RT-PCR method using the LightMix R gene assay. For statistical analysis, data were collected and analyzed using SAS software, version 9.4 and Prism, Version 8.4. Sensitivity and specificity statistical measures were used to compare the indirect RT-PCR results with the direct RT-PCR assay. Tables for the qualitative results were analyzed by agreement confidence intervals, McNamara's test and T Paired test. Boxplots and correlation graphs were used to show the distribution of Ct values between these two methods. All p-values reported are two-sided and were considered to be statistically significant at alpha <0.05. The results of the direct RT-PCR compared with the indirect RT-PCR method as a qualitative test is summarized in Table 1 . Overall the agreement between the direct method and the indirect method was 72% (p < 0.0001), which is considered a moderate-high agreement. The developed direct RT-PCR method reported a sensitivity of 79.34% and specificity of 100%, furthermore, the probability of the negative likelihood was estimated to be 0.20. In the case of the Ct obtained by both methods, quantitative analysis showed late Ct values detected in the direct RT-PCR method compared with the indirect RT-PCR method. The Ct values distribution by methods of detection are shown in Figure 1 , overall lower Ct values were shown in the indirect RT-PCR, which is expected, and the change between the methods was significant (t = 5.93; p < 0.0001). The median Ct value reported in the indirect RT-PCR method was 25 and the median Ct value for the direct RT-PCR method was around 30. The mean of the change obtained in the form of Ct values was 5, which is considered high with 95% confidence intervals of (3.4-6.74). The levels of the Ct values by classification show a lower agreement as shown in Table 2 . The correlation for the cycle threshold for the samples tested by both methods is shown in Figure 2 . The correlation between both methods by Ct values shows a moderate correlation with R 2 = 0.755 (p < 0.0001). This correlation was shown to be the strongest for low Ct value positive samples. For the pilot experiment, adding 3 μl of nine samples showed a better S-shape and steep-slope amplification curve, which are indications of a healthy reaction. The presence of sample inhibitors impedes the RT-PCR reaction and interferes with efficiency, therefore, results in a lower curve slope [15] , and Ct value although the difference in mean Ct value reported was not significant (Supplementary Materials 1.1) . For the heating procedure, 17 samples out of 40 COVID-19 samples showed no significant difference either on Ct value or amplification curve shape. A total 23 sample results showed no amplification curve since their Ct values were >33, therefore we proceeded without the heating step ( Supplementary Materials 1.2 ). Overall, our study indicates that the use of direct RT-PCR provides concordance testing results compared with the indirect RT-PCR method. The developed detection assay had a 79% sensitivity and 100% specificity. These results indicate that direct RT-PCR can be a reliable method in difficult times when the extraction step is not applicable. However, the test showed high false-negative rates in samples with late Ct values. Lower Ct values were reported in the indirect method, which is expected, and the change in Ct values was significant between these two methods (t = 5.93; p < 0.0001). The median Ct value reported by the indirect method was 25 and the median Ct value for the direct method was 30. The Ct values distribution by the two methods shows that in the early Ct value positive samples high agreement was detected, however in samples with late Ct most of the results were negative. Furthermore, in positive samples, a moderate correlation was observed which indicates that although Ct values were significantly different, the two assays were correlated together. The direct RT-PCR method provided good specificity, but its sensitivity is still lacking, however, in times of need, the usage of this test can be useful in terms of TAT and emergency conditions such as lack of supply and pandemic high demand for testing. Direct RT-PCR was validated in a preprint during the COVID-19 pandemic [16] . Investigators reported a high sensitivity of 93% and used samples without VTM, in our study we got less sensitivity which could indicate a PCR inhibition. Another study from Washington University also used a similar method starting with preheating the samples before the RT-PCR, their assay was able to detect 92% of the positive samples [17] . Both studies reported higher sensitivity results for the indirect method than ours; however, both studies used samples with lower Ct values. In our study, sample Ct values were ranged from 14 to 35 to avoid bias. High CT (>33) samples mostly give negative results with direct RT-PCR, which affects the sensitivity value significantly. The Ct value gives an estimate of the amount of target nucleic acid and correlates inversely with viral load, a high Ct value indicates low viral load and vice versa [18] . Higher volume inputs did not improve sensitivity, as with direct RT-PCR the higher the VTM volume the higher the chance of introducing inhibitors as found also in another study [19] . Furthermore, the use of commercial VTMs affects SARS-CoV-2 and Influenza PCR detection as shown in one study [20] . In our direct method, we introduced the VTM components directly into the master mix, which may partially inhibit or interfere with the RT-PCR and cause a delay in Ct values especially samples with low viral load (Ct > 33), which have affected our assay sensitivity. The direct RT-PCR in the application of DNA viral detection has shown promising results in hepatitis B and Mycoplasma organisms [21, 22] . Other studies using direct RT-PCR on detecting viral RNA suggested that the use of PCR enhancers could lead to early RNAase release, which compromises the RNA integrity [23] . PCR inhibition is the leading cause of reporting low testing sensitivity in various fluids, in which some are reported as resistant to heat and demand the extraction step with sample dilution to eliminate the inhibition [24, 25] . The main advantages of using this assay are: shortening the diagnostic test time from 2.5 to 1.5 h, increasing the cost-effectiveness, minimizing hands-on labor, which decreases the probability of contamination, and coping with extraction reagent supply challenges [26] . The importance of our work rises in the time of limited supplies, however, the direct assay can be improved by using adjusted lysis reagent and validating the different types of VTMs as they play an important role in introducing PCR inhibitors. In another recent study to assess direct RT-PCR, they found that by using specific types of VTM and controlled inactivation of the virus, the direct RT-PCR is a viable option [18] . For future work, we aim to assess the different types of VTMs and lysis buffers to improve the sensitivity of our assay. The main limitation of our study is shortage of resources, because of the current time it is hard for many laboratories worldwide to research diagnostic testing. Resources and labors were scarce with high priority for diagnostic testing. The majority of the samples received at the public health laboratories were mostly positive, which could potentially bias the results, but we pooled a sufficient number of negative samples to use in our study. Another limitation of our study is that we used a single gene target to detect the virus, which could have played a role in our high false-negative rate. We were able to perform and evaluate a direct and simple RT-PCR test that does not require extraction or step during the SARS-CoV-2 pandemic crisis. The direct RT-PCR approach delivered sufficient results with good concordance with the gold indirect test. Direct RT-PCR could be a valuable solution to identify COVID-19 suspects, although with a lower Ct value, in a shorter time. Therefore, the indirect RT-PCR test could be reserved for individuals with an initial negative result. Our finding provides a simple solution to overcome the RNA extraction supply shortage. • In our study, we proposed the use of direct real-time polymerase chain reaction (RT-PCR), this test does not require extraction or preheating step, which saves a lot of cost, labor, processing time and provides a solution for supply shortage. • The direct RT-PCR assay was validated by comparing it to the nationally validated method in Saudi Arabia, to determine its sensitivity and specificity. • A total of 185 nasopharyngeal samples stored in validate viral transport media were used for the validation. Both of the indirect RT-PCR protocol, which is based on RNA extraction, and the direct RT-PCR were evaluated on commercially validated SARS-CoV-2 kit targeting E gene. • Overall, the agreement between the direct method and the indirect method was 72% (p < 0.0001), which is considered a moderate agreement. • The developed direct RT-PCR method reported a sensitivity of 79.34% and specificity of 100%. • The test showed high false-negative rates in samples with late cycle threshold (Ct) values. Lower Ct values were reported in the indirect method, which is expected, and the change in Ct values was significant between these two methods (t = 5.93, p < 0.0001). • The correlation between both methods by Ct values shows a moderate correlation with R 2 = 0.755 (p < 0.0001). • The main limitations of our study were resources-shortage, sample selection and RT-PCR target. • Future work should be done to improve the assay sensitivity and validate viral transport media that provide the best results, thus limit the effect of PCR inhibitors. • This direct RT-PCR approach can be utilized as a qualitative diagnostic tool for emergency SARS-CoV-2 surveillance in countries with limited resources. To view the supplementary data that accompany this paper please visit the journal website at: www.futuremedicine.com/doi/sup pl/10.2217/fvl-2020-0187 Author contributions T Alaifan is the principal investigator of this study. T Alaifan, A Altamimi and T Alshehri are laboratory technicians who ran the molecular assays. D Obeid and A Altamimi analyzed and interpreted. D Obeid and T Alaifan wrote the manuscript. A Albarrag and S Almatrrouk reviewed and edited the manuscript. All authors helped in writing and reviewing the manuscript. In addition, all authors read and approved the final manuscript. Financial & competing interests disclosure COVID-19) Situation Reports A pneumonia outbreak associated with a new coronavirus of probable bat origin Probable pangolin origin of SARS-CoV-2 associated with the COVID-19 outbreak Effects of vaccine program against pandemic influenza A (H1N1) virus, United States If a coronavirus vaccine arrives, can the world make enough? Centers for Disease Control and Prevention. Interim guidance for antigen testing for SARS-CoV-2. Updated December Interim Guidelines for Collecting, Handling, and Testing Clinical Specimens from Persons for Coronavirus Disease 2019 (COVID-19) Test sensitivity is secondary to frequency and turnaround time for COVID-19 screening Effects of COVID-19 on the laboratory turn-around time of vaccine-preventable disease surveillance: the case of measles in South Sudan Rapid point-of-care testing for SARS-CoV-2 in a community screening setting shows low sensitivity Overcoming the bottleneck to widespread testing: a rapid review of nucleic acid testing approaches for COVID-19 detection Testing on the Move South Korea's rapid response to the COVID-19 pandemic Universal direct PCR amplification system: a time-and cost-effective tool for high-throughput applications An evaluation of direct PCR amplification Interpretation of qPCR curve shapes Validation of an extraction-free RT-PCR protocol for detection of SARS-CoV2 RNA Direct RT-qPCR detection of SARS-CoV-2 RNA from patient nasopharyngeal swabs without an RNA extraction step No correlation between Ct values and severity of disease or mortality in patients with COVID 19 disease Massive and rapid COVID-19 testing is feasible by extraction-free SARS-CoV-2 RT-PCR The impact of viral transport media on PCR assay results for the detection of nucleic acid from SARS-CoV-2 Direct qPCR is a sensitive approach to detect Mycoplasma contamination in U937 cell cultures Rapid quantification of hepatitis B virus DNA by direct real-time PCR from serum without DNA extraction Direct DNA and RNA detection from large volumes of whole human blood Inhibition of PCR by aqueous and vitreous fluids An internal control for routine diagnostic PCR: design, properties, and effect on clinical performance Comparative evaluation of commercially available manual and automated nucleic acid extraction methods for rotavirus RNA detection in stools The authors have no relevant affiliations or financial involvement with any organization or entity with a financial interest in or financial conflict with the subject matter or materials discussed in the manuscript. This includes employment, consultancies, honoraria, stock ownership or options, expert testimony, grants or patents received or pending, or royalties.No writing assistance was utilized in the production of this manuscript. Ethical approval for this study was obtained from King Fahad Medical City Institutional Review Board and REC Research Ethics Committee (APPROVAL NUMBER: 20-326E).