key: cord-0047178-2tvw3y9h authors: Sur, Amitava; Paria, Anshuman title: Histogram analysis for bedside respiratory monitoring in not critically ill preterm neonates: a proposal for a new way to look at the monitoring data date: 2020-07-08 journal: Eur J Pediatr DOI: 10.1007/s00431-020-03732-2 sha: 6b1635e3f763d356405b747e993d02132c22e0c2 doc_id: 47178 cord_uid: 2tvw3y9h Despite robust evidence in favour of maintaining optimal oxygen saturation targets in the preterm infants, the titration of oxygen is largely dependent on manual observations and transcription. Similarly, notwithstanding the gaining popularity of non-invasive modalities like high-flow nasal therapy, the practices of weaning and escalating support are largely individualized and based on point of care observations. These are often erroneous and lack objectivity. Histogram analysis from patient monitors is an easy and objective way of quantifying vital parameters and their trends. We review the technology and evidence available behind this practice. Conclusions: Though there are no randomized controlled trials on this practice solely, we identify several quality improvement studies implementing this into practice with benefit. We also cite studies which have implemented histogram analysis in methodology, thus concluding that this is a useful clinical tool worth incorporating into clinical practice to reduce manual errors and bring more objectivity into decisions. The last decade in neonatology has seen a change in principles of respiratory care for the preterm infant. The two major shifts of practice have been in use of early noninvasive ventilation (NIV) as respiratory support and altered saturation targets for optimal oxygenation. While CPAP has been the most commonly used form of NIV in the past, high-flow nasal cannula (HFNC) has emerged as the preferred option for many neonatal units [1] . The major criticism around HFNC, in spite of perceived improved patient comfort, has been the uncertainty around consistency in delivering PEEP. Nasal intermittent positive pressure ventilation (NIPPV) has been reported to have improved rates of success in extubation when compared with CPAP [2] . The two main forms of NIPPV commonly used are bi-level NIPPV and ventilator-driven NIPPV (CMV), the latter having the advantage of being able to deliver higher peak inspiratory pressures and shorter inspiratory times [3] . Nasal high-frequency oscillatory ventilation (HFOV) and NIV-NAVA (neurally adjusted ventilatory assist) are other, albeit less commonly used, modalities of NIV. Nasal HFOV has been shown to have superior clearance of CO 2 and reduce risk of intubation in preterm infants in a recent meta-analysis when compared with common modes like CPAP or bi-level NIPPV [4] . NAVA improves synchrony in patients with respiratory support by detecting the electrical activity of the diaphragm and may offer potential benefits in neonatal ventilation. Again, it has been shown to be superior to CPAP in facilitating extubation in preterm infants in small studies [5] . So, it seems that a synchronised modality of consistent pressure delivery might be superior to conventional modalities of NIV, but their use across neonatal units is still limited. Irrespective of the modality, subsequent weaning of non-invasive support is not evidence based and requires streamlining based on physiology. Current guidelines are at best a compendium of expert consensus opinions [1] . Decision to wean is generally influenced by regimented weight-based approach, perceived FiO 2 requirement, snapshot counting of respiratory rate (RR) and clinically examined work of breathing. Measured PCO 2 in blood gas analysis is also often used to guide the weaning process but it is a poor physiological reflection of pressure requirement. Optimal saturation targeting for the preterm infants has been long-standing debate with widely variable practices. The data from the NeOProM meta-analysis has clearly demonstrated less risk of mortality and necrotizing enterocolitis when saturation targets are between 91 and 95% as compared with lower targets [6] . Preterm infants with evolving or established chronic lung disease are known to have swinging saturations; spending a sizable proportion of time outside saturation limits with cumulatively unacceptable periods of hypoxia and hyperoxia. While prolonged periods of hypoxia can contribute to increased risk of mortality and necrotising enterocolitis, hyperoxia has been associated with retinopathy of prematurity and chronic lung disease [6] . Sustained periods of hypoxia are also known to increase risk of pulmonary hypertension. Manual transcribing of saturations by nursing staff is still the gold standard for bedside monitoring in neonatal units but they tend to underestimate these fluctuations and time spent outside desirable range [7] and is also associated with considerable alarm fatigue. While closed loop oxygen feedback systems can alleviate some of these problems while on invasive ventilation, infants of non-invasive support continue to have such episodes. A recent study on use of automated oxygen control via IntellO2 has shown significantly improved control in saturation targets in babies on HFNC [8] , but again this is not used in wider practice and is still limited to research. In summary, although more preterm neonates are being supported with NIV, tight control of saturations and objective escalation and de-escalation are yet to be optimized. Data from electronic patient monitoring can provide an important missing link in this regard. Histograms are a graphical way of representing patient's measurements over an extended time period. It basically answers the question: "For how much of a certain period of time was my patient within or outside a certain range of values?" For respiratory monitoring of preterm neonates, we rely on a set of clinical parameters and observations to remain within prespecified ranges, e.g., saturation between 91 and 95%, and histogram analysis allows us to observe this. Generally, it is represented as a cluster of bars, with each bar representing time spent within a particular time interval. The picture below (Fig. 1a) is an example of a saturation histogram where the Xaxis represents the parameter range and Y-axis the distribution. Each blue bar represents the percentage of time spent in the range. Histograms can be used for both real-time data and trended data like heart rate and respiratory rate. Being a numerical and graphical representation, they allow us to easily appreciate effects of any intervention by studying a change in pattern. All standard pulse oximeters and monitors will have the algorithm to produce histograms, but in this piece we will focus on the Philips IntelliVue monitors. The inbuilt pulse oximetry module (M1020B #A01) uses a motion-tolerant signal processing algorithm, based on Fourier Artefact Suppression Technology (FAST). It provides four measurements: oxygen saturation of arterial blood (SpO 2 ), plethysmograph waveform, pulse rate and perfusion indicator. The histogram feature can be selected from the main screen of the monitor and it yields two types of data: real time and trended. The SpO 2 histogram is calculated over 12 or 24 h with a 1 s real-time sampling rate. For parameters like respiratory rate and heart rate, the data is built over similar time period but with a sampling rate of 1 or 5 min. For histogram analysis of any parameter, the bars in the foreground (blue) show the percentage of time the values fall into each unit on the scale. The bars in the background (shaded area) show the cumulative percentage-of-time value, which results from adding each of the foreground bars to the sum of those to its left. The vertical axis shows the percentage of time. The monitor can be customized so that maximum patient information is displayed on the same screen allowing the clinician to obtain a comprehensive picture of the variability and median values. This can be through the "3 histogram view" (Fig. 1b) which also displays the histogram analysis of saturation, RR and HR over a chosen time period thus providing a more objective idea of the trend. It is worthwhile noting the fourth parameter displayed on screen, although we will not discuss it in details-perfusion numeric (Perf). The accuracy of the readings for this oximeter is well validated for perfusion values between 0.3 and 1. Figure 2 depicts the real example of a patient with suboptimal histogram with 29% of the time in the last 12 h spent below 90% SPO 2 . The review of the technology obviously brings to front certain limitations. For example, the data will be influenced by sampling time and relative stability of the patient. For a critically ill infant with rapid and continuous change in parameters, a stable consistent histogram will not be available, and decisions based on that risk being erroneous. Also, other parameters like delivered mean airway pressure (MAP), (which can be impacted by the leak) and physiological changes like temperature and pain perception can alter respiratory and heart rate. Although monitoring based on PaO 2 is the gold standard for HRF, certain non-invasive trends have been used successfully like oxygen saturation index, respiratory severity score, SPO 2 /FiO 2 ratio (S/F) and transcutaneous monitoring of PO 2 (TcPO 2 ). Detailed review of these is beyond the scope of this article. But it is worth mentioning that monitoring based on SPO 2 values can often be fallacious as SPO 2 values are affected by various factors including fetal Hb. In a large ARDS network trial, S/F of 315 correlated well with PaO 2 /FiO 2 of 300 [9] . TcPO 2 has been recommended by various studies and its feasibility has been well established in adult studies. TcPO 2 might be an added advantage in infants without arterial access as its values can have good correlation with pAO 2 if the sensor is pre-warmed. A recent study which combined a new fluorescent quenching technique of monitoring PO 2 with pulse oximetry has shown good feasibility and minimal value drift [10] . Adding this to histogram data in patient monitors might bring more accuracy and less bias to monitoring. What is available in current literature on use of histogram Although it is a relatively recent addition to the clinical repertoire of preterm respiratory monitoring, several groups have already reported on the utility of using this. We conducted a search on medical databases namely Medline, Ovid and CINAHL with the Mesh terms (Histogram OR Histogram analysis OR histogram monitoring OR oxygen saturation histogram) AND (neonat* OR preterm OR prematu*). Search yielded 7707 articles. After removing duplications and applying relevant search filters, 45 articles and conference papers including dissertations were screened and 6 full-text articles were included for qualitative synthesis ( Table 1) . It was not possible to provide a meta-analysed data as the use of histogram in these studies were not always the primary intervention, rather in certain instances just a part of methodological tool kit. But the range of time spent outside saturation target that has been deemed as significant in the studies ranges from 5 to 20% (mean 10%). The lower acceptable saturation limit for this assessment ranges from 81 to 90% (mean 86%). Based on the review of literature, the authors propose that > 20% time spent below 91% or above 95% is unacceptable and should lead to clinical intervention. From the physiological point it also seems rational to consider > 20% time spent with respiratory rate > 80/min to be unacceptable. Based on the above, the authors propose the following algorithm for titration of non-invasive respiratory support based on histogram. This algorithm is just a physiological extrapolation and current practice in the authors' unit. QI initiative to implement histogram monitoring into daily clinical rounds and training of residents, nursing staff and respiratory therapist in histogram analysis. Aim: To reduce rate of death/severe ROP in < 29 weeks and improve time spent within target SPO 2 limit 90-95%. In infants < 29 weeks' gestation (n = 518), the rate of death or severe retinopathy of prematurity prior to hospital discharge decreased from 32.1 to 18.0%. Time within goal saturations (90-95%) increased from 48.7 to 57.6% (p < 0.0001). The study outlines a QI process and states the importance of multiple PDSA cycles and training at multiple levels to influence sustainable change. Borenstein-Levin et al. (2019) [12] Analysis of 6000 histograms from 73 infants of mean GA 28 ± 3 weeks was carried out and categorized into 5 types based on cumulative time spent below SPO 2 81% and number of bars between cumulative 10 and 90%. [13] To examine the correlation between interval vital signs recorded by nursing staff and continuous monitor recordings, and to determine whether aggregated monitor recordings could better predict impending escalation of respiratory support in premature infants. Ninety-four infants with median GA 32 weeks were enrolled and events of hypoxia (< 90%) and tachypnea (RR > 70/min) were recorded. Correlation between nursing records and monitor events was analysed. Philips IntelliVue MP50 Neonatal monitors were used for recording. Infants with monitor-recorded hypoxia for > 5%/day (p < 0.0001) or tachypnea for > 30%/day (p < 0.0001) were more likely to require an increase in respiratory support within the next 3 days. The overall correlation coefficient between nursing and monitor records was 0.507 for hypoxia events and 0.431 for tachypnea (RR > 70). For infants who had escalation of support, nursing documentation of hypoxia recorded by monitors was not present in 46.6% (p < 0.0001), and tachypnea in 27.8% (p < 0.0001) of infants. Manual documentation of data points for respiratory events were less in number and less accurate as compared with that generated from monitors. Histogram analysis of SPO 2 and RR could predict escalation of respiratory support. Analysis of 24 h histograms in 31 infants between 24 and 32 weeks to predict failure of transition from CPAP or HFNC. Hypothesis: Infants with ≥ 15% of time spent with SpO 2 < 86% before transitioning from CPAP or high-flow nasal cannula (HFNC) to low-flow nasal cannula, oxyhood or room air were more likely to fail. Twenty-four infants transitioned successfully, but 7 failed. Two of 7 (28.6%) who failed experienced SpO 2 < 86% ≥ 15% of the time pre-transition compared with none in the success group (p = .045). The failure group experienced SpO 2 < 86% 10.7 ± 11.9% of time pre-wean compared with 3.3 ± 4.7% of time in the success group (p = .02). Histogram analysis can predict readiness for transition from respiratory support. Limitation: Small observational study. (2016) [15] RCT comparing strategies to wean off CPAP for preterm infants. One hundred one infants < 32 weeks who met criteria to wean off CPAP were randomized to HFNC or reduction of pressure on CPAP to 4 cm H 2 O. Histogram analysis was used as methodology to label which infants were "stable" on CPAP to wean. If time spent in SPO 2 < 85% was < 5%, they were deemed stable. Though not used as primary intervention, this study also used histogram analysis as background assessment of respiratory support. Evaluation of oxygen exposure among very low birth weight infants. Infants were assigned target saturation ranges and signal extraction technology and histogram analysis was implemented to temporally quantify achievement of these ranges. Outcome analysis included BPD/death/ROP. Target saturations were altered from 88-96% to 85-93% for those requiring supplemental oxygen and from 88-100% to 85-100% for those in room air. Target was to maintain all infants within SPO 2 target range for > 75% of time. In this article, we highlight the difficulties and resulting inaccuracies in manual transcription of vitals for respiratory monitoring in the NICU. In light of this and the importance of tight control of saturations in this vulnerable cohort of infants, often with prolonged respiratory support, we have discussed how patient monitors can provide us with objective data. We have also provided a clinical range of acceptable deviation from target range beyond which a clinician should ideally look to intervene. Histogram is not a new tool and is a well-accepted statistical modality of data analysis, but as a clinical decision-making aid it is relatively recent. Small number of studies and the emphasis on preterm population only are the limitations of the review. The latter is understandable as the effects of hypoxia and hyperoxia are more pronounced and definitive on the preterm population and hence of greater clinical concern. The authors are not aware of any previous article which reviews the technology and evidence behind histogram analysis and its application in the neonatal intensive care unit. Author contributions AS and AP conceptualized the article. AP and AS conducted the background literature search. AS devised the manuscript and both authors agree to the formatting and contents. Conflict of interest The authors declare that they have no conflict of interest. Ethical approval This article does not contain any studies with human participants or animals performed by any of the authors. Evidence support and guidelines for using heated, humidified, high-flow nasal cannulae in neonatology Nasal intermittent positive pressure ventilation (NIPPV) versus nasal continuous positive airway pressure (NCPAP) for preterm neonates after extubation Neonatal non-invasive respiratory support: synchronised NIPPV, non-synchronised NIPPV or bi-level CPAP: what is the evidence in Noninvasive high-frequency oscillatory ventilation as respiratory support in preterm infants: a meta-analysis of randomized controlled trials Comparison of NIV-NAVA and NCPAP in facilitating extubation for very preterm infants Association between oxygen saturation targeting and death or disability in extremely preterm infants in the Neonatal Oxygenation Prospective Meta-analysis Collaboration Transcribed oxygen saturation vs oximeter recordings in very low birth weight infants Randomised crossover study of automated oxygen control for preterm infants receiving nasal high flow National Institutes of Health, National Heart, Lung, and Blood Institute ARDS Network (2007) Comparison of the SpO2/ FIO2 ratio and the PaO2/FIO2 ratio in patients with acute lung injury or ARDS Novel transcutaneous sensor combining optical tcPO2 and electrochemical tcPCO2 monitoring with reflectance pulse oximetry Oxygen saturation histogram monitoring to reduce death or retinopathy of prematurity: a quality improvement initiative Clinical quantification of SpO2 instability using a new histogram classification system: a clinical study Continuous pulse oximetry and respiratory rate trends predict short-term respiratory and growth outcomes in premature infants The objective use of pulse oximetry to predict respiratory support transition in preterm infants: an observational pilot study Heated humidified high-flow nasal cannula for weaning from continuous positive airway pressure in preterm infants: a randomized controlled trial Temporal quantification of oxygen saturation ranges: an effort to reduce hyperoxia in the neonatal intensive care unit Publisher's note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations