key: cord-0054109-kni7fu2h authors: Cao, Yuan; Liu, Qiang; Wang, Ruifeng; Liu, Kun; Chen, Weidong; Wang, Guishi; Gao, Xiaoming title: Development of a 443 nm diode laser-based differential photoacoustic spectrometer for simultaneous measurements of aerosol absorption and NO(2) date: 2020-12-09 journal: Photoacoustics DOI: 10.1016/j.pacs.2020.100229 sha: eb4487e73d2721a5f0efa423ccba2099246b39a5 doc_id: 54109 cord_uid: kni7fu2h Measurement of ambient aerosol and nitrogen dioxide (NO(2)) is important as they are major pollutants from the burning of fossil fuel and biomass. In the present work, a differential photoacoustic spectrometer (D-PAS) was developed for simultaneous, online measurements of aerosol optical absorption and NO(2) concentration. A novel photoacoustic resonator was designed and employed in the D-PAS for controlling a large flow rate, improving response time, and keeping the flow noise at a low level. The detection limits of 1.0 Mm(−1) and 0.87 ppb for aerosol absorption and NO(2) concentration measurements were achieved with a lock-in amplifier time constant of 1 s. The D-PAS accuracy was demonstrated by performing a long-time, continuous measurement of aerosol, and NO(2) in ambient air. The measured results of NO(2) are consistent with the NOx analyzer and environmental monitoring station results. Aerosols are one of the important pollutants in the atmosphere, and their existence impacts atmospheric visibility, human health, the hydrological cycle, and radiative forcing [1] [2] [3] . Atmospheric aerosols directly affect radiative forcing by absorbing and scattering solar radiation, the radiative forcing of the total aerosols in the atmosphere is − 0.9 [− 1.9 to − 0.1] W m − 2 . Compared with greenhouse gases, the impact of aerosols on radiative forcing has greater uncertainty [4] . The measurement of aerosol light absorption characteristics is of great significance for the estimation of radiative forcing. NO 2 plays an important role in complex atmospheric chemical reactions, which affect air quality and climate [5] . For example, NO 2 is the main precursor of tropospheric ozone formation and participates in the formation of secondary organic aerosols [5] [6] [7] . An evidence episode was observed in a field observations of haze event in Beijing that SO 2 was rapid oxidated by NO 2 [8] . The main source of NO 2 is from anthropogenic combustion processes, such as automobile exhaust, power plants, and industrial emissions [5, 9] . In recent years, an increase in atmospheric NO 2 has been observed in South Asia, East Asia, and the Middle East [7] . Therefore, simultaneous measurements of aerosol absorption and NO 2 in the atmosphere are of great significance to the comprehensive management of ambient air quality and climate research [10] . There are few instruments capable of measuring aerosol absorption and NO 2 simultaneously. The most common method for measuring the aerosol absorption coefficient is filter-based technology using instruments such as the Aethalometer, the particle soot absorption photometer, and the multi-angle absorption photometer [11] . Although these instruments are widely employed, they are limited by factors associated with filter-loading, extraction, and multiple scattering effects as the optical properties are not measured on natural suspended status [1, 12, 13] . The concentration of NO 2 is usually measured by chemiluminescence-based techniques, but it is cross-sensitive to other nitrogen compounds such as peroxyacetyl nitrate and nitric acid [14] . Simultaneous measurement of the absorption characteristics of aerosols and the concentration of NO 2 helps monitor the dynamics of the two main pollutants in the atmosphere in real-time. In addition, the simultaneous measurement of aerosol absorption and NO 2 helps to analyze the influence of atmospheric oxidation capacity on aerosol absorption [15] . Photoacoustic spectroscopy (PAS), discovered by A. G. Bell in 1880 [16] , is a sensitive "zero baseline" method for weak absorption measurement [17] [18] [19] [20] and is regarded as the best candidate for in situ aerosol absorption measurement. PAS is based on the detection of sound waves that are generated due to the absorption of modulated optical radiation related to the photoacoustic effect. PAS can directly measure the absorption of suspended particles; this effectively avoids the problems related to the filter-based method. The aerosol light absorption coefficient can be obtained with a well-calibrated PAS spectrometer. The calibration can be usually performed with gas (ozone [21] , nitrogen dioxide [22] , oxygen A-band [23, 24] ) or aerosols [25] in order to determine the photoacoustic cell constant. Due to the high sensitivity and accuracy of the PAS spectrometer, it gradually came to be used to measure the absorption characteristics of aerosols [12] . Wiegand et al. reported a UV-vis photoacoustic spectrophotometer for the measurement of gas-phase and aerosol absorption over the UV-vis region [13] . The photoacoustic spectrophotometer can measure the absorption coefficient of aerosols in eight bands. Bluvshtein et al. used a multi-pass PAS spectrometer, calibrated with nigrosine particles, to directly measure the light absorption of atmosphere aerosol [25] . Cremer et al. reported PAS measurements of an optically trapped single droplet, which can be used for the direct observation of nanofocusing in aerosol photokinetics [26] . Although there are many reports on the use of PAS to measure aerosols absorption and trace gases [27] [28] [29] [30] [31] [32] , to our knowledge, there are few photoacoustic spectrometers with sufficiently high sensitivity that can simultaneously measure the absorption characteristics of atmospheric aerosols and the concentration of atmospheric NO 2 . In addition, when PAS is used for atmospheric measurements, the sampling flow rate usually needs to be less than 0.5 L/min (for example, the sampling flow rate in reference [11] is 200 mL/min, the sampling flow rate in reference [12] is 330 mL/min, and the sampling flow rate in reference [27] is 400 mL/min). A low flow rate will affect the PAS sensor's response time, but a high flow rate creates turbulence in the photoacoustic cell, resulting in considerable noise affecting the performance of the PAS sensor [11] . The development of a PAS spectrometer that can simultaneously measure the aerosol absorption and NO 2 concentration in the atmosphere at large flow rates will quickly respond to changes in the measurement environment. In this work, the development of a differential photoacoustic spectrometer (D-PAS) sensor based on a 443 nm blue laser diode (LD) for real-time, online, in situ measurement of aerosol optical absorption and NO 2 was developed and demonstrated. Two acoustic resonators with the same resonant frequency were carefully designed. To increase the total flow rate to improve the response time of the D-PAS sensor, four additional airways were arranged in the acoustic resonator, which enables the D-PAS to work with a large flow rate (>1 L/min) without being disturbed by flow noise. The D-PAS developed was used to simultaneously measure the aerosol absorption coefficient and NO 2 concentration in the atmospheric environment continuously. The measured results of NO 2 were compared with the NOx analyzer and the environmental monitoring station, and good agreement between them was observed. A blue-LD source (PL-TB450B, 1.6 W) was used for excitation of the PAS signal. The LD emission spectrum was measured using a spectrometer (HR2000), and the actual central wavelength was determined to be 443 nm. The emission spectrum of the LD is a Gaussian distribution, and the full width at half maximum (FWHM) was determined to be 1.6 nm. An LD controller (ITC110, Thorlabs) was used to modulate the current and temperature. A thermoelectric cooler (TEC) was inserted between the LD and heat sink [33] . The LD's temperature was measured using a 10 KΩ thermistor (MEB 10 K NTC) controlled at approximately 25℃, which helps to achieve long-term stability of the laser output power. The laser beam was collimated using a lens with a radius of 6 mm and a focal length of 6 mm. According to the datasheet, the used LD output power can reach to 1.6 W when it working at maximum current (1.6 A). However, for LD safety, working current of the LD was not set at maximum, but set at 1 A during the experiment. In such case the corresponding laser output power was measured to be 600 mW. An optical chopper (MC2000B-EC, Thorlabs) was used to modulate the light at the resonant frequency of the acoustic resonators. The photoacoustic cells and optical chopper were "soft" mounted to the bottom plate via a rubber sheet for decreasing chopper vibration noise [34] . The PAS signals were detected with a microphone (BSWA, MP201, 50 mV/Pa) and then demodulated using two lock-in amplifiers (LIA-BVD-150-L) with 1 s time constant. The PAS signals were digitally sampled by a data acquisition card (DAQ, NI-USB-6212) and then displayed on a laptop via LabVIEW (National Instruments) software. A power meter (NOVA II, OPHIR) was placed behind the PAS spectrometer to monitor the laser power. In the following sections, the PAS signals were normalized with respect to laser power to eliminate signal fluctuation due to laser power jitter in long-term measurements. A temperature and humidity sensor (TH22R-EX) was connected to the D-PAS system to measure the temperature and relative humidity (RH) of the sample. A LabVIEW program was written to record laser power, RH, temperature, and PAS signals simultaneously. Acoustic resonators are widely used in the PAS spectrometer to amplify PAS signals [35] . Thus, an acoustic resonator is a critical component for a high-performance PAS spectrometer. In the present work, the acoustic resonator was designed and optimized using finite element method (FEM) simulation. The material used in the finite element model is air, and the air density value comes from the material library of the simulation software, and its value is 1.205 kg/m 3 . The temperature and pressure set in the finite element model are 293.15 K and 1 atm respectively. Using the numerical calculations, the acoustic resonator's basic characteristics, such as eigenfrequencies and Q FEM Q FEM -factors, can be predicted and will provide a reference to optimize the structure of photoacoustic cell [36, 37] . In addition, the photoacoustic signal can also be calculated by the finite element method [36] . In order to design a suitable photoacoustic cell for aerosol measurement, the impact of the acoustic resonator size (radius and length) on the PAS spectrometer performance (such as frequency and photoacoustic signal) was simulated. The simulated results are shown in Fig. 2 . Fig. 2 (a) plots the resonant frequency of the longitudinal (100) mode of an acoustic resonator with different inner radii and lengths. Fig. 2 (b) plots the Q FEM -factor of the longitudinal (100) mode of an acoustic resonator with different inner radii and lengths. Fig. 2 (c) plots the simulated PAS signals of the longitudinal (100) mode of an acoustic resonator with different inner radii and lengths. As shown in Fig. 2 (a) , the photoacoustic cell's resonant frequencies decrease monotonically as the resonator length increases, while there is no apparent relationship between the resonant frequencies and the resonator radius. Secondly, Fig. 2(b) shows that the Q FEM -factors of the photoacoustic cell decrease as the length of the acoustic resonator increases, and increase as the radius of the acoustic resonator increases. Fig. 2 (c) shows the PAS signal increasing with the increasing of the resonator length and decreasing with an increase in the resonator's inner radius. This is consistent with the theory that the PAS signal is inversely proportional to the effective cross-section V L (V and L are the acoustic resonator's volume and length, respectively) [35] . Hence, it is easier to improve the detection sensitivity of an acoustic resonator with a small radius. Considering the beam size of the LD employed and the 1/f f noise, an acoustic resonator with a radius of 5 mm and a length of 110 mm (corresponding to a resonant frequency of 1474 Hz and a Q FEM -factor of 60) was selected. By choosing these parameters, the laser will not irradiate the inner wall of the resonator, and thus will not produce a background signal. Simultaneously, the acoustic resonator has a high resonance frequency, which avoids the interference of low-frequency noise. Two buffers with a radius of 35 mm and a length of 55 mm are connected to the both sides of the resonator (Fig. 3) . The buffers act as a soft acoustic field boundary, and the PAS signal forms a pressure node at the interface between the resonator and the buffer. The buffers can effectively filter the noise resulting from light absorbed by the window [35] , flow noise, and environment noise. Fig. 4 shows the normalized simulation PAS signal distribution along the optical axis of (100) mode of the designed photoacoustic cell. The microphone is usually placed in the middle of the resonator, which corresponds to the sound pressure's antinode. It can be seen in Fig. 4 that the PAS signal gradually decreases to zero in the buffer area, which is caused by the end correction of the acoustic resonator. This observation can be explained by the effects of a mismatch between the one-dimensional sound field in the resonator and the three-dimensional sound field in the buffers [35] . Fig. 5 shows the normalized D-PAS signal as a function of the frequency. The experimental data were fitted to a Lorentz contour, which describes the power in a classical driven oscillator as a function of frequency. The optimum resonant frequency of the photoacoustic cell is 1430 Hz. This corresponds to the first-order longitudinal resonance frequency of the photoacoustic cell. This frequency will be used in the following measurements as the amplitude modulation frequency of the laser diode. The error between the resonance frequency measured by the experiment and the resonance frequency calculated by the finite element simulation is 3%. This result shows that the finite element method is a reliable method to calculate the resonance frequency of the photoacoustic cell. The calculated Q factors of photoacoustic cell 1 and photoacoustic cell 2 are 37 and 36, respectively. When using a PAS spectrometer for actual atmospheric detection, the sampling flow rate should generally not be too high. Otherwise, the photoacoustic cell will generate considerable flow noise, which will affect the accuracy of the measurements (for example, the sampling flow rate in reference [11] is 200 mL/min, the sampling flow rate in reference [12] is 330 mL/min, and the sampling flow rate in reference [27] is 400 mL/min). To increase the total flow rate to improve the time response of the D-PAS sensor, four additional airways were arranged in the acoustic resonator, as shown in Fig. 6 . The radius of the airways was 4 mm, such that the particle deposition loss in the sampling lines can be neglected [38] . Fig. 6(a) and (b) are simulations of the flow velocity field in the photoacoustic cell without and with additional airways, as indicated by arrows. Although the flow rate at the inlet of each photoacoustic cell is the same, the total aerosol flow is shunted by the additional airways, which reduce the flow rate in the resonator. After these four additional airways in the acoustic resonator have been inserted, the photoacoustic system can be operated with high flow rates without being affected by the flow noise, while it also reduces the sample residence time and hence improves the time response of the D-PAS. Fig. 7 (a) shows the noise level measured in the D-PAS system when operated at different flow rates with and without four additional airways. The experimental results show that in the acoustic resonator with additional airways, the flow noise will not increase substantially until the flow rate exceeds 1400 sccm. However, in the acoustic resonator without additional airways, the flow noise was substantially increased when the flow rate exceeded 600 sccm. Considering that the photoacoustic cell with four additional airways can work at higher flow rates without being disturbed by flow noise, this improved acoustic resonator is used in the actual atmospheric measurement in this work. The sample flow rate was set to 1000 sccm. Fig. 7 (b) is an investigation of the sensor response speed at a gas flow rate of 1000 sccm. High purity N 2 and 2 ppm NO 2 were periodically passed through the photoacoustic cell. Since NO 2 will undergo adsorption and desorption processes after entering the photoacoustic cell, it takes a long time to reach equilibrium in the photoacoustic cell. Before using the developed D-PAS sensor to measure the aerosol optical absorption coefficient, NO 2 was used to calibrate the PAS system to obtain the product of microphone sensitivity and cell constant (Eq. 4, namely, M•C). Different concentrations of NO 2 were obtained by diluting 5 ppm of NO 2 /N 2 (Nanjing Special Gases) with high purity N 2 (≥ 99.999 % ≥ 99.999%, Nanjing Special Gases). The accurate NO 2 concentrations were measured by the NOx gas analyzer (Model 42i). An effective absorption cross-section σ(λ) within the broadband LD emission region is calculated [39] : where σ(λ i ) (cm 2 /molecule) represents the absorption cross-section of NO 2 obtained from the HITRAN database [40] , and g(λ i ) is the normalized optical power density distribution function. Based on Eqs. (1) and (2), an effective absorption cross-section of 4.21 × 10 − 19 cm 2 /molecule over the full laser emission spectral region was calculated. The absorption coefficient of NO 2 can be calculated with the following equation [39, 41] : where N L = 2.479 × 10 19 molecule•cm − 3 •atm − 1 , T (K) and P 1 (atm) are the temperature and pressure of the NO 2 sample, respectively, and c is the concentration of NO 2 . The value of α (Mm − 1 ) converted into the NO 2 mixing ratio (ppb) is 1.04 Mm − 1 /ppb. The calibration will be performed based on the following equation: where S (mV) is the photoacoustic signal with background removed, α (Mm − 1 ) is the corresponding NO 2 absorption coefficient, P 2 (mW) is the LD power, M (mV/Pa) is the microphone sensitivity, and C (Pa/ (mW•Mm − 1 )) is the cell constant. The calibration results are displayed in Fig. 8 . The upper and lower abscissas represent the concentrations and the absorption coefficient of NO 2 , respectively, and the ordinate is the PAS signal after power normalization. The difference between the calibration results of the two photoacoustic cells is related to machining errors and used microphone sensitivity of M (mV/Pa) of each cell, which cause the cell constants of the two photoacoustic cells to be different. From Fig. 8 , it is obvious that the photoacoustic cell 2 offer larger signal value than photoacoustic cell 1 for the same absorption. The detection limits and measurement precision of the developed D-PAS were evaluated using a time series measurement of N 2 filled in the photoacoustic cell [42, 43] . The time resolution of each measurement was 1 s. Fig. 9(a), (b) plots the detection limits from photoacoustic cell 1 and photoacoustic cell 2, respectively. Allan deviation analyses provide detection limits of about 1.0 Mm − 1 and 0.9 Mm -1 (0.87 ppb) within 1 s for the two photoacoustic cells, respectively. These detection limits could be further improved down to 0.14 Mm − 1 and 0.09 Mm − 1 (0.09 ppb) when using integration times up to 150 s and 480 s, respectively. Histogram plot of the measured data is show in Fig. 9(c), (d) , which can be used to evaluate the measurement precision of the D-PAS sensor. The data distributions were Gaussian profile, and the measurement precision is the half width at half maximum (HWHM) of the Gaussian fitting curve. For a 1 s integration time, the measurement precision of photoacoustic cells 1 and 2 are 1.7 Mm − 1 and 1 Mm − 1 , respectively. The difference of Allan deviation between photoacoustic cell 1 and cell 2 is mainly caused by the cell constant. Photoacoustic cell 2 offer better performance than photoacoustic cell 1 as photoacoustic cell 2 offer higher cell constant than photoacoustic cell 1, as shown in Fig. 8 . These detection limits could allow the developed D-PAS to measure ambient aerosol and NO 2 in real-time. Potential spectral interferences for D-PAS to measure aerosols and NO 2 at 443 nm, mainly resulted from the molecules (NO 3 , O 3 , N 2 O 5 , O 4 ) [44] are discussed below. (1) The absorption cross-section of NO 3 is about 3.82 × 10 − 19 cm 2 / molecule at 443 nm, and its concentration is about 1-200 pptv in the atmosphere [45] ; accordingly, the absorption coefficient can be estimated as 9.41 × 10 -4 -1.88 × 10 -1 Mm -1 . (2) O 3 is a major pollutant in the atmosphere, and its concentration ranges from 10 to 500 ppbv [45] . Given its absorption cross-section of 1.82 × 10 − 22 cm 2 /molecule at 443 nm, its absorption coefficient is about 4.48 × 10 -3 -2.24 × 10 -1 Mm -1 . (3) The absorption cross-section of N 2 O 5 for λ > 420 nm is missed in the database. Therefore, its absorption cross-section at 420 nm is used to estimate its absorption coefficients to be between 2.46 × 10 − 7 -7.39 × 10 -5 Mm -1 , with concentrations in the range of 2-600 pptv [46] . (4) O 4 bands are used in differential optical absorption spectroscopy (DOAS) to infer information about aerosols [47] . The absorption coefficient of O 4 of about 4.37 × 10 − 2 Mm -1 at 443 nm [47] was calculated using the following equation: where α O4 is the O 4 absorption coefficient, σ O4 (cm 5 ⋅molecule − 2 ) is the corresponding absorption cross-section, O 2mr is the oxygen mixing ratio (20.95 %v/v), N d is the number density. Table 1 summarizes the absorption coefficients of the trace gases (NO 3 , O 3 , N 2 O 5 and O 4 ) discussed above. Fig. 10 shows the upper absorption coefficients of these molecules. It is clearly shown that the absorption coefficients of these gases in the atmosphere are under the detection limit of the developed D-PAS sensor. Therefore, the interference from these gases can be neglected. Potential interference from water vapor absorption in the laser emission wavelength of 443 nm was also considered. For this purpose, Fig. 11 . As shown in Fig. 11 , the PAS signals remain almost unchanged when the water vapor concentration ranges from 0.38 % to 1.38 % (RH is between 20 % and 75 %, the temperature is about 16℃). Therefore, potential interference from water vapor absorption could be neglected for the developed D-PAS. Measurement of aerosol optical absorption and NO 2 concentration in the atmosphere was carried out at Anhui Institute of Optics and Fine Mechanics (AIOFM) in Hefei (China). During measurement, the NO 2 was also monitored by using a NOx analyzer for comparison. The sample inlet was about 2.5 m above ground level. The sampling location is between two roads with fewer vehicles, which is about 70 m from the front road and 20 m from the back road. Since the measurement location is close to urban and rural areas, the measurement data will be affected by straw burning in rural areas, vehicle exhaust emissions in urban areas, and industrial emissions. To measure the absorption coefficient of aerosols, another special issue should be considered. In high RH, the evaporation of water vapor will affect the photoacoustic signal [48] [49] [50] . Therefore, the RH in measuring the absorption characteristics should be less than 30 % to ensure measurement accuracy. Meanwhile, the resonant frequencies vary with the change of environment temperature and RH. Hence, a relatively stable temperature and RH are necessary for accurate atmospheric measurements. A diffusion drying tube was mounted on the sampling line to maintain the RH at about 20 %. Furthermore, the measurement process was carried out in an air-conditioned room with a temperature stabilized at 20 • C. Using the calibrated D-PAS, measurements of ambient aerosol optical absorption coefficients and NO 2 concentration were performed. Fig. 12 shows the time series measurement results of the aerosol absorption coefficient and NO 2 concentration in the winter of 2019. The measured data were averaged in 5-minute intervals. Fig. 12(a) shows the aerosol optical absorption coefficient measured by D-PAS. The NO 2 concentration monitored by a near environmental monitoring station, which was approximately 2 km from the laboratory, is also shown in Fig. 12 (b) for comparison. To further confirm the developed D-PAS's performance, comparison measurements of ambient NO 2 with the D-PAS and NOx analyzer were carried out during a clear day in the summer of 2020. The measured NO 2 by D-PAS shows good agreement with an independent NOx analyzer ( Fig. 13 (b) ). For convenience comparison, the NO 2 concentration recorded by the environmental monitoring station is also shown in Fig. 13 (b) . It can be seen from Fig. 13 (b) that the NO 2 concentration of the environmental monitoring station is higher than the results measured by the D-PAS and NOx analyzer, but the general trend is the same. This difference may be caused by the fact that the laboratory and the environmental monitoring station (~2 km away from laboratory) are not in the same location. In Fig. 13(a) , the aerosol absorption is weak, which agrees with the fact that there was less particulate matter in the air. The concentration of NO 2 gradually increased from 6:00 to 7:00. This fact may be related to the emissions of automobile exhaust. The aerosol absorption coefficient appeared at a broad peak from 2:00 to 3:00. This observation may be related to the atmospheric chemical reaction process at night. D-PAS will be further used in conjunction with a particle counter to monitor aerosols' light absorption characteristics and the concentration of NO 2 for a long time and analyze their changing trends. In this work, the development of a 443 nm LD-based D-PAS sensor is reported that is capable of measuring simultaneously ambient aerosol and NO 2 with a high flow rate. Minimum detection limits of 1.0 Mm − 1 (1 s) and 0.87 ppb (1 s) were obtained for aerosols (using photoacoustic cell 1) and NO 2 (using photoacoustic cell 2), respectively. In comparison, Haisch et al. developed a PAS spectrometer for artifact-free parallel measurements of soot and NO 2 in engine exhaust [14] . The instrument has a detection sensitivity of 0.3 ppm for NO 2 and a detection sensitivity of 0.54 μg m -3 (4 Mm -1 ) for soot. Sample flow noise was decreased by arranging four additional airways in the acoustic resonator that enable the D-PAS to work at high flow rates while improving the response time. Potential absorption interference analysis has shown that no other gases will affect the measurement results. Aerosol optical absorption and NO 2 concentration in the ambient air were measured by the D-PAS. The experimental results are consistent with the results of the NOx analyzer and environmental monitoring station. This fact demonstrates that the developed D-PAS offer good performance and has the capacity to measure atmosphere aerosol and NO 2 simultaneously. In the future, the D-PAS will combine with other reference instruments, such as the scanning mobility particle sizer (SMPS) and the cavity ring-down spectrometer, for more detailed atmospheric studies and intercomparison investigation. The authors declare that there are no conflicts of interest. photoacoustic spectroscopy and laser spectroscopy for application in atmospheric photochemistry and environmental science. Aerosol light absorption and its measurement: a review A new approach for retrieving the UV-vis optical properties of ambient aerosols Photoacoustic optical properties at UV, VIS, and near IR wavelengths for laboratory generated and winter time ambient urban aerosols Climate Change 2013: The Physical Science Basis. Contribution of Working Group I to the Fifth Assessment Report of the Intergovernmental Panel on Climate Change Development of a mid-infrared nitrogen dioxide sensor based on Faraday rotation spectroscopy Abrupt decline in tropospheric nitrogen dioxide over China after the outbreak of COVID-19 Abrupt recent trend changes in atmospheric nitrogen dioxide over the Middle East Fast sulfate formation from oxidation of SO 2 by NO 2 and HONO observed in Beijing haze Sub-ppb nitrogen dioxide detection with a large linear dynamic range by use of a differential photoacoustic cell and a 3.5 W blue multimode diode laser Towards low-cost QEPAS sensors for nitrogen dioxide detection A novel multiwavelength photoacoustic spectrometer for the measurement of the UV-vis-NIR spectral absorption coefficient of atmospheric aerosols A portable, four-wavelength, single-cell photoacoustic spectrometer for ambient aerosol absorption A UV-vis photoacoustic spectrophotometer Photoacoustic analyzer for the artifact-free parallel detection of soot and NO 2 in engine exhaust Effects of photochemical oxidation on the mixing state and light absorption of black carbon in the urban atmosphere of China On the production and reproduction of sound by light Broadband detection of methane and nitrous oxide using a distributed feedback quantum cascade laser array and quartz-enhanced photoacoustic sensing A novel photoacoustic spectroscopy gas sensor using a low cost polyvinylidene fluoride film Multi-resonator photoacoustic spectroscopy Photoacoustic spectroscopy for gas sensing: a comparison between piezoelectric and interferometric readout in custom quartz tuning forks Aerosol absorption measurement using photoacoustic spectroscopy: sensitivity, calibration, and uncertainty developments Nitrogen dioxide and kerosene-flame soot calibration of photoacoustic instruments for measurement of light absorption by aerosols Simultaneous photoacoustic spectroscopy of aerosol and oxygen A-band absorption for the calibration of aerosol light absorption measurements Standard photoacoustic spectrometer: model and validation using O 2 A-band spectra Calibration of a multi-pass photoacoustic spectrometer cell using light-absorbing aerosols Signorel, Photoacoustics of single lasertrapped nanodroplets for the direct observation of nanofocusing in aerosol photokinetics Aerosol absorption measurement at SWIR with water vapor interference using a differential photoacoustic spectrometer Photoacoustic and nephelometric spectroscopy of aerosol optical properties with a supercontinuum light source ppb-level SO 2 photoacoustic sensors with a suppressed absorptiondesorption effect by using a 7.41 μm external-cavity quantum cascade laser Beat frequency quartz-enhanced photoacoustic spectroscopy for fast and calibration-free continuous trace-gas monitoring Atmospheric CH 4 measurement near a landfill using an ICL based QEPAS sensor with V-T relaxation self-calibration QEPAS spectrophones: design, optimization, and performance Incoherent broadband cavity enhanced absorption spectroscopy for in situ measurements of NO 2 with a blue light emitting diode Aircraft instrument for comprehensive characterization of aerosol optical properties, part 2: black and brown carbon absorption and absorption enhancement measured with photo acoustic spectroscopy Application of acoustic resonators in photoacoustic trace gas analysis and metrology Finite element calculation of photoacoustic signals FEM analysis of cylindrical resonant photoacoustic cells Particle Loss Calculator -a new software tool for the assessment of the performance of aerosol inlet systems Application of a broadband blue laser diode to trace NO 2 detection using off-beam quartz-enhanced photoacoustic spectroscopy The HITRAN database Multi-wavelength measurements of aerosol optical absorption coefficients using a photoacoustic spectrometer Sensitive methane detection based on quartz-enhanced photoacoustic spectroscopy with a high-power diode laser and wavelet filtering Highly sensitive acetylene detection based on multi-pass retro-reflection-cavityenhanced photoacoustic spectroscopy and a fiber amplified diode laser Simultaneous monitoring of temporal profiles of NO 3 , NO 2 and O 3 by incoherent broadband cavity enhanced absorption spectroscopy for atmospheric applications Sensing atmospheric reactive species using light emitting diode by incoherent broadband cavity enhanced absorption spectroscopy An aircraft based three channel broadband cavity enhanced absorption spectrometer for simultaneous measurements of NO 3 , N 2 O 5 and NO 2 Temperature dependent absorption cross-sections of O 2 -O 2 collision pairs between 340 and 630 nm and at atmospherically relevant pressure Photoacoustic and filter-based ambient aerosol light absorption measurements: instrument comparisons and the role of relative humidity Limitations of the photoacoustic technique for aerosol absorption measurement at high relative humidity Evaporation-condensation effects on resonant photoacoustics of volatile aerosols This work was financially supported by the National Natural Science