key: cord-0039839-rzzjrs0j authors: nan title: ECR 2006 - B - Scientific Sessions date: 2006-05-25 journal: nan DOI: 10.1007/s10406-006-0175-4 sha: bc0b6e929da0b1d0bf1155fac0e067dab587868e doc_id: 39839 cord_uid: rzzjrs0j nan us to find out whether delayed complications occurred. Adverse events were divided into major and minor complications (further therapy/no further therapy mandatory). Results: 905 procedures were performed in 442 patients. The size of the treated tumors ranged from 5 to 60 mm (mean 28 mm). Major complications were diagnosed after 10 of 905 treatments (3 pneumothorax, 2 bleeds, 4 abscesses, 1 stenosis of the common bile duct) and minor complications were diagnosed after 37 of 905 treatments (24 small capsule hematoma, 10 prolonged pain, 3 small pneumothorax). No treatment related death happened. Conclusion: RFA with an overall complication rate in our series of about 5 % and a major complication rate of about 1% is a safe therapeutic option for patients suffering from primary or secondary liver tumors. Monopolar saline-enhanced radiofrequency ablation of malignant liver tumors: Two year experience Z. Brnic, B. Brkljacic; Zagreb/HR (zoran.brnic@zg.htnet.hr) Purpose: To evaluate technical success, safety and short-term patient outcome of radiofrequency ablation (RFA) of malignant hepatic tumors under local anesthesia and sedation vs. general anesthesia. Methods and Materials: During 2 years a total of 61 patients underwent salineenhanced RFA of liver tumors (59 secondary, 2 primary; mean size 41 mm, range 15-80, 1-4 nodules treated per session); 34 patients under local anesthesia with sedation (group 1) and 27 under general anesthesia (group 2). Results: Mean effective ablation time (min), ablation volume index (ablated volume/total nodule volume), % of patients with complete tumor ablation and % of patients free of recurence after 6-12 months in groups 1 vs. 2 were: 12 vs. 22; 61 vs. 94, 48 vs. 70, and 29 vs. 48, respectively. Incomplete tumor ablation and recurrence were more frequent in patients with larger tumors and in group 1. Apart from minor complications in 13%, only 5% of patients had major complication (1 liver abscess, 1 track seeding, 1 gallbladder injury). No procedure-related mortality was observed. Conclusion: RFA is a safe and efficient technique for local control of growth of liver tumors < 5 cm in diameter. Pain during procedure is a limiting factor and RFA under general anesthesia results in a better therapeutic response. Additional concentric expansion of RF-coagulation volume: Efficacy of hypertonic saline injection in area of vaporization around the electrode: Ex vivo porcine liver study A. Shimizu, H. Ishizaka, T. Tatezawa, S. Awata; Maebashi/JP (simiakiko@yahoo.co.jp) Purpose: When coagulation volume is insufficient in initial radiofrequency (RF) ablation additional concentric expansion of coagulation would be an ideal solution. However vaporization occuring around the electrode after RF application that hinders current delivery limits further coagulation. Whereas hypertonic saline injection prior to RF ablation is known to increase current conductivity and thereby enlarge coagulation volume. We tested whether hypertonic saline injection in the area of vaporization, followed by additional RF application could enlarge the coagulation volume. Methods and Materials: RF coagulation of excised porcine liver was performed in five different areas. First an RF electrode (Cool-tip) was coaxially placed through a cannula and coagulated a maximal area at output power of 40 W. Temperature of the liver near the electrode was continuously measured by a thermocouple and the margin of maximal coagulation was determined as a 49 °C area after 15 minutes. The RF electrode was then removed, 1.5 mL of 36% saline was infused through a coaxially placed PEI needle and RF application was resumed. Margin of maximal coagulation was then determined in the same way. Finally, coagulation areas were observed on cross-sections. Results: Enhancement of coagulation volume (by a distance of 4-15 mm; mean: 9 mm from the electrode) was confirmed in all cases after injection of 36% saline. The coagulation area was left-right symmetrical with a margin immediately adjacent to the thermocouple. Conclusion: Injection of a small amount of 36% saline in the area of vaporization followed by additional RF application can concentrically expand the coagulation volume, simplifying the treatment of larger lesions. Purpose: To evaluate clinical experience of percutaneous ablation of renal cell carcinoma using combined radiofrequency coagulation and ethanol injection. Methods and Materials: Nine patients (5 males, mean 68 years) presented with renal tumours. The tumour size was < 2 cm in 3, 2-3 cm in 4 and 3-4 cm in 3. All patients were either considered unfit for surgery or had refused it. Absolute ethanol (0.5-3 ml, mean 1.4 ml) was injected into the tumour immediately before treatment with radiofrequency. Radionics cool-tip electrodes were used in 8 patients and a Rita electrode in 1. CT follow-up was carried out the day after the procedure and three monthly thereafter for the first year. Follow-up 4-38 months, mean 20 months. Results: Complete coagulation necrosis was observed on CT in six patients after the first procedure. Two patients were retreated because of enhancing areas on CT, which may have represented residual disease. The mean hospital stay was 2 days. All the patients are well, except for one patient who has developed a colonic carcinoma. One patient had a subcapsular haematoma. Two patients had transient loin pain, treated conservatively. Conclusion: Combined therapy of percutaneous radiofrequency ablation and ethanol injection of renal tumours is an effective method of treatment, which destroys the tumour. It is tolerated well and is associated with few complications. Laser-induced thermal ablation of lung metastases: Follow-up imaging and feasibility C. Rosenberg, C. Weigel, S. Langner, C.-P. Fröhlich, N. Hosten; Greifswald/DE (christian.rosenberg@uni-greifswald.de) Purpose: Laser-induced thermal ablation of pulmonary metastases as a new option in multimodal cancer therapy shows first technical and clinical results in an ongoing study. In 42 patients 64 lesions were treated (CT fluoroscopy, Monocath 5.5 F water-cooled applicator, MeoMedical, Germany; 14 W for up to 15 min, 1 to 3 applicators per lesion). Different primaries, mainly colorectal carcinomas (10/42 patients), had caused metastatic disease. Lesion size ranged from 8 to 85 mm (median 27.6 mm). Follow-up CT was performed monthly for the first three months, then after 3 and 6 months. Median follow-up was 7.6 months. Results: Ablative therapy was technically succesful in 62/64 lesions. In 7 cases therapy was completed after 2 sessions and in 2 cases after 3 sessions. An initial increase in size of the visualized correlate was seen in 25 cases and secondary cavities in 8 cases. Marked size reduction of the imaging correlate on further follow-up was demonstrated in 32 lesions. These patients were diagnosed successfully treated. Residual CT findings were: ground-glass opacity or cavitary defects. Complications were not significantly different to diagnostic lung biopsy. Incidence of recurrent tumor growth was significantly higher in lower segments (7-10), whereas there was no difference between central vs. peripheral location. Conclusion: Laser-induced thermal ablation is a promising option in multimodal therapy of non-resectable lung metastases. Feasibility is correlated to target location. Interval reduction for therapy success evaluation is a future task. Background: It is well described that "vulnerable" atheromatous plaque has a thin, fibrous cap and extensive lipid core with associated inflammation. This inflammation can be detected as areas of signal drop on MR imaging using a novel contrast medium, Sinerem, an Ultrasmall Super-Paramagnetic Iron Oxide (US-PIO). Studies using USPIO have, thus far, not reported its additional T1 effect of enhancing fibrous cap. Methods: 20 patients with carotid stenosis (10 symptomatic, 10 asymptomatic) were imaged with 1.5 T, pre and 36 hours post USPIO infusion. Images were anonymised, plaque seen was manually segmented into quadrants (CMR tools, London) and focal regions of signal drop post contrast. Signal change was normalised to adjacent muscle. Results: There were 9 males and 11 females and mean ages were 71 (symptomatic) and 73 (asymptomatic). Mean stenoses were 82% (symptomatic) and 72% (asymptomatic). Symptomatic patients had significantly more quadrants with signal drop post USPIO than asymptomatic patients (75% vs 32%, p < 0.01)and more focal regions of signal drop (0.68 vs 0.26 regions/image, p < 0.01). Asymptomatic subjects had more quadrants with signal enhancement than those with symptoms (68% vs 25%, p < 0.05). Conclusion: Symptomatic plaques had more focal areas of signal drop than asymptomatic plaques, suggesting that they harboured a greater inflammatory burden. Asymptomatic plaques showed more enhancement than symptomatic plaques, suggesting thicker fibrous caps and greater stability, although some focal areas of signal drop were noted suggesting vulnerability. If validated by larger studies, USPIO may be a useful dual contrast medium, able to enhance risk stratification of patients with carotid stenosis, improving selection for intervention. Purpose: Our objective was to evaluate the diagnostic potential of MD-CTA and 3D CE-MRA with parallel acquisition technique in evaluating carotid artery stenosis. Methods: 30 previously untreated symptomatic patients (24m,6w) suffering from suspected carotid artery stenosis were included. Standardized MD-CT-angiography was performed using 16-row-and 64-row-scanners. 3D-CE-MRA was performed using a 1.5 Tesla scanner with 12-channel surface-coils. PAT using SENSE reconstruction with an acceleration factor of 2 was applied. MIP and MPR were calculated for CTA and MRA. Two blinded neuroradiologists independently quantitatively assessed the degree of stenosis separately for each modality following modified NASCET criteria. Furthermore visual assessment and evaluation of plaque morphology were performed. Image reading results were compared to surgical and duplex sonographic findings which served as a standard of reference in all patients. For evaluation of plaque morphology, correlation to histopathologic data was performed. T-test and Kappa-test were applied for statistical evaluation. Results: No significant differences were found between either examination modality and the two standards of reference with regard to the quantitative evaluation of carotid artery stenosis (p> 0.05). Mean degree of stenosis was 87.9 ± 9.0% for MD-CTA, 84.8 ± 10.3% for MRA, and 90.2 ± 4.9% for Doppler sonography. Evaluation of the visual impression of the degree of stenosis showed no significant difference between MD-CTA and MRA, with a tendency toward underestimating the degree of stenosis with both modalities. Plaque ulcerations were detected better with MD-CTA than with MRA (p < 0.05). A high interrater correlation was found (kappa=0.91). Conclusion: Vascular imaging using both CE-MDCT angiography and 3D-MRangiography with parallel imaging technique have a good diagnostic potential in the evaluation of carotid artery stenoses. Regarding plaque delineation, MD-CTA appears to be superior. Common carotid artery intima-media thickness is independently associated with risk for transient ischemic attacks in addition to main risk factors and plaque presence G. Terzis, A. Kalogeropoulos, A. Chrysanthopoulou, G. Gioldasis, D. Alexopoulos, D. Siablis; Patras/GR (riondoc@otenet.gr) Purpose: Sound evidence supports the independent value of common carotid artery intima-media thickness (CCA-IMT) as a risk factor for stroke. However, the association of CCA-IMT with risk for transient ischemic attacks (TIA) is less clear. Moreover, recent data suggest that incidence of stroke is related to CCA-IMT even in the absence of carotid plaque. We investigated whether CCA-IMT and presence of carotid plaque are associated with risk for non-embolic TIA after adjusting for major risk factors in a case-control series. Methods and Materials: Our study population comprised of 80 consecutive patients undergoing carotid ultrasonography during hospitalization for documented first TIA (TIA Working Group Definition 2002) and 245 asymptomatic control subjects undergoing screening evaluation. Patients with known cardioembolic sources were excluded. We selected 160 controls (2:1) by propensity score matching in order to estimate with minimal bias the independent effect of CCA-IMT and plaque presence. Results: Patients and controls were matched for age (66 ± 10 vs. 65 ± 13 yrs, p = 0.492), gender (55.0% vs. 56.2% male, p = 0.854), smoking (28.8% vs. 30.6%, C D E F B A G p = 0.686), hypertension (57.5% vs. 61.2%, p = 0.576), diabetes (18.8% vs. 20.0%, p = 0.817), and dyslipidemia (39.4% vs. 36.2%, p = 0.639). Compared to control subjects (0.68 ± 0.13 mm), CCA-IMT in TIA patients (0.73 ± 0.13 mm) was significantly higher (p = 0.01). The incidence of plaques did not differ in the two groups (63.8% vs. 71.2%, p = 0.31). In multivariate analyses, only CCA-IMT emerged as independent predictor (p = 0.01). Conclusion: Increased CCA-IMT is associated with higher risk for non-embolic TIA in addition to established risk factors. This effect is independent of the presence of carotid plaque. The hemodynamic effects of carotid desobstruction in patients with symptomatic carotid artery stenosis. Purpose: To evaluate whether pre-treatment CT perfusion analysis in patients with symptomatic carotid artery stenosis (CAS) can be used to distinguish between groups of patients that differ with respect to improvement in cerebral perfusion after carotid intervention. Methods and Materials: Thirty-six patients with symptomatic CAS and no significant contralateral disease were scanned before and 1 month after primary stent placement (n = 21) or carotid endarterectomy (n = 15) using a multi-detector CT scanner. We calculated the difference in mean transit time (dMTT) and ratio of cerebral blood flow (rCBF) between symptomatic and asymptomatic hemispheres. We classified patients into three groups according to the pre-treatment dMTT: group 1, dMTT < 1s, group 2, dMTT 1-2s and group 3, dMTT > 2s. We studied changes in brain perfusion for the whole group and separately for the three subgroups by comparing pre-and post treatment dMTT and rCBF using a paired ttest. Results: For the total group mean dMTT decreased from 1.3s before to 0.1s after treatment (p < 0.001), mean rCBF increased from 0.85 to 0.93 (p = 0.02). In group 1 (n = 16) dMTT and rCBF did not significantly improve, in group 2 (n = 12) only dMTT improved significantly, and in group 3 (n = 8) both dMTT and rCBF improved significantly. Conclusion: Pre-treatment dMTT is a suitable CT perfusion measure to distinguish between groups of patients with different improvement of cerebral perfusion after carotid intervention. In 22% of patients significant improvement in dMTT and rCBF was seen. This might hold promise for selecting patients that benefit most from carotid intervention. Purpose: Patients with critical ICA stenosis benefit from endarterectomy but those with occlusion or long-segment tight stenosis (trickle-flow) do not. Distinguishing between these two on conventional ultrasound is problematic as it is insensitive to very slow flow. The current reference standard is digital-subtraction angiography (DSA). The reliability of other non-invasive investigations is unproven. We prospectively studied the accuracy of contrast-enhanced ultrasound (CE-USS), 2D time-of-flight (2D-TOF) magnetic resonance angiography (MRA) and contrast-enhanced MRA (CE-MRA) against DSA in patients with apparent carotid occlusion on conventional ultrasound. Dual reporting of CE-USS (unblinded) and MRA (blinded) was performed. Results: 60 vessels were studied. 41 vessels underwent CE-MRA and DSA. On CE-MRA, 2 vessels were patent and 37 were occluded. It was difficult to differentiate critical stenoses from trickleflow on MRA due to lack of dynamic flow information but no critical stenosis on DSA was misclassified as occluded by CE-MRA. One study was non-diagnostic and there was inter-observer disagreement in one study (trickle-flow on DSA). On 2D-TOF a critical stenoses on DSA was misclassified as occluded. 54 vessels underwent CE-USS and DSA. Technical problems prevented confident assessment of vessel patency with CE-USS in 7 patients. A critical stenosis on DSA was misclassified as occluded by CE-USS. In the presence of apparent carotid occlusion on conventional ultrasound, occlusion on CE-MRA can confirm the presence of a non-surgical lesion but patency does not imply the presence of a surgical lesion. 2D-TOF and CE-USS can misclassify surgical lesions as being occluded. How effective is a 3D-time-resolved echo-shared angiographic technique Results: High field MR imaging with pelvic phase-array coils should include three T2 weighted sequences in sagittal, axial and coronal oblique orientation (short and long axis of the uterine body) of the pelvic content. High resolution post contrast acquisition in the optimal image orientation should always be applied for the diagnosis of myometrial invasion and acquired between 90 and 150 seconds after intravenous contrast injection; optimal timing of image acquisition depends on the chosen 2D or 3D technique. If cervical invasion is suspected additional slice orientation according to the axis of the endocervical channel is recommended. Due to the limited sensitivity of MR imaging to detect lymph node metastasis retroperitoneal lymph node screening with pre-contrast sequences is optional. Possible lymph node invasion should be suggested indirectly based on tumor grade at endometrial tissue sampling and myometrial and cervical invasion detected at MR imaging. Purpose: Despite excellent contrast /spatial resolution at 3 T Magnetic Resonance imaging (MRI) in neuro-imaging, its potential for body area remains controversial. This study aimed to evaluate MR imaging of the female pelvis at 3 T/ 1.5 T on high-resolution T2 weighed image (T2WI), HASTE, and Diffusion weighted image (DWI). Material and Methods: MR images of 19 healthy women were obtained by 3 T/ 1.5 T (Trio/Symphony:Siemens) MR unit using phased array coil with the following imaging parameters; T2WI (Axial/Sagittal): TR/TE=3200-3600/94-103 ms, FOV/voxel size = 260x195/0.8x0.6x2.0 mm, NEX=2. HASTE (sagittal): TR/ TE=3000/100 ms, FOV/voxel size = 250x250/1x1x5 mm, NEX=1. DWI (axial): TR/ TE=2200/80 mm, FOV/voxel size = 200x200/2.1x1.6x3.0 mm, NEX=3, b=0.500.1000 s/mm 2 . Parallel imaging (sense factor:2) was used. Qualitative evaluations of image quality (overall, uterus, uterine zonal structure, cervix, ovary) and artifacts were made independently by two radiologists using a 5-point scale and mean scores were used for analysis. Quantitative evaluations were made for contrast of uterine zonal structure and signal intensity of anterior/ posterior fat (for signal intensity homogeneity). Intra-individual comparisons between 3/1.5 T image evaluations were performed. Results: Image quality was superior at 3 T to 1.5 T for uterus (T2WI sagittal), uterine cervix (T2WI axial/sagittal and HASTE) and ovary (T2WI axial). Images were more inhomogeneous at 3 T (T2WI axial), while motion artifact was severer at 1.5 T (T2WI axial/sagittal)(p < 0.05). No significant difference was found between 3/1.5 T MR imaging for overall image quality on all the sequences. Conclusion: Our results suggest that MR imaging of the female pelvis at 3 T has a potential to provide excellent image for uterine cervix, and ovary on high-resolution T2WI. No significant improvement in overall image quality might be due to inhomogeneity at 3 T images. Purpose: To evaluate the diagnostic accuracy of 3 T Magnetic Resonance Imaging (MRI) in assessing the depth of myometrial infiltration in endometrial cancer. Methods and Materials: Twenty-six patients (60 ± 13.4 years old) with histologically proven endometrial carcinoma underwent preoperative MR imaging. Axial FSET1, axial, sagittal and para-coronal FSET2 and para-coronal FMPSGR Gadolinium-enhanced sequences were performed by a high field strength magnet (3 T, Philips Medical Systems, Best, The Netherlands) using an external phased-array coil. Within one month from MR imaging, all patients underwent hysterectomy. The exams were evaluated by two different radiologists and the results of MR imaging in assessing the depth of myometrial infiltration were compared with the histological results. A statistical evaluation was performed. Results: Histology showed intramucosal neoplasm in 4 patients, myometrial infiltration less than 50% in 6 patients, more than 50% in 12 patients and transmural cancer in 4 patients. At MR imaging, intramucosal cancer was diagnosed in 3 cases, myometrial infiltration less than 50% in 8 patients, more than 50% in 10 patients and transmural in 4 patients. Concordance between MR imaging and histological diagnosis was 88.4% (23 over 26 cases), with mean sensitivity, specificity and accuracy respectively of 89.5%, 97.5% and 95.1%. Conclusion: Our preliminary results show that 3 T MR imaging proved to be a reliable and accurate tool in assessing the depth of myometrial infiltration in endometrial cancer. Its accuracy falls in the highest rate of diagnostic accuracy achievable by 1.5 T MR imaging. Our inclusion MR imaging criteria for adnexal malignancy were: contrast-enhanced solid lesions, contrast-enhanced solid component in mixed lesions (except those with low-signal intensity solid component on T2WI), contrast-enhanced papillary projections in cystic lesions (except those with low-signal intensity papillary projections on T2WI) or septal thickness ≥ 3 mm. Ascites, peritoneal metastasis and pelvic enlarged lymph nodes were also inclusion criteria for malignancy. Results: By MR imaging evaluation 90 adnexal lesions were malignant and 64 C D E F B A G were benign. Histologically, 85 masses were confirmed to be malignant (ovarian: 62 carcinomas, 15 epithelial borderline tumors (BT), 2 tumors with uncertain malignant potential; 5 fallopian tube carcinomas and 1 mesosalpinx BT); 69 were benign (63 ovarian tumors, 4 functional cysts), 1 peritoneal pseudocyst and 1 epiploic appendix esteatonecrosis. Correlating MR imaging with histology results we confirmed 83 were malignant (true positive) and 62 were benign (true negative) lesions. The MR sensitivity and specificity for malignancy were 90% and 98%, respectively. MR reached an accuracy of 94% with a 92% positive predictive value and 97% negative predictive value for malignant adnexal lesions. Conclusion: MR has an excellent accuracy with a very high predictive value for the diagnosis of malignant adnexal lesions. Purpose: Ovarian tumours with borderline malignant potential have a favourable outcome. Characterisation of mixed solid and cystic ovarian masses pre-operatively can aid in surgical planning and may allow preservation of fertility in younger patients. This study examines the imaging characteristics of borderline ovarian tumours in comparison with other encapsulated ovarian masses. Methods and Materials: 49 patients with 59 encapsulated ovarian lesions were retrospectively evaluated. All patients underwent MR evaluation of suspicious pelvic adnexal masses between June 2001 and June 2004. MR imaging was performed at 1.5 or 3 T (GE LX Signa systems) using a flexible 4 or 8 channel torso coil respectively. FRFSE images were acquired sagittally, axially and obliquely followed by T1W FSPGR ± fat suppression as determined clinically. Buscopan and a dielectric pad at 3 T were used to minimise artefacts. Histopathology was available in all subjects. Results: Ovarian cysts ranged in size from 0.9 to 26 cms. All cystadenomas (14), fibromas (2), a haemorrhagic cyst (1) and 2/3 cystadenocarcinomas were correctly diagnosed. MR correctly called 49% of cystadenofibromas, but 51% was misclassified as cystadenomas or fibroma (1/31). 3/6 borderline tumours were correctly identified, 2 classified as cystadenoma or indeterminate cyst and one termed cystadenocarcinoma, although omental deposits were present in this case. Typical features seen in the correctly diagnosed borderline tumours included endocystic projections separately identified from an intact ovarian capsule. Conclusion: This study demonstrates the potential of MR imaging in pre-operative assessment of ovarian masses and indicates the need for high resolution imaging preferably at 3 T. Multislice CT and MR imaging evaluation of ovarian masses A.C. Tsili, C. Tsampoulas, E. Arkoumani, E. Paraskevaidis, S.C. Efremidis; Ioannina/GR (athinatsili@in.gr) Purpose: To evaluate the accuracy of multislice CT and magnetic resonance (MR) imaging in the detection and differentiation of benign and malignant ovarian masses. Methods and Materials: Preoperative CT of the abdomen and MR imaging of the pelvis were performed in 40 women presenting with clinically or sonographically detected ovarian masses. CT examinations were performed on a 16-slice CT scanner, using a detector collimation of 16 X 0.75 mm and a pitch of 1.2. MR examinations were performed on a 1.5 Tesla unit and T2 and T1-weighted sequences, before and after intravenous administration of gadolinium chelate compounds were obtained. The accuracy of multislice CT and MR imaging in the detection and characterization of ovarian masses was evaluated, using surgical and histological results as reference. Results: Both multislice CT and MR imaging detected 50 (93%) out of 54 adnexal masses, as proven histologically. The sensitivity, specificity, positive predictive value and negative predictive value of multislice CT and MR imaging in the differentiation between benign and malignant ovarian masses were 83%, 89%, 71%, 94% and 92%, 95%, 85% and 97%, respectively. Conclusion: Multislice CT and MR imaging demonstrated a similar accuracy in the detection of ovarian masses, but MR imaging performed better in the characterization of adnexal masses, mainly due to its superior tissue characterization. Role of endocavitary MR imaging for posterior deep endometriosis with rectosigmoid involvement. Preoperative prospective study with surgical correlation C. Roy, A. Biblocque, V. Thoma, A. Wattiez, J. Leroy, C. Saussine; Strasbourg/FR (Catherine.Roy@chru-strasbourg.fr) Purpose: US and MR imaging are able to detect rectosigmoid infiltration of deep endometriosis. The purpose was to assess added diagnostic value of TVUS with MR imaging obtained by phased-array (bMRI) and endovaginal (evMRI) to detect intestinal involvement. Methods and Material: From 2004 to 2005, 47 women underwent preoperative TVUS, bMRI and evMRI before surgical resection of the nodule. MR imaging was performed using T2-weighted FSE and T1-weighted SE images before and after Gd. Examination results were correlated with findings during surgery and pathological analysis. Results: At pathology, 43 women demonstrated deep muscle layer invasion, 2 an isolated serosal involvement and 2 no intestinal extension. All nodules had well-defined borders, low signal intensity with small hyperintense spots on T2weighted MR (18 cases). All were homogeneous and hypoechoic on TVUS. Spiculated strandings were more often seen on MR (44 cases) than on TVUS (35 cases). Deep wall extension was correctly diagnosed as a hypoechoic or hypointense homogeneous thickening of muscle layer in 32 cases on TVUS (Se : 74%) and 41 cases on MR (Se : 95%). An additional pattern as a hyperintense layer of mucosa surrounding the nodule was found on MR in all cases. EvMRI was clearly better at diagnosing intestinal involvement for small lesions (less than 2 cm) localized inside the pouch of Douglas or recto-vaginal septum. Conclusion: evMRI gives additional information over TVUS in the diagnosis of rectal involvement preoperatively for small sized nodules. A hyperintense signal layer surrounding the nodule is specific for this purpose. MR imaging of the pelvis in mullerian duct abnormalities: Is there a role in selecting patient for hysteroscopic metroplasty? P. Paolantonio 1 , M. Rengo 2 , M. Anzidei 2 , A. Carnevale 2 , F. Iafrate 1 , C. De Angelis 2 , A. Laghi 1 ; 1 Latina/IT, 2 Rome/IT (paolantoniopasquale@hotmail.com) Purpose: Selection of patients for hysteroscopic metroplasty requires a complete assessment of uterine morphology, usually achieved by laparoscopy which represents the gold standard in the evaluation of uterine morphology. The aim of our study was to assess the role of MR imaging in selecting patients with mullerian duct fusion defects for hysteroscopic metroplasty. We performed a prospective study on 25 women with known mullerian duct defects assessed at hysteroscopy. The examination was performed using an 1.5 T magnet equipped with a phased-array coil acquiring a high-resolution TSE T2W sequence (TR: 3000 msec/TEeff 190 msec/ETL: 29/ Purpose: FAIR is a novel technique to evaluate blood perfusion. Compared with first pass dynamic contrast enhance MR imaging, it is non-invasive, more convenient and cost-efficient. This study aims to determine whether quantification at FAIR imaging can be used to assess regional cerebral blood volume (rCBV) in a manner similar to CE MR imaging. Methods and Materials: Using a GE Signa Excite II 3.0 T MR scanner, 34 patients, including 17 gliomas (10 low grade WHO I, II and 7 high grade WHO III, IV), 7 meningiomas, 3 metastases, 3 acoustic schwannomas, 2 medulloblastomas, 1 craniopharyngioma and 1 hemangioblastoma were examined by FAIR and CE MR imaging. Three ROIs were selected on the solid part of each neoplasm on the post-processing images. Mean FAIR value and rCBV were calculated. The results of benign and malignant gliomas were compared with independent samples T test. The results of all cases were analysed with linear correlation analysis. Results: There was statistically significant difference between the FAIR value and rCBV of benign and malignant gliomas (P < 0.01). The FAIR value and rCBV of all cases showed positive linear correlation (P < 0.05). Conclusion: FAIR and CE MR imaging methods are both helpful in differentiating benign from malignant glioma. The significant positive linear correlation in detecting tumoral blood perfusion between FAIR and CE MR imaging methods suggests that FAIR imaging, like CE MR imaging perfusion, is complementary to routine MR imaging in the assessment of tumoral perfusion. Furthermore, because there is no need for contrast enhancement, FAIR imaging, a non-invasive technique, will be more acceptable in clinical application. Purpose: Pathologically, there is microscopic infiltration of tumoral cells across the corpus callosum (CC) in patients with high grade gliomas. The aim of this study is to determine if there are DTI changes within the normal appearing CC and contralateral white matter (WM-CONT) and if there are differences between glioma grades. Methods and Materials: 30 patients with primary glioma -classified using the WHO classification as low-grade (n = 10), anaplastic (n = 12) and glioblastoma multiforme (n = 8) -had DTI performed using a pulsed gradient, double spin echo, echo planar imaging (TR/TE=5300/74; 128x128 matrix; 220x220 mm FOV; fortyfive 3 mm contiguous slices; b= 1000 s/mm 2 ) at 1.5 T. ROIs were placed in the CC in three locations, the ipsilateral callosum (CC-IPSI), central callosum (CC-CENT) and contralateral callosum (CC-CONT) to the side of the glioma, as well as within the WM-CONT. FA and ADC were measured. Results: There is significantly increased ADC (p = 0.003) and decreased FA (p < 0.001) in the CC-IPSI and in the CC-CENT (p = 0.007, and p < 0.001 respectively) in grade III gliomas compared with grade II. There is also significantly decreased FA in the normal appearing WM-CONT (p = 0.017), as well as the CC-IPSI (p = 0.02), CC-CENT (p = 0.001) and CC-CONT (p = 0.012) in grade IV compared to grade II. Conclusion: DTI is able to demonstrate changes in FA and ADC in the normal appearing corpus callosum of grades III and IV and contralateral white matter of grade IV, indicating potential microscopic tumoral infiltration. Purpose: New treatments for patients suffering from glioblastoma have shown promising results. This study evaluates recent MR imaging follow-up techniques on newly diagnosed glioblastoma patients and analyzes the potential for perfusion and diffusion sequences in anticipating disease progression. Various MR imaging parameters are further studied as potential prognostic factors of disease progression. This study included patients from a phase II clinical trial of temozolomide combined with surgery and radiotherapy. MR imaging follow-up was carried out 1 month after radiochemotherapy and every 2 months thereafter. T1-and T2-weighted SE series as well as diffusion and perfusion EPI series were systematically performed on a 1.5 T magnet. The analysis used clinical parameters combined with classical criteria on tumour size variation to focus on disease progression. Results: Among the 41 patients included in the study (311 MR imaging examinations), 32 demonstrated disease progression at some time in their follow-up. Perfusion and diffusion imaging did not anticipate disease progression. Further multivariate analysis determined T1 largest diameter and T2 size variation as significant survival parameters at disease progression (p < 0.02, p < 0.05). Conclusion: Our study presents a simple time scheme combined with statistical analysis for the evaluation and potential scoring of brain tumours and treatments. T1 and T2 techniques remain the gold standard in setting the date of disease progression. At disease progression, T1 contrast-enhancement larger than 3 cm in size combined with increased T2 hyper signal is a marker of inferior prognosis. Perfusion and diffusion techniques showed a complex behaviour and could not be used to anticipate tumour progression. C D E F B A G B-095 10:57 ! Functional mean diffusion map: A noninvasive MR imaging biomarker for evaluating cell proliferation of brain gliomas Z.X. Li, J.P. Dai, T. Jiang, S.W. Li, Y.L. Sun, P.Y. Gao, J. Ma; Beijing/CN (tshirtjean@163.com) Purpose: To investigate the correlation between the mean diffusivity (MD) measured using diffusion tensor imaging (DTI) and the Ki-67 labelling index in gliomas. Methods and Materials: DTI was performed using a Siemens Trio MR scanner in fifty-six consecutive patients with histologically proven gliomas. The regions of interest (ROIs) were positioned on the solid portion of the lesion. The MD values were calculated. Ki-67 staining of paraffin embedded sections with monoclonal antibody MIB-1 was done. In the area with the highest number of labelled cells, the percentage of Ki-67 positive nuclei (labeling index: LI) was determined. Then, the results of Ki-67 LI were divided into three groups: High (LI > 40 %), median (15 % ≤ LI ≤ 40 %) and low (LI < 15 %). Then, the Ki-67 LI were compared with MD values. The MD values of the high Ki-67 LI group (0.863 ± 0.197 × 10 -3 mm 2 /s) were significantly lower than those of the median Ki-67 LI group (1.130 ± 0.275 × 10 -3 mm 2 /s) (p = 0.015) or low Ki-67 LI group (1.278 ± 0.298 × 10 -3 mm 2 /s) (p < 0.001). Negative correlation existed between MD values and Ki-67 LI (r s = -0.570, p < 0.001). Conclusion: In brain gliomas, Ki-67 LI bears a significant relationship to MD values. Its molecular pathological base is that Ki-67 can precipitate the tumor cell proliferation. The MD value of DTI is regarded as a noninvasive MR imaging biomarker for evaluating cell proliferation in gliomas. Purpose: To asses the validity of a treatment response evaluation system established for malignant gliomas to patients with primary CNS lymphoma (PCNSL). Methods and Materials: MR imaging examinations of 112 immunocompetent patients with biopsy-proven PCNSL were retrospectively evaluated for response to primary MTX-based chemotherapy according to the Macdonald criteria. The response classification was compared to the further course and treatment of the disease with the aim of identifying patients who did not achieve complete response (CR) but nevertheless remained progression-free for at least 6 months without receiving further therapy. Results: Sixty-eight patients had persisting contrast-enhancing lesions after primary therapy. Four patients of this cohort did not receive further tumour-directed therapy. After follow-up of 9, 24, 32 and 54 months, all 4 patients did not present with progressive tumour growth. In all 4 patients, the residual lesions were 5 mm in diameter or smaller. In all patients, the residual lesion was found in the vicinity of the biopsy site. Conclusion: Residual contrast enhancement may be observed in PCNSL patients after successful therapy and does not necessarily indicate residual, biologically active lymphoma. On the basis of the morphology of the non-progressive residual lesions in the patients described above, we propose a modification of the Macdonald response criteria which, in the absence of perilesional oedema, classifies contrast-enhancing lesions < 3 mm in general, or contrast-enhancing lesions < 5 mm in the region of biopsy, haemorrhage, infection or a previously larger (> 5 cm) tumour lesion as CR. Purpose: To evaluate with MR imaging the spinal epidural and leptomeningeal infiltration in patients with lymphoma and leukemia and correlate the results with lumbar puncture. Methods and Materials: 50 patients with ALL, AML and NHL and with clinical suspicion of spinal extra-medullary intra-dural spread underwent MR imaging using a 1.0 T superconductive magnet. We performed sagittal-SE-T1-weighted scans (TR: 500 ms; TE: 20 ms; n-acq: 3; acq-time: 3'40") and FSE-T2-weighted scans (TR: 2840 ms; TE: 96 ms; acq-time: 3'30") at the level of the dorsal and lumbar spine (sl. thickn: 4 mm). Then we performed contrast-enhanced sagittal-T1-weighted scans. In positive cases we also performed axial FSE-T2-weighted scans and contrast-enhanced axial T1-weighted scans. All the patients underwent lumbar puncture within two days after MR examination. Results: We detected 10 patients with positive features of spinal epidural and leptomeningeal infiltration on MR imaging. Among these 10 patients only 7 had a positive lumbar puncture ("blast" forms or lymphomatous cells in CSF). Lumbar puncture was negative when the neoplastic infiltration involved the epidural space and the nerve roots. The most sensitive MR imaging-sequence in the detection of spinal epidural and leptomeningeal involvement has been considered the sagittal T1-weighted scans performed after administration of Gadolinium-DTPA. Conclusion: In our limited experience MR imaging was more sensitive than lumbar puncture in the detection of spinal epidural and leptomeningeal spread by hemopoietic tumors. The diagnosis of leptomeningeal spread is fundamental because it modifies the therapeutic measures and its early symptoms are often nonspecific. In patients with a hemopoietic neoplasm and strong clinical suspicion of spinal epidural or leptomeningeal infiltration, when lumbar puncture is negative, MR imaging should be performed. Detection of brain metastases with 3 T MR imaging: Comparison among three contrast-enhanced sequences M. Hanamiya 1 , S. Kakeda 1 , Y. Korogi 1 , N. Ohnari 1 , J. Moriya 1 , T. Sato 1 , Y. Hiai 2 ; 1 Kitakyushu/JP, 2 Kumamoto/JP (hanamai@med.uoeh-u.ac.jp) Purpose: 3 T MR systems have some advantages of a greater relative T1-shortening effect and an increased signal-to-noise ratio for contrast-enhanced MR imaging. However, it has not been clarified which T1-weighted sequence is appropriate for detection of brain metastasis at 3 T. The purpose of this study is to compare the detection and conspicuity of enhancing brain metastases among three contrast-enhanced sequences, 3D fast SPGR with 1.4 mm isotropic voxel, conventional spin-echo sequence (SE), and inversion recovery fast SE (IR-FSE). In 20 consecutive patients with brain metastases referred for enhanced brain imaging, axial images of SE, IR-FSE and 3D fast SPGR at 3 T were obtained after administration of gadoteridol (ProHance; Bracco) at 0.2 mmol/kg. A 3D fast SPGR was acquired with 1.4-mm thick sections, and both SE and IR-FSE were acquired at the same section locations with section thickness of 6 mm. Two neuroradiologists assessed the image quality of three sequences with regard to the lesion detectability of brain metastases. Results: In nine (45%) of 20 patients, 3D fast SPGR revealed additional metastases. Significantly more metastases were detected with contrast-enhanced 3D fast SPGR than with SE and IR-FSE (a total of 114 metastases with 3D fast SPGR vs.70 with SE and 69 with IR-FSE). In particular, 3D fast SPGR was superior to the other two sequences in detection of the small lesions (< 5 mm). The post-contrast fast SPGR sequence with 1.4 mm isotropic voxel is clinically valuable for detecting brain metastases at 3 T. High-resolution susceptibility weighted MR imaging in patients with brain tumors during the application of exogeneous gaseous agents J. Sedlacik, A. Rauscher, C. Fitzek, B. Walter, A. Hochstetter, R. Kalff, W.A. Kaiser, J.R. Reichenbach; Jena/DE (jan.sedlacik@med.uni-jena.de) Purpose: With susceptibility weighted imaging (SWI) it is possible to obtain functional and anatomical information with high spatial resolution. The aim was to investigate the response of brain tumors to a modulation of venous blood oxygenation by breathing air, carbogen and oxygen. Methods and Materials: Data from six patients (four with glioblastoma multiforme (grade IV), one with astrocytoma (grade II), and one with a cerebral metastasis; all histologically confirmed) were acquired using a 3D gradient-echo, first-order velocity compensated sequence on a 1.5 T scanner (TE/TR/ alpha=40 ms/67 ms/25, FOV=256x192x64, matrix=512x192x64). One scan was acquired for each breathing condition. SWI was computed by multiplying the magnitude images with phase masks calculated from the same scans. Additionally, venous vascularization was assessed on minimum intensity projections of the SWI scan acquired during air breathing. Signal changes were evaluated over ROIs of typically 0.5 cm3 (15x15x10 voxels) and maps of relative signal changes were computed. Results: Signal intensities in the tumor decreased during carbogen breathing compared to oxygen inhalation in 3 cases. The glioblastomas displayed the strongest response and high heterogeneity (signal changes between -7% and +28%). The astrocytoma and cerebral metastasis showed a negative response to carbogen in all ROIs ranging from -9.3% to -2.0% (astrocytoma) and -8% to -16% (cerebral metastasis). C D E F B A G Methods and Materials: A total of 495 sets of screen-negative mammograms (prediagnostic mammogram and the immediate previous mammogram) were collected from women participating in the biennial Dutch screening programme (250 control-and 245 cancer cases). These mammograms were read by 15 experienced radiologists who were blinded to outcome. Results were used to calculate the relationship among recall rate, detection rate and numbers of false-positives. Results: Increasing the recall rate from 1% to 2% would increase the detection rate from 4.2% to 4.5% due to the earlier detection of interval cancers. Moreover, 0.5%of the screen-detected cancers would be detected 2 years earlier (late-detected cancers). For each 1% incremental increase in recall rate above 5%, the detection rate would increase by approximately 0.03% only, with positive predictive values decreasing to below 10%. Conclusion: Breast cancer can be detected earlier by lowering the threshold for recall, especially for recall rates of 1% -4%. To benefit most from a higher recall rate, radiologists should be able to identify the cancers with subtle mammographic features, which are often based on temporal changes. [JNCI 2005; 97:748] Purpose: We aimed to investigate the hypothesis that interval cancers arising soon after the previous screen and true interval cancers are biologically aggressive and have a relatively poor prognosis. We also wished to assess which prognostic features are relevant to interval cancers. We have analysed prognostic pathological features (grade, lymph node stage, size, vascular invasion, oestrogen receptor (ER) status, and histological type), radiological features (comedo/non-comedo calcification and spiculation) and survival for 538 invasive interval breast cancer cases by subtype and time since previous screen. Results: Women presenting in the first year were more likely to be lymph node positive (47%vs40%, p = 0.02). Women presenting with true interval cancers were more likely to be grade 3 than minimal signs and false negative intervals (52 vs35%, p = 0.01). False negative cancers were more likely to have lobular features than other intervals (47%vs20%, p < 0.0001). There was no significant survival difference by classification sub-group (p = 0.64) or interval year (p = 0.83). At univariate analysis of all interval cancers; tumour size, grade, nodal stage, ER status, vascular invasion and comedo calcification correlated with survival. On multivariate analysis of prognostic features significant at univariate analysis, nodal stage (p-value = 0.009), tumour size (p = 0.004), ER status (p < 0.0001), and vascular invasion (p < 0.0001) were independently significant prognostic features. Conclusion: Our study shows that true intervals and interval cancers arising quickly after screening do not have a worse prognosis and that interval cancers have a unique set of prognostic features. The Purpose: It is known there is an excess of breast cancer in women with radiographically dense breasts. Previous studies have shown poorer prognostic features in these tumours. No study has looked at survival. We have classified the background pattern (BI-RADS system) of 759 screened women who presented with a screen detected or interval invasive breast cancer. Pathological details (tumour size, grade, stage, vascular invasion and histological type) and mammographic appearances were recorded. Breast cancer specific survival was ascertained with a mean follow-up of 9 years. Results: There was an excess of interval cancer in women with dense breasts (p-value < 0.0001). Screen detected (but not interval) tumours were significantly smaller in fatty breasts (p = 0.014). Tumour grade, stage, vascular invasion and histological type did not vary significantly with mammographic background pattern in screen detected or interval cancers. Screen detected cancer in fatty breasts were more likely to appear as ill defined (p = 0.003) or spiculate masses (p = 0.002) in contrast to cancer in dense breasts which more commonly appear as distortions (p < 0.0001). However there was no significant breast cancer specific survival difference by mammographic background pattern for screen detected cancers (p = 0.75), interval cancers (p = 0.82), or both groups combined (p = 0.12). The prognosis of screened women presenting with breast cancer is unrelated to dense mammographic background pattern despite an excess of interval cancers and larger screen detected tumours in this group. This data supports the mammographic screening of women with a dense background pattern. Delay in breast cancer diagnosis in the Netherlands: Performance of hospitals involved in the evaluation of women with a positive mammographic screening examination F.J. Nobrega 1 , L.E.M. Duijm 1 , J.H. Groenewoud 2 ; 1 Eindhoven/NL, 2 Rotterdam/NL (LEMDuijm@hotmail.com) Purpose: To determine the inter-hospital variability in the diagnostic delay of breast cancer after breast cancer screening. We included all women aged 50-75 years who underwent biennial screening mammography in the southern breast cancer screening region of the Netherlands between January 1, 1996 and July 1, 2003. Clinical data, breast imaging reports, biopsy results and breast surgery reports were collected of all women with a positive screening result with a minimum of 2-year follow-up. Results: Of 196,317 mammographic screening examinations, 2,106 (1.1%) were positive screens. Seven regional hospitals performed the workup of 2,097 women with a positive screen examination and complete follow-up. Breast cancer was diagnosed in 992 women. A total of 58 (5.8%) women with a true positive screen experienced a delay in breast cancer diagnosis exceeding three months. This delay was due to erroneous mammographic interpretation of suspicious lesions as benign or probably benign lesions (40 cases), benign biopsies from a malignant lesion (13), and omission to biopsy or remove a lesion, that was suspicious at breast imaging (4) or core biopsy (1). The proportion of women who experienced a delay exceeding three months in breast cancer diagnosis after a true positive screen ranged from 0-9.4% between the seven hospitals. Conclusion: Hospitals that are involved in the workup of women with a positive screen examination show a large variation in the percentage of patients who experience a delay in breast cancer diagnosis. A majority of the delays are caused by radiological misinterpretation of suspicious lesions. Role of routine breast ultrasound in women with negative radiologically dense mammography S. Ciatto 1 , V. Corsetti 2 , A. Ferrari 2 , M. Ghirardi 2 , R. Bergonzini 2 , S. Bellarosa 2 , O. Angelini 2 , C. Bani 2 ; 1 Firenze/IT, 2 Calcinato (BS)/IT (s.ciatto@cspo.it) Breast ultrasound (US) was performed routinely in 6,449 of 17,883 (36.5%) subjects with mammography reported as negative and radiologically dense (BI-RADS D3-4). While 138 cancers had been detected out of 257 suspicious mammographies, 29 additional cancers were detected (overall detection rate: 0.44%, or 17.3% of total cancers detected) at US of D3-4 mammography negative breasts: 25 of 29 US detected cancers were available for internal and external review, and negative mammography and asymptomatic status was confirmed in 15 (detection rate: 0.23%, or 8.9% of total cancers). Cancer detection rate was 0.11%, 0.22%, 0.32% and 0.14 % in age groups < 40, 40-49, 50-59 and > 59, respectively. The cost per additional carcinoma detected by additional US alone was € 25,847.85, whereas that per examined woman in the total series was € 21.68. The study confirms that US may detect breast carcinoma occult to mammography in dense breasts. Internal and external review of mammograms, which is not commonly adopted in not controlled retrospective studies, is important to confirm that mammograms were really negative, and to avoid overestimation of US benefits. Our data are not sufficient to support a general recommendation of adding US to mammography in dense breasts, which should be based on the results of ongoing randomized trials (ACRIN, RIBES). Nevertheless we confirm that adding US in dense breast may be useful in the current practice, though associated with substantial costs. Purpose: The aim of our study was to evaluate the advantages of reporting VC using a colon CAD added to primary 2D approach compared to simple primary 2D approach. Methods and Materials: Three radiologists with different levels of experience (expert, intermediate, non-expert) blindly evaluated 50 cases chosen from a standard population of FOBT+ subjects. A primary 2D approach was performed on a Vital Images workstation equipped with a Vitrea 3.7 software; 3D reconstruction was performed for problem solving. A further 2D approach with the help of a CAD system was performed on a Medicsight workstation equipped with Colon CAR 1.3 software. We evaluated per patient accuracy, sensitivity and specificity for each radiologist as well as mean reporting time. Results: No differences in accuracy, sensitivity and specificity for expert and intermediate radiologists using the two approaches were detected. Improvement in accuracy (94% only 2D approach, 100% for 2D + CAD) and sensitivity (80% only 2D approach, 100% for 2D + CAD) with the same results for specificity (100%) was revealed in the non-expert radiologist. Mean reporting time was 13 min for primary 2D approach and 7 min 30 sec for 2D + CAD system. Conclusion: In our preliminary experience 2D + CAD approach does not improve accuracy, sensitivity and specificity for expert radiologists, but it reduces mean reporting time. Accuracy and sensitivity improve for the non-expert radiologist. Purpose: Can CAD-systems detect mammographically non-detected malignant breast lesions on priors and if so, how early? Methods and Materials: All available prior mammograms (PM) of patients suffering from breast cancer later detected at our department were retrospectively analyzed by CAD and by radiologist with full knowledge (and availability) of the malignancy containing mammogram (MCM), that yielded cancer detection. 278 mammograms (97 patients) were analyzed in consensus in order to verify whether a lesion was already visible earlier. Images were analyzed using Second Look 6.0 (iCAD) and correlated to MCM. The earliest mammogram with the correct C D E F B A G detection of the (later diagnosed) malignancy was used for statistical analysis. Time intervals PM-MCM was scored: G1:< 0.5; G2:< 1y; G3:< 1.5y; G4:< 2 y; G5:< 3y; G6:> 3y. Results: 53/97 and 17/97 patients had a retrospectively visible mass and microcalcification, respectively. 30/53 masses were detected by CAD in 1 and 4 in both views (64%). 8/17 MC were detected in 1 and 7/17 in both views (88%). Time interval vs. detection rate for masses was G1: 10/14 (71%), G2: 4/7 (57%), G3: 3/ 5 (60%), G4: 5/7 (71%), G5: 3/9 (33%), G6: 9/11 (81%); for microcalcifications G1: 1/2, G2: 3/4, G3: 1/1, G4: 2/2, G5: 2/2, G6: 6/6. Conclusion: About half of the initially non-detected cancer containing mammograms had a mammographically visible lesion, being not suspicious according to BI-RADs. CAD detected the majority of such MC and masses. Earliest detections of later diagnosed malignancies were dated more than 3 years prior to the radiological detection. Thus, CAD might be a promising tool in this application. Clinical role of a CAD-system for dynamic analyses of breast lesions in MR imaging A. Malich 1 , K. Huskobla 2 , J. Boettcher 2 , D.R. Fischer 3 , A. Hansch 2 , R. Gorna 1 ; 1 Nordhausen/DE, 2 Jena/DE, 3 Berne/CH (ansgar.malich@med.uni-jena.de) Purpose: Breast MR-CAD systems automatically analyze and colour-code contrast uptake of breast lesions. This was compared to ROI-based radiological dynamic calculation. Methods and Materials: 110 patients undergoing breast MR imaging (6/2003-12/2003 ) with histopathological verification were included. Wash in was differentiated into < 50% (CAD: No colour coding), 50-100% and > 100% uptake; later enhancement type into continued enhancement, plateau and wash out, as determined by 3 radiologists and compared to CAD and histopathology. Worst curve evaluation analysis by CAD (Confirma, USA) was used. Results: 11/110 examinations were not transferable to CAD (inoptimal data storage, incomplete data). Out of 117 malignant and 58 benign lesions enhancement was judged to be congruent in 71/117 and 23/58 lesions regarding wash in and 74/117 and 20/58 regarding later enhancement type (radiological interpretation vs. CAD result). Radiologists scored enhancement of 5 malignancies/DCIS as unsuspicious, CAD only 1 (thus CAD detected 4 more malignancies/DCIS). Benign lesions were scored as rapidly enhancing in 20 cases (radiologist) and 19 cases (CAD); wash out rate was 9 (radiologists) and 19 (CAD) (thus CAD detected 10 more false positive cases). Using updated software transferability increased, volume calculation was possible in all except one lesion, and dynamic coding was similar; no miscalculations of cardiac enhancement occurred. Conclusion: CAD offers for the very first time a reliable, automated analysis of enhancement types within a breast lesion. Analysis of enhancement types is not always congruent among radiologists and radiologists vs. CAD, suggesting a higher rate of pathological enhancement types in benign lesions using CAD; CAD is however more sensitive for any enhancement, including that of premalignancies. Purpose: Malignant melanoma is a tumor with rising incidence, frequently associated with multiple lesions throughout the body. Therapy monitoring requires reliable imaging and post processing methods for measuring tumor burden and therapy response. The RECIST criteria are the current standard for therapy monitoring of solid tumors, with inherent disadvantages. This pilot study evaluates the feasibility of volumetry as a more accurate measurement. Methods and Materials: Multislice-CT was performed in 25 patients with metastatic malignant melanoma, covering neck, chest, abdomen and pelvis with 1 mm collimation using a standardized imaging protocol. Post-processing was performed using semi-automated 3D-volumetric analysis software (OncoTreat, MeVis, Germany). Suspicious lymph nodes (LN), (n = 120) were evaluated by two independent readers using volume [ml] , time [s] , quality of segmentation [unacceptable-excellent] and number of corrections [n] . Additionally, 20 LN were segmented manually. The software performance allowed a reliable and faster volumetry than manual segmentation. Quality of segmentation was rated acceptable -excellent in 81% by reader one and 79% by reader two, respectively. The correlation of the volume was highly significant between both readers (r = 0.88). 15-20% of the LNs had to be manually corrected by drawing representative ROIs. The average time for automated segmentation per LN was 70-100s, including the manually corrected ones. Average segmentation time for the primary manual segmentation was 180-200s per LN. Conclusion: Semi-automated 3D-volumetric analysis allows a reliable, fast and convenient segmentation of lymph nodes and should be further evaluated for monitoring tumor burden in a larger patient population. Purpose: To develop a prototypical software assistant for the volumetric measurement of metastases and tumors in lung, liver, brain and lymph nodes to support planning and monitoring of chemotherapy. Methods and Materials: Hybrid segmentation algorithms have been developed that make use of contrasts in density as well as of morphology information to segment the different types of lesions. For follow-up examinations the task of locating the corresponding lesion in the new scan is simplified by automatic matching of corresponding lesion positions. Additionally, all information about the processed lesions is summarized in a report, which includes a table containing volumes, volume growth and doubling time for each lesion. Results: A reproducibility study was performed with 96 lung metastases (minimum diameter 4.6 mm, average 11.1, maximum 59.6) and showed a low interobserver variability of the volumetry (median: 0.1%) and a low inter-scan (i.e. volume of the same lesions in different scans obtained within a few minutes and measured by different observers) variability (median: 4.7%). A second reproducibility study was carried out with 86 liver metastases processed by 3 radiologists with the developed application and showed a low median inter-observer variability of 7 %. The average processing time for each lesion is about 1-2 sec. Conclusion: The proposed software tool is able to measure oncological lesions in lung, liver and brain as well as enlarged lymph nodes in the same application, and has the potential to improve treatment planning and monitoring with respect to accuracy, robustness, speed, and convenience. Neural network based segmentation of multispectral MR imaging data for brain volume quantification in patients with multiple sclerosis A. Wismueller 1 , A. Meyer-Baese 2 , J. Behrends 1 , M. Blazevic 1 , D. Auer 3 , M.F. Reiser 1 ; 1 Munich/DE, 2 Tallahassee, FL/US, 3 Nottingham/UK (axel@wismueller.de) Purpose: In the literature, the percentage of brain volume (PBV) has been identified as an important marker of disease progression in multiple sclerosis (MS). The aim was to develop, test, and evaluate a neural network-based quantification system for PBV by computer-assisted segmentation of multispectral MR imaging data. Methods and Materials: MR examination was performed in six women with relapsing-remitting MS (aged 23-46 years, 1.5-T system, T1w±C, T2w/PDw SE, FLAIR, MT±C). After preprocessing (image registration, automatic inhomogeneity correction), PBV was computed by automatic CSF segmentation using a threelayer feed-forward radial-basis-functions (RBF) neural network: The voxel-specific gray-level intensity spectrum forms a 7-dimensional feature vector which is classified by the neural network as belonging to CSF or not. As a reference, PBV results were compared to the conventional angle-image method which only uses T2w and PDw data. For quality assessment w.r.t. observer-induced effects, training data for the neural network were defined by two different observers independently. The same holds for threshold definition using the angle-image method. Results: Neural network-based PBV computation outperformed the conventional angle-image method as expressed by inter-/intra-observer agreement indices (British Standards Institution) of 0.971/0.982 for RBF-neural network compared to 0.937/0.952 for angle-image. The method-induced differences proved to be statistically significant (univariate F-test, N=6, p < 0.01). The neural network specifically performed better by avoiding misclassifications in white-matter-lesions that can hardly be distinguished from CSF by retrieving only T2w and PDw signals. Conclusion: Computer-assisted machine learning based PBV quantification using RBF neural network segmentation based on multispectral MR imaging data yields competitive results w.r.t. inter-and intra-observer agreement levels. 1.5 T and 3.0 T and to correlate these measurements with imaging peculiar parameters like SNR and contrast quality. Methods: Twelve calcaneus were imaged at 1.5 T and 3.0 T with a gradient echo (Flash) sequence and with a spin echo (TSE) sequence, using identical sequence parameters. Relaxation times were measured. SNR in corresponding sections of the calcaneus were evaluated and correlated with relaxation times (T 1 , T 2 , and T 2 * ). Results: Qualitative analysis showed a better contrast at 3 T than at 1.5 T. Quantitative analysis demonstrated that magnetic susceptibility effects reduce the T 2 * and, because of diffusion effects, also reduce the apparent T 2 relaxation time. Conclusion: Our results explain quantitatively the SNR gain obtainable in Flash and TSE images of trabecular bone at 3 T with respect to 1.5 T. Generally, in soft tissues like brain tissue, the SNR gain at 3 T with respect to 1.5 T is about 2, while in trabecular bone tissue is less than 30% in Flash images. This is because of the internal gradient effects due to magnetic susceptibility difference between trabecular compact bone and bone marrow in spongy bone. Comparison between digital radiogrammetry (DXR), dual energy X-ray absorptiometry (DXA) and peripheral quantitative CT (pQCT) in children with chronic inflammatory bowel disease H.-J. Mentzel, J. Blume, D. Tuchscherer, G. Lehmann, J. Böttcher, E. Kauf, W.A. Kaiser; Jena/DE (hans-joachim.mentzel@med.uni-jena.de) Purpose: Patients with chronic inflammatory bowel disease are at increased risk of developing osteopenia and osteoporosis. The aim of this study was to evaluate a radiogrammetrically based bone densitometric technology on children and adolescents. Methods and Materials: 25 patients (11 female, 14 male; 10y -18y) suffering chronic inflammatory bowel disease were included. Bone mineral density (BMD) was estimated from a plain radiograph of the non dominant hand using the Pronosco X-Posure System for Digital radiogrammetry (DXR). DXR-BMD and DXR-MCI were evaluated at the three middle metacarpals. BMD was also evaluated by DXA (Hologic, QDR 4500 A) at lumbar spine and femur and by pQCT at the distal radius (Stratec, XCT 900) . Spearman correlation between the results of the different methods was performed for statistical analysis. Results: There was a significant correlation (p < 0.01) between DXR-BMD and DXA-BMD at the lumbar spine (r = 0.79) and DXA-BMD at the femur (r = 0.56). DXR-MCI correlated significantly (p < 0.05) with DXA-BMD at the lumbar spine (r = 0.75) and DXA-BMD at the femur (r = 0.50), and cortical BMD at the distal radius (r = 0.43). Conclusion: This study suggests that digital radiogrammetry shows a significant correlation to the standard method DXA. Practicable, non-invasive DXR-technology may be a possible adjunct to give information about the development of the individual BMD and risk of fracture in patients with chronic inflammatory bowel disease. Interdisciplinary approach of kyphoplasty in fresh and traumatic fracture: The Heidelberg concept G. Nöldge, M. Libicher, K. Da Fonseca, I. Grafe, C. Kasperk, G. Kauffmann, P.-J. Meeder; Heidelberg/DE (gerd_noeldge@med.uni-heidelberg.de) Purpose: To evaluate long-term results (up to 3 years) for pain relief and stabilization of traumatic and osteoporotic fractures treated with CPC and PMMA by kyphoplasty. Methods and Material: Between June 2002 and September 2005, 355 patients ranging from 18 to 89 years (mean: 64.8 yrs) were treated. Results: CPC (Calcibon) was used in 103 and PMMA in 252 patients. 3.7 -9.9 ml of bone cement was used in 585 vertebral bodies. 92 fresh fractures (osteoporosis /trauma) and 8 unstable fractures (internal fixation prior to kyphoplasty) were treated. Good functional results (EVOS) of the treated segments with pain relief were achieved in nearly 93% 48 hours after the procedure and could be maintained in about 70% for 12 months and in 48% after 36 months under permanent accompanying medical antiosteoporotic therapy. Vertebral augmentation of the fracture could be achieved between 1 -3 mm in about 65% and in fresh traumatic fractures up to 4 mm. Recurrence of pain occurred in about 30% (12 months) and in 65% (36 months) was due to new fractures of neighbouring vertebral bodies. Cement leakage could be visualized in 18.5% without neurological symptoms. Conclusion: The Heidelberg interdisciplinary concept of more than 3 years experience seems to be very convincing for the treatment of osteoporotic and traumatic fractures with very good initial results in terms of stabilization, pain relief, and functional and clinical outcome. (CT) and magnetic resonance imaging (MRI) were performed at baseline. MR imaging was performed at least twice during the 12 months treatment and at the end. MR imaging and CT were reviewed by a senior musculoskeletal radiologist. Morphological changes of abscesses, signal intensity changes of vertebral bodies and discs were analysed and recorded. Results: Initial MR imaging showed oedema of 75 to 100% of the vertebral body in all cases. All patients improved clinically except one with cervical dislocation, who died immediately after surgery. Weight gain was almost totally achieved at 6 months and pain relief at 9 months. Biological inflammatory syndrome vanished at 3 months. On MR imaging, paravertebral abscesses disappeared respectively at 3, 6 and 12 months in 24%, 35% and 88% of the cases. Improvement of vertebral body bone marrow oedema with conversion to a fatty signal occurred in 30% of cases at 6 months and 53% at 12 months. Epidural abscesses disappeared at 9 months in 100% of cases. Conclusion: Significant imaging lesions persist at the end of treatment despite a good outcome. Therefore MR imaging should not be used to define the duration of therapy if clinical and biological evolution are favourable. 3D-CISS MR myelography for assessment of adult traumatic brachial plexus injuries M.A. Png, F.C. Yong; Singapore/SG (gdrpma@sgh.com.sg) Purpose: Differentiation of pre-ganglionic from post-ganglionic brachial plexus injuries (BPI) is necessary for adequate pre-operative planning. This study reviews our experience using MR myelography, based on a 3-D constructive interference in steady state (CISS) technique, to assess for pre-ganglionic injuries in clinically suspected BPI. We retrospectively reviewed the 3D-CISS myelography scans of 9 patients with suspected traumatic BPI who underwent MR imaging before surgical exploration. All scans were performed on a 1.5 T Siemens Vision and were reviewed by an experienced musculoskeletal radiologist. The status of the C5 to T1 roots on every MR myelogram was evaluated for: intact rootlets, nerve root avulsion or injury (partial or complete), and pseudomeningocele formation. The MR imaging findings were recorded and correlated with the findings at surgical exploration for each root level. Results: Altogether 35 roots evaluated had surgical correlation. For detection of root avulsion injury, the overall sensitivity and specificity was 100% and 71% respectively, with positive and negative predictive values of 84% and 100%. For complete avulsions, the sensitivity and specificity was 88% and 71% respectively. Twelve of 28 (43%) proven root avulsions had no associated pseudomeningocele on MR myelography. In this study, all cases with pseudomeningoceles detected by 3D MR myelography, had root avulsions found at surgery. Conclusion: Our experience shows that 3D-CISS MR myelography is accurate for detecting root avulsions in adult traumatic BPI. The ability of this technique to demonstrate root avulsions in the absence of pseudomeningoceles significantly improves the accuracy of MR imaging in detecting pre-ganglionic injuries. 14:00 -15:30 Room C The Health Optimum project: The standards application to telemedicine workflow C. Dario 1 , S. Giovannetti 1 , G. Bettiol 1 , A. Ramelli 2 , C. Saccavini 2 , F. Di Paola 1 ; 1 Treviso/IT, 2 Padova/IT (aramelli@rad.unipd.it) Purpose: Health Optimum is a telemedicine project to validate systems interoperability and the management of patient documents (creation, digital signature, transmission and publication in EHR), and to disseminate the results. Methods and Materials: Health Optimum, approved and co-funded by the European Community, began in May 2004 and will end in January 2006. It involves Spain-Aragona (oncology, haematology, radiology, dialysis and laboratory), Denmark-County of Funen (cardiology and endocrinology) and Italy-Veneto region as coordinator (12 healthcare structures for neurosurgical counselling, POCT-laboratory and regional EHR), with other technological and organisational partners. Neurosurgical tele-counselling wants to satisfy the need to give fast and formalized answers for urgent cases of cranial and vertebro-medullary trauma, linking central neurosurgeons and neuroradiologists to peripheral ER, neurology and intensive-care. The form management system is based on XML-CDAv2 and HL7 standard; digital signature follows law and rules; CT images are in DICOM format. To share telemedicine documents we choose the IHE XDS profile (Cross Enterprise Document Sharing). To manage security we use IHE ATNA profile (Audit Trail Node Authentication) and XUA profile (Cross Enterprise User Authentication). Results: Physicians have reorganized workflow using a telemedicine approach, and they have defined a standard clinical data set. This demonstrates that the avoidance of unnecessary transfer of patients and documents, and the use of EHR speed up diagnostic and therapeutic decisions. Conclusion: Telemedicine allows reorganization and optimization of processes and resources, with clinical and technological standardization. National and international rules and law, medico-legal liabilities, security and privacy must be defined and respected. Purpose: Increasingly, radiology studies are distributed on compact discs (CDs) to referring physicians for review; CDs are cheaper and a potentially better option than producing X-ray film, however the software viewing interfaces used differ considerably. Therefore, we analyzed the most common image viewing software programs' functionality and ease of use. We prospectively analyzed 250 consecutive CDs sent to our Institution for review over a one month period in 2005. Functionality was assessed by analyzing 5 basic viewing tasks 1) open a case, 2) navigate through series, 3) change window/level settings, 4) magnification, 5) open and compare studies. The 5 most commonly used image viewers accounted for 193 (77%) of all CDs submitted. In descending order of frequency: 42%, 19%, 9%, 5%, 2%. Fifteen (6%) of CDs had no viewer or could not be opened. The opening display formats ranged from 1 to 16 images. Left and right mouse button functions varied in 4 of 5 viewers. Image navigation steps were different in 5 of 5 viewers. Pre-set window settings were found on 2 of 5 viewers. Objective global functionality scores ranged from 26 to 38, mean 32 (out of 50) for the 5 most common viewers. Conclusion: There is great variability in commonly used functions among image viewing programs. Additionally, many programs are not intuitive and are difficult to use. This variability, lack of standardization, and poor functionality hampers a physician's ability to assess radiology studies, interrupting workflow and routine clinical care. ASP for managing the PACS archive E. Talini, A. Bertini, G. Cardia, R. Dell'Osso, D. Caramella, C. Bartolozzi; Pisa/IT (e.talini@med.unipi.it) Purpose: Picture Archiving and Communication Systems (PACS) have proved able to assist radiologists by streamlining the digital workflow. However, the management of PACS complex technology may cause additional strain to the hospital personnel. Application Service Provisioning (ASP) may be a solution to this issue, by means of the total or partial outsourcing of PACS functions. At our Institution we evaluated the feasibility of using ASP for managing the PACS archive. Methods and Materials: Four high-volume DICOM modalities were configured to send diagnostic exams to a cache server (Centricity Archive, GEMS) that forwarded them to a remote GEMS PACS by using a broadband wide area network. In order to evaluate the performance in different clinical and networking situations we monitored the system when up to 14 reporting workstations were concurrently used and by setting the network speed at 2 Mbps, 4 Mbps, 6 Mbps, 10 Mbps, and 100 Mbps. Results: Our assessment was done on the basis of objective measurement and of subjective evaluation collected by means of detailed questionnaires completed by radiology residents. We found that with a network speed of 6 Mbps the number of concurrent reporting sessions ceased to have an impact on the delay in retrieving exams from the remote ASP archive. Conclusion: Our experience suggests that the PACS archive can be managed with ASP. We found that no significant impact is perceived on the radiological workflow, provided that a guaranteed bandwidth of 6 Mbps is available. C D E F B A G Increased quality of patient care and balanced workload for technologists: Leveraging industry tools to analyze and optimize patient flow in a full digital regional hospital H. Bartel 1 , E. Pisar 1 , A. Mueller 2 , P. Ahrens 1 , S. Liedke 1 , L. Mikolajewski 1 , K.-H. Zerbst 2 ; 1 Düsseldorf/DE, 2 Stuttgart/DE (dr.pisar@kaiserswerther-diakonie.de) Purpose: Analyze "current state" and identify waste in target process. Develop "future state" vision and implement optimized process within one week. Methods and Materials: Florence Nightingale Hospital in Kaiserswerth (Germany) has a complete digital environment (PACS/RIS/HIS). However, patient waiting time to perform X-ray procedures was excessive and technologists' workloads were not balanced. In a "Gemba" session workflow specialists observed the situation, then analyzed RIS data and developed the "Current State Value Stream Map" and the "Future State Map "necessary to create the new process. "Target Sheets" showing major measurements/targets gave full visibility to the team. After training "on the job" a KAIZEN was developed, piloted and deployed by the radiology staff, implementing the new process. Results: Thanks to the "KAIZEN" tool, changes were implemented in one week and improved performance is constantly monitored. The team moved from a "batchordering" process via "intercom" to a "single-piece-flow" of patients scheduled and picked-up by staff. Patient waiting queue was reduced from 3 to 1, waiting time was significantly decreased; cycle-time was reduced by 60%, resulting in a potential 50% patient throughput increase as technologists workload was balanced. Results were obtained with no investment in additional resources/technology; on the contrary, future department expansions may leverage the freed resources. Conclusion: Significant improvements can be implemented quickly provided staff understand and share the objectives. The KAIZEN concept was taught, understood and leveraged by staff to define, test and implement the new process directly "on the job", which was successfully "up and running" within one week. Introducing a redundant mini-PACS as a backup and migration tool in an already existing large PACS environment U. Fronz, M. Nuefer, M. Forsting; Essen/DE (uwe.fronz@uni-essen.de) Purpose: Beginning in April 2002 a PACS was introduced into our radiology department, supporting our already existing RIS. In March 2004 we installed a smaller system as a redundant mini-PACS to support backup and upgrade needs in the large PACS system. The purpose of this study was to evaluate the advantages and disadvantages of such a system in a large radiology department. Methods and Materials: Since 1999 we have used Medora from Innomed, Germany, as our RIS. In December 2001 we decided to introduce PACS Centricity from GE Medical Systems, USA, into our department. In March 2004 we introduced a second smaller system for backup and upgrade purposes in the original PACS. This smaller PACS consists of an EMC CX200 computer system equipped with 1.2 TB storage capacity in RAID technology. Based on the mini-PACS, reports can be generated and web-distribution of images and reports is possible. Results: We already used this system for two migration steps to upgrade the normal PACS system. We also used it in one system failure with a downtime of 3 hours. The mini-PACS proved to be fully functional and allowed the radiology department to examine patients, generate reports and distribute the images and reports in the same way as with the normal PACS system. Conclusion: Running the redundant mini-PACS in a large scale environment proved to be feasible and became useful in case of errors or upgrades within the normal PACS system. We think that a filmless hospital needs to maintain a backup and migration system. Usability of speech recognition in a trauma hospital: One year follow-up study T. Kauppinen, J. Ahovuo; Helsinki/FI (tomi.kauppinen@hus.fi) Purpose: It has been demonstrated earlier by several study groups that speech recognition (SR) reduces the time between imaging and final report. Because SR affects the whole radiological process and turnaround, it would be important to total the number of reports which are created by SR. The purpose of this work was to evaluate usability and usage of SR in one radiological pilot unit, and try to find reasons why the number of reports created by SR is not 100%. The aim was also to follow and collect experiences during this follow-up study. The radiological dictation process was followed in Töölö trauma hospital in Helsinki, Finland, which has emergency unit working 24/7. At the beginning of this follow-up study three radiologists started to use SR and at the end of follow-up time all senior radiologists are able to dictate using SR. Results: It was found by this study that the attitude of radiologists affects the usage of SR. At first the proportion of reports which were dictated by SR compared to the conventional way was 0%, after 2 months use by the first three users the proportion was over 50 %, and after 3 months over 60%. Conclusion: It was noticed that SR changes workflow and increases the quality of the reports. The proportion of reports created by SR increased after the first months. The total number of workstations, usability and dictation environment have an impact, increasing the usage. Impact of PACS/RIS integrated speech recognition on report availability in a three-year follow-up study C.G. Trumm 1 , C. Glaser 1 , C. Grosse 1 , B. Küttner 1 , S. Nissen-Meyer 1 , P. Popp 2 , M.F. Reiser 1 ; 1 Munich/DE, 2 Erlangen/DE (christoph.trumm@med.uni-muenchen.de) Purpose: To assess the long-term effect of a PACS/RIS integrated speech recognition system on radiology report availability (RA) in a university hospital over a 3-year period. Methods and Materials: In a retrospective PACS/RIS database analysis, report turnaround times (RTT) were calculated for 3 comparable 0.5 year time periods: 1. 01-06/2003 (tape-based dictation with transcription), 2. 01-06/2004 (0.5 year experience with speech recognition system [SRS]), 3. 01-06/2005 (1.5 year experience with SRS). Only remote (RSR) and online (OSR) speech recognition with transcriptionist and self-correction respectively were used in period 2 and 3. Cumulative RA (ratio of available reports vs. total number of examinations) within 2/6/12h after examination completion was calculated for period 1/2/3 respectively including all modalities. For period 3, cumulative short-term RA was differentiated into reports available 10/30/60 minutes after completion of conventional radiography (CR), computed tomography (CT) and ultrasound (US) examinations. Results: 33155/45136/60209 datasets were analyzed for time period 1/2/3. Including all modalities, the number of reports available in less than 24h was 3796/ 13125/36031 respectively. Within 2/6/12h after examination completion cumulative RA was 2.3/7.9/11.2% for period 1, 14.9/24.5/28.3% for period 2 and 31.6/ 54.5/59.1% for period 3. Cumulative short-term RA after 10/30/60 minutes was 4.7/19.3/23.8% for CR, 1.3/12.2/18.7% for CT and 6.2/31.3/47.3% for US examinations. Conclusion: Institution-wide use of a PACS/RIS integrated SRS facilitates a continuous improvement of report availability over a 3-year period. Reasons are an increasing acceptance of OSR by radiologists and a shift from RSR to OSR for dictation of routine reports. The status of structured reporting in imaging: Special focus on remote reporting H.K. Pohjonen 1 , P. Ross 2 , J.G. Blickman 3 , L. Christensen 4 ; 1 Espoo/FI, 2 Tallinn/EE, 3 Nijmegen/NL, 4 Svendborg/DK (hanna.pohjonen@rosalieco.fi) Purpose: Structured reporting has been studied for at least 10 years to increase report creation speed and the clarity, communicative value, and overall quality of the imaging information supplied for a modern hospital. Today, the digital workplace knows no boundaries: Remote reporting, out-of-hours coverage and the 'virtual' radiologist are all ways to offer imaging services in a non-traditional way. The purpose of the paper is to study the status of structured reporting with a special focus on cross-border eRadiology. The presence of lists and hierarchical relationships, coded or numeric content in addition to plain text, relationships between concepts and embedded references to images and similar objects is characteristic of structured reporting. The study involves a survey on commercially available report templates in general, and especially those supporting DICOM structured reporting. Special attention is paid to templates in multilingual reporting across borders. Results: To our knowledge there are no commercially available clinical templates that support DICOM structured reporting. A few companies exist -especially in North-America -that offer structured reporting as a service, but these are not implemented in DICOM. A few research-oriented projects do exist however that address structured reporting in the DICOM world. The Baltic eHealth project uses structured multilingual templates in remote reporting between Denmark, Estonia and Lithuania. for remote reporting across borders. It enables the utilization of radiologist manpower from different countries, while using their own native language. Commercial applications to support these tools, however, are lacking. Taking advantage of PACS/structured reports resources to generate teaching files A. Delistamatis 1 , M. Mantatzis 1 , N. Tapus 2 , P. Prassopoulos 1 ; 1 Alexandroupolis/GR, 2 Bucharest/RO (mmantatz@med.duth.gr) Purpose: To develop a web-based system for teaching files (TF) creation using information derived from DICOM structured reports (SR) documents and DICOM images from a PACS. Methods and Materials: The TF system retrieves all the information from diagnostic reports, in the form of SR documents, from a PACS to store it in a local database via a synchronization procedure. Queries are applied to the system local database using keywords related to specific pathology, imaging findings, diagnostic procedures or techniques for the selection of cases with educational value. The system automatically retrieves from the PACS the images corresponding to the selected cases/group of cases. Representative images along with the diagnostic reports, commentaries and MESH-keywords are edited to form the teaching file that is stored in the final teaching file database. Results: Review of teaching files is included in every-day training of medical students and radiology residents accessing the TF system database through a web-interface in the university intranet. In addition, the system incorporates an interface for possible integration with the RSNA MIRC web network. The TF system expands by 2-3 new teaching files per day generated by attending radiologists. Conclusion: The proposed TF system offers a platform for teaching files creation utilizing information already stored in a PACS. A filmless radiology department in a full digital regional hospital: Quantitative evaluation of the increased quality of patient care A. Nitrosi 1 , G. Borasi 1 , F. Nicoli 1 , A. Botti 1 , G. Modigliani 2 , P. Notari 1 , E. Costa 1 , S. Bronzoni 1 ; 1 Reggio Emilia/IT, 2 Cinisello Balsamo/IT (andrea.nitrosi@asmn.re.it) Purpose: To quantitatively measure, assess and document the business and organizational benefits that result from the transition to digital imaging. Methods and Materials: Reggio Emilia regional hospital installed PACS as the final step towards a completely digital clinical environment, completing the HIS/ EMR and 1200 web/terminals for patient information access. Financial benefits throughout the hospital were assessed upfront and measured every 4 months after the system's deployment. Key indicators (radiology exam turn-around-time (TAT), number of radiology procedures performed, inpatients length of stay before and after the PACS implementation, etc). were analysed and values were statistically tested to assess workflow and productivity improvements. All raw data used for the analysis was extracted from the hospital HIS and finance systems. Results: The hospital went "filmless," with the exception of analogue mammography in 28 weeks and within its original budget. Overall radiology department productivity increased by 15%, TAT improved by more than 65%. Timelier patient care resulted in decreased lengths of stay. Neurology alone experienced an 11% improvement in average patient stay. Annual financial upsides have exceeded €1.5 millions/year and PACS is being extended to 5 smaller county hospitals with a common clinical data repository that will be shared with the entire region next year. Conclusion: A well-planned PACS deployment simplifies imaging workflow and improves patient care throughout the hospital while delivering substantial financial benefits. Staff buy-in was key in this process and on-going training and process monitoring are a must. Purpose: In acute abdominal pain, plain abdominal radiographs are often obtained to identify bowel obstruction and free intra-abdominal gas. The purpose of this study was to compare low-dose computed tomography (CT) with plain abdominal radiographs in this setting, using the same radiation dose. The radiation risk for the patient is thus the same with both methods. Methods and Materials: Fifty-eight patients were imaged with both methods. The effective dose for plain abdominal imaging was calculated at 1 mSv with Monte Carlo simulation. Low-dose CT was performed at 120 kV and 20 mAs/ slice, giving the same effective dose. All cases were reviewed independently by three radiologists. A gold standard was obtained by reviewing medical records, laboratory data and further radiological imaging. Results: There was good agreement between the two methods for diagnosis of bowel obstruction. In three cases the cause of obstruction was found on lowdose CT (inguinal hernia, a low bowel obstruction and an abdominal wall hernia). Three cases were correctly diagnosed as having a diverticulitis and one as having a pancreatitis using low-dose CT while these diagnoses could not be made on abdominal radiographs. One case of inflamed bowel was diagnosed on plain abdominal imaging but missed on low-dose CT. One case of diverticulitis and one appendicitis were missed with both methods. No case of free intra-abdominal gas was found. Conclusion: Low-dose CT using the same radiation dose as plain abdominal radiographs is well suited as a replacement. The diagnostic potential is increased and more information is gained. The multislice CT-enteroclysis in bowel endometriosis detection E. Biscaldi, S. Ferrero, V. Remorgida, G.A. Rollandi; Genova/IT (ennio.biscaldi@fastwebnet.it) Purpose: To demonstrate the effectiveness of multislice CT-enteroclysis (MSCTe) in detecting bowel endometriosis wall involvement. Methods and Materials: 40 patients (aged from 25 to 46) with ovarian endometriosis already known (by MR imaging or laparoscopy) and symptoms suggesting colorectal endometriosis (pelvic pain, dyspareunia, tenesmus) were evaluated. The colon was distended with methylcellulose and water (50%) before the contrast enhanced MSCT (16 rows, collimation thickness 0.625 mm, reconstruction 1.25 mm, pitch 1,375, 120 KV, 330 mAs; smart mAs; 1.5 cc per Kg of patient weight of iodinated c.m. (Iodixanol 320 mg I/100 ml) at a flow/rate of 3 cc/s, with a delay time of 40s after the arterial peak). Within 20 days after the radiological examination, independently from the findings of MSCTe, all women underwent laparoscopy. Results: Out of 40 women evaluated, 3 were found to have a normal colon on MSCT: These findings were confirmed at surgery. In 37 cases the MSCTe detected the site of endometriosis and the colorectal wall involvement depth; the deeper nodules infiltrated the serosa in 13 cases, the muscularis in 17, the submucosa in 7. In one patient localisation in the last ileal loop was identified. One patient refused surgery. In 8 cases the CT detected some lymphoadenopathy near the colon. Results were compared with endoscopic data. MR imaging was performed after oral administration of a superparamagnetic contrast-agent, with colon air insufflation, using T2-weighted and T1-weighted sequences, before and after intravenous gadolinium injection on axial and coronal planes. MR images were evaluated by two radiologists blind of endoscopic results. The following parameters were assessed and graded (0-3); extent of lesions, bowel wall thickness, T1 wall gadolinium-enhancement, T2 wall signal and T2 signal outside of the wall. MR imaging score of UC activity ranged between 0-15. In particular; 0-1 non active, 2-4 moderately active, 5-10 active and 11-15 severe active pancolitis. Spearman nonparametric correlation was used for statistical analysis. To establish the efficacy of ultrasound of the abdomen as a diagnostic test in Typhoid Fever. The Widal test is the most commonly used method of detecting Typhoid fever, but does not provide results until a week after the onset of fever due to the need for enough antibodies to develop to render a positive result. Abdominal Ultrasound was performed within three days of the onset of fever in 80 cases suspected to be having Typhoid fever. Subsequently all 80 cases were found to be Widal positive. Subsequent follow-up scans were performed at five days, ten days and fifteen days. Results: The US findings were as follows: splenomegaly (n-80,100%); Bowel wall thickening (n-69, 86.25%); mesenteric lymphadenopathy (n-55.68.75%); hepatomegaly with normal parenchymal echotexture (n-22.27.5%); thickened gall bladder (n-53.66.25%); biliary sludge (n-8.10%); positive US Murphy's sign (n-14.17.5%); pericholecystic edema with increased vascularity (n-8 ,10%); mucosal ulceration in the wall of the gall bladder (n-1. Purpose: Quantitative perfusion techniques are advocated for assessing tumour perfusion, in particular to assess the effects of vascular modulating therapy. However, analysis methods are not standardised causing problems for comparison across studies. We aimed to determine the level of agreement between tumor blood volume and permeability measurements using two commercially available CT perfusion analysis methods. Methods and Materials: 20 patients (mean age 67.7 years; range 35 to 87 years, 10 male and 10 female) with proven colorectal cancer were examined prospectively using a four detector row CT (LightspeedPlus, GE Healthcare technologies). A 65-second perfusion study (4 X 5 mm slice thickness; 1 acquisition/second) was acquired through the tumour epicentre following intravenous contrast administration (100 mls 340 Niopam; 5 ml/s). Tumour blood volume and permeability was determined using two analysis methods: Distributed parameter analysis and Patlak analysis. Measurement agreement between perfusion parameters was assessed using Bland-Altman statistics. The mean and standard deviation (SD) for blood volume and permeability were 6.4 (1.3) ml/100 ml and 14.5 (3.4) ml/100 ml/min respectively using distributed parameter analysis and 11.3 (3.1) ml/100 ml and 17.8 (9.4) ml/100 ml/ min respectively using Patlak analysis. The mean difference (95% limits of agreement) were -4.9 (-10.6 to +0.71) ml/100 ml and -3.3 (-22.7 to +16.2) ml/100 ml/ min for blood volume and for permeability respectively. The within subject coefficient of variation was 45.5% and 44.6% for blood volume and permeability respectively. Commercially available CT perfusion software may produce discrepant results. The level of agreement between methods must be taken into account in cross comparison studies. The mean values of the bronchial lumen were 5.2 mm in OCT and 4.9 mm in histology. The tunica mucosa (TMUC) with ciliated epithelium was detectable by OCT as a medium-intense line with a mean thickness of 0.14 mm (0.11 mm in histological specimen). The high signal tunica muscularis (TMUS) (mean thickness 0.24 mm in OCT, and 0.16 mm in histology) could be separated from the low signal adventitia (0.23 mm in OCT, and 0.17 mm in histology). Visualization of the cartilage in the adventitia, that produced no reflexions, was possible in all 124 sections. Taking into account the magnification factor in OCT images due to shrinkage of the tissue by fixation, there was no significant difference in detection of the different layers between the two methods. Conclusion: Intraluminal OCT allows a nearly microscope-like resolution of the different bronchial wall layers in vitro. Therefore, utilization of this method in patients appears very promising. Free-breathing volumetric CT lung imaging using prospective non-contact gating in a large animal experiment J. Zaporozhan 1 , S. Ley 1 , R. Unterhinninghofen 1 , Y. Saito 2 , M. Fabel-Schulte 1 , G. Szabo 1 , H.-U. Kauczor 1 ; 1 Heidelberg/DE, 2 Tochigi/JP (j.zaporozhan@dkfz.de) Purpose: Investigation of prospective respiratory gating for CT of the lung using two prototype gating devices in a large animal experiment and evaluation of image quality. Methods and Materials: Eight anesthetized and ventilated healthy pigs underwent MDCT (1 mm, 120 kV, 300 mA) with prospective respiratory gating. Two non-contact gating devices were used: (1) CCD camera in all animals and (2) laser sensor in five pigs. The output signal of both gating devices was connected to the scanner instead of the ECG gating unit. Under free-breathing conditions the image acquisition was set to maximum inspiration and expiration. The quality of images and MPRs was analyzed in axial, coronal and sagittal orientation for sharpness of bronchi, diaphragm, lung fissures, parenchyma and pericardial structures using a 4-point-score (1 -excellent to 4 -severe artifacts). Results: Prospective respiratory gating worked perfectly well for inspiration and expiration in all animals when using the CCD camera, and a little less convincingly with the laser sensor: 100 % for inspiratory, but only 60% for expiratory scans. All acquired images showed excellent sharpness (CCD camera vs. laser sensor) for bronchi (1.3 ± 0.5 vs. 1.6 ± 0.7), lung fissures (1.0 vs. 1.1 ± 0.3), parenchyma (1.0 ± 0.2 vs. 1.4 ± 0.6) and diaphragm (1.5 ± 0.8 vs. 1.9 ± 0.9). The pericardial lung structures often showed minor to major cardiac motion artifacts (1.9 ± 0.6 vs. 2.3 ± 0.5). Conclusion: Free-breathing volumetric CT of the lung using prospective noncontact respiratory gating is feasible with high image quality. The CCD camera provided better results and slightly higher image quality than the laser sensor device. High-resolution CT of the chest in cystic fibrosis: Is simplification of scoring systems reliable? A. Oikonomou 1 , E. Hatziagorou 2 , M. Mantatzis 1 , J. Tsanakas 2 , P. Prassopoulos 1 ; 1 Alexandroupolis/GR, 2 Thessaloniki/GR (aoikonom@med.duth.gr) Purpose: The proposed scoring systems for the assessment of cystic fibrosis (CF) on HRCT, although accurate, are cumbersome and time-consuming, thus not widely applied in every day clinical practice. The aim of this study is to simplify these scoring systems without compromizing their reliability, by selecting representative variables. Methods and Materials: Forty-two consecutive patients with CF (mean age: 12 years) underwent a baseline and a follow-up chest HRCT (mean interval: 31 months). Three radiologists deciding by consensus, evaluated 84 HRCTs, employing 5 scoring systems (Bhalla, Hellbich, Santamaria, Brody, Oikonomou)each one with 7 to 11 variables -resulting in 5 baseline and 5 follow-up "scores" for each patient. Subsequently four variables were selected (severity and extent of bronchiectasis, severity of thickening, air-trapping) from each scoring system to form 5 new shorter "subscores". Each "subscore" was correlated with the corresponding complete score, using Pearson correlation, and the baseline scoressubscores were compared with the follow-up ones. Results: There was excellent correlation between each score and its corresponding subscore (0.922 70%) in all stents. In the Liberté stent high-graded stenoses were correctly diagnosed in 100% of all cases (Driver: 83%; Vision: 66%; Lunar Rox: 33%). Purpose: To evaluate the ability of 64 MDCT for the detection of significant restenosis in the follow-up of drug eluting coronary stent (DES). Methods and Materials: 51 patients (32m, age range 42-75 yrs) who previously underwent coronary percutaneous transluminal angioplasty (PTA) and stent placement (112 stents) were evaluated. After 9 months all patients were studied with MDCTA one week prior to selective coronary angiography (SCA), in order to evaluate stent patency. ECG-gated MDCT of coronary artery was performed before and after iv administration of contrast agent with the following acquisition parameters: 64x0.6 mm slice collimation with a slice thickness 0.6 mm, 120 kV, 700 mAs. Images were analyzed by two radiologists on a dedicated workstation using a real time interaction approach. Stent patency was evaluated qualitatively and quantitatively by measuring the density value proximally, inside and distally to the stent. MDCT results were compared with SCA, considered the gold standard. Results: 3 patients (5 stents) were not evaluated because of heart rate (HR) artefacts due to abnormalities of HR during the administration of iodine contrast. 107 stents were patent at SCA. MDCTA was able to detect in-stent re-stenosis in 13/16 cases by the first observer and in 15/16 cases by the second observer (overall sensitivity 87%). At MDCT 86/91 stents were judged patent by the first observer and 89/91 by the second observer with an overall specificity of 96%. Purpose: The aim of this study was to evaluate the diagnostic accuracy in assessment of coronary artery bypass grafts using 64-slice computed tomography technology. Methods: CT coronary angiography was performed for 96 bypasses in 31 patients with suspected coronary artery disease using a Siemens Sensation 64slice CT-scanner and compared with invasive coronary angiography. ECG-gated cardiac CT was performed at a mean heart rate of 63.3 ± 7.3 beats per minute (range 29 to 164). Patients with an irregular or fast heart rate despite ß-blocker administration were not excluded from the study. Results: All bypass grafts and 94% of the distal bypass anastomoses could be visualized by CT, non-evaluable distal arterial anastomoses was either due to clip material or calcification artifacts. 42 bypass graft occlusions and 3 significant stenoses were detected by CT and confirmed by invasive coronary angiography. Two venous grafts were missed and one arterial graft was not evaluable with invasive coronary angiography, but both were clearly depicted by MSCT. One false negative and two false positive CT-findings resulted in a sensitivity of 97.8%, a specificity of 89.3%, a positive predictive value of 90% and a negative predictive value of 97.7%. Conclusion: State-of-the-art MSCT coronary angiography allows detailed evaluation of venous and arterial grafts with high diagnostic image quality even in patients with higher heart rates. CT reveals complex graft anatomy readily and reliably. Calcification, poor opacification and clip material artefacts can still hamper the evaluation of distal arterial anastomoses despite improved spatial and temporal resolution. MDCT and conventional coronary angiography within 15 days after bypass surgery to assess the graft patency (9 venous grafts and 27 arterial grafts). One of the arterial grafts and two of the venous grafts were excluded because of failure in catheter engagement during conventional CAG. The graft status was evaluated according to the presence of significant stenosis (including obstruction). CT findings were compared with CAG. Results: In 32 grafts, there was no significant stenosis in both CAG and MDCT. In two grafts, one significant stenosis and one obstruction was concurred in both modalities. Only two grafts were interpreted differently. With CAG findings as a standard of reference, the sensitivity, specificity, and positive and negative predictive values were 100%, 97.0%, 75%, 100%, respectively, with an overall diagnostic accuracy of 97.3%. Conclusion: Many problems such as respiratory motion and fast, irregular cardiac rhythm is present in early period after CABG. Although in a small study population, 64-slice MDCT showed high sensitivity and specificity compared with invasive conventional angiography. CT coronary angiography is a promising alternative diagnostic modality in evaluating the patency of coronary bypass graft. Purpose: To compare US lymphography efficacy of microbubble formulations of varying diameter and composition using a normal dog model and to assess the effect of tumor replacement on nodal contrast uptake in a swine melanoma model. C D E F B A G Methods: Regional lymph nodes of 6 normal dogs were imaged using power Doppler US following subcutaneous injection of one of three hard-shelled bubbles (POINT Biomedical Corporation, mean particle size = 0.75, 0.91 and 1.33 µm) and one lipid-shell microbubble formulation (DefinityË, perflutren, mean particle size = 1.8 µm) into the distal front or hindlimb. Regional nodes were imaged 15, 30 and 45 minutes post-injection. Sixteen swine with cutaneous melanomas were imaged in a similar manner following SQ injection of perflutren suspension around each primary lesion. Lymphscintigraphy studies were performed on all dogs and pigs prior to US imaging. Results: Contrast enhancement was evident at a minimum of one time-point in 95% (1.8 µm diam), 85% (1.33 µm diam), 75% (0.75 µm diam) and 70% (0.91 µm diam) of normal canine nodes and occurred within 15 minutes in most instances. Enhancement involved 5-100% of cross-sectional area in all perflutren-enhanced nodes. Regional nodes contrast enhanced in all pigs with excellent correlation with lymphoscintigraphy even in nodes predominantly replaced by tumor. Conclusion: Microbubbles of varying mean diameter and shell composition are efficacious for sentinel node localization with contrast enhancement being more consistent when using larger particles. Furthermore, sentinel node detection using this approach compares favorably to current localization techniques even when significant tumor replacement has occurred. Purpose: A new ultrasound method of image guided drug delivery using targeted nanoparticles will be described. The in-vitro efficacy of ultrasound mediated drug release from specially formulated nanoparticles (NPs) containing chemotoxic agents will be presented. Release of encapsulated drug from the nanoparticles using ultrasound was accomplished using a specially modified Aloka SSD5500 PHD RF ultrasound system using a 5.0 -10.0 MHz linear ultrasound transducer (Aloka UST-5539). To determine the efficacy of this point and shoot ultrasound mediated of drug delivery method using our NPs we cultured a set of cancer cells for 1). a control condition, 2). a raw drug condition and 3). an ultrasound mediated NP drug release condition. Amount of cell death was determined using standard staining methods. The results of this experiment revealed that ultrasound mediated NP drug release produced the same amount of cell death compared with the application of raw drug -doxorubicin; 80% raw drug; 79% ultrasound triggered NP drug release. No cell death was observed in the control group. Conclusion: Ultrasound mediated release from our NPs produces cell death in a similar manner as raw drug in-vitro. Consistent with published literature, there appears to be an ultrasound advantage after six hours most likely due to the enhancement of the effects of ultrasound and drug uptake by the cells. Early functional evaluation by contrast enhanced ultrasonography using perfusion software of isolated limb perfusion with melphalan and TNF alpha for in-transit melanoma metastases L. Chami, A. Cavalcanti, M. Lamuraglia, N. Elie, S. Bonvalot, J. Leclere, A. Roche, N. Lassau; Villejuif/FR (chami@igr.fr) Purpose: The aim of this study was to evaluate if the tumor microvascular changes on contrast enhanced ultrasonography (CEUS) after a previous isolated limb perfusion (ILP) could predict the response. Methods and Materials: From march 2004 to august 2005 twenty patients qualified for first ILP were included in this prospective study. CEUS examination using SonoVue (Bracco) was performed for each patient at D-1 before treatment, at D+1, D+7 and D+90. Using a 14 Mhz probe, the tumor size was measured and the contrast enhancement was evaluated both visually and from raw data using the CHI-Q software (Toshiba). The results were compared with clinical response at 3 months backward. Results: At this time, 72 CEUS were performed on 23 target lesions. One patient was lost for follow-up, one had no visible lesion, one tumor did not exhibit any contrast uptake and for 6 patients data at D+90 are not yet available. The median follow-up was 6 months (min 1; max 18). Regarding target lesions, the response rate was complete in 28.6% (n = 4), partial in 64.3% (n = 9) and progression occurred in 7% (n = 1). At D+1, 56% of target lesions showed an improvement of contrast uptake may due to inflammatory reaction. At D+7, a decrease of more than 10% in tumor contrast uptake was observed in 69% (n = 9) of the responders. media consisting of DMEM, FCS, Resovist and protamine at different concentrations and increasing incubation times of 1, 2, 6, 10 and 24 hours were used to label mesenchymal stem cells (MSC). After labeling, samples of 150,000 cells in ficoll solution were imaged with T1, T2 and T2*-weighted sequences at 1.5 T and 3 T. Respective relaxation times were calculated using self-written software. Iron uptake was confirmed by electron microscopy and quantified by spectrometry (ICP-AES). Results: Labeling medium consisting of DMEM, protamine and Resovist lead to formation of complexes that were too large for cellular uptake. Adding FCS into the solution prevented any complex formation and allowed for efficient labeling of MSCs. Simple incubation with Resovist only resulted in efficient labeling after incubation for 24 hours while assisted labeling with protamine sulfate resulted in even better results after only one hour. The labeling resulted in significantly shortened T2 and T2*-relaxation times which were more pronounced at 3 T than at 1.5 T. Electron microscopy confirmed intracellular iron oxide uptake. Purpose: To evaluate the safety and efficacy of 1 molar gadobutrol as alternative contrast agent for computed tomography. In nine domestic pigs (52 ± 4 kg body weight) native and contrast enhanced multidetector CTs (Volume Zoom, Siemens, Erlangen, Germany) of the chest and abdomen were randomly performed using either 1.2 or 3 mL/kg BW gadobutrol (1 mol/L; n = 3 each). Changes in blood chemistry and hematologic parameters were monitored before, 24h, 48h, 72h, and 7 days post application. Urine samples were evaluated pre and 7 days post application. Before and 7 days post contrast media application, a 99mTc-MAG3 renal scintigraphy was performed. The animals were sacrificed and kidneys were removed for examination by light microscopy one week after contrast media administration. Results: Differences in enhancement of vascular structures and parenchymal organs was statistically significant between 1 mL/kg BW compared with2 and 3 mL/kg BW. No serious adverse event occurred, and no clinically relevant changes in hematologic results, blood chemistry or urine analysis were detected during the observation period. Markers for glomerular filtration (creatinine, cystatin C, Beta2-microglobulin, creatinine clearence) and tubular function (Alpha1-microglobulin) were unaffected in all groups. No morphologic changes of the renal parenchyma was found in histology. Conclusion: Contrast enhanced CT with use of gadobutrol resulted in an excellent vascular and parenchymal enhancement. Gadobutrol did not affect renal function up to a dose of 3 ml/kg BW. Even in high doses, gadobutrol can be administered as alternative contrast media in case of contraindications for iodinated contrast media, e.g. for CT. Purpose: To label human monocytes with gadofluorine M by simple incubation for subsequent cell depiction at 1.5 and 3 Tesla. Gadofluorine M is a gadolinium-based T 1 -contrast agent with a high R1 relaxivity, which forms micelles in aqueous solutions. The micelles are spontaneously phagocytosed by macrophages. Samples of human monocytes were incubated with gadofluorine M-Cy3 at varying concentrations, incubation times and either investigated directly or after culturing over increasing time intervals. All cell samples and non-labeled controls were investigated at 1.5 and 3 Tesla. R1-relaxation rates of the cells were determined and tested for significant differences using a t-test. Cellular contrast agent uptake was proven by fluorescence microscopy and quantified by ICP-AES. Results: Efficient cell labeling was achieved after cell incubation with gadofluorine M at a concentration of 25 µmol Gd/ml and an incubation time of 24 hrs resulting in a maximal Gd uptake of up to 3.0 fmol Gd/cell without impairment of cell viability. Fluorescence microscopy confirmed an intracellular uptake of the fluorescent contrast agent inside the cells. The r1 relaxivity of the labeled cells was 137 mM -1 s -1 at 1.5 T and 80.46 mM -1 s -1 at 3 T. Follow-up imaging studies showed a stable labeling for at least 7 days. The peak areas of NAA, Cho, Cr before Gd-DTPA administration were more than those after Gd-DTPA administration in most cases. Only the decrease of peak area of Cho had statistical significance (t = 2.401, P < 0.05). The ratios of Cho/Cr, Cho/NAA before and after administration of Gd-DTPA showed no statistical difference. Conclusion: According to the fact that it is ratios such as Cho/Cr, Cho/NAA but absolute quantitation that are popular parameters used to analysis proton spectra so far, administration of paramagnetic contrast agent will not influence the result of 1 H MRS. To examine an ill-defined lesion, 1 H-MRS could be performed after contrast agent administration to put the ROI precisely without the concern over influence of contrast agent on spectrum analysis. The total effective rate of the p53 treatment group was 58.3%, and that of the control group was 26.5% with significant difference (P < 0.05). The occurrence of gastrointestinal symptoms was lower than that in the control group (P < 0.05). 3 month and 6 month survival in the p53 treatment group was significantly higher than in the control group (P < 0.01). The approach for patients with HCC injected intratumorally with rAdp53 was safe and effective. Our results show rAdp53 is a potentially effective gene therapy agent in HCC treatment. Human growth factor (HGF) evaluation before liver intervention: Long term results of a prognostic significance for liver regeneration T. Purpose: To evaluate and compare the complexity of liver regeneration in correlation with the regeneration factor HGF post liver resection and MR-guided laser induced thermotherapy (LITT). Sixty-three patients were evaluated: 31 patients post surgical liver intervention and 32 patients post ablation (LITT). During a period of 18 months the effects of therapy on the cytokine HGF were evaluated and com-pared to the initial, pre-interventional HGF value. In addition, the effect on liver volume after surgical and interventional radiological procedure was evaluated using unenhanced and contrast enhanced MR imaging and CT. Volumetric examination (CT-volumetry) of liver parenchyma pre-and post-intervention was accomplished and the regional growth of liver segments was evaluated. Results: A significant influence on the HGF process could be confirmed on loss of liver parenchyma resulting form surgery or LITT induction. Pre-interventional HGF values < 1500 pg/ml (n = 33) led to a significant increase in liver volume (115-584 ml) . Patients with initial HGF-values > 2000 pg/ml (n = 21) showed a decrease in liver volume (5-409 ml) after intervention. Pre-interventional HGFvalues between 1501 pg/ml and 1999 pg/ml (n = 9) showed no definite reaction of liver parenchyma. All patients younger than 55 years showed stronger dynamics than older patients. The technique used in surgery per se had no significant influence on the cytokine process and regeneration dynamics. Conclusion: Pre-interventional human growth factor (HGF) value has a significant impact on liver regeneration post MR-guided laser induced thermotherapy (LITT) and surgical resection. Pre-interventional low HGF-values correlate with increased liver regeneration and stronger regeneration dynamics. Experimental study of transarterial administration of GRGDSP combined with transarterial chemoembolization in rats with hepatic carcinoma J. Qian, J. Yin, Y. Wang, G. Feng; Wuhan/CN (qianjun_hcc@yahoo.com) Purpose: To evaluate the effects of transarterial administration of GRGDSP (Gly-Arg-Gly-Asp-Ser-Pro) combined with transarterial chemoembolization (TACE) against hepatic carcinoma in rats. Methods and Materials: Subcapsular implantation of a solid Walker-256 tumor in the liver was carried out in 26 male Wistar rats. The following agents were injected in the hepatic artery: Group A -GRGDSP + TACE (mitomycin + lipiodol + ligation of the hepatic proper artery) (n = 9); Group B -TACE alone (n = 8); Group C -0.9% NaCl (0.2 ml) (n = 9). Tumor volume was assessed by MR imaging before (V1) and after (V2) the interventional treatment. The expression of factor VIII in tumor tissue and the tumor microvessel density (MVD) was analyzed by immunohisto-chemical examination. The mean ratio of V2/V1 in Groups A, B and C was 4.42 ± 0.48, 6.98 ± 1.09 and 13.00 ± 1.68 respectively. There were significant differences in V2/V1 between different groups (p < 0.05). After interventional therapeutic procedures, the optical density value related to the positive expression of MVD in Groups A, B and C was 0.18 ± 0.02, 0.22 ± 0.02 and 0.23 ± 0.02 respectively. There were significant differences between Group A, and Groups B and C (p < 0.05) but not between Group B and Group C (p> 0.05). Conclusion: Infusion of GRGDSP via the hepatic artery combined with TACE noticeably inhibits the growth of hepatic carcinoma and its intrahepatic metastasis in rats. Purpose: Intraarterial injection of yttrium 90 microspheres is used in the treatment of otherwise non responding hepatic metastases from various tumors. This article presents an analysis of the incidence of liver toxicities associated with selective internal radiation therapy (SIRT). Methods and Materials: 34 patients with extensive hepatic metastases qualified for SIRT. Liver toxicities were coded with use of standard oncology criteria (Southwest Oncology Group Criteria). Laboratory (total bilirubin, aminotransferases, alkaline phosphatase) and clinical parameters (ascites, encephalopathy, liver failure, hemobilia) were recorded and compared with pre-treatment levels. The SWOG criteria grades liver toxicity from 0 (no toxicity) to 5 (death) with a clinical significance of a toxicity grade ≥ 3. Descriptive statistical methods were applied to determine the incidence of liver toxicities after SIRT. Results: 22 patients (65%) experienced 33 liver toxicities. At 3 month follow-up 17 grade 0 and 16 grade 1 toxicities were observed. No toxicity grade ≥ 3 was recorded. At 6 month follow-up 6 grade 0, 15 grade 1, 4 grade 2 and 1 grade 3 toxicities were noted. The most frequent toxicity was increased aminotransferase levels, followed by alkaline phosphatase levels and bilirubin levels. Eight of the 33 toxicities resolved on follow-up, 5 became worse. The grade 3 toxicity was attributed to tumor progression. Major side effects were seen in 10% (versus 2% in Group 2); median hospital stay was 26 days in the original group (versus 5 days for Group 2). The median number of treatments was 2 in both groups. Twice as many group 2 cases (37%) were identified by routine surveillance. Adjunctive percutaneous ethanol injections (PEI) were given to 10 patients in Group 1, while 9 in Group 2 had radiofrequency ablation (RFA). Median survival in patients with good liver function improved from 8 months (Group 1) to 21.5 months (Group 2). Conclusion: Despite an increase in the incidence of HCC there has been an improvement in survival and complication rates, probably as a result of earlier detection and better case selection. Other factors may include lead-time bias, evolving changes in the TACE procedure and the impact of RFA. Purpose: To evaluate the efficacy of paclitaxel loaded nanoparticles administered by local administration over the channelled balloon catheter at preventing restenosis after balloon angioplasty in restenotic iliac arteries of New Zealand white rabbits. Methods and Materials: Eight male New Zealand White rabbits were fed a 1% cholesterol diet and received denudation of both common iliac arteries with a 3 mm balloon catheter to induce an arterial denudation. Four weeks after stenosis induction, local application of paclitaxel loaded nanoparticles was performed randomly on one common iliac artery. As a control saline was injected into the contralateral side under the same conditions. The channelled balloon catheter allows simultaneous balloon dilation (8 atm) of the stenosis and local application of a drug solution (2 atm). Four weeks after local drug delivery the animals were killed, the vessels were excised and computerized morphometric measurements were performed. Results: In none of the animals acute complications e.g. dissection, vessel thrombosis or perforation were seen. Application of paclitaxel loaded nanoparticles resulted in a statistically significant reduction of neointimal area compared to the control group of 50%. The neointimal area was 0.80 ± 0.19 mm 2 using nanoparticles compared to 1.58 ± 0.6 mm 2 in the control group. Conclusion: Local drug delivery of paclitaxel loaded nanoparticles using the channelled balloon catheter could reduce neointimal hyperplasia in restenotic rabbit iliac arteries under the predominant conditions. How do patients perceive the benefits and risks of peripheral angioplasty? Implications for informed consent S.B. Habib 1 , A.M. Groves 2 , T.C. See 3 , L. Sonoda 3 , P.J. Ell 2 ; 1 Nottingham/UK, 2 London/UK, 3 Cambridge/UK (saidsuha@aol.com) Purpose: In order for patients to give their informed consent they should fully understand the benefits and risks of the procedure involved. A survey was performed to explore patients' perceptions of benefits and risks of peripheral angioplasty. Methods and Materials: Over 8 months, 206 patients undergoing peripheral angioplasty were surveyed. 30 patients formed a pilot study leaving 176 patients (123 male, 53 female) average age 69.6. The patients were randomly assigned into 2 groups. One group was given a recognised "Risk-Assessment-Chart", the other group answered without this tool. Patients were questioned regarding the risks and benefits of the procedure immediately pre and post-angioplasty. Patients were asked to express their likelihood of benefit and complication in terms of a percentage (definite benefit being 100%). Whether the patient had received information or had undergone angioplasty prior was noted. Statistical differences were calculated using students-t-test. Results: 158 (90%) patients thought that they would have at least a 75% (3 in 4 chance) of benefiting from angioplasty. 39% thought they would definitely (100%) benefit. 85 (48%) patients thought the test would be painful, 59 (34%) patients thought there would be no radiation exposure, 28 (16%) patients realised there was a chance of uncontrolled bleeding and 143 (81%) patients thought the test was easier than they had originally thought. The use of the "risk-assessmenttool" significantly altered patient's perceptions (t-test-p = 0.040). Conclusion: Most patients undergoing angioplasty were optimistic about the benefits, but were unaware of the risks. The risk assessment tool would appear to be useful for the consent process. An audit of pain control in interventional radiology G.G.J. McKenzie, A. Downie; Glasgow/UK (gmckenzie@doctors.org.uk) Purpose: In the United Kingdom the majority of interventional cases are performed without an anaesthetist so making the interventional radiologist responsible for pain control and sedation as well as performing the procedure. The aim of this audit was to assess the adequacy of pain control in interventional radiology. Method: Over a 6 month period, pain control was assessed using a Visual Analogue Scale. Both the pre-procedure and the highest intra-operative pain scores were recorded for a wide variety of procedures. The type of anxiolytic/analgesic administered were also recorded. Statistical analysis was performed with the Minitab computer package. Results: Eighty patients took part in the audit. Analysis shows that 46 (57.5%) of these patients had insufficient intra-operative pain control. Particularly, there was a significant increase in pain levels (p < 0.05) experienced intra-operatively, compared to the pre-procedure levels in angiography, angioplasty, Hickman line insertion, gastrostomies, biliary stenting and varicocele embolization. Furthermore, pain control was shown to be insufficient (p < 0.05) in Hickman line insertions, gastrostomies and biliary stenting. An anaesthetist was not involved in any of the cases and pre-medication was never given. Conclusion: Pain control is poorly managed in the majority of interventional cases, particularly Hickman line insertions, gastrostomies and biliary stenting. There is therefore considerable scope to improve patient analgesia perhaps by improving the type and dose of drug administration including the use of pre-medication and ideally if resources allow the presence of anaesthetic colleagues. Background: PTA of severely stenotic arteries is hampered by high restenosis rate. The effects of PTA on vascular wall as well as the anti-restenotic properties of endovascular brachytherapy (EVBT) remain unclear. We aimed to elucidate the vascular effect of PTA and PTA+EVBT in vivo by serial MR imaging. Methods: Twenty symptomatic patients with severe stenosis of the femoropopliteal artery were randomly assigned to PTA (n = 10) or PTA+EVBT (n = 10, 14 Gy by γ-source) and imaged by high-resolution MR imaging before and 24 hours, 3 months and 2 years after intervention. An independent observer blinded to the procedure analyzed the MR imaging data. Results: At 24 hours, MR imaging revealed that lumen area (+86% and +67%) and total vessel area (+47% and +34%) similarly increased in the PTA and PTA+EVBT group. All patients showed severe splitting of the atherosclerotic plaque. At 3 months, MR imaging revealed a significant difference in lumen area change between the PTA and PTA+EVBT group (+40% and +106%, respectively; p = 0.026) and in the total vessel area (+14% and +39%, respectively; p = 0.018). At 24 months, signs of restenosis and inward remodeling were also present in patients treated with PTA+EVBT. PTA induced plaque disruption persisted at 3 months in 50% and at 2 years in 10% in PTA+EVBT patients only. Conclusion: PTA induces deep disruption of the atherosclerotic plaques and extensive remodeling process of the arterial wall. Luminal loss after PTA is partially due to inward vessel remodeling. Brachytherapy initially prevents restenosis, induces outward remodeling of the vessel and delays the process of restenosis. Purpose: To prospectively evaluate feasibility and safety of day-case interventional angiography in a self-contained radiological nurse-administered unit. Methods and Materials: Data was recorded prospectively over an eight-year period (1996 to 2004) . Radiology Specialist Nurses (RGN), in a pre-angiography evaluation clinic, assessed patient suitability for day-case angiography based on a protocol. Immediately prior to the procedure patients were reassessed and an interventional radiologist performed all procedures. The RGNs were responsible for patient care before, during and after angiography and for follow-up post-procedure. The data were summarised into either diagnostic or interventional (balloon dilatation/stent insertion/embolization) cases. Complications (minor or major) were divided into immediate or delayed ( were treated with PTA (Group A) whereas the last 28 patients (32 lesions) underwent "cutting" balloon angioplasty (Cutting-Balloon Ultra; IVT/Boston Scientific, USA; 3.5-5-mm diameter/10-15-mm length) (Group B). Baseline patient demographic data, pre-and post-procedural patient clinical data and procedural results of Group A and B were recorded. Follow-up consisted of clinical assessment and color duplex ultrasonography (CDU) 1, 3 and every three months after the procedure. Results: All treatments were successfully performed via an anterograde approach with a technical success rate of 100%, without any major complications. In two treated lesions of Group A (2/29, 6.9%) a stent was implanted because of a dissection, whereas no patient of Group B required placement of a stent because of recoil, dissection, or arterial tears. In Group A (PTA) primary and secondary patency rates were 85.2% and 93% at 6-months, 73.8% and 77.9% at 12-months, respectively. In Group B (CBA) primary and secondary patency rates were 90% and 96.8% at 6-months, 80.5% and 87.7% at 12-months, respectively. Conclusion: CBA seems to be a valuable tool in the endovascular treatment of short femoro-popliteal stenotic lesions allowing an increased patency and better middle-term results as compared to conventional PTA. Purpose: Comparison of routine and low-dose CTU protocols regarding diagnosis of upper urinary tract (UUT) pathologies. Methods and Material: During a 1.5-year period, 92 patients with diseases suspected to affect the UUT underwent contrast-enhanced CT of the abdomen, including a portovenous/nephrographic and a delayed excretory phase. CTU was obtained with 4x2.5 mm collimation after administration of i.v. saline. In group 1, CTU was performed with a standard dose protocol (n = 42, 120 kV, average 175.6 mAs/slice) and in group 2 with a low dose examination protocol (n = 50, 120 kV, 29 mAs/slice). Pathologies of the UUTs were corroborated by a combination of portovenous/nephrographic scans and clinical information. CTU scans were reviewed by two independent readers (R1, R2) blinded to corroborating data. UUT pathology (dilatation, compression, intraluminal lesions) was diagnosed using a 5-point confidence scale: pathology definitely/probably absent (1.2), undecided (3), probably/definitely present (4.5). Statistics were performed using ROC-analysis and Chi-Square testing. Results: Sensitivity and specifity for UUT dilatation was equal in routine and lowdose CTU (group 1, sensitivity 92.9%, specificity 99.7%; group 2, sensitivity 96.2%, specificity 99.4%; sensitivity, chi-square = 0.01, p = 0.92; specificity, chisquare = 0.062, p = 0.8). For UUT compression, no significant differences between the groups were found (group 1, sensitivity 96.4%, specificity 98.8%; group 2 sensitivity 90.9%, specificity 99.0%; sensitivity, chi-square = 0.013, p = 0.91; specificity, chi-square = 0.23, p = 0.6). Number of intraluminal lesions was low: In group 1 5/5, in group 2 1/1 (R1) and 0/1 (R2) lesions were detected. Conclusion: Dose-reduced CTU is adequate to determine UUT dilatation and compression. However, it is unlikely to depict intraluminal lesions reliably. Low-dose CT-urography: Image quality analysis during gradual tube current reduction. Purpose: To intra-individually evaluate the possibility of reducing X-ray exposure during multi slice CT urography (MSCTU) with consideration of image quality using a porcine model. Methods and Material: MSCTU was performed in 8 healthy pigs under general anesthesia (weight ± 70 kg). Examinations were performed with a gradual reduction of the tube current time product at 120 kV (200, 125, . Three blinded observers independently evaluated the image data for anatomical detail, subjective image quality and image noise using a five point scale. Overall image quality was compared to radiation dose. Diagnostic reliability was tested using the intraclass-correlation-coefficient (ICC). Stratified comparisons of mean scores were calculated using Friedman and Wilcoxon-tests. Results: Adequate image quality for detailed visibility of the upper urinary tract was obtained by dropping the tube current time product to 70 eff.mAs at 120 kV. Image noise did not impair image quality to a relevant degree using these parameters. Mean score for image quality did not significantly advance with the use of higher currents (p > 0.05). A further reduction of the current time products below 70 mAs leads to a significant impairment of image quality (p < 0.05). There was a high agreement between the 3 observers (ICC = 0.95). In the experimental setting, radiation dose in MSCTU could be significantly reduced. To maintain a sufficient image quality, a current time product reduction to 70 mAs at 120 kV proved to be feasible offering an advantageous dosage reduction. Systematic clinical studies are required for the assessment of the diagnostic impact of such protocols. Purpose: To evaluate the image quality of two-phase MDCTU protocol for evaluation of the urinary system. Methods and Materials: 110 patients with haematuria, flank pain and or suspected urinary tract abnormalities were examined on a 16-slice CT-scanner (Somatom Sensation 16, Siemens). MDCTU protocol consisted of two phases: Unenhanced and triple-bolus iv contrast injection. 1) patients were given 800 ml of water 20 min before the study and unenhanced (low-dose) scan of the full abdomen was acquired; 2) 30 ml 2 ml/s of IV contrast (Visipaque TM 320 mg I/ml, Amersham) was injected as a first bolus, 7 min. later a second bolus of 50 ml 1.5 ml/s and after 20 seconds the third bolus of 65 ml 3 ml/s injected and the full abdomen scanned. Coronal 5 mm MPR reconstructions were generated as standard for both scans and in selected cases oblique MPR or VR. Results: Examination revealed urolithiasis (n = 24), hydronephrosis (n = 9), renal mass (n = 36), TCC (n = 2), accessory renal arteries (n = 7). High quality parenchymal enhancement was achieved in 94 (85.4%) patients. Streak artefacts were seen only in 4 (3.6%) cases caused by dense opacification of the calyces. Opacification of the collecting system and ureters were scored on a 5grade scale (1=not filled, 3=half filled, 5=all filled). Median score for upper urinary tract filling was 4 and 3 for lower segments of the ureters. In 2 cases MR followed MDCTU, but was not problem solving. Conclusion: Two-phase MDCTU with triple-bolus single scan IV contrast injection is a simple protocol acquiring native, arterial, nephrographic and excretory phases in two series. It provides satisfactory imaging scores, reduces radiation dose and radiological workload. Purpose: MR-pyelography is a technique to evaluate the renal collecting systems, performed after drug induced distention of the urinary tract. Method and Materials: 10 normal volunteers and 271 patients with renal diseases underwent drug induced MR-pyelography using 0.5 and 1.5 T superconductive magnets. The examination was performed with a 3D non-breath holding fat-suppressed TSE-sequence (TR:3000 ms; TE:700 ms; N.Ex.:4; ETL:128; acquisition time:3 min) in a coronal plane. These acquisitions were post-processed with a MIP algorithm. To obtain maximum filling of the collecting system, diuresis was pharmacologically induced by administering 250 ml of saline solution i.v. together with 20 mg of furosemide. One MR acquisition was performed 10 minutes after diuresis induction. 5 normal volunteers and 230 patients had undergone an IVU and ascending pyelography was performed in 15 cases. Results: We always obtained excellent anatomical evaluation of the renal collecting systems with drug induced MR-pyelography that was considered comparable to that obtained with an IVU. It was always possible to study the renal collecting system malformations. Renal calculi larger than 2 mm were always identified by analysis of the 3D coronal scans. Drug-induced MR-pyelography is particularly important for the detection of ureteric stones: In 23 cases with obstructive uropathy the site of the obstruction was determined and using conventional MR images, abdominal plain radiographs and urinary cytology, we were always able to determine the cause. Conclusion: Drug induced MR-pyelography allowed a morphological study of obstructed and non obstructed renal collecting systems and is suggested for patients with contraindications for i.v. administration of contrast agents; in nonfunctioning kidneys it is considered the best diagnostic imaging modality to perform after US. Relation between cerebral blood flow, cerebrovascular reactivity and severity of leukoaraiosis F. Todua, D. Gachechiladze, M. Beraia; Tbilisi/GE Purpose: Leukoaraiosis (LA) is a frequent finding among the elderly and in patients with cerebrovascular disease. The aim of our study was to assess the relationship between cerebral blood flow (CBF), cerebrovascular reactivity (CVR) and severity of LA. Methods and Materials: 114 patients (mean age 63.6 ± 12y) underwent brain MR imaging, MR-angiography and color Doppler of extra-intracranial vessels, and routine neurological examinations. Patients with prior stroke and high-grade carotid stenosis were excluded from study. Total CBF was determined by adding volume flow rates in both the internal carotid and vertebral arteries. CVR was evaluated by means of TCD Diamox 1 g test. The severity of LA was rated by semiquantitative rating scale by adding scores for periventricular and subcortical areas. Results: Presence of multifocal hypointense cerebral lesions, revealed on MR imaging, was strongly correlated with LA. FLAIR sequences were most informative for assessment of pathological changes. Examination revealed carotid artery moderate stenosis in 43 (38%) of cases. CBF in LA patients was decreased (468 -1078 ml/min, mean 764 ml/min). A significant negative correlation was seen between periventricular white matter changes and CBF (r = -0.76). TCD showed increase of pulsatile index (PI) -1.02 ± 0.22 and impaired CVR (+38.6% to Diamox test). Conclusion: Our data gives us the possibility to conclude that LA is not associated with ICA atherosclerosis severity. There is a significant negative relationship between total CBF and severity of LA. Due to increased vasomotor resistance in brain circulation and decreased vasoreactivity the slowed blood flow may play an important role in the development of LA. Quantitative perfusion measurements in acute ischemic stroke using functional CT and a distributed-parameter tracer kinetic model: Preliminary results and comparison with deconvolution-based perfusion measurements S. Bisdas 1 , T.S. Koh 2 , L. Cheong 2 , M. Hartel 3 , J. Gurung 1 , T.J. Vogl 1 ; 1 Frankfurt/DE, 2 Singapore/SG, 3 Katowice/PL (s.bisdas@med.uni-frankfurt.de) Purpose: To evaluate the potential of a distributed-parameter (DP) tracer kinetic model for assessing cerebral perfusion in acute ischemic stroke using perfusion-CT. Methods and Materials: 15 patients with acute supratentorial stroke symptoms underwent perfusion-CT measurements. Perfusion images were analyzed with a region-of-interest-based method using a newly introduced DP model. Blood flow (F), fractional intravascular (V1) and extracellular extravascular blood volume (V2), transit time in vascular compartment (T), first pass extraction ratio (E) and permeability-surface-area product (PS) maps were generated. A time-lag parameter (Tlag) was introduced so as to denote the time difference between the initial rise of the arterial and tissue density curves. Furthermore, a vascular pixel elimination method was applied to avoid an overestimation of the F values. The kinetic parameters as well as the infarct and penumbra size were compared to those derived from a deconvolution-based tracer kinetic model. The DP perfusion maps provided better resolution than the deconvolution-based ones. Infarct and penumbra regions were delineated and demonstrated distinct F, V1, and T values, which were significantly (P < 0.01) correlated in both models. E and PS parameters were not specific to infarct and penumbra delineation. Significantly elevated E and PS values were found in two patients with hemorrhagic transformation. Tlag maps allowed rapid identification of the areas with varying degrees of hypoperfusion and correlated better than T maps with the F abnormalities (P < 0.01). V2 maps demonstrated abnormalities within the ischemic lesions. Conclusion: Our proposed DP model can reliably depict the hypoperfused brain lesions in acute ischemic stroke. Purpose: To determine the accuracy of CT-perfusion (CTP) and angio-CT (CTA), and correlate results with the clinical evaluation during emergency admission to predict tissue outcome. Methods and Materials: Noncontrast-CT (NCCT), CTP and CTA were obtained from 36 patients (mean age 64 years) with non-hemorrhagic hyperacute stroke and within 24 hours after beginning the episode (between 1 hour and 24 hours). The regional cerebral blood flow (CBF), cerebral blood volume (CBV) and mean transit time (MTT) maps were calculated to discriminate between infarcted, non infarcted and penumbra areas. Head and neck CTA was performed from the origin of the aortic arch to the Circle of Willis. CTA was reviewed with MIP, MPR and vessel analysis program. Patients were clinically evaluated according to the National Institutes of Health Stroke Score at the time of admission and discharge. Statistical analysis was performed by using the weighted kappa index (KI) with its 95% confidence interval (CI ) with first-ever acute ischaemic stroke underwent diffusion-weighted MR imaging (DWI) and SPET within the first 48 hours and at day 8 after onset of stroke. CCD was considered to be present when SPET interhemispheric asymmetry index (AI) was > 5% in cerebellum. ADCvalues of bilateral cerebellar hemispheres were obtained from each patient and 15 age-matched healthy volunteers (67 ± 6 years). Results: Within first 48 hours, the bilateral cerebellar ADC-values did not significantly differ from volunteers or between the ipsilateral and contralateral cerebellar hemispheres (p = 0.867). However, on day 8, the ADC-value in the contralateral cerebellum was significantly higher than on the ipsilateral side (71 ± 5×10 -5 mm 2 / s vs. 69 ± 4×10 -5 mm 2 /s, p = 0.037), being also significantly higher than the averaged (right/left) ADC-value of volunteers (67 ± 4×10 -5 mm 2 /s, p = 0.013). No such difference was found on the ipsilateral side. CCD in SPET was present in 11 (50%) patients at acute stage, and in 14 (64%) patients on day 8. The difference in ADC between two sides significantly correlated with the AI in SPET within 48 hours (r = 0.611, p = 0.003) and on day 8 (r = 0.617, p = 0.002). Purpose: Carotid artery stenting (CAS) is an alternative method to endarterectomy in the revascularization of carotid artery stenosis. The long term efficacy of CAS is still debated. The aim of our study was to assess the survival/neurological free event rates and the incidence of long term intrastent restenosis after CAS and to identify some clinical or radiological predictors of restenosis. We conducted a retrospective study selecting from our data base 191 patients treated between 1997 and 2004 that had a clinical and CD-US follow-up. The survival Kaplan-Meier curves were obtained for overall survival, for freedom from restenosis and freedom from stroke-TIA/all death. We used a univariate analysis to correlate clinical, radiological and procedural data to stenosis recurrence and a multivariate analysis to determine independent predictors of restenosis. The overall peri-procedural complication rate was 3.1% (1.6% of minor, 0.5% of major and 1 % of fatal stroke). In-stent restenosis was observed in 11.5% of procedures with a cumulative primary patency rate of 88. Results: Mean interval between initial study and follow-up was 14.7d. Initially CAD was proven in 21 vessels (14xICA, 7xVA). 18 vessels showed a stenosis, 3 were occluded. On follow-up study the degree of stenosis increased in 3 vessels, remained unchanged in 11 vessels and decreased in 4. All 3 occluded vessels were recanalized on follow-up. In 3 vessels a pseudoaneurysm was visible on the initial study and remained unchanged on follow-up; in 1 vessel a new pseudoaneurysm was observed. In 3 vessels new dissections were identified on follow-up studies. Conclusion: High resolution MR imaging permitted excellent analysis of morphological features of CAD and clinically important information could be gained. Thus, short-term follow-up in pts with acute CAD is recommended. Further studies are needed to assess the relationship between short-term results and definite outcome. Purpose: To analyze the usefulness of perfusion computed tomography in the imaging diagnosis of brain death prior to organ donation. We prospectively examined 12 studies in 8 patients (17-67 years, 6 men) with suspected brain death where electrophysiological diagnosis of brain death was impossible because of drug intoxication. A perfusion CT and angiographic CT study was performed and analyzed as confirmatory tests, and compared with subsequent clinical evolution of these patients. Results: Perfusion CT demonstrated a very good accuracy in assessing brain death diagnosis, as a complimentary tool of the angiographic study performed with CT. Angiography demonstrated non-filling of intracranial arteries at the entry of the skull in 6 studies. Perfusion CT confirmed an arrest of cerebral blood flow in both carotid and posterior circulation, and persistent flow in external carotid artery territories. In 4 patients reduced but persistent cerebral blood flow was demonstrated, and in three of them a new CT angio and perfusion study was performed several hours later that confirmed then a complete absence of blood flow. Conclusion: Perfusion CT is a very good technique to study brain death because it is rapid and easy to interpret. Diagnostic parameters may be discussed in future studies in order to obtain similar standard parameters between different techniques classically used in these patients. in 12, IIB in 23, IIIA/B in 26) underwent breast MR imaging followed by corebiopsy. MR imaging was performed at 1.5 T acquiring morphological (T2-weighted sequence) and dynamic (3DFSPGR sequence before and after intravenous administration of Gadolinium chelates) studies. MR imaging morphofunctional findings were criteria described by Fischer in 1999 (shape, margins, contrast material kinetic, early contrast uptake [ECU] and signal/intensity curve morphology) evaluated as single criterion and as scoring system. Core-biopsy was performed for determination of histopathological characteristics: histological type, nuclear grade, estrogen and progesteron receptors (ER, PgR), Ki-67 and cerbB2 expression. Cut off for positivity was 20% for Ki67, 10% for ER and PgR and 3+ for cerbB2. Results: MR imaging identified 68 single node breast cancers (57DCI, 11LCI). Fischer's scoring system resulted 2 in 1 case, 3 in 1, 4 in 10, 5 in 11, 6 in 16, 7 in 12 and 8 in 17. ER, PgR, Ki67 and cerbB2 were positive in 38, 30, 40 and 15 lesions, respectively. Significant association was found between higher Fischer's scoring system and cerbB2 negative status (p = 0.002). No significant correlation between MR imaging single criteria and histopathological characteristics were found. Conclusion: Our study shows that combining MR imaging morphofunctional findings may identify breast cancer with different biological characteristics. In particular a significant correlation was found between Fischer's scoring system and expression of cerbB2 which is a marker of chemoresponsiveness and prognosis. Combination of fast initial wash in followed by a wash out was most common in G3: 12.9%; G2: 8.9%, G1: 2.1%; Gx: 5%. Mean curve peaks were: 172% (G1); 121% (G2); 133% (G3) and 98%(Gx). Conclusion: Despite still existing technical problems CAD offers for the first time a reliable, automated analysis of percentage distribution of enhancing types within a suspicious breast lesion. A discrimination of the grading seems, however not yet possible using the dynamic enhancement distribution offered by the CADsystem. One hundred and forty-four patients with pure DCIS without micro-infiltration or association to another infiltrating carcinoma (n = 146) were selected. Standard international dynamic protocol was used. Subtracted images were used to increase lesion conspicuity, and ROIs analysis to evaluate contrast uptake. Normalization of signal values was used in interpatient comparison. PI evaluation includes: wash-in, wash-out, time-to-peak, positive enhancement integration, standard deviation over time. These parameters were calculated in all cases. Findings in MRM were compared to a range of histological and biomolecular factors including: ki-67, p-53, hormonal receptors (estrogen and progesterone), HER2neu, VEGF and CD-31. Results: In 134 lesions enhancement was found (92 %). High grade DCIS showed significantly higher enhancement values than lower and intermediate grade and a more characteristic signal intensity curve. In parametric evaluation high grade DCIS is related with high wash-in and short time to peak values. Delayed subtracted images are needed in order to depict DCIS with low enhancement profile. High signal intensity and high velocity contrast uptake are related with high grade lesions This group of lesions shows overexpression of VEGF and increased capillary density in the surrounding stroma. Overexpression of HER2-neu is associated with continuous and dendritic growing pattern. Conclusion: Signal behaviour in DCIS appears to be related with biomolecular markers and quantitative evaluation is a promising tool in the diagnostic process. Purpose: To study the diagnostic value of multi-slice CT angiography for imaging of the spinal cord feeding arteries in patients with complex thoracic aortic pathologies undergoing endovascular aortic repair (EVAR). The impact of proximal posterior intercostal artery occlusion for the visualization of the spinal blood supply was evaluated. Method and Materials: Eighteen patients (5 women, mean age 60 years) with thoracic aortic type B dissection or aneurysm underwent two CTA examinations (pre-and post-EVAR, 120 kV, 120-150 mAs, 1 mm collimation, 16-slice CT scanner). Contrast agent (400 mgI/ml) was administered at a flow rate of 5 ml/s. Pulse rate and neurological status were documented. Qualitative analysis was performed by two experienced radiologists counting the posterior intercostal arteries, their rami dorsales and the Adamkiewicz artery (AKA). Density measurements included the ascending and descending aorta. Results: Contrary to expectations visualization of the posterior intercostal arteries and rami dorsales could be visualized post-EVAR due to retrograde perfusion. On average 10/11 posterior intercostal arteries, 8/9 rami dorsales were retrogradly perfused post-EVAR within the stented aortic segment (mean length 237 mm). Additionally, 9/10 AKA were visualized. ROI analysis in the thoracic aorta confirmed high mean HU values of 445 HU pre-and 385 HU post-EVAR. High pulse rates and partially thrombosed lumen in type B dissection and aortic aneurysm limited the visualization of the AKA in some patients. Neurological events were not observed in any patient. Conclusion: Multi-slice CT angiography is capable of visualizing even small spinal cord feeding arteries. High contrast imaging showed patent post-EVAR spinal vessels due to retrograde perfusion. After contrast application, arterial phase was focused on the suspected aneurysm. Delayed arterial phase was again extended from the aortic arch to the femoral arteries, whenever the pre-contrast phase demonstrated an unexpected aortic dilatation on the opposite side of diaphragm. According to the International Society for Cardiovascular Surgery guidelines, both suspected and unexpected aortic dilatations were classified as ectasia, or aneurysm with diameter below or above cut-off for surgical intervention. Results: Overall, 45 unexpected aortic dilatations were detected (21/45 ectasia, 15/45 aneurysm below cut-off, 9/45 aneurysm above cut-off); unexpected dilatations of thoracic aorta were observed in 28.3% of patients with suspected AAA, unexpected dilatations of abdominal aorta in 25.6% of patients with suspect AAT. Nineteen unexpected AAT (8/19 above cut-off) and 5 unexpected AAA (1/5 above cut-off) were observed: consequently in 9/163 (5.5%) cases the unexpected AA had the potential to change surgical planning. The purpose was to compare the diagnostic performance of 4-and 16detector MDCT angiography for evaluating peripheral arterial occlusive disease. Methods and Materials: 50 cases with symptomatic peripheral arterial occlusive disease underwent both MDCT angiography and DSA. In 28 of the 50 cases, MDCT angiography was performed by 4-detector system with 2-mm collimation with image reconstruction of 2-mm slice thickness at 1-mm intervals, and in the remaining 22 cases, it was performed by 16-detector system with 1-mm collimation and 1-mm slice thickness at 1-mm intervals. In evaluating MDCT data, crosssectional images perpendicular to the vascular course were observed by scrolling method. The arterial system from the iliac to the calf was divided into 9 segments. Pedal arteries were not included. Each segment was classified as normal, mildly stenotic, moderately stenotic, severely stenotic, or occluded. The diagnostic performance of MDCT angiography was determined using DSA as the reference standard. Results: Of the 911 segments studied, 892 were assessable on both MDCT and DSA. 312 steno-occlusive segments (73 mildly stenotic, 37 moderately stenotic, 60 severely stenotic, and 142 occluded) were detected on DSA. In the subgroup performed by 4-detector system, the sensitivity, specificity and accuracy of MDCT angiography for detecting segments with more than mild stenosis were 99.3%, 98.6%, and 98.2%, respectively, whereas those in the 16-detector system subgroup were 98.9%, 98.0%, and 98.2%, respectively. These did not reveal a significant difference. Conclusion: 4-as well as 16-detector MDCT angiography had high diagnostic accuracy for evaluating peripheral arterial occlusive disease. Peripheral angiography with 16-row MDCT: Optimization of acquisition speed using arterial enhancement quantification dependent on Purpose: To determine the potential of multidetector CT angiography (MD-CTA) in the evaluation of peripheral arterial occlusive disease (pAOD). Digital subtraction angiography (DSA) served as a standard of reference. Methods and Materials: 50 patients referred for DSA of the peripheral arteries due to known or suspected pAOD were prospectively included. All patients underwent MD-CTA prior to DSA using a 16-row multidetector CT scanner. All CT angiograms were reconstructed using a semi-automated reconstruction algorithm allowing visualization of entire vascular tree. For assessment, every leg was divided into 21 vascular segments, and scored in 5 categories of disease severity. Additionally, length and multiplicity of lesions within every vascular segment was assessed and compared to DSA. Purpose: To evaluate the optimal timing of delayed phase imaging for detecting low-flow endoleaks. Methods and Materials: Thirty-four patients with unruptured abdominal aortic aneurysm treated with endovascular repair (EVAR) underwent follow-up MDCT performed during unenhanced, arterial, and delayed phase (4x1-mm, a 1.25-mm slice width, and a pitch of 6). Delayed phase imaging, focused on stent-graft, was performed with a delay of 60 seconds (early delayed phase enhanced: set A) and 300 seconds (late delayed phase enhanced: set B) after intravenous injection of 120 ml of iodinated non-ionic contrast medium (Iomeprol 300 mgI/mL, Iomeron; Bracco, Milan, Italy), at a flow rate of 3 ml/s via an antecubital vein. Each set of delayed images was interpreted separately by two independent readers for the presence of endoleaks. Sensitivity, specificity, and diagnostic accuracy of both reading sessions were compared. The standard of reference was contrast-enhanced ultrasound, as suggested by recent literature. Results: Sensitivity, specificity, and diagnostic accuracy of set B (late delayed phase enhanced) were significantly higher (P < 0.05) than those obtained with set A (early delayed phase enhanced), for both readers. In patients undergoing MDCT after EVAR of abdominal aortic aneurysms, delayed scans should be performed 300 seconds after injection of contrast medium, in order to improve detection of low-flow endoleaks. Purpose: Dynamic stages of carpal instability are manifest only during wrist movement (dyskinematics). The aim of this study was to evaluate kinematography for detecting the radiographically occult stages of carpal instability. Methods and Materials: Twenty-four patients suffering from tenderness, click and/or snap phenomena and grip weakness were examined with kinematography using pulsed fluoroscopy (15 pulses/s). The kinematographic movies were stored on the PACS. Fifteen patients underwent arthroscopy, and in 9 patients surgical treatment was performed. Results: Standard radiograms were inconspicuous in all cases. Nineteen pathological movement patterns were found: 12 scapholunate dissociations (SLD), 2 lunotriquetral dissociations (LTD) and 5 midcarpal instabilities (MCI), 2 of these bilaterally. In the remaining 5 cases, no abnormal movement patterns were seen. There was a close correlation of kinematography to arthroscopic and surgical findings: 11/12 of the SLD and 2/2 of the LTD were confirmed, whereas in 1 case the SL ligament was assessed in arthroscopy to be only incompletely torn. One partial tear of the LT ligament was detected in the 5 patients with normal kinematograms. No information about the ligamentous state was available in the 5 patients with MCI patterns. Conclusion: Aside from physical examination manoeuvres, kinematography is the method of choice for detecting dynamic subcategories of carpal instability non-invasively. Purpose: In dynamic stages of carpal instability, there is a normal arrangement of the wrist at rest, although the intrinsic SL or LT ligaments are completely ruptured. The aim of this study was to evaluate contrast-enhanced MR imaging when dynamic carpal instability is suspected. Methods and Materials: Forty-five patients suspicious of having dynamic patterns of SL or LT dissociation were examined with contrast-enhanced MR imaging (0.2 mmol/kg) using axial T2*-w GRE, coronal fat-saturated PD-w FSE, plain and contrast-enhanced coronal T1-w SE, contrast-enhanced sagittal T1-w SE. All patients underwent arthroscopy. Results: Focal enhancement areas were found at the SL ligament in 25 patients, with the dorsal and volar segments involved in 16 patients and only one segment involved in 9 patients. Arthroscopy revealed 24 complete tears of the SL ligament, but could not confirm 6 lesions (sensitivity 79%, specificity 71%, PPV 76%, NPV 75%, accuracy 76%). In 12 patients, enhancement patterns were seen at the LT ligament, either dorsovolarly (7 cases) or volarly (5 cases). Arthroscopy revealed 14 tears of the LT ligament, and did not confirm 4 lesions (sensitivity 57%, specificity 87%, PPV 67%, NPV 82%, accuracy 78%). Statistics for MR imaging evaluation were significantly better in the acute/subacute rupture stage compared to long-standing instability patterns. Conclusion: Contrast-enhanced MR imaging allows moderately good non-invasive visualization of dynamic SL and LT by enhancing fibrovascular reparation tissue at the rupture site. This mechanism is less significant in chronic instability stages when carpal malarticulation has induced diffuse synovitis. Purpose: To evaluate wrist functioning, the analysis of dynamic 3D movement of the carpal bones in the wrist joint is crucial. Video fluoroscopy and animated 3D imaging of the wrist joint will not reveal abrupt changes as seen often with dynamic joint pathology. We created a system for 4-Dimensional rotational X-ray (4D-RX); using a 3D-RX system to image an object that is forced into a cyclic motion to obtain a series of time resolved 3D reconstructions. The image quality of our setup is investigated by imaging a post mortem wrist. A prototype image intensifier based 3D-RX system (BV Pulsera, Philips Medical Systems, Best, NL) is used together with a movement device. For 4D-RX, projection images from around a constant cyclic moving object are taken. The projection images are reconstructed and multiple 3D-RX reconstructions of the moving object at different motion phases are obtained. Results: The best quality was obtained using 41 projection images, obtaining 9 3D reconstructions. Although both the volume renderings and slice views show artefacts, the anatomical structures of the carpal bones are clearly visible. Moreover, a clear representation of the 3D joint kinematics is obtained. Conclusion: A first prototype shows the feasibility of 4D imaging of the wrist joint. We anticipate that dynamic imaging 4D-RX of the wrist and other joints for diagnostic purposes may be of great help for assessing functional disorders. Purpose: The rupture of the retinaculum of the extensor carpi ulnaris (ECU) tendon is an acute injury, more common because of lift and two hand technique, in professional tennis players. We study the value of US and MR imaging for the diagnosis and follow-up. Methods and Materials: Twenty-one professional tennis players were investigated by US and MR imaging after an acute injury of the wrist during a matchplay with suspicion of a lesion of the ECU tendon or its retinaculum. All patients had a dynamic assessment with pronosupination imaging. Imaging follow-up was performed over 2 to 3 months. Results: The most common injury combines a medial detachment of the ECU retinaculum and an avulsion pouch along the medial aspect of the ulnar head (n = 17), sometimes extending volarly to the pronotor teres muscle (n = 5). The radial avulsion is less common (n = 4), with a risk of interposition of the retinaculum beneath the tendon and lack of healing. Tendinopathy is common with chronic lesions (n = 17) and tenosynovitis (n = 8) in these professional players. The supi- Purpose: The aim of the study is to evaluate the potential MR imaging findings of painful postoperative carpal tunnel syndromes (CTS) compared with electromyography (EMG). Methods and Materials: At 1.5 T, 34 wrists in 34 patients with previous CTS release were evaluated by two observers for median nerve (signal, morphology and enhancement), for flexor retinacular regrowth, for fibrosis, for volar nerve migration and tunnel release measure, and for compressive abnormalities. The population is constituted of 11 controlled patients with normal electromyography (EMG), and 23 with electric abnormalities. There were 27 female and 7 male patients, mean age 55.7 years. Mean time between surgery and MR imaging was 29.9 months, and 3.8 months between EMG and MR imaging. Results: Retinacular section was seen in 12/23 with and 5/11 without recurrent CTS (p > 0.05). High T2 signal of the median nerve was seen in 20/21 with, and 7/11 without recurrent CTS (p = 0.036), enhancement was seen in 10/23 with, and 1/11 without recurrent CTS (p = 0.06). Mean ratio was 2.40 with and 1.89 without recurrent CTS (p = 0.018). Compressive abnormalities were present in 7/ 23 patients with, and in 5/11 without recurrent CTS (p > 0.05). Fibrosis was seen in 14/23 with, and 2/11 without recurrent CTS (p = 0.029). Conclusion: Fibrosis, proximal flattening and high signal of the median nerve are significant findings in recurrent CTS with electric abnormalities compared to a controlled population. Sonographic diagnosis of carpal tunnel syndrome: Diagnostic value of the triangular cross-section sign K.T. Szopinski, T. Mazurczak-Pluta, B. Lakomiec, R.Z. Slapa; Warszawa/PL (kazszo@poczta.onet.pl) Purpose: In some patients with carpal tunnel syndrome, the median nerve squeezed between the flexor retinaculum and flexor tendons takes a triangular shape. The purpose of this study was to determine the value of the triangular cross-section sign in the diagnosis of carpal tunnel syndrome. In 50 patients with hand pain (41 females and 9 males), 82 sonographic examinations of the median nerve were performed with an ATL HDI 5000 scanner, using a linear 7.5 MHz probe. All patients underwent an electromyographic (EMG) examination. Patients with bifid median nerve were excluded from this study. Results: Triangular shape of the median nerve in the carpal canal was observed in 7/63 (11%) cases of EMG-positive diagnosis of carpal canal syndrome. No such deformation was seen in patients with normal results of the EMG examination. Mean cross-sectional area of deformed median nerves in the carpal canal was 17.7 mm 2 (range 12-25 mm 2 ), mean cross-sectional area of EMG-positive median nerves was 15 mm2 (range 6-30 mm 2 ), mean cross-sectional area of median nerves with normal EMG results was 10.1 mm 2 (range 6-17 mm 2 ). Sensitivity, specificity, positive predictive value (PPV) and negative predictive (NPV) value of the triangular cross-section sign in diagnosis of carpal tunnel syndrome as compared with the electromyography were 0.11, 1.0, 1.0 and 0.25 respectively. Conclusion: Triangular shape of the cross-section of the median nerve in the carpal canal is a highly specific sign of carpal canal syndrome. The role of US in affecting the management of patients referred to a hand surgeon G.M. Allen, M. Tan, M. Dhillon, V. Rajaratnam; Birmingham/UK (gina_m_allen@btinternet.com) This was a prospective study of all patients referred for US of the hand, wrist and forearm conditions (including the elbow) in a hand unit. Referrals were made by three hand surgeons. The US was performed by one dedicated musculoskeletal radiologist. The clinical diagnosis was recorded from the request card and clinic notes. The ultrasound findings were recorded and it was noted whether the US made a difference to the decision in further management. The findings were compared with the final histological diagnosis for those patients who went to surgery. Results: A total of 31 patients are included in the study. The majority (24 patients, 77%) of these were referred to US with a single proposed diagnosis. Of these cases US confirmed the diagnosis in 15 patients (63%) but in the remaining 9 patients (37%) other diagnoses were made. Of the remaining 7 patients, 6 were referred with more than one clinical diagnosis and a single patient had no provisional clinical diagnosis. US agreed with one of the possible differentials in one third, and excluded several of the possibilities in patients who had previously undergone complex hand surgery in another third. In the final third of patients within this subgroup the clinical and US diagnoses were at variance. Conclusion: Ultrasound is a very useful adjunct to clinical assessment, altering preoperative diagnoses in a significant number of patients potentially saving the patient unnecessary surgery. Value of pulse-inversion harmonic imaging of inflammation activity in rheumatoid arthritis C. Schueller-Weidekamm, G. Schueller, C. Krestan, F. Kainberger; Vienna/AT (claudia.schueller-weidekamm@meduniwien.ac.at) Purpose: To evaluate the value of contrast enhanced pulse-inversion harmonic imaging (PI-HI) on both the synovial vascularization and the therapeutic effects on the inflammation of finger joints in rheumatoid arthritis (RA). Two hundred and eighty finger joints (MCP I-V, IP, PIP II-V) of 14 patients (10 females, 4 males) with RA were investigated with Power Doppler and PI-HI ultrasound after i.v. application of a second generation ultrasound contrast media (SonoVue, Bracco, Milan, Italy). With power Doppler, one finger joint per patient representing the strongest hypervascularization before steroid therapy was determined. Dynamic examination with PI-HI of the selected joint was carried out before and after 7 days of middle-high dosed steroid therapy. Vascularization was quantified by calculating the area underlying time-intensity curves [dB] . The changes of the signal intensities before and after therapy were correlated with clinical examinations (DAS-score). Results: All patients showed a statistically significant reduction of PI-HI signals after steroid therapy (p < 0.05). The baseline and follow-up median values of the area underlying time-intensity curves were 8.56 ± 1.28 and 7.65 ± 0.66 respectively. The median values of the DAS decreased significantly from 4.9 ± 0.86 to 3.6 ± 1.0 (p < 0.01) 7 days after the steroid therapy. Conclusion: PI-HI enables the detection of synovial perfusion alterations after steroid therapy, and therefore may be an additional useful tool in the evaluation of active inflammation in RA and in the assessment of the therapeutic response. Purpose: To assess the PIP alterations in rheumatoid arthritis of the finger using a high-resolution microscopy coil at 3 Tesla. Methods and Materials: High-resolution 3 Tesla MR imaging was performed in 15 patients with rheumatoid arthritis, using a microimaging gradient insert with gradients up to 200 mT/m. Coronal STIR (TR/TE 1800/14.6, TI 170 ms, 4 averages, FOV 70x35 mm, matrix 256x96, slice thickness 2 mm), axial T1w (TR/TE 538/15.3 ms, 10 averages, FOV 30x30 mm, matrix 256x192, slice thickness 1 mm), and contrast enhanced, fat-suppressed coronal and axial T1w sequences (TR/ TE 808/15.4, 4 averages, FOV 70x35 mm, matrix 512x148, slice thickness 2 mm) were performed. The images were evaluated according to modified EULAR-OMERACT criteria. Results: Synovial swelling and enhancement, bone marrow edema, erosions, and the involvement of collateral ligaments, tendons and tendon sheaths were clearly visualized. Interestingly, the most common findings were not only synovial swelling (15/15) and synovial contrast enhancement (14/15), but also thickening and enhancement of the collateral ligaments (14/15). In all patients with bone marrow edema (9 /15), the insertion zone of the collateral ligaments was involved. The erosions which were seen in 9/15 patients were exclusively located in the insertion zone of the collateral ligaments in 3/9 patients, partly in 5/9 patients, and distant to this zone in only 1/9 patient. C D E F B A G Conclusion: High-resolution 3 Tesla MR imaging of PIPs in rheumatoid arthritis allows detailed information about more anatomic structures involved than has been described previously. The collateral ligaments and the bone of the insertion zone of the collateral ligaments are commonly involved. Purpose: To evaluate therapeutic response of percutaneous radio-frequency thermal ablation (RFTA) of hepatocellular carcinoma (HCC) by SonoVue-enhanced pulse-inversion ultrasonography (CEUS) in comparison with multislice computed tomography (MSCT). Methods and Materials: 280 patients with 337 nodules (mean size 3.2 cm) diagnosed as HCC by biopsy and treated with RFTA were evaluated by CEUS and contrast-enhanced multiphase MSCT one month and three months after treatment respectively. On MSCT or CEUS we considered: a) complete ablation-when there was no enhancing portion within or at the margin of the ablation zone during the hepatic arterial phase; b) residual unablated tumor-an irregular peripheral-enhancing focus in the ablation zone after one month; c) local tumor progressionthe presence of growing enhancing tumors at the margin after three months but with no evidence of residual tumor in the previous control. Results: CEUS and MSCT revealed residual tumor in 67 (19.8%) and 72 (21.3%) nodules respectively, local tumor progression in 18 (5.3 %) and 20 (5.9 %) respectively and complete response in 252 (74.8%) and 255 (75.7%) respectively. Local tumor progression was confirmed by US-guided biopsies of the areas of growing enhancing tumors in all the 18 nodules assessed by CEUS. Conclusion: SonoVue-CEUS can be considered a reliable alternative technique to contrast-enhanced multiphase MSCT to assess therapeutic response after RFTA of HCC. Association between a wound-response gene expression program and hypodense tumor halos on CT in hepatocellular carcinoma M. Kuo 1 , J. Gollub 2 , C. Sirlin 1 , C. Ooi 3 ; 1 San Diego, CA/US, 2 Stanford, CA/US, 3 Hong Kong/HK (mkuo@ucsd.edu) Purpose: Elements of cancer have been compared to a wound healing response gone awry. A canonical "wound-response" gene expression program has been previously identified. We sought to determine whether hypodense tumor halos seen on CT were associated with the wound response gene expression program in HCC. Methods and Materials: 29 HCCs were analyzed across 29 CT scans for the presence or absence of hypodense tumor halos by 2 radiologists in consensus. Tumor halo scores for each tumor were then evaluated for enrichment of the wound-response gene expression program against matched microarray gene expression measurements across 3960 clones for the same 29 HCCs using supervised bioinformatic analyses. The strength of the enrichment for the wound response gene expression program was then determined and statistical significance evaluated. Results: Hypodense halos were present in 12/29 HCCs and absent in 17/29. Hypodense tumor halos were strongly associated with regulation of the woundresponse gene expression program (p < 0.05). Further, HCC with absent halos were associated with up-regulation of the wound-response gene expression program whereas presence of halos correlated with down-regulation. Additional analysis also demonstrated that absence of hypodense tumor halos was also correlated with regulation of a p53-associated HCC gene expression program. Conclusion: Hypodense tumor halos may represent an imaging surrogate for a "wound-response" tumor transcriptional program in HCC and may also incorporate elements of a p53 gene expression program. Purpose: To compare three-phase helical CT findings of hepatocellular carcinoma (HCC) in cirrhotic and non-cirrhotic livers of Korean patients, and to find out the difference of CT findings between noncirrhotic livers of Asian and non-Asian patients. Methods and Materials: During a 5-year period, 191 patients (165 men and 26 women, mean age; 52 years) who had three-phase helical CT were collected among 527 patients with surgically proven HCC. At pathology, 89 patients had cirrhotic liver (88 HBV infection, 1 HCV infection) and 102 patients had noncirrhotic liver (68 HBV infection, 5 HCV infection, 4 unknown chronic hepatitis, 25 normal liver). All CT images were analyzed retrospectively for various morphologic and enhancement patterns of tumor, tumor thrombosis and intrahepatic metastasis, and were correlated with pathologic findings. Results: More frequent morphologic patterns in noncirrhotic livers were ill-defined margins (p=.001), fibrosis (p <.001), necrosis (p <.001), and hemorrhage (p=.201). More frequently found patterns were capsule (p <.001), fatty metamorphosis (p=.127), and mosaic pattern (p=.118) in cirrhotic livers. For enhancement pattern, arterial enhancement and delayed washout were more frequently seen in non-cirrhotic liver without statistical difference. Intrahepatic metastasis and tumor thrombosis were more frequently found in non-cirrhotic liver (p > .05). Conclusion: Some morphologic findings of HCC in cirrhotic liver were found differently in non-cirrhotic livers. Reviewing the literature, our results of patients with non-cirrhotic livers were sufficiently similar to those of non-Asian patients. Therefore, we suggest that morphologic appearances of HCC are not dependent on the racial differences but on the presence of cirrhotic change. CTs obtained in all patients for initial staging were followed by a pharmacoangio-CT. Pharmaco-angio-CT technique: 0.05 mg of adrenaline hydrochloride diluted in 10 ml of normal saline are injected through a 3 French catheter into the common hepatic artery, four seconds prior to injection of a contrast bolus (4 ml of non-ionic contrast medium in 20 ml of normal saline at a flow rate of 3-4 ml /sec). Multi-detector spiral CTs were obtained with a four second delay. A second CT series was generated in the portovenous phase. Purpose: To assess multi-detector row helical computed tomography (MDCT) findings with histopathology in patients with hepatocellular carcinoma (HCC) who had undergone living-donor liver transplantation (LDLT). This prospective study comprised 41 patients (28 men, 13 women, age 30-68 years, mean 55 years) with HCC admitted for LDLT. Results of preoperative MDCT were correlated with histopathologic results after 5mm-slicing of the explanted liver. After 2 mL/kg of 300 mg I/mL of contrast medium were injected, late arterial and equilibrium phase images were obtained. MDCT was performed using following parameters: detector row configuration, 2 x 8 mm, reconstruction interval;1 mm, 200 mAs; and 120 kVp. Image analysis was performed in consensus by three abdominal radiologists using 2 mm thickness with 1 mm overlap images with paging method as well as multi-planner reformat images (i.e. axial, coronal and sagittal images). Results: A total of 134 HCC lesions (mean diameter: 21 mm) were confirmed pathologically. The sensitivity, specificity and positive predictive value for the detection of HCC was 87%, 96% and 84%, respectively for all lesions; 98%, 99% and 97%, respectively for lesions larger than 1 cm; and 46%, 76%, 41%, respectively for tumors smaller than 1 cm (n = 28). There were 12 false negative lesions whereas only five false positive findings. Four tiny HCCs were misdiagnosed as arterial-portal shunts. Conclusion: Multi-detector row CT by viewing images on digital monitors is very effective in detecting HCCs larger than 1 cm with only a few false-positive lesions. Detection of hepatocellular carcinoma ( Purpose: To compare Gd-BOPTA-enhanced MR imaging with MSCT for the detection of HCC in patients with cirrhosis. Methods and Materials: 52 patients with 94 HCCs underwent MR imaging on a 1.5 T scanner (Magnetom Vision Plus, Siemens) and CT on a multislice scanner (Somatom Plus 4 Volume Zoom, Siemens). For MR imaging, unenhanced T2weighted HASTE and T1-weighted FLASH sequences were acquired. After intravenous injection of 20 mL of Gd-BOPTA (0.1 mmol/kg; MultiHance, Bracco), arterial, portal-venous, and equilibrium phase T1-weighted images were obtained at 25, 60, and 180 sec, respectively. A hepatospecific phase was obtained 1 hour after the injection of contrast material. For MSCT, unenhanced, arterial, portalvenous, and equilibrium phase images were obtained before and 35, 70, and 180 sec, respectively, after intravenous injection of 120-150 mL of nonionic contrast material (300 mgI/mL). CT imaging parameters were: collimation, 4 x 2.5 mm, section thickness, 3.0 mm; effective mAs, 160; and kVp 120. Three observers independently interpreted the images obtained with each technique. The alternative-free response receiver operating characteristic (A (z)) method was used to analyze the results. Sensitivity and positive predictive value were also calculated. Results: Mean sensitivities of MR imaging and CT were 84% and 85.1%, respectively (p not significant). Mean positive predictive values of MR imaging and CT were 86.2% and 87%, respectively (p not significant). There was also no significant difference in A (z) values between MR imaging (0.81) and CT imaging (0.78) (p not significant). Conclusion: Gd-BOPTA-enhanced MR imaging and MSCT showed similar diagnostic accuracy for the detection of HCCs in patients with cirrhosis. Gd-EOB-DTPA enhanced MR imaging versus triphasic CT: Detection and classification of focal liver lesions C. Kölblinger, A. Ba-Ssalamah, F. Längle, T. Grünberger, F. Wrba, W. Schima; Vienna/AT (claus.koelblinger@meduniwien.ac.at) Purpose: To compare the performance of triphasic CT and Gd-EOB-DTPA (Primovist, Schering, Berlin, Germany) enhanced MR imaging in detecting and classifying focal liver lesions. Methods and Materials: CT and Primovist enhanced MR imaging scans of 29 patients, recruited from phase II and III studies, with focal liver lesions (a total of 106 lesions) were evaluated. The gold standard was either defined by histopathologic correlation, intraoperative ultrasound or imaging follow-up. Images were obtained by using triphasic CT and dynamic Gd-EOB-DTPA enhanced MR scans (1.5 Tesla). Additionally late phase MR scans 20 and 40 minutes after contrast media injection were acquired. Two radiologists, blinded to the gold standard, evaluated the images in consensus reading with respect to size, number, location and morphological characteristics of the detected lesions. Results: From a total of 106 lesions 68% could be revealed with CT and 89% with MR imaging, respectively. The correct diagnosis of the detected lesions was accomplished in 74% with CT and in 94% with MR imaging. Conclusion: MR imaging with Gd-EOB-DTPA markedly improves detection and classification of focal liver lesions compared to triphasic CT. Enhanced examination included the dynamic study (arterial, portal and equilibrium phases) and the delayed study performed 20 minutes after contrast injection. Three radiologists independently interpreted the images for detection of HCCs, according to a four point confidence scale. The diagnostic accuracy was evaluated using the receiver operating characteristic method and the sensitivity of each imaging technique was compared. The mean diagnostic accuracy of dynamic MR imaging was higher than that of delayed MR imaging (Az = 0.94 vs Az = 0.89), although the difference was not statistically significant. The mean sensitivity of dynamic MR imaging was significantly higher (p < 0.05) than that of delayed MR imaging. Conclusion: Gd-EOB-DTPA enhanced dynamic study seems to be more accurate than Gd-EOB-DTPA enhanced delayed study in the detection of HCC. Small focal hypervascular lesions seen in the arterial phase of Gdenhanced MR imaging in cirrhotic patients: Differentiation of benign from malignant based on pre-Gd SPIO-enhanced signal characteristics K.S. Mangat, D. Aguirre, C. Sirlin; San Diego, CA/US (kam_mangat@hotmail.com) Purpose: Unidentified bright objects (UBO) are < 2 cm, circumscribed foci of arterial enhancement seen frequently on Gadolinium (Gd)-enhanced magnetic resonance (MR) images of cirrhotic livers. We tested the hypothesis that we can differentiate benign (BUBO) from malignant (MUBO) lesions based on super paramagnetic iron oxides (SPIO) uptake. CC failed in assessment of 246/408 potentially visible colonic segments. Stop Point on CC examination was predominantly in the sigmoid. Additional lesions identified in proximal colon in 19 % of patients, include3 primary carcinoma, 12 polyps in 8 patients, 2 lipomas. In one patient with incomplete colonoscopy at sigmoid a mesenterial liposarcoma was found. The main causes of incomplete CC was; redundant colon, tortuous colon loops and occlusive mass. Level of tolerance was rated good as well as patient compliance and patient preparation. Mean duration time of CTC was 9 minutes. Conclusion: CTC is feasible and effective technique in evaluating entire colorectal segments of the colon not visualized during colonoscopy. CTC has an undeniable role in elderly and frail patients in whom conventional colonoscopy is difficult to be fully performed. Purpose: To prospectively compare the diagnostic performance of MSCTC without cathartic preparation with that of colonoscopy for the detection of colorectal polyps in asymptomatic average-risk patients. Methods and Materials: 180 asymptomatic patients referred to undergo screening colonoscopy were recruited. For fecal tagging, an oral iodinated contrast agent (diatrizoate meglumine and diatrizoate sodium) was administered beginning 48 hours prior to CT. No cathartic preparation was used prior to CT. All CT examinations were acquired using a MSCT scanner (Somatom Plus 4 Volume Zoom, Siemens) using collimation, 4 x 2.5 mm; section thickness, 3.0 mm; effective mAs, 10; and kVp, 140. Colonoscopy with segmental unblinding was performed 3-7 days after CT as reference standard. Three radiologists, unaware of colonoscopic findings, analyzed CT colonographic images using a dedicated software (Vit-rea, Vital Images). Sensitivity was calculated on both a per-polyp and a per-patient basis. In the latter, specificity was also calculated. Interobserver agreement was calculated using the k statistic. Results: Colonoscopy detected two carcinomas and 76 polyps in 42 patients; the remaining 138 patients showed no abnormality. CT colonography yielded an average per-polyp sensitivity of 48.6% (92% for polyps ≥ 8 mm). Average perpatient sensitivity and specificity were equal to 92.8% and 94.2%, respectively. Interobserver agreement was high on a per-polyp basis, and high to excellent on a per-patient basis. In asymptomatic patients at average-risk for colorectal cancer, MSCTC without cathartic preparation has a diagnostic performance similar to that of colonoscopy for the detection of colorectal polyps ≥ 8 mm. Automatic unseen colon surface detection in 3D MDCT colonography A. Graser 1 , C.R. Becker 1 , L. Guendel 2 , M.F. Reiser 1 ; 1 Munich/DE, 2 Forchheim/DE (anno.graser@med.uni-muenchen.de) Purpose: In primary 3D read of CT colonography (CTC), visualization of the entire colonic wall remains a problem. Even after fly-through from rectum to cecum and vice versa not all of the colon might have been visualized. We evaluated the percentage of the unseen surface after forward and backward fly-through and performance of a computerized tool that automatically displays unseen areas to the reader. Methods and Materials: 40 data sets with different patient positions (20 prone, 20 supine) were randomly selected from a database and investigated on a CTC workstation (syngo Colonography, Siemens). The percentage of unseen surface was noted after one single flight from rectum to cecum and after an additional flight from rectum to cecum. Direct navigation to unseen areas was performed in all cases. Results: After one single fly-through from rectum to cecum, only 71.7-88.2% (mean, 81.3%) of the colonic surface had been visualized. A 2 nd flight increased these values to 91.4-98.8% (mean, 96.2%). The "unseen areas" tool implemented in the CTC software sorted unseen areas according to their size and displayed them to the reader. Conclusion: 3D reading with 2D for problem solving leads to an incomplete view on the colon surface even if bidirectional fly-through is performed. A tool which measures the percentage of colon surface that has been seen by the radiologist and automatically displays unseen areas prevents perception errors and may cut reading time. CT colonography: Automated diameter measurement of colonic polyps compared with a manual technique: In vitro study using a human colectomy specimen S.A. Taylor 1 , A. Slater 2 , L. Honeyfield 1 , M. Roddie 1 , S. Halligan 1 , J. Dehmeshki 1 ; 1 London/UK, 2 Oxford/UK (csytaylor@yahoo.co.uk) Purpose: To investigate the accuracy of automated and manual CT polyp measurement using a colectomy phantom. Methods and Materials: A colectomy specimen from a patient with familial adenomatous polyposis containing 27 polyps was insufflated, submerged, and scanned using a MDCT. A histopathologist measured the maximum dimension of all polyps in the opened specimen. Digital photographs and line drawings were produced to document exact polyp location. Two observers (novice and expert) independently estimated polyp diameter using three methods: software calipers 2D ('manual 2D'), software calipers 3D ('manual 3D') and automated polyp measurement software (automatic) [Medicolon 1.3, Medicsight plc]. The software requires the placement of 2 seed points in the polyp vicinity prior to automated segmentation and measurement. Data was analyzed using a paired t-test and Bland-Altman limits of agreement. Results: 6 polyps (all flat and ≤ 6 mm) could not be extracted by the software and were excluded, leaving 21 polyps (4-15 mm) for analysis. Automated polyp measurement was not significantly different from true histological size for either reader. However both observers significantly underestimated polyp size using manual 2D (mean difference [md] -1.3 mm, sd 1.3, p = 0.001, and md -1.6 mm, sd 1.2, p < 0.001) and using manual 3D for the expert only (md -0.6 mm, sd 1.4, p = 0.04). Bland Altman limits of 95% limits of agreement were similar for all three methods (automated -3 mm to 3.5 mm, manual 2D -3.8 mm to 1 mm, manual 3D (-3.2 mm to 2 mm). Conclusion: Automated polyp measurement is technically feasible, does not require reader expertise, and is potentially more accurate than manual measurement. Purpose: To evaluate the role of CT colonography (CTC) in the follow-up of asymptomatic patients that underwent polypectomy. Methods and Materials: Forty patients that underwent a previous polypectomy (mean follow-up at 1 year), were studied with CTC at low dose, with no contrast and a standard preparation (no feacal tagging); no muscle relaxants were used. Examination was performed in supine and prone decubitus. Patients were asked to undergo colonoscopy after CTC. Results: CTC found 34 polyps in 17 patients, that were all submitted to colonoscopy; in the remaining 23 patients (negative for CTC findings), only 9 underwent colonoscopy that confirmed CTC results (no FP or FN results). The per polyps sensitivity was 71% for those with diameter < 6 mm (6 FN) and 92% for larger polyps (1 FN). The per polyps specificity was 65% for those with diameter < 6 mm (7 FP) and 91% for larger polyps (2 FP). Conclusion: CTC represents a valid alternative to colonoscopy in the follow-up of asymptomatic patients with previous polypectomy. Purpose: The risk of malignant transformation of an adenomatous polyp is related directly to its diameter but conventional estimates of diameter made from CT may vary due to measurement error. We investigated whether automated segmentation and diameter assessment was feasible for CT colonography. Methods: Two experienced observers independently measured the maximum diameter of fifty polyps (colonoscopic reference size range 5-12 mm) from colonography data sets using conventionally placed software calipers and a variety of 2D window settings (colon, abdominal, bone, lung CT), and also 3D perspective rendering. Polyps were re-measured subsequently using in-house automated polypsegmentation software. Agreement between observers and with the colonoscopic reference were determined using Bland-Altman, Wilcoxen, and Mann-Whitney-U analyses. Results: Inter-observer agreement was similar for all window displays: Mean difference in mm (SD difference; 95% limits of agreement) ranged from 0 (1.7; -3.3, 3.3) for 2D colon to -1.1 mm (1.6; -4.3, 2.0) for 3D, compared to -0.5 (2.09; -4.6, 3.6) for automated measurement. When compared to colonoscopy; the largest mean size difference obtained by both observers occurred using the 3D display (1.3, 2.5 mm). There was also significant difference between observers and reference size when using the 2D abdominal and 3D displays (P =.03, p <.001). Subset analysis of tagged cases revealed that one observer overestimated diameter for untagged cases measured using 3D (median difference 3 mm versus 2 mm, p = 0.01). Conclusion: Automated and manual methods of polyp measurement results in similar inter-observer agreement. The greatest measurement error is encountered when using a 3D display for estimates of diameter. Impact of computer-aided diagnosis (CAD) on radiologists' performance to detect colonic polyps at CT colonography M. Cadi, O. Lucidarme, D. Touitou, P. Grenier; Paris/FR (macadi@mac.com) To determine potential improvement provided by computer-aided diagnosis (CAD) system in the detection of polyps at CT colonography (CTC). Two different readers, one senior radiologist (R1), and one junior radiologist (R2), independently evaluated CTC exams of 40 patients referred for screening colonic polyp or cancer, obtained with a MDCT scanner (16 x 1.25 mm, 1 mm reconstruction interval, 120 kV, 80 mAs). Readings were performed first without CAD using a 2D first approach (Colonography software, ADW4.2, GEMS), then with CAD (Mediantechnologies Sophia Antipolis, France). The standard of reference was established from the results of classical colonoscopy that had been also performed in the same patients. Because the CAD tool was not designed for the detection of polyps smaller than 5 mm, readers considered only polyps above that threshold. The classical colonoscopy identified 19 polyps greater than 5 mm in 14 patients. CTC sensitivities to detect polyps without and with the CAD tool were respectively 89.4% and 94.7% for R1, and 57.8% and 94.7% for R2. CAD alone selected 195 candidates for polyps. 158 and 162 of these candidates were correctly identified as false positives of CAD (bulbous fold 41%, stools 16%, ileocecal valve 9.7%) by both readers respectively. Conclusion: CAD increased sensitivity of CTC to detect colonic polyps. Effect of CAD was greater with the beginner reader. CAD may help reduce false negative diagnoses at CTC, and improve the learning curve of junior radiologists. Performance of a prototype computer-aided detection (CAD) system in MDCT colonography A. Graser 1 , L. Bogoni 2 , M.F. Reiser 1 , C.R. Becker 1 ; 1 Munich/DE, 2 Malvern, PA/US (anno.graser@med.uni-muenchen.de) Purpose: To evaluate the accuracy of a CAD system in CT Colonography as compared to expert readers. Methods and Material: 145 individuals (82 men, 63 women, mean age 59 years) underwent MDCT colonography for colon cancer screening. Scans were performed using 16-detector row scanners (Somatom Sensation 16, Siemens Medical Solutions, Forchheim, Germany) at 0.75 mm collimation. Individuals were scanned supine and prone without fecal tagging. 1 mm axial slices were reconstructed for 3D visualization. Two expert readers read the scans using a primary endoluminal approach. A prototype CAD system (polyp-enhanced view, PEV, Siemens Medical Solutions, Malvern, PA) was used as "second reader". The system's sensitivity for the detection of polyps and the number of false-positive findings per dataset was evaluated. Average running time of CAD was documented. Polyps were classified as small (≤ 5 mm), medium (6-9 mm) and large ≥ 10 mm). In the study population, a total of 98 polyps (small-51, medium-32 and large-15 polyps) were found in 42 patients. CAD detected 69 of these polyps (overall sensitivity of 70 %) with an average of 3.8 false positives per dataset. Average CAD running time was 4.8 minutes. Sensitivity was 51 % (26/51) for small, 94 % (30/32) for medium and 87 % (13/15) for large polyps. Conclusion: Computer-aided detection shows high sensitivity in detecting clinically significant polyps > 5 mm with acceptable false-positive rates. These numbers are within the range of expert readers. The short calculation time of the CAD system makes it useable in routine MDCT colonography. Effect of computer aided detection as a second reader in multidetector CT colonography: A multiobserver study T. Mang 1 , P. Peloschek 1 , C. Plank 1 , A. Maier 1 , C.J. Herold 1 , L. Bogoni 2 , W. Schima 1 ; 1 Vienna/AT, 2 Malvern, PA/US (thomas.mang@akhwien.at) To assess the effect of computer aided detection (CAD) as a second reader for lesion detection in MDCT colonography. To compare the influence of CAD on the performance of multiple readers with different levels of experience. Methods and Materials: 33 MDCT colonography data sets (21 patients: 30 endoscopically confirmed polyps ≥ 5 mm, 5 carcinomas, 12 patients with no polyps) were reviewed by four radiologists (2 experienced, 2 inexperienced) on a 3D workstation. Virtual endoscopy was used for primary data evaluation with additional 2D views for lesion confirmation. After primary data evaluation, second reading in consensus with findings of a CAD system (polyp-enhanced view, Siemens) was performed. Sensitivities and false positive rates were calculated for the different readers who were and were not in consensus with CAD findings. Results: Sensitivity of experienced readers was 97% each (29/30) for the detection of polyps. Inexperienced readers reached a sensitivity of 90% and 77%. CAD found 27/30 lesions (90%) and increased sensitivity of inexperienced readers to 97% each, with no effect on sensitivity of experienced readers. All carcinomas were found by 4 readers, 3/5 was found by CAD. False-positive rate was one for both experienced, and 5 and 6 for unexperienced readers, respectively. CAD decreased false positive rate of inexperienced readers from 5 and 6 to 2 each. CAD consensus reading increased reading time by a mean of 3 minutes. Conclusion: CAD used as a second reader improves sensitivities and reduces false positive rates for polyp detection of inexperienced readers. There was no significant effect on the performance of experienced readers. CAD in virtual CT colonoscopy without bowel cleansing: Initial experiences Z. Tarján, J. Koloszár, G.C. Kovács, J. Wacha; Budapest/HU (tarjan@radi.sote.hu) The thorough cleansing of the colon is usually a prerequisite of virtual colonoscopy similarly to conventional colonoscopy. Nevertheless, the cleansing may be one of the main causes of patient discomfort. We evaluated the C D E F B A G possibility of a CAD system in detecting polyps in preparationless CT colonography. Methods and Materials: 21 patients with colonic polyps and 20 patients without colon pathologies underwent preparationless CT colonography and optic colonoscopy within a week. The patients had no food restrictions or colonic cleansing in the days before the CT examination, only 6 times 20 ml of ionated radiologic contrast material (not containing meglumin) was administered orally 48 hours prior to the examination. Native supine low-dose 2.5 mm collimated scans were obtained on a 4 row scanner. The data were evaluated using an academic software (ColVis) equipped with electronic cleansing and CAD. The results generated by the CAD and that of a radiologist were compared. Results: 4/41 patients had diarrhea due to the preparation. CAD detected 15/15 polyps > 10 mm and 15/19 polyps between 5-10 mm with 17.3 ± 12.6 (mean, SD) false positives/case. The radiologist could detect 15/15 polyps > 10 mm and 16/19 polyps between 5-10 mm with 0.5 (mean) false positives/case. Conclusion: Virtual colonoscopy without colonic cleansing minimizes patients' discomfort before the examination. CAD combined with electronic cleansing detects large and small polyps similarly to a radiologist but with more false positives than in a clean colon. The aim of the current study was to show if CNR can be improved without loss of diagnostic accuracy for the assessment of myocardial infarction, if a PSIR single shot trueFISP sequence is used at 3.0 T instead of 1.5 T. Methods and Material: 10 patients with myocardial infarction were examined at a 1.5 T MR system (Avanto, Siemens, Medical Systems) and at a 3.0 T MR system. Imaging of delayed contrast enhancement was started 10 minutes after application of contrast media. A phase-sensitive inversion recovery (PSIR) single shot trueFISP sequence was used at 1.5 and 3.0 T and compared with a segmented IR turboFLASH sequence at 1.5 T, that served as reference method. Infarct volumes and CNR of infarction and normal myocardium were compared with the reference method. The PSIR single shot trueFISP technique allows to image 9 slices during a single breathhold without adaptation of the inversion time. The mean value of CNR between infarction and normal myocardium was 5.9 at 1.5 T and 9.5 at 3.0 T. The CNR mean value of the reference method was 8.4. The CNR mean value at 3.0 T was significantly (p = 0.03) higher than the mean value of the reference method. The correlation coefficients of the infarct volumes, determined with the PSIR single shot trueFISP technique at 1.5 Tesla and at 3.0 Tesla and compared to the reference method, were r = 0.96 (p = 0.001) and r = 0.99 (p = 0.0001). The use of PSIR single shot trueFISP at 3.0 Tesla allows for accurate detection and assessment of myocardial infarction with significantly improved CNR. Methods: Sixteen patients with acute MI, and 21 patients with chronic left ventricular ischemic dysfunction underwent MDCT and CE-MR. CE patterns on MDCT images acquired early (at the time of coronary imaging), and 10 minutes later were compared with those obtained with MR. Results: Early MDCT images demonstrated areas of contrast hypo-enhancement in 11/16 patients with acute and in 7/21 patients with chronic MI. On segmental basis, there was moderately good concordance (92%, κ=0.54, p < 0.001) in the identification of early hypo-enhanced regions between MDCT and MR, Quantitative size of early hypo-enhanced regions was similar (6 ± 16 vs. 7 ± 16 g, p = 0.25), and highly correlated (r = 0.93, p < 0.001) to MR. Late MDCT images revealed areas of delayed hyperenhancement 15/16 patients with acute MI and in 19/21 patients with chronic LV ischemic dysfunction. There was good concordance (82%, κ=0.61, p < 0.001), in the identification of these hyper-enhanced regions between both techniques on segmental basis. In absolute terms, size of hyperenhanced area by MDCT was also similar (6 ± 34 vs. 31 ± 40 g, p = 0.14) and highly correlated (r = 0.89, p <.001) to MR. Conclusion: CE-MDCT demonstrates similar contrast enhancement patterns in acute and chronic myocardial infarcts to CE-MR. It may thus provide important information on infarct size and viability at the time of non-invasive coronary imaging. were evaluated using a segmental model with two different methods: a visual 5points scale score (0=non hyperenhancement, 4=hyperenhancement > 76% of myocardial wall) and with quantitative analysis based on manual contours tracing followed by automatic threshold analysis (Cardiac Analysis, ViewForum 4.1, Philips, NL). Each segment was also assigned a wall motion score on B-TFE images ranging from 0-normokinesia to 4-dyskinesia. Statistical evaluation was performed. Results: Mean LV ejection fraction was 49.3 ± 14.9 with EDV and ELV respectively of 149 ± 47.8 and 80.7 ± 44.3. Of 1280 segments 322 (25.1%) showed wall motion abnormalities with enhancement in 327 (25.5%) evaluated with visual score and in 414 (32.3%) quantitatively. Among segments with normal or mild hypokinesia 89.2% had a delayed enhancement score ≤ 1 whereas 80.2% of akinetic a dyskinetic segments had a score ≥ 3, confirming the relationship between transmurality of infarction and contractile impairment. Mean time requested by visual and quantitative approach was respectively 7 ± 3 and 18 ± 9 minutes. There was high agreement between visual and quantitative methods (K=0.92; p < 0.01); visual analysis underestimated 104 (8.1%) and overestimated 19 (1.5%) segments. Conclusion: Visual analysis of delayed enhancement is a time saving approach, sufficient to assess the transmurality and spatial extent of infarction; moreover it has high correlation with wall motion abnormalities. Purpose: To evaluate if interventional MR imaging allows targeting intramyocardial injections at the border of myocardial infarcts after intravenous administration of the small particle of iron oxide SHU555 A. In six pigs with reperfused infarction three hours after intravenous injection of SHU555 A a needle-catheter was advanced into the left ventricle under real time MR imaging guidance using a steady state free precession sequence. 2 ml of a 0.1 mmol/ml gadodiamide solution, mixed with 3 % Evans blue dye, were injected at two regions at the border of the infarct. T1 values were serially measured prior and 3, 15 and 30 min after each intramyocardial injection. Results: Injection of SHU555 A caused a decrease of the T1 value of the infarcted myocardium (778 ± 63 ms before, and 588 ± 54 ms 3 h after injection). The contrast between the infarct and remote myocardium was sufficient for delineation of the infarct on real-time images (0.43 ± 0.02). In all animals, the catheter could be directed into the border region of the infarct. In each case, two injections were performed at different locations. T1 values at the injection sites decreased (240 ± 25 ms) and slowly increased (468 ± 37 ms 30 min after intramyocardial injection). Conclusion: Percutaneous, MR-guided intramyocardial injection of a gadodiamide containing solution into the border of infarcted myocardium is feasible after the ischemically injured region is delineated following i.v. injection of SHU555 A. Due to the strong decrease of T1 values at the injection sites repeated injections into the same region can be avoided. Purpose: The purpose of this study was to compare energy metabolism in noninfarcted and infarcted myocardium using 31P-MR-Spectroscopy (MRS) and to correlate these changes with the outcome of the revascularization. Methods: 30 patients with subacute anterior myocardial infarction (MI) underwent MRS and 31P-spectra were obtained from non-infarcted/septal as well as infarcted/anterior myocardium (voxel size 25 ml each) by means of 3D-CSI. Phospocreatin/adenosine triphosphate (PCr/ATP) ratios were determined using "AM-ARES". Left ventricular (LV) wall thickness was evaluated by cine magnetic resonance imaging (MRI). To assess myocardial viability, MR imaging was repeated 6 months after revascularization. 15 age matched volunteers served as control. Results: 15 of 30 patients regained normal wall motion (VIAB), whereas the other 15 not (NVIAB). PCr/ATP-ratios of infarcted/anterior myocardium were significantly reduced for both groups (1.17 ± 0.25 (VIAB) and 0.71 ± 0.70 (NVIAB); 1.51 ± 0.32 (control)). Regarding non-infarcted/septal myocardium, a reduction of PCr/ATP ratios was found only in NVIAB. Reductions of PCr/ATP ratios did not correlate with decreased wall thickness in any group. Conclusion: Irreversibly damaged myocardium is associated with severe depression of energy metabolism, which in turn affects even non-infarcted myocardium independent of wall thickness at the subacute stage of MI. MR imaging before CABG surgery: Impact of scar extent in contrastenhanced MR imaging on coronary graft flow P. Hunold, P. Massoudy, T. Schlosser, C. Böhm, J. Forkel, J. Barkhausen; Essen/DE (peter.hunold@uni-essen.de) Purpose: To relate myocardial late enhancement (LE) in contrast-enhanced MR imaging (CE-MRI) to intraoperative graft flow in patients with impaired LV function undergoing CABG surgery. Methods and Materials: Thirty-three CAD patients (mean EF 29%) underwent CE-MRI before CABG surgery. LE scans were acquired after 0.2 mmol/kg Gd-DTPA using an inversion-recovery turboFLASH sequence (TR, 8 ms; TE, 4 ms; FA, 25°; TI, 200-260 ms). Myocardial scar extent was evaluated using a LE score 1-4 (score 1, no LE; 2, LE < 50% of wall thickness; 3, LE > 50%; 4, transmural LE) for each vessel territory (RCA, LAD, LCX). Intraoperative flow as determined by Doppler sonography in grafts supplying vessel territories with a score 1 or 2 (group A) was compared to flow in grafts supplying vessel territories with a score ≥ 3 (group B) which are not expected to improve function after revascularization. Results: Of the 99 vessel territories, 89 yielded score 1 or 2 (group A), 10 score ≥ 3 (group B). In group A, 68 of 89, and in group B, 9 of 10 territories were grafted using left intrathoracic artery (LITA) or venous graft. Mean LITA graft flow was 54 ± 6 in group A and 28 ± 11 cc/min in group B (p=.04). Mean vein graft flow was 74 ± 6 and 42 ± 8 cc/min in groups A and B, respectively (p=.05). Conclusion: Graft flow to vessel territories with no or little scar tissue is higher than flow to territories with > 50% scar tissue. Since sufficient flow is a prerequisite for graft patency, the prognosis of those grafts may be better. Between clinical parameters measured at admission we choose the following: Qwave presence, CK-MB peak, symptoms-to-balloon time (min). Cine and contrast enhanced MR images were evaluated with a segmental semi-quantitative score: wall motion score (WMS) and segmental scar score (SSS). Patients with multivessel disease and/or previous infarction. Univariate and multivariate statistical evaluation was performed in the remaining 38 subjects. C D E F B A G The mean symptoms-to-balloon time was 192.8 minutes. Q-wave was present in 27 subjects (71%). Mean CK-MB peak was 235 ± 98.8 ng/ml in patients who exhibit Q-waves and 171.4 ± 102.3 in the remaining. Most of segments with contractile impairment exhibit delayed enhancement (p = 0.03). Q-wave presence correlates with infarct extension as quantified by segmental scar score (25.1% VS 12.8%; β=7.99; p = 0.01) and wall motion abnormalities (6.3% VS 18%, β=7.32; p = 0.02). Also high CK-MB peak positively correlates with SSS (β=0.05; p = 0.00) and WMS (β=0.04; p = 0.00). MR imaging parameters did not significantly correlate with Symptoms-to-balloon time (p = 0.70 and p = 0.36 respectively). Multivariate analysis confirmed those correlations. Conclusion: Q-waves presence and CK-MB peak are better predictors of both spatial and transmural extension of infarction and contractile impairment with respect to other clinical parameters. Multislice multiecho T2* MR imaging approach for the detection of myocardial iron distribution: Validation in healthy subjects and thalassemia major patients A. Pepe 1 , E. Cracolici 2 , V. Positano 1 , L. Pitrolo 2 , L. Prossomariti 3 , D. De Marchi 1 , M. Midiri 2 , M. Lombardi 1 ; 1 Pisa/IT, 2 Palermo/IT, 3 Napoli/IT (alepepe@email.it) Purpose: To validate the multislice multiecho T2*MR imaging approach, for the detection of myocardial iron distribution, in healthy subjects and in thalassemia major (TM) patients. Three short axis views of the left ventricle were obtained using T2* multiecho sequences in 20 healthy subjects and in 53 TM patients. The myocardium was segmented into 12 segments. T2* value on each segment and the global T2* value were calculated. To assess the endomyocardial and epimyocardial iron distribution, each segment was divided in the inner and outer half. Results: The coefficients of variation for intra-, inter-observer and inter-study reproducibility were 3.9%, 5.5%, 4.7% respectively. In healthy subjects the percentage of deviation (%D) of the slices from the global heart T2* mean was not statistically significant. The %D of the segments from the global heart T2* mean ranged from -20 to 18. Significant differences were observed only among the segments with higher %D (5/66 cases). In TM patients, the epimyocardial iron concentration was significantly higher in the slices and in 10/12 segments. Three groups of patients were identified: 15% with homogeneous myocardial iron overload (MIO); 47% with no MIO; 38% with a heterogeneous MIO (T2* segments < 20 ms and T2* segments > 20 ms). Serum ferritin levels, liver iron concentrations and urinary iron excretions were significant different among groups. Conclusion: Our data did not account for rejecting certain slices or segments. Multislice multiecho T2* MR imaging approach provides a reproducible way for assessing myocardial iron distribution, which may play a role in clinical arena. (24) and toxic adenomas (41). Nodular volumes were 2.5-38 ccm (mean 20.8). 96% ethanol was used, with total volumes injected (4-57 ml) of 1.5 times nodule volume. Procedure was performed twice a week until total volume injected. 3-6 months after completion of PEI outcome evaluation was performed by hormonal status (T3,T4,TSH), scintigraphy and US. The outcome was cathegorized as complete recovery, partial recovery or failure. In 21 patients regular follow-up was > 5 years (5-8.5 years) with hormonal status assessed regularly annually/biannually, US at least biannually, and scintigraphy performed at least once. Results: Technically successful PEI was performed in 61/65 patients (94%). Volume reduction > 50% was observed in 57/61 patients (93%). Complete recovery was noted in 37/61 patients (60.7%), partial recovery in 18/61 patients (29.5%) and failure in 6/61 (9.8%). Failure was noted in three patients with toxic nodular goiters and three patients with toxic adenomas. Follow-up > 5 years in 21 patients showed three cases of hyperthyroidism recurrence, after three, four and six years, with scintigraphically hot nodules in treated (2) and non-treated (1) thyroid gland parts. All were successfully retreated with PEI. 18/21 patients were euthyroid (85.7%), with no new scintigraphic hot nodules. Conclusion: US-guided PEI treatment of thyroid autonomous nodules is an efficient method with very good initial and long-term results. Evaluation of a novel navigation system with a patient fixation device for interventional CT-guided procedures M. Purpose: Evaluation of the clinical practicability and accuracy of a novel navigation system for interventional CT-guided procedures with a patient fixation device and virtual navigation in a phantom and patient study. Methods and Material: The new navigation system (CAPPA IRad) was applied to CT-guided biopsies in 3 patients and to 15 in a phantom. The system consists of a standard PC, navigation software, a passive optical tracking system and a patient frame. For the phantom study we used a prepared spine phantom. First of all we fixed the patient or respectively the phantom with a patient-fix-kit on the CT table, positioned a patient frame with optical and CT markers above and the setup was scanned with CT. All data were transferred to the navigation system inside the scanner room. The target approach was defined and the trajectory was adjusted to the surface of the corresponding CT marker. The needle holder was orientated towards the trajectory. Finally a coaxial biopsy needle was pushed towards the planned target. The accuracy of the procedures was evaluated using navigation software for measurement target-needle tip. The deviation between planned and documented needle tip position was 0.9 mm in the phantom study (n = 15) and 0.8 mm in the patient study (n = 5). Time between the end of planning CT and needle positioning using the system was 29 minutes. There were no complications. Conclusion: CAPPA IRad allows interventions with high accuracy. Advantages are the precision and less traumatic intervention even of the smallest lesions with an angulated 3D approach and reduction of radiation exposure. Purpose: Insertion of a permanent vena cava filter has been shown to reduce occurrence of pulmonary embolism despite an increased risk in DVT over the long-term. Retrievable or optional caval filters offer the ability to be removed or not, according to PE exposure. A bi-centric retrospective study was performed to assess for filter efficacy and for ability to remove the ALN filter in patients presenting with a severe acute venous thromboembolism associated with at least one indication for a filter such as: contra-indication to anticoagulant (mainly due to an haemorrhagic phenomena), acute major surgery or recurrence of venous thromboembolism despite anticoagulant. Results: This study included 285 (225 + 60) patients presenting with DVT (26%), pulmonary embolism (13%) or both (61%). After a median follow-up of 11.4 months, there were two episodes of pulmonary embolism and three thromboses in the filters. A decision to remove the filter was taken in 70 (55+15) patients (24.5%) and all filter retrievals were successful without any complications. The median implantation period was 48 days (range 6 days-14 months). Conclusion: These preliminary results confirm the efficacy of the ALN filter. They also demonstrate the feasibility and safety of retrieval up to 14 months after implantation. Trellis device which allows clot break-up/clearance. To outline its advantages: significant reduction in patient symptomatology, potential reduction of long term morbidity, shortened hospital stay. Methods and Materials: Twelve patients presenting with acute IFV/SFV thrombosis diagnosed on Doppler Ultrasound were referred for treatment with the Bacchus Trellis Device. Symptoms were present for an average of 3.5 days prior to diagnosis. Prospective color Doppler ultrasonographic follow-up is on-going. Results: Subjective impression of reduction of thrombus load varied from 75-100% (6 patients 100%, 3 patients 75%, 3 patients 60%). Radiological resolution of obstruction, reduction in pain within 24hrs and measurable reduction in leg swelling within 72hrs, was seen in all. Based on retrospective hospital data, mean hospital stay reduced from 7 to 2 days. No symptomatic recurrence was noted at 3-month. Conclusion: Standard DVT anti-coagulation regimes merely prevent further clot propagation and diminish risk of pulmonary embolus. Catheter directed thrombolysis is well established for IFV DVT particularly with phlegmasia cerulea dolens (PCD), but is time consuming, usually requires ICU admission and involves the risk of systemic passage of thrombolytic agents. The Bacchus Trellis device has the major advantage of completion of treatment in a single sitting, without the need for indwelling venous sheaths, ICU stay, repeat blood testing, systemic anticoagulation or thrombolytic administration. We show the feasibility of major clot clearance with no significant risks. Although the cost may not be inconsiderable, much quicker discharge from hospital is a major advantage. Purpose: Pancreatic-islets represent an alternative to whole-pancreas transplantation in diabetic kidney-transplanted patients (IAK). Recently, IA transplantation (Edmonton Protocol) has been adopted for non-uremic patients to prevent diabetic complications, with a different immunosuppressive regimen. Our aim was to compare IAK and ITA regarding technical feasibility, complications and clinical outcomes. Ultrasonography and fluoroscopy-guided intraportal islets transplantation was performed in 24 kidney-transplanted (IAK) and 10 nonuremic (ITA) diabetic patients. Immunosuppression was based on cyclosporine, azathioprine, mycophenolate-mofetil and low-dose-steroids for IAK and on a steroid-free regimen (sirolimus, tacrolimus) for ITA. For each procedure we evaluated technical success, number of puncture attempts, complications and, at one year, the following clinical outcomes: well-function (C-peptide > 0.5 ng/ml), insulin-independence, transaminases, creatinine, urinary albumin excretion (UAE). The success of the percutaneous procedure was 100%, with a singlepuncture attempt in 90% of the cases. Procedure associated complications were: 1 peripheral portal thrombosis (ITA), 1 candida infection (IAK), 2 hemoperitoneum (1-IAK, 1-ITA). IAK and ITA showed similar 1-year rate of well-function and insulin-independence (IAK: 58%, ITA: 62%). Transaminases were significantly higher in ITA than IAK at 1 month, normalizing during the follow-up. Creatinine and UAE were stable after islets transplantation (excepting 1 ITA with pre-existing mild nephropathy, in which they increased). Conclusion: Percutaneous intraportal islets transplantation is a safe, low-complication procedure using a combined ultrasonography and fluoroscopy-guided technique. A similar clinical outcome was found in IAK and ITA patients. First experimental evaluation of an expandable percutaneous catheter pump for left ventricular support T. Schmitz-Rode, J. Graf, J.G. Pfeffer, C. Brücker, R.W. Günther; Aachen/DE (smiro@hia.rwth-aachen.de) Purpose: To evaluate the performance of a newly designed percutaneous catheter with expandable pump as a temporary support system for patients with acute left ventricular failure. The pump catheter (introduction diameter 9 F) is positioned in the left ventricle. The rotor is driven by an external motor through a flexible drive shaft. A model circuit was used to assess pump performance, hemolysis tests, and particle image velocimetry. The feasibility of the catheter placement and pump operation were examined in 12 anesthetized sheep. In 7 of the animals cardiogenic shock was induced. Cardiac output (CO) and mean aortic pressure (MAP) were recorded before and during shock, and during catheter pump action. The catheter pump delivered a flow of 4.1 l/min at a differential pressure of 60 mmHg. The average modified index of hemolysis (MIH) was 11.6 (optimum 1.8). Fluoroscopically and echocardiographically guided in-vivo placement and deployment of the device were quick and uncomplicated. Under simulation of acute left ventricular failure (CO 43 ± 22% and MAP 55 ± 16% of the baseline value) the catheter pump significantly improved CO to 67 ± 12% and MAP to 74 ± 18%. Maximum in-vivo duration of operation was 6 hours (average 3.1 ± 1.4 hours). Animal studies revealed: 1. no significant hemolysis (average plasma free hemoglobin 26 ± 4 mg/l after 3 hours), 2. no thrombotic deposits at rotor or pump housing, and 3. no damage of endocardium or aortic valve. Conclusion: A percutaneously insertable, expandable catheter pump is technically and clinically feasible. First experimental results are encouraging. Soft sclerosis for treatment of hepatic cystic diseases Purpose: To describe a new percutaneous treatment of symptomatic cystic diseases of the liver, with lower discomfort for patients and better handling for the operator. Methods and Materials: 16 patients with a total of 19 cysts were treated with a soft method instead of classic alcoholization. Cysts ranged between 11 cm and 19 cm in diameter. After US-guided percutaneous puncture 50% of the fluid was extracted from the cyst, and a solution of minocycline hydrocloride (1 mg/ml of native volume of the cyst) was introduced and left in the cavity (single-shot). 2 patients had polycystic liver disease (5 cysts). Follow-up was performed at 3, 6, 9 and 12 months after treatment. Results: At 6 months treatment was successful in 17/19 cysts (89.4%). In 2/ 16 pts (10.6%) a re-treatment was needed, with a total success rate of 100% at 12 months. Multiple shots technique through a small catheter was used in 1 of the recurred cases (brown, corpuscolated fluid at diagnosis). No complications were noted. No recurrences were noted at 2 years in 10 patients. Conclusion: This method is superior to alcoholization: The procedure is more rapid and less expensive (catheter set is not required). The discomfort for the patient is minimal, the procedure is easier for the operator and results are better, without risks in case of extravasation into the peritoneal cavity or into the biliary tree. All patients had symptomatic varicose veins with 80% (45/55) of patients having CEAP Class 3 and above varicose veins, including 17 patients with non healing venous ulcers. Treatment was performed with the Diomed 810-nm diode laser via a 600-µm fibre introduced percutaneously into the GSV under local anaesthesia using ultrasound and fluoroscopic guidance. No adjuvant injection sclerotherapy or avulsion phlebectomy was performed. Patients were followed clinically and with duplex ultrasound. Mean follow-up period was 3.7 months (range 1 week to 15 months). Results: There was one technical failure as the GSV could not be cannulated due to severe venospasm following guidewire insertion. The remaining 55 GSVs were successfully occluded with no evidence of recanalisation on follow-up duplex ultrasound. All patients reported significant improvement or resolution of symptoms (cramps, swelling) by 1 week. The majority of venous ulcers (14/17) were healed by 3 months. Mild to moderate bruising and tenderness along the course of the treated vein, which resolved within 2 weeks, was noted in all patients. There was one case of skin burn (blisters over the lower thigh) which healed within 1 week. Conclusion: Our early results suggest that EVLT is a safe and effective treatment for symptomatic varicose veins secondary to GSV reflux. Implantable blood Purpose: To prove the feasibility of invasive arterial blood pressure (ABP) monitoring with a miniature pressure sensor designed for use in an implantable telemetry system for continuous ABP monitoring. After arterial puncture a miniature pressure sensor on the tip of a 3 F catheter was inserted into the femoral arteries of 6 sheep. A micro cable inside the catheter allowed the read-out of the measured values with an external reading unit. As a reference, a wire-mounted commercially available pressure sensor was placed into the contralateral femoral artery. To minimize the loss of blood during the insertion procedure, and to reduce bleeding complications after removal of the catheter and the wire, a new peel-off arterial closure device was designed during this study. In each sheep a continuous ABP curve could be obtained with the miniature pressure sensor. The measured ABP values correlated well with those of the reference sensor. The arterial closure device is capable of preventing bleeding during insertion, led to a stable puncture site throughout several hours and prevented complications like haemorrhage or subcutaneous haematoma after extraction of the catheter. Conclusion: Invasive ABP measurement with the miniature pressure sensor is feasible. Further studies are needed for further miniaturization of the system, subcutaneous implantation of the telemetric unit, wireless data transfer and longterm results. Results: Biopsy was successful in 29 patients. One was rejected because the lesion was too close to pectoralis muscle. 9 biopsies revealed malignant lesions (31%): 6 invasive carcinomas and 3 Ductal carcinoma in situ. One revealed fibrous changes with lobular carcinoma in situ (LCIS). 19 biopsies were benign. Of the malignant lesions, 5 presented as an enhancing mass, one as ductal enhancement and 3 as multinodular enhancement with segmental distribution. The LCIS lesion presented as a mass and benign lesions presented as: 6 enhancing masses, 4 ductal enhancements, 3 multinodular enhancements with segmental distribution, 2 linear enhancements and 1 multinodular enhancement. The positive predictive value upon the shape of enhancement is: 45% for masses, 50% for multinodular with segmental distribution, 20% for ductal enhancement. Linear, focal or multinodular regional enhancements were benign in our study. Conclusion: MR guided vacuum assisted biopsy is a useful tool in management of incidental enhancing lesions with negative second look ultrasound. It revealed 31% malignant lesions among 29 suspicious enhancements (categories 4-5). Comparison of ultrasound-guided fine needle aspiration of infracentimetric thyroid nodules according to sonographic criteria R. Butros, F. Boyvat, C. Aytekin, B. Banu, N. Guvener Demirag, B. Demirhan; Ankara/TR (boyvatf@yahoo.com) Purpose: Referrals to endocrinologists for incidentally discovered thyroid nodules have become increasingly common. The proper management of non-palpable thyroid nodules has been debated over recent years. The purpose of our study was to compare the biopsy results of two groups of infracentimetric nodules according to sonographic criteria. Methods and Materials: Ultrasonography of 491 patients showed 584 infracentimetric thyroid nodules. Nodules were classified as positive or negative for sonographic features that suggest malignancy. These included: an irregular margin, microcalcifications, lymphanedopathy and marked hypoechogenicity. Fine needle aspiration biopsy was performed on 317 positive and 272 negative nodules. Results: Of 317 nodules that were classified as positive, 9 were malignant. Of 272 nodules that classified as negative, 4 were malignant. Sensitivity, specifity, and accuracy were calculated. Conclusion: Although we selected two groups according to the sonographic findings, we found a low prevalence of malignancy in both study groups. This raises the question; do we need to biopsy infracentimetric nodules? Studies are needed correlating pathological results with ultrasound appearance in order to identify nodules with a higher risk of malignancy. Purpose: To evaluate MSCT-guided biopsy of focal liver lesions and to determine the rate of diagnostic biopsies using 16G semi-automated core biopsy systems compared to 18G systems. Method and Materials: From July 2001 to April 2005, 163 patients underwent MSCT-guided biopsy of the liver. Indication was based on the detection of a focal liver lesion on a preceding CT or MR imaging. In most cases a 16G automated core biopsy system was used (n = 121), partially in coaxial technique using a 15G guidance needle (n = 68), partially without guidance (n = 53). An 18G core biopsy system was used in 42 patients, under coaxial guidance in 27 cases. Results: Sensitivity, specificity and accuracy for MSCT-guided biopsy were 93.3%, 100% and 94.5% respectively. Sensitivity and accuracy values were higher for 16G biopsy compared to 18G biopsy: 97.2% vs. 78.6% and 97.5% vs. 85.7% respectively. A definitive histological diagnosis was obtained in 90.0% including 94.0% in the 16G group and 75.8% in the 18G group. Using Chi-square-test all these differences between 16G and 18G proved to be highly significant (p < 0.01). Median biopsy duration was 27 minutes. There was only one major complication: expansive subcapsular bleeding. Conclusion: MSCT-guided core biopsy provides fast and accurate diagnosis of focal liver lesions. The statistically significant higher rate of diagnostic results justifies the use of biopsy systems with larger core diameter (i.e. 16G). Ultrasound-guided biopsy of thoracic masses M.A.M.S. Abusedera; Sohag/E.G. (malaa2@yahoo.com) To determine the utility of ultrasound as a guiding tool for biopsy of thoracic masses. Sa Sa Sa Satur tur tur tur turda da da da day y y y y B 203 C D E F A G The criteria for inclusion in the study were: thoracic masses that were diagnosed by preliminary contrast enhanced CT scan and were accessible by ultrasound, and good patient coagulation profile. Results: Twenty-five patients had thoracic masses (two posterior mediastinal, one anterior mediastinal, two pleural, one chest wall, nineteen parenchymal). The lesion diameters ranged from 3-7 cm with average diameter of 5.3 cm and number of passes ranged from 2-4 with average of 2.9 passes per patient. Histopathological diagnosis of biopsy specimen was achieved in 92% (n = 23), in 4% (one case) histology was inconclusive and the sample was insufficient in one case (the biopsy was repeated). Flexible fibro-optic bronchoscopy was done in 19 cases. It showed a central mass in 37% (n = 7) and extrinsic compression in 10% (n = 2), could not be completed in 21% (n = 4) and was negative in 32% (n = 6). Cytological examination following bronchoalveolar lavage was positive in 32% (n = 6) and negative in 47% (n = 9). No samples could be obtained in 21% (n = 4). Pleural effusion was found in 10 cases. Cytological examination of the pleural effusion was positive for malignant cells in 40% (n = 4) and negative in 60 % (n = 6). There were no major complications or fatalities but minor complications occurred in 8% (n = 2): mild hemoptysis in one patient and a small pneumothorax in another (neither required further intervention). Conclusion: Ultrasound can be used as image guidance for biopsy of thoracic lesions safely, with high diagnostic yield. Navigation system for CT-guided percutaneous punctures: First phantom experience R. Bale 1 , M. Lang 1 , C. Lederer 1 , T. Lang 1 , W. Jaschke 1 , M. Nagel 2 ; 1 Innsbruck/AT, 2 Erlangen/DE (reto.bale@uibk.ac.at) Purpose: Evaluation of a novel CT-guided navigation system for percutaneous punctures. Method and Materials: The navigation system (CAS innovations AG, Erlangen, Germany) consists of an aiming device and a reference frame which allows for fully automatic image-to-patient registration. For evaluation of the accuracy a phantom consisting of a base plate and 20 rods with various lengths with glass spheres at the top of every rod was used. Helical CT scans of the phantom with the reference frame with different slice thickness (1.25 mm, 2.5 mm and 5.0 mm) were obtained and the dataset was sent to the navigation system. Results: 300 punctures were performed. CT-scans were obtained and the distance from the needle tip to the previously planned target point was measured. The distance was defined as normal deviation from the achieved position to the planned target. The accuracy of all 300 punctures was 1.2 ± 0.67 mm. The mean accuracy for 1.25 mm scans was 1.0 ± 0.6 mm (n = 100), for 2.5 mm scans 1.1 ± 0.6 mm (n = 100) and for 5.0 mm slice thickness 1.5 ± 0.8 mm (n = 100). The deviation in the needle direction as evaluated in a separate study was 0.31 ± 0.18 mm in 90 punctures. Conclusion: The presented device seems to be an ideal tool for various percutaneous CT-guided interventional procedures including biopsies, local tumor therapy, vertebroplasty etc. Purpose: Experiment to assess accuracy of a novel CT-navigation system for biopsy procedures in a phantom. Methods: The navigation system (CAPPA) consists of an optical tracking system (Polaris), device holder, a reference frame with CT-detectable markers as well as reflective markers visible for the optical tracking and user interface connected to the CT scanner. The frame is placed over the target area prior to CT-scan. The acquired CT volume data set is transferred to the user-interface for navigation planning. To measure the technical accuracy a calibrated 10G needle was placed with its tip exactly on 60 different targets of an individual acrylic phantom. The offsets calculated by the navigation system were noted. To evaluate the accuracy of the needle-positioning of the navigation system 6 radiopaque targets (d=2 mm) located in different positions in the phantom were punctured with a 7.5G needle (l=11 cm); control CT scans (Sensation 16) were made to verify needle tip positions. This procedure was repeated 10 times with a total of 60 punctures. The deviation from the needle tip position to the target in the CT scans was analyzed. The measurement of the technical accuracy of the system showed a mean tip error of 0.66 mm ± 0.20 mm. In the needle positioning experiment the deviation from the needle tip location to the target was mean 0.96 mm ± 0.26 mm. The navigation system provided excellent targeting accuracy in the phantom experiments. It may offer advantages over conventional procedure while eliminating the need for repetitive fluoroscopy. Comparative study of FNAB and core needle biopsy of pulmonary and mediastinal lesions A.I. Ikramov, N.M. Djuraeva, M.F. Maksudov; Tashkent/UZ (ikramovmed@list.ru) Purpose: To compare the effectiveness of percutaneous fine needle aspiration and cutting biopsy of pulmonary and mediastinal lesions. We retrospectively studied 105 CT-guided percutaneous interventional diagnostic lung (78) and mediastinal (27) (37), tuberculoma (3), hemangiopirocytomas (2), hemodectomas (1), inflammatory lesions (17). Negative results of FNAB were 32.5%, verified after surgery. Specificity of core biopsy was 87.5%. In 3 cases occurred complications (pneumothorax). Conclusion: Percutaneous transthoracic CT-guided core biopsy is a fast, safe and effective method for different localization of pulmonary and mediastinum lesions and provides useful diagnostic information in the majority of cases. It represents a valid alternative to more invasive techniques as thoracoscopy and mediastinoscopy. Does non-sterile coupling gel used prior to disinfection process in ultrasound guided percutaneous biopsy contribute to bacterial growth at the biopsy site? K. Gurel 1 , O. Karabay 1 , S. Gurel 1 , C.F. Hildebolt 2 ; 1 Bolu/TR, 2 St. Louis, MO/US (kamilgurel@hotmail.com) Purpose: The aim of this study is to determine the extent to which non-sterile coupling gel, used prior to disinfection process of US-guided percutaneous biopsies, results in bacterial contamination of the biopsy site. Method and Materials: Patients referred for US-guided percutaneous biopsies were included in this study. Transmission material used for sonographic evaluation before the disinfection process was either non-sterile coupling gel or sterile saline. Patients were randomly assigned to two groups: gel (n = 30) or saline (n = 30). Specimens were obtained from the skin within a 10 cm 2 area at the biopsy site before using transmission material for sonographic evaluation and after the disinfection process. Specimens were also obtained from the coupling gel, saline and povidine-iodine. Inoculated specimen plates were incubated at 37º C in aerobic conditions and the numbers of colony forming units were recorded. Nominal logistic regression analysis was used to calculate the odds of post-disinfection bacterial growth based on group, gender, coincidental disease, biopsy site and specific factors (skin fold, hairy skin). Results: Controlling for the other factors, there was no significant difference in post-disinfection growth between the gel and saline groups (p = 0.10). Biopsy site and specific factors did, however, significantly affect bacterial growth (p < 0.05). No bacterial growth resulted from samples obtained from the coupling gel, povidine-iodine, nor saline. Conclusion: Non-sterile coupling gel used prior to percutaneous biopsy does not significantly contribute to bacterial growth. . For the monitoring of the procedure T1-weighted FLASH sequences (TR/TE=100/9; FA= 70°) in all patients and additional different sequences were used. After positioning of the needle tip in the tumors repeated biopsy specimens were acquired with cutting needles (Somatex) during the interventional procedures using coaxial technique. The visibility of the needles and the precisions of the biopsy procedure were evaluated. Peri-and post-interventional complications were observed. Results: The size of the biopsied lesions ranged from1 to 15 cm. Adequate specimens for histological interpretation were obtained in 343 cases (96 %). In 14 patients (4 %) the biopsy results were non-specific. Follow-up showed that the lesions were missed in 25 patients (7 %). The possibility of free slice orientation during the biopsy facilitated the interventional procedures. The different organs, the tumors and vessels were determined safe and all biopsies were performed without major vascular or organ injuries. Our results showed that MR-guided biopsy can be performed safely and precisely and appears to be an alternative to CT-guided biopsy. Purpose: Post-mortem MSCT and MR imaging is being prospectively evaluated in correlation with autopsy. Excellent visualisation of skeletal (MSCT) and soft tissue (MRI) injuries and pathology as well as application of angiographic techniques has been recently demonstrated. Non-invasive full-body cross-sectional imaging is so far insufficient for microscopic tissue analysis. Non-targeted needle autopsies have already been reported. We attempted to implement navigated targeted post-mortem tissue sampling to enhance the potential of virtual autopsy. Clinically approved ACN-III biopsy core needles (14 gauge x 16") with automatic pistol device (Bard Magnum, Medical Device Technologies, Denmark) were navigated and tested in MSCT (Emotion 6, Siemens, Germany) using a navigation system (MEM Center/ISTB Medical Application Framework, Berne, Switzerland) on ex-vivo porcine models and subsequently five human bodies. Results: The navigation system enabled accurate placement of the needles within the brain, heart, lung, liver, spleen and muscle tissue. The ACN-III needles allowed for sampling ("biopsy") of tissue cores with a mean width of 1.7 mm (range 1.2-2 mm) and a maximal length of 20 mm at all locations. The obtained tissue specimens were of sufficient size and adequate quality for histological analysis. Conclusion: Post-mortem navigated "biopsy" can reliably provide tissue samples from different locations of the body. The tissue cores sampled using ACN-III needles seem to be adequate for microscopic tissue analysis. The minimally invasive navigated tissue sampling might significantly improve the information from post-mortem imaging alone in cases that, for whatever reason, are not amenable to autopsy. Purpose: To assess the value of 64-MDCT cystography and virtual cystoscopy in the detection of bladder neoplasms smaller than 10 mm. Methods and Materials: Fifteen patients referred for hematuria, or follow-ups after tumor resection, underwent examination of the pelvis on a 64-MDCT (Sensation 64, Siemens, Germany) with the following protocol: 50 mAs, 120 KV, thickness 1 mm, collimation 0.6, recon increment 0.6, kernel B30 smooth. Using a Foley catheter the bladder was completely emptied of urine and distended with approximately 350-400 cc of room air. Subjects were examined in supine and prone position. In patients with lesions larger than 1 cm, an additional scan was performed in the prone position after contrast injection. Datasets were downloaded to a dedicated workstation (Vitrea 3.7 Vital Images, Plymouth, USA) for image analysis. CT results were compared with those of conventional cystoscopy performed 1 week later. Results: Virtual cystoscopy showed 27 lesions, whilst conventional cystoscopy correctly visualized 28. Twenty-three lesions were in the range of interest (1-10 mm). A 3 mm lesion situated on the bladder dome, in a patient with poor bladder distension, was missed. Overall CT sensitivity for lesion detection was 91% with a specificity of 100%. Conclusion: Virtual cystoscopy is a non-invasive and accurate technique for the detection of bladder lesions smaller than 1 cm. The attenuation gradient between air and bladder wall allows a reduction of tube current, lowering patient exposure. Conventional cystoscopy is still required for biopsy and lesion removal. Results: In comparison with other sequences, FLAIR-sequence was more sensitive in the detection of bladder neoplasms. This sequence demonstrates the hyperintense signal of bladder neoplasms from the filled bladder lumen with no signal. The sensitivity in the identification of bladder neoplasms was 100% with FLAIR sequences, 89.6% with TSE T2-weighted sequences and 86.2% with SE T1-weighted sequences. That was due to the higher signal to lesion ratio of the FLAIR sequences in comparison with the others. In fact on FLAIR-sequences the mean value of contrast to lesion ratio of bladder neoplasm was 33.1 while on SE T1-weighted sequences and TSE T2-weighted sequences was respectively 15.2 and 29.2. FLAIR-sequences allowed the detection of small papillomas (less than 2 mm). TSE T2-weighted sequences were more sensitive than other sequences in the study of bladder wall infiltration. Conclusion: FLAIR-sequences were more sensitive than others in the detection of bladder neoplasms, thanks to their higher contrast to lesion ratio and can be very helpful in the visualization of small papillomas, especially when multifocal. Detection of vesico-ureteral reflux with MR cystography using dynamic bolus tracking technique: Preliminary result H.S. Teh 1 , B.M. Yeh 2 , S.J.J. Gan 1 , F.C. Ng 1 ; 1 Singapore/SG, 2 San Francisco, CA/US (cyber_xray@yahoo.com) Purpose: To evaluate the feasibility of using MR as an imaging technique for detection of vesico-ureteric reflux (VUR). Methods and Materials: Study received institutional review board approval, and written informed consent was obtained from all patients. Six patients (age range, 24 -35 years, 5 female and 1 male) with VUR detected on fluoroscopic voiding cystourethrography (VCUG) were recruited into the study. One patient was excluded when MR imaging revealed an intra-uterine pregnancy. These yielded 10 kidney ureter units (five were normal, and 5 had VUR detected on VCUG). MR cystography was performed using a 1.5 T MR scanner. Patients were catheterized and gadolinium enhanced saline was infused into the bladder. A gadolinium detection pulse sequence technique was used to allow real time visualization of the ureters as a means to detect VUR. When the contrast was seen refluxing into the ureter, a three-dimensional (3D) MR gradient echo sequence was triggered. The result was compared to VCUG. Results: Findings at MR cystography and VCUG were concordant in 8 (Four normal; 4 with VUR) of the 10 kidney ureter units (80%). There was discordance between the two techniques in the remaining 2 kidney ureter units; one had VUR detected on MR cystography only and the other on VCUG alone. There was also good concordance for the extent of hydronephrosis. Purpose: To evaluate the diagnostic accuracy of voiding MR cystourethrography in visualization of the male urethra. Methods and Materials: 10 volunteers and 86 patients underwent voiding MR cystourethrography. The MR examination was performed with a 1.5 Tesla superconductive magnet. The urinary bladder was filled with paramagnetic contrast agent by administering 20 mg furosemide followed by ¾ of the normal dose of a paramagnetic contrast agent. When the bladder was full, the patient was asked to urinate. An urisheath was placed on the penis in order to collect urine. During micturition T1-weighted spoiled 3D-GE acquisitions in the sagittal plane were performed (TR: 12 ms; TE: 2.7 ms; flip-angle: 40°; slice thickness: 2 mm; acquisition time: 12 s). Two consecutive acquisitions were performed and the 3D images were post processed with a MIP algorithm. In 10 patients we also performed urethral virtual endoscopy using VR-technique. Results: 10 patients were unable to perform the MR examination. In other cases we obtained homogeneous opacification of the bladder lumen and perfect evaluation of the male urethra. The visualization of the urethra with MIP reconstructed images was considered comparable to conventional cystourethrography. We detected 24 cases of bladder neck obstruction, 36 cases of urethral strictures, 2 urethral papillomatosis and 4 cases of benign prostatic hypertrophy. The site, length and the number of urethral strictures were accurately determined. The analysis of 3D sagittal scans and urethral virtual endoscopy allowed a better evaluation of the morphology of the urethral strictures compared with conventional cystourethrography. Conclusion: Voiding MR cystourethrography can replace standard retrograde and micturating cystourethrograms. This novel technique avoids both radiation exposure to the gonads and urinary catheterisation. Purpose: To evaluate the role of dynamic MR imaging in pelvic floor disorders, in comparison with conventional defecography. As it is impossible to reproduce a real defecation in a supine position in a MR unit, we propose a dynamic MR technique performed without defecation, by distending the rectum with air. Methods and Materials: Fifty-five patients with evidence of posterior pelvic floor disorders underwent conventional defecography and dynamic MR imaging in a supine position. After rectal air insufflation, the Foley catheter was the bench mark of the ano-rectal junction. T2-weighted HASTE sequences were acquired at rest, contraction and straining in axial and sagittal planes. A dynamic study was performed using a modified T2-weighted HASTE sequence (1image/second) in rest, squeeze and straining phases. Conventional defecography was considered the gold-standard. We adopted the same standard reference lines and angles of conventional defecography, as well as modified values specific for MR imaging, obtained from a control group (16 normal volunteers). MRI images were analyzed by two radiologists blind of the results of conventional defecography. Results: MR vs defecography identified 26/28 rectoceles (86%), 23/28 (82%) ano-rectal junction descents, 5/9 invaginations (56%), 6/6 dyskinetic pubo-rectalis syndrome; 4 enteroceles, 9 uterine descents, 10 cystoceles, undetected or underestimated at conventional defecography. Conclusion: MR imaging of the pelvic floor is a rapid and well tolerated examination. In comparison with defecography, it was valuable in the diagnosis of functional and pathological pelvic floor disorders, with a tendency to underestimate invaginations. On the other hand it provided additional information on the sphincters, levator ani, puborectalis muscles, uterine prolapses, cystoceles and enteroceles, underestimated at conventional defecography. Purpose: To evaluate the magnetic resonance spectroscopy patterns of testes in prepubertal and postpubertal males. Methods and Material: Fourteen patients who were referred to the urology department with various symptoms and signs were included in the study group. The conventional MR imaging sequences of all revealed normal findings. 22 MRS of testes were obtained from the remaining 12 patients. There were 6 adults (age: 19-35 years, mean age: 26 years) and 6 pediatric patients (age: 9-14 years, mean age: 12 year). Lipid and choline peaks were measured and choline/lipid ratios were calculated. Mann Whitney-U test was used for statistical analysis. Results: After all MR imaging and spectroscopy findings were obtained, they were evaluated by two radiologists. Two peaks were prominent and diagnostic: choline (3.21 ppm) and free lipid peak (1-1.5 ppm). There were statistically significant differences between adult and pediatric age groups (p < 0.001). Choline/ lipid ratios were ranging between 0.47-4.20 (Mean: 1.64) in adults and between 0.06-0.94 (Mean: 0.32) in the pediatric age group. Blood testosterone levels were evaluated in both study groups. Conclusion: Choline/lipid ratios were high in adults and low in the pediatric age group. This statistically significant difference represents the increasing amount of spermatogenesis in adults. We recommend using choline/lipid ratio for the evaluation of unilateral testicular spermatogenesis. To our knowledge, there was no relevant study in the literature dealing with the normal MR spectroscopy pattern of testes. This article will provide the preliminary findings and be the initiator of MRS studies of testes with various pathologies. Purpose: There is limited data available in the monitoring of brain arteriovenous malformations (AVM) after stereotactic radiosurgery (STRS) due to the need for repeated catheter angiography. Using a combination of static and dynamic MR angiographic sequences we have studied the rate of AVM obliteration at 12 months and correlated this with baseline AVM characteristics. Methods and Materials: Forty patients (41 AVMs) consented to a dedicated 1.5 T MR protocol 12 months after receiving STRS for a brain AVM. CE-SLINKY (static) and MR-DSA (dynamic) were performed. Nidal volumes were calculated using CE-SLINKY data in patients with a persisting arteriovenous shunt. The rate of change of nidus volume was computed by statistical methods using the planning angiographic data and the MR angiographic data. The AVM factors including 1) maximum linear dimension, 2) nidal volume, 3) AVM location, 4) nidal morphology, 5) venous drainage, 6) "high-flow angiographic change" and 7) prior embolization were used in the statistical analysis to determine their role in nidus obliteration. Results: Complete nidal obliteration was found in 9 patients, 26 showed greater than 50% nidal reduction and 6 had less than 50%. Without Bonferroni correction for multiple comparisons, only AVM venous drainage (p = 0.018) and AVM location (p = 0.032) were significantly correlated with rate of obliteration (Fisher's Exact Test). Purpose: To determine the relationship between functional MR imaging cortical organization and DTI white matter architecture in patients with AVMs; to predict functional network reorganization and to improve preoperative planning. Methods and Materials: BOLD functional MR imaging and DTI data were obtained in 7 patients (14-60y; 4M, 3 F) presenting with AVMs. Functional sensorimotor, language and visual systems were tested. A cross-correlation analysis of the fMRI data was used, with a p < 0.01. Supraselective WADA testing was performed in 5 patients for functional assessments. Fractional anisotropy (FA) and direction sensitive colour-coded FA maps were used to study white matter tracts. Results: Four of the seven patients showed functional MR imaging activation patterns within language, sensori-motor or visual cortex that were consistent with cortical reorganization. Altered architecture of white matter tracts was demonstrated in each of these four patients. Cortical reorganization was supported, though not proven, by WADA testing. In one case clear white matter tract abnormalities were identified without altered cortical organization, while two patients showed no cortical reorganization or altered white matter architecture. Conclusion: There is high association between cortical reorganization detected with functional MR imaging and altered architecture of white matter tracts detected with DTI in AVMs, with impact on preoperative planning. Purpose: To evaluate venous drainage patterns of cavernous sinus dural arteriovenous fistula (CSDAVF) with an intracranial lesion seen on MR imaging. Methods and Materials: DSA and MR imaging findings of 18 patients (4 males and 14 females, age range 38-77 years, mean age 63.8 years) with angiographically proven CSDAVF were retrospectively reviewed to evaluate the relationship between retrograde leptomeningeal venous drainage (RLVD) and abnormal signal intensity of the brain parenchyma on axial T1-weighted images before and after intravenous administration of Gd-DTPA and T2-weighted fast spin-echo images (T2WI). Results: RLVD was identified on DSA in 10 of 18 patients (55.6%), including drainage into the superficial middle cerebral vein (n = 5), uncal vein (n = 3), petrosal vein (n = 3), lateral mesencephalic vein (n = 1) and anterior pontomesencephalic vein (n = 2). In 4 of 18 patients (22.2%), T2WI revealed abnormal hyperintense lesions in the region with RLVD. Of these 4 patients, one presented with putaminal hemorrhage, while the other three presented with hyperintensity of the pons or the middle cerebellar peduncle on T2WI, reflecting venous congestion, and 2 presented with enhancing lesions of the pons on the contrast study. All of these 4 patients showed concentration of venous outflow from cavernous sinus into RLVD and other venous drainage routes were obliterated. The patients without abnormal signal intensity of the brain parenchyma on MR imaging had other patent venous drainage routes in addition to RLVD. Conclusion: Brain parenchymal hyperintensity on T2WI or the contrast enhancement in CSDAVF is a crucial finding that indicates concentration of venous drainage into RLVD, causing venous congestion or parenchymal hemorrhage. bined standard of reference, two neuroradiologists assessed the MDCTA data sets using thin-slice multiplanar reformats, thin-slab MIP and volume rendering techniques. They rated in consensus whether simple endovascular coil placement was considered possible (i.e. without assisting techniques such as remodelling or stenting), based on the morphology of the aneurysm neck and on the presence of vessels originating from the aneurysm. The ratings were compared to the actual therapy results in each aneurysm. Results: Out of 84 aneurysms diagnosed per standard of reference, 72 were considered as potentially ruptured and amenable to acute treatment. 49 aneurysms were successfully coiled without stenting, 5 required additional stenting, 9 were primarily clipped, 5 were clipped after unsuccessful coiling attempts, 4 received no further therapy after unsuccessful coiling attempts. The sensitivity, specificity, PPV and NPV of MDCTA for successful coil embolization is 94%, 92%, 96% and 88%. Conclusion: 16-row MDCTA with advanced postprocessing techniques is highly sensitive in the identification of intracranial aneurysms that can be successfully treated by endovascular coil embolization. Purpose: Although aneurysmal recanalization after coil embolization is a wellknown phenomenon, information on the enlargement of embolized aneurysms is limited. The purpose of this study was to evaluate the fate of embolized aneurysms with a focus on the incidence and risk factors of aneurysmal growth after coil embolization. Methods: A total of 529 patients harboring 591 intracranial aneurysms were treated by endovascular coil embolization from 1992 to 2004. We reviewed 586 angiographic studies of 214 cases who underwent follow-up angiographic studies. Radiological and clinical data were reviewed to identify the risk factors for enlargement of aneurysms after embolization. Results: Thirty-four cases showed enlargement of embolized aneurysms, and all of them showed major recanalization as well. Aneurysms showing minor recanalization revealed no increase in their sizes on the follow-up angiographic studies. The incidence of aneurysmal growth was 81% among cases with major aneurysmal recanalization, 35% among cases with any aneurysmal recanalization, and 8.5% among cases in which follow-up imaging studies were performed. The location of aneurysm at the terminal bifurcation was an important risk factor for aneurysm enlargement at the time of major recanalization. All bifurcation aneurysms except one showed enlargement in the presence of major recanalization. Older age and presence of a secondary pouch were found to be related to no growth of recanalized aneurysms. Conclusion: Aneurysm enlargement after coil embolization is common in the presence of major recanalization. Our data suggest that bifurcation aneurysms have a higher risk of enlargement after coil embolization when they are recanalized. The role of micro-Doppler ultrasound in aneurysm clipping intraoperatively E. Kapsalaki 1 , K. Fountas 2 , I. Fezoulidis 1 , P. Toulas 3 , J.S. Robinson 2 ; 1 Larissa/GR, 2 Macon, GA/US, 3 Athens/GR (ekapsalaki@excite.com) Purpose: We present our experience with the utilization of micro-Doppler ultrasonography during aneurysmal clipping. In a retrospective study, 54 patients surgically treated for 69 intracranial aneurysms were studied. Five patients harboring 7 aneurysms were surgically treated on an elective basis while 49 patients acutely presented with subarachnoid hemorrhage. Nine patients (9 aneurysms) had a posterior circulation aneurysm while the remaining harbored anterior circulation aneurysms. Blood flow velocities of the parent, adjacent vessels and aneurysmal sac were measured by using a Consorma shapable micro-Doppler flow probe (Cook Vascular, Inc., Leechburg, PA, USA). Results: In 11 aneurysms (15.9%) there was decreased or absent flow in the parent vessel or on one of the adjacent vessels after clipping. The clip was repositioned and in 9 of those the blood flow in the previously compromised vessel was restored. In the remaining 2 cases repositioning of the clip could not improve the MDU findings. Interestingly, in 9 aneurysms (13%), MDU demonstrated flow through the aneurysmal dome although the aneurysmal neck appeared to be totally obliterated. After repositioning of the applied clip in 7 cases or application of a booster clip in 2 cases, the obtained MDU showed no blood flow. Conclusion: MDU seems to be a non-invasive, reliable alternative methodology to intra-operative angiography. This inexpensive method might lend itself to the C D E F B A G routine usage of aneurysm clipping for confirming patency of the parent and adjacent vessels, as well as complete obliteration of the aneurysm. Comparison of perfusion in patients with a perimesencephalic hemorrhage and patients with aneurysmal subarachnoid hemorrhage I. van der Schaaf, M. Wermer, B. Velthuis, G. Rinkel; Utrecht/NL (irenemarieke@yahoo.com) Purpose: The origin of the bleeding in perimesencephalic hemorrhage (PMH) is still unknown but clinical and radiological characteristics suggest a venous source. If PMH is of venous origin, less elevation of the intra-cranial pressure and less perfusion deficits will occur than in aneurysmal subarachnoid hemorrhage (aSAH Purpose: Delayed cerebral ischemia (DCI) is an important cause of death and disability after subarachnoid hemorrhage (SAH). We studied the additional prognostic value of brain perfusion to three established predictors (age, clinical condition on admission and amount of subarachnoid blood) for the development of DCI. Methods and Materials: 69 patients scanned with perfusion CT within 72 hours after SAH were included. For each patient we determined cerebral blood flow (CBF) ratios of prespecified opposite regions of interest and the three established predictors. We calculated adjusted hazard ratios (HRs) for the CBF ratio and the three established predictors by means of multivariate analysis (Cox regression). The additional prognostic value of CBF ratios was assessed by comparing the area under the ROC curve (AUC) of two models: One with and one without addition of the CBF ratio to the three established predictors. Purpose: To test the diagnostic value of an increased ipsilateral vascular maps as MR imaging (D-MRI) sign of breast cancer. Subjects and Methods: 204 consecutive D-MRI exams using a 1.5 T equipment entered the analysis. Seventy-two exams were excluded due to unilateral imaging or previous treatment. For the remaining 132, maximum intensity projections (MIPs) of the first subtracted images of a T1-weighted 3D GE sequence (coronal 1-mm 128 par titions; TR/TE=11/4.8 ms; FA=25°; FOV=384 mm; matrix=384x192 mm; 1-mm 3 voxel; 0.1 mmol/kg Gd-DOTA; 120-s time resolution; 1 pre-and 4 post contrast phases) were reviewed. A one-side increased vasculature was reported when two or more vessels (of 30 mm or greater in length and 2 mm or greater in maximal transverse diameter) was found per breast. Gold standard was pathological examination or at least negative 1-year follow-up (for negative exams). Results: Excluding two bilateral cancers (with symmetrical vascular maps), a one-side increased breast vascular maps was found in 46 exams: 39 malignancies of 22.9 ± 12.9 mm in diameter and 7 benignancies ( Purpose: To prospectively and intra-individually compare equivalent 0.1 mmol/ kg doses of Gd-BOPTA and Gd-DTPA for accuracy of detection and characterization of breast lesions on contrast-enhanced MR mammography (MRM). Methods and Materials: Twenty-six consecutive women with a suspected breast tumor on mammography and sonography underwent two identical MRM examinations at 1.5 T separated by 48-72h. A T1w 3D gradient-echo sequence was used and images were acquired pre and several minutes after injection of either Gd-DTPA or Gd-BOPTA at an identical flow rate of 2 ml/s. Separate and combined assessment of non-enhanced, enhanced and subtracted images was performed blindly by two readers in consensus. The accuracy for lesion detection was determined against final diagnosis, based on findings from conventional mammography, sonography and surgery. The sensitivity, specificity, PPV, NPV and overall accuracy for malignant lesion identification was determined against histology. Purpose: The capability to quantify biological activity of atherosclerosis using MR imaging holds great promise for non-invasive risk-stratification based on plaque composition. Macrophages are known to play a central role in atherosclerosis pathogenesis. Gadolinium-containing immunomicelles (micelles linked to a specific antibody targeting macrophages) and micelles have been shown to improve in-vitro and ex-vivo assessment of macrophages using MR imaging. The goal of the current study was to evaluate whether immunomicelles improve invivo imaging of atherosclerotic plaque in Apolipoprotein-E-Knockout (ApoE-KO) mice using MR imaging. Methods and Materials: Gadolinium-containing micelles were synthesized. Immunomicelles were made using a biotin-avidin-bridge to bind micelles to a specific antibody targeted to macrophage-scavenger-receptor (MSR). Immunomicelles, micelles and standard (Gd-DTPA/Gd-DOTA) paramagnetic contrast-agents were tested in ApoE-KO-mice. Mice were imaged at baseline with a 9.4T MR system using a high-spatial-resolution sequence. The mice were then imaged at inter vals following a tail injection of micelles (n = 6), immunomicelles (n = 7), or standard contrast agent (n = 4). Results: Using micelles the ratio of signal-intensity in the aortic wall post and pre-contrast, normalized to muscle, was an average of 1.4 ± 0.3 (40%enhance- Results: Histological analysis revealed low grade histological necrosis/thrombosis (non-responder) in 5 out of 9 animals, while 4 animals showed a good treatment response. Responders as defined by histology showed a significant decrease in ∆R2* compared to the control group (∆R2* control group: 0.0175 ± 0.0021 msec -1 versus responder: 0.0039 ± 0.0018 msec -1 ; p < 0.01). However for non-responders the ∆R2* remained unchanged (∆R2* control group: 0.0175 ± 0.0021 msec -1 versus non-responder: 0.0167 ± 0.0029 msec -1 ; p > 0.5). Moreover the VVF of the non-responders was significantly higher compared to the responders (nonresponder: 0.0167 ± 0.0029 msec -1 versus responder: 0.0039 ± 0.0018 msec -1 ; p < 0.02). Conclusion: Parametric MR imaging allows a non-invasive, early assessment of anti-angiogenic treatment efficacy using a thrombogenic vascular targeting agent. This imaging paradigm requires further evaluation for other anti-angiogenic treatment modalities. Purpose: To depict differences in leukocyte migration to arthritic joints with optical imaging, before and after anti-inflammatory therapy. Methods and Materials: An immune-mediated monoarthritis was induced in the right knee of 12 Sprague Dawley rats. Six rats were investigated before therapy, six rats were investigated after seven days of systemic cortisone therapy. Legs were shaved prior to imaging. Optical images of both knees were acquired before, at 5 minutes, 1 hour, 4 hours, and 24 hours after injection of DiD-labelled leukocytes. Images were evaluated by calculating signal intensity ratios between the arthritic (A) and contralateral normal (N) knee. A/N-ratios were tested for significant differences between baseline values and values after cell injection using a paired t-test and between the untreated and cortisone-treated group using an unpaired t-test. Imaging data was compared with fluorescence microscopy. Results: Optical images before leukocyte injection showed a minor, but significantly increased autofluorescence of arthritic knees compared to normal knees. At 4 h and 24 h after intravenous injection of DiD-labeled leukocytes, the fluorescence increased significantly in the arthritic knees (p < 0.05), but not in the normal knees. Arthritic-to-normal knee ratios were significantly higher before treatment than after cortisone treatment (p < 0.05). Fluorescence microscopy confirmed an accumulation of DiD-labelled leukocytes in arthritic knees, but not normal knees. The arthritic knees contained a higher quantity of fluorescent leukocytes before than after cortisone treatment. Conclusion: An accumulation of DiD-labelled leukocytes in autoimmune-arthritis can be detected and quantified with optical imaging. This leukocyte accumulation is significantly reduced after anti-inflammatory therapy. The highest stem cell fractions were always found in liver, spleen and kidney. There were no significant differences in the progenitor cell accumulation between low and high malignancy tumors, and no differences between vital and irradiated progenitor cells. However, in both tumor models fixation reduced the amount of progenitor cells in tumors by a factor of > 10. While 20 ± 9% were still positive for AC133 in HaCaT-ras-tumors, no positive cells were found in HaCaT-A5-tumors. 24 ± 19%/24 ± 3% (HaCaT-ras/HaCaT-A5) of the progenitor cells were positive for smooth muscle actin. Surprisingly, only 12 ± 10%/8 ± 2% of stem cells were positive for CD31 (HaCaT-ras-tumors/HaCaT-A5-tumors). No positive cells for human lysozyme were found in both tumor models. Conclusion: Accumulation of human stem cells in both tumor models does not seem to underlay active migration, however, adhesion may be receptor mediated. The results further indicate that the progenitor cells predominantly differentiate to smooth muscle cells or myofibroblasts and only to a lesser degree to endothelial cells. Purpose: To non-invasively detect early stages of peritonitis with fluoro-optical methods by in vivo labelling of macrophages using a near infrared fluorescent dye. Optical properties of the near infrared fluorochrome DY676 (Dyomics, Jena, Germany) and uptake by mouse macrophages J774 A.1 have been tested. Peritonitis was induced in 6 NMRI-mice by intraperitoneal injection of 3 mg zymosan A. After intravenous injection of 55nmol fluorochrome per kg body weight, imaging was performed using the whole body small animal near infrared fluorescence (NIRF) imaging system (bonSAI, Siemens, Erlangen, Germany) at short acquisition times of 0.5s. In vivo data were confirmed by ex vivo investigation of isolated peritoneal macrophages with NIRF imaging and confocal laser scanning microscopy. Results: Absorption and emission maxima of DY676 were determined at 676 nm and 701 nm respectively. The molar extinction coefficient was approximately 90,000 M -1 cm -1 . Fluorochrome was found to be taken up into macrophages by endocytosis. Qualitative and quantitative analysis of NIRF images of whole mice with peritonitis revealed a significant increase in fluorescence intensities in the ventral cavity by a factor of 4.6 from 188 ± 41 arbitrary units in control mice to 869 ± 151 arbitrary units in diseased mice that have received fluorochrome intravenously. Confocal laser scanning microscopy of isolated peritoneal macrophages showed evidence of involvement of those cells in imaging. Conclusion: Diagnosis of peritonitis by near infrared optical imaging of in vivo labelled macrophages is feasible, indicating that early stages of inflammation could be detected by the proposed diagnostic method. In-vivo detection of magnetically labelled mesenchymal stem cells in kidneys after selective intra-arterial injection: Results in a glomerulonephritis model in rats at 3 T H. Ittrich, F. Purpose: To evaluate MR imaging for in-vivo monitoring of selective intra-arterially injected iron oxide-labelled mesenchymal stem cells (rMSCs) in a rat model of glomerulonephritis. Methods and Materials: 10 6 bone marrow derived and magnetically labelled rMSC (Resovist) were injected in the left renal artery in rats after induction of an anti-Thy1 glomerulonephritis (2 control groups: Saline or free SPIO injection). Invitro viability of labelled rMSC was checked by trypan blue exclusion and proliferation assays. In-vivo MR Imaging (T2*wGRE) and T2*/R2* measurements were performed before and repetitively up to day 22 after administration using an animal coil in a clinical 3 T scanner (Philips Intera). MR imaging data were matched with histology (HE, Prussian blue). Statistical analyses of signal intensity (SI) and T2*/R2* changes were performed (student's t-test). Results: Magnetic labelling of rMSC with Resovist was highly effective. In-vitro no changes in cell viability were found. Significant SI decrease (96.8-58%) and shortening of T2* (3.5-21.4 ms, day 0-22) were observed in the cortex and the outer medulla of left kidneys in rMSC group up to 3 weeks after cell administration (control 1 : SI decrease 2.9 ± 3.3%, T2* 39.4 ± 3.1 ms, p < 0.05) with glomerular predominance of labelled rMSC in histology. Aberrant rMSC were detected in spleen, liver and lungs. Free SPIO injection (control 2 ) resulted in a significant SI and T2* decrease only in liver and spleen. Conclusion: Effective cell labelling of mesenchymal stem cells is feasible with clinically approved SPIO. This initial study demonstrates the potential of MR imaging for in-vivo cell-tracking of therapeutically administered, magnetically labelled rMSC in the kidneys after selective intra-arterial perfusion. The aim of this study was to label macrophages with a bimodal contrast agent. This approach facilitates detection of inflammation, both in MR imaging and near infrared fluorescence (NIRF) spectroscopy, and combines the sensitivity of NIRF with the anatomical resolution of MR imaging. Methods and Materials: Bacterial magnetosomes with a diameter of 42nm were used as the contrast agent. To make them observable in NIRF spectroscopy, fluorochrome DY676 (Dynomics, Jena) was bound on their surface. T1-and T2relaxation times were determined on a 1.5 T MR-scanner (Siemens, Erlangen) with IR and multi-echo spin-echo sequences, respectively. Different concentrations of pure magnetosomes as well as samples with different cell counts of magnetosome-containing macrophages were investigated. The magnetosomes had only a small effect on T1, while a distinct shortening of T2 was observed. Furthermore, NIRF spectroscopy showed that fluorescence labelling of the macrophages was successful. Labelled macrophages had lower relaxivities than samples containing pure magnetosomes. For pure magnetosomes, values of R1 = 2.3 ± 0.4 (mM*s) -1 and R2 = 407 ± 86.8 (mM*s) -1 were obtained, for labelled macrophages R1 = 0.57 ± 0.07 (mM*s) -1 and R2 = 170.4 ± 20 (mM*s) -1 . Conclusion: Macrophages were successfully labelled by a bimodal contrast agent. The different relaxivities obtained for pure magnetosomes and labelled macrophages may be caused by shielding effects of the surrounding cytoplasm or a conglomeration of the nanoparticles inside the cells. Further studies with mice shall verify the applicability of this contrast agent in vivo. Purpose: To evaluate the efficiencies of three recent ultrasound technologies, compound (CI), tissue harmonic (tHI), speckle reduction imaging (SRI) and their combinations (CI+tHI and CI+tHI+SRI) compared to the conventional B-mode imaging, in the evaluation of solid breast masses. Methods and Material: 126 solid breast masses (mean size 14 mm) with BI-RADS 3-to-6 categories were evaluated. Cine views obtained with single and combined technologies for each lesion was recorded. SRI function was used at the maximum level. Static images were obtained from the same planes of cine views for each lesion. Images were reviewed and graded by three observers for delineation of lesion boundaries, clarity of internal texture, and conspicuity of posterior echo pattern, edge-shadowing and peripheral echogenic halo. Friedman's test was used for multiple statistical comparisons between the techniques. Results: In the evaluation of lesion boundaries and peripheral halo, tHI+CI+SRI was the most valuable method, followed by SRI, tHI+CI, tHI, CI, and B-mode (p < 0.001). In the evaluation of internal texture, CI was the most sensitive method followed by tHI, while SRI, single or combined, was the least valuable one (p < 0.001). In the evaluation of posterior echo pattern and edge shadowing, SRI was the most, and tHI+CI+SRI was the second most useful method (p < 0.001). Conclusion: SRI, CI, and tHI have significant advantages over B-mode imaging in the evaluation of solid breast masses. SRI is useful in delineating lesion boundaries, peripheral halo, posterior echo and edge shadowing. CI and tHI are useful in clarifying internal texture. Comparison of three-dimensional and two-dimensional ultrasound for differentiating benign and malignant solid breast masses: A performance study by radiologists and computer-aided diagnosis systems W. Moon 1 , R.-F. Chang 2 , D.-R. Chen 3 , N. Cho 1 ; 1 Seoul/KR, 2 Taichung/TW, 3 Chiayi/TW (moonwk@radcom.snu.ac.kr) Purpose: To evaluate the performance of radiologists and CAD systems on 3D US and conventional 2D US for the characterization of solid breast masses. Methods and Material: 150 solid breast masses including 68 nonpalpable lesions were prospectively collected using both 2D and 3D US with a 5-10 MHz dedicated volume transducer. Sixty-one cancers and 89 benign lesions were confirmed with biopsy. Six radiologists who specialize in breast imaging independently and blindly reviewed static 2D US images and stored 3D volume data. For a CAD study, the extracted texture features of the 2D and 3D US images were used to classify the tumor as benign or malignant using the artificial neural network. Results: For all radiologists, 3D US was superior to 2D US in sensitivity (97. Purpose: To determine how accurately freehand elasticity images and measurements correlate to the pathological appearances and dimensions of breast cancers. An earlier study suggested that it is more accurate than conventional greyscale ultrasound for demonstrating breast tumour extent. Reported series show reoperation rates of as high as 50% when wide local excision is based on conventional breast imaging techniques. Methods: Freehand elasticity images (using real time elasticity software implemented on a Siemens Elegra ultrasound scanner) were obtained at regular intervals through the tumour the day before surgery. The skin of the breast was marked to allow orientation sutures to be placed on the tumour specimen as it was excised. Histology sections were made in the same plane as the imaging planes at right angles to the greatest diameters of the greyscale tumour appearance. The edges of the specimen were examined to ensure tumour margin extent was accurately measured. The skin marking at imaging, and specimen marking at surgery, allowed ultrasound and specimen image correlation. Methods and Materials: 102 cases of infiltrating breast cancer have been reviewed, involving pre-surgical axillary echographical evaluation and subsequent pathological verification following lymph-adenectomy (gold standard). The echographical study was performed with Mode B echography, high-frequency sound (7.5-10 Mhz), and was considered "positive" or "negative", according to the presence/absence of pathological findings in the nodes. No percutaneous biopsy was performed on the nodes in the "positive cases". The following parameters have been evaluated: Sensitivity, Specificity, Positive Prediction Value, Negative Prediction Value and Exactitude. The echographical study considered 41 cases to be "positive" and 61 cases "negative". In all "positive" cases, disease was confirmed by surgery. However, metastases was observed in 9 of the 61 cases considered "negative". Contrast-enhanced US (CEUS) with a microbubble contrast agent (SonoVue) was performed in 26 patients with suspicious findings in axillary US and/or negative fine needle aspiration biopsy (FNAB) in the context of breast cancer staging to confirm or rule out the presence of axillary metastases. In all positive lymph nodes a FNAB was performed. Findings of FNAB, sentinel node procedures or lymphadenectomies were considered the gold standard. Results: CEUS showed hypoechoic areas consistent with metastatic deposits in 11 patients and homogeneous enhancement consistent with normal reactive lymph nodes in 9 patients. The smallest metastatic deposit seen with CEUS was 4 mm. There were four false positive cases (one due to a granuloma and three attributed to the low resolution of the images) and two false negative cases. Depiction of metastatic deposits with CEUS changed therapeutic approach from surgery to neoadjuvant chemotherapy in four patients (15%). Sensitivity was 73% and specificity 82%. Conclusion: These preliminary results show that perfusion imaging with CEUS can enhance the already high positive predictive value of axillary US guided biopsy in initially negative cases and increase diagnostic confidence in negative exams. Low resolution of the images may limit its interpretation and further studies as well as longer temporal windows are needed to gain experience and increase its accuracy. 14:00 Purpose: To describe US and MR imaging findings in patients with peroneal intraneural ganglia. Methods and Materials: 17 consecutive patients with dysfunction of the common peroneal nerve were evaluated with 12-5 MHz and 17-5 MHz US and 1.5 T MR imaging. All patients had abnormal nerve conduction studies. Results: In all but 1 patient, US and MR imaging identified ganglia of variable size and shape in relation to the superior tibiofibular joint. US and MR imaging divided these cysts into extraneural ganglia (n = 9/16), which developed outside the nerve, and intraneural ganglia (n = 7/16) developing within the nerve. Intraneural ganglia had a stereotypical appearance: US found the bulk of the ganglion on the anterolateral aspect of the superior tibiofibular joint, located remotely from the position of the peroneal nerve. The ganglion was invariably associated with a dilated articular branch. This branch appeared as a long tubular process coursing along the posterolateral aspect of the fibula to join the bulk of the ganglion with the nerve. It was markedly enlarged and assumed a cystic appearance without detectable fascicles. More proximally, the fascicles of the deep and common peroneal nerve were displaced eccentrically by the cyst growing within the epineurium. The overall size of the ganglion and the entity of nerve deficit varied with time. Two intraneural ganglia regressed at 6-months follow-up. Conclusion: In the evaluation of peroneal neuropathy caused by superior tibiofibular joint ganglia, US and MR imaging provide unique information on the intraneural location of the cyst and the status of the involved nerve. Purpose: To determine the outcome of bone marrow edema (BME) and subchondral lesions of the knee after treatment with either the prostacyclin analogue Iloprost or Tramadol in a double-blind, randomized MR imaging study. Methods and Materials: Thirty-one patients with painful BME of the knee caused by early-stage osteonecrosis, osteoarthritis, bone bruise or stress were included, and Institutional Review Board approval was obtained. Coronal STIR images of the affected knees were acquired. After the initial MR examinations, patients were randomized either to Iloprost (n = 16, group 1) or Tramadol (n = 15, group 2). The treatment duration was 4 weeks. Follow-up MR images were obtained after 3 months and again after 1 year. For baseline and follow-up examinations, the mean volume and signal contrast of the BME were quantified using a computer-assisted method. The presence of subchondral lesions immediately adjacent to the subchondral bone was also assessed. Results: In affected bones, mean BME volume regression at the 3-month follow-up indicated a better treatment effect for Iloprost (Iloprost: -47.1%; Tramadol: -38.6%). Mean BME signal contrast regression was also higher in group 1 (Iloprost: -42.8%; Tramadol: -38.5%). Subchondral lesions in the Iloprost group decreased from 11 at baseline to 1 at final follow-up, while in the Tramadol group, subchondral lesions only decreased from 5 at baseline to 3 at final follow-up. Classification and regression tree analysis demonstrated an increase of the retraction index of 1% in 40% of lesions at ≤ 24 months, of 7% in 32% of lesions at 24-40 months, of 52% in 17% of lesions at 40-60 months and of 17% in 11% of lesions at > 60 months (F =51.9, p =.0001). The liver changes from FCF occur substantially and rapidly in some patients and more slowly in others. Follow-up imaging may be beneficial in characterizing these lesions. Computed tomography for the assessment of macrovesicular hepatic steatosis in living liver donors: Comparison of visual grading and liver attenuation index S. Lee, S. Park, P. Kim, K. Kim; Seoul/KR (amaranthlsw@yahoo.co.kr) Purpose: This presentation is to determine and compare the performance of visual grading and liver attenuation index (LAI) of unenhanced computed tomography (CT) in the diagnosis of ≥ 30% macrovesicular steatosis (MaS) in potential donors for living donor liver transplantation (LDLT). Of the LDLT donor candidates who underwent sameday unenhanced CT and ultrasound-guided liver biopsy, two case (i.e. ≥ 30% MaS)-control (i.e. < 30% MaS) combinations between a case group (n = 24) and two control groups (n = 24, respectively), i.e. an overtly-negative control (mean ± SD of MaS=4 ± 6%) and a rather inexplicitly-negative control (mean ± SD of MaS=10 ± 8%), were chosen to represent diagnostically easy and difficult situations, respectively. Two radiologists independently measured liver attenuation index (LAI) on CT, derived and defined as the difference between mean hepatic and mean splenic attenuation, and graded hepatic steatosis using a five-point scale. Interobserver agreement for visual grading and LAI was determined. Receiver operating characteristic (ROC) analysis was done to compare the accuracy of the two methods in the diagnosis of ≥ 30% MaS. Purpose: To assess the feasibility of determining the degree of liver fibrosis with MR elastography. Methods and Materials: Forty-two consecutive patients who had liver biopsy for chronic liver disease were included in the study. The degree of fibrosis on the biopsies was assessed according to the METAVIR scoring system from F0-no fibrosis, to F4-cirrhosis. MR elastography was performed with a 1.5 Tesla scanner. Low-frequency (65 Hz) mechanical waves were transmitted into the liver by a transducer placed at the back of the patients. A motion-sensitized spin echo sequence, phase-locked to the mechanical excitation was used. The phase maps were processed to obtain elasticity and viscosity maps. The elasticity and viscosity values were compared to the results of the biopsies with a Kruskal Wallis test followed by two by two comparisons with Wilcoxon rank-sum tests. The mean elasticity was 2.25 ± 0.22 kPa in the 16 patients without substantial fibrosis (F0 -F1), 3.15 ± 1.39 kPa in the 8 patients with substantial fibrosis (F2 -F3), and 4.38 ± 1.26 kPa in the 18 patients with cirrhosis (F4). The differences between groups were statistically significant (P < 0.05). The mean viscosity was significantly higher in the patients with cirrhosis (4.57 ± 1.75 Pa·s) than in the patients without cirrhosis (2.12 ± 0.74 Pa·s in F0 -F1 and 2.69 ± 1.66 Pa·s in F2 -F3 patients). Conclusion: Non-invasive MR elastography is a feasible method to assess the degree of liver fibrosis. In vivo phosphorus-31 magnetic resonance spectroscopy of the human liver in patients with cirrhosis S.S. Bagnenko, G. Trufanov, V. Fokin; St.Petersburg/RU (bss_1979@mail.ru) Purpose: To investigate the value of in vivo phosphorus-31 magnetic resonance spectroscopy (31P-MRS) in the assessment in patients with cirrhosis using 1.5-T scanner. Methods and Materials: Nineteen patients with compensated cirrhosis were examined with 31P-MRS, which was performed in the right hepatic lobe. Reference data were acquired from 20 healthy volunteers who had no history or evidence of liver disease. Spectra were acquired using a 1.5 T spectroscopy system. Peak area ratios of phosphomonoesters (PME), inorganic phosphate (Pi), and phosphodiesters (PDE) and Adenosine triphosphate (ATP) were calculated. Results: The PME/PDE (0.82 ± 0.5), PME/Pi (5.08 ± 2.7) ratio and the PME signal height ratio were significantly higher in patients with cirrhosis compared with control group (0.25 ± 0.15 and 1.7 ± 0.91 respectively). These changes suggest regenerative activity in cirrhotic liver. Furthermore in patients with small liver the PDE/βATP (0.66 ± 0.14) was significantly lower than the healthy volunteers (3.28 ± 1.19), while this ratio in patients with hepatomegaly was normal (3.28 ± 1.1). These changes reflecting the energy state in the cells. The results of this study indicate that biochemical changes in hepatic energy and phospholipid metabolism exist in patients with cirrhosis and 31P-MRS can help us to detect them. Of course a longitudinal study with larger number of patients is required to better confirm these findings and to understand the biochemical changes underlying this phenomenon. Diffusion-weighted MR in evaluation of liver fibrosis: Role of apparent diffusion coefficients (ADC) measurement in diagnosis and determination of a threshold for pathology R. Girometti, A. Furlan, G. Como, M. Bazzocchi, F. Toso, C. Zuiani; Udine/IT (rgirometti@sirm.org) Purpose: To investigate diagnostic performance of ADC measurement as marker of hepatic fibrosis in cirrhotic patients and to establish a threshold for pathology. Methods and Materials: 12 cirrhotic patients with histologically proven liver fibrosis (8 male and 4 female; mean age 59.6; 11 A and 1 B Child-Pugh score) and 13 healthy volunteers (9 male and 4 female; mean age 43) underwent Diffusionweighted MR of the upper abdomen using a 1.5 T magnet (maximum gradient strength 40 mT/m), equipped with a six-elements phased-array body coil. Diffusion study with parallel imaging (GRAPPA) was performed with a single breathhold, single-shot Spin Echo Echo-planar sequence (TE 74 msec), using increasing b-values: b= 0 sec/mm 2 , b= 150 sec/mm 2 , b= 250 sec/mm 2 , b= 400 sec/mm 2 and applying a unidirectional diffusion gradient along slice direction. ADCs of hepatic parenchyma were calculated on ADC maps. Results: ADC was significantly lower in cirrhotic livers than in controls (mean 1.12 ± 0.17 vs 1.54 ± 0.11 x 10 -3 mm 2 /sec) (p < 0.0001). ROC analysis showed an area under the curve (AUC) of 0.949 (C.I. 95%: [0.84; 1.05]), showing best sensitivity and specificity (respectively 91% and 100%) for an ADC cut-off of 1.34 x 10 -3 mm 2 /sec. Conclusion: Liver Diffusion-weighted MR is a potentially highly accurate tool in evaluating and monitoring liver fibrosis in cirrhotic patients. Using a single-shot SE Echo-planar sequence with maximum b= 400 mm/sec 2 , the best ADC threshold in discriminating liver fibrosis resulted 1.34 x 10 -3 mm 2 /sec. Comparison of two single-shot spin-echo echo-planar diffusion-weighted sequences with increasing b-values in diagnosing liver fibrosis R. Girometti, A. Furlan, G. Como, M. Bazzocchi, F. Toso, C. Zuiani; Udine/IT (rgirometti@sirm.org) To determine the agreement of two Diffusion-weighted sequences with increasing b-values in diagnosing liver fibrosis, in order to assess optimal b-values set. Methods and Materials: Diffusion study on 12 cirrhotic patients (histologically proven fibrosis) and 13 healthy volunteers was performed on a 1.5 T magnet using a breath-hold, single-shot SE echo-planar sequence with two sets of increasing b-values, respectively: 0 sec/mm 2 , 150 sec/mm 2 , 250 sec/mm 2 , 400 sec/ mm 2 (first sequence); 0 sec/mm 2 , 100 sec/mm 2 , 200 sec/mm 2 , 400 sec/mm 2 , 600 sec/mm 2 , 800 sec/mm 2 (second sequence). Hepatic parenchymal ADCs were calculated on ADC maps. Signal-to-noise ratio was measured at b=400 sec/mm 2 and b=800 sec/mm 2 images. Results: A significant difference in ADCs was found comparing cirrhotic patients to controls with the first sequence (ADCs lower in cirrhotic than in controls) (mean 1.12 ± 0.17 vs 1.54 ± 0.11 x 10 -3 mm 2 /sec) (p < 0.0001), but not with the second (mean 0.93 ± 0.15 vs 1.06 ± 0.21 x 10 -3 mm 2 /sec) (p > 0.05). First and second sequences significantly differed in ADC measurements (p < 0.0001) and in diagnostic performance at ROC analysis, with AUC respectively of 0.94 (p < 0.001) (best cut-off: 1.34 x 10 -3 mm 2 /sec) and 0.71 (p > 0.05). According to Bland-Altman analysis, the limits of agreement between sequences were 0.76 and 0. (T2-w TSE, T1-w FLASH 2D, MRCP) as well as dynamic (T1-weighted) and static SPIO-enhanced studies. MR angiography was performed 10 minutes post SPIO-injection using a double contrast technique (i.v. bolus of Gd-DTPA post SPIO). Delayed static double contrast T1weighted imaging was performed to conclude the study. Three independent radiologists evaluated raw data and 3D-images with regard to detection and characterization of liver lesions, liver parenchyma as well as vascular and biliary system. The combined MR-protocol was well tolerated in all patients. Good correlation with intraoperative findings was observed in 21 patients. Best imaging protocol for detection of lesions was the SPIO-enhanced T2-weighted sequence (95%). Improved characterization was documented using dynamic SPIO-enhanced combined with delayed T1-weighted double-contrast imaging. MR angiography depicted anatomical variants in 6 patients (right hepatic artery out of superior mesenteric artery) and portal thrombosis in 3 patients. Dilated bile ducts were delineated in 5 cases. The combined MR-protocol with the use of double contrast technique is an excellent tool for the diagnosis of the liver to identify the exact morphology, vascular system, biliary anatomy as well as detection and characterization of focal liver lesions in one approach. Purpose: To compare the abdominal MR findings between cirrhotic and schistosomotic forms of chronic liver diseases. Abdominal MR was performed in 24 patients with chronic hepatosplenic schistosomiasis and 27 patients with alcoholic or viral cirrhosis (Child-Pugh class A). Images were evaluated for the presence of periportal fibrosis, irregularity of hepatic contours, fissure widening, heterogeneity of hepatic parenchyma, splenic siderotic nodules, ascites and collateral circulation. Measurements of the right, left and caudate lobes of the liver were performed along with measurements of the longitudinal, anteroposterior and transverse diameters of the spleen allowing for calculation of the splenic index. Results: Chi-square tests showed that schistosomotic patients more frequently presented periportal fibrosis and splenic siderotic nodules (P <.001) while ascites was more frequent in cirrhotics (P <.01). Tukey-Duncan tests showed that schistosomotics had smaller right lobes (P <.0006) and larger caudate-right lobe ratios (P <.0079) than cirrhotics and larger splenic diameters and index (P <.01). The presence of periportal fibrosis, splenic siderotic nodules, smaller right lobe and larger caudate-right lobe ratio and splenic diameters can help differentiate between schistosomotic and cirrhotic forms of chronic liver diseases on MR imaging. Purpose: To assess the measurement variability in lung nodule volume that is associated with the image plane. Methods and Materials: 121 non-calcified pulmonary nodules (≤ 30 mm) from 44 consecutive patients were included. CT scans were acquired using a 16-detector row CT scanner with 16 × 0.75-mm collimation. Two sets of overlapping images were reconstructed by two section thicknesses (1-, and 3-mm) at 0.2-mm reconstruction intervals. From these highly overlapping data sets, at each section thickness, we sub-sampled and prepared multiple contiguous, non-overlapping image sets by shifting the z-position of the images. For the non-overlapping data set of 1-mm section thickness, the starting point in z-axis was either 0 or 0.6 mm. Likewise, for the non-overlapping data set of 3-mm section thickness, the starting point in z-axis was either 0 or 1.6 mm. Thus, in each set, a nodule is centered at a different image plane. A commercially available software package (Vitrea 2; Vital Images) was used for the volume measurements. The volume measured from two different image planes along the z-axis was compared for two section thicknesses respectively with a paired t-test. Results: Due to failure of measurements in some pairs, 82 pairs of 1-mm section thickness and 71 pairs of 3-mm section thickness were compared. There were no statistically significant differences between two different image planes along the z-axis for both 1-mm and 3-mm section thickness (P > 0.05). Conclusion: Measurement variability by different image planes along the z-axis was not a significant factor in volumetric measurement of lung nodules. Inadequacy of manual measurements compared to automated CT volumetry in assessment of treatment response of pulmonary metastases using RECIST criteria Overall correlation was significantly better for AV than for MD (p < 0.05). Intraobserver 95%LoA were -1.85-1.75 mm for MD and -11.28-9.84 mm 3 for AV. The interobserver 95%LoA were -1.46-1.92 mm for MD and -11.17-9.33 mm 3 for AV. There was total intra-/interobserver agreement on response using AV (κ=1). MD intra-and interobserver agreements were 0.73-0.8 and 0.77-0.80, respectively. 28/200 MD response ratings (14/50 patients) were discordant. Agreement using MD dropped significantly from total remission to progressive disease (p < 0.05). Conclusion: AV allows for better reproducibility of response evaluation in pulmonary metastases and should be preferred to MD in these patients. Volumetric evaluation of therapy response in patients with lung metastases. Preliminary results with a computer system (CAD) and comparison with unidimensional measurements. F. Fraioli, L. Bertoletti, A. Napoli, F. Calabrese, C. Catalano, R. Passariello; Rome/IT (francesco.fraioli@uniroma1.it) Purpose: To assess the efficacy of a CAD system in the identification of lung metastases and to compare CAD volumetric with observer unidimensional measurements in the evaluation of treatment response in oncology patients. Methods and Materials: Two observers (A, B) evaluated 9 patients undergoing lung CT just before and immediately after having received chemotherapy. The MDCT scans were performed before and after the injection of contrast media, with a high resolution protocol (collimation 4x 1 mm, 100 mAs, 120 Kv). RECIST criteria have been used to consider the disease as stable, increased, or decreased. Subsequently, target lesions (most significant lesions identified before and after chemotherapy) were evaluated with a CAD system to establish volumetric measurements. Observers' unidimensional and CAD volumetric measurements were analyzed for comparison. Results: 24 nodules (DT:5-18 mm in the first study and 4-20 mm in the follow-up study) were included in this study. Observers agreed in the assessment of therapy response in 21 nodules:8 were considered increased in dimensions and 13 were judged stable. Observers and CAD measurements disagreed in three nodules; two considered as stable by the radiologists and increased by CAD; one considered increased by the radiologists and stable by CAD. In judging patients response, radiologists disagreed in two cases. CAD and observers did not agree in one patient. Purpose: To assess the effect of 3D lossy image compression on computer aided detection. Methods and Material: 120 cases were selected from MDCT chest scans acquired with collimation of 1.25 mm, and collected from different sites. The database was composed of 66 low-dose cases and 109 cases with a high-frequency reconstruction kernel. Two expert radiologists established that 37 cases had no actionable nodules and the remaining 83 cases contained 169 nodules. The mean nodule size was 5.8 mm ± 3.0 [SD] (3.8 -35 mm). All cases were compressed using the 3D SPIHT algorithm to 24:1, 48:1, 96:1 levels. A study of the effect of compression on sensitivity was performed at different operating points (2.5, 5, 10 FP/case) using McNemar's test. Logistic regression models were used to evaluate the impact of compression level, nodule and scanner characteristics on CAD sensitivity. Results: Compared to no compression, there was no significant degradation in sensitivity found at any of the proposed compression levels and operating points. However, between compression levels there was marginal association with sensitivity. Specifically, 24:1 level was significantly better than 96:1 at all operating points, and occasionally better than no compression at 10 FP/case. Based on multivariable analysis, nodule location was found to be a significant predictor with a lower sensitivity associated with pleural nodules. Nodule size, dose, reconstruction filter, and contrast medium were not significant predictors. Conclusion: CAD detection performance of solid lung nodules was found to be robust across all compression levels up to 96:1. CTs were acquired with a 16-row MDCT using 1 mm slice thickness. Three radiologists independently reviewed in a random order on a workstation 3 types of images: axial 1-mm images, axial 6-mm MIP, CAD system's detections (Lung-Care, Siemens). An independent experienced chest radiologist decided for each detection method which nodules were true positive nodules. Results: 287 nodules of 711 (40.4%) nodules ≥ 1 mm were identified as true positive. The detection rates of the three independent observers with 1-mm axial images were 64/287 (22.3%), 86 (30%), and 136 (47.3%). When MIP images were added to 1-mm images, these rates were 123/287 (42.9%), 140 (48.8%), and 205 (71.4%) (p < 0.0001). When CAD was added to 1-mm images, these rates were 145/287 (50.5%), 154/287 (53.7%), and 190/287 (66.2%) (p < 0.0001). CAD detected 298 opacities; 107 of them were true positive nodules (detection rate: 107/287; 37.3%). CAD showed 6.82 false positive per CT scan. 31 nodules were detected by CAD only; they were smaller than other nodules (2.24 ± 0.9 vs 3.14 ± 1.7, p < 10-4). Conclusion: MIP and CAD reduced the number of overlooked nodules on 1-mm images. The additional nodules detected by these 2 techniques were different. Is the computer-aided detection scheme for lung nodules also effective in detecting lung cancer? Purpose: To evaluate the performance of a computer-aided detection (CAD) system in the detection of both lung nodules and lung cancer. Methods and Materials: 150 chest CTs were evaluated. Database included 23 lung cancers (long diameter < 20 mm), nodules that were stable for at least 2 years, and normal cases. 5 chest radiologists and 5 radiology residents independently recorded the locus of each nodule candidate, and assigned each confidence scores for the likelihood of nodule and malignancy without CAD; then they repeated analysis with use of CAD. Jackknife alternative free-response receiver operating characteristic (JAFROC) analysis was used to compare the observers' performance of detecting nodules and cancer with and without CAD. The JAFROC analysis demonstrated that the performance of detecting lung nodules was significantly increased with CAD for all radiologists and subgroups of chest radiologists and radiology residents (P < 0.01). CAD itself detected 18 of 23 lung cancers. Four lung cancers missed by three radiology residents on initial reading were additionally detected with CAD. However, because the number of false positive detections for lung cancer increased with the use of CAD, the overall performance of detecting lung cancers was not significantly different with and without CAD for all radiologists and subgroups (P > 0.05). The detection rate of lung nodules was significantly improved with CAD. However, the overall performance of detecting lung cancer by radiologists was not significantly affected with CAD, although the use of CAD contributed to detecting additional lung cancers by less experienced radiologists. Purpose: Assessment of the influence of reader's experience on detection and characterization of pulmonary nodules (PN) in ultra-low-dose MSCT (ULD-MSCT). Methods and Materials: 135 patients referred to chest CT for suspected PNs entered the study. The scoutview was replaced with a chest ULD-MSCT at following parameters: 3.5eff. mAs, 120 kV and 1 mm collimation (Toshiba-Aquilion), resulting in a radiation-dose of 0.2-0.3 mSv (same dose-range as CXR in 2 views). A standard dose MSCT followed with 52.2eff.mAs. Two blinded radiologists (R1experienced; R2-resident) read axial images first in ULD-and consecutively in SD-technique: number, size, location, morphology and characterization of each PN was analyzed. If more than 5, PNs were grouped in "many" (6-20), "multiple" (21-50), "diffuse" (> 50) and max./min. Ø was determined. In each technique, subjective diagnostic confidence for detection and characterization was scored between 1-10. After analyzing images in SD-technique, the readers were asked to say whether ULD technique hindered either PN detection or characterization. Results: The findings of SD-technique served as reference standard: sensitivity and specificity of ULD-MSCT were 100%, 95.7% (R1) and 96.3%, 100%(R2) respectively. After evaluating images in SD the readers believed ULD-technique to be a hindrance to PN detection (over-and undercounted PNs < 2 mm) and characterization in 9.7%(R1), 5.2%(R2) and 12.7%(R1) and 1.5%(R2) respectively. Significant differences in reader's subjective confidence in PN detection and characterization between the 2 methods were found for R1 and R2 (p < 0.05 Purpose: The purpose of this study was to use MR imaging to detect perfusion and viability defects in patients with chest pain and report on its utilization in a tertiary hospital. Methods: Patients evaluated for chest pain at the University of Florida/Shands Jacksonville with a cardiac MR imaging extending from 05/03 to 03/05 were included. A Siemens Sonata 1.5 Tesla magnet was used for all studies. Localizer images were acquired followed by rest perfusion using TurboFLASH. Gadolinium contrast (Magnevist 9 ml, 0.03 mmol/kg) was injected with and serial images were obtained to capture the first pass of contrast through the myocardium. Adenosine was then infused (140 mcg/kg/min) for 4 minutes and another dose of contrast administered. After 10 minutes, a delayed enhancement scan was acquired. Exams were interpreted qualitatively and reviewed for perfusion and viability defects. Results: 357 patient exams were acquired and reviewed (age was 60 ± 13 years with 53% female). 46% of these exams were normal. Ischemia was detected in 65 patients (18%) and acute or chronic infarction in 68 patients (19%) additionally of these, 14 patients had both ischemia and infarction. Twenty-two (6%) patients had hypoperfusion in a circumferential subendocardial distribution indicating microvascular disease. This study demonstrates the potential utility of MR in diverting some patients from unnecessary coronary catheterization (with 46% being normal) as well as detecting microvascular disease that would evade detection by angiogram (6%). Additionally, ischemia was detected in 18% that could be referred for further diagnostic studies including CT angiography or catheterization. Purpose: The aim of this prospective study was the optimization of a gradientecho (GRE) MR tagging sequence at 3.0 T. Methods: 17 consecutive healthy subjects (mean age 43.2 yrs) and 3 patients were scanned at a 1.5 T and a 3 T whole body scanner (Siemens). A prospectively triggered GRE sequence was used. Sequence parameters were as follows: TR 5.2 ms, TE 3.7 ms, matrix 192, field of view 350x350 mm, temporal resolution 46 ms. Sequence parameters like flip angle (8-16°), tag size (4-8 mm) and slice thickness (4-8 mm) were varied systematically at both field strengths. For each parameter, the signal-to-noise (SNR), contrast-to-noise (CNR) and relative contrast myocardium-to-tag (RCMT) were evaluated individually. Results: While the conspicuity of the tags is greatly reduced in the diastolic phase at 1.5 T, tags are clearly visible in the myocardium in all phases at 3 T. Based on initial results, a slice thickness of 6 mm and a tag thickness of 8 mm were found to be suitable. At 1.5 T, the optimal flip angle was found to be 12 deg while that at 3 T was found to be 8 deg based on overall image quality, SNR and tag conspicuity (RCMT). Blood SNR was significantly higher at 3 T despite the lower flip angle. Conclusion: Myocardial tagging at 3 T has shown a superior image quality in comparison to 1.5 T due to a higher baseline SNR and an improved CNR as well as RCMT. As a consequence of suppressed fading of the tags, the accessibility to the diastolic phase of the cardiac cycle becomes possible. Purpose: Myocardial MR-perfusion imaging (MPI) is increasingly gaining attention for the assessment of hemodynamically relevant coronary stenoses. Shortcomings of MPI at 1.5 T are limited SNR and image contrast. The aim of our study was to intraindividually compare MPI at 3.0 T and 1.5 T. Methods: MPI at rest is performed in 20 patients with proven chronic myocardial infarction at 3.0 T and 1.5 T in our ongoing study. For MPI a Turbo FLASH 2D sequence (1.5 T/3.0 T: TR 2.2/1.9, TE 1.08/1.04, flip 12°/12°, matrix 94x192/ 115x192) is used after injection of 0.05 mmol gadobenate dimeglumine/kg bw at 4 ml/s (MultiHance, Bracco-Altana, Konstanz, Germany). Contrast dynamic was analyzed visually and quantitatively applying least square fit of the signal-time curve. SNR measurements were performed in precontrast and contrast enhanced myocardium. Image quality was rated by two independent observers (score 0-3). Results: The image score was significantly higher at 3.0 T (2.0 ± 0.3/2.4 ± 0.3 p < 0.05). SNR of contrast enhanced myocardium was significantly higher at 3.0 T (14.9/20.9, p < 0.005). In three of six patients perfusion deficits at rest correlating with myocardial infarction were detected at 3.0 T whereas only two perfusion deficits were seen at 1.5 T. Conclusion: MPI at 3.0 T yields significantly improved image quality with superior SNR and CNR compared to1.5 T. Results: In our first patients image quality at 3.0 T was significantly higher than at 1.5 T (score 1.5 T/3.0 T=2.1 ± 0.2/2.6 ± 0.3 (p < 0.05) with high image quality at both systems. SNR/CNR of infarcted myocardium was 28.2 ± 6.7/24.1 ± 6.5 at 3.0 T and 18.4 ± 4.4/14.7 ± 4.2 at 1.5 T (p < 0.001/p < 0.001). SNR of normal myocardium did not differ significantly (4.1 ± 0.9 at 3.0 T and 3.7 ± 0.7 at 1.5 T (p > 0.05)). Conclusion: DE MR imaging at 3.0 T yields significantly increased SNR, CNR and image quality compared to 1.5 T allowing potentially for acceleration of acquisition time higher spatial resolution or reduction of contrast agent. Purpose: Anastomotic leakage after esophagoplasty using a gastric tube is one of the most serious complications. One of the major causes is the sudden decrease in gastric blood supply leaving the right gastroepiploic artery as the sole supply of the gastric tube. Our purpose was to evaluate the effect of embolization in preoperative redistribution of gastric blood supply in the prevention of anastomotic leakage. The study was prospective from July 2002 to November 2004. Transarterial embolization (TAE) of the left gastric artery (LGA), right gastric artery (RGA), and splenic artery (SA) was performed in 17 consecutive patients. The procedure was carried out 2-3 weeks prior to surgery. Coeliac arteriography was performed before and after embolization with 0.035'' platinum coils and 0.018'' microcoils through a 4 or 5 F catheter and a 2.7 Fr coaxial microcatheter. Embolization was considered successful when the LGA, RGA and SA were occluded. The results were compared with those previously obtained in our institution in 100 patients who underwent this type of surgery without TAE: 14 cases of anastomotic leakage and one death. Results: In 16 out of 17 patients embolization was successfully performed. It was unsuccessful in 1 patient because of anatomical difficulties. There were no early complications. Many patients experienced mild local pain, nausea or vomiting. One developed acute self limited pancreatitis. No anastomotic leak was observed. Conclusion: Preoperative vascular redistribution by TAE is a safe procedure that contributes to decreasing the incidence of postoperative anastomotic leakage after esophageal reconstruction. Superselective preoperative embolization of bone neoplasia facilitates extended tumour resection and stabilization C. Waldherr, R. Kickuth, K. Ludwig, M. Beck, K. Siebenrock, J. Triller; Berne/CH (christian.waldherr@insel.ch) Purpose: The aim of this study was to evaluate the impact of preoperative devascularization of hypervascular bone tumours before resection and orthopaedic stabilization in relation to the degree of postembolization tumour vascularity and the following intraoperative blood loss. We retrospectively analyzed charts and images of 17 consecutive patients (m: 11, f: 6; median age: 65y, range: 31-85y), who underwent preoperative transluminal embolization of long bone, hip and spinal tumours before resection and stabilization. The following bone tumours were identified: Metastases of renal cancer (n = 12), of malignant melanoma (n = 1), of leiomyosarcoma (n = 1), of hepatoma (n = 1), of prostate cancer (n = 1) and a giant cell tumour (n = 1). Embolization was performed with polyvinyl alcohol particles (n = 13), with coils (n = 1), and with both coils and particles (n = 3). Each tumour C D E F B A G was assigned a postembolization vascularization grade (I-III) according to tumour blush after contrast agent injection in the main feeding arteries. Results: Intraoperative blood loss was correlated with the degree of postembolization vascularization. In patients with complete devascularization (grade I; n = 8), it was median 500 ml (range: 200-4000 ml), whereas in patients with partial embolization (grade II; n = 8), it was 1750 ml (range: 350-3500) and 2500 ml (grade III, n = 1). The mean intraoperative blood loss was 1585 ml. Embolization was complicated by tissue necrosis in 1 patient. Conclusion: Preoperative embolization of bone neoplasia is an effective method to reduce intraoperative blood loss, which facilitates and enables extended tumour resection and bone stabilization. Purpose: Transarterial embolization for preoperative induction of contra-lateral liver hypertrophy had been proposed as an alternative to portal vein embolization. The aim was to evaluate clinical outcome and efficiency of this technique. In 11 mini-pigs all but the right-lateral segmental hepatic arteries were embolised with lipiodol/histoacryl-mixture using transfemoral access. In-vivo angiography and ex-vivo specimen angiography were repeated at termination after 4 weeks. Arterial patency and flow-parameters were obtained by intraoperative ultrasound before and after the embolization and after 4 weeks. The radiological intervention was successful in 10/11 cases. In one animal subtotal occlusion of all arterial branches occurred. To avoid right-lateral arterial occlusion, proximal minor sub-segmental branches of the right-middle segment remained patent in 30% of angiograms and 40% of ultrasound examinations performed peri-interventionally. After 4 weeks all 9 surviving animals had developed extensive collaterals with arterial reperfusion of all segments in angiography as well as ultrasound. The animal with subtotal embolization died after two days of liver failure, one other animal after 7 days. 78% of the surviving animals presented with bile duct necrosis with hepatic abscesses at autopsy. All had signs of cholangitis with irregular dilatation of the bile ducts in the specimen cholangiography. Conclusion: Compared to portal vein embolization, the distinctive arterial collateralisation observed limits chances of effective hypertrophy in the contra-lateral segment. Because of the high rate of necrotic and inflammatory damage to the biliary system, hepatic artery embolization for induction of hepatic regeneration is not considered suitable for human application. Methods and Materials: 24 consecutive patients (men 15, women 9, mean age 73 years) between September 1999-January 2005 underwent arterial examination and embolization, for acute upper gastrointestinal tract bleeding due to gastro-duodenal ulcers, following endoscopic treatment failure. In 16 cases, selective embolization of the thin arterial pancreatico-duodenal branch, that was supplying the ulcer was feasible with coils and/or histoacryl (glue). For the 8 others patients, the "sandwich" technique was systematically performed with coils in the gastroduodenal artery, especially in the absence of contrast medium extravasation. The coaxial technique was mainly used (17/24) with a 3 F microcatheter. Results: In this group (24), no major complications related to the procedure were observed. Immediate, durable and permanent haemostasis was achieved in 22 patients, without a second catheterisation during the following days; only 2 patients were operated on with success after rebleeding. On the other hand, regardless of their clinical conditions at the endovascular intervention, several variables did impact on the primary success, including multiorgan system failure (6/24), with a pejorative outcome during the first three months. Conclusion: Angiographic selective embolization with a microcatheter may be an effective way to stop massive life threatening bleeding from gastro-duodenal ulcers, especially in the case of debilated non operable patients. For the mid and long term outcome (3 months-6 years), the patient's survival is directly dependent on their initial clinical condition, at the embolization procedure. At the end of the follow-up period, the size of thyroid gland had decreased by one third to one half of its original volume, but 4 patients continued to require antithyroid drugs. Histologic analysis of thyroid tissue from 6 patients who had embolization followed by surgery showed occlusion of both the superior and/or inferior thyroid arteries plus most of their branches. Consequently, embolization of the majority of the thyroid glands was achieved, and evidence of chemical inflammation, necrosis and fibrosis were observed in embolized thyroid tissue. Conclusion: TAE is an effective, minimally invasive and safe method for the treatment of Graves' disease patients who cannot, or choose not to, accept current therapies. There were 5 males and 17 females, ranging in age from 12 to 62 years (mean 38.9 years). Eight patients presented with bleeding and 4 with loin pain. The remaining 10 patients were asymptomatic at presentation but were treated either for having tumours that were large (more than 4 cm diameter) or increasing in size. The mean largest diameter of the tumours was 10 cm (range 2.5 to 24.4 cm). Super selective embolization of the feeding arteries was performed using a mixture of absolute ethanol and lipiodol with microcatheter techniques, to achieve capillary level ablation of the tumours. The procedures were all technically successful with no significant procedural complications. On a mean follow-up of 35.7 months (range 2 to 65 months), no bleeding episodes were encountered. Thirteen of the patients had tumours that decreased in size by a mean of 41%. Eight patients had tumours that remained stable in size while an increase in tumour size by 17% was documented in one patient. Conclusion: Transarterial embolization using absolute ethanol is an effective and safe alternative to surgery in the management of renal angiomyolipoma. Purpose: HepaSphere™ is a new permanent embolic material that adapts perfectly to the vessel wall and reaches small diameter arteries (> 60 microns). There is little information about the wall damage that these particles induce. The aim of this study is to elucidate the mechanism and timing of the vessel wall alteration and its possible influence if HepaSphere™ is used as a drug carrier. Methods and Materials: Kidneys and left hepatic lobes of nineteen pigs were partially embolized using two different dry sized particles (50 and 150 microns). Animals were sacrificed right after the embolization (4) or four weeks later (15). Histologic studies were performed with hematoxilin-eosin and orcein staining. Results: There was little or no destruction of the arterial wall structures in animals sacrificed immediately after the embolization. Four weeks after the procedure, the internal elastic lamina was disrupted in all cases. Hypertrophy of the muscular layer and damage of the adventitia depended on the amount of the inflammatory reaction and the extension of the infarct (mainly in kidney). All these findings were only observed with orcein staining. Perfect definition of the portal space structures and preservation of biliary tracts were also observed in all cases. Conclusion: HepaSphere™ is a deformable particle that perfectly attaches to the arterial wall. Four weeks after embolization damage of the arterial wall layers is found. This can be studied with special staining techniques such as orcein. The knowledge of arterial disruptions is crucial if the particle is to be used as a drug carrier. 14:00 Purpose: To evaluate the feasibility of 3 T proton MR spectroscopic imaging (1H-MRSI) with solely a body-array coil and determine optimal metabolite ratios discriminating between healthy and prostate cancer tissue. Methods and Materials: Twenty-two patients underwent 1H-MRSI at 3 T (Siemens TRIO) before radical prostatectomy. After acquiring conventional T2-weighted MR images, a PRESS sequence with optimized 180° pulses and 2 dual-band frequency selective excitation pulses to suppress water and lipids was applied. Up to 8 outer volume saturation slabs were used. Voxel size was 7x7x7 mm 3 . In consensus, a radiologist and MR spectroscopist reviewed whole-mount section histopathology on which tumours were outlined. Blinded to metabolite spectra, healthy peripheral zone (PZ), periurethral (U), central gland (CG) tissue and tumor voxels were selected by matching the T2-weighted images with histopathology. Purpose: To evaluate a dynamic contrast enhanced (DCE) sequence with low temporal resolution (and thus good spatial resolution and slice coverage) for the detection and localization of prostate cancer foci. Methods and Materials: 46 patients underwent prostate MR imaging before rad-ical prostatectomy. In addition to axial, sagittal and coronal T2-weighted sequences, an axial DCE sequence (14 slices, 30 seconds) was repeated 3 times after injection of gadolinium (20 cc at 2 cc/s). For image analysis, the prostate was divided into 20 segments. DCE images were assessed only visually, all early enhancing lesions being considered malignant in the peripheral zone. In the central gland, only early enhancing lesions with a homogeneous appearance and or invading the peripheral zone were considered malignant. Three independent readers, blinded to clinical data and patients names, evaluated T2 and DCE images separately. They had to specify cancer involvement within each segment and to record tumor nodule number and location. Results were compared with wholemount step-sections of the surgical specimens. Results: Of the 920 segments examined histologically, 227 were invaded (25%). Sensitivity and specificity for detection of segments involvement were respectively, depending on the readers, 0.13-0.23 and 0.93-0.98 for T2 images and 0.46-0.56 and 0.94-0.99 for DCE images. Forty-nine malignant nodules (> 0.3 cc) were present at histology: 49-60% were detected and correctly located on T2 sequence with 19-50 false positive; 77-81% were detected and correctly located on DCE-MR imaging with 19-45 false positive. Conclusion: DCE-MR imaging allows better detection and localization of prostate cancer foci than T2-weighted MR imaging, even if a low-temporal resolution is used. Detection of prostate cancer using a combined approach of contrast enhanced color Doppler, targeted and systematic biopsy with a prostate specific antigen of 4 to 10 ng/ml J. Gradl, L. Pallwein, D. zur Nedden, F. Frauscher; Innsbruck/AT (johann.gradl@uibk.ac.at) Purpose: Transrectal gray scale ultrasound guided biopsy is the standard method for diagnosing prostate cancer (PC). Improved cancer detection with ultrasound contrast agents is related to a better detection of tumor vascularity. We evaluated the impact of a combined approach of contrast enhanced, color Doppler targeted biopsy (CECD) and systematic biopsy (SB) for the PC detection rate in men with prostate specific antigen (PSA) 4.0 to 10 ng/ml. Methods and Materials: We examined 380 screening volunteers with a total PSA of 4.0 to 10 ng/ml (percent free PSA less than 18). CECD was always performed before SB. Another investigator blinded to the contrast enhanced findings performed 10 SBs. The cancer detection rate for the CECD, SB and combined approaches was assessed. Results: PC was detected in 143 of 380 patients (37.6%, mean total PSA 6.2 ng/ ml). The PC detection rate for CECD and for SB was 27.4% and 27.6%, respectively. The overall cancer detection rate with the 2 methods combined was 37.6%. For targeted biopsy cores the detection rate was significantly better than for SB cores (32.6% vs 17.9%, p < 0.01). CECD in a patient with cancer was 3.1 times more likely to detect PC than SB. Conclusion: CECD allows for the detection of lesions that cannot be found on gray scale ultrasound or SB. CECD allows for the assessment of neovascularity associated with PC. However, the combined use of CECD and SB allows for maximal detection of PC with a detection rate of 37.6% in our patients with a PSA of 4 to 10 ng/ml. Is there a relation between pain and prostate volume during transrectal ultrasound guided prostate biopsy? A retrospective linear regression model T. Yun, H. Lee, S. Kim; Seoul/KR (yuntj@radiol.snu.ac.kr) Purpose: To determine the relationship between pain and prostate volume during transrectal ultrasound guided prostate biopsy. Methods and Materials: Between March 2004 and September 2004, the total prostate volume, peripheral zone volume, and transition zone volume of 71 patients, who had prostate biopsies, were retrospectively analysed. Pain was assessed with a visual analogue scale, during periprostatic nerve block (VAS 1), during the biopsy procedure (VAS 2) and 20 minutes afterwards (VAS 3). Associations between prostate volume and pain during transrectal prostate biopsy were retrospectively evaluated using linear regression models. Results: Total prostate volume was positively associated with VAS 2 (p = 0.0220) and VAS 3 (p = 0.0012). Peripheral zone volume was also positively associated with VAS 2 (p = 0.0231) and VAS 3 (p = 0.0001). However no significant association could be demonstrated between total prostate volume or peripheral zone volume and VAS 1. There was not any significant relationship between transition zone volume and VAS values. Purpose: To evaluate the feasibility of finding lymph node metastases outside of the normal (= limited) surgical dissection area, using Ferumoxtran-10 enhanced MR imaging, in patients with prostate cancer. This is in order to select which patients are best suited for an 'extended' lymphadenectomy or CT-guided biopsy. Methods and Material: 300 consecutive patients with biopsy proven prostate cancer (PSA > 10, or Gleason > 6 or DRE stage T3) were enrolled in this study. T1-weighted TSE and T2*-weighted GRE MR images of the pelvis were obtained, 24 hours after Ferumoxtran-10 administration. Positive nodes on MR examinations were called inside, or outside the normal surgical area. Based on MR imaging results, a lymph node dissection, a MRL guided minimal extended lymph node dissection, or CT-guided lymph node biopsy was performed. Results: Forty-nine (16%) patients had malignant nodes on histological examination. In 21 patients (43 %), malignant nodes were found exclusively outside the limited surgical area. In 3 (6%) patients malignant nodes were found both inside the limited dissection field and outside this area. In the remaining 25 (51%) patients malignant nodes were found in the area inside the limited dissection field. The feasibility of quantitative T2* measurement has been demonstrated in the fetal brain. Although there are some differences between SO 2 values determined by MR imaging and blood samples, there is reasonable correlation between these two methods. Therefore MR imaging has potential to non-invasively assess oxygen saturation and consequent fetal risk. The role of fetal MR imaging in the detection of gyration and migration anomalies before 24 weeks gestation D. Horák 1 , J. Hyjánek 1 , P. Vlašín 2 , M. Herman 1 ; 1 Olomouc/CZ, 2 Brno/CZ (davidhorak@post.cz) Purpose: Most prenatal MR imaging in the Czech Republic is done before 24 weeks gestation (WG) for medicolegal reasons (possibility to terminate pregnancy up to 24 WG for medical causes). The fetal brain during this period is undergoing early developmental changes, especially in terms of gyration and sulcation. The authors want to assess the benefit of fetal MR imaging for detection of migration and gyration anomalies of fetuses up to 24 WG. Methods and Material: 48 CNS examinations of fetuses up to 24 WG were performed on a 1.5 T scanner. Heavily T2W images of the brain were acquired in three orthogonal planes and at least in two planes for the spine. T1W images were routinely used for the brain and spine in one plane. We acquired all images without fetal or maternal sedation. MR imaging results were compared to autopsy or postnatal imaging, clinical follow-up and the US study. Results: We found 8 (16.7%) cases of migration and gyration anomalies: 3 complete callosal agenesis, 1 callosal dysgenesis, 1 schizencephaly, 2 complex migration anomaly and 1 parenchymal atrophy. US found mostly indirect changes such ventriculomegaly, dilation of subarachnoid space or thin cortex. We found 25 (52.1%) other anomalies (ventriculomegaly, posterior fossa anomalies, spinal dysrhaphisms etc), 15 (31.3%) examinations were normal. Autopsy/follow-up confirmed the MR diagnosis in 87.5%. Conclusion: Despite the presence of only early developmental changes in the examined brain, fetal MR imaging before 24 WG is capable of detecting CNS migration and gyration anomalies and provides valuable diagnostic information compared to US. Ultrasound and MR imaging of paediatric neuropathies (brachial plexus excluded) M. Valle 1 , P. Toma' 1 , G. Bodner 2 , S. Bianchi 3 , E. Silvestri 1 , C. Martinoli 1 ; 1 Genova/IT, 2 Innsbruck/AT, 3 Geneva/CH (mvalle@zeus.newnetworks.it) Purpose: To describe and explain the anatomical basis, the clinical characteristics as well as the US appearances for a variety of neuropathies (brachial plexus excluded) which occur in children. Methods and Materials: Thirty consecutive patients with either dysfunction or abnormal nerve conduction studies or a palpable mass over the course of a nerve of the upper or lower limb were retrospectively evaluated. US was performed with 12-5 MHz and 17-5 MHz US transducers. All patients had correlative 1.5 T MR imaging. Results: US allowed detection of a spectrum of nerve abnormalities. Traumatic nerve lesions due to penetrating injuries included: partial tear of the ulnar nerve at wrist (n = 2) and elbow (n = 1); combined tear of ulnar, median and radial nerves at the forearm (n = 1); complete tear of the median nerve at the arm (n = 1). Closed stretching injuries selectively involved the peroneal nerve in the popliteal fossa with formation of a fusiform neuroma (n = 4). Other lesions included: entrapment of the peroneal nerve at the fibular neck by bone exostoses (n = 3) or superior Sa Sa Sa Sa Satur tur tur tur turda da da da day y y y y Scientific Sessions B 229 C D E F A G tibiofibular joint ganglia (n = 2); carpal tunnel syndrome in hemophilia (n = 1); radial nerve entrapment at the spiral groove by fibrous band (n = 1) and postoperative scar encasement of the sural nerve (n = 1). Nerve tumors included: fibrolipomatous hamartomas of the median (n = 3) and plantar (n = 1) nerves, plexiform schwannoma (n = 1), diffuse neurofibroma (n = 2) and type-1 neurofibromatosis (n = 6). Conclusion: US is a promising means for evaluating a wide range of nerve abnormalities in children. Purpose: Ferumoxtran is administrated for in-vivo MRI monitoring of macrophagic cells on experimental adult animals. Several new born experimental models contain macrophagic activated cells (periventricular leukomalacia for example). Intravenous injection is impossible in these animals (< 10 grams). Thus, we test a new type of administration for Ferumoxtran on healthy new born rats: the intraperitoneal administration (IP). Ferumoxtran IP (2400 micromole/kg) was evaluated after a delay of 0 to 105 hours in 61 rats (age 1-7 days) with: 1/ in-vivo 1.5-T abdominal MRI (TR=2000, TE=50, liver, spleen, and bone marrow signal) 2/ plasma and peritoneal iron concentrations 3/ liver, spleen and bone marrow histology (Perls coloration for iron intracellular deposit). 5 rats were control rats (NaCl injection). Results: Ferumoxtran IP was resorbed from the peritoneum before 11 hours, and detected in plasma between 0 to 75 hours (concentration peak at 8 h, 660 mg/ l). A significant (P < 0.05) decrease of liver, spleen and bone marrow signal was observed from 2 hours (liver, spleen) and 4 hours (bone marrow) after IP and beyond that. Intracellular iron deposits were significantly observed (P < 0.05) in liver and spleen from 3 hours and beyond that and at 96 hours in bone marrow. Purpose: To determine the feasibility of multimodality imaging in paediatric patients suffering from neuroblastoma, hepatoblastoma and Ewing sarcoma. Methods and Materials: In the last three years 16 paediatric patients underwent multimodality imaging in our institution. Among them 4 patients with neuroblastoma (aged 9 months to 2 years) were imaged 8 times, 2 patients with hepatoblastoma (1 and 4 years) 3 times, and 2 patients with Ewing sarcoma (11 and 14 years) 3 times for staging and restaging. Overall 7 superimpositions of CT and SPECT, 4 of CT and PET, 2 of MR and SPECT and 2 of MR and PET were performed on a TREON StealthStation® Navigation System (Medtronic, USA). Up to 4 years of age, all datasets were obtained one by one in one session during general anaesthesia. Movement was additionally inhibited using a Bluebag® vacuum mattress (Medical Intelligence, Germany). Results: Superimposition of datasets was successful in all examinations (n = 15). Follow-up and restaging after therapy was possible in measuring tumour extent and functional activity of tumour tissue. Conclusion: Multimodality imaging is possible even in young patients using general anaesthesia and immobilization devices. Tumour extent and function can be determined in one dataset leading to appropriate adjustment following surgical, chemo or radiation therapy. 14) and MRI (n = 11) examinations were performed in six pediatric patients with histopathologically proven PTLD in the initial staging and during the course of therapy with an anti-CD20 chimeric antibody (Rituximab). One (n = 3), two (n = 2) or three (n = 1) follow-up studies were performed in all patients. Images were retrospectively correlated on a lesion by lesion basis. Results: In five patients PTLD occurred 12 (± 7) months after transplantation of the heart and lungs (n = 1) or the kidney (n = 4). In one patient PTLD was diagnosed 84 months after transplantation of the heart (n = 1). PTLD showed organ involvement of the lung (n = 2), the liver (n = 1), the kidneys (n = 2), the digestive tract (n = 2), the lymphatic nodes (n = 6) and the skeletal system (n = 1). All sites of organ involvement as demonstrated by CT or MRI showed markedly increased tracer uptake in FDG-PET in the initial staging. During therapy, FDG-PET was able to demonstrate response early whereas in CT and MRI the lesions showed only delayed changes. Conclusion: [ 18 F]-FDG-PET (/CT) aids with the accurate initial staging of pediatric patients suffering from PTLD. It allows response to therapy to be demonstrated earlier than CT or MR imaging. Purpose: Although Hodgkin's disease (HD) in children has an event free survival (EFS) of about 95%, the rate of secondary malignancies is about 25% after 30 years. Radiation therapy seems to be the most important risk factor. Therefore the GPOH-HD 2002 study group attempts to reduce the indication for radiation therapy. Methods: Until now only the patients with complete remission on CT/MRI after chemotherapy get no radiation therapy. This strategy had no influence on the EFS. The next study generation is intended to further reduce the indication for radiation therapy using FDG-PET for response evaluation. In our study in 41 patients with early stage HD CT/MRI was paralleled by FDG-PET in initial staging and in the follow-up. Results: 7 of the 41 patients were in CR based on PET and CT/MRI criteria, one patient with CT/MRI-based CR had still a positive PET result. In 17 patients PET was completely negative after chemotherapy whereas residual masses were found by CT/MRI. In 2 cases PET was not able to be evaluated in its entirety. In summary, using a PET-based strategy, 27/41 patients (66%) with early stage HD would not have to be irradiated in the future in contrast to 9/41 patients (21%) according to the current strategy. Conclusion: By using FDG-PET-based stratification in addition to the established CT/MRI, the number of non-irradiated patients will increase dramatically. We estimate that about 60 -70% of all patients with early stage HD will no longer receive radiation therapy without reduction of EFS rates. Evaluation of airway obstructive lesions in complex congenital heart disease: Using composite volume rendered images from MSCT K.S. Choo, H. Cho, H. Lee, S. Sung, C. Kim; Pusan/KR (kschoo0618@medimail.co.kr) Purpose: To investigate clinical usefulness of composite volume rendered images in the evaluation of the relationship between cardiovascular structure and airways in children with complex congenital heart disease. Methods and Materials: 4-slice or 16-slice MSCT scanning was performed consecutively in 77 children (age range 5 days to 5 years, mean age: 6.4 months) with congenital heart disease and respiratory symptoms, chest radiograph abnormality or abnormal course of the pulmonary artery on echocardiography. MSCT C D E F B A G scanning was performed during breathing and after sedation with chloral hydrate in all patients. Contrast medium was administered through a pedal or arm venous route in all patients. The CT volume data was transferred to commercially available software (voxelplus2, Seoul, Korea). The VR technique was use to reconstruct the cardiovascular structures and airways, and then both volume-rendered images were composed using the software. Results: Stenoses were seen in the trachea in 1 patient and in bronchuses in 14 patients (19%). Their causes were abnormal course of the pulmonary artery in 2 patients, circumflex retro-esophageal aortic arch in 2, pulmonary artery dilation in 1, anterior displacement of the descending aorta or posterior displacement of the ascending aorta following operation in 5, posterior displacement of the pulmonary artery in 4, and sub-pulmonary sling in 1 patient. Other patients with complex congenital heart disease did not have significant airway stenoses. Conclusion: Composition of both volume-rendered images with MSCT can provide more exact airway images in relationship to surrounding cardiovascular structures and can thus offer optimal management strategies in treating congenital heart disease. Purpose: To compare the performance of a rawdata-based synchronization signal to the commonly used ECG for a 64-slice CT scanner. The kymogram represents the temporal variation of mass and provides a motion function which is correlated to the heart rate. It is generated by a center-of-mass tracking of the scanned object [Med. Phys., 29 (7):1489-1503, 2002]. We evaluated this purely rawdata-based synchronization signal as an alternative to the ECG-signal for phase-correlated image reconstruction using a 64-slice CT scanner (Sensation 64, Siemens, Forchheim). For evaluation data of 90 consecutive patients were used scanned with 2•32*0.6 mm collimation, 0.33 s rotation time and concurrent ECG recording. As a quality measure the correlation of the ECG with the kymogram was used for validation. Additionally, visual assessment of the kymogram-based and the ECG-based reconstructed images was performed by a double-blind comparative rating. Results: In 70 out of 90 patients a high correlation between the kymogram and the ECG was observed with a mean difference of the detected heart rate below 4 bpm. In 6 out of 90 patients a strong improvement in image quality was achieved by using a kymogram-based reconstruction. This was due to bad ECG recordings. Conclusion: Synchronization signals on the basis of a kymogram correlated well with the ECG for most patients and improved image quality in some cases. We conclude that a kymogram-based phase-correlated image reconstruction in cardiac CT is practical and may provide advantages. Image quality improvements in cardiac imaging with step-and-shoot volumetric CT J. Hsieh, J. Londt, M. Vass, J. Li, X. Tang, D. Okerlund; Waukesha, WI/US (jiang.hsieh@med.ge.com) Purpose: Although nearly all cardiac CT protocols utilize low-pitch helical data acquisition, previous studies have shown advantages of using step-and-shoot mode acquisition to significantly improve dose efficiency. We propose new reconstruction approaches to effectively improve the temporal resolution of the scanner by reducing cardiac motion artifacts and to suppress cone beam artifacts. In helical acquisition, cardiac phases and scan locations are closely coupled, and the projection view range is selected primarily by the EKG signal. Since these parameters are completely independent in the stepand-shoot case, we utilize a set of projection consistency conditions to further adjust the view selection to minimize the banding artifacts caused by the residual phase misalignment. Consistency conditions include both H-L condition and correlation pattern. In addition, a complimentary cone beam reconstruction algorithm is used to compensate for the reduced FOV due to cone beam effect in z. Two adjacent scans are used in combination to produce images. Results: Both phantom and animal experiments were conducted. Phantom studies with high contrast objects have shown that the complimentary reconstruction algorithm is effective in combating cone beam artifacts due to extrapolation. Animal and balloon experiments have shown that optimization of view selection by the proposed algorithm can significantly reduce banding artifacts and improve the effective temporal resolution. Conclusion: We proposed new approaches for the step-and-shoot cardiac imaging. Both animal and phantom studies have shown significant improvements in image quality in terms of motion and cone beam artifacts reduction. Clinical experiments are currently under way. CEnPiT: Helical cardiac CT reconstruction C. Bontus, P. Koken, M. Grass, T. Koehler, R. Proksa; Hamburg/DE (claas.bontus@philips.com) Purpose: Evaluation of a retrospectively gated helical cardiac CT reconstruction method, which is based on an exact algorithm. Reconstruction results from a CT system with 64 rows as well as simulated data for systems with 128 rows are used. For cone-beam CT scanners with a large number of detector rows, exact reconstruction algorithms are necessary. Otherwise conebeam artifacts in the images reach an unacceptable level. When applying ECG based gating in a naïve way to the acquired helical projection data, the acquisition becomes incomplete. Considering cardiac CT, the question, how an exact algorithm can be combined with a gating approach is addressed. The used approach splits the data into contributions of different kinds of Radon-planes. In particular, we combined different gating-functions with different kinds of Radonplane contributions. Radon-planes intersecting only once with the helix are measured without redundancy. Therefore, the corresponding contributions are back-projected without gating. For Radon-planes with the maximum number of intersections, the full gating-scheme is applied. The reconstruction results are very promising. The images show that the algorithm is capable of delivering diagnostic image quality for large cone angle CT systems, though gating is applied. Neither the inclusion of data from the complete scan, nor the variability of the heart rate introduce additional artifacts. The reconstructed images were analyzed regarding image quality and artifacts. Conclusion: An exact filtered back-projection algorithm can be combined with a cardiac gating approach. Thus the method is a good candidate for future systems with a large area detector. Purpose: When metal implants present, low signal-to-noise ratio is caused in projection and the noises are with high energy. Conventional methods reduced the noises by smoothing or interpolating the affected region in the projection, which inevitably cause errors and blur reconstruction. This study provides a new technique to amend the wavelet coefficients of the projection by developing a shrink algorithm based on statistical evaluation to isolate the noises from the projection. Methods and Materials: Two dimensional phase shift wavelet decomposition is adapted to decompose the projection into several components. The energy of each component in the affected region is compared with that in the rest part of the projection, and a shrink algorithm is derived to reduce the correspondence of the noises in the projection according to statistical evaluation. A repaired broken bone of a leg of pork is scanned with a stainless steel plate and screws to demonstrate the method. Results: A pig's leg with a lead nail inside the bone was scanned with the parameters of 140.0 kV and 40 mA. FBP with interpolation causes low frequent disturbance, normal wavelet method cannot reduce streaking completely. The method in this study reduces the streaking effectively without significant reconstruction blurring. The decomposition of projection using phase shift wavelet describes the noises with high energy effectively, and a shrink algorithm is derived according to the distribution of the noises in wavelet domain to isolate the correspond- Purpose: Dose and image quality assessment in CT are affected by the variety of CT scanners and imaging protocols in use. Our aim is to assess image quality related to patient dose indexes. Dose and image quality were all measured with a standard CT phantom for the main protocols (head, chest, abdomen and pelvis). Results: CT dose indexes (CTDIw, DLP) for each protocol were compared with values displayed by the scanner. The differences were always below 7%. All indexes were below the DRL defined by the European Council directive 97/42. Effective doses were measured for each protocol with thermoluminescent dosimeters inserted in an anthropomorphic phantom and compared with the same values computed by the ImPaCT CT patient dosimetry calculator software and corrected by a factor taking in account the number of slices (16 to 64). The differences were always below 20%. The effective doses range from 1.5 mSv (head) to 12.4 mSv (abdomen). The dose reduction algorithm of the scanner was assessed comparing the effective dose measured in a phantom to the dose computed on 46 patients. The reduction factors range from 5% to 32%, depending on the protocol and patient size. The possibility of a further dose reduction was investigated measuring spatial resolution, contrast and noise as a function of CT-DIw. This curve shows a quite flat trend decreasing the dose approximately to 80% and a sharp fall below that value. Conclusion: All the information allowed optimizing the protocols obtaining an overall mean dose reduction up to 35%. R.W.R. Loose 1 , M. Wucherer 1 , R. Adamus 1 , T. Brunner 2 ; 1 Nuremberg/DE, 2 Forchheim/DE (loose@klinikum-nuernberg.de) Purpose: To evaluate the low-contrast resolution of a dynamic CsI-aSi flat panel detector for 3D-reconstructions from rotational angiography (CT-mode) for phantoms and clinical examinations. Methods and Materials: 3D datasets of low contrast objects and patients were acquired with a dynamic digital 40x30 cm FD angiography system (Axiom Artis dTA, Siemens) with rotation speeds of 60°/sec and 30 frames/sec image frequency. The detector matrix of the FD is 2480x1920 with 154 µm pixel size. The projection data were reconstructed with a CT cone beam algorithm by summing of 2 contiguous detector lines and columns (binning). A spatial resolution of 0.3 mm at the detector level and 0.2 mm at the object level was obtained with up to 1000 reconstructed slices from one 220° rotation. Results: With detector entrance doses of 0.36 µGy and 1.2 µGy we obtained doses at the 16 cm CTDI phantom of approx. 13 mGy and 43 mGy which are below the diagnostic reference values for CT head examinations. At the 16 cm CATPHAN phantom objects with 10 mm diameter at 5 mm slice thickness with ≤ 5 HU are visible. Cerebral structures with liquor spaces and dilated bile ducts were visualized for the first time in head and abdominal examinations without contrast and with high image quality. Conclusion: The significant higher image quality of a dynamic FD in comparison to an image intensifier enables for the first time 3D acquisitions with excellent low contrast resolution in clinically acceptable acquisitions times without exceeding the CT dose reference values. Quantitative analysis of coronary arteries with computer tomography coronary angiography: Methodology and preliminary results using intracoronary ultrasound as reference standard using a software prototype (QCT) F. Cademartiri, G. Runza, L. La Grutta, A. Palumbo, N.R. Mollet, N. Bruining, R. Hamers; Rotterdam/NL (filippocademartiri@hotmail.com) Purpose: To compare a newly developed software (QCT) for the quantitative analysis of coronary lumen and plaque with computer tomography coronary angiography (CT-CA) with intracoronary ultrasound (ICUS). Methods and Materials: 15 patients with suspected coronary artery disease underwent retrospectively ECG-gated CT-CA (Sensation 16, Siemens, Germany) and conventional coronary angiography (CAG) with retrospectively ECG-gated ICUS (Intelligate, CURAD, The Netherlands). A dedicated software (QCT, CURAD, The Netherlands) was used to extract the CT-CA dataset for the evaluation of the coronary arteries. The CT-CA dataset and ICUS dataset were loaded into a software platform for quantification purposes. Then, a target vessel region was identified for both CT-CA and ICUS. An operator performed all QCT quantification and another one all ICUS quantification. The parameters analysed were: target vessel length, lumen volume, vessel volume, plaque volume. The difference between the measurements were tested with a paired t-test and a p value of < 0.05 was considered significant. Results: Overall 15 coronary vessels were studied (12 left anterior descending, 2 left circumflex and 1 right coronary artery). The mean target vessel length was 24.7 ± 12.9 and 25.7 ± 14.4, the mean lumen volume was 242.7 ± 121.5 and 255.6 ± 157.3, the mean vessel volume was 428.7 ± 212.7 and 476.1 ± 292.8, the mean plaque volume was 189.4 ± 94.6 and 220.5 ± 149.9, for ICUS and QCT, respectively (p > 0.05). There were no significant differences when the repeated QCT measurements were compared. Conclusion: Quantitative parameters can be extracted from CT-CA dataset. The new software allows to obtain quantitative assessment of coronary artery parameters without significant differences with ICUS. Purpose: Beam hardening is regarded as one of the key factors causing artefact and impeding a properly post-operation evaluation using computed tomography (CT), especially when a metal implant presents. This research provides an algorithm to approximate beam hardening in the projection of CT to deal with the illposted problems, amend the projection, and obtain a clearer reconstruction. Based on the initial reconstruction, ideal correspondence of the metal implants and human tissues is generated separately by forward modelling, which is used as reference of beam hardening approximation. A group of curves are selected for beam hardening adjustment and the parameters are identified using neural networks under the supervised by consistency test, to obtain an optimal beam hardening approximation and computational efficiency. The projection is amended according to the beam hardening approximation prior to reconstruction. This method is demonstrated by an experimental scanning. Results: A repaired broken bone of a leg of pork is selected as the scanning object with a stainless steel plate and scews, with the parameters of 140.0 kV and 40 mA, using a GE CT scanner. In the original reconstruction, serious artefact was caused. Using the method in this research, the information lost in the original reconstruction was recovered and the reconstruction was significantly clearer. Conclusion: Beam hardening approximation is transferred into parameters identification, which is performed automatically using neural networks under the supervision of consistency test. The amended projection provides clearer reconstruction where the lost information caused by beam hardening is successfully recovered. Purpose: Although examinations of heart function are widely conducted in CT scanning and nuclear medicine, there is no existing dynamic type of phantom which can be used for both purposes. In this experiment we created a new phantom with heartbeat movement, which can be used in CT scanning as well as nuclear medicine examinations. We created a completely new type of phantom, which exactly imitates heart action. We injected a contrast agent into the heart lumen and radioactive isotope into the heart muscle. The aortic valve began to move, triggered by the heart movement, thus causing the ventricle lumen to distort at contraction. With this phantom we studied the heart function by both CT scanning and nuclear medicine examination and compared the results. Results: We measured the cardiac output rate by both CT scanning and nuclear medicine examination using the phantom, operating under the same conditions. Compared with the primary cardiac output rate of 45%, we observed a 43% cardiac output rate in the CT scan and 35% in the nuclear medicine. C D E F B A G The heartbeat-operating phantom we used in this experiment precisely duplicated actual heart function. In the CT scan, we were able to obtain accurate data of variations in the position of the aortic valve as well as the condition of the ventricle lumen. However, the nuclear medicine examination failed to produce accurate data of cardiac output rate. These results, therefore, have proven that CT scanning is capable of accurate volume measurement. Purpose: This phantom study was conducted to compare the ability of two different versions of automatic exposure control (AEC) systems to reduce effective dose in abdominal CT at different patient sizes. Methods and Materials: Effective dose was measured in an Alderson-Rando-Phantom with 100 thermoluminescence dosimeters, representing the 23 organs required by ICRP. Abdominal CT-scans were performed using a 16-slice multidetector-CT (Somatom Sensation16, Siemens). The original AP-diameter of the phantom (22 cm) was increased to 28 and 34 cm using two additional layers of fat-equivalent material to simulate slim, normal and obese patients. The three phantom sizes were scanned without AEC, with mAs-modulation only in the axial plane (CareDose2D®, Siemens) and using mAs-modulation in the axial plane as well as along the Z-axi (CareDose4D®). Each measurement was performed three times. The mean effective dose for the slim phantom was 12.5mSv without AEC, 11.4mSv with CareDose2D and 6.3mSv with CareDose4D. For the normal sized phantom these measurements were 18. Purpose: Intravascular optical coherence tomography (OCT) is a new imaging modality providing micro-structural information of atherosclerotic plaques with a resolution of 10-20 µm. Intravascular ultrasound (IVUS), current reference standard in vivo, has limited resolution of 150-300 µm. Purpose of the present study was to compare OCT, histopathology and IVUS in differentiating atherosclerotic plaques and quantifying vascular dimensions in crural arteries ex vivo. Methods and Materials: 50 atherosclerotic segments were derived from 5 amputated lower limbs. Different plaque types (fibrous, lipid-rich, calcified) were assigned by 2 examiners and sensitivity, specificity as well as intra-and interobserver agreement were calculated. Quantitative measurements comprised luminal area (LA) and vessel wall area (VA). Results: 50/50 segments were available for comparison. Sensitivity and specificity for OCT criteria were 81% and 89% for fibrous plaques, 100% and 93% for lipid-rich plaques, and 80% to 89% for calcified plaques (overall agreement 83%). Inter-and intraobserver agreement was high (κ ϖαλυεσ οφ 0.86 ανδ 0.89, π < 0.001). There was also a high correlation between quantitative measurements (Bland- Purpose: To evaluate and compare the angiogenic and arteriogenic capacity of thrombin with other well-established growth factors (bFGF, VEGF) in the New Zealand rabbit, in hindlimb ischemia. Methods and Materials: Bilateral hindlimb ischemia was invoked by surgical excision of the femoral artery. On day 20, after quiescence of endogenous angiogenesis 5,000 IU of Thrombin or 10 µg of VEGF or 10 µg of bFGF was intramuscularly injected in the medial thigh of one ischemic hindlimb per rabbit (n = 4-5 in each group), while the contralateral limbs were injected with equal volumes of normal saline and served as the control groups. On day 40, functional CT (fCT) at the level of the saphenous artery below-the-knee and intraarterial DSA followed by computerized quantitative analysis of the depicted vasculature were performed. Results: fCT studies calculated significantly increased peak enhancement values and shorter time-to-peak periods in all groups of agents as compared to control hindlimbs. Quantitative DSA detected a significantly increased total area of large collaterals only in the Thrombin group (p < 0.05). Histopathology of the adductor muscle in the medial thigh documented a significantly increased arteriogenic outcome in the thrombin group (p < 0.05) and only a mild non-significant elevation of small tunica-covered vessels in the bFGF and VEGF groups. Conclusion: Thrombin augments arteriogenesis and ischemic tissue reperfusion and seems to outperform bFGF and VEGF in establishing new vessel conduits with a developed tunica media in the ischemic rabbit model. Results: MR imaging sharply demonstrated the superficial temporal arteries allowing for an evaluation of their lumen and vessel wall. 31 of the 64 patients were finally diagnosed as having giant cell arteritis. Scoring of the mural inflammation with MR resulted in a sensitivity of 80.6 % and a specificity of 97.0 % as compared to a sensitivity of 77.8 % and specificity of 100 % based on histology. The mean wall thickness increased significantly from 0.39 (+-0.18) mm to 0.74 (+-0.32) mm (p < 0.001) and the lumen diameter decreased significantly from 0.84 (+-0.29) mm to 0.65 (+-0.38) mm (p < 0.05) for the patients with giant cell arteritis. Conclusion: Contrast enhanced high resolution MR imaging is feasible for noninvasive assessment of mural inflammation in giant cell arteritis and assisting in the diagnosis of the disease, the selection of a biopsy site, and long-term monitoring of therapeutic intervention. Purpose: To prospectively compare the costs and effects of three non-invasive imaging tests as the initial imaging test in the diagnostic workup of patients with peripheral arterial disease. A total of 514 patients with peripheral arterial disease were randomized to magnetic resonance angiography (MRA) or duplex ultrasound (DUS) or computed tomographic angiography (CTA) in four hospitals. The outcome measures included the clinical utility, functional patient outcomes, quality of life, and costs related to the initial imaging test during 6 months follow-up. Clinical utility was assessed with therapeutic confidence (0-10) in the initial imaging test. Functional patient outcomes included change in ankle-brachial index, maximum walking distance, and clinical status. Quality of life was assessed using 3 questionnaires. The diagnostic cost included the initial imaging test and all additional vascular imaging. The therapeutic cost included percutaneous vascular interventions and vascular surgery. We analyzed the differences between group means adjusted for clinically important baseline characteristics with multivariable linear and logistic regression. Results: A significant higher clinical utility was found for MRA and CTA compared to DUS. There were no statistically significant differences in improvement in functional patient outcomes and quality of life between the groups. The total costs were significantly lower in the CTA group compared to the MRA group and the DUS group. The results suggest that both CTA and MRA are clinically more useful than DUS and that CTA leads to cost-savings compared to both MRA and DUS in the initial imaging evaluation of peripheral arterial disease. Purpose: To determine the prevalence and significance of arterial inflow stenoses in dysfunctional hemodialysis access fistulas (AVFs) and grafts (AVGs). Contrast-enhanced magnetic resonance angiography (CE-MRA) was performed of 60 failing AVFs and 33 AVGs in 54 men and 39 women (mean age 62 years; age range 31-86 years). Complete arterial inflow (from the subclavian artery), shunt region and complete venous outflow (including subclavian vein) were depicted at CE-MRA. In addition to standard digital subtraction angiography (DSA) of the shunt region and outflow, DSA of the complete inflow was obtained if CE-MRA had depicted a significant (> 50%) arterial stenosis. Endovascular intervention of significant lesions was performed after DSA. Results: CE-MRA demonstrated 16 significant arterial lesions in 11 (12%) patients. DSA confirmed 15 of these lesions in 10 patients and depicted no additional inflow lesions. Of these 10 patients, seven had an isolated arterial stenosis and three had accompanying stenoses in the shunt region and/or outflow. Referral criteria for the 10 patients to undergo access evaluation had been decreased flow rates (7 cases), steal symptoms (2) and insufficient access maturation (1). Access flow of the seven patients with low flow access improved from 496 ± 73 ml/ min to 860 ± 212 ml/min after angioplasty. One patient with steal symptoms became symptom free after angioplasty. Endovascular intervention of two patients proved to be unsuccessful. Purpose: To evaluate the diagnostic performance of water-excitation trueFISP in cartilage abnormalities of the knee. Methods and Materials: 29 patients (30 knees, 13 men and 17 women) with a mean age of 56 years (range 18-86 years) were prospectively evaluated with a preoperative MR examination including a sagittal 3D trueFISP sequence and a standardized evaluation of cartilage surfaces during surgery. The mean time interval between MR imaging and surgery was 1 day (range 0-9 days). During arthroscopy the medial and lateral femoral condyles and tibial plateaux, the femoral groove and the patella were separately evaluated using a modified Outerbridge score. The MR images were evaluated separately and blinded by two readers on two separate occasions, using the same score. Results: If the cutoff between normal and abnormal was set between grades 1 and 2, sensitivity, specificity and accuracy were 54%-64%, 76%-92% and 70%-73% respectively. For a cutoff between grades 2 and 3, sensitivity, specificity and accuracy were 53%-64%, 91%-95% and 80%-83% respectively. Mean interobserver agreement was 0.73 and 0.65 for the first and second evaluations respectively at the first cutoff, and 0.78/0.79 at the second cutoff. Intraobserver agreement was 0.84/0.78 for Reader 1 and 0.60/0.83 for Reader 2. Conclusion: 3D water-excitation trueFISP represents a viable alternative to imaging sequences currently used for detection of cartilage abnormalities. Purpose: To compare the reproducibility of patellar cartilage volume and thickness measurements with a validated 3D-FLASH WE and an optimized 3D-true-FISP WE sequence at 3.0 T. The patellar cartilage of 6 healthy volunteers was examined. The measurements were performed on a 3.0 T whole body imager (Magnetom Trio, Siemens Medical Solutions) using a transmit-receive extremity-coil. The patella was covered by 40 axial partitions (0.31x0.31x1.5 mm). Image data were acquired with a 3D-FLASH WE and a 3D-trueFISP WE sequence. To assess the reproducibility, 3 consecutive data sets of each volunteer were acquired for both sequences, the knee joints being repositioned. Cartilage was delineated by a segmentation routine. Volume, mean and maximum thickness were calculated, reproducibility and interindividual variability were determined. Results: Patellar cartilage volume and thickness calculated from the trueFISP images were smaller than from the FLASH images (p > 0.05). Reproducibility of cartilage volume and thickness showed slightly lower values for the trueFISP sequence, the inter-individual variability was comparable. Conclusion: In healthy volunteers, patellar cartilage volume and thickness from 3D-trueFISP images were smaller than from the 3D-FLASH data. The lower values resulting from the trueFISP sequence might be attributable to high SNR/ CNR of synovial fluid that might lead to underestimation of cartilage. Reproducibility was slightly better for the FLASH sequence, reflecting a complex signal behaviour of the trueFISP-sequence with a slightly lower reliability of cartilage delineation. SSFP techniques however, may improve cartilage segmentation in OA patients where cartilage-joint differentiation is difficult. Imaging articular cartilage of the ankle at 3.0 and 1.5 Tesla J.S. Bauer, C. Barr, L. Steinbach, D. Malfair, R. Krug, C. Ma, T.M. Link; San Francisco, CA/US (jan.bauer@sippe-midgard.de) Purpose: To optimize MR imaging of the ankle cartilage at 1.5 Tesla (T) and 3.0 T in volunteers and to compare these optimized sequences concerning image quality and in assessing cartilage pathology in fresh human cadaver specimens. Methods and Materials: Bandwidth, repetition time (TR), echo time (TE) as well as flip angle were optimized for (i) a fat-saturated (fs) fast spin echo (FSE), (ii) a fs spoiled gradient echo (SPGR), and (iii) a balanced free precession steady state (bSSFP) sequence in three volunteers at 1.5 T and 3.0 T. Optimization outcome measures were effective signal-and contrast-to-noise-ratios (SNR, CNR), artifacts and image quality assessed by two radiologists in consensus. Using the optimized sequences 74 regions of interest in thirteen fresh human cadaver ankle joints with (n = 33) and without (n = 41) cartilage pathology were imaged. Four radiologists independently assessed cartilage pathology, separately for every sequence according to a five-level-confidence-scale. Macroscopic findings after dissection served as standard of reference. Receiver-operator-characteristic (ROC) curves were calculated to investigate diagnostic performance for the different sequences at 1.5 T and 3.0 T. The optimized sequences at 3.0 T were characterized by higher BW and TR as well as SNR and CNR. The FSE sequence at 3.0 T showed the highest area under the curve (A Z value 0.89), significantly better than all other sequences (at 3.0 T: A Z bSSFP=0.83, A Z SPGR=0.80; at 1.5 T: A Z FSE=0.77, A Z bSSFP=0.75, A Z SPGR=0.67, p < 0.05). Conclusion: Among all protocols used, the 3.0 T-FSE sequence was suited best to assess cartilage pathology at the ankle. Comparing 3.0 T and 1.5 T, all sequences consistently showed a better diagnostic performance at 3.0 T. Reproducibility of patellar cartilage T2 relaxation times at 1.5 T and 3 T C. Glaser 1 , T. Mendlik 1 , S. Wagner 1 , W. Horger 2 , M.F. Reiser 1 ; 1 Munich/DE, 2 Erlangen/DE (christian.glaser@med.uni-muenchen.de) Purpose: To compare reproducibility of T2 relaxation time quantitation in articular cartilage at 1.5 and 3.0 T. Methods and Materials: Three consecutive Multiecho (3000 at 1.5 T; 4500 at 3 T/13.2-105.6 ms) and Flash datasets (0.6² x 3 mm³) were acquired from patellar cartilage of the right knees of 6 healthy volunteers with repositioning of the knees between the measurements. Segmentation of cartilage (from Flash) was superimposed on the Multiecho sequence and T2 values were calculated on a pixel by pixel basis. Reproducibility was determined as coefficient of variation from the 3 consecutive measurements for the total patella, superficial, middle and deep layers and for both facets and ridge. Results: The intra-individual reproducibility for the lower, middle and upper layer ranged from 0.9 to 7.5% (1.5 T)/0.7 to 5.0% (3.0 T), from 1.3 to 5.1% (1.5 T)/2.5 to 7.5% (3.0 T) and from 3.0 to 9.6% (1.5 T)/1.4 to 9.5% (3.0 T) respectively. The average reproducibility was 4.1, 3.8 and 6.3% (1.5 T)/3.7, 4.6 and 5.4% (3.0 T) respectively. Inter-individual variability was 2.7, 4.7 and 7.2% (1.5 T)/7.2, 4.5 and 5.6% (3.0 T) respectively. Cartilage T2 values at 1.5 T data were significantly higher than at 3.0 T data for all three cartilage layers (p1 = 0.00002, p2 = 0.0009, p3 = 0.03). Conclusion: Reproducibility of regional patellar cartilage T2 times varies between 1 and 9% and is comparable between 1.5 T and 3 T. It is small compared to described changes in OA (up to 100% increase). Data give a basis for follow-up exams in OA. Three specimens were exposed to papain-solution for proteoglycan depletion, to collagenase-solution for collagen depletion and to buffer solution without enzymatic treatment. Then, patellae were placed in 2.5 mM of Gd-DTPA solution and C D E F A Sunda Sunda Sunda Sunda Sunday y y y y G high resolution (in plane resolution 300 µm) T1-weighted SE, T1-and T2-parameter images were acquired on a 1.5 T scanner every 30 minutes for 11hrs. T1 and T2-changes were calculated throughout cartilage thickness and compared for intact, proteoglycan-and collagen-depleted cartilage. Furthermore matrixdegradation was correlated with histological findings. Results: Histological findings demonstrate that these enzymatic treatments successfully altered collagen and proteoglycan concentration selectively. For proteoglycan and collagen degraded cartilage a significant decrease in T1 and T2 in the presence of Gd-DTPA was observed in comparison to intact cartilage. The T1-and T2-effects were of the same magnitude and depended on the depth of the cartilage layer, being most pronounced in superficial layers (1 mm) and less pronounced in the deeper calcified layers. There was no difference, neither in T1 nor in T2 in the presence of Gd-DTPA between proteoglycan and collagen degraded cartilage. Conclusion: T1 as well as T2 can be used for Gd-DTPA enhanced cartilage imaging to demonstrate early matrix-degradation, but it is not possible to differentiate between proteoglycan and collagen loss. Delayed gadolinium-enhanced magnetic resonance imaging ( For macroscopic classification the resected femoral head was divided in different regions and analysed, then pin-marked and imaged with an isotropic MR imaging protocol. Afterwards the femoral head was fixed and semi thick slices of 0.3 mm with toluidin blue staining were histopathologically registered by the pin locations to the MR data. Results: We found correlation of macroscopic changes to MR imaging in 3D-DESS and SS-DWI for normal, moderate and severe grades of OA. For histological changes in the articular cartilage we found specified alterations in signal intensity for 3D-DESS and the trace weighted diffusion. For the 3D-DESS we found correlation with the amount of proteoglycans represented by the toluidin blue staining, but a high percentage of false negative for mild changes. The SS-DWI did not correlate directly with the staining and the cartilage thickness, but showed significantly better results for matrix structure changes in mild grades of OA. Our results show the potential for SS-DWI in the detection of structural defects in articular cartilage. It was possible to correlate macroscopic and histological findings to SS-DWI and use a not quantitative approach for validation of the method. Especially in clinically important, mild grades of OA, SS-DWI could be beneficial. Purpose: To compare the performance of different MR pulse sequences for highcontrast and high-spatial-resolution imaging of collagen matrices. Chondral resection 2 cm² in size was carried out resembling a grade-IV defect of a porcine femoral condyle. A bilayer matrix consisting of collagen I/III (Chondro-Gide®, Geistlich Biomaterials, Switzerland) was dissected and fixed with fibrin glue. MR-imaging was conducted using a 1.5 T device (Siemens, Germany). Twenty-seven pulse sequences were varied with regard to their main properties (T1-and T2-weighted spin-echo or gradient echo), mode of volumetric data acquisition (2D, 3D), image matrix, slice thickness (1.3-2 mm) and the choice of fat saturation techniques. Signal intensities were measured in the subchondral layer, collagen matrix, adjacent cartilage and fluid. Contrast-ratios were computed. Results: Contrast-ratios of the matrix vs. adjacent fluid/subchondral layer were calculated with values 0.05-0.88 ± 0.04 and 0.03-0.99 ± 0.05 respectively. While highest contrast was achieved using PD-weighted TSE-sequences with low echotime, high image matrix and thin sections without fat saturation, T1-weighted IRsequences yielded an intermediate contrast-ratio with optimal depiction of the defect depth. Five sequences were unable to differentiate between subchondral layer and matrix. Conclusion: Appropriate MR imaging parameter selection is crucial for optimized visualization of the matrix and has the potential to provide information about matrix integrity and success of cartilage repair. Purpose: To assess the diagnostic effectiveness of multidetector spiral CT arthrography (MDSCTA) in detecting cartilage abnormalities in the gleno-humeral joint. Methods and Materials: MDSCTA images obtained in 22 consecutive patients (mean age 50 years, range 23-74 years, 12 female and 10 male) were evaluated for glenohumeral cartilage lesions. Two musculoskeletal radiologists who were blinded to the arthroscopy report independently analysed the humeral head and glenoid cartilage (9 areas for each surface). The sensitivity and specificity of MDSCTA for grade 2 (substance loss < 50 % of cartilage thickness) or higher, and grade 3 or higher cartilage lesion (substance loss > 50 % of cartilage thickness), the Spearman correlation coefficient between MDSCTA and arthroscopic grading, and K statistics for assessing intra-and interobserver reproducibility were determined. Results: Sensitivities and specificities of MDSCTA ranged between 90 and 95 % for the detection of grade 2 or higher cartilage lesions and ranged between 80 and 98 % for the detection of grade 3 or higher lesions. There was moderate to substantial agreement between grading at arthroscopic examination and grading at MDSCTA (Spearman correlation coefficients from 0.532 to 0.651). Interobserver agreement was substantial for detecting grade 2 or higher, and almost perfect for detecting grade 3 or higher cartilage lesions (K of 0.618 and 0.876). Intraobserver agreement was almost perfect for detecting grade 2 or higher and grade 3 or higher lesions (K of 0.845 and 0.851). Conclusion: MDSCTA is an accurate tool for the study of the articular cartilage surface in the shoulder joint. , and 18 patients post local regional chemotherapy (TACE). During a period of 12 months the effects of the therapy on normal liver parenchyma and possible liver regeneration after surgical and interventional radiological procedure were evaluated using unenhanced and contrast enhanced MR imaging and CT. Volumetric examination (CT-volumetry) of non-involved liver parenchyma and tumor/metastasis pre and post intervention was examined. Results: A significant influence on the HGF process could be confirmed on loss of liver parenchyma resulting form surgery or LITT induction. Pre-interventional low HGF-values correlate with increased liver regeneration. Patients younger than 55 years show stronger dynamics than 55+ year old patients. After 6 months, original liver volume can be observed in 58.1 % (n = 18) of all patients post surgical intervention, and in 65.6 % (n = 21) of LITT patients. Patients post chemoembolization presented an increase in liver volume of 13.1 % (3.2 % -17.8 %) after 6 months by reduction of tumor volume. Interventional radiologic therapy methods such as MR-guided laser induced thermotherapy (LITT) and chemoembolization (TACE) basically stimulate liver regeneration and cause a significant increase of cytokine HGF, on average 1.7 weeks post intervention. No variation of TNF-a value can be observed. Purpose: To assess the value of triggering abdominal and torso CT scanning based on liver parenchyma enhancement in patients with potential hypodense liver lesions. Methods and Materials: 58 oncological patients underwent CT torso on 16-row and 64-row multidetector (MDCT) scanners. 21 patients were scanned with a fixed 60 second delay for the abdomen, while in 37 patients the abdominal scan was initiated when liver parenchyma enhancement reached 50 HU. In each patient attenuation values from liver, portal and hepatic veins were obtained from both the non-contrast and contrast-enhanced scans, as for 37 hypodense, presumed metastatic, liver lesions. Results: Mean delay for scanning in the triggered group was 56.9 seconds (range 43-70). Mean post contrast liver attenuation was 127.6 ± 21.3 HU (range 93.2-155.3) in the triggered and 101.7 ± 24 HU (range 63.6-132.8) in the non-triggered group. Mean parenchyma enhancement was 26 HU greater in the triggered group (p < 0.0001). The mean portal and hepatic vein enhancement was 31.5 HU and 30 HU greater in the triggered group (p = 0.0026 and p = 0.017, respectively). The mean value of hypodense lesions enhancement was 8.7 HU greater in the triggered group (p = 0.3; not significant) with a higher conspicuity (normal liver/lesion HU) in the triggered group (mean ratio 3.8) than in the non-triggered group (mean ratio 2.8), with a significant difference (p = 0.0346). Conclusion: Liver triggering is a simple technique which produces consistent and optimal liver parenchymal and vascular enhancement in oncological patients independent of their cardiovascular status while increasing the conspicuity of hypodense liver metastases. G from macroscopic measurements of the actual specimen and (2) which CT phase is most accurate to assess the true size of liver metastases, as measured in the surgical specimen. Methods and Materials: 15 patients with a total of 24 hepatic metastases from colorectal carcinoma underwent 16-row MDCT before and after 150 ml IV contrast agent administration in the arterial, venous, and 5 minutes delayed phase. Following surgery within 5 days the metastases in surgical specimens were digitally photographed and measured on magnified images. Results: The mean metastasis size was 3.4 cm as measured on the specimens. Four lesions were not identified on CT. Mean size of metastases was 3.2 cm in the unenhanced, arterial, and venous phase and 3.3 cm in the delayed phase. Mean difference in size measurement between CT phase and specimen was 28%, 20%, 17%, and 22% for the unenhanced, arterial, venous, and delayed phase, respectively. Conclusion: Neither unenhanced nor contrast-enhanced CT are very accurate to measure the size of liver metastases in comparison to the surgical specimen. CT measurement obtained in the venous phase were most accurate. Due to the high variability in size measurements between different CT phases progression or regression of metastatic disease should only be judged by comparing metastatic lesions in the same CT phase and using the same windowing. Purpose: To evaluate diagnostic impact of different section thicknesses using 64-slice CT to detect hepatic colorectal metastases. Methods and Materials: 43 patients with known colorectal metastases were examined with a 64-slice scanner (Sensation 64, Siemens). CT parameters were: collimation 0.6 x 64 mm, mAs, 250, and kVp, 120. All patients received non-ionic contrast medium (400 mgI/mL; Iomeron 400, Bracco) at a rate of 5 mL/sec. A triple-phase protocol was acquired with hepatic arterial (delay time determined with bolus tracking technique), portal-venous (60 sec), and delayed (180 sec) phases. CT datasets were reconstructed with four different section thicknesses-0.6, 1, 3, and 5 mm. Each dataset was interpreted separately by three observers in four reading sessions. Sensitivity, positive predictive value (PPV), and area under the receiver operating characteristic curve (A (z)) were calculated. Results: Mean sensitivity and PPV were 79% and 77% on 0.6-mm images, 78% and 77% on 1-mm images, 74% and 78% on 3-mm images, and 72% and 80% on 5-mm images, respectively. No significant difference in sensitivity and PPV among images was detected. Mean A (z) values were 0.79, 0.78, 0.77, and 0.75 for 0.6, 1, 3, and 5 mm images, respectively (no significant difference). With regard to small (≤ 1 cm) metastases, mean sensitivities of 0.6 (72%), 1 (71%), and 3 mm (71%) images were statistically significantly superior to that of 5 mm (62%) images (p < 0.05). Conclusion: When using 64-slice CT, a section thickness ≤ 3 mm should be chosen to maximize detection of small (≤ 1 cm) hepatic colorectal metastases. Oral manganese as contrast medium in detecting liver metastases with MRimaging at 1.5 and 3 T H.M. Dekker 1 , C. van Herpen 1 , Y. Hoogeveen 1 , J.G. Blickman 1 , H. Thomsen 2 , T. Ruers 1 , J.O. Barentsz 1 ; 1 Nijmegen/NL, 2 Copenhagen/DK (h.dekker@rad.umcn.nl) Purpose: Evaluation of the diagnostic performance of oral manganese as a new contrast medium in liver MR-imaging in patients with liver metastases. Methods and Materials: 18 patients with known liver metastases were examined with MR imaging at 1.5 T and 3 T before and 3 hours after oral administration of Mn-contrast diluted in 400 ml water. MR imaging included T1-w FLASH breathold sequences in coronal and in transversal planes. At 1.5 T contiguous 5 mm slices and at 3 T 3 mm were made. Additionally, a T2w true-FISP sequence was performed to recognize liver cysts and haemangiomas. Contrast between liver tissue and metastases was determined on the pre-and post Mn-contrast scans. The homogeneity of liver enhancement was evaluated. In addition, the number of detected liver metastasis was evaluated. Results: There were no side-effects after the intake of oral Mn-contrast. The mean liver-metastases contrast improved at 1.5 and 3 T respectively with a factor 2.1 and 1.5. Higher liver-metastases contrast increased the number of liver detected metastasis with more than 50% at both 1.5 and 3 T. In patients with a history of chemotherapy, liver enhancement was inhomogeneous, probably due to disturbance of the portal circulation, nonetheless this did not influence the improved metastases detection. Conclusion: This pilot study shows oral Mn-contrast to be a simple and promising contrast agent which results in improved visualization of liver metastases by selective increase of liver signal. Purpose: The purpose of this study is to evaluate the MR imaging findings of hepatic eosinophilic infiltration in patients with gastric cancer. Methods and Materials: Ten gastric cancer patients who had focal hepatic eosinophilic infiltration were included in this study (pathologic confirmation in 9 cases, clinical confirmation in 1 case). There is no any evidence of cause provoking peripheral blood eosinophilia in these patients. Seven patients of gastric cancer were undergoing staging workup and 3 gastric cancer patients undergoing postoperative follow-up. We evaluated the size, number and distribution of these hepatic lesions on MR scan. We also evaluated the signal intensity of focal lesions in T1-weighted, T2-weighted images and the pattern of enhancement in a dynamic contrast study. Results: A total 22 focal hepatic lesions were observed. The mean size of hepatic lesions is 1.2 cm (0.86-1.9 cm). Nineteen lesions (86%) showed subcapsular distribution. The lesions were isointense (50%) or hypointense (50%) on T1-weighted images and isointense (19%) or hyperintense (81%) on T2-weighted images. The arterial phase of contrast study revealed 11 (50%) hyperintense lesions, 10 isointense lesions, and 1 hypointense lesion. During the portal and delayed phase scans, 18 (82%) and 21 (95%) lesions were hyperintense with fuzzy margins, respectively. The focal eosinophilic infiltrations showed homogenous enhancement with fuzzy margin in the portal and delayed phases in the dynamic contrast MR study. These findings should help to distinguish focal eosinophilic infiltration from metastasis in patients with gastric cancer patient undergoing pre-operative staging or post-operative flow up. , and STIR) and three NSE (T2-TSE PACE with and without FS, STIR PACE) were obtained. Images were analysed quantitatively (lesion-liver contrast-to-noise ratio [CNR], liver-spleen CNR and liver signal-to-noise ratio [SNR]) and qualitatively (detectability of hepatic structures and lesions, presence of artefacts). Results: Highest liver SNR was found using T1-GRE sequences. Best liver-spleen CNR was obtained with T2-TSE-FS and with both HASTE-sequences, best lesion-liver CNR with T2-TSE-FS. There was no significant difference between BHSE and NSE in lesion-liver CNR. Comparing T2 weighted BHSE with the corresponding NSE, BHSE showed a better liver-spleen CNR and liver SNR. However, in the qualitative analysis T2-TSE PACE displayed better lesion clarity especially for small liver lesions, probably due to a higher acquisition matrix. Qualitative comparison of all sequences showed an advantage of HASTE, followed by the T1-GRE sequences. Conclusion: Although no significant quantitative difference in lesion-liver CNR was found, qualitatively the HASTE, the T1-GRE and the T2-PACE sequences showed the best results. Particularly in detection of lesions smaller than 1 cm, T2-PACE can be advantageous. Thus, for clinical practice, BHSE should be preferred due to reduced acquisition time and inferior artefact susceptibility. NSE may be helpful examinating respiratory insufficient patients who cannot sustain sufficiently breathing. Purpose: To describe the utility of diffusion weighted MR imaging in patients undergoing loco-regional treatment of liver tumours. Methods and Materials: Diffusion-weighted MR imaging (DWI) was performed before and after transarterial chemoembolization (TACE) (n = 22) or radio-frequency ablation (RFA) (n = 18) for hepatocellular carcinomas (n = 28) and metastatic lesions of liver (n = 12). DWI was performed with single shot echo planar sequence and unidirectional gradient sequences with two b values (200, 500) and apparent diffusion coefficient (ADC) maps generated. The lesion to liver signal intensity ratio and ADC values were calculated. Routine T2-W and T1-W sequences were also obtained. One month follow-up DWI were also performed for 6 lesions. Results: All hepatic tumours were hyperintense to liver on DWI before TACE and RFA (lesion: liver = 2.2 ± 0.43). Post TACE and RFA, the lesions demonstrated a general decrease in signal intensity (1.5 ± 0.32) on DWI. RFA lesions were either hypointense with a variable hyperintense rim or mixed-intensity. The post-TACE and RFA lesions demonstrated higher mean ADC values (190 ± 30; 155 ± 25) as compared to the tumours prior to treatment (160 ± 20; 120 ± 22) (p < 0.02). The ADC values increased after 1-month follow-up suggestive of further necrotic changes. In three patients recurrence was noted which appeared hyperintense to the treated lesion with low ADC. Conclusion: There is a trend for ADC of the liver lesions to increase following loco-regional treatment signifying necrotic changes. DWI with ADC maps may be useful for monitoring response to treatment and detect recurrence. Purpose: Assessment of the variation of ventricular volume measurements in a moving heart phantom on a 64-detector MDCT and EBT-system compared to absolute values. C D E F B A Scientific Sessions G A respiration device operated moving heart phantom was used on a 64-detector MDCT and a 100 ms EBT unit. The heart rate (HR) was varied between 50 and 100 bpm. End diastole and end systole ventricular volume of the heart phantom were determined by two independent readers using CardIQ (GE) with data of both modalities, from which ejection fraction was derived as EF=(EDV-ESV)/EDV. Stroke volume (SV=EDV-ESV) of the heart was determined with a gas-flow analyzer and EDV was calibrated by filling the phantom with water. Results: An EDV = 258 ± 1 ml was obtained from the water calibration. From the gas-flow analysis, a decrease of SV=76 ml at 50 bpm to 60 ml at 100 bpm was obtained, corresponding to a decrease of EF from 30%-23% (∆=7%). From the MDCT-data a decrease in EF was obtained from 20%-15% (∆=5%) at 50 and 100 bpm, respectively. From the EBT-data a decrease in EF was obtained from 23%-17% (∆=6%) at 50 and 100 bpm, respectively. The average underestimation of EF using CardIQ with MDCT-data yields 39 ± 9% and with EBT-data 27 ± 5%. Conclusion: EBT is more sensitive than MDCT for decrease in EF with increase of HR. Determination of ventricular volume in the heart phantom, as derived from MDCT and EBT, shows a significant underestimation at HR between 50 and 100 bpm, for EF between 23-30%. EBT more accurately determines the absolute ventricular volume as compared to MDCT. Purpose: The aim of this study was to assess left ventricular (LV) volumetric and functional parameters by using 64-slice Computed Tomography (CT) compared with cardiac-Magnetic Resonance (MR). Methods and Materials: Thirty-four patients (30 male; mean age 59.6 ± 10.4 years) with a previous acute coronary syndrome and scheduled for left ventricular function assessment with MR were retrospectively enrolled to perform within 48h a 64-slice CT (Sensation 64, Siemens, Germany) by using the standard protocol for coronary angiography. Only patients in sinus rhythm, spontaneous HR < 75 bpm, and able to breath-hold for 15 sec, were included. Patients with contra-indications to intravenous iodinated contrast material were excluded. A dedicate semi-automated software (Argus, Siemens, Germany) was used to assess end-diastolic, end-systolic LV volumes (LVEDV, LVESV) LV stroke volume, and ejection fraction (LVSV, LVEF) by using LV short axis images. The results were compared with those of cardiac MR using a standard protocol. Purpose: To quantify and compare global left ventricular (LV) perfusion and coronary flow reserve (CFR) in patients with chronic heart failure and healthy control group by measuring coronary sinus flow with velocity-encoded cine (VEC) magnetic resonance (MR) imaging. Methods and Materials: MR measurements were performed in 20 consecutive patients with chronic heart failure due to coronary artery disease and in the control group (n = 11). Global LV perfusion was quantified by measuring coronary sinus flow in an oblique imaging plane perpendicular to the coronary sinus with non-breath hold VEC MR imaging. LV mass was measured by means of cine imaging that encompassed the heart. CFR was measured at rest and after infusion of dipyridamole. Results: At rest, LV perfusion was not significantly different in patients with chronic heart failure and control group (0.85 ± 0.3 ml/min/gr vs 0.83 ± 0.26 ml/min/gr, p < 0.001). After administration of dipyridamole, LV perfusion and CFR were significantly decreased in patients with chronic heart failure compared with that of the control group (LV perfusion 1.19 ± 0.4 vs 2.39 ± 0.69; p < 0.001, and CFR 1.45 ± 0.32 vs 3.00 ± 0.99; p < 0.001). Conclusion: Combined cine and VEC MR imaging revealed that patients with chronic heart failure have normal LV perfusion at rest but severely depressed LV perfusion after vasodilatation. Impaired CFR may contribute to progressive decline in LV function in patients with chronic heart failure. G were calculated from axial contrast enhanced 64-detector row CT images in 21 patients. Patients whose heart rates exceeded 60 bpm received 5 mg bisoprolol orally. In case of insufficient heart rate reduction, up to 20 mg of metoprolol were injected intravenously. The end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume (SV), ejection fraction (EF) and cardiac output (CO) were analyzed and compared to MR measurements based on continuous short axis images. Results: Mean heart rate was 56 ± 5 bpm during the MDCT and 73 ± 9 bpm during the MR imaging examination (p < 0.05 . LAV strongly correlated with oxygen consumption (VO2 peak p = 0.024, r= -0.47, VO2 AT p = 0.045, r= -0.42, VO2%N relative to age and sex p = 0.008, r= -0.54) and pathologic hyperventilation parameter VE-VCO 2 slope (p = 0.038, v=0.43). No other of LV function parameters correlated with CPET results. Conclusion: MDCT of the heart together with CPET allow early diagnosis and risk stratification in patients with type A and B (AHA)heart failure. Volume of left atrium strongly correlates with parameters of diastolic and systolic left heart function. Patients with LAV > 120 ml are especially endangered by heart failure. LA volume is an accurate and precise morphological parameter of physiological pathologies in patients with low exercise tolerance. LA volume can be considered as an independent non-invasive imaging marker of heart failure in patients with atypical chest pain and low exercise tolerance. Purpose: Conduction delay in patients with left bundle branch block (LBBB) may result in reduced LV performance due to an asynchronous contraction pattern. In these patients cardiac resynchronisation therapy (CRT) by biventricular pacing can be beneficial. Several approaches by echocardiography and MR imaging are used to identify these patients. In the present study we evaluated the ability of Cine-MR imaging to assess mechanical inter-and intraventricular asynchrony. Methods and Materials: 28 patients with reduced LV function (16 patients with, and 12 patients without LBBB ≥ 120 ms) were imaged at 1.5 Tesla. A sharedphases trueFISP sequence using parallel imaging enabling a reconstructed temporal resolution of 6.3 ms was used. Electromechanical delay (EMD) was defined as the interval between R-wave and the peak of systolic shortening. EMD was assessed for the septal and lateral LV-and lateral RV-wall. Inter-and intraventricular delay was calculated as the difference in EMD between these regions. Patterns of contraction in patients with and without LBBB were compared. Interand intraventricular delay was correlated with QRS width. Results: In patients with LBBB, inter-and intraventricular delay (130.6 ± 17.3 ms and 322 ± 66.4 ms, resp) was longer than in patients without LBBB (18.8 ± 12.4 ms (p < 0.0001) and 51.1 ± 84.3 ms (p < 0.0001), resp). Correlation of inter-and intraventricular delay with QRS width was only moderate (r = 0.47; p < 0.0001 and r = 0.67; p < 0.0001, resp). Conclusion: Cine-MR imaging identifies mechanical inter-and intraventricular asynchrony in patients with LBBB and may be a valuable tool for the identification of patients eligible for CRT. Purpose: To evaluate the feasibility of head and neck tumor perfusion CT measurements using a two-compartment distributed-parameter (DP) physiologic model. Methods and Materials: 26 patients with extracranial head and neck neoplasms and metastatic nodes underwent dynamic contrast-enhanced CT scans obtained from a 16-detector row multislice CT scanner. After normalization of the area under the arterial curve to the concentration curve at a vein voxel, perfusion parametric maps that corresponded to blood flow (F), fractional intravascular (V1) and extracellular extravascular blood volume (V2), transit time in vascular compartment (T), first pass extraction ratio (E) and permeability-surface area product (PS) were generated. A time delay or time lag parameter (Tlag) was introduced so as to denote the time difference between the initial rise of the arterial and tissue density curves. Results: After optimization of resolution and speed, low-inherent noise parametric maps provided adequate anatomical and functional information of the slice that was scanned in vivo. The average values of F, V1, V2, E, and PS were significantly higher in primary tumors and metastatic nodes than in adjacent normal muscle structures. The mean values of T and Tlag were significantly reduced in the neoplasms than in normal tissue. The V1 and V2 maps showed wide interand intratumoral variability. F, V1, Tlag, E, and PS maps provided the best tumor delineation. Conclusion: Perfusion CT parameters obtained from a DP physiologic tracer kinetic model can be used to differentiate extracranial head and neck neoplasms and metastatic lymph nodes from adjacent normal tissue. Whole-body dual modality PET/CT for primary tumor detection in patients with lymph node metastases in the head and neck region K. Nassenstein 1 , P. Veit 1 , H. Stergar 1 , A. Gutzeit 2 , J. Barkhausen 1 , G. Antoch 1 ; 1 Essen/DE, 2 Arau/CH (kai.nassenstein@uni-essen.de) Purpose: To assess the diagnostic accuracy of combined positron emission tomography (PET)/computer tomography (CT) for primary tumor detection in cervical metastases of unknown origin compared to PET, CT, and PET+CT side-by-side evaluation. Methods and Materials: Between December 2001 and April 2005 39 consecutive patients (8 women, 31 men, mean age 59.9 ± 11.2 years, ranging from 39 to 89 years) with cervical lymph node metastases of an unknown primary (CUP) were enrolled in this study. PET/CT images were obtained 1 hour after injection of 350 MBq of fluorodeoxyglucose. Oral and intravenous contrast agents were administered in all patients to ensure diagnostic CT data. Fused PET/CT data were evaluated for primary tumor detection. The diagnostic accuracy was calculated and compared with CT, PET, and side-by-side PET+CT evaluation. Statistical analysis of obtained differences in diagnostic performance was based on the McNemar test. Results: Fused PET/CT depicted the primary tumor in 11 of 39 (28%) patients. In 28 (72%) patients the primary tumor remained occult. CT revealed the primary in 5 (12.8%), PET alone in 10 (25.6%), and side-by-side evaluation of PET+CT in 10 (25.6%) of 39 patients. Statistical analysis showed no significant differences between the imaging modalities. Conclusion: PET, side-by-side PET+CT, and PET/CT revealed similar detection rates for primary tumors in cervical CUP patients. Therefore, cervical metastases of an unknown primary may be assessed with either of these imaging modalities. However, detection rates of CT were substantially lower. Thus, inclusion of functional data for assessment of cervical CUP patients must be recommended. PET/CT. PET/CT was evaluated by a nuclear medicine physician and a radiologist in consensus, PET+CT was evaluated by a different team and the CT was evaluated by two radiologists. The readers were aware of the clinical background. Histopathological results served as the standard of reference for all lesions found in all patients. The accuracy of TNM staging was assessed and compared with the reference standard. Differences between the staging procedures were tested for statistical significance by McNemar's test. Results: Overall TNM-staging proved more accurate (p < 0.05) with combined PET/CT compared to PET+CT and CT alone. This result was based on a more accurate assessment of the T-stage with PET/CT (p < 0.05). No statistically significant difference was detected for N-and M-staging. Tumors were overstaged with combined PET/CT in 1, with PET+CT in 6, and with CT alone in 4 cases, whereas PET/CT understaged 8, PET+CT 10, and CT alone 15 tumors. Conclusion: Combined PET/CT proved to be significantly more accurate in assessing the overall TNM-stage than the compared modalities. These results are based on a higher accuracy concerning the T-stage. Since these findings are initial results, larger patient cohorts are needed to define role of PET/CT in this patient population. To establish the normal range of FDG uptake in palatine tonsils and to compare these values with uptake in known tonsillar primaries using PET/ CT D.J. Gibson; London/UK Purpose: 1. To establish normative data for FDG PET-CT uptake in the palatine tonsil and to evaluate this against tonsillar malignancy. 2. To define normal range of uptake between left and right tonsil for control group and those with tonsillar carcinomas. We retrospectively analysed 49 consecutive, non aerodigestive tract oncology patients with whole body PET-CT. This population consisted of 24 females and 25 males. This patient group had a low risk of head and neck cancer. A second group of six patients with tonsillar carcinomas was similarly analysed. Patients who had chemotherapy were excluded. FDG PET was obtained using a Discovery ST PET-CT. Images were acquired 1 hour ± 5 minutes post injection using an average of 381 MBq FDG. Two experienced observers identified palatine tonsil using axial CT images from PET/CT data. Localisation G of tonsillar FDG uptake was refined with fused SUV maps. Regions-of-interest were placed over selected slices to determine the maximum uptake for each tonsil. Results: In the control group studied, the SUV right tonsil = 4.44 ± 1.95 (M ± SD), left tonsil = 4.33 ± 2.06. Average percentage difference SUV between left and right tonsil uptake was 13% ± 9.3 (M ± SD), which did not reach significance. Five of six tonsillar primaries had SUV values outside physiological range (5 th , 95 th centiles). Conclusion: Within the control group, no significant difference in FDG uptake between left and right tonsil was observed. We quantitatively discerned 5 of 6 tonsillar carcinomas. Clinical evaluation of real-time tissue elastography (RTE) for the differential diagnosis of thyroid follicular lesions N. Fukunari 1 , K. Tanaka 2 , K. Arai 1 , A. Nakamura 1 , K. Yamazaki 1 , K. Ito 2 , T. Mitake 2 ; 1 Yokohama/JP, 2 Tokyo/JP (fukunari@fc5.so-net.ne.jp) The diagnosis and management of follicular carcinoma of the thyroid remains a controversial topic. The aim of this study was to evaluate the clinical usefulness of real-time tissue elastography (RTE) imaging for the differential diagnosis of thyroid follicular lesions. The 56 follicular tumors obtained from 49 patients were examined by gray-scale US imaging, colour Doppler imaging and RTE. A comparative study of these images was made and final diagnosis was determined by histological diagnosis. The images of thyroid follicular lesions obtained by RTE were recognized as four typical patterns as follows: 1) nodule is relatively homogenous and colored with light green, 2) the centre of nodule is colored with green, and its periphery is colored with blue, 3) light green and red mixed inside blue zone and 4) the whole tumor is hard and displayed in blue. ) and were treated again with vertebroplasty. These were distributed with 29 adjacent to pretreated vertebrae at 2.6 months (± 2.5, range 0.5-11), and 26 in distant vertebrae at 5.8 months (± 5.1, range 0.5-22). New adjacent fractures were located inferior to pretreated vertebrae in 15/29, superior in 10/29 and between two pretreated fractures in 4/29 (sandwich fracture). There were cement leakages from the pretreated vertebrae into adjacent disc spaces in 11/ 29 fractures, 18/29 occurred without intradisc cement leaks. However, 3/4 sandwich fractures occurred in cases with pre-existing intradisc cement leaks in adjacent disc spaces. The distribution of new fractures represented the fracture distribution of the whole collective. Purpose: We propose describing the procedure performed under CT-guidance of trans-articular sacroiliac application of screws on patients with posttraumatic sacroiliac disruption in 12 patients. We performed this procedure on patients with unstable but reducible sacroiliac disruption: classified as Tile B (vertically stable but rotationally unstable) and Tile C (vertically and rotationally unstable). With CT-guidance each patient then underwent percutaneous placement of sacral and posterior iliac screws. An 8.0 mm diameter cannulated screw (Stryker, Asnis) was used. Depth and angulation measurements for placement of the materials were determined from an axial image. A small skin incision was made at the entry sites. The exact length of the screw was calculated by the CT-scan. The mean time for the procedure was 2 hours. No complications, in particular infectious, occurred during the follow-up of the patients. The hemorrhage during the procedure was negligible, less than 25 cc. All patients had a satisfactory sacroiliac stabilization. Less than 20% had residual pain with sacroiliac arthrosis. Conclusion: CT-guided fixation of reducible sacroiliac disruptions is a safe technique and a good alternative to classic surgery. Percutaneous laser diode disc decompression (PL3D) P. Purpose: Oxygen-ozone injection for disc nucleolysis is a minimal invasive therapy for patients with chronic low back pain. This study assessed potential dependencies of the outcome regarding different underlying morphologic disorders, multilevel degenerative disease, age and gender. A total of 612 patients were treated with an intradiscal (4 mL) and periganglionic (6 mL) injection of an oxygen-ozone mixture at an ozone concentration of 30 µg/mL, followed by a periganglionic injection of long lasting corticosteroid and anesthetic. Patients were divided into 5 groups according to underlying disease. Questionnaires regarding the Oswestry disability score (ODS) and a visual pain scale (VPS) were evaluated 6 months after treatment. Results: At 6 months follow-up, improvement in ODS was statistically significant, in particular in patients with disc protrusion (mean value before and after treatment 46% and 31% respectively, Wilcoxon test, Kruskal-Wallis test p < 0.0001). Improvement in VPS was also statistically significant (mean value before and after treatment 8.5 and 5.4 respectively, Wilcoxon Test p < 0.0001). Patients with only one affected segment showed a significantly better outcome compared to patients with multilevel disease, and patients under 50 years had more benefit from treatment. Conclusion: Disc nucleolysis by ozone injection is an effective, minimally invasive treatment in patients with chronic low back pain who fail to respond to noninvasive means, especially if they have single level disease, disc protrusion, and are under 50 years of age. . There was one minor complication consisting of L5 root irritation after RFA plus vertebroplasty of a sarcoma metastasis, which subsided without treatment after 8 weeks. There was one major complication involving intraspinal cement leakage during tumour vertebroplasty, causing T5 root compression and requiring laminectomy for cement removal. The overall rate of major complications requiring treatment was 0.99%. Conclusion: Vertebro-and kyphoplasty can be safely performed under MSCTfluoroscopy guidance alone. The rate of major complications is very low. We saw a high rate of small asymptomatic cement leakages which would likely be overlooked on conventional fluoroscopy (CF). This raises the question of whether CF is indeed the standard of guidance. Further studies comparing CF and MSCTfluoroscopy are therefore required. There was no statistically significant advantage for kyphoplasty concerning cement leakage and technical success, leaving the increased costs without an obvious benefit. Purpose: To assess the efficacy, accuracy and safety of foraminal and direct intracanalar epidural steroid injections under CT guidance for sciaticas and cruralgias. The long term analysis of the clinical results is reported and discussed. Methods and Materials: Five-hundred patients were treated for sciatica and/or cruralgia due to a disk herniation, without sensory-motor deficit, resisting conventional medical treatment of 3-4 weeks duration, between January 2000 and November 2004. The patient is in the prone position; the incriminated level (disk herniation) is first controlled with a lateral scout-view (CT). After local anesthesia, the extremity of a thin needle (22G) is positioned under CT guidance as follows: in the proximal part of the foramen in the case of lateral herniations (cruralgias,1/4 of the patients series) and directly in the lumbar canal, through the ligamentum flavum, in a strict epidural location, near the root. A contrast CT control verifies the epidural diffusion space (even with a foraminal approach) before the corticoid infiltration (Cortivazol 7.5 mg). The clinical follow-up is realized at 1, 3, 6 and 18 months. GP questioning, professional activity recovery and medical treatment stoppage confirm 65 % of good results after a mean delay of 26 months. The treatment failed in 35% of cases (175/500) but less than half of these patients underwent surgery (85/500 = 17%). Conclusion: Delayed corticoid injection under CT guidance allows, in all anatomic situations, marked infiltration of the root, the ganglion and also of the intracanalar epidural space. This rapid, safe and accurate procedure now represents the treatment of choice in the case of resisting radiculalgia particularly in ambulatory patients. The operated failure rate (17%) after procedure, mainly during the first month, is acceptable. Results: The measurement method proved to be reliable in control vertebrae. Mean height increased significantly after VTP (5-30% of initial height, p < 0.001). The best elevation was obtained in the center of the vertebrae, while the posterior wall, being less compressed initially, displayed only mild augmentation. Conclusion: VTP is able to increase the height of compressed vertebral bodies treated for symptomatic osteoporotic fractures. Percutaneous nucleoplasty in the treatment of discal lumbar pain S. Masala, A. Ursone, R. Fiori, F. Massari, G. Simonetti; Rome/IT (salva.masala@tiscali.it) Purpose: To assess the effectiveness of percutaneous nucleoplasty in the treatment of lumbar pain due to contained discal herniation. Background: Ninety-two patients with positive visual analogue scale (VAS) (mean score: 8) and lumbar pain were selected for treatment. In all patients lumbar MR imaging revealed a contained discal herniation (3 pts: L1-L2; 15 pts: L2-L3; 26 pts: L3-L4; 42 pts: L4-L5; 6 pts: L5-S1). All patients were refractory to medical and rehabilitative therapy for more than three months. The main physical principle of this procedure is a bipolar radiofrequency energy emitted by an electrode-needle, with consequent volumetric reduction of the nucleus polposus and elastic return of the anulus fibrosus. The treatment is performed under biplanar fluoroscopic guidance. Results: Postprocedural evaluation confirmed reduction of the VAS (mean score: 3). MR imaging performed 6 and 12 months after the procedure demonstrated complete resolution of the treated discal pathology in all patients. No post-procedural complications or disease recurrences were observed. In our experience percutaneous nucleoplasty demonstrated to be an easy, effective and feasible technique. This minimally invasive procedure is ideal for treatment of patients with painful contained discal herniation, refractory to conservative medical therapy, not candidates for surgical treatment. Purpose: To evaluate the therapeutic benefit and morphologic changes after CTguided intradiscal and intraforaminal ozone-oxygen injection combined with an intraforaminal drug application. Methods and Materials: 69 patients with lumbar radiculopathy were treated with a combined intradiscal and intraforaminal injection of ozone-ozygen, Betamethasone-21-dihydrogenphosphate, Betamethasone-21-acetate and Bupivacainehydrochloride under CT-guidance. During a period of 3 months the effects of the therapy on lumbar radiculopathy were evaluated. Volumetric examination (MR imaging-volumetry) of disc herniation was accomplished, changes on lumbar radiculopathy were examined with a 0-10 numerical pain rating scale. Results: At 7 days after injection 15 patients (21.7%) showed a complete relief of pain, in 12 patients (17.4%) we had a reduction of pain of 1 point, in 10 patients (14.5%) a reduction of 2 points, in 8 patients (11.6%) a reduction of 3 points and in 20 patients (29%) a reduction of ≥ 4 points. In 4 patients (5.8%) no significant change in lumbar radiculopathy was demonstrated. Three months after intervention the follow-up revealed in 17 patients (24.6%) a complete relief of pain, a strong relief (reduction of > 4 points) in 23 patients (33.3%), in 22 patients (31.9%) a mild relief (reduction of 1-3 points) and in 7 patients (10.1%) no relief of pain was demonstrated. Volumetric examination showed a reduction of disc herniation, an average of 13.4% after three months. Conclusion: CT-guided combined intradiscal and intraforaminal injection of ozoneozygen represents a therapeutic alternative for lumbar radiculopathy with promising results. Its ease of execution permits the successful outpatient treatment of lumbar sciatic pain. Posterior arthrodesis of the spine by percutaneous CT-guided application of screws Purpose: The intention of the trial was to compare differing teaching and learning methods (classroom teaching, print media, and digital media) conducted in standardized conditions. This was carried out in order to achieve an increase in efficiency by utilising the best teaching method possible. In order to standardise the information that was to be imparted, the digital media was compiled on the same basis as that of the print media. The lecturers adhered to these media during their classroom instructions. Thus, medical students in the second semester, who possessed some anatomical but no radiological qualifications, were lectured by this method in the subject of X-ray anatomy. These teaching methods were evaluated objectively with multiple choice tests and subjectively with a student survey. The evaluation of the objective data showed no detectable differences concerning the teaching methods. The subjective data, however, displayed that the students' preferred a combination of digital media and classroom teaching by dedicated lecturers. Learning by print media was mostly depreciated. The lack of experienced and committed lecturers, who are superior to digital media, necessitates compromises. The trial indicates, therefore, that the combination of digital media and highly qualified, dedicated lecturers would provide an ideal solution to the current problems. Purpose: The purpose of the paper was to review eHealth initiatives to improve healthcare and lower costs with information and communication technology in Europe and North America, and to define implications on the business of radiology. and North America to augment provider-based systems with regional and national solutions to improve accessibility to imaging information. This work includes comparing approaches in different geographic regions including technologies and business impacts. In 2004 an eHealth initiative accompanied by an action plan was launched by the European Commission. In Canada, there is an initiative called Canada Health Infoway. The US government has called for a national health record to be available in a decade and has begun assisting states and communities to implement regional solutions linking together. Results: eHealth programs seem to be the key drivers to change the business of radiology. In summary the implications are 1) there will be regional and national projects, 2) projects will be driven by high level governmental decision makers, 3) purchasing will be focused on solutions and services rather than software and hardware components, 4) PACS will be integrated to national level healthcare record summaries, 5) images will be utilized across boundaries and 6) long-term archiving will include all clinical data, not just images. There are different technical approaches, but similar business implications, of the process to improve health and lower costs in the eHealth programs in Europe and North America. . ERDDS aims to integrate the national radiological nomenclatures through a software platform. It will represent a strong pillar in achieving a European shared classification of radiological services. Methods and Materials: The system requirements have been defined through the analysis and modelling of real data (clinical and administrative nomenclature, etc). collected from content providers. The project bottom-to-top approach is assured thanks to the identification of the content providers in the final users, and to the IT modelling directly built upon their realities. Results: Through best practice principle, a European Radiological TransNomenclature (ERTN) has been defined. A platform prototype has been developed. It is composed of two modules, the first transcodes the different nomenclatures and allows the second to perform benchmarks (on prices, volumes of activities, technologies, etc). among the healthcare providers belonging to the network. The prototype offers to MD and patients the radiological appropriateness guidelines. Conclusion: The platform will be able to communicate directly with the hospital information system. These tools will allow health professionals and patients to access, through web technology, comparable information in the radiological sector, and so be of use for clinical studies, department management and for steering the patient to the appropriate health services. Evaluation of formal quality of radiological educational resources on the web T. Adla, P. Kasal, M. Hladíková, A. Janda; Prague 5/CZ (ta@slymak.cz) Purpose: Expansion of use of the internet has brought genesis of many web educational resources, including those directed at radiologists. The quality of each individual website varies considerably. The aim of the study was to investigate formal qualities among radiological educational resources. The main inclusion criterion was radiological educational content of a website (whole or a considerable part). Secondly, resources were selected with respect to covering a whole spectrum of online CME resources (from big portals to one-man-sites). 10 resources were selected. Rankmed.cz was used for the evaluation, which is based on internationally accepted standards. Websites were evaluated by 20 criteria, which are divided into 4 groups describing the main formal qualities of website: Presentation, navigation, functions and credibility. Results: Overall average score was 59.8 points from 100 available (min. 38.0, max. 76.5). Regarding the groups of criteria, presentation contributed to the average score with 20%, navigation 21%, function 23% and credibility 36%. Average results for particular types of resources were: portals 65.4, databases of cases 63.8, online books 59.5 and one-man-sites 47.7 points. Conclusion: Hypothesis of the presence of measurable quality differences among tested websites was confirmed. Surprisingly the average credibility of websites was relatively high (making up more than one third of the average score). Difficulty in creating websites and keeping them up-to-date is evident -the best score was achieved by websites which are provided by professional teams, as opposed to one-man-sites and websites provided by small groups. The course included open questions and simulated patient cases (www.prewise.fi/radiationsafetytraining). 210 participants (96 female, 114 male, aged 25-64 years) fulfilled the questionnaire (a Likert-type format with five potential responses) including questions regarding their expectations, the scope, schedule and the benefit of the course. In addition, we asked if they found using the internet to be a difficult learning method, whether e-learning saved time and whether they felt that they learned better compared to conventional lectures. Results: 19% of the respondents had previous experience in e-learning. They found e-learning to be easy (74%) and timesaving (61%). 57% stated that they learned better using e-learning compared to conventional lectures. 91% found that the course fulfilled their expectations. 84% stated that they benefited from the course and 94% were willing to study using the internet in the future. When using multiple ordinal regression no sex or age differences in answers were found. A significant positive Spearman correlation was found between the participants' expectations and the questions regarding the scope (p < 0.001) of the course, the benefit derived from the course (p < 0.001), the time-saving aspect (p < 0.001) and better learning (p < 0.002). Conclusion: e-Learning seems to be a well accepted and practical learning method for radiation protection training. Purpose: Wide distribution and re-use of standardised eLearning content is essential for today's education strategies. The aim of our scenario is to show that while using open international standards and standard interfaces eLearning content creation can be almost as easy as writing a text document and therefore can be done by all knowledge providers who are familiar with standard office applications. Nevertheless for installation and maintenance of the system IT professionals are needed. The Sharable-Content-Object-Reference-Model (SCORM) defines a standard for eLearning objects and is strongly required for the mentioned purpose. The workflow implies a teaching file database using XML-WebServices, Microsoft's PowerPoint®-application with a Simple-Object-Access-Protocol (SOAP)-adapter and the Producer-for-PowerPoint extension to generate SCORM-compatible eLearning objects. Alternatively we used the Thesis® Professional 2.5 Software which has additional features. To evaluate the resulting documents we used a freeware SCORM player and the OpenSource LMS "moodle". Results: In our scenario the knowledge provider of eLearning content just needs to be familiar with the commonly used PowerPoint application. The teaching file database is seamlessly integrated in the user workflow via WebServices. The output of the Producer/Thesis Software is usable in SCORM-ready learning management systems (LMSs). Conclusion: While using de-facto standard office applications together with the inherent interoperability features of international standardised interface protocols and data formats eLearning documents can be easily created by most of the knowledge providers themselves and not only by IT professionals. Therefore the scenario we present might be a solution to provide the emerging LMS with its necessary content. A remote access training system for the detection of colorectal polyps at computed tomography colonography V. Luauté; Dublin/IE (luautev@eeng.dcu.ie) Purpose: Computed tomography colonography (CTC) is emerging as an alternative to conventional colonoscopy (CC). However CTC is not yet in widespread use due in part to the lack of suitably trained radiologists. We have developed a novel remote access system to train radiologists for colorectal cancer screening using CTC. In order to ensure that radiologists gain the relevant experience without the need for any specialist equipment or software, we opted for designing a system that is accessible via the internet using a standard browser. The interface lets the user locate and characterise polyps with the help of appropriate tools such as windowing, polyp measurement and zooming. Each user has an account in order to allow monitoring of their training. They can also run an automatic evaluation of their work based on gold standard information previously gathered from specialists. Our system has been developed using a combination of Java servlets and applets. The interface allows the user to start working on the initial images while the rest of the dataset is being transferred. The system is able to offer fast response time. An iteration of lumen tracking using the system takes approximately 40 seconds. We have developed a system which allows a trainee radiologist to gain proficiency in an emerging colorectal cancer screening technique. This research has yielded an operational system that meets the needs of remote access users. Performance of an integrated web-based RIS/PACS system to facilitate effective mobile diagnostic imaging services in the UK C. Stavrianou, P. Tually, J. Walker; Wimbledon/UK (c.stavrianou@globaldiagnostics.co.uk) Purpose: Faster delivery of medical services and an increase in capacity has been the focus of a recent NHS scheme which has been implemented to address the excessive waiting lists for diagnostic services in the UK. The mobile operations will enable delivery of tests to regions of greatest need but require high bandwidth networks and sophisticated software applications to distribute large data volumes, interpret results and integrate with existing hospital systems. To accomplish this, a comprehensive, digital, web-based patient information system has been installed to manage patient care. The technology allows on-line scheduling, exam requisition, patient tracking and off site sub-specialist opinion in realtime. We describe the predominate factors that affected the implementation, acceptance, technical parameters to attain optimal system functionality and present results from a formal evaluation on user satisfaction. Purpose: Assessment of skeletal development is frequently performed in paediatric radiology. The most commonly applied clinical method is the atlas matching method which compares a hand radiograph with a film-based hand atlas. A computer assisted bone age assessment method has been developed. It matches automatically a patient image with the digital hand atlas. Methods and Materials: Four components are integrated in order to obtain a computer aided diagnosis system. First, a set of standard images has been selected from 500 clinical cases. A comparison of each image with hand patterns, included in an atlas of radiologically approved standard, has pointed the closest match in each age group. Then, the application software extracts radiological features describing the size, shape, and density of epiphyses and metaphyses, as well as parameters describing the stage of epiphyseal fusion. Next, standard images in diagnostic and compressed resolution along with their extracted features are organized into a hand e-atlas. At the radiological image analysis stage features are extracted and compared with the e-atlas to find the closest match. Finally, a graphical user interface has been designed. It views the image analysis C D E F A Sunda Sunda Sunda Sunda Sunday y y y y G results and displays 2 closest matches from the e-atlas for the radiological approval. Results: A set of 250 clinical hand images has been tested. The agreement between radiological and computerized assessment has reached 90%. Conclusion: The CAD system offers a fast and objective tool able to assist radiologists in performing the bone age assessment. In case of failure, a manual eatlas-based assessment is available. Purpose: Determination of skeletal development in children is important. The most used evaluation method is to study left hand X-ray based on the standards depicted by Greulich & Pyle (G&P). The aim of our study was to compare the accuracy of a new sonographically based method with the standard method. Methods and Materials: 160 consecutively evaluated children and adolescents (77m, 83f) who received a X-ray of the left hand were evaluated. Ultrasound examination of the same hand was performed on the same day using the Bon-Age™ system (Sunlight Medical Ltd., Israel). This system evaluates the relationship between the velocity of the wave and growth using gender and ethnicity based algorithms. Three investigators (U1-U3) analysed the X-ray and assigned bone age based on the G&P atlas. Results: In 152 patients BonAge™ measurement was successfully. The correlation between the three investigators using the G&P method was between 0.977 and 0.980; correlation between BonAge™ and the investigators (U1-3) was 0.902 and 0.920. The paired t-test showed no significant differences between BonAge™ and the investigators and also for the comparison between U1 and U2. There were significant differences between U1 vs. U3 and U2 vs. U3 (p < 0.05). The BonAge™ device demonstrates the ability to produce a sufficient assessment of bone age using an ultrasound method. Advantages of the ultrasound device are objectivity, lack of ionizing radiation, and easy accessibility. In the case of first investigations, X-ray is necessary to look for dissociated skeletal age, dysplasia, and mineralisation. Sonographic evaluation of nasal bone fractures in children S.W. Lee, J.Y. Whang; Seoul/KR (sonic323@mm.ewha.ac.kr) Purpose: To describe the sonographic findings of nasal bone fractures and evaluate the diagnostic value of sonography (US). Methods and Materials: Nineteen children who had sustained nasal trauma underwent plain radiography, US and CT and the CT findings were used as the gold standard. The US findings were retrospectively reviewed and compared with these of other modalities. Results: Sonographic findings of nasal bone fractures were irregularity (n = 3), discontinuity (n = 12), displacement with discontinuity (n = 14)of the thin smooth echogenic line formed by ventral cortex of the nasal bone. Twenty-nine nasal bone fractures (19 patients) were identified by CT and US but only 14 by plain radiographs. US and CT demonstrated unilateral involvement of nasal bone fracture in seven bones and bilateral involvement in 11, while plain radiograph revealed unilateral involvement in four and bilateral involvement in five. The US findings of nasal bone fractures were similar to those of CT in detecting the location and degree of depression or displacement of nasal bone fractures. We believe that US is very useful for the evaluation of such fractures, especially in children with equivocal abnormalities on plain radiographs. Purpose: MR imaging measurement of the thickness of cartilage cap of osteochondroma -a sign of malignant transformation -is regarded as the most accurate non-invasive modality. We evaluated the new BalancedFFE sequence (balFFE) in comparison to an established double-echo gradient echo sequence (3D-GE) in measuring cap thickness with histopathological correlation. In 17 patients with solitary or multiple osteochondroma, we performed either 3D-GE (n = 14), balFFE (n = 9) or both MR imaging sequences (n = 6) -in case of multiple lesions, a single lesion of each patient was measured. Lesions were benign, proven by histopathology. Thickness of cartilage was measured perpendicular to cap surface. MR imaging was correlated to histopathological measurements. Statistical analysis was done using 97.5% confidence intervals for mean of differences between MR imaging and histopathology. Results: Histologically determined cap thicknesses ranged from 2 -4 mm (mean 3 mm). BalFFE measurements (n = 9) could be compared to histology with mean of differences 0.9 mm (CI -0.7 mm through 2.5 mm). Mean of differences between first echo of 3D-GE and histology (n = 14) was -0.11 mm (CI -1.1 mm through 0.9 mm), between second echo and histology (n = 14) it was -0.3 mm (CI -1.2 mm through 0.7 mm). All confidence intervals included zero (0 mm) indicating an excellent correlation to histology. , were enrolled in the study. The MR imaging protocol consisted of: sagittal T2-weighted turbo spin-echo (2019/90) scans, sagittal and coronal, plain and contrast-enhanced, fat-suppressed, T1-weighted spinecho (425/15) scans and gradient-echo (65/14/20°) scans with incremental mouth opening. A grading score was used for the evaluation of the condyle (1=normal, 2=bifidity, 3=absence of the anterior half of the condyle), articular eminence (AE) (1=normal, 2=shallow), articular disc (AD) (1 = "bow-tie" shape, 2 = abnormal), acute pannus (1=normal, 2= pannus), intraarticular fluid (1=normal, 2=fluid) and the range of condylar motion (RCM) (1=90º, 2=75º, 3=45º, 4=0°). Results: 70 joints were examined. Higher grades were more frequent in systemic arthritis (condyle3=55%, AE2=60%, AD2=65%, RCM4=5%) than in oligoarthritis (condyle3=17.9%, AE2=0%, AD2=0%, RCM4=0%) and polyarthritis (condyle3=27.3%, AE2=13.6%, AD2=18.2% RCM4=0%) (p < 0.001). There was no significant difference between systemic arthritis, oligoarthritis and polyarthritis in the frequency of pannus (25%, 52.4%, 45.5% respectively) and intraarticular fluid (14.3%, 10.7%, 13.6% respectively). Patients with abnormal condyles had longer disease duration (11.8 ± 5.1 years) than patients with normal condyles (5.5 ± 6.12 years) (p < 0.001). Conclusion: MR imaging reveals TMJ involvement of different severity and frequency in the 3 subtypes of JIA. MR imaging may be useful for early detection of the inflammatory process before irreversible lesions occur. Up to now, 5 children who complained of sudden onset of bone pain were examined. Imaging included an X-ray of the painful region and additionally an MR imaging scan of the whole body (T2w TIRM and T1w GE sequences -pre and post-contrast (the latter with fat sat)). 4 children had at least one follow-up examination during 6 months. Results: None of the baseline X-ray images showed pathological changes. MR imaging revealed typical ON bone marrow lesions in 3/5 children. Additional to the sites with bone pain, MR imaging revealed other ON locations. All of the lesions had bone edema and contrast enhancement. Follow-up examinations showed slight reduction of bone edema during 6 months. Conclusion: Whole body MR imaging allows rapid detection of ON, much earlier than plain X-ray diagnostics. MR imaging reveals additional ON even before clinical symptoms occur. T2w TIRM seems to be the most sensitive sequence. Development of fetal adipose tissues as seen with fetal magnetic resonance imaging P.C. Brugger, D. Prayer; Vienna/AT (peter.brugger@meduniwien.ac.at) Purpose: Adult fat is characterized by T1-weighted hyperintensity in magnetic resonance imaging (MRI). The present study investigates the development of these MR imaging signal characteristics in fetal adipose tissues. [20] [21] [22] [23] [24] [25] [26] [27] [28] [29] [30] [31] [32] [33] [34] [35] [36] [37] [38] ] were imaged on a 1.5 T magnet using T2-, T1-weighted and steady-state free-precession (SSFSE) sequences. Signal intensities of adipose tissues on different sequences were assessed with respect to topographical regions. Results: T1-weighted hyperintensity of fetal adipose tissues was not evident until the beginning of the 23 rd GW when the buccal fat pad was the first to display it. By 25 GW moderate hyperintensity appeared in the ischio-rectal fossa and face, the retropatellar fat pad and the deep extensions of buccal fat pad could be recognized. By 27 GW subcutaneous tissues showed hyperintense signals, followed by popliteal (28 GW), intraorbital (30 GW) and perirenal fat (32 GW). SSFSE sequences showed marked hypointensity of buccal fat between 25-32 GW. From GW 32 onwards, chemical shift artefacts delineated subcutaneous fat from dermis and underlying structures. Conclusion: Fetal adipose tissues gain their T1-weighted hyperintense signal properties in a regular temporary sequence depending on their composition and topography. Though abundant, perirenal fat displays no T1-weighted hyperintensity until 32 GW, since it is made of brown adipose fat in the fetus which has different signal properties. These findings may have a bearing on the prenatal MR imaging detection of (intracranial) lipomas, and assessment of fetal macrosomia or intrauterine growth restriction. SSFSE sequences may be an alternative in detecting lipoma. Purpose: To study the tonotopic organization of plasticity effects in the auditory cortex (AC) of unilateral tinnitus patients when stimulated with their experienced tinnitus frequency or another frequency by means of fMRI. Methods and Materials: Fifteen Patients were scanned on a 3 T MR imaging scanner. A T2*w EPI silent gap sequence (TR = 5s) was used during the stimulation paradigm. The stimulation paradigm consisted of a blocked design of 18 epochs in which white noise filtered through a narrow (freq ± 500 Hz) or wide bandpass (freq ± 1000 Hz) and tones were presented 3 times binaurally each. The stimuli were either the tinnitus frequency experienced by the patients e.g. 8000 Hz (TF-session) or a different frequency e.g. 1000 Hz (OF-session). Data were analyzed using SPM99 software and activation patterns were visualized. Quantitative analysis consisted of calculating an activation ratio of activated voxels in the AC and assessing differences between the left and the right side with a paired samples t-test. Results: Significant fMR signal change was found bilaterally in the AC in both sessions. Signal change was symmetrical in patients when exposed to the OFsession. When exposed to the TF-session, a decrease of signal change (plasticity effect) was visualized in the contralateral AC. Comparing the activation ratio in the left and right AC showed a significant difference during the TF-session, but not during the OF-session. Conclusion: By means of 3 T fMRI we were able to demonstrate tonotopic plasticity effects in unilaterally affected continuous tinnitus patients. Acknowledge Purpose: During a face matching task healthy subjects selectively activate cortical areas along the ventral visual stream, whereas patients with amnestic mild cognitive impairment (MCI) show activation along the ventral and the dorsal visual stream. We studied the morphological substrate of this differential activation of visual areas during face matching. Sixteen subjects with amnestic mild cognitive impairment (MCI) and 17 healthy subjects underwent structural and functional MR imaging examinations. The object matching task consisted of two faces presented simultaneously and participants were asked to decide on each trial if a pair of faces were identical or not. The data were analyzed using AFNI (Cox, 1996) and FSL (FMRIB Software Library). For morphometric analysis a high resolution T1w 3D-MPRAGE sequence was performed. The processing of the 3D sequences followed a protocol implemented through Statistical Parametric Mapping (SPM). Results: Both in MCI and healthy subjects greater fusiform activation in fMRI was correlated with better preserved grey matter volume along the ventral visual pathway. Smaller fusiform activation was correlated with better preserved cortical grey matter along the dorsal visual pathway in both groups. The morphological separation between ventral and dorsal visual streams may result from system-specific neurodegeneration in MCI. The interpretation of activation pattern in fMRI should take into account the degree of cortical atrophy along the entire functional system underlying the task under study. Covert language production could not touch the nature of stuttering: An fMRI study Y. Yang, K. Li, D. Peng, C. Lu; Beijing/CN (yyh_wj@sohu.com) Our previous fMRI studies showed that the striatal-thalamus-cortical neural circuit contributes exclusively to stuttering. While it was responsible for sequencing and timing of movement in the traditional view, could speech tasks without movement touch the nature of stuttering? We wanted to test this hypothesis by fMRI. Methods and Materials: Nine normal controls and ten stuttering subjects were scanned in a 1.5 T MR imaging scanner. While the subjects named pictures covertly or overtly in two separate runs (block design), fMRI images were acquired. The data was processed using the Analysis of Functional NeuroImages (AFNI) programs. In agreement with previous results, we detected the abnormal activity of the striatal-thalamus-cortical circuit in the stuttering brains in the overt naming condition. There were also some areas in the stutterers' brains exhibiting failing deactivation, including some superior and middle frontal areas. In the covert naming condition, the areas which manifested differences between the two groups were bilateral superior and middle frontal cortex, left superior temporal cortex and right insula. Although, to some extent, it indicated different activity patterns between the two groups related to pre-articulatory procedures, it failed to reveal the important aspects of dysfunction of the striatal-thalamus-cortical circuit in the stuttering brain. Conclusion: Because of the complexity of stuttering pathology and its close relationship with language production, it was necessary to involve almost every critical aspect in the testing tasks so as to touch the nature of stuttering. Tasks such as covert naming might fail this criterion. G Results: The mean percent CBF changes at activated motor cortex were significant during handgrip (46 ± 23%, p = 0.006) and finger tapping (48 ± 17%, p = 0.001). The mean percent CBF changes at the contralateral motor cortex were 9 ± 6% during handgrip and 7 ± 23% during finger tapping. The mean percent CBF changes at the ipsilateral sensory cortex were 0 ± 17% during handgrip and 21 ± 23% during finger tapping. The CBF change at activated motor cortex showed a negative correlation during handgrip motion (r = -0.92) and a positive correlation (r = 0.56) during finger tapping to the contralateral motor cortex. The CBF change at activated motor cortex showed a positive correlation to the ipsilateral sensory cortex during finger tapping (r = 0.70). Conclusion: The significant CBF change (> 40%) during continuous hand motion detected by CASL may suffice the mapping of motor cortex with this technique. The negative correlation of the contralateral motor cortex during handgrip motion as opposed to a positive correlation during finger tapping suggests that handgrip exercise should be a better motor task than finger tapping in motor cortex mapping. The cortical representation of taste in the human brain: An MR imaging study G. Polonara 1 , M. Fabri 1 , G. Mascioli 1 , G. Tassinari 2 , G. Berlucchi 2 , T. Manzoni 1 , U. Salvolini 1 ; 1 Ancona/IT, 2 Verona/IT (g.polonara@ao-umbertoprimo.marche.it) Purpose: To identify the human primary gustatory cortex using fMRI and to establish whether this cortical representation is mainly ipsilateral, as in non-human primates, or bilateral, as recent neuropsychological studies of patients with callosal resection seem to indicate. Methods and Materials: Data were obtained from 9 healthy subjects. Functional images were acquired with a 1.5 Tesla machine. The stimulation protocol lasted 5 min and consisted of 60 sec rest, 30 sec stimulation, 90 sec rest, 30 sec stimulation, and 90 sec rest. Gustatory stimulation was applied by placing a small cotton pad soaked in a 1M NaCl solution (the salty stimulus), distilled water (the neutral stimulus) or 10% sucrose (the sweet stimulus) on the side of either hemitongue. Data was analyzed with BrainVoyager and SPM2 software. Results: Salty stimulation of the left hemitongue evoked bilateral activation of area GI in all subjects, more pronounced in the ipsilateral hemisphere. Salty stimulation of the right hemitongue also evoked bilateral activation of area GI in all subjects, but without appreciable side differences. Sucrose stimulation evoked cortical activation in both hemispheres in areas similar to those activated by the salty stimulus. Neutral stimulation activated more posterior and more dorsal cortical regions likely corresponding to the tactile cortical representation of the tongue in the first somatic sensory area. Conclusion: These findings show that the organization of taste afferents in man differs from the organization described in non-human primates, and are in agreement with previous neuropsychological studies performed in control subjects and callosotomized patients. Differences in cerebral activation between females in distinct menstrual phases during viewing of erotic stimuli: An fMRI study E.R. Gizewski, A. Baars, M. Schlamann, M. Forsting; Essen/DE (elke.gizewski@uni-essen.de) Purpose: There is evidence that men experience more sexual arousal than women, but also that women in mid-luteal phase experience more sexual arousal than women outside this phase. Recently, a few functional brain imaging studies have tackled the issue of gender differences of reactions to erotica. The question of possible differences in reactions to erotica with women in distinct phases has not yet been answered from a functional brain imaging perspective. Methods and Materials: fMRI was performed in 22 female volunteers measured twice (mid-luteal and menstrual phase). Subjects viewed erotic film excerpts alternating with emotionally neutral excerpts in a standard block-design. Arousal to erotic stimuli were evaluated using standard rating scales after scanning. Results: Two-sample t-Test with uncorrected p < 0.001 values for a priori determined ROIs involved in processing of erotic stimuli revealed differences between both scanning times: Comparing women in mid-luteal phase and during their menses, superior activation was revealed for women in mid-luteal phase in the anterior cingulate, left insula, and orbitofrontal cortex. However, only the superior activation in the orbitofrontal cortex and left insula was maintained using the sexual arousal as a covariate. No significant superior activation was revealed for women scanned during their menses when compared with women in mid-luteal phase. Our results indicate that there are few differences in cerebral activ-ity during viewing of erotic stimuli with women in mid-luteal phases and outside the mid-luteal phase. Women in mid-luteal phase revealed a superior activation in some brain areas specific for emotional processing. Purpose: The purpose of our work was to assess the usefulness of fMRI technique for comparative study of bipolar-I patients before and after treatment. In a double-blind manner, after one-week washout, 30 treated bipolar-I patients were compared before and after an 8-week trial. No refractory or recently lithium-treated patients were included. 22 patients were in depressive, 4 in mixed, and 4 in manic phases (DSM-IV-TR). MADRS, YMRS, MDQ, and CGI ratings were compared with functional MR imaging results. The control group consisted of 15 healthy volunteers. The Stroop test was implemented as the fMRI stimulus. Final and initial fMRI brain activations were 'subtracted' and used for group analysis. Two sample T-test analysis as well as ANOVA analysis were used to compare means and variations of activations. Results: Patients improved significantly in nearly all clinical ratings. The common fMRI findings in the patient group were improvement in dorsolateral prefrontal (T=5.5, p-value < 0.001) and dorsal cingulate gyrus (T=6.4, p-value< 0.001) activations. Signal increased both at cluster-level and voxel-level analyses, however remission and positive cognitive response were also correlated with some reductions in global activation (subcortical, temporal and rear frontal areas). There was statistically significant correlation between the values of activation in regions of interest and clinical rating tests (p-value < 0.001). Purpose: In recent years the use of digital imaging systems for mammography has become widespread. The assessment of image quality continues to be of paramount importance as part of the routine testing of equipment performance and images produced with test objects are commonly used for this purpose. In this study the performance of 5 test objects was evaluated on a range of digital imaging systems covering most technologies currently available for digital mammography. The 5 test objects evaluated were TOR (MAM), TOR (MAX), Dmam, CDMAM and DIGIMAM. Images were produced on 3 digital mammographic systems (GE, Lorad, Fischer) and one CR system (Konica). Additionally, screen film images were also acquired to obtain reference data. Each test object was imaged over a range of dose values, aiming at variations of image quality. All images were evaluated as softcopy by 4 experienced observers following recommended methodology. The results were analysed using "contrast-detail curves" and single image quality figures. The various test objects were compared in terms of absolute and relative performance. Additionally, the image quality performance rating of the various systems was investigated with each test object. The results shown noticeable variations in the performance of the various test objects and indicate which may be most useful. Conclusion: Contrast detail analysis using test objects is straightforward method for the assessment of technical image quality. The understanding of the suita-bility and performance of various tools available provides useful information to buyer and facilitates the comparison of results obtained with different test objects. Discussion of fog standards in mammography R.E. Dickerson; Rochester, NY/US (robert.dickerson@kodak.com) Purpose: This paper discusses the rationale for existing fog standards in mammography and suggests that these standards are based on performance from older mammography films using outdated emulsion technology. New mammographic films using modern AgX microcrystals no longer exhibit the same performance as older mammographic films. A new proposal for fog standards is suggested that is more relevant to existing mammographic films. Methods and Materials: This paper describes details of older and more modern mammography films and their response to film aging and resulting fog levels. Results: Increased fog values in mammographic films are due to continuing chemical sensitization and cosmic energy radiation. Older pre-1950's mammographic films by virtue of several factors result in higher film fog, lower toe contrast and higher film granularity. These include halide composition, grain morphology and distribution of grain diameter and volume. Modern mammographic films do not have the same response to film aging as older films systems. Purpose: To determine the optimal exposure factors for different types of CR and DR digital mammography system. Methods and Materials: Mean glandular dose (MGD) and contrast-to-noise ratio (CNR) were measured using PMMA phantoms simulating breasts 20-90 mm thick, with a 0.2 mm thick square of aluminium for contrast. For each thickness, a range of mAs, kV and target/filter combinations were used, for various DR and CR systems. The optimal exposure factors for each thickness were determined as those giving a target CNR for minimum MGD. Results: Aiming for a target CNR resulted in the selection of different exposure factors from those which give constant detector dose. The use of higher energy spectra generally resulted in the lowest patient dose, but with a higher detector dose to compensate for lower contrast. For smaller breast thicknesses (21 or 32 mm), the optimal beam quality was usually 25 or 28 kV and Mo/Mo target/filter combination. For thicker breasts a target/filter combination such as Rh/Rh or W/ Rh were usually optimal, leading to a dose saving of about 30% compared to 28 kV Mo/Mo. Systems with a relatively large amount of structure noise were optimal using lower energy spectra. Conclusion: A simple method of optimizing digital mammography systems is described, and results indicate that current systems aiming for a fixed detector dose are not optimal. Maintaining a target level of CNR usually results in the selection of higher energy spectra (Rh/Rh or W/Rh) for breast thicknesses above 45 mm. MoniQA, monitor quality assurance: Initial experience with a dedicated software tool to perform general quality control of medical screen devices with daily renewed test images J. Jacobs, T. Deprez, G. Marchal, H. Bosmans; Leuven/BE (jurgen.jacobs@uz.kuleuven.ac.be) Purpose: To develop a complete environment (MoniQA) for quality control of radiological screen devices. We report on the practical work-out of the software tools and on the first validation on a large series of different monitors in a digital radiology department. Methods and Materials: MoniQA (developed in JAVA) uses a modular approach in which patterns are packed into plug-ins. Therefore MoniQA is able to present G any available test pattern, including a newly created pattern for daily quality control. This pattern allows the evaluation of luminance, gradient, resolution and geometric distortion. Certain parts in this image are randomly created to avoid memorization. The validation of the software tool was performed in our radiology department on 68 medical screen devices. Radiologists scored the patterns on a daily basis. All scores were automatically sent to a central computer for analysis as a function of time by another tool, QAMPR. We compared the different scores and we assessed the learning curve and the reproducibility of the scores. An evaluation of the MoniQA pattern against the AAPMtg18 QC pattern is on-going. Results: We tested our tools successfully in a medical environment. The personnel accepted the evaluation procedure well (learning curve is less than one week). It takes roughly 300 s to evaluate two monitors. The results were different for different monitor types and known artefacts were detected. Conclusion: Our preliminary results demonstrate that we can appropriately monitor the quality of medical screens on a daily basis for a large series of monitors spread over different locations. Optimization of image quality and radiation dose in digital radiography D. Lavin 1 , B. Tuohy 1 , E. Keavey 1 , P. Woulfe 1 , J. Gallet 2 , W. Van Der Putten 1 ; 1 Galway/IE, 2 New York, NY/US (david.lavin@mailn.hse.ie) Purpose: Conventional film screen (FS) imaging over recent years is gradually being replaced with digital imaging alternatives such as computed radiography (CR) and direct digital radiography (DR). One of the inherent benefits of this approach is the ability to optimise image acquisition and image display independently. CR/DR and FS have inherently different imaging characteristics and therefore images should not be acquired using the same exposure factors. Insufficient attention however, has been directed in the scientific literature, at optimisation studies in diagnostic radiography using CR and DR systems. Methods and Materials: Phantoms to simulate various body parts were constructed. CDRAD contrast detail and Huttner resolution phantoms were placed in the middle of a patient phantom to simulate realistic clinical conditions. Starting points for each X-ray examination series were the kVp/mAs, filtration and SID values which had previously been used with the FS technique. Entrance doses to the phantoms were also measured. For each X-ray examination type, an image quality factor (IQF) was obtained. Each image series consisted of; the original FS image, and subsequent images acquired using CR/DR technology with exposure parameters/filtration/SID settings changed. Image optimization was achieved by determining the IQF as a function of entrance dose. Results: Dose reduction factors of between 1.3 and 4 have been found between the FS and CR/DR technology over a range of X-ray examination types. Conclusion: It is clear from our findings that, with CR/DR technology, there is potential for significant reduction in patient entrance dose with no negative impact on image quality. Radiation dose reduction by efficient 3D image restoration B. Svensson, M. Andersson, Ö. Smedby, H. Knutsson; Linköping/SE (orjan.smedby@imv.liu.se) The trade-off between radiation dose and image quality in CT motivates development of image restoration methods for maintaining image quality while reducing the radiation dose. This study concerns 3D restoration applied to low-dose CT images, using filter networks. Post-mortal volume CT images of the orbit were acquired at 80, 160 and 320 mAs. By post-processing, the volumes were restored by searching local 3D neighborhoods for planar structures. Although CT datasets are 3D, post-processing is usually carried out in 2D due to the computational load of 3D filtering. The load is however significantly reduced by applying smaller filters in sequence and exploiting similarities of the filters, i.e. using filter networks. Original and restored images acquired at 80 mAs and 160 mAs as well as original images with 320 mAs, were shown to 4 radiologists who assessed the subjective degree of noise and overall image quality. Results: A significant improvement of the noise score by the restoration was seen at 80 mAs (the mean score increased from 2.5 to 3.3) but not at 160 mAs (from 3.6 to 3.8). For general image quality, none of the datasets differed significantly from the 320 mAs reference. Conclusion: Filter networks make 3D post-processing feasible and seem to improve the subjective image quality with respect to noise in low-dose CT datasets. To further evaluate the dose reduction potential, larger studies are needed. Image repeat analysis for a digital radiography system A. Steffen, U. Neitzel, K. Förger; Hamburg/DE (ulrich.neitzel@philips.com) Purpose: Image reject/repeat analysis is an important quality control tool with increasing importance for digital radiography systems. We report on methods for and results of analysing repeat rates with clinically used digital radiography systems. Methods and Materials: Two different methods were applied to record and analyse the rejected and repeated images on digital flat-panel detector radiography systems in routine clinical use. Method I used the logfile entries generated for each X-ray examination. For method II additionally all images (even rejected) were stored on a separate storage device during the study period for later analysis. In total, method I was applied to 162,000 image acquisitions in 27 hospitals, located in different parts of the world; method II was applied to 13,500 image acquisitions in 3 hospitals located in Germany. The overall repeat rate determined with method I was 6.3 ± 3.1%, however the rate varied from 0.9 to 12.2% for individual hospitals. For the three hospitals analyzed according to method II the repeat rate ranged from 4.6 to 7.4%. The mean repeat rate per examination type reached from values below 2% for thorax to almost 25% for sacrum. The main reasons for repeating image acquisitions were positioning and collimation errors in 65-75% of the cases; technical problems or malfunctions accounted for about 5%. Conclusion: Examination log files can be used for repeat analysis in clinical routine. Repeat rates for the digital radiography system evaluated are mainly determined by positioning/collimation errors, whereas technical reasons form only a minor part. Is it time to re-introduce formal radiation protection teaching for all specialties? M. Nadarajah 1 , N. Jayasekera 2 ; 1 Manchester/UK, 2 London/UK (n4das@hotmail.com) Purpose: To assess level of knowledge, training and safety in the use of the image intensifier by orthopaedic trainees in England. A questionnaire survey was conducted among orthopaedic specialist registrars with operating responsibility for the image intensifier. Results: 75 trainees responded. 72% had received some tuition in ionising radiation protection. Only 15 correctly located X-rays on an energy spectrum depicting microwaves to gamma rays. 21% were unable to correctly identify the X-ray tube end of the c-arm on a picture of an image intensifier. 71% admitted to inadvertent screening of their hands at least once a month. 85% routinely held their periphery within 60 cm of the flare of the greater trochanter during screening. 16% routinely wore insufficient radiation protection. The original ionising radiation regulations made it compulsory for all staff working with X-ray equipment to receive formal instruction on the hazards of ionising radiation and the safe use of X-ray equipment. In 2000 with the advent of IRMER the radiation protection course was disbanded without alternate compulsory instruction. Therefore there is a lack of consensus regarding the level and content of knowledge in ionising radiation protection issues required by a trainee. The majority of orthopaedic trainees do not use the image intensifier safely. This may result in unjustified ionising radiation exposure of patients, assistants, other theatre staff and themselves to unnecessary ionising radiation. Instruction in ionising radiation protection should be reintroduced as a prerequisite for higher training in orthopaedics in order to stem this dangerous trend. Purpose: Based on the MED-directive, 97/43/Euratom, STUK has the responsibility to set the national Diagnostic Reference Levels (DRLs) for the most common radiological examinations in Finland. Paediatric patients deserve a special attention due to the higher radiation risk compared with adults. The purpose was to present a method which takes into account patient size when setting DRLs in paediatrics. G aminations and 10 -30 other conventional or fluoroscopy examinations. Different methods for taking the patient size into account were compared by statistical methods. Both dose quantities, entrance surface dose (ESD) and dose area product (DAP) were used. Results: The method established by the National Radiation Protection Board (UK) for setting DRLs was found to produce extra uncertainty in the procedure and was not easy to apply to Finnish practise. Patient doses correlated exponentially (R 2 = 0.86 -0.99) with both the projection thickness, which is measured directly from the patient, and the estimated cylindrical diameter (ECD) of the patient. Conclusion: For the specification of the size of patient, a projection thickness is more practical than the calculated ECD. Reference levels for paediatric radiological examinations could be given as a function of patient projection thickness, presented as a linear curve on half logarithmic scale. Purpose: To prospectively assess diagnostic accuracy of non contrast renal magnetic resonance angiography (MRA) using a navigator gated free breathing 3D SSFP sequence, compared to contrast-enhanced (CE) MRA. Methods and Materials: Thirty hypertensive patients (mean age 57 years) with suspected renal artery stenosis underwent navigator-gated 3D SSFP MRA during free breathing (acquired voxel 0.9x1.2x2.0 mm) at 1.5 T, followed by a high resolution (acquired voxel 1.1x1.3x2.5 mm), breath-hold CE-MRA (0.15 mmol/ kg Gd-BOPTA) using parallel imaging technique. Renal artery stenosis was graded on a 5 point scale. Maximal visible vessel length was measured for both techniques. Consensus reading of CE-MRA and DSA (n = 8) were used as standard of reference. Two blinded, independent observers evaluated non-contrast SSFP versus CE-MRA. Results: Eleven stenoses of renal arteries > 50% were diagnosed. Sensitivity of non-contrast SSFP MRA for detection of significant renal artery stenosis (> 50%) was 100% for both observers (CI 0.82-1.15 and 0.82-1.18). Specificity was 90% (CI 0.85-0.95) for reader 1 and 95 % (CI 0.91-1.0) for reader 2, and accuracy was 92% (CI 0.88-0.97) and 96% (CI 0.92-1.01), respectively. Maximal visible renal artery length was significantly greater in CE-MRA (66.7 ± 24.5 mm) compared to SSFP MRA (51.6 ± 21.9 mm) (p < 0.001). Conclusion: Compared to CE-MRA, free breathing navigator gated SSFP MR angiography has a high diagnostic accuracy for assessment of renal artery stenosis without the need for contrast media, therefore reducing the cost of MR imaging for screening of renovascular disease. However, severity of renal artery stenosis may be overestimated by SSFP MR angiography due to signal loss distal to the stenotic lesion. Feasibility of contrast-enhanced MR angiography during preoperative evaluation of donor kidney for transplantation J.J. Lee, Y. Kim, T. Kwon; Daegu/KR (jonglee@knu.ac.kr) Purpose: During renal transplantation, the feasibility of CE-MRA as a modality for preoperative donor kidney exam was evaluated. Methods and Materials: Prospectively, 69 kidney donors (mean age of 41) were examined using both CE-MRA and DSA. CE-MRA preceded DSA without delay. The interpretation of CE-MRA was blinded from DSA. In 55 subjects, hand assisted laparoscopic nephrectomy was performed. As check points, four facts of early branching of renal artery, accessory artery, accessory vein, and renal arterial stenosis were selected. The check points of CE-MRA were compared with DSA and surgery. Statistical analysis of CE-MRA based on either DSA or surgery was performed. Results: Based on surgery, sensitivity, specificity and accuracy of CE-MRA to depict "early branching of renal artery" were 50%, 98% and 95% (p < 0.05, k=0.54). The score for "accessory renal artery" was 80%, 96% and 95% (p < 0.001, k=0.70), respectively. The depiction of "accessory renal vein" using CE-MRA was statistically insignificant. Based on DSA, sensitivity, specificity and accuracy of CE-MRA to depict "early branching of renal artery" were 79%, 100% and 98% (p < 0.001, k=0.87). The depiction of "accessory renal artery" showed 97%, 100% and 99% (p < 0.001, k=0.98), respectively. The depiction of "accessory renal vein" was also statistically significant (100%, 99% and 99%, p < 0.05, k=0.49). In evaluation of renal arterial stenosis, Spearman' rho (r s ) between CE-MRA and DSA was 0.95 (p < 0.001). Purpose: Purpose of this study was the evaluation of a high-resolution MRA of the kidneys at 3.0 T in healthy volunteers and patients with suspected renal artery stenosis (RAS). Methods and Materials: 20 healthy volunteers and 12 consecutive patients suspected of having RAS (mean age 47.3 years) were scanned with a 3D-GRE Flashsequence (TR/TE 2.9/1.2 ms, flip 23°, matrix 576, voxel size 0.6x0.9x1.2 mm³, GRAPPA 2, acquisition time 22 sec). Using a phased-array surface coil on a 3.0 T whole body scanner (Siemens) a high-resolution MRA was implemented after injection of 0.1 mmol/kg BW gadodiamide. Images were assessed subjectively based on visibility and delineation of vessel wall and image quality (0-3 points). Quantitative evaluation was done by measuring the contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR). Results: Our technique was able to visualize renal arteries (RA) in all subjects (100%) with mean visibility score of 2.3 ± 0.3. RA were identified up to the second order branches in all subjects. In 4/12 patients a haemodynamic relevant narrowing of the RA was found and proven by conventional angiogram. No motion degradation of the RA manifest by blurring and reduction of arterial signal intensity (SI) were found. No significant difference concerning mean CNR and SNR were found (CNR abdominal aorta 45.4 ± 19.2, right RA 44.3 ± 16.9, left RA 44.4 ± 17.1). Conclusion: Using parallel imaging techniques renal artery MRA at 3.0 T is feasible and initial results are promising. Signal gain at 3.0 T imaging is an important factor to further increase spatial resolution and improves the visualization of small vessel segments. Purpose: The aim of this study was to assess the influence of the acquisition technique and field strength on the variability of semiquantitative renal perfusion parameters. Methods and Materials: 8 healthy volunteers were examined at 1.5 T (Siemens AVANTO) using a 2D-SR-turboFLASH sequence (voxel size 2.3x2.1x8 mm³, temporal resolution 1s/4 slices) for the assessment of renal perfusion after the bolus administration of 7 ml 0.5-molar Gd-chelates. At 3.0 T (Siemens TRIO), another 8 healthy volunteers were examined using the above 2D-SR-turboFLASH after the bolus administration of 7 ml 0.5-molar Gd-chelates. Another 8 healthy volunteers were examined at 3.0 T using TREAT (voxel size 1.7x1.2x5 mm³, temporal resolution 1.4s/3D data set) after bolus administration of 6 ml of 0.5-molar Gdchelates. Postprocessing based on a gamma-variate fit, yielded the renal functional parameters mean transit time (MTT), time to peak (TTP), maximal upslope (MUS) and maximal signal intensity (MSI). C D E F B A G Results: MTT and TTP did not show significant differences between the different techniques and field strengths. MSI and MUS were significantly (p = 0.002) higher with TREAT (range of means (ROM) 591 A.U. and 103.5 A.U./s) than with SR-turboFLASH at both field strengths (MSI ROM 400.5 A.U. -362 A.U. and MUS ROM 65.4 A.U./s -68.7 A.U./s). Image quality was rated better for the 3 T images. Conclusion: Semiquantitative renal perfusion measurements are feasible with TREAT at 3.0 T and turboFLASH techniques at 1.5 T and 3.0 T. MTT and TTP are independent of the technique and field strengths applied. Temporal sampling greater than 1.4 seconds does not seem to be required. Contrast enhanced magnetic resonance angiography of renal arteries and aorta: Comparison of peak signal intensity using 1.0M and 0.5M gadolinium chelates B.C. Anil Kumar 1 , T.A. Sudarshan 2 , W. Milne 2 , G. Houston 2 ; 1 Nottingham/UK, 2 Dundee/UK (anil.kumar@doctors.org.uk) To compare the effect of 1M Gadolinium chelate to that of three 0.5M chelates on peak signal intensity (PSI) in contrast enhanced magnetic resonance renal angiography (CE-MRA). H. Ittrich 1 , A.N. Priest 1 , P.G. Begemann 1 , C. Jahntz 1 , C. Weber 1 , B. Misselwitz 2 , G. Adam 1 ; 1 Hamburg/DE, 2 Berlin/DE (ittrich@uke.uni-hamburg.de) Purpose: To evaluate the potential of 3 T MR imaging for detection of high-risk atherosclerotic plaques after injection of different USPIO dosages. Methods and Materials: High resolution MR imaging of the aorta was performed on 15 WHHL (ages 27.2 ± 5.9 months) and 6 NZW rabbits (control, 22.7 ± 3.2 months) before and 24-120h after USPIO administration (DDM 43/34,Schering) in a 3 T scanner (Philips Intera) using an 8-channel head coil. 3 groups with dosages of 1.0, 0.25 and 0.1 Fe/kg were used (each 5 WHHL/2 NZW). T2*w3D GRE (TR/TE=30/2.7-11.0 ms, FA20°, res.0.35×0.35×3 mm) and 3D MRA (TR/ TE 5.6/2.1 ms, FA25°, res.0.7×0.7×0.7 mm, USPIO/Gd-DTPA-enhanced) were used to investigate areas of signal intensity (SI) reduction within the aortic wall demonstrating the uptake of USPIO. MR Images were matched with histology (HE, Prussian blue, RAM11). Results: Areas of SI reduction within the aortic wall were detected after injection of 1.0 mmol/kg (4/5 animals) correlated with USPIO uptake in inflammatory plaques (double-immunostaining). No significant SI reduction was detected after injection of 0.25 and 0.1 mmol/kg in WHHL animals with negligible USPIO uptake by macrophages despite the high burden of atherosclerotic plaques. NZW animals showed no aortic SI changes corresponding to the absence of plaques. Increasing TE (2.7-11.0 ms) decreased the detection threshold for aortic susceptibility artefacts, but increased artificial image distortion from outside the aorta (USPIO-uptake in bone marrow, lymph nodes). Conclusion: Contrast enhanced MR imaging techniques at 3 T using experimental dosages of USPIO (1.0 mmol/kg) allow identification of vulnerable atherosclerotic plaques. The absence of aortic SI reduction after lower dose USPIO administration does not correspond to an absence of plaque formation ((0.25),0.1 mmol/kg). Molecular imaging of atherosclerosis-associated matrix metalloproteinases (MMPs) using a novel and specific MR imaging contrast agent -P947 S. Amirbekian 1 , V. Amirbekian 1 , J.S. Aguinaldo 1 , F. Hyafil 1 , E. Vucic 1 , E. Lancelot 2 , C. Corot 2 , Z.A. Fayad 1 ; 1 New York, NY/US, 2 Paris/FR (vardan@jhmi.edu) Purpose: Matrix Metalloproteinases (MMPs) play an important role in atherosclerotic plaque progression, rupture, and thrombosis. Targeting MMPs with molecular MR imaging Probes may facilitate detection of atherosclerosis and assessment of plaque stability. P947 is a short peptide ligand for MMPs. P947 has a Gadolinium attached to it turning it into an MR imaging contrast agent. The object of the study is to test the ability of P947 to detect and quantify atherosclerosis. Methods and Materials: 15 month-old Apolipoprotein-E-Knockout mice (n = 10) underwent in vivo MR imaging of the abdominal aorta using a 9.4T MR system. Pre-contrast enhanced (CE) and post-CE MR imaging was performed at 1, 2, 3, and 24 hrs post-injection using a T1W black blood sequence. P947 (100 µmol/ kg) was injected via the tail vein. For control, a group ApoE-KO mice (n = 4) were injected with Gd-DOTA (Dotarem-standard Gadolinium agent) using an equivalent dose of Gd (100 µmol/kg). After MR imaging, the aortas were isolated and fixed; pathology, MMP zymography and immunohistochemistry was performed for MMPs. MRI images of the matched (pre and post) slices were used for analysis. In comparison to controls, in ApoE-KO mice that were injected with P947 there was heterogeneous enhancement seen on MR imaging, with significant increase in contrast-to-noise ratio (CNR) for wall/lumen&wall/muscle in the post-CE images (1, 2 & 3 hrs post-injection) with no significant increase at 24 hrs vs. pre-CE images [ANOVA, p < 0.05]. The ratio of the post to pre-contrast signal intensity of the wall (normalized to muscle) with P947 was 1.92 ± 0.19 (91% enhancement) in KO group at 1 hr, 1.51 ± 0.14 (52%) at 2 hrs, 1.23 ± 0.13 (22%) at 3 hrs, and 1.13 ± 0.06 (14%) at 24 hrs. Conclusion: Noninvasively, P947 provided good contrast enhancement of aortic atherosclerotic plaque in ApoE-KO mice. P947 offers potential for novel noninvasive detection of atherosclerosis. Comprehensive diabetes imaging with whole body MR imaging at 1. Purpose: To implement and evaluate an MR protocol for comprehensive diabetes imaging on a 1.5 and 3.0 T whole body scanner using iPAT. Methods and Materials: 30 patients with type 1 or 2 diabetes (> 10 years) were examined on a 1.5 T (12 pts) or 3.0 T (18 pts) whole body MR system (Magnetom Avanto/Trio, Siemens Medical Solutions). Imaging of the brain with T2-w-, diffusion weighted and FLAIR-images was performed. Cardiac function was assessed with a single breath-hold multi-slice real time trueFISP technique (TSENSE factor 2-4). For DCE imaging, segmented trueFISP PSIR images were performed. 3D-Gd-MR-Angiography of the carotids, abdominal aorta (iPAT factor 3), thighs, calves and pedal arteries (iPAT factor 2) was obtained. Images of the feet were acquired. Correlation with a group of 200 healthy adults who received whole body MR imaging for cardiovascular screening was performed. Results and Discussion: Vascular pathologies were substantially more often found in diabetics than in healthy controls. Carotid artery stenosis was observed in 21% (6% of healthy adults). Renal artery stenosis was seen in 11 % of diabetics (controls 0.25%). Stenoses of peripheral arteries were seen in 47% -58% (healthy controls 14%). In four diabetes patients (21%) silent myocardial infarctions were detected (control group 1%). Chronic ischemic brain lesions were more often found in diabetics. Osteomyelitis was detected in 1 patient, neuropathic foot found in 3 patients. Conclusion: Whole body MR imaging is a highly promising method for comprehensive disease specific imaging in diabetes since typical pathologies can be identified and treated earlier. Purpose: To estimate the dose savings which could result from removing the anti-scatter grid in full-field digital mammography, without degrading image quality, as a preliminary step for a clinical experiment. Methods and Materials: Contrast-to-noise ratio (CNR) provided by 100 mm square aluminium foils, 0.1 and 0.2 mm thick, positioned on a PMMA phantom, was measured for PMMA thickness in the range 20-70 mm using a clinical digital mammography unit. For each thickness, one image was acquired in automatic mode with the grid present and the Average Glandular Dose determined; then six images were acquired without grid in manual mode, retaining the same technique factors and varying the mAs. A flat-fielding correction was applied to the grid-less images to correct the non-uniformity of the scattered radiation. For each thickness, the mAs number providing the same CNR as with the grid was computed. Using a distribution of PMMA-equivalent breast thickness measured previously, the average dose to the population was computed in both standard and grid-less cases. Results: For phantom thickness between 20 and 40 mm, the mAs were reduced by 30% to 15%, and by less than 10% between 40 and 50 mm. The breakeven occurred around 60 mm. The dose to the population at constant CNR would be reduced by 6%. Conclusion: Dose reduction achievable by removing the anti-scatter grid at constant CNR is small for most of the phantom thickness range; the resulting dose reduction to population does not justify a clinical application. Purpose: Tomosynthesis, through the reduction of structure noise caused by overlying breast tissue, may improve mammographic sensitivity and specificity. In particular, tomosynthesis provides improved contrast and conspicuity of masses. We have studied the visibility of low contrast objects using tomosynthesis images acquired at reduced radiation doses compared to full field digital mammography (FFDM). Methods and Materials: Mass visibility as a function of radiation dose was studied using phantoms. The mammography contrast detail (CD-MAM) and other phantoms were imaged in the presence of complex backgrounds and breast tissue. FFDM of the objects was performed at conventional doses and tomosynthesis images were taken at varying doses ranging down to 45% of the FFDM exposures. Contrast detail and object visibility performance for the FFDM and tomosynthesis images were compared. The prototype tomosynthesis system used an amorphous-selenium image receptor, and the FFDM system was the commercially available Hologic Selenia. Both image receptors have an imaging area of 24 x 29 cm, and a native resolution of 70 microns. Results: Object detection for the tomosynthesis images at 45% dose was statistically superior to the equivalent detection task for the FFDM images at 100% dose. Tomosynthesis detection performance was increasingly superior as dose increased to match the FFDM dose. Conclusion: Tomosynthesis is superior to 2-D mammography in the detection of low contrast objects in the presence of structure noise, even at lower dose. Optimal mammography performance may indicate the use of tomosynthesis. These results are being used to design our clinical trials. Phantom experiments with a digital breast tomosynthesis prototype R. Schulz-Wendtland, E. Wenkel, M. Lell, C. Böhner, W. Bautz, T. Mertelmeier; Erlangen/DE (ruediger.schulz-wendtland@idr.imed.uni-erlangen.de) Purpose: To compare the image quality of digital breast tomosynthesis prototype with digital projection images employing tungsten and molybdenum X-ray spectra. Methods and Materials: The research tomosynthesis system is based on a fullfield digital mammography system (NovationDR, Siemens) modified for a wide angular range for tomosynthetic image acquisition. The large-area a-Se detector enables fast readout and can be switched between 2-D and tomosynthesis mode. Images of a mammographic random phantom (RMI 152 A) were acquired in both 2-D and tomosynthesis mode. The phantom structures simulate masses, calcification, and fibres. From the measured projection data slice images were reconstructed with filtered backprojection. The slice images were assessed and compared to the 2-D images in a observer study with five observers. For the used X-ray spectra, the dose was estimated with Dance's method. The MTF, NNPS and DQE were determined as a function of spatial frequency and detector exposure. The measurements were performed using the available W/Al anode/filter combination at various tube voltages with PMMA or Aluminium added in the X-ray beam. Results: The system was found to have a relatively constant DQE > 0.60 at zero spatial frequency for detector exposures ranging from 15 to 150 µGy with a halfvalue layer in the 0.70-1.1 mm Al range. The 10% MTF was 10 mm -1 in the scanning direction and 12 mm -1 in the slit direction measured 60 mm from the chest wall edge and 45 mm above the breast support table, which resulted in a lower DQE in the scanning than the slit direction for spatial frequencies above zero. Conclusion: The system has a wide dynamic range (15 -150 µGy) in which the DQE at zero spatial frequency is above 0.60. An image compression algorithm based on a combination of discrete cosine (DCT) and wavelet transforms has been assessed at different compression ratios. Compressed images from lossless to 120:1 were subjected to blind analysis by a group of experienced readers who evaluated and ranked the images in order of quality. Results: Visually lossless format levels of up to 80:1 compression were judged as indistinguishable from the original; higher ratios such as 120:1 proved identifiable although pixilation became apparent when the image was subjected to x3 magnification. Microcalcification and architectural changes were among the features highlighted and compression at 65:1 was suggested as a suitable limit for transmission to other centres. Reduction in interval cancers could follow wider dissemination of difficult cases which provide up to 40% of those missed cancers. The same algorithm has also been applied to European Quality Control test phantoms (CDMAM version 3.4, Nijmegan University) with results indicating satisfactory ratios of 40/60:1 for these very exacting analyses. Conclusion: Mammographic images can be safely compressed to take up only 1/65 of the original storage space/bandwidth. The resulting reduction in terms of transferring the images over electronic media proves useful in online reporting applications. Future work will investigate all the factors that influence the visual/imaging pathway to form conclusions on a variety of influences, from electronic to ambient lighting and relevant to the decision pathway by reporting radiologist. Computer-aided detection in digital mammography: Comparison of craniocaudal, mediolateral oblique and mediolateral views W. Moon, S. Kim, N. Cho; Seoul/KR (moonwk@radcom.snu.ac.kr) Purpose: To compare the sensitivity of a CAD system for the detection of breast cancer in three digital mammographic views. A commercially available CAD system was applied to the craniocaudal (CC), mediolateral oblique (MLO), and mediolateral (ML) digital mammographic views of 83 women with 83 histologically proven breast cancers. Findings were 59 masses and 41 microcalcifications (17 lesions showed both findings; 42 lesions, mass only; and 24 lesions, microcalcification only). The sensitivities of the CAD system in the three views were 92% (76 of 83) for CC, 83% (69 of 83) for MLO, and 86% (71 of 83) in ML, which were not statistically significant (P =.07). Its sensitivity increased to 96% (80 of 83) in CC plus MLO views, and to 99% (82 of 83) in CC, MLO plus ML view. For masses, the sensitivity of the CAD system was 76% (45 of 59) in CC view and 75% (44 of 59) in MLO view and this increased to 93% (55 of 59) when in combination with MLO and CC views (P =.0003). For microcalcifications, sensitivity was 98% (40 of 41) in CC view and 95% (39 of 41) in MLO view, and this increased to 100% (41 of 41) when in combination with MLO and CC views (P =.31). The sensitivities of the CAD system were not significantly different in these three digital mammographic views. The sensitivity for masses was increased when the craniocaudal view was added to the mediolateral oblique view. Computer-aided detection (CAD) of breast cancer in screening mammography: Early signs of breast cancer detected on prior mammograms Z. Huo 1 , J. Lewin 2 , W. Menhardt 1 ; 1 Rochester, NY/US, 2 Denver, CO/US (zhimin.huo@kodak.com) Purpose: To recognize early signs of breast cancer detected by CAD and to learn to work effectively with CAD to improve the efficacy of screening mammography. While clinical use of CAD has shown to be helpful, greater benefit can be achieved if radiologists could recognize more early cancers that had been detected by CAD. We retrospectively identified 47 visible prior cancers (28 masses, 17 Ca ++ , and 2 with both) on mammograms that were taken an average of 15 months before examinations with biopsy was confirmed cancers. These priors were randomly mixed with 150 normal and 29 positive cancer examinations. Four readers, in two sessions, were asked to mark suspicious areas and provide a BIRADS rating for each breast examination with and without CAD. Results: The CAD system correctly marked 30 (15 masses, 13 Ca ++ , and 2 with both) of the 47 prior cancers. Of the 120 readings (30 prior cancers read by four readers), 38% were not considered suspicious enough for biopsy and were simply dismissed when the CAD correctly prompted the locations. Of the 30 priors correctly marked by CAD, 29 were considered suspicious enough for biopsy, by at least one reader, when the CAD correctly prompted the locations. Conclusion: It is important to recognize early signs of breast cancer detected by a CAD system to improve the benefits of mammography screening. In the presentation, we review the characteristics of the prior cancers and the cancers that were not detected, even when the cancer locations were correctly prompted. The use of computer aided classification (CAC) in BI-RADS category 4 A lesions F. Sperber 1 , R. Lederman 2 , S. Fields 2 , S. Buchbinder 3 , P. Bamberger 2 , I. Leichter 2 ; 1 Tel Aviv/IL, 2 Jerusalem/IL, 3 Staten Island, NY/US (isaac.leichter@siemens.com) Purpose: To explore the performance of a CAD device with computer aided classification (CAC) capabilities, on mammographic lesions that were assigned BI-RADS category 4 A by conventional interpretation. Methods and Materials: BI-RADS category 4 A is used for findings requiring biopsy, but with a low suspicion of malignancy. A CAC scheme was used to retrospectively analyze 76 cases forwarded for stereotactic biopsy (6 masses, 70 clusters), which were prospectively assigned BI-RADS Category 4 A. Of these cases, 5 masses and 64 clusters proved to be benign at pathology. The CAC scheme automatically extracted from the digitized mammograms, quantitative features that characterize the lesions. Based on these features, the system used a classification scheme to score the lesions by their likelihood of malignancy, and stratified the lesions according to their score into low, intermediate and high levels of suspicion. Results: For the 7 malignant cases included in the study, the computer assigned a high level of suspicion to all lesions and of the 69 (5 masses, 64 clusters) benign cases the computer assigned a low level of suspicion to 17 cases (3 masses, 14 clusters). The CAD analysis, thus, correctly classified all malignant cases, yielding a sensitivity of 100%. The computerized analysis could correctly classify 17 benign cases, improving the specificity for BI-RADS Category 4 A lesions by 24.6%. Conclusion: In a case series consisting predominantly of clusters of microcalcifications, the use of computerized classification could have correctly characterized all the malignant lesions assigned BI-RADS Category 4 A, and could possibly eliminate 24.6% of unnecessary biopsies. Do CAD-systems pass the official mammography tests required for screening and do they substantially support the radiologists to pass these tests? A. Malich 1 , A. Beier 1 , P. Bank 2 ; 1 Nordhausen/DE, 2 Jena/DE (ansgar.malich@shk-ndh.de) The role of CAD-systems as primer diagnostic tools and as second reader is discussed controversially. This study evaluated whether CAD-systems are able to pass official tests offered to radiologists (using similar criteria) and whether CAD-results assists the radiologists in a significant way to pass the tests. C D E F B A G An official test collection including 200 images (both sides; two views; 50 patients) was scanned (Second look, v.6.0, iCAD, USA) and analyzed (each side separately) blinded to the histological results by 3 radiologists (two experienced, one assistant radiologist) with and without CAD-printouts. Test-collection included (officially given number to all participating radiologists) 20-27 malignancies, at least one bilateral cancer and a few benign lesions but no BI-RADs3 cases. Test was failed in case of either > 2 false negative or > 5 false positive interpretations. Results: CAD-system highlighted 26/27 cancers and not marked (correctly) 14/ 73 unsuspicious mammograms (Sen: 96.3%, test passed; Spec: 19.2%, test failed). Both experienced radiologists passed the test with and without CAD, the assistant doctor failed twice. In mean, sensitivity increased by 1%, specificity increased by 0.6% with CAD. Per image CAD revealed 0.43 and 0.44 FP-markers (MA/ MC). Conclusion: Due to the overall high number of false positives, CAD-system failed the official mammographic test despite the highly sensitive detection of malignancies. A significant decrease of FP-markers for masses using version 6.0 was verified. The effect of CAD on the analyzing radiologist, however, did not influence the test results substantially. Purpose: To assess the diagnostic accuracy of arthro-MR imaging versus arthroscopy in 13 athletes with clinical suspicion of non-outlet impingement. We enrolled 13 patients (19-34 years, mean age 28), all athletes. All patients presented with clinical evidence of non-outlet impingement of the shoulder. Before surgery, all patients underwent arthro-MR imaging and arthroscopy. We acquired SE T1-w and FSE T2-w sequences on standard examination and SE T1-w and fat saturated SE T1-w after intra-articular injection of 10-15 ml of contrast medium (dilution of 0.6 ml of Gd-DTPA in 250 ml of saline). We compared arthro-MR imaging with arthroscopic findings. Results: In 11 patients (84.6%) MR imaging showed shoulder abnormalities. Abnormalities were: 5 type-II SLAP lesions, 2 posterosuperior impingements, 2 supraequatorial capsular enlargement, 1 SGHL, 1 Buford complex. Two patients had non-pathologic findings. Arthroscopy confirmed MR imaging findings in 10 cases (76.9%). In one patient with MR imaging findings suggestive for type-II SLAP lesion, arthroscopy showed a type-III bicipital anchor associated to sublabral hole. At arthroscopy the two patients with normal MR imaging had posttraumatic enlargement of the rotator interval. Conclusion: Arthro-MR imaging appears to be the method of choice in the preoperative study of articular lesions associated with non-outlet impingement in athletes. Rotator cuff tears: Assessment with MR arthrography in 250 patients with arthroscopic correlation S. Waldt, K. Woertler, M. Bruegel, A. Burkart, E.J. Rummeny; Munich/DE (waldt@roe.med.tu-muenchen.de) Purpose: To assess the diagnostic accuracy of MR arthrography in the diagnosis of rotator cuff tears of the shoulder. Methods and Materials: MR arthrograms obtained in 250 patients including a study group of 120 patients with arthroscopically proved rotator cuff tears and a control group of 130 patients with intact rotator cuff tendons were reviewed in random order. MR imaging was performed on a 1.0 T system (Magnetom Expert, Siemens). MR arthrograms were analyzed by two radiologists in consensus for partial-and full-thickness tears of the supraspinatus, infraspinatus and subscapularis tendons. Results: At arthroscopy 189 rotator cuff tears were diagnosed, including 67 partial-(58 supraspinatus and 9 infraspinatus) and 122 full-thickness (62 supraspinatus, 36 infraspinatus, 24 subscapularis) tendon tears. For full-thickness tears the sensitivity, specificity, and accuracy were 92%, 98%, and 97% respectively, and for partial tears 73%, 97%, and 88% respectively. False negative assessments in the diagnosis of partial-thickness tears were predominantly (85%) observed with very small articular-sided tendon tears. Conclusion: MR arthrography is highly accurate in the diagnosis of full-thickness rotator cuff tendon tears, but is less accurate in the diagnosis of partialthickness tears, mainly due to limitations in the differentiation between very small articular-sided partial-thickness tears and frayed or friable tendon margins and/ or synovitic changes at the tendon margin. Purpose: The purpose of this study was to evaluate the association of rotator cuff abnormalities on the acromiohumeral distance (AHD). The AHD was measured on conventional radiographs and MR images. Three age and sex matched patient groups each including 21 patients were stratified by the AHD on conventional radiographs (group I ≤ 7 mm, group II 8-10 mm, group III > 10 mm). The AHD was related to the presence, location and size of a rotator cuff tear, and fatty degeneration of the muscle assessed on MR arthrography. The relative influence of the different MR arthrography findings on the AHD was assessed. Spearman rank correlation and stepwise regression was used for statistical analysis. Results: In group I (AHD ≤ 7 mm) full thickness tear in supraspinatus tendon was present in 90% (19/21), in infraspinatus tendon in 66% (14/21), and in subscapularis tendon in 42% (9/21), respectively. The size of rotator cuff tendon tears and the degree of fatty degeneration in all rotator cuff muscles showed a significant negative correlation with the AHD (P < 0.05). After stepwise regression a significant relative influence on the AHD remained for the size of rotator cuff tear (P < 0.0001) and for the degree of fatty degeneration of the infraspinatus muscle (P = 0.013). Conclusion: Tendon tears and fatty muscle degeneration of the rotator cuff correlate with a reduced AHD. The size of the rotator cuff tear and fatty degeneration of the infraspinatus muscle have the most pronounced influence on AHD. Correlation between size of rotator cuff tear and degree of tendon retraction A.M. Venkatesan 1 , B. Fritz 2 , A. Kassarjian 1 , W.E. Palmer 1 ; 1 Boston, MA/US, 2 Duesseldorf/DE (ben@fritz.md) Purpose: To determine the relationship between anteroposterior extent of rotator cuff tear (RCT) and measurement of tendon retraction, and to test if tendon delamination causes a disproportionate degree of retraction of bursal or articular cuff fibers. In 59 patients (32 male, 27 female, ages 18-79) with full-thickness RCTs at surgery, MR images were retrospectively reviewed to measure the anteroposterior length of tendon tear. Two reviewers reached consensus. After 4 months the cuff retraction from the osseous attachment site to the proximal tendon margin was measured. Whenever tendon delamination resulted in different degrees of bursal or articular retraction, the more retracted margin was measured. MR images were acquired at 1.5 T using a dedicated shoulder coil and routine protocol. IRB approval was obtained. Results: Anteroposterior length of RCT ranged from 2.0-5.0 cm (mean = 2.8 cm), and retraction of tendon margin ranged from 3.0-5.5 cm (mean = 3.2 cm). The anteroposterior length of tear showed a statistically significant positive association with tendon retraction (P <.0001). There were 49 tendons without cuff delamination and 10 tendons with surgically confirmed delamination. Comparing the ratio of RCT length to retraction, tendons without delamination (1.2:1) showed significantly less retraction (P=.04) than tendons with delamination (1:1.1). Conclusion: Anteroposterior length and tendon retraction in full-thickness RCTs correlate closely approximating a 1:1 relationship. An anteroposterior length less than the greatest amount of tendon retraction (ratio less than 1:1) indicates possible cuff delamination of the supraspinatus tendon. Cuff delamination is important to recognize because it is associated with poorer surgical outcome. . Each medical record for labral status was reviewed retrospectively and classified into three grades (A: normal, B1: tear, B2: non-tear abnormalities such as degeneration, fraying, cracking, concealed avulsion, etc). Musculoskeletal radiologists retrospectively reviewed the MRA and recorded the presence of labral lesions, regardless of differentiation between B1 and B2, according to the location. Results: Arthroscopy revealed 146 B1 lesions (37 superior, 90 anterior, 19 posterior) and 78 B2 lesions (31 superior, 11 anterior, 36 posterior). Relative incidence of B2 lesion in the posterior labrum was significantly higher than in the of superior (p = 0.001) or anterior (p = 0.012) labrum. Sensitivity, specificity, and accuracy according to the location were 54%/90%/79% for superior, 80%/93%/ 87% for anterior, and 44%/90%/79% for posterior labrum. The sensitivities for B1 and B2 lesions according to location were 77%/27% for superior, 84%/41% for anterior, and 61%/35% for posterior labrum. Sensitivity for B1 lesion of the posterior labrum was significantly lower than that of the superior (p = 0.2384) and anterior (p = 0.003) labrum. The diagnostic values of MRA for posterior labral lesions were relatively poor compared to the anterior labrum, which may be caused by a higher proportion of non-tear abnormalities and location of the posterior labrum itself. Purpose: To present the optimal concentration of iodinated contrast for CT arthrography of the shoulder comparing four different concentrations of iodinated contrast. Methods and Materials: Forty consecutive CT arthrograms of the shoulder were performed were randomly assigned to either 25% (n = 10), 50% (n = 10), 75% (n = 10), or 100% (n = 10) iodinated contrast. CT arthrograms were assessed quantitatively (CT number ratio of intraarticular fluid to pectoralis major muscle) and qualitatively (visual grading scale of 1-5 for overall image quality, image contrast, degree of joint distension, and beam hardening artifacts). Results: The CT number ratio (mean ± SD) of intraarticular fluid to pectoralis major muscle was 18.9 ± 3.2 for 25% iodinated contrast, 28.6 ± 2.2 for 50% iodinated contrast, 37.9 ± 6.1 for 75% iodinated contrast, and 41.0 ± 5.7 for 100% iodinated contrast. CT arthrograms with 75% and 100% groups showed significantly higher CT number ratio than other groups (p < 0.05, one-way ANOVA). In qualitative assessment, 50% iodinated contrast was significantly better than other concentrations (p < 0.05, one-way ANOVA). Conclusion: CT arthrograms of the shoulder obtained with more than 75% iodinated contrast provided higher CT number ratio. However, we preferred the image quality of CT arthrograms with 50% iodinated contrast. The horizontal component of partial thickness tears was detected more frequently on ABER CT arthrography; eight were visible only in the ABER position. The type of lesion was: type A= 2, type B= 6, type C= 4 and type D= 3. Full thickness tears were diagnosed in neutral and ABER position. ABER images allowed the visualisation of unstretched tendon floating inside the articulation. Conclusion: ABER CT arthrography allows a better detection and depiction of horizontal partial tears. CT arthrography with ABER and neutral position is easily realized without loss of time during the same examination. Results: Seven patients demonstrated labral tears on MR imaging and this was confirmed surgically in all cases. Thirty-seven patients (97%) had labral-chondral transitional zone lesions on MR imaging compared with 38 surgically confirmed cases. The site of labral-chondral transitional zone pathology at arthroscopy was as follows: 50% anterosuperior, 36% anterosuperior/superolateral, 11% superolateral and 3% superolateral/posterosuperior. The site was identified correctly in between 92% and 95% of cases on MR imaging. Acetabular cartilage abnormality was surgically identified in 39% and femoral cartilage lesions were found in 20% of cases. The acetabular lesions were correctly identified in between 89% to 94% and femoral lesions in 100% of cases. Conclusion: Non-contrast MR imaging provides an accurate method of assessment of patients with a clinical suspicion of femoroacetabular impingement. It is non invasive and provides useful preoperative information regarding the presence and anatomical site of labral and cartilage abnormality. MR imaging findings of acetabular dysplasia S.L.J. James 1 , M. Miocevic 2 , F. Malara 2 , J. Pike 2 , D. Young 2 , D. Connell 1 ; 1 London/UK, 2 Melbourne/AU (jamesslj@email.com) Purpose: To evaluate the diagnostic accuracy of MR imaging in the identification of labral and articular cartilage lesions in patients with acetabular dysplasia. Methods and Materials: Pre-operative MR imaging was performed in 25 consecutive patients (27 hips, 16 males, 9 females, age range 19-52 years, mean age 31.2 years) with radiographic evidence of acetabular dysplasia (centre edge angle of Wiberg < 20 degrees). The average duration of symptoms was 16.2 months. Two musculoskeletal radiologists assessed MR images for the presence of abnormality involving the acetabular labrum and adjacent acetabular articular cartilage. A high resolution technique was used to assess the labrum and labral chondral transitional zone. Surgical correlation was obtained in all cases by a single surgeon experienced in hip arthroscopy. The acetabular labra in the dysplastic hips demonstrated abnormal signal intensity and had an elongated appearance when compared with the control group (mean length 10.9 mm v 6.4 mm). Morphological appearances in the labra included surface irregularity, fissures and cleft formation. MR imaging correctly identified the severity of chondral abnormality in 24 of 27 hips (89%) when compared with arthroscopic findings. Conclusion: MR imaging demonstrates an elongated labrum, focal intra-substance signal change and irregularity and fissuring of the margins in patients with acetabular dysplasia. Abnormality is also identified at the labral chondral transitional zone where fissuring, focal clefts, chondral deficiency and subchondral cyst formation may be apparent. Purpose: Labral lesions in patients with hip dysplasia could indicate biomechanical decompensation of the hip joint and often cause early clinical symptoms. Our objective was to assess the condition of the acetabular labrum and its association with corresponding cartilage quality based on MR imaging. Methods and Materials:15 patients in an age range from 18 to 37 years with clinical (groin pain, impingement) and radiological (CE-angle) signs of hip dysplasia but no signs of moderate or severe osteoarthritis underwent MR imaging. As a control group we assessed MR imaging of 15 age and sex matched healthy volunteers. For MR imaging a 3D-Dess sequence (resolution 0.6x0.6x0.6 mm, Siemens Trio) and a radial sectioned PD-TSE sequence perpendicular to the acetabular labrum were performed. For double-oblique planning of the 12 radial planes through the acetabular opening, isotropic images were used. The acetabular labrum morphology was compared to the corresponding femoral and acetabular cartilage quality in the radial portions. The acetabular labrum was larger and the acetabular cartilage thicker in dysplastic hips. All patients had portions with signal intensity changes within the labrum. In 12 patients MR imaging suggested degeneration or a tear of the labrum or both. Labrum lesions and corresponding cartilage inhomogeneity were observed mainly in the anterior-superior sections. Conclusion: It was possible to compare cartilage morphology to acetabular abnormalities by using an isotropic cartilage sequence for positioning of the acetabular opening and the radial section. The presented findings are important for early detection and accurate diagnosis, and may impact optimal treatment planning and prognosis. Follow-up examinations in these young patients can help to clarify the natural course of labral disorders. ) was used to compare the confidence of detection and characterization for the reference standard. The images were compared quantitatively by calculating the liver signal-to-noise ratio (SNR) and lesion-to-liver contrast-to-noise ratio (CNR). Results: The PACE-TSE had significantly higher Az values in lesion detection (P < 0.05 for two reviewers) and lesion characterization (P < 0.05 for two reviewers) compared with PACE-HASTE. For two reviewers, the Az values in lesion detection and characterization of PACE-TSE were superior to those of RT-TSE, but there was no significant difference (P > 0.05). Mean SNR for liver and CNR for hepatic lesions were highest on PACE-HASTE than those of other sequences. Mean acquisition times of the PACE-TSE, PACE-HASTE, and RT-TSE sequences were 3.18 minutes, 3.17 minutes, and 6.17 minutes, respectively. The PACE-TSE sequence could provide the highest tumor detection and characterization, although the mean lesion-to-liver CNR was inferior to those of PACE-HASTE and RT-TSE. Value of contrast-enhanced CT in combined PET/CT protocols for oncological and non-oncological imaging C. Pfannenberg, S.M. Eschmann, P. Aschoff, K. Brechtel, R. Bares, C.D. Claussen; Tuebingen/DE (christina.pfannenberg@med.uni-tuebingen.de) To determine the value of contrast-enhanced CT in comparison to lowdose non-contrast CT in combined PET/CT protocols for oncological and nononcological imaging. Methods and Materials: PET/CT scans of 154 consecutive patients, 138 patients with malignancy and 16 patients with vasculitis or infection were evaluated retrospectively. PET/CT was performed with a high resolution LSO detector and 16-row CT. The CT protocol varied depending on the specific clinical question and was multi-phasic in 101 patients and single-phasic in 53 patients. PET/CT studies were analyzed by different categories to determine the value of contrastenhanced CT on the PET/CT interpretation. The results were validated either by histopathology or clinico-radiological follow-up. The clinical impact of contrastenhanced PET/CT was evaluated with respect to changes in patient management. Results: Diagnostic CT was of additional value in 93/154 patients altering the PET/CT interpretation, in the remaining 51 patient no additional benefit was derived from the diagnostic CT. 10 patients were excluded due to inconclusive diagnosis in both methods. The greatest benefit of diagnostic CT was found in the following categories: improved localization of pathologic FDG uptake; improved local tumor staging; and detection of additional relevant findings. PET/CT interpretation was altered in 25% of patients in each of these groups. The benefit of diagnostic CT was influenced by the tumor type and changed clinical management in 35% of these patients. Conclusion: Diagnostic CT as part of the combined PET/CT examination provides additional value in specific clinical conditions with resultant change of therapy in a substantial proportion of patients. Monda Monday y y y y lesion per patient. When more than one lesion was seen, the performing radiologist chose the lesion with the best possible acoustic window. Ultrasound was performed continuously (5 minutes) at a low mechanical index after a 2.4-4.8 ml bolus injection with microbubbles containing sulphurhexafluroride. CT exams (4 phases) with injection of iodixanol 270 mg/mL and MR exams with hepatocellular specific 0.01 mmol/mL mangafodipir. The reference standard was defined as either pathology or a combination of clinical follow-up of at least 6 months, comprising any imaging and clinical features. Results: Sensitivity for a malignant versus benign lesion was 97% for CEUS (31/ 32), 94% for MDCT (30/32) and 72% for MR imaging (23/32); specificities were 70% (23/33), 75% (24/33) and 91% (30/33), respectively. Accuracy with respect to lesion type was highest for MR imaging (75%; 49/65), CT correctly characterised 48 (74%) of all lesions, whereas CEUS had an accuracy of 68% (44/65). Conclusion: CEUS and MDCT are equally diagnostic with respect to ruling out a lesion, whereas MR imaging is more specific for ruling in a malignant lesion. However, CEUS does not perform as well as CT and MR imaging in specific lesion type characterisation. Study of a diffusion-weighted single-shot spin echo echo planar imaging sequence using parallel imaging in the detection of focal liver lesions: Comparison with a T2-weighted turbo spin echo technique K. Coenegrachts 1 , J. Delanote 1 , L. ter Beek 2 , M. Haspeslagh 1 , F. Van Kerkhove 1 , L. Steyaert 1 , H. Rigauts 1 , J.W. Casselman 1 ; 1 Bruges/BE, 2 Best/NL (kenneth.coenegrachts@azbrugge.be) To compare diffusion-weighted Single-Shot Spin Echo Echo Planar Imaging (SS SE-EPI DWI) with T 2 -weighted Turbo Spin Echo (T2w TSE) in the detection of focal liver lesions, with focus on detection of small (< 10 mm) lesions. Twenty-four consecutive patients with suspected malignant liver lesions underwent routine MR imaging examinations. A respiratorytriggered T2w TSE sequence and a respiratory-triggered SS SE-EPI DWI with b-values of 0, 20, 300 and 800 s/mm 2 were compared quantitatively by doubleblind measurements and calculation of the lesion-to-liver Contrast-to-Noise Ratio (CNR). All sequences were compared qualitatively by grading image quality, lesion conspicuity, and artifacts. Rank order analysis (RIDIT-analysis) and nonparametric tests were used to compare the ordinal data. The lesion-to-liver CNRs were highest on T2w TSE when evaluating biliary cysts and on SS SE-EPI DWI with b=20 s/mm 2 when evaluating hemangiomas and metastatic lesions (p < 0.05). Concerning qualitative lesion detectability, biliary cysts were best detected with T2w TSE and hemangiomas and metastases were best detected with SS SE-EPI DWI with b=20 s/mm 2 . The best image quality was achieved with T2w TSE. The best lesion conspicuity was achieved with T2w TSE for biliary cysts and with SS SE-EPI DWI using b=20 s/ mm 2 for hemangiomas and metastases. Image artifacts were lowest with T2w TSE (p < 0.05). The application of SS SE-EPI DWI with low b-values can provide useful information regarding the detection of focal liver lesions in general and small lesions in particular when compared with T2w TSE. Purpose: To evaluate the efficacy of ferucarbotran in T2w-and T2*w sequences for the characterization of focal liver lesions. In 68 patients pre-contrast T2w-TSE and contrast-enhanced T2w-TSE and T2*w-GRE images after i.v.-injection of 1.4 ml ferucarbotran (Resovist, Schering AG) were obtained on a 1.5 T MR-system. The final diagnoses included 46 malignant and 22 benign liver lesions established by histology (35/68) and long-term follow-up (33/68). SNR and CNR was calculated for all sequences. The percentage signal-intensity loss (PSIL) of focal lesions was calculated for the T2w-TSE sequence. Qualitative analysis was performed for image quality and lesion conspicuity. The mean SNR of solid benign lesions showed a decrease from 34.1 to 21.0 (p < 0.05). In malignant lesions, the mean SNR showed only a minor decrease from 33.3 to 32.5. The mean CNR of the malignant lesions was the highest in the contrast-enhanced-T2*w sequence as compared to the post-and the pre-contrast T2w-TSE sequence (p < 0.01). With a threshold PSIL of 25 %, the sensitivity and specificity for the characterization of malignant lesions was 97.8% respectively 92.9%. Contrast-enhanced-T2*w images showed a superior image quality and lesion conspicuity (p < 0.05). Conclusion: Solid benign lesions showed SPIO uptake with a significant signal loss in contrast to malignant lesions. When a threshold of 25% is used, PSIL is an accurate tool for the characterisation of benign and malignant lesions. For the detection of malignant lesions, the add-on of a contrast-enhanced-T2*w-GRE sequence is helpful due to the higher CNR, a superior lesion conspicuity and better image quality. The Purpose: To evaluate the diagnostic performance of double contrast MR imaging with sequential administration of gadobenate-dimeglumine (Gd-BOPTA) and ferucarbotran (Resovist) for detection and characterization of focal hepatic lesions. Unenhanced Gd-BOPTA-enhanced dynamic MR imaging using 3D-gradient echo sequence, and then Resovist-enhanced MR imaging using T2-weighted turbo spin-echo and T2 * -weighted gradient echo sequences were obtained in 57 patients with focal hepatic lesions. Among total of 121 lesions, 89 lesions were confirmed by pathology. Thirty two benign lesions were diagnosed by a consensus panel of the all-clinical and imaging data, and at least 6 months follow-up. Two radiologists independently reviewed the dynamic MR imaging set and the double contrast set, respectively. In both sets, sensitivity and accuracy of detection and characterization were compared by alternative ROC analysis. Results: Overall detection accuracy (A z ) of double contrast set (0.902 for reader 1, 0.870 for reader 2) was better than that of dynamic set (0.836 for reader 1, 0.828 for reader 2) (P <.05 for reader 1). Both readers were more accurate in differentiating benign from malignant lesions with double contrast set (A z : 0.882 for reader 1, 0.929 for reader 2) than with dynamic set (A z : 0.832 for reader 1, 0.804 for reader 2) (P <.05). The sensitivities of detection and characterization for the double contrast set were higher than dynamic set. Purpose: To identify the benefit of FDG PET-CT over FDG PET with diagnostic CT in staging thoracic oesophageal and gastro-oesophageal junction cancer. Methods and Materials: Following ethical approval, 43 consecutive patients (15 male, 6 female; mean 68 years) with proven oesophageal cancer were examined prospectively with PET-CT (Discovery ST, GE Healthcare Technologies). Following IV injection (4.5 MBq/kg FDG), scan acquisition (5 minutes emission scan/ bed position) was commenced 1 hour post injection; iterative reconstruction was performed to view images. All studies were read in consensus by two radiologists with PET experience. Initially the PET alone was read with the diagnostic CT; then the combined PET-CT was read; overall stage was noted at each instance. Comparison was made with the original clinical stage and any change in management was subsequently recorded. Results: There were 34 adenocarcinoma, 11 squamous carcinoma; 29 thoracic oesophagus, and 14 gastro-oesophageal junction. In 12 (23%) patients PET-CT changed stage from M0 to M1; in 5 this was due to unexpected nodal involvement (M1a) and in 7 due to distant metastases (M1b-liver (4), lung (2), bone (1)). In 4 M1 disease was shown only on PET-CT; this included 1 M1a, and 3 M1b disease (liver (2), bone (1)). In 2 further patients PET-CT showed a asymptomatic colorectal cancer, subsequently confirmed histologically. Thus PET-CT specifically influenced management in 5 cases (12%). In thoracic oesophageal and GOJ cancer, FDG PET-CT confers significant additional benefit to PET read with diagnostic CT, providing valuable incremental information which results in change of clinical management. The CT patterns of disease recurrence following oesophagectomy for carcinoma of the gastro-oesophageal junction Optimal oral preparation for gastric cancer staging I. Britton 1 , S. Chang 2 , D. Wright 2 ; 1 Stoke-upon-Trent/UK, 2 Boston/UK (ingrid.morley-davies@virgin.net) Purpose: Optimal gastric distension is essential to ensure normal mucosal folds are not mistaken for tumour growth. Whole milk contains 4% fat, providing negative contrast and delaying gastric emptying to produce gastric distension. This study compares different gastric preparations, to determine which oral preparation provides optimal distension in gastric cancer staging. Methods and Materials: All post-surgical gastric cancer pathology reports between April 2002 to April 2004 were retrospectively reviewed (n75). Of these, 50 patients had preoperative staging CT scans. Three different oral preparations were in use over this time period; (1) 500-1000 mls of water (n 14), (2) 1L of full fat milk (n 24) or (3) carbex with prone scanning (n 8). All images were acquired in 3 mm increments. The 8 mm hardcopy images were reviewed retrospectively. For each patient, maximum normal wall thickness was measured in each slice and graded; grade 1 (1 -4 mm), grade 2 (5 -10 mm) and grade 3 (> 10 mm). The percentage of images per patient demonstrating optimal distension (grade 1) was calculated. Four patients with linitus plastica were excluded for the purposes of statistical analysis. Results: Milk achieved the greatest percentage of images with grade 1 distension (74%) compared with water (38%) and carbex/prone (41%) P < 0.04. Conclusion: Milk provides optimal reproducible distension as a negative contrast agent, is palatable to patients and delivers cost benefits when compared to barium or calogen preparations. Methods and Materials: 148 consecutive patients with biopsy-proven PGC underwent MDCT scan prior to gastrectomy. We evaluated the CT images separately for the site, size, depth, lymph node and enhancement characteristic of each case. Each post-surgical stomach specimen was axially sectioned and comparison was made to determine the correlation between the CT findings and the pathological examination of each tumor bearing slice. Results: MDCT correctly predicted 72 of the 94 tumors with obliteration of the fat plane between the gastric tumor and an adjacent organ or invasion of an adjacent organ. MDCT correctly identified 65/76 patients with GBA invasion and 31/ 40 patients with metastatic lymph node greater than 1 cm in GBA. Patients with PGC of the posterior wall (case group, n = 90) showed significant CT features of GBA invasion as compared to the patients with PGC of the anterior wall (control group, n = 58); increased tumor infiltration into subphrenic fat, marked gastrophrenic ligament invasion, and greater lymph nodes along the GBA or subphrenic retroperitoneal space (p < 0.05). In vivo measurement of the surface distance from the pylorus to the early gastric carcinoma using CT gastrography J. Choi 1 , H. Chung 2 , H. Ryeom 2 , H. Choi 2 , W. Yu 2 , J. Lee 2 , K. Cleary 1 , S.K. Mun 1 ; 1 Washington, DC/US, 2 Daegu/KR (choi@isis.georgetown.edu) Purpose: Early gastric carcinomas (EGCs) are often not perceptible by surgical palpation or inspection. Even gastroscopy does not provide the accurate position for carcinomas, so determining the extent of surgical resection is extremely difficult. The aim of this study is 1) to develop software measuring in vivo surface distance from anatomic landmarks to carcinomas using CT gastrography, 2) to compare the distances measured by the developed software and pathologic findings. Methods and Materials: Twenty-one consecutive patients (male = 19, female = 2, mean age: 59) with surgically proven EGC were included in this study. Preoperative CT scans (Slice thickness/Reconstruction interval: 0.625 mm, kVp: 120, mAs: 200) were performed after insufflations of room air via a 5-F nasogastric catheter. Three-dimensional volume rendering and surface distance measurement from pylorus to EGC were performed with the developed software. We assessed the correlation between the distances measured by pathologic specimen and the developed software. The difference between the lesion to pylorus distances measured from pathological specimens and CT gastrography was 5.04 ± 2.97 mm (range, 0-11 mm). CT gastrography-based measurement overestimated the lesion to pylorus distance in 18 patients and underestimated in three patients. The lesion to pylorus distance measured from CT gastrography was well correlated with that measured from the pathological specimens (r = 0.995, p < 0.01). These results suggest that the surface distance from an anatomic landmark to the EGC can be measured accurately by CT gastrography. This technique could be used for preoperative localization of early gastric carcinomas to determine the optimal extent of surgical resection. Purpose: To evaluate whether indolent and aggressive gastric and duodenal lymphoma can be differentiated by hydro-multidetector CT (CT) or endoscopic ultrasound (EUS) and to assess their correlation of findings. The study group consisted of 14 patients with histopathologically proven lymphoma: 9 had an aggressive lymphoma and 5 an indolent lymphoma. Lesion thickness was measured in mm and invasion of the gastric or duodenal wall was morphologically described by a 5-point-scale according to the methodological capacities (CT: 1=focal, 2=diffuse, 3=blurring of fat plane; EUS: 1=mucosa, 2=muscularis, 3=serosa, for both; 0=normal, 4=infiltration of adjacent organs). Results: With both imaging methods, wall thickness and infiltration was larger in aggressive lymphoma than in indolent lymphoma, but differences were not significant. In aggressive lymphoma, thickness of diseased wall was 0.8-12 cm with CT but only 0.3-1.3 cm with EUS. In indolent lymphoma the wall thickness was 0.7-1.3 cm with CT and 0.3-0.9 cm with EUS. The mean scores for wall invasion were 2.9 and 2.4 for aggressive lymphoma, measured with CT and EUS, respectively, and 1.5 and 0.8 for indolent lymphoma. For wall invasion, EUS and CT showed a high correlation (r s = 0.8, p < .05) although the criteria used differed. For wall thickness no linear correlation was found (r = 0.1). Conclusion: Although aggressive lymphoma tends to cause higher wall thickness and more advanced infiltration, no clear distinction can be made based on morphology provided by EUS or CT. Both methods showed a high correlation in assessment of wall infiltration but did not find a correlation in wall thickness. In vitro imaging of gastric cancer: Phase contrast imaging with synchrotron radiation X.P. Zhang, L. Tang, Y.S. Sun, J. Li, G. Li; Beijing/CN (zxpabc@263.net) Purpose: To investigate the ability of phase contrast imaging with synchrotron radiation (SR-PCI) in demonstrating gastric wall layers and invasion depth of gastric cancer in vitro. Six specimens of gastric cancer were fixed in formalin for 2-7 days. Then the specimens were cut into strips with a thickness of 1 mm and performed SR-PCI examination. The synchrotron radiation came from 4 W1 A beamline of Beijing Synchrotron Radiation Facility (BSRF). Double-crystal monochrometer Si (111) was employed to obtain the coherence and monochromatic X-ray. The X-ray energy was 9 keV, and the distance from the X-ray source to the specimens was 45m. A X-ray CCD with 10 µm spatial resolution was used as the detector. The results were compared with histopathology. Results: It could differentiate 4-5 layers of normal gastric wall with SR-PCI, which were mucosa, submucosa, muscularis propria and subserosa/serosa. All of the six specimens could be correctly judged about the cancer invasion depth. It could also show clearly the foveolae gastricae and vessels in the submucosa whose diameter exceeding 20 µm. Conclusion: SR-PCI can demonstrate the tiny architecture of normal gastric wall and gastric cancer in vitro. Gastric cancer: Diffusion-weighted MR imaging with sensitivity-encoding technique X.P. Zhang, L. Tang, Y.S. Sun, J. Li, Y. Li; Beijing/CN (zxpabc@263.net) To determine the feasibility of diffusion-weighted MR imaging using sensitivity-encoding technique (ASSET-DWI) in depicting gastric cancer. Twenty-three patients with gastric cancer were examined at 1.5 T. One DWI sequence and three ASSET-DWI sequences with different NEXs (Se1: 1NEX; Se2: 2NEXs; Se3: 4NEXs) were employed to generate diffusion-weighted images. When the acquisition time exceeded 15 seconds, the sequence will be separated to 2-3 breath-holds using a "pause scan" function key. The b values were 0 and 1000 sec/mm 2 . Two radiologists evaluated the obtained images from the viewpoints of tumor detectability, gradation of artifacts and contrast-to-noise ratio (CNR). The apparent diffusion coefficients (ADCs) of tumors and water in the gastric lumen were measured and compared. The EPI-related artifacts and image distortion of ASSET-DWI was reduced compared to that without ASSET (p < 0.01). There was no influence of separate breath-holds on the ADC measurement (p > 0.05) of both water and tumors. Separate breath-holds with multi-NEX technique could reduce the AS-SET-related artifacts (p < 0.01) and achieve higher CNR (Se1: 12.16 ± 3.75, Se3: 17.65 ± 7.59, p < 0.01). The mean ADC of gastric cancer was 1.17 mm 2 /s×10 -3 ± 0.22. Conclusion: High quality ASSET-DWI imaging of gastric cancer can be achieved through the combination of separate breath-holds and multi-NEX technique. The diffusion-weighted MR imaging may provide functional parameters for the diagnosis of gastric cancer. patients, 33 male and 20 female, mean age 65.3, were evaluated with conventional endoscopy with biopsies, EUS and MDC. We considered only patients with gastric adenocarcinoma. All the CT examinations were performed after the oral administration of 700/900 ml of water for gastric wall distension; each examination was performed with triphasic technique (basal, arterial and portal phase). The images were post-processed using Multi-Planar Reformatting (MPR). The lesion morphology and thickness, the degree of gastric wall infiltration, locoregional lymph nodes and systemic metastases were evaluated by two radiologists blinded to the EUS findings. All the patients enrolled in the study underwent surgery; then the preoperative EUS and MDCT staging was evaluated against the pathology findings. Results: Gastric cancer was confirmed in all patients. Overall accuracy, sensitivity, and specificity in the evaluation of T stage for EUS and three-dimensional MDCT were 86%, 87%, 90% and 85%, 77.5% and 92.8% respectively. For the evaluation of N factor for EUS and three-dimensional MDCT were respectively, 77%, 67.6%, 80% and 72.8%, 57% and 88%. Conclusion: Although EUS seem to have a better overall accuracy than MDCT in the evaluation of T1 and T3 tumors, three-dimensional MDCT still presents higher specifity and remains the cross sectional imaging technique of choice in the preoperative evaluation of gastric cancer. To investigate the possibility of assessing the underlying respiratory disease as well as cardiac function during ECG-gated CT angiography of the chest with 64-slice multidetector-row CT (MDCT). Methods and Materials: 133 consecutive patients underwent an ECG-gated CT angiographic examination of the chest without ß-blockers using the following parameters: (a) collimation: 64x 0.6 mm; pitch: 0.3; (b) 120 mL of a 35% contrast agent. Data was reconstructed: (a) to evaluate the underlying respiratory disease (1-mm thick lung and mediastinal scans reconstructed at 55% of the R-R interval); and (b) to determine right and left ventricular ejection fractions (short-axis systolic and diastolic images). The mean heart rate was 73 bpm and the mean scan time was 18.11 ± 2.67 sec. A total of 123 examinations (92%) had both lung and mediastinal images rated as diagnostic whereas 10 examinations (8%) had nondiagnostic images altered by respiratory-motion artefacts (n = 4) or cyclic artefacts related to the use of a pitch value of 0.3 in patients with a very low heart rate (n = 6). Assessment of right and left ventricular function was achievable in 124 patients (93%, 95% CI: 88%-97%). In the remaining 9 patients, an imprecise segmentation of right and left ventricular cavities was considered as a limiting factor for calculation of end-systolic and end-diastolic ventricular volumes. The mean (± SD) DLP value of the examinations was 279.86 (± 117.50) mGy.cm. Conclusion: Assessment of underlying respiratory disease and cardiac function from the same data set was achievable in 92% of patients with ECG-gated 64slice MDCT. . The presence, location and severity of cardiac motion artifacts at the level of pulmonary arteries were analyzed on 1 mm-thick CT angiograms obtained in strict apnea. Results: Considering central pulmonary arteries, the frequency of cardiogenic artefacts in Group 3 was significantly lower than that observed in Group 1 and Group 2 at the level of : (a) the pulmonary trunk: n = 6 (9%; Group 3) vs n = 4 (12.5%; Group 2), n = 25 (68%; Group 1) (p < 0.001); (b) right main pulmonary artery: n = 0 (0%; Group 3) vs n = 5 (16%; Group 2), n = 18 (49%; Group 1) (p < 0.001). Considering intraparenchymal pulmonary arteries, the frequency of cardiogenic artefacts was significantly lower in Group 3 (n = 9; 14%) compared to Group 2 (n = 22; 69%) and Group 1 (n = 34; 92%) (p < 0.001). Whereas cardiogenic motion artefacts were responsible for pseudo-filling defects in 20 patients in Group 1 (54%) and in 8 patients in Group 2 (25%), only 1 patient showed pseudo-filling defects in Group 3 (1.5%) (p < 0.0001). Conclusion: Motion artefacts in the pulmonary arteries causing interpretive difficulties on multislice CT scans can be significantly reduced by ECG gating. and evaluation (n = 13) were obtained with a 4-slice scanner with a pixel size of 0.6 mm, a spacing between slices of 0.6 mm and a slice thickness of 1.3 mm. The study group consisted of 13 CTPA studies with a mean number of 4 thrombi (range 1-10). Datasets were selected to demonstrate a variety of thrombus load, considerable breathing artifacts (n = 2), sub-optimal contrast (n = 1) and additional parenchymal disease (n = 6). One experienced radiologist was asked to annotate all thrombi. The most proximal location of the thrombus defined its anatomic classification (main, lobar, segmental or sub-segmental), the thrombus size defined its conspicuity (obvious, medium or subtle). Results: Without CAD, the radiologist annotated 37 emboli located in the main (n = 7), lobar (n = 6), segmental (n = 14) and sub-segmental arteries (n = 10). With CAD, 12 additional thrombi were detected (increment of 32%), located in segmental (n = 2/16, 12%) and sub-segmental arteries (n = 10/20, 50%). Thrombi additionally detected by CAD were classified as being of subtle conspicuity (9/ 12, 75%) or medium conspicuity (3/12, 25%). Per dataset the CAD system showed on average 8.7 false positive calls (median = 7). Conclusion: Compared to the radiologist, the prototype CAD system found additional emboli in segmental and sub-segmental arteries. These preliminary results suggest that CAD may become a helpful tool for diagnosing low grade acute pulmonary embolism in peripheral arteries. Purpose: To evaluate the performance of a prototype CAD-tool using artificial intelligence for the detection of pulmonary embolism (PE) and the benefit for radiologists. Methods and Materials: 40 MDCT datasets using 100 kVp, 100 mAs effective per slice and 1 mm thick axial reformats in a low frequency reconstruction kernel were evaluated. 80 cc iodinated contrast agent was injected with a 5 cc/sec flow rate. Primarily, 6 community level radiologists marked any PE using dedicated lung evaluation software for visualization with consecutive processing with the PE-CAD-prototype. An expert panel consisting of two dedicated chest radiologists had to confirm any PE marked by the readers and/or the CAD-prototype and to mark any additional PE primarily missed by both, creating the reference standard. Purpose: The aim of this study was to find out whether late enhancement (LE) in magnetic resonance imaging (MRI) as a marker of myocardial fibrosis is a good parameter to establish a success of enzyme replacement therapy (ERT) in patients with Fabry disease. Methods and Materials: 17 patients (13 men/4 women; age 41 ± 8) with genetically proven Fabry disease received ERT (agalsidase ß) and were followed for 12 months. At baseline and after a year all 17 patients underwent cardiac MR imaging and additionally colour Doppler myocardial imaging to confirm an improvement of the regional function under ERT. Results: 8/17 (47%) patients showed LE at baseline (1.3 ± 1.1% of left ventricular (LV) mass). The regional function was reduced in segments with LE. The LV mass of patients without LE at baseline was higher than in patients with LE (160 ± 23 g vs. 211 ± 58 g; p = 0.046). The LV mass at the follow-up after a year of ERT only decreased in patients without LE (145 ± 27 g; p = 0.008). An improvement of the regional function was limited o segments without LE. The share of LE did not decrease under ERT. Conclusion: Patients without LE showed a decreased LV mass and an improvement of the regional function, as a sign for the benefit of ERT. LE cannot be reduced by 12 months of ERT as well as the regional function in segments with LE did not recover. Purpose: Anderson-Fabry disease (AFD) is rare X-linked disorder of glycosphingolipid metabolism. Cardiac involvement represents one of the severest aspects of this disease. Our aim was to investigate the occurrence and distribution of lateenhancement (LE) in the myocardium of these patients and to correlate LE presence with histological features. Methods and Materials: 10 AFD patients underwent cardiac magnetic resonance (CMR) on a 1.5-T system (Philips). CMR included the acquisition of delayed images 10-15 minutes after gadolinium injection, with 3D-T1w-TFE sequence, adjusting inversion time to null normal myocardium. LE was expressed as percentage of left ventricle (LV) myocardium. All patients underwent LV endomyocardial biopsy with automated morphometric quantification (Nikon) of fibrosis on section stained with Masson's trichrome. Results: 7 patients (70%) showed the presence of LE areas ranging from 1 to 17.5% (mean 6.5 ± 5.8%) of LV myocardium. The fibrosis measured on histological samples ranged between 4 and 18% (mean 9.1 ± 4.8%). LE percentage at CMR was strongly related to the percentage of fibrosis calculated at histology (R=0.93; P < 0.0005). LE areas were located in the medium-basal segments of lateral LV wall in all patients. Only two patients with severe LV hypertrophy showed additional LE areas in apical segments. Conclusion: LE located in the medium-basal lateral wall of LV might be considered as a specific sign of cardiac involvement in AFD patients. The strong relation between LE extension and histological fibrosis quantification suggests that LE percentage could represent an important non-invasive marker of the severity of cardiac involvement in Fabry disease. Therapy follow-up via cardiac MR imaging in patients with cardiac sarcoidosis M. Heller 1 , M. Schiemann 1 , N.D. Abolmaali 2 , T.J. Vogl 1 ; 1 Frankfurt/DE, 2 Dresden/DE (crhcrx@gmx.net) Purpose: To assess the follow-up in patients with cardiac sarcoidosis under therapy with corticoids using cardiac MR imaging. Methods and Materials: 18 patients with cardiac sarcoidosis underwent an initial and 3, 6 and 12 month follow-up MR imaging under therapy with corticoids. Short axis views covering the entire heart and 4-chamber views were done with T2-w images using a breath hold turbo inversion recovery sequence. Additionally delayed enhancement imaging was done with a 3d turbo flash sequence using an optimized TI-time of 210 ms in the short and long axis. Dynamic left ventricular analysis was done with contiguous short axis views covering both ventricles using a True-FISP cine sequence. Additionally 3-and 4-chamber views were performed. MR imaging findings were categorized into stable (± 50 cardiac infiltration volume), worsening (> 50 % more infiltration volume) or improving situations (> 50 % less cardiac infiltration volume). Results: 4 patients showed an improving situation with 3 patients presented a complete clearing of patchy granulomatous infiltrations with no areas of late enhancement while in 9 patients the MR imaging showed a stable situation with persistence of cardiac infiltrations in the same extension as in the initial imaging. Worsening situations of cardiac involvement was seen in 5 patients who all died due to therapy refractory ventricular tachyarrhythmias within 2 years. Conclusion: Cardiac MR imaging offers a rapid and non-invasive screening method with high spatial resolution for the detection of myocardial involvement and fibrosis occurring in cardiac sarcoidosis and is a useful method in follow-up under treatment. Results: Adequate image quality could be achieved in all but one patient. MR imaging analysis revealed hypertrophic cardiomyopathy (n = 3), residual myocarditis (n = 2), myocardial infarction (n = 2), dilated cardiomyopathy (n = 1), right ventricular dilatation (n = 1) and focal dyskinesia (n = 1). Delayed enhancement with spotted, streaky, subepicardial or transmural distribution could be found in 3 of our 12 patients. A reduction of the ejection fraction could be detected in 4 patients resulting in a mean shortening fraction of 45% and hypo-to akinetic areas in up to 3 patients. Only two patients showed no pathologic abnormalities. Conclusion: Although no specific MR pattern was found in our study cohort, MR allowed characterizing the underlying cardiac pathologies in 82% of HIV patients with elevated BNP levels. Purpose: For the diagnosis of Left ventricular non-compaction (LVNC) four main criteria have been proposed, both in echocardiography and more recently in cardiac MR imaging. In this study we specifically evaluated the "non-compacted to compacted myocardium" ratio (NC/C) in distinguishing LVNC from other cardiomyopathies. Methods and Materials: Nine patients with suspected LVNC were enrolled. We performed the following MR imaging sequences (Siemens Symphony 1.5 T): 1) TSE T1 black blood and cine TrueFISP sequences for morphological and functional analysis in both ventricles; 2) a perfusion study with first pass sequence during gadolinium injection (0.05 mmol/kg) on short axis at the basis, medium and apical level of the left ventricle; 3) delayed enhancement at10 minutes was evaluated on short axis, two chambers long axis and four chambers long axis to detect myocardial fibrosis. Results: In all patients we identified the typical pattern of thin-layered epicardium and hypertrabeculation of the endocardium in the antero-lateral, medialdistal wall and in the apex. The NC/C ratio measured during diastolic phase was > 2.3 in all the cases, but in 3 patients this value was not confirmed from the measurement obtained in the systolic phase. Systolic function of left ventricle was reduced in 5 patients. First pass study revealed delayed perfusion both in the non compacted area and the compacted one. Only in one case myocardiac fibrosis was detected. Conclusion: In our experience the NC/C ratio may be not a sufficient index in the differential diagnosis of LVNC. Other MR imaging criteria should be prospectively evaluated. Results: In 45 patients a primary neoplasm of parotid gland could be histologically verified. Seven types of parotid gland tumor were histologically discriminated. Pleomorphic adenomas (2.14 x 10 -3 mm 2 /sec ± 0.11 x 10 -3 mm 2 /sec) (mean ± standard deviation) and Warthin tumors (0.85 x 10 -3 mm 2 /sec ± 0.1 x 10 -3 mm 2 / sec) showed significantly different ADC values in comparison to all other evaluated tumors (p < 0.001), and also between each other (p < 0.001). ADC values presented by lipomas were statistically significant compared to all other entities (p < 0.001 to 0.015). Among primary malignant parotid gland tumors no statistically significant ADC values could be observed (p = 0.18 to 1). Conclusion: Diffusion-weighted echo-planar MR imaging seems to be a valuable tool for differentiating benign from malignant primary parotid gland tumors. Purpose: To investigate the feasibility of using diffusion-weighted (DW) echoplanar imaging (EPI) for measuring different functional conditions of the parotid gland and to compare different field strengths. Methods and Materials: Parotid glands of 27 healthy volunteers were examined with a DW EPI sequence at 1.5 T and 3 T before and after oral stimulation with commercially available lemon-juice. The b-factors used were 0, 500, and 1,000 sec/mm 2 . Apparent diffusion coefficient (ADC) maps were evaluated with a manually placed region of interest (ROI) including the entire parotid gland. For comparison of the results, Student's t-test was used, based on the mean of the volunteer median values. To compare both field strengths, the Pearson's correlation coefficient (r) was calculated. Results: DW EPI MR imaging successfully visualized the parotid glands of all volunteers. Using 1.5 T, the mean ADC prior to stimulation was 1.12 x 10 -3 mm 2 / sec ± 0.08 x 10 -3 mm 2 /sec, whereas after stimulation with lemon juice, the ADC increased to 1.18 x 10 -3 mm 2 /sec ± 0.09 x 10 -3 mm 2 /sec. For 3 T, the ADC prior to stimulation was 1.14 x 10 -3 mm 2 /sec ± 0.04 x 10 -3 mm 2 /sec, with an increase to 1.17 x 10 -3 mm 2 /sec ± 0.05 x 10 -3 mm 2 /sec after stimulation. For both field strengths, the increase in ADC after stimulation was significant (P < 0.001). High correlations between both field strengths were found prior and post stimulation (r = 0.955, and 0.936 respectively). Diffusion-weighted echo-planar MR imaging allows quantification of functional changes in the parotid glands with high correlation between 1.5 T and 3 T. Perineural extension along the infraorbital nerve was defined as swelling or thickening of the infraorbital nerve with contrast enhancement. The frequency of perineural extension along the infraorbital nerve was retrospectively studied. Its coexistence with tumors in the pterygopalatine fossa and superficial musculoaponeurotic system was also evaluated. Results: Perineural extension along the infraorbital nerve was observed in 16 patients (8.6%). The primary sites of the tumors were the maxilla (n = 8), orbit (n = 2), hard palate (n = 2), nasal cavity (n = 2), and upper gingival (n = 2). The pathological diagnoses were squamous cell carcinoma (n = 10), adenoid cystic carcinoma (n = 3), malignant melanoma (n = 2), and malignant lymphoma (n = 1). All cases except one, an orbital adenoid cystic carcinoma, showed coexistence with tumors in the pterygopalatine fossa and superficial musculoaponeurotic system. The present study demonstrated that perineural extension along the infraorbital nerve was not rare in patients with head and neck tumors. Perineural extension usually coexists with tumors in the pterygopalatine fossa and superficial musculoaponeurotic system. Coronal MR images are excellent for demonstrating perineural extension along the infraorbital nerve. Purpose: To assess influence of partial neuromuscular blockade on the upper airway system using realtime TrueFISP MR imaging and volumetry of the upper airways. Methods and Materials: 10 healthy volunteers were studied before, during (trainof-four [TOF]-ratio: 0.5 and 0.8) and after (TOF-ratio: 1.0 and 15 minutes later) rocuronium induced partial neuromuscular transmission failure which did not evoke dyspnea or oxygen desaturation. Imaging was performed on a 1.5 T scanner using a head/neck surface-coil for signal reception. Functional exploration of the upper airway system was performed using sagittal real-time TrueFISP sequences (temporal resolution 250 ms). Datasets were acquired at rest and forced inspiration before, during and twice after relaxation. Minimal distance from back of tongue to posterior wall of pharyngeal space was evaluated. Additionally, sequential T1w images were collected in transverse orientation from the top of the hard palate to the vocal cords for the volumetric airway measurements. Results: During partial neuromuscular blockade, upper airways partially collapsed at end-inspiration. Measurements of minimal distances during forced inspiration showed a mean reduction from 20 mm before relaxation to 14 mm during maximal relaxation. Comparing pre-relaxation datasets to TOF-ratio 0.5 datasets, mean inspiratory volumes correspondingly decreased to 78 ± 11 % of baseline (p < 0.01). Complete recovery was not found until 10 minutes after termination of relaxation. Purpose: To examine prospectively the clinical effects of renal artery stenting (RAS) in patients with global renal artery stenosis and chronic renal insufficiency. Methods and Materials: A successful RAS was performed in 30 patients with chronic renal impairment (serum creatinine > 1.5 mg/dl) and global atherosclerotic obstruction (bilateral renal artery stenosis or unilateral stenosis in a presence of a solitary or single functional kidney). To assess a significant change of RF, we managed to compare the slopes of the regression lines derived from the reciprocal of creatinine versus time, plotted before and after stent deployment. We also evaluated changes in renal length and interlobar resistive index. Median follow-up was 24 months (range 9-52). Results: Before stenting all the patients showed a negative slope, indicating progressive renal failure. After stenting the mean of slopes became positive (mean pre -0.0249 ± 0.0527 vs post 0.0048 ± 0.0196, p < 0.008). The Viabil is a self-expandable stent-graft with an inner ePT-FE/FEP lining and an outer supporting structure of nitinol wire. Biliary obstruction was secondary to pancreatic carcinoma (n = 38), cholangiocarcinoma (n = 10), gallbladder cancer (n = 6) and metastatic lymphadenopathy (n = 11). The stricture was located in the upper third of the common bile duct in eight patients, in the middle portion in 43 and in the lower third in 14. Sixty-nine stent-grafts were implanted: 49 with side holes and 20 without holes. In 62/65 patients the distal end of the device was deployed in the duodenum. Clinical evaluation and assessment of serum bilirubin and liver enzyme levels were done before stent placement and at 1, 3, 6, 9 and 12 months. Results: Stent-graft placement was successful in all cases. The 30-day mortality rate was 11%. The survival rates were 40% at 6 months and 20.2% at 12 months. The stent patency rates were 91%, 78% and 78% at 3, 6 and 12 months respectively. After a mean follow-up of 37.2 months we reported acute cholecystitis in 6 cases (9.2%). Complications other than stent occlusion occurred in 12 patients (18.4%). Conclusion: e-PTFE covered stent-graft seems to be an effective tool in cases of malignant strictures of the common bile duct. men and 13 women, comprised and were followed-up in this study. We used the 10 mm wide Viabil© stent (WL GORE, USA) of 6 or 8 cm length, with/without proximal side holes. Biliary obstruction was found in the lower CBD (in 17), upper CBD including cystic duct (in 9), lower CHD including cystic duct (in 4), CHD (in 2), upper CHD and left main bile duct (in 2), and left main biliary duct (in 1). One stent was placed in each patient, while in 2 patients a second Viabil was inserted after occlusion of the first one. In another 2 cases uncovered stents were placed after Viabil occlusion. All stents were successfully placed except one, which migrated distally (technical success rate 97%). Thirty day mortality was 24% and not procedure related. Eleven out of 35 patients are still alive (mean survival 105 days). Mean primary and secondary patency was 82% and 99%, respectively. No serious complications were noted despite a relatively large delivering catheter diameter. No pancreatitis occurred in cases when the papilla was fully covered. Conclusion: Patency, survival and complication rates of the Viabil stents are comparable with those of the known uncovered stents. Larger patient numbers and longer follow-up, as well as correlation studies with patient groups with uncovered metallic stents are needed for more conclusive results. The development of a retrievable stent for the recanalization of thrombusoccluded vessels J. Koizumi 1 , A. Kitamura 2 , K. Tokuyasu 3 , R. Hatomi 3 , Y. Imai 1 ; 1 Isehara-city/JP, 2 Matto-city/JP, 3 Sagamihara-city/JP (jkoizumi@is.icc.u-tokai.ac.jp) Purpose: For massive thrombus-occluded vessels such as pulmonary embolism with cor pulmonale or deep venous thrombosis, direct stenting alongside the pulmonary emboli may be effective for immediate decompression. However, the lack of long-term durability and possible thrombogenic effect are theoretical disadvantages. Thus we developed a retrievable stent to obtain an immediate but temporary recanalization of the thrombus-occluded vessels by compressing the clot. The conical stent was developed using one string of nitinol wire including a hook at one end for loop snare retrieval. The 15 x 65 mm stents composed of three different diameters (0.20, 0.25, 0.30 mm) of wire filaments were deployed in clot-filled test tubes (inner diameter; 13 mm). For various age clot-filled test tubes from day 0 to 7, the minimal recanalization ratios (instent diameter/total lumen diameter) were measured using CT. After measurement the stents were retrieved and the existence of in-stent thrombus was observed. Results: All stents' recanalization ratios decreased with clot age and the stronger two stents (0.25, 0.30 mm) had significantly higher ratios (70-80% vs. 50-60%) than the weakest one (0.20 mm). In-stent thrombus was not observed. Conclusion: A conical shaped retrievable stent was feasible for immediate recanalization by compressing thrombus in vitro. Endobronchial dilatation and stent placement for bronchial stenosis in patients after lung transplantation N. Abi-Jaoudeh 1 , G. Soulez 2 , V.L. Oliva 2 , M.-F. Giroux 2 , E. Therasse 2 , R. Jean-Francois 2 , C. Poirier 2 ; 1 Quebec, PQ/CA, 2 Montreal, PQ/CA (nadine.abi-jaoudeh@sympatico.ca) Purpose: To evaluate the outcome of balloon dilation and/or stent placement in patients with bronchial strictures after lung transplantation. Methods and Materials: Forty-one lung recipients underwent balloon dilation and/or stent placement for bronchial strictures between January 1997 and January 2005. Clinical assessment, pulmonary function testing and bronchoscopic evaluation were obtained before and after each procedure. Clinical and bronchoscopic follow-up ranged from 1 to 87 months (mean 17 months, median 11 months). Results: Of the forty-one patients treated, twenty-four were alive on the last follow-up (58%). All patients had primary balloon dilatation; endobronchial stent placement was done if balloon dilation had failed to re-expand the bronchus during the procedure or at a later time. The average number of procedures per patient was 2.8 (median 2). During a procedure an average of 1.75 lesions were treated (median 1). Twenty-three patients (56%) received a stent because of failure of balloon dilatation. Among those, the mean number of stents per patient was 1 (median 1). Primary patency (after the first procedure) was noted in thirteen patients (32%). Secondary patency (after the last procedure) was noted in thirty-eight patients (93%). There was significant improvement in clinical and mean pulmonary function testing after each procedure, with an amelioration of 8% of the VEMS (P < 0.00003). Purpose: To study the long-term therapeutic efficacy of temporary partially-covered metal stent insertion by follow-up in achalasia. We inserted 103 temporary partially-covered metal stents in 103 patients with achalasia, which were removed 3-5 days later. The placement and removal of stents were successful in 100 % of cases. Immediately after placing the stent, we performed an esophageal contrast examination, which confirmed good stricture patency. The mean follow-up time was 12-120 months. Results: Technical and clinical success was achieved in 98 and 94%; 97 and 92%; 98 and 94 % of cases respectively for, oesophageal, gastrojejunal and large bowel strictures. Better results occurred when dedicated stents with specific features for each anatomical site were used. The overall complication rate was 21% including, occlusion (12%), migration (3%), jaundice (1.2%), perforation (0.3%), stent rupture (0.3%) and others. The mean follow-up time for patients with neoplastic strictures was 4.5 months. Conclusion: Stenting of the alimentary tract has a high procedural and clinical success rate. In inoperable patients it provides a prompt reduction or disappearance of symptoms, good quality of life and survival improvement. It can be considered a safe and effective procedure and represents a less expensive treatment compared to palliative surgery which is affected by a higher mortality and morbidity rate. Purpose: To evaluate whether the motor impairment seen in ALS patients can be attributed to vascular and/or neuronal defects in the brain using BOLD fMRI. Methods and Materials: 33 ALS patients and 21 age-and sex-matched controls underwent fMRI examination using a 3 T MR imaging scanner. In the first session, a motor task was performed with the right hand to assess the activation pattern in the brain when executing voluntary movements. In the second session, an alternating hyperventilation -breath-hold task was performed to assess the vascular reactivity in the brain without interference of task-induced activations. Data was analyzed using SPM2 and activation patterns of both tasks were visualized. Additionally, a ROI analysis of different areas was performed on the images of both sessions. Results: In the motor task both groups showed activation of areas typical for voluntary movement. In ALS patients, there was a recruitment of both motor and extramotor areas. The ROI analysis showed significant increase of percentage signal change of motor areas (except M1) in ALS patients. Vascular reactivity seemed to be impaired in ALS patients in the motor cortex and frontal cortex, whereas there was no difference in vascular reactivity in the visual cortex. Purpose: Parkinson's disease is diagnosed in more than 1/200 worldwide; however 20-25% of these are estimated to be misdiagnosed. As the average population age is increasing, the need for accurate diagnosis and treatment is highlighted. We retrospectively reviewed the radiological features, and in correlation with clinical background, determined the clinical and radiological utility of DaTSCAN's in diagnosis and management of 40 diverse patients with atypical presentation's, suspicious for Parkinson's. Purpose: The differential diagnosis between typical idiopathic Parkinson's disease (PD) and the striatonigral variant of multiple system atrophy (MSA-P) is often difficult because of the presence of signs and symptoms common to both forms of Parkinsonism, particularly at symptom onset. To find the correct diagnosis early in the course of the disorders is, however, of major importance with respect to prognostic aspects and therapeutic strategies. This study investigated striatal, midbrain and cerebellar findings in MSA-P and PD patients in comparison with normal controls with three dimensional magnetic resonance imaging (3D MR imaging) based volumetry, together with DTI quantification in the nigrostriatal projection to increase the differential diagnostic accuracy between both disease entities. Methods and Materials: 11 patients with MSA-P within 5 years from symptom onset, and 12 patients with PD of similar duration were studied by 3D based MR imaging volumetrics and diffusion tensor imaging (DTI) using a GE excite II MR machine. 10 age-matched healthy controls were also studied. Direct measurement of the striatum, midbrain and cerebrellum were made and data were normalized by 3D brain volumetry. The fractional anisotropy (FA) values of magnetic resonance diffusion tensor imaging in ROI of the extrapyramidal system including globus pallidus, putamen, head of caudate and substantia nigra, were compared among the groups using one-way ANOVA. Results: MSA-P differed significantly from PD patients in terms of decreased putaminal and cerebellar volume. The FA significantly decreased in substantia nigra and globus pallidus in both MSA-P and PD groups, compared with control, and the decrease was more prominent in MSA-P patients than in PD patients (Table 1) . MR imaging volumetry of the midbrain region did not contribute to the differential diagnosis between PD and MSA-P. The extent and spatial distribution of morphological changes in the striatum and cerebellum permit the differentiation of MSA-P from PD. Assuming that the loss of FA parallels the neuronal change in the brain, the results indicate that neuronal loss in nigrostriatal projection is more severe in MSA-P patients than idiopathic Parkinson's disease in the early stage. Close comparison of striatal and cerebellar volume in addition to FA in the nigrostriatal projection may contribute to the early differential diagnosis of MSA-P and Idiopathic Parkinson's disease. Purpose: Posterior reversible encephalopathy syndrome (PRES) is a recently recognized entity, characterized by specific neurological deficits, a characteristic imaging pattern, and usually a benign course. Primarily related to hypertensive conditions, many other etiologies, notably neurotoxic agents, have been associated with PRES. The aim of this study was to determine differences in the MR features of PRES based on the underlying etiology. Methods and Materials: Fifteen patients with clinical and imaging findings consistent with PRES were included. All exams were performed on a 1.5 T superconducting system (Gyroscan NT-Intera, Philips). Based on their underlying etiology, we classified the patients into 2 groups: one with exposure to neurotoxic substances such as immunosuppressants (n = 6) or chemotherapeutic agents (n = 3), and the other with hypertension-related conditions (n = 6), including 4 patients with eclampsia. Results: In all but one patient, FLAIR imaging revealed hyperintense lesions in the subcortical white matter of the parieto-occipital regions bilaterally, partially in an asymmetrical distribution (n = 7). All patients with hypertension-related PRES had additional lesions in the basal ganglia (n = 9), whereas this region was not affected in the other patients. Brainstem involvement was seen only in the neurotoxic group (n = 3). There was no predominance of involvement of the corpus callosum (n = 2) and the cerebellum (n = 4) in either group. In one patient with eclampsia, diffusion-weighted imaging (DWI) revealed high signal lesions with a low apparent diffusion coefficient (ADC), indicating progression to infarction. Conclusion: Our preliminary results suggest that there are differences in the MR appearance of hypertension-related and drug-induced PRES, possibly reflecting differences in underlying pathophysiology. C D E F B A G Visual attention deficits in Alzheimer's disease: An fMRI study J. Hao 1 , K. Li 2 , K. Chen 1 , K. Li 2 ; 1 Shanghai/CN, 2 Beijing/CN (nancyhaojing@yahoo.com.cn) Purpose: To define the anatomical areas of activation associated with visual attention processing and to define deficits or changes that may occur in AD patients compared with control group. Methods and Materials: Thirteen AD patients and thirteen age-and educationmatched normal controls were tested in two visual search tasks (one was a conjunction task, where feature binding is required, the other was a subset task, where group stimuli are needed without feature binding) using fMRI techniques. Data analysis was performed by Statistical Parametric Mapping software (SPM 99). Voxel-by-voxel t statistics were used to compare activated brain areas between patients and controls. Results: Our findings suggest that both search tasks are controlled by partially overlapping cerebral networks, including parietal, frontal and occipito-temporal cortical regions. The AD patient group showed less activation in both parietal lobes and the left frontal regions, while increased activation was found in the right frontal lobes and the right occipito-temporal cortical regions with the conjunction task. In the subset task, decreased activation in AD patients was seen in the left parietal lobe and bilateral frontal lobes, while increased activation was seen in both medial temporal lobes. In addition, for the comparison between tasks, the difference is very small for AD patients. Control group showed a higher amplitude in the right prefrontal region, temporal cortical regions and parietal lobe. Conclusion: AD patients have a particular impairment in both search tasks. We suggest attention deficits in AD patients may be attributed to both binding problem and grouping inefficiency. A voxel-based morphometric study of Alzheimer's disease J. Hao 1 , K. Chen 1 , K. Li 2 ; 1 Shanghai/CN, 2 Beijing/CN (nancyhaojing@yahoo.com.cn) Purpose: Up till now, the study of regional gray matter atrophy in Alzheimer's disease (AD) has been assessed with regions of interest (ROI), but this method has many shortcomings. In this study we used the fully automated voxel-based morphometry (VBM) approach in order to provide a full-brain assessment of AD morphology. Methods and Materials: 19 patients with AD and 15 healthy subjects of similar age and gender ratio were included in the study. Magnetic resonance imaging was performed on a 1.5-T Siemens Sonata MR imaging system (Siemens, Erlangen, Germany). Whole brain T1-weighted 3D MPRAGE (magnetization-prepared rapid-acquisition gradient echo) data sets were acquired in the sagittal plane (TR, 1900 Purpose: To assess treatment effects in patients with Wilson's disease using an MR proton spectroscopy method. Methods and Materials: Fourteen patients diagnosed with WD and presenting only with neurological symptoms were studied before and after fourteen-month chelation therapy. A control group consisted of 12 healthy volunteers (age and sex matched). Single-voxel spectra were acquired on a 1.5 T scanner, using PRESS sequences with parameters of 22 and 2000 ms (TE/TR). Voxels (15x15x15 mm) were located in the center of the left and right globi pallidi and left and right thalami separately for each examination. All results were expressed as metabolite/creatine ratios. Data was analyzed using nonpaired t-test for significance testing; p < 0.05 indicated a statistically significant difference. The most significant metabolic changes in WD patients before treatment were located in the globi pallidi. There were notably increased Lip/Cr (p < 0.004), Glx/Cr (p < 0.025), Cho/Cr (p < 0.03) and decreased NAA/Cr (p < 0.008) ratios as compared to the control group. In post-treatment follow-up studies, decreased Lip/Cr ratio (p < 0.04) for globi pallidi and increased NAA/Cr ratio (p < 0.04) for thalami were observed when compared to pre-treatment results. The differences in findings between the left and right globi pallidi and thalami for the same patient were found to be insignificant. The reversible changes in NAA/Cr and Lip/Cr ratios may be due to functional and probably anatomical neuronal disturbances as an end result of copper deposition. The chelation therapy of WD reduces copper deposition. The presence of decreased NAA/Cr in WD patients does not necessarily reflect cell death. The study showed new potential value of MRS in the assessment of treatment effects. 1984) . There was also no significant difference in visually assessed image quality (thorax-soft tissue p > 0.17, thorax-lung p > 0.16, abdomen p > 0.17, pelvis p > 0.36). Compared to weight-adapted pediatric standard protocols, using the IG resulted in further mean dose reduction in abdominal (mean 41.7% ± 11.3%) and pelvic (mean 44.1% ± 11.3%), but not in chest CT (mean 0.7% ± 18%). The IG is a feasible tool for routine 16-MDCT, which allows for a pre-exam overview of the expected image quality. Radiation exposure can be optimized with regard to patient weight. It allows for problem-oriented examination planning and may be used as a guideline for pediatric MDCT-examinations. Purpose: Frequently in BPD patients lung function impairment and respiratory symptoms persist in childhood. Little is known about the structural substrate responsible for these abnormalities. To describe long term structural changes on CT scans of BPD patients, developing a scoring system to define these abnormalities in a systematic and reproducible fashion. Methods and Materials: Retrospective analysis of patients with a diagnosis of BPD and who had a thorax CT scan was performed. CT's were evaluated to identify and record all structural abnormalities. Next, a scoring system was developed and 3 independent readers scored all scans to determine inter-observer and intra-observer reproducibility. Results: 37 BPD patients were identified. Age at CT was 2 months to 17 years. A score was developed evaluating the 5 lobes, that included the following lesions: decreased density of parenchyma; interstitial fibrosis, thickening of the interlobular septa, triangular sub pleural opacities; bullae; air trapping; atelectasis and consolidation; ground glass; bronchiectasis and airway wall thickening. Mean CT score was 23 (3 to 45). Interobserver agreement assessed by intraclass correlation coefficient was good (ICC from 0.83 to 0.89); intra-observer agreement was excellent (ICC 0.933). A significant negative correlation between the mean score and the FEV1 was observed (R=0.612, P < 0.002). Conclusion: CT scans of a selection of BPD patients show considerable structural abnormalities. A CT-scoring system was developed that was highly reproducible. A correlation between structural and functional abnormalities was found. Magnetic resonance imaging of the fetal gallbladder and bile P.C. Brugger, D. Prayer; Vienna/AT (peter.brugger@meduniwien.ac.at) The appearance and development of the fetal gallbladder was studied with magnetic resonance imaging (MRI) with special reference to fetal cholelithiasis. Methods and Materials: 200 fetuses [20 th -40 th gestational weeks (GW)] were imaged on a 1.5 T magnet using T2-, T1-weighted (w) and steady-state freeprecession sequences in at least two section-planes. Visibility was assessed on all sequences and gallbladder dimensions (maximum length and width) were recorded. Signal intensities of bile were scored and measured on T1w images (compared to liver) and a bile/liver ratio was computed. The gallbladder could be visualized in 98% of cases. Both dimensions increased in a linear manner, though widths showed more variation. Until 30 GW, T2w hyperintensity of bile corresponded to T1w hypointensity. Afterwards an increase of the bile/liver ratio was found, with bile being iso-or hyperintense to liver or even isointense to meconium, which corresponded to T2w hypointensity. This increase was more pronounced in males. The fetal gallbladder can be visualized with MR imaging by virtue of the signal properties of fetal bile, which shows age-dependent signal characteristics. An increase in signal intensity is a normal finding after 30 GW and more pronounced in male fetuses. This may cause non-visualization of the gallbladder on T2w sequences. It is proposed to consider bile iso-or hyperintense to hepatic tissue as sludge and refer to bile isointense to meconium as gallstones. These new findings may offer new insights into the physiology of fetal bile and aid interpretation of fetal and neonatal abdominal MR imaging scans. Purpose: To evaluate the feasibility and diagnostic potential of 3DUS based virtual cystoscopy in infants and children, presenting our initial and preliminary experiences. Methods and Materials: 10 patients (age: 1 to 14 years) who presented with bladder pathology on conventional 2DUS (Sequoia 512, Acuson/Siemens, Mountain View, USA) underwent a subsequent 3DUS study with reconstruction of a "virtual cystoscopy" from the 3DUS data set using gradient light shaded surface rendering (Voluson Expert Pro, Kretztechnik/GE, Zipf Austria). The results were then evaluated for artifacts and diagnostic resolution and compared to routine transurethral cystoscopy findings. Results: 3DUS acquisition with sufficient image quality for virtual cystoscopy was feasible in all patients without sedation. Patient diagnoses included intravesical polyps or ureteroceles, ureter ostium pathology, bladder wall trabeculation and bladder diverticula. In all patients the virtual cystoscopy images and findings matched the result from transurethral cystoscopy; however, a significant change in diagnosis of the initial 2DUS report was achieved only in 1 patient, though 3DUS conspicuity was higher and presentation of the findings was more convincing using 3DUS based virtual cystoscopy display. Conclusion: 3DUS based virtual of the urinary bladder is feasible in children. It offers a great visualization tool that matched transurethral cystoscopy findings in all patients and has the potential to reduce diagnostic cystoscopies in those pediatric patients who do not need biopsy or endoscopic therapy. Sonographically guided reduction of meconium ileus in preterm neonates M. Riccabona, M. Haim, J. Kutschera; Graz/AT (michael.riccabona@meduni-graz.at) Purpose: To report our preliminary experience with ultrasound (US) guided reduction of meconium ileus in preterm neonates. Methods and Materials: 5 preterm neonates with bowel obstruction due to meconium ileus underwent bedside enema reduction using US (Sequoia 512, Acuson/ Mountain View, USA) for guidance of the prewarmed and saline diluted Jopami-ro® (Schering, Berlin, Germany) enema, as they were too ill to be brought to the fluoroscopy suite for the conventional fluoroscopic procedure. Results: In all 5 patients US managed to guide the transrectal catheter into the sigmoid or descending left colon for effective and gentle instillation of the enema. With repeated enema instillation up to 40 mls, all 5 meconium plugs were dissolved and obstruction was relieved, even in a child that later demonstrated a distal small bowel stenosis and on the initial US was thought to show an unused microcolon. No complications were observed during and after these procedures. Conclusion: US guided reduction of meconium plugs appears to be a safe and effective method that reduces radiation to these preterm babies and can easily be performed at the patient's bed side in the NICU. Purpose: Evaluation of a flat-panel X-ray detector, based on amorphous selenium with respect to skeletal radiography. Conventional images were compared with digital radiographs using identical and reduced radiation doses. Methods and Materials: 84 consecutive patients were studied prospectively using conventional screen-film radiography (detector dose 2.5 microGy). Digital images were taken from the same patients with detector doses of 2.5, 1.25, 0.625 and 0.3125 microGy, respectively, using a flat panel detector with implemented Ptone/EVP image processing software. The active-matrix detector had a pixel size of 35 x 43 cm, a matrix of 2.5 x 3K, and a pixel size of 139microm. All hard copies were presented in a random order to five independent observers, who rated image quality according to subjective and objective quality criteria. Results were assessed for significance. Results: A statistically significant preference for digital over conventional images was revealed for all quality criteria, expect for over-exposure (detector dose 2.5 microGy). Digital images with a dose reduction of 25% showed a statistically significant supremacy compared to screen-film technique. The screen-film radiography was significantly inferior to flat-panel detector-technique at 50% dose reduction regarding bone cortex and trabecula, contrast and overall impression. Conclusion: Due to its high detective quantum efficiency and dynamic range compared with traditional screen-film systems, a dose reduction of up to 50% and more is possible by using a flat panel detector with implemented P-tone/EVP image processing software without loss of image quality. Radiographers' professional knowledge related to good patient care B.T. Andersson 1 , Å. Purpose: Many patients cannot raise their arms when undergoing CT examination. In these cases, the correct diagnosis cannot be obscured by the resulting artifact. We introduced an easy scanning method to solve this problem. We arranged the patient's CT examination, their arms are not put on the body side but both arms are arranged in front of the body. The artifact was evaluated with standard deviation (SD) of CT number. Results: When the arms arranged in side of the body, SD of CT number was 19.40. The arms arranged in front of the body, SD was 9.57. The artifact from the both arms was able to be decreased by new patient positioning. Conclusion: It was a cause of the artifacts that the both arms and the spine existed straight. This originates in a specific direction the fact that the amount of the projection data decreases. The both arms and the spine become not straight by putting the both arms on the front of the body. As a result, the artifact has decreased. What should radiographers know about contrast agent induced nephrotoxicity G.A. Marshall; Lancaster/UK (g.marshall@ucsm.ac.uk) Purpose: Radiographers work with contrast agents routinely, and although much has been written on this topic, little has been focussed particularly at radiographers. This is becoming increasingly important as, via role extension, radiographers especially in the U.K., are increasingly performing, under written schemes of work, that were previously designated medical tasks e.g. intravenous injection of contrast agents. Consequently there is a need for radiographers to receive further education regarding these agents, and it is incumbent on them to understand the possible deleterious effects of these drugs. Methods and Materials: An extensive literature review was performed to establish key facts re contrast agent induced nephrotoxicity (CIN); i.e. incidence, pathogenesis, cost implication, protocols to reduce CIN, as this information would be valuable to injecting radiographers. Results: The reduction of CIN via the identification of patients at high risk, consideration of the dose and osmolality of contrast agent injected, water and sodium intake regimens peri-procedurally, the role of pharmacological agents and the use of alternative modalities will be evaluated. as defined by the local phase. The M-vectors were allowed to precess and the complex sum over all M was computed and assigned to L. Simulated images L (x, y, z) were computed for different imaging geometries (in-plane resolution, slice thickness, slice orientation), correction methods (z-shimming) and echotimes. Results: Simulated and measured EPI images showed very similar patterns of signal loss. Acquisition schemes could be optimized in terms of SNR and BOLD sensitivity= TE*Signal, based on a single phase map. The method is scalable to other field strengths and echo times than those used for the acquisition of the 3D phase-map. Signal behavior can be investigated in the presence of static magnetic field inhomogeneities and as a function of echo time and other acquisition parameters. This provides a tool for off-line optimization of scan parameters for, e.g., BOLD-fMRI. Results: Distance measurements: mean accuracy to 1/50 th of a pixel (dr/r) at 1% low-contrast edge. Area/volume measurements: mean accuracies to 1/30 th and 1/10 th of a pixel (dr in 2prdr/pr 2 ) in area and volume (volume=area*slice-thickness) at high and 1% low-contrast edges respectively. MTF at high-contrast edge improves from 5.83 lp/cm to 13.90 and 15.03 lp/cm at 10% normalised modulation level after 5x and 7x magnifications and enhancement. Low-contrast resolution: detected 0.3% low-contrast to 2 mm rod at 300 mA, 120 kVp, 10 mm slice; random noise effect on the enhanced linear edge-profile was reduced as indicated by da 1 and da 0 in Y=(a 2 ± da 2 )X 2 +(a 1 ± da 1 )X+(a 0 ± da 0 ) from 57.50% to 12.40% and 1.47% to 0.38% respectively by a dynamic filter. Purpose: Dear-Mama is a project funded by EU FP5 programme for a period of 4 years to develop digital imaging X-ray machine, for general radiology and mammography, that reduces the dose, and increases the image contrast. Methods and Materials: Based on direct capture of X-ray photons with room temperature solid-state pixel detectors coupled to photon counting readout electronics, Medipix2 chip, via bump bonding. Medipix2 chip is a matrix of 256x256 pixels (55 microns pixel size). The effective area of one Medipix2 assembly is 14.08x14.08 mm. Medipix2 assemblies are mounted on small PCBs, later mounted on large PCB that we call the "Motherboard". Two different high voltage power supplies and two different X-ray tubes will be used accordingly for each type of clinical evaluation. The clinical trials will be carried out for mammography at UDI-AT (Barcelona), chest radiology at Hôpital d'Efant Armand Trousseau (Paris), and osteoporosis at radiology department of Vienna medical school. Results: Allows discrimination of signal after amplification and pulse shaping, from the intrinsic detector noise with the electronics. This leads to superb signal to noise and a high contrast image. This implies that with little X-ray dose one can achieve a very good image with practically zero electronic noise. One frame is captured and displayed in less than 10 seconds and then sent to the PACS. The total active area of the sensor is 18x24 cm. Conclusion: In the presentation we will discuss the overall performance of the X-ray machine and the outlook for an industrialization plan. Evaluation of the Medipix2 read-out system capabilities for applications in X-ray computed tomography M. Quattrocchi 1 , R. Avitabile 2 , N. Belcari 1 , M.G. Bisogni 1 , C. Carpentieri 1 , E. Cicalini 1 , A. Del Guerra 1 , P. Delogu 1 , M.E. Fantacci 1 , G. Mettivier 2 , M.C. Montesi 2 , D. Panetta 1 , V. Rosso 1 , P. Russo 2 , A. Stefanini 1 ; 1 Pisa/IT, 2 Napoli/IT (mariagrazia.quattrocchi@pi.infn.it) Purpose: The aim of this work is to evaluate the capability of the acquisition system based on the Medipix2 read-out chip for future application in computer tomography (CT). The X-ray source has a W anode and a 30 micrometers focus size. The imaging system is based on the Medipix2 read-out chip, bumpbonded to a 300 um and 700 um silicon pixel detector. The read-out chip geometry is a matrix of 256 x 256 cells, 55 x 55 um 2 . The detection system operates in single photon counting mode and is able to detect low contrast object (less than 1%). Results: It has been initially investigated how to optimize the equalization process of a planar image for a uniform and a variable thickness phantom by varying the flat field acquisition conditions. Due to the possibility of the detection system to apply a variable energy windowing, the contrast resolution as a function of the energy has been studied. For a given thickness and material, the contrast resolution has a minimum at a different energy. This device is now being tested in CT acquisition. The improvements of the planar image quality should be reflected in a better quality of the reconstructed CT images. The CT image quality will be investigated as a function of different energy windows. A comparison with existing CT systems will be made. In order to evaluate the system performances at low doses, a comparison between analytical and iterative reconstruction algorithms will be also presented. Purpose: CT reconstruction FOV is 50 cm, which is much smaller than the 70 cm for PET. When CT images are used to perform attenuation correction, significant artifacts result when scanned objects are outside 50 cm. Although H-L consistency is used to perform corrections for multi-slice CT, it cannot be applied to VCT due to large cone angles. The total missing attenuation is first calculated based C D E F B A G its continuity over views. The projection distribution is estimated using properties that the projection profile does not change abruptly as a function of angle. When combined with the boundary projection values, the size and location of a water cylinder best fitting these parameters are obtained. To ensure the existence of the projection first derivative, the estimated projection is blended with the mirror image of the measured projection at the boundary. Results: For evaluation, various phantom experiments were conducted. Phantoms include oval cylinder with angled Teflon rods, GE Performance, 48 cm poly, and shoulder phantom. They were positioned near the edge of 70 cm FOV. Both axial and helical scans were collected on a 64-slice scanner. Results have shown that the proposed algorithm produces nearly perfect reconstruction for portions of the object inside 50 cm, while the shape of the object outside are nicely restored. A new FOV expansion algorithm is proposed. Unlike previously proposed algorithms, it makes use of the continuity of the total attenuation and gradual change of the projection profile. Extensive phantom experiments have demonstrated the efficacy of the proposed algorithm for PET-VCT applications. Inter-and intra-radiograph reproducibility of bone mineral density (BMD) and metacarpal index ( . The histopathological diagnoses were adrenocortical adenoma (n = 26), adrenocortical cancer (n = 13), adrenocortical hyperplasia (n = 8), pheochromocytoma (n = 6), metastasis (n = 3) and tumours of non-adrenal origin (n = 19). The sensitivity and specificity for MTO-PET in proving adrenocortical origin of the lesions were 89% and 96%, respectively. Non-adrenocortical lesions were all MTO-negative. PET measurements using standardized uptake values (SUV) could distinguish adrenocortical tumours larger than 1 -1.5 cm from normal adrenocortical tissue. Five PET examinations proved false negative -two profusely necrotic adrenocortical cancers and three benign adrenocortical lesions with a size (1 cm) near the resolution limit of the PET camera. One false positive result of distal metastasis to a groin lymph node was noted in a patient with a leiomyosarcoma, which by contrast was MTO-negative. Conclusion: MTO-PET is a specific and sensitive method for diagnosing adrenocortical tumours. MTO-PET is useful in the imaging work-up of adrenal incidentalomas and may be beneficial for examination of patients with primary aldosteronism. In adrenocortical cancer extensive necroses may produce false negative results which hamper the imaging results in these patients. Purpose: Colorectal cancer is one of the leading cancers in western countries. To provide whole-body tumor TNM-staging in a single session, PET/CT protocols for whole-body tumor staging and CT-colonography were integrated into one examination. Methods and Materials: 14 prospective patients underwent whole-body PET/ CT after water-based bowel distension and pharmacological bowel relaxation. PET/CT data was acquired using oral and intravenous contrast agents. PET/CT data sets were read by an experienced radiologist and a nuclear physician in consensus, and histopathology from conventional colonoscopy as well as surgical histopathology served as the standard of reference. Results: Integration of CT-colonography into whole-body PET/CT staging proved feasible and was performed without technical errors in all patients. PET/CTcolonography revealed 18/19 colonic lesions detected by conventional colonoscopy, only one flat lesion was missed. Additionally, 6 extracolonic tumor sites were detected with the integrated PET/CT exam. Based on the imaging findings 11 patients were treated surgically, whereas three received chemotherapy. For the surgically treated patients PET/CT revealed correct T-stage in 8/11 patients and a correct N-stage in 9 patients. Conclusion: Whole-body PET/CT with integrated CT-colonography is feasible and provides accurate results for TNM staging in patients with colorectal cancer. Therefore it might be used as an "all-in-one" staging procedure, especially in patients with incomplete colonoscopy. Purpose: To study the role of PET/CT in changing therapy in patients with primary or recurrent pancreatic cancer. A retrospective analysis of prospectively collected 18 FDG-PET/CT scans in 24 patients was performed. All patients underwent PET/ CT for staging primary pancreatic tumours or for recurrent disease. All PET/CT scans were assessed by a radiologist in cooperation with a nuclear medicine physician. Pancreatic lesions were interpreted based on visual assessment and by calculation of a semi-quantitative evaluation (SUV). These results were correlated with pathologic diagnosis and/or clinical course. In each patient therapeutic change was assessed. Results: In the staging group (n = 13), PET/CT changed therapy in 6 patients (46%). One patient was correctly down-staged by PET/CT and underwent surgery with curative intent. Five patients were correctly up-staged (liver metastases) and surgery was cancelled. In 7 patients (54%) PET/CT didn't change therapy. In 2 patients PET/CT was unable to discriminate between pancreatitis and malignancy. Four patients underwent surgery because PET/CT showed operability, but vascular in-growth (n = 1), liver (n = 2) or peritoneal (n = 1) metastases appeared intra-operatively. In 1 patient PET/CT suggested liver metastasis but peroperatively no lesion was found. In the recurrence group (n = 11), PET/CT changed therapy in 10 patients (91% They were already receiving intramuscular medication and were scheduled for a CT abdomen and pelvis scan. There were near equal numbers of both genders (age ranged from 21-87 years). Prior to the CT scan, the intramuscular medication and 1 mL of air was administered into the upper outer quadrant of the buttocks using a standard 23 gauge, 30 mm needle. The CT scans were analyzed by 2 independent radiologists establishing the location of the air bubble. Body mass index (BMI), gender preponderance, distance to injection site, and thickness of subcutaneous fat and muscle were also calculated. Results: Only 12 (25%) received an intramuscular injection. There was not a linear relationship with BMI. However, most of the patients who received subcutaneous injections were female, had an average fat thickness of more than 2 cm and an average muscle thickness of less than 3 cm. In contrast, those who received an intramuscular injection were all male, had an average fat thickness of less than 2 cm and an average muscle thickness of more than 3 cm. Conclusion: 75% of patients did not actually receive an intramuscular dose in the gluteal region. The efficacy of intramuscular injections is related to gender as well as subcutaneous fat and muscle thickness. The results of this study may have an impact on the administration of intramuscular injections. Purpose: To evaluate the pennation angles of lower limb muscles (vastus lateralis and medial head of gastrocnemius) in trained track and field athletes with HR-US to investigate whether the arrangement of muscle fascicles is related to different muscle activities (aerobic vs. anaerobic). Methods and Materials: Thirty-six track and field athletes trained in anaerobic (100-meters runners) and aerobic (marathon runners) muscle activities were evaluated at the end of their training session. In each subject, HR-US evaluation of the vastus lateralis (VL) and the medial head of the gastrocnemius (MHG) was performed on both limbs by means of a high-resolution broadband linear array transducer (12-5 MHz). Muscle pennation angles were calculated on scanning planes oriented along the long axis of the muscle. Results: Athletes performing anaerobic training showed greater fascicle angles (VL, mean 23°; MHG, mean 24°) than the aerobic trained ones (VL, mean 16°; MHG, mean 18°). Conclusion: HR-US evaluation of pennation angles is an easy technique to achieve information on the architecture of muscles. Our data indicate a significant difference of orientation of muscle fascicles between trained aerobic and anaerobic athletes. Perfusion maps of the forearm musculature were acquired in eight healthy volunteers (24 to 42 years old) before and after intense exercise. All measurements were performed in a 3.0 Tesla whole-body MR unit in combination with an 8-channel head coil. Pulsed arterial spin labeling and signal detection were performed with an adapted FAIR-trueFISP technique. Quantitative perfusion maps were calculated on a pixel-by-pixel basis by means of the extended Bloch equations. Results: The acquired perfusion images with an in-plane resolution of 1 mm showed no significant distortions or blurring. Perfusion time curves could be acquired with a temporal resolution of 6.4 seconds. Maximum perfusion in the musculature was found approximately two minutes after exercise, reaching values of up to 220 ml/min/100 g tissue with good delineation between the active muscles and the musculature not involved in the exercise. Purpose: The aim of this study was to evaluate the imaging parameters necessary to perform DT-MRI and fibertractography at 1.5 T and 3 T in normal human skeletal muscle using different b-values and to calculate and to determine a number of quantitative parameters to characterize diffusion anisotropy in organized tissue. Five healthy adult volunteers were included in this study. All volunteers underwent MR imaging at 3 T and 1.5 T (GE Healthcare, Milwaukee, WI) within five days. Imaging on both MR systems was performed utilizing a standard knee coil and using the same imaging protocol. Three routine diffusionweighted spin-echo EPI pulse sequences (with different b values) were applied to collect a series of 2D images through the calf muscles of each subject. The image analysis and tensor calculations were performed using the DTIStudioprogram. The eigenvalues (Λ 1 ,Λ 2 ,Λ 3 ), the mean diffusivity 〈D〉, the fractional anisotropy (FA), the relative anisotropy (RA) and the volume ratio (VR) were calculated. The mean values for (Λ 1 ,Λ 2 ,Λ 3 ) and D were 1.47 ± 0.08, 1.30 ± 0.09, 1.06 ± 0.08 and 1.86 ± 0.09 at 3 T and 1.77 ± 0.07, 1.90 ± 0.09, 1.83 ± 0.08 and 2.23 ± 0.09 at 1.5 T. (All values are reported in 10 -5 cm 2 ·s -1 ; mean ± SD). There was no significant difference between calculated parameters at 3 T and 1.5 T (P < .001). The study demonstrates useful parameters to perform DT-MRI and fibertractography at 1.5 T and 3 T. DT-MRI at 1.5 and 3 T provides in vivo validation of quantitative structural analysis of human skeletal muscle. Fiber tracking at 3 T MR provides information about muscle fiber orientation and architecture of the diffusion pathways. Purpose: To characterize the MR appearance of the medial collateral ligament in asymptomatic subjects. Methods and Materials: Standard ankle MR examinations of 56 asymptomatic subjects (23-60 years) were analyzed by two musculoskeletal radiologists. For all five components of the medial collateral ligament complex the visibility, the optimal imaging plane and the signal characteristics were analyzed. The thickness of all ligaments was measured. The posterior tibiotalar and tibiospring ligament were visible in all asymptomatic subjects, the tibiocalcaneal ligament was seen in 49 (88%) subjects, whereas the anterior tibiotalar and tibionavicular ligaments were only visible in 31 (55%) subjects. The coronal imaging plane was considered to be always the best imaging plane for all components except for the tibionavicular ligament (55% coronal, 45% transverse). The anterior and posterior tibiotalar ligament had a variable signal on T2-weighted images and mean thickness of 1.4 mm and 8.2 mm respectively. On T2-weighted images the posterior tibiotalar ligament had a typically striated appearance in all subjects under 45 years (89% of the whole study population). The tibionavicular ligament had a variable signal on T2-weighted images and a mean thickness of 1.6 mm. The tibiospring and the tibiocalcaneal ligament had a low signal on T2-weighted images (75% and 78% respectively) and a mean thickness of 2.0 mm and 1.2 mm respectively. Conclusion: The posterior tibiotalar ligament was always visible with a typically striated appearance on T2-weighted images in all subjects under 45 years. The tibiospring and tibiocalcaneal ligaments are the consistently visible superficial components with a typically low signal on T2-weighted images. Correlation between X-ray and MR imaging of ankle fractures with regard to syndesmotic injury J.J. Hermans, N. Wentink, P.G.H. Mulder, A. Beumer; Rotterdam/NL (j.j.hermans@erasmusmc.nl) Purpose: To validate three fracture classification systems with respect to syndesmotic injury. In 33 patients X-rays of the ankle fractures were classified according to Weber, AO-Müller and Lauge-Hansen. Both fracture type and additional measurements of the ratio of the medial and superior clear space (MCS/ SCS) and tibiofibular overlap (TFO) were used to assess syndesmotic injury. An MCS/SCS > 1 mm and a TFO < 0 mm are indicative for syndesmotic injury. Dual TSE images in three orthogonal planes and an additional oblique plane were acquired with a 1.5 T magnet. The syndesmotic ligaments were classified as normal, thickened, possibly or definitely ruptured. Syndesmotic injury was defined as a definite rupture or avulsion of the anterior tibiofibular ligament or a complete rupture of the entire syndesmosis (anterior and posterior). The agreement score (κ) concerning the presence of syndesmotic injury between the X-ray and MR imaging was determined. Results: With Lauge-Hansen, syndesmotic injury could be assessed in all patients. Using Weber resp AO-Müller in 30% resp 9% of patients syndesmotic injury could not be assessed either because the fracture could not be classified and/or the measurements were normal. The agreement score (κ) between the three fracture classifications and MR imaging with respect to syndesmotic injury was: Weber ( recipients were examined to exclude postoperative biliary obstruction. The examination protocol included the intravenous short-infusion of 100 ml of a biliary contrast agent. CT cholangiography data was acquired with a slice thickness of 1 mm. Data sets were evaluated quantitatively by measurement of the biliary opacification, and qualitatively on the basis of a scale ranging from 1 (non-diagnostic) to 4 (excellent). The opacification was correlated with the post-operative serum bilirubin level. Results: CeMDCT-CA provided diagnostic delineation of the biliary tree in all 15 donors and seven of 11 recipients; in four recipients the degree of biliary opacification was rated non-diagnostic. The biliary opacification was generally higher in the donor collective. Four donors and three recipients presented a moderate focal biliary constriction without elevation of laboratory values indicating biliary obstruction. Six patients presented with postoperative fluid collections suggestive of perihepatic biloma, however no biliary fistula could be visualized. Conclusion: CeMDCT-CA allows for non-invasive assessment of postoperative biliary anatomy and complications in living liver donors and in the majority of liver transplant recipients. Direct visualization of biliary leakage appears to be limited to fistulae of bigger size. Preoperative T-staging of the carcinoma of the gallbladder using MDCT: In 90 cases with normal bile duct, area and caliber of the bile duct were prospectively measured by 1.5-T MR system using 3D T2 image of HASTE sequence. The provided image voxel was 1.0x0.9x1.0 mm. The following sites were measured; upper boarder of pancreas (A), common hepatic duct (B), left hepatic duct (C) and right hepatic duct (D). Among the 90 subjects, 10 cases were measured for main pancreatic duct using the same procedures. The pancreatic duct was measured in head (E) and neck (F) of the pancreas. Results: The measured areas (mm 2 ) were 40.4 ± 20.1 in A, 43.2 ± 24.0 in B, 27.9 ± 14.9 in C and 27.4 ± 16.0 in D, respectively. The diameters (mm) were 6.9 ± 1.9 in A, 7.1 ± 2.2 in B, 5.7 ± 1.6 in C and 5.6 ± 1.8 in D, respectively. There was no correlation between age and size of the bile ducts. As for the pancreatic duct, the areas were 10.8 ± 2.9 in E and 8.8 ± 2.8 in F. The calibers were 4.4 ± 0.8 in E and 3.8 ± 0.7 in F. Conclusion: MRC provided useful parameters to define standard size of only bile duct. According to this results, upper limit of normal value of intrahepatic bile ducts were 8.9 mm (left) and 9.2 mm (right). Purpose: To determine whether Mn-DPDP-enhanced MRC could improve the diagnostic confidence provided by conventional T2w MRC in patients with suspected biliary complications after liver transplantation. Methods and Materials: Thirty liver transplant subjects with suspected biliary complications (abnormal liver function tests and/or biliary ductal dilation at ultrasound) underwent MRC at 1.5 T (Signa, GE Healthcare). After acquisition of axial T1w and T2w sequences, conventional MRC was performed through a respiratory-triggered, thin-slab (2 mm), heavily T2w FSE sequence and a breathhold, thick-slab (10/50 mm), single-shot T2w sequence in the coronal plane. In each patient coronal T1w fat-suppressed breath-hold 3D gradient-echo imaging (FAME) was also obtained 30-40 minutes after mangafodipir trisodium (Mn-DPDP, Teslascan®) intravenous administration. All source images and 3D MIP and volume rendered reconstructions were reviewed in conference by two observers that were asked to determine presence and site of biliary tract complications. Imaging results were correlated with direct cholangiography (n = 21), surgery (n = 4) and/or follow-up (n = 5). Results: In 5 patients no biliary tract abnormality was identified. The remaining 25 subjects were affected by ischemic-type biliary lesions (n = 6), non-anastomotic strictures (n = 6), anastomotic strictures (n = 8), inflammatory ampullary stenosis (n = 2), intrahepatic biloma (n = 1) and stones (n = 11). Mn-DPDP-enhanced MRC tended to out-perform conventional T2w MRC in delineating 3 slight anastomotic stenoses and 8 out of 12 strictures involving the biliary tree of the graft. Additionally, it allowed the correct diagnosis of an intrahepatic biloma. Seven surgical clips were embedded in a gel phantom and positioned at predefined distances from a fluid-filled tube designed to simulate the common bile duct. In-vitro MR imaging protocol included T1-w GRE, RARE, HASTE, and 3D TSE T2-w sequences on a 1.5 T and a 3 T MR system. Within each magnet class, the maximum clip distance was noted where susceptibility artifacts obscured the fluid-filled tube. Susceptibility artifact size was also calculated for each sequence within each magnet class. In-vivo analysis included 21 patients with a history of cholecystectomy who underwent MRC at 3 T. Presence of susceptibility artifacts was noted for each sequence. MRC image quality was also graded with specific emphasis on whether image quality was impaired by susceptibility artifacts. Results: In-vitro, mean area of susceptibility artifacts was 104 mm 2 on the 3 T MR system and 75 mm 2 on the 1.5 T MR system. While surgical clips within a 2 mm range impaired visualization of the fluid-filled tube on 1.5 T imaging, this range increased to 4 mm on 3 T MR imaging. In-vivo, MRC image quality was impaired by susceptibility artifacts in three of 21 cases by obscuring adjacent biliary ductal branches. Conclusion: Susceptibility artifacts from surgical clips are larger with ultra-high field 3 T MR imaging than on standard 1.5 T MR imaging. Adjacent biliary ductal branches will be obscured by these susceptibility artifacts in selected cases. Purpose: To evaluate subjects affected by esophageal motility disorders using Dynamic TURBO-FLASH sequence during positive oral contrast agent administration and to compare MR findings with videofluoroscopy and manometry results. Methods and Materials: Thirty patients who previously underwent videofluoroscopy and manometry for suspected esophageal motility disorders were studied using a 1.5 T magnet (Magnetom Vision: Siemens, Erlangen, Germany) equipped with phased-array coil. Imaging protocol was based on the employment of Dynamic TURBO-FLASH sequence on sagittal and axial planes to achieve motility evaluation (TR 416 mS; TE 1.2 mS; Flip Angle 8°; Thickness 20 mm; FoV 350; Matrix 90x128; N. acquisition 45; TA=25 sec; Position: prone) during oral administration of positive contrast agent (yoghurt + Gd-DTPA 0.5 M, 1:100 boluses). Images were reviewed by two radiologists with long experience in reporting barium fluoroscopy. MR imaging findings were compared with manometry and videofluoroscopy results. Results: Good quality images were obtained in all patients, with adequate lumen contrast. Frame rate was 4 fps, allowing a near real-time evaluation of esophageal motility. Comparing with videofluoroscopy and manometry, achalasia was correctly diagnosed in ten cases, "corkscrew esophagus" in six and diffuse esophageal spasm in nine; in three cases esophageal spasm was incorrectly diagnosed as reflux; in two cases there was were evidence of normal findings. Conclusion: MR fluoroscopy approach represents a promising radiation-free modality in the evaluation of esophageal motility disorders. Exists elevate agreement between MR imaging findings and standard modality findings. pudding with/without paramagnetic contrast medium. Thirty serial images were obtained using four real-time trueFISP (AcqT/slice 0.6 s) and three cine turbo-FLASH (AcqT/slice 0.2 s) sequences with varying parameters (field-of-view, matrix, slice thickness, etc) to assess the best spatial and temporal resolution. Functional and anatomical information obtained with the different acquisitions and the clarity of swallowing-related motion were evaluated. Additionally, signalto-noise ratios (SNRs) were measured in each used sequence. Results: Among turboFLASH sequences, the 5 mm acquisition showed the best compromise between SNR and spatial resolution, whereas 8 mm sequence revealed to be the optimal trueFISP acquisition. Considering the most performing trueFISP and turboFLASH sequences, the last provided the best temporal and contrast resolution, with a sufficient spatial resolution during motion. Conclusion: High-speed kinetic MR imaging is a very promising tool for evaluating deglutition. Real-time trueFISP is well suited for this purpose, providing good image contrast and sufficient temporal and spatial resolution, but turboFLASH acquisition currently provides the best anatomical and functional information for the evaluation of the swallowing process. Purpose: The application of radiopaque markers with conventional abdominal overview and nuclear medical procedures are easy methods assessing colonic transit time, but both methods repeatedly expose patients to radiation providing only insufficient transit characterisation. Therefore, the aim was to develop an assessment for the determination of colonic transit time using functional cine MR imaging. In a pilot study 8 healthy volunteers (4 women, 4 men, mean age 37.5 ± SD14.4y) without anamnesis of gastrointestinal diseases were examined using a 1.5 Tesla system with a protocol for respiratory gated 3D T1 and T2w-coronal oriented sequences with 12 minute duration after a 3-day diet for the standardization of the gastric-colonic-content. The volunteers swallowed 5 cups filled with Gd-NaCl-ratio prior to the first exam which was repeated after 3, 6, 12, 24 and 48 hours. The resulting MR-images were searched for the appliqued capsules. Similar results were obtained to those described for conventional methods in current literature. The male transit time was significantly shortened compared to women (28 ± 2h vs. 48 ± 3h). Conclusion: Functional cine MR imaging is an adequate method for the assessment of colonic transit time offering the advantages of non-existing radiation, minor expenditure of time and the possibility of 3D image reconstruction. In terms of future adoption in clinical routine and for surgical planning an automatic software tool for evaluation is in development. Endosonographic prediction of external anal sphincter function in faecally incontinent women M. Purpose: To evaluate the position, configuration, dimensions of the thoracic duct with fat-suppressed T2-weighted sequences and MR lymphography (MRL) and the clinical significance. Methods and Materials: MRL by heavily T2-weighted techniques and fat-suppressed T2-weighted sequences was performed in 139 patients. Forty-two patients were excluded because of poor reconstructional images. On T2-weighted images, the position of thoracic duct with reference to the vertebrae was defined, and the coronal and sagittal MIP reconstructional images were obtained for measurement of the dimension. The differences of the parameters in the groups including a control group, portal hypertension group and choledochal obstruction group were compared. Results: The cisterna chyli was shown in 91.37% of patients on fat-suppressed T2-weighted images. Most were located at the vertebral level between L1 and L2 in 80.31%, and in the midline of the vertebrae in 75.59%. On the MIP images, the thoracic duct was shown in 69.78%, and the most common configuration of the cisterna chyli was tubular or saccular. The mean longitudinal, anteroposterior and transverse diameters of the cisterna chyli were 26.04 (± 10.45)mm, 5.08 (± 2.08)mm, and 5.02 (± 2.13)mm respectively. The anteroposterior and transverse diameters of the thoracic duct were 3.08 (± 0.44)mm and 3.58 (± 0.13)mm in control group. There was a significant difference of the transverse diameter of the thoracic duct between control and disease groups (F=5.638, P=0.005). Conclusion: Noncontrast-enhanced MR techniques may be used to detect the thoracic duct, and portal hypertension may influence its dimension. Purpose: To test the feasibility of in-vivo T1 measurements of the entire lung parenchyma, to quantify T1 measured in the entire lung, and to determine the effect of inspiration and expiration on the measured T1 values. Methods: Ten healthy volunteers underwent MR imaging investigation of the lung using a 1.5 T scanner. Image acquisition was based on a SNAPSHOT-FLASH sequence. The acquired series of images were used for calculation of T1 relaxation time on a pixel-by-pixel basis. For each volunteer image acquisition was performed both in inspiration and expiration. Results: The average of the mean T1 values was in inspiration 1199 ± 117 ms and in expiration 1333 ± 167 ms. T1 values were higher in expiration than in inspiration in all 10 volunteers. The difference was statistically significant (P=0.005). In inspiration, the difference between the average dorsal T1 value and the average ventral T1 value was 81 ms with higher ventral values. This difference was statistically significant (P=0.022). In expiration, the difference between the average dorsal T1 value and the average ventral T1 value was 33 ms, with higher dorsal values. This difference was not statistically significant. Conclusion: Our study documents the feasibility of T1 measurements of the entire lung parenchyma. These measurements resulted in significantly lower T1 values in inspiration, and in significant differences between inspiratory T1 values from the ventral and the dorsal parts of the lung. Our findings underline the need for a close monitoring of the inspirational status during MR examinations of the thorax, notably when T1 measurements of the lung are performed. Oxygen-enhanced MR imaging of the lung using a pneumotachograph for respiratory monitoring F. Molinari, G. Corbo, S. Valente, T. Pirronti, L. Bonomo; Rome/IT (fmolinari@rm.unicatt.it) Purpose: To demonstrate the benefits of respiratory monitoring with a pneumotachograph in oxygen-enhanced MR imaging of the lung. Methods and Materials: 7 healthy-volunteers (age-range: 26-32) were imaged on a 1.5 T MR-scanner. To evaluate the impact of monitoring respiration with a pneumotachograph, 5 series of IR-SSFSE images (TE=28.2 ms; TI=1200 ms) were obtained in each subject during the ventilation-paradigm room air-100% oxygen-room air: a) Series no.1, respiratory-triggered [R-tr]; b) Series no.2, cardiac-triggered [C-tr]; c) Series no.3, cardiac-triggered with respiratory-monitoring based on a pneumatic-belt [C-tr/BELT-m]; d) Series no.4, cardiac-triggered and respiratory-monitoring with a pneumotachograph [C-tr/PNTC-m]; e) Series no.5, respiratory-monitoring with a pneumotachograph [PNTC-m] . Standard deviations of the lung (SIvar) and diaphragm mismatch (DM) were measured on room-airimages. The relative SI-change (∆SI) was computed from room-air and oxygenenhanced-images. Parametric maps were obtained from cross-correlation analysis of the ventilation-paradigm. Mean correlation-coefficients (cc) and the percentage of activated pixels over the lung (Act%) were calculated from these maps. All five parameters were compared among the five series (Friedman-ANOVA, Dunn's post-hoc test). In series no.4 [C-tr/PNTC-m], DM and SIvar were significantly lower than in respiratory and cardiac-triggered series (DM= 4.7% vs 12.4% and 14.4%; SIvar = 7.6% vs 14.1% and 12.8%). In the same series cc and Act% were also significantly higher than in series 1 and 2 (cc= 0.86 vs 0.70 and 0.68; Act%= 71.3% vs 44.7% and 41.2%). ∆SI was not significantly different among all series. Conclusion: Continuous monitoring of breathing has potential for: a) reducing signal variability and diaphragm mismatch in IR-SSFSE images; b) improving the quality of oxygen-enhanced ventilation-maps of the lung. Intrapulmonary ³He gas distribution depending on bolus position K. Purpose: Dynamic ³He-MRI of ventilation is a new method for quantitative assessment of intrapulmonary gas distribution. This study intends to investigate the temporal and regional differences in intrapulmonary distribution of a gas bolus depending on its volume and application delay. Methods and Materials: 12 healthy domestic pigs were anaesthetized, intubated and mechanically ventilated. ³He boli of 100 ml, 150 ml and 200 ml were applied via a gas application device followed by room air. The 150 ml bolus was additionally applied with delays of 0, 200 … 1200 ms after initial inspiration. The mean global tracheoparenchymal delay (td90) was 120 ms in the 100 ml bolus. It was prolonged to 190 ms with increasing bolus size (150, 200 ml). Introducing a 200 ms delay, td90 was 250 ms decreasing to 140 ms with increasing delay (up to 1200 ms). In the euthanized pig the td90 was 160 ms without application delay. Dynamic ventilation ³He-MRI shows differences in inspiratory gas distribution depending on volume and delay of the contrast-gas. The increase of td90 with bolus size is probably caused by the longer application duration of a larger bolus. The euthanized pig shows accelerated gas inflow with post mortem loss of compliance of the lung. Gas application with increasing delay after initial inspiration leads to an accelerated gas inflow due to facilitated convection and diffusion by widening of airways during later inspiratory phases. Acknowledgements: The ³He project is supported by the German Research Council (DFG, FOR474) and the research program of Mainz University (MAIFOR). This work contains data from K. Hawig's doctoral thesis. Pulmonary imaging using hexafluorethane (C 2 F 6 ): Initial experience U. Wolf, A. Scholz, C. Düber, K. Markstaller, C.P. Heußel, W.G. Schreiber; Mainz/DE Purpose: The actual strategy to improve SNR in 19 F-MRI is a high number of averages using gases with short T1 such as SF 6 (1.65 ms), necessitating short echo times and high bandwidths. The aim was to improve SNR through lowering receiver bandwidths using C 2 F 6 (T1 5x SF 6 ). Methods and Materials: 3 pigs were ventilated with 70% C 2 F 6 (or SF 6 ) and 30% O 2 . FLASH projection was done during and after gas wash-in. Dynamic imaging with 1 frame per breath every 5s (TA=2s) and static imaging after completion of the wash-in (TA=10s) at three different bandwidths 200, 488, 700 Hz/pixel during an inspiratory hold was done. Furthermore, triples of 1, 3 and 100 acquisitions were taken after wash-in at 488 Hz/pixel, exclusively. Results: C 2 F 6 was clearly superior to SF 6 at all tested bandwidths. The SNR for C 2 F 6 was maximal at 200 Hz/pixel (488 Hz/pixel for SF 6 ). For the first time, subsecond 19 F-MRI of the lung proved feasible using C 2 F 6 at 488 Hz/pixel. A peak SNR of 11 was observed for a scan time of 780 ms. Conclusion: MR imaging of C 2 F 6 yielded good image quality in healthy pig lungs. In the medium bandwidth range tested here, C 2 F 6 was clearly superior to SF 6 . This is a result of the relatively low bandwidths applied here which are well scaled C D E F B A G to the T1 of C 2 F 6 . Therefore, SNR improvement through lowering of the receiver bandwidth proved to be efficient using C 2 F 6 . In summary, subsecond 19 F-MRI of the lung proved feasible using C 2 F 6 at the medium bandwidth of 488 Hz/pixel. Navigator triggered oxygen-enhanced MR imaging in patients with interstitial lung disease F. Molinari 1 , M. Eichinger 2 , F. Risse 2 , L. Bonomo 1 , H.-U. Kauczor 2 , C. Fink 3 ; 1 Rome/IT, 2 Heidelberg/DE, 3 Munich/DE (fmolinari@rm.unicatt.it) Purpose: To demonstrate a correlation between navigator-triggered oxygen-enhanced MR imaging of the lung and pulmonary function tests (PFT) in patients with interstitial lung disease (ILD). Methods and Materials: 12 healthy-volunteers (age range: 20-32y) and 10 patients (age range: 37-87y) with ILD underwent pulmonary function tests (PFTs) and oxygen-enhanced MR imaging. Static lung volumes, forced expiratory flows and diffusion capacity were obtained in all subjects. Arterial blood gases were obtained only in the patient-group. Oxygen-enhanced MR imaging was performed on a 1.5 T MR-scanner. The paradigm room-air-O2-room-air was acquired with a nonselective-IR-HASTE (TE=12 msec; TI=1200 msec). Navigator and cardiactrigger were combined in a double-triggering-method to control simultaneously cardio-respiratory motion. Image acquisition was performed in end-expiration. Inhouse-developed software was used to further reduce motion artifacts. Crosscorrelation analysis was performed in ROIs encompassing both lungs (r-value = 0.5; p-value = 0.01 as thresholds). The percentage of activated pixels over the total number of pixels in the ROIs (OAP%) was calculated from both lungs. Mean OAP%s calculated from ILD-patients and volunteers were compared (Mann-Whitney test). The correlation between OAP%s and PFTs was assessed with Spearman-test. The double-triggering method reduced artifacts from cardio-respiratory motion. Free-breathing end-expiratory acquisition was time-effective and not tiring for all subjects. Image post-processing further reduced the signal variability. The mean OAP% of patients was significantly lower than that of volunteers (61.3 vs 84.0;p = 0.04). OAP%s correlated significantly with ar terial-pO2 (r = 0.67;p = 0.02), and with arterial-O2-saturation (r = 0.72;p = 0.01). Conclusion: Oxygen-enhanced MR imaging of the lung using navigator and cardiac-triggering is an efficient imaging method and has a potential role in the functional assessment of ILD. Free-breathing non-contrast VIBE MR-angiography of the lung using ECG triggering and respiratory gating: Image quality assessment J. Zaporozhan 1 , S. Ley 1 , C. Purpose: Aim of this study was to evaluate the feasibility and image quality of a "work in progress" ECG triggered and respiratory gated volume interpolated body examination sequence (VIBE) for non-contrast MR-angiography of the pulmonary arteries. The whole lung was imaged in 36 subjects (10 volunteers, 26 patients) during quiet respiration on a 1.5 T MR scanner. The sequence was applied in a coronal orientation (TR/TE=9/3.32 msec; flip angle 8°; 1.3x1.2x2 mm 3 ; GRAPPA 2; 88 slices). In each heartbeat, the navigator for respiratory gating is executed after an appropriate trigger delay time, followed by a fat suppression pulse and the data acquisition bloc. Images were assessed for sharpness and quality of pulmonary arterial branches and thoracic structures and presence of artifacts using a 4-point-score (4-excellent to 1-insufficient/severe artifacts). Results: The mean acquisition time was 10 ± 3 min. 86% of the images were of good to excellent quality with no or minor artifacts. In 8% some parts of the lung were not diagnostic. In 6% the images could not be evaluated due to severe artifacts. The mean score for sharpness of pulmonary arteries was 3.2 and 3.6 for thoracic structures. Overall the image quality was fine, mean score 3.3. Most artifacts were found in the lung parenchyma or close to the diaphragm, overall mean score was 3.6. The new ECG triggered and respiratory gated VIBE sequence was successfully used in 86% of patients during free-breathing with a high image quality and high level of sharpness of vascular and thoracic structures. Quantitative whole tumour perfusion assessment using MDCT: Reproducibility and observer variability in advanced lung cancer V.J. Goh 1 , Q. Ng 1 , E. Klotz 2 , H. Fichte 2 , M. Saunders 1 , P.J. Hoskin 1 , A. Padhani 1 ; 1 Northwood/UK, 2 Forchheim/DE (vicky.goh@paulstricklandscannercentre.org.uk) Purpose: Perfusion techniques have been confined to a small tumour volume to date, which may not be representative of the entire tumour. We aimed to determine the reproducibility and observer variability of a quantitative whole tumour perfusion CT technique. Methods and Materials: Following IRB approval and informed consent, 10 patients with advanced lung cancer (mean size 8.3 cm) were scanned prospectively using a 16-detector row scanner. The perfusion studies, consisting of sequential volumetric acquisitions encompassing the entire tumor, were acquired following IV contrast infusion. Studies were repeated within 24 hours to assess reproducibility. All 20 studies were analysed by two independent observers. The whole tumour was analysed pixel-by-pixel using perfusion software based on Patlak analysis; median values of permeability and blood volume were obtained. Measurement reproducibility and observer agreement were assessed using Bland-Altman statistics. The mean difference (95% limits of agreement) for repeated studies was 0.07 ml/100 ml/min (-3.73 to +3.87) and 1.30 ml/100 ml (-2.63 to +5.23) for permeability and blood volume respectively; the coefficient of variation was 9.49% and 26.31% respectively indicating good reproducibility. For inter-observer agreement, the mean difference (95% limits of agreement) was 0.07 ml/100 ml/min (-1.01 to +1.08) for permeability, and 0.44 ml/100 ml (-0.59 to +0.68) for blood volume. The coefficient of variation between observers, and within observers was 3.47%, 6.34% and 3.50% respectively for permeability, and 3.64%, 3.30% and 3.95% respectively for blood volume, indicating good agreement. Conclusion: Good measurement reproducibility and inter-and intra-observer agreement can be achieved with whole tumour perfusion CT, boding well for clinical application. Purpose: To test the feasibility and accuracy of 64-slice CT coronary angiography (gantry rotation time 330 msec) in patients with acute coronary syndromes, with invasive coronary angiography and intravascular ultrasound (IVUS) as reference standard. Methods and Materials: Twenty-five patients with unstable angina (UA) or suspected acute myocardial infarction (AMI) underwent CT coronary angiography within 24 hours prior to cardiac catheterization. IVUS studies were conducted in 16 patients. Subjects in sinus heart rhythm were included, provided they were able to hold their breath for 12-14s. Patients with heart/renal failure or allergy to contrast agent were excluded. Beta-blockers were administered to patients with heart rates above 65 beats per minute (bpm). Results: 11/25 patients presented with UA and 14/25 with AMI. Beta-blockers were administered to 15 patients. Average heart rate dropped from 67 to 58 bpm. All subjects were correctly identified as normal (n = 1), single-vessel (n = 8) or multi-vessel (n = 16) disease patients. Five/25 patients were overdiagnosed with 3-vessel instead of 2-vessel disease. In the analysis of lesion characteristics performed blinded of the clinical context and IVUS results, the culprit lesion could be defined in all patients with 1-vessel disease (n = 8), in 6 patients with 2-vessel disease (n = 8) and in 5 patients with 3-vessel disease (n = 8). Conclusion: 64-slice CT in acute coronary syndromes may be helpful in identifying single-vessel versus multi-vessel disease. Assessment of the culprit lesion seems feasible based on plaque characteristics and could be enhanced by knowledge of the clinical context and clinical tests results. Purpose: The aim of this study was to evaluate the value of 64-detector row computed tomography for ruling out high-grade coronary stenoses in patients with a low to moderate pre-test probability of significant coronary artery disease. The study included 179 patients with a suspicion of coronary artery disease on the basis of atypical angina or ambiguous findings in non-invasive stress testing. All patients were examined using 64-detector row computed tomography angiography (CTA). All Patients with stenoses or uncertain lesions at CTA (n = 61) underwent invasive coronary angiography (ICA). On a coronary segmental level, the presence of high-grade (≥ 50% diameter) stenoses was compared. Results: In a total of 915 segments, CTA detected 62 significant stenoses. Thirty-four significant stenoses were confirmed by ICA, whereas 28 stenoses could not be confirmed by ICA. 22 of those 18 lesions showed stenoses < 50% at ICA whereas 6 segments were angiographically normal. Accordingly, on a coronary segmental basis, 28 false positive and 0 false negative findings resulted in a sensitivity of 100%, a specificity of 96.8%, a positive predictive value of 54.8%, and a negative predictive value of 100%. The diagnostic accuracy was 96.9%. Conclusion: Sixty-four-detector row computed tomography reliably detects significant coronary stenoses in patients with suspected coronary artery disease and appears helpful in the selection of patients who need to undergo ICA. However, grading of stenoses is still challenging and stenoses are frequently overestimated by 64-detector row computed tomography. 64-slice CT angiography reliably detects and excludes significant coronary artery stenosis on a per-patient but not on a per-vessel or per-segment basis G. Savino 1 , J.U. Schoepf 1 , P. Costello 1 , P. Zwerner 1 , T.J. Vogl 2 , C. Herzog 2 ; 1 Charleston, SC/US, 2 Frankfurt/DE (savino@musc.edu) Purpose: To investigate the diagnostic accuracy of 64-slice multi detector-row CT (MDCT) in the assessment of coronary artery disease (CAD). Methods and Materials: 55 patients with suspected CAD underwent both, conventional coronary catheter angiography and contrast enhanced, retrospectively ECG-gated 64-slice MDCT of the coronary arteries. Two experienced observers analyzed all MDCT-scans for signs of CAD (degree of stenosis = 0%; < 50%; < 70%; < 99%; 100%). Diagnostic accuracy compared to cardiac catheter was determined on a per-segment, per-vessel and per-patient basis. Results: Globally, all CT coronary angiograms (55/55) were considered of diagnostic quality. On a per-segment basis, 92.2% (762/826) of all segments could be clearly evaluated. In 7.8% of segments, image quality was compromised either by misregistration (15.6%), motion artefacts (31.2%) or small vessel size (53.1%). The correlation coefficients for the detection and grading of stenosis were r = 0.65 on a per-segment, r = 0.83 on a per-vessel and r = 0.88 on a per-patient basis. Background: Recent improvements in computed tomography have expanded its application to coronary artery imaging. Using a flat panel CT research system (GE Global Research), a spatial resolution of 0.2 mm can be achieved, which raises the possibility of depicting the intramyocardial coronary circulation of human hearts. The post mortem flat panel CT examination of excised human hearts displays the coronary arteries lumen throughout all branches. The coronary arteries were filled with a barium sulphate contrast agent to achieve an intraarterial opacification that is equivalent to results obtained with venous injection of iodinated contrast agents in humans. The sixth generation coronary artery branches could consistently be discerned, and intramyocardial vessel segments were clearly visualized. Discussion: The consistent visualization of intramyocardial coronary artery segments opens new avenues for the noninvasive diagnosis of cardiac disease. Although the method is presently limited to a static situation, it may serve as proof of principle that new CT techniques with superior spatial resolution are capable of investigating the intramyocardial coronary circulation. Also, research on the role of vascular growth factors for arterial myocardial architecture would potentially benefit from the new technique. Multi-slice CT-angiography of the coronary arteries in patients with atrial fibrillation A. Kovacs, C. Probst, T. Sommer, P. Naehle, H. Schild, S. Flacke; Bonn/DE (attila.kovacs@ukb.uni-bonn.de) To determine the optimal reconstruction window during the cardiac cycle in patients with atrial fibrillation (AF) for each of the three major coronary arteries using a new frequency adapted reconstruction algorithm. Methods and Materials: 20 patients with permanent AF were imaged on a 16 slice scanner (slice coll.16x0.75 mm; rotation time 0.42 s; 140 kV; 380 mAs). The patients had not received previously any drugs for heart frequency regulation. Acquisition was started after bolus tracking of a biphasic bolus of 120 ml Ultravist 370® injected intravenously. Each coronary segment was reconstructed at 0%-90% of the cardiac cycle in increments of 10%. Two blinded independent readers assessed the image quality of the coronary segments in terms of visibility and artifacts (five point rating scale 1=very poor up to 5=excellent) and the degree of stenosis (five point rating scale 1=0%, 2=1%-49%, 3=50%-74%, 4=75%-99%, 5=100%) on axial slices, multi-planar reconstructions and three-dimensional volume rendered images. The heart rate during the examination ranged between 42 and 156 beats per minute, the average heart rate was 78 bpm ± 23. Each of the two readers evaluated 300 segments in 20 patients. Visualisation of all coronary artery segments was superior (p < 0.01) at 40% (mean score of the image quality 2.79) as compared to the standard diastolic reconstruction window at 80% (image quality 2.33). The second best image quality (2.57) was acquired at 0% of the cardiac cycle. Conclusion: Using a frequency adapted delay algorithm with the choice of an endsystolic reconstruction window provided diagnostically valuable images in patients with AF. Purpose: To evaluate the influence of ejection fraction (EF), stroke volume (SV), heart rate, and cardiac output (CO) on coronary artery opacification with 64-slice computed tomography (CT). Methods and Materials: 60 patients (22 women, 38 men; mean age 61.9 ± 10.5 years) underwent retrospectively electrocardiography-gated 64-slice CT coronary angiography. Ten data sets were reconstructed in 10% steps of the R-R interval for quantification of left ventricular EF, SV, and CO using software capable of semi-automated contour detection. Attenuation values were measured and contrast-to-noise ratios (CNR) were calculated in the proximal right coronary artery (RCA) and left main artery (LMA). CNR was defined as difference of mean attenuation in the vessel from the mean attenuation in perivascular tissue divided by image noise in the ascending aorta. Results: Mean EF during scanning was 61.5 ± 12.4%, SV was 63.2 ± 15.6 ml, heart rate was 62.5 ± 11.8 bpm, and CO was 3.88 ± 1.06L/min. No significant correlation was present between the EF and heart rate and the attenuation and CNR in both coronary arteries. A significant negative correlation was found in both arteries between SV and attenuation (RCA: r=-0.26, P < 0.05; LMA: r=-0.34, P < 0.01) and between SV and CNR (RCA: r=-0.26, P < 0.05; LMA: r=-0.26, P < 0.05). Similarly, a significant negative correlation was found between the CO and attenuation (RCA: r=-0.42, P < 0.05; LMA: r=-0.56, P < 0.001) and between the CO and CNR (RCA: r=-0.39, P < 0.05; LMA: r=-0.44, P < 0.001). The actual hemodynamic status of the patient influences the coronary artery opacification with 64-slice CT, in that vessel opacification decreases as SV and CO increases. Prospective ECG-pulsing and retrospective ECG-gating in 64-slice cardiac MDCT: Evaluation of image quality of coronary arteries in 37 patients J. Schmitt 1 , K. Koenig 1 , N. Abolmaali 2 , H. Ackermann 3 , G. Görge 1 , G. Brill 1 ; 1 Saarbrücken/DE, 2 Dresden/DE, 3 Frankfurt am Main/DE (jschmittj@web.de) To evaluate the quality and dose differences between prospective pulsing and retrospective gating in coronary MDCT. Methods and Materials: 37 consecutive patients in a stable clinical condition and sinus rhythm were examined. Patients with heart rates above 60 b.p.m. received 100 mg atenolol p.o. and up to 50 mg esmolol i.v. Expiratory and contrast enhanced (80 ml) MDCT was performed using 64 slices at 0.75 mm collimation, 120 kV, and ECG-gated tube current modulation. The images were visually analysed by two readers in consensus using the 16-segment coronary artery model of the American Heart Association (AHA). The quality of the visualization of each coronary segment was rated (1= excellent, 2= evaluable 3= not evaluable). 21 patients (group A) were examined with prospective ECG-pulsing, 16 patients (group B) were examined with retrospective ECG-gating. The difference in image quality was done by the Mantel-Haenszel-test on a 95%-level. The radiation exposure was compared using the DLP. Results: Mean quality of all coronary segments in group A was 1.71 and 1.89 in group B, respectively. In all coronary segments the difference in quality between groups A and B did not reach significance. The p-values varied between 0.1 and 0.9. The DLP was 447 mSvcm in group A and 854 mSvcm in group B, respectively. Mean heart rate in both groups was 59 b.p.m. during MDCT. Conclusion: ECG-pulsing is a valuable tool for 64 slice coronary MDCT that allows reduction of radiation exposure by about 50% preserving complete image quality. Multislice computed tomography of the coronaries: Intraindividual comparison of radiation dose exposure with conventional coronary angiography C. Weber 1 , A. Deetjen 2 , O. Ekinci 2 , J. Hansel 2 , A. Kluge 2 , M. Weber 2 , G. Bachmann 2 , C. Hamm 2 , T. Dill 2 ; 1 Hamburg/DE, 2 Bad Nauheim/DE (C.Weber@uke.uni-hamburg.de) Purpose: The aim of this study was the intra-individual comparison of the effective radiation dose applied for Multislice spiral computed tomography (MSCT) and conventional coronary angiography (CXA). Methods and Materials: 56 patients (68 ± 8 years, 49 male) underwent MSCT and CXA. For MSCT, the standard protocol consisted of pre-monitoring, monitoring, calcium scoring and MSCT coronary angiography using a 16 slice CT scanner. CXA was performed on a current system with automatic selection of X-ray beam filtration following a protocol with 8 standard projections (LCA 6, RCA 2). To compare the radiation doses of both investigations, the effective dose was chosen as a parameter. A conversion factor was used for the equation of the effective dose: E=k x DLP. For conventional angiography, the effective dose was calculated using the dose-area product and a conversion factor according to data published by Le Heron. The mean effective dose for MSCT including pre-monitoring and monitoring scans was 10.36 ± 2.18 mSv (n = 56); if calcium scoring was added to MSCT, the dose was 12.67 ± 2.58 mSv (n = 45). In comparison, the mean effective dose of CXA was 3.33 ± 2.42 mSv (n = 56). A subgroup of patients (n = 10) with bypass grafts were analyzed in addition. The effective dose for MSCT was 12.95 ± 1.75 mSv. In comparison, the mean effective dose of CXA was 6.27 ± 4.04 mSv. Conclusion: This intraindividual comparison demonstrates that MSCT is still associated with a relatively high radiation dose exposure. This study emphasizes that indications for MSCT have to be clearly defined. Pediatric cardiovascular 64-slice CT angiography: Radiation dose reduction by automated anatomical tube current modulation G. Savino 1 , J.U. Schoepf 1 , P. Costello 1 , T.J. Vogl 2 , D. Mulvihil 1 , C. Herzog 2 ; 1 Charleston, SC/US, 2 Frankfurt/DE (savino@musc.edu) To assess the effect of weight-based scan protocols and automatic tube current modulation on radiation dose and image quality at pediatric 64-slice CTA. Our pediatric cardiovascular 64-slice CT protocols use a weight based algorithm to determine nominal tube current settings with 80-120 kV. Automated anatomic tube current modulation was used for each case. mAs, CTDI vol and DLP values were recorded and the effective dose calculated. Based on the selected nominal tube current, CTDI vol , DLP and effective dose, which would have been delivered without tube current modulation, were also calculated. Two radiologists independently rated image quality on a 5-point scale. Image noise was objectively assessed. Findings at CT were clinically correlated with cardiac ultrasound, angiography or surgery. Results: 38 CT scans were evaluated. Mean diagnostic quality was rated at 3.7 (range 3-4.4), with mean image noise of 9.1 HU STD (range 3.3-15.9). Compared to scanning without tube current modulation, mAs, CTDI vol , DLP and mean effective dose values were significantly (p < 0.01) reduced by 57.8% (54.1/ 128 mAs), 57.1% (2.5/5.9 mGy), 55.4% (70.1/156.9 mGy*cm) and 59.9% (2.38/ 5.9mSv), respectively. Body weight and mAs-values were significantly (p < 0.01) lower for 80 kVp than for 100 kVp and 120 kVp scans and radiation dose was significantly (p < 0.01) higher for 120 kVp than for 100 kVp and 80 kVp scans. Image quality between 120 kVp, 100 kVp and 80 kVp scans was not significantly different. CT diagnoses correlated well with clinical findings. Conclusion: Automated tube current modulation combined with low tube voltage settings significantly reduce radiation exposure and are generally recommended in pediatric cardiovascular 64-slice CT. Purpose: FpVCT offers a higher contrast resolution then MSCT. Until now, all published comparisons were performed by scanning small, exiled specimens such as cut temporal bones. However, with the recent availability of clinical scale fpVCT scanners, it was possible to test the performance on whole human skull bases and faces. Methods and Materials: Three embalmed human cadaver heads were scanned in a state of the art MSCT (GE Lightspeed 16) and an experimental fpVCT (GE). MSCT scanning was performed to optimize the spatial resolution (120 kV, 80 mA, 0.625 mm collimation) as were reconstruction (9.6 cm rFOV, 512x512 matrix, 0.3 mm z-spacing, "boneplus"). The fpVCT scanner consisted of two flat-panel detectors (1024 2 elements at 200 2 µm 2 ) that were mounted side by side on a standard CT gantry. Scan FOV was 33 cm with a z-coverage per rotation of 3.4 cm. Step and shoot scanning (1000 projections, 8 s rotation time, 140 kV, 25 mA) and modified FDK reconstruction were performed. The image appearance of 21 structures of the skull base and maxillo-facial region were rated on a 4 point scale by two independent raters. Results: The comparison revealed a significantly (p=< 0.05) higher average sum score for fpVCT (60.8) then for MSCT (41.5). Structures such as the stapedial muscle and the bony walls of the facial nerve canal were much more crisply displayed in fpVCT. FpVCT showed more streak artifacts than MSCT. Conclusion: Using fpVCT imaging of the whole human skull base and maxillofacial region is possible and superior to MSCT. Preclinical evaluation of root-canal-anatomy of human teeth with conebeam flat-panel volume-CT (FP-VCT) G. Heidrich, F. Hassepass, C. Dullin, T. Attin, E. Grabbe, C. Hannig; Göttingen/DE Purpose: For successful endodontic diagnostics and therapy, adequate depiction of root canal anatomy with multimodality diagnostic imaging is necessary. The aim of the present study was the evaluation of the presentability of the endodont with a flat-panel volume-CT device (FD-VCT). Methods and Materials: 13 human teeth were examined with a prototype of a FD-VCT. After acquisition of the data and generation of volume data sets in Volume Rendering Technology (VRT) the findings were compared to X-rays and cross-sectional preparations of the teeth. Results: Anatomical structures of the endodont such as root-, side-canals and communications between different root canals as well as denticles could be detected precisely with FD-VCT. Also the length of curved root canals was determined accurately. The spatial resolution of the system amounted to 10µm. Only 73% of the main root canals, detected with FD-VCT and 87% of the roots could be visualized with a conventional dental X-ray. None of the side canals, shown with FD-VCT were detectable in conventional X-rays. In all cases enamel and dentin of the teeth could be differentiated well. Stored and freshly extracted teeth as well as primary and adult teeth showed no differences in image quality. The FD-VCT is an innovative diagnostic procedure in preclinical and experimental use for non-destructive three-dimensional analysis of teeth. The smallest structures such as side canals can be detected and evaluated due to the high isotropic spatial resolution, in contrast to conventional X-rays. Purpose: To evaluate the utility of a new diagnostic radiographic technique for morphologic and functional evaluation in the lacrimal outflow system of patients with suspected nasolacrimal duct obstructions. Methods and Materials: Dacryocystography (DCG) was performed by gentle manual injection of nonionic water-soluble contrast media (Ultravist 300, Schering, Berlin, Germany) through a micro catheter inserted into the inferior cannaliculus. Imaging included digital flat-detector DCG (2 frames/sec) in frontal and lateral view (Alura Xper FD 20 flat detector System, Philips, The Netherlands). Additionally, 3D rotation scan was performed using C-arm roll movement at 55 degrees/ sec acquiring 120 images. Images were analyzed using functional and morphological criteria based on 3D surfaced shaded and multiplanar reconstructed 2D images. Additional radiation exposures were measured and analyzed. Results: In two cases no tear duct obstruction was found. In ten cases a stenosis and in four cases occlusion was found. Soft tissue swelling within the nasolacrimal duct was the most common cause of obstruction. In three cases additional dacryocele, respectively nasal pathology was depicted. The radiation dose applied was lower compared to conventional post-DCG computer tomography, providing comparable diagnostic information. Conclusion: 3D-rotation-flat-detector dacryocystography provides detailed imaging of the soft tissue lacrimal outflow system and surrounding bony structures within one diagnostic tour. This new technique combines morphologic and functional diagnostic information that improves planning of surgical interventional procedures. Five teeth with root fillings and clinical symptoms such as fistulas and isolated periodontal pockets of 8 mm or more were extracted after dental X-rays indicating lateral or periapical lesions. Vertical root fractures or cracks suspected due to the symptoms and clinical findings were evident after extraction in all cases but fracture lines were not visible on routine dental radiographs acquired before extraction. The extracted teeth were explored with a prototype of a flat-panel volume-CT (FP-VCT). Results: Using the FP-VCT, in all cases vertical root fractures or crack lines could be detected clearly in different views, depiction-modes and cross-sections at a spatial resolution of 140 µm. The evaluation of the fracture lines and teeth could be performed in three dimensional views. The FP-VCT findings were confirmed by detailed inspection of the extracted teeth. The FP-VCT is an innovative, diagnostic tool for non-destructive, three dimensional evaluation of extracted teeth in pre-clinical and experimental studies. The FP-VCT allows precise visualization and evaluation of vertical root fractures or cracks in extracted teeth. Clinical application of the system may be possible if technical modifications reduce the exposure dose: the high resolution detector-systems of the FP-VCT should be combined with radiation-systems that collimate the beam to the area of interest. Flat-panel based volume-CT (fpVCT) improves intraoperative navigation accuracy: Innovative surgery concepts such as navigated, minimally invasive cochleostomy might become possible S.H. Bartling 1 , O. Majdani 1 , J. Graute 1 , C. Dullin 2 , T. Lenarz 1 , H. Becker 1 ; 1 Hannover/DE, 2 Goettingen/DE (soenkebartling@gmx.de) Purpose: Multislice-CT based intraoperative navigation is a well established method in ENT surgery. However, currently the best possible navigation precision (0.7-1 mm) is limited compared with small structures of the lateral skull base. The resolution of the underlying CT dataset could be one significant influential factor; therefore we tested whether high resolution imaging provided by flat-panel based Volume-CT translates to higher navigation precision. Methods and Materials: An acrylic glass phantom with a rigidly attached reference star providing 40 navigation markers in the form of inverted cones was C D E F B A G designed. It was scanned in a state of the art MSCT (GE Lightspeed 16) and an experimental fpVCT (GE). MSCT scanning (~350 µm feature size in plane, 5 00 µm in z) and fpVCT scanning (~ 200 µm feature size, isotropic) were optimized for high resolution. The navigation system VectorVision2 (BrainLAB) was used. Six cones were used for registration, seven for measuring the average target registration error (TRE) (5 reiterations). Furthermore, a stereotactic, minimally invasive cochleostomy was tried on four excised temporal bones that were scanned in fpVCT. Results: Average TRE using fpVCT was 0.46 (SD 0.22), significantly lower (p < 0.1, two-sided t-test) than using MSCT: 0.82 mm (SD 0.35). Stereotactic cochleostomy was successful in all four temporal bones. Conclusion: High resolution fpVCT imaging improves intraoperative navigation accuracy. As shown here in ex-vivo experiments, this could be highly beneficial for skull base surgery, where precise image guided interventions such as cochleostomy or stereotactic drilling biopsies could become possible. Contrast-enhanced computed tomography examination may provide noninvasive differentiation between cholesteatoma and granulation tissue A. Trojanowska, P. Trojanowski, E. Czekajska-Chehab, W. Olszanski, W. Golabek, A. Drop; Lublin/PL (agnieszka30@yahoo.com) To determine the role of delayed post-contrast computed tomography imaging in the differentiation between residual or recurrent cholesteatoma and granulation tissue in patients who have undergone middle ear surgery. Methods and Materials: Sixteen patients who underwent resection of cholesteatoma were included in the study. In each case, CT examination of the temporal bone was performed in the following protocol: slice thickness 0.6 mm, overlap 0.3 mm, FOV 9.8 mm, matrix 512 x 512, reconstruction kernel -soft tissue, 80 kV, 160 mAs. Examination was performed before and after intravenous injection of iodinated non-ionic contrast medium: 60 ml, flow rate 1.5 ml/s, post-contrast delay 70 s. During post processing, the density of middle ear masses was measured in Hounsfield units (HU) with the use of specially designed color-coded maps. An experienced radiologist evaluated images searching for post-contrast enhancement. Imaging findings were correlated with surgical findings. Results: Based on computer created maps coding HU in colours, it was possible to observe non-homogenous enhancement of a middle ear mass in twelve cases, which were proven to be granulation tissue at surgery. In four cases the middle ear mass showed no significant post-contrast enhancement; these were proven to be cholesteatomas. Conclusion: Evaluation and comparison of pre-and post-contrast CT images can confirm residual or recurrent cholesteatoma in patients who underwent middle ear surgery by showing non-contrast enhancement. In cases with granulation tissue, delayed post-contrast enhancement of more than 15 HU is typical. This will help in non-invasive differentiation between post-inflammatory and cholesteatoma lesions. High resolution multislice CT (MSCT) for predicting basal cochlear length for electric-acoustic stimulation M.G. Mack 1 , O. Adunka 2 , M. unkelbach 1 , A. Wetter 1 , T. Lehnert 1 , J.O. Balzer 1 , T.J. Vogl 1 ; 1 Frankfurt a. Main/DE, 2 Chapel Hill, NC 27599-7070, NC/US (m.mack@em.uni-frankfurt.de) Purpose: To assess the feasibility and accuracy of predicting electrode insertion depth necessary in cochlear implantations (CI) for electric acoustic stimulation by means of pre-operative MSCT. Methods and Materials: 8 fresh human temporal bones were examined with MSCT with a 0.5 mm collimation (increment 0.2 mm). The basal cochlear structures were reconstructed and measured using a 3D-program (vessel view, Siemens). Standard cochlear implantations with a free fitting array were performed. A special technique, which allows sectioning of undecalcified bones was used. After embedding, all bones underwent conventional X-ray analysis with further insertion measurements. Pre-implantation HRCT data was compared to X-ray and histology. Results: Pre-implantation MSCT reconstruction was feasible and easy-to-use in every bone. Linear reconstruction revealed all relevant anatomical details of the cochlea and adjacent structures. Using the described linear reconstruction method, an average basal cochlear length for the first 360° of 20.3 mm (range 18.8 to 22.0 mm) was measured. For the 8 specimens, values for the implanted length until the 360° point ranged from 19.2 to 21.9 mm with an average of 20.3 mm. Pre-implantation CT-based measurements correlated very well with post-implantation X-ray data. Comparing the results of the pre-, and post-implantation meas-urements, a mean failure of 0.3 ± 0.28 mm in predicting the correct length of the basal turn of the cochlea was observed. The maximum failure was 0.7 mm. Conclusion: MSCT allows the good prediction of the cochlea lengths before CI. This is especially useful when aiming for hearing preservation, where insertion depths of 360° are necessary. Vascular mapping for reconstructive surgery: Comparison of CTA and DSA M. Lell 1 , U. Baum 1 , B. Tomandl 2 , E. Nkenke 1 , W. Bautz 1 ; 1 Erlangen/DE, 2 Bremen/ DE (michael.lell@idr.imed.uni-erlangen.de) Purpose: To compare the potential of computed tomography angiography (CTA) with digital subtraction angiography (DSA) in vascular mapping of the external carotid artery (ECA) branches for planning of microvascular reconstructions of the mandible with osteomyocutaneous flaps. Methods and Materials: In 35 patients CTA and DSA was performed prior to surgery. 23 patients had prior surgery or radiation therapy because of cancer of the floor of the mouth/oral cavity. Selective common carotid angiograms were acquired in 2 projection for both sides of the neck. 16-or 64-slice-spiral-CT was performed with a dual phase protocol, using the arterial phase images for 3D-CTA reconstruction. Volume rendering (VR) was employed for postprocessing of CTA data. The detectability of the different ECA branches in CTA and DSA was evaluated by two examiners. Results: No statistically significant differences between CTA and DSA were found for identifying branches relevant for microsurgery. DSA was superior to CTA if more peripheral ECA branches were included. All patent parent vessels of the transplants could be visualized with CTA. Conclusion: CTA proved to be a promising non-invasive alternative to DSA in vascular mapping for planning microvascular reconstruction of the mandible. One hundred and sixty-five individuals were randomly selected in a community located in southeast Brazil. We took a clinical history in each subject and determined serum thyrotropin, antiperoxidase antibodies, TRAb and thyroid volume by US. Subjects with thyroid disease, and those under 20 years of age were excluded. 84 representative subjects (30 men and 54 female) remained. Systolic peak velocity (SPV), resistive index (RI) and pulsatility index (PI) of superior and inferior thyroid arteries were measured by a 5-12 MHz linear probe. Results: Except for RI, the distribution of all Doppler parameters was not gaussian. The median and mean references for SPV, RI and PI were 24.80 cm/s and 25.85 cm/s, 0.60 and 0.62, and 0.98 and 1.04, respectively for superior thyroid arteries; these reference values for the inferior thyroid artery were 20.92 cm/s and 21.50 cm/s, 0.57 and 0.57, and 0.84 and 0.88, respectively (p = 0.0004). Females had greater SPV values (p < 0.007). There was a linear correlation between SPV and age (r = 0.41 and p < 0.0001). We have determined reference values of thyroid Doppler parameters in our iodine non-deficient population and prepared tables by sex and age. Purpose: To evaluate the reliability of contrast-enhanced US and dynamic MR imaging in determining the response to percutaneous radiofrequency ablation of hepatocarcinoma. Methods and Materials: Fifteen patients with hepatocarcinoma treated by radiofrequency ablation underwent contrast-enhanced (Sulfur Hexafluoride (SF 6 )) and gadolinium-enhanced MR imaging four weeks after treatment. Any focus of intense contrast uptake within the tumor in the arterial phase was considered as persistence of the tumor. All patients were biopsied after contrast-enhanced US. Results: Biopsy showed persistence of the tumor in ten hepatocarcinomas and complete ablation in five. Both contrast-enhanced US and MR imaging found complete ablation in thirteen hepatocarcinomas and persistent tumor in two, with one false positive in each technique. Sensitivity, specificity, PPV and NPV were 10%, 80%, 50% and 30% for both techniques. Conclusion: Biopsy to measure the response of hepatocarcinoma to radiofrequency ablation is rarely performed and histopathological criteria are uncertain. There is a discrepancy between the response to radiofrequency ablation of hepatocarcinoma measured by imaging techniques and histopathological study. Purpose: Endovascular stent grafting has a proven role in treating diseases of the thoracic aorta. The mortality and morbidity associated with the procedure is far lower than open surgery. However many patients are unsuitable for treatment due to the absence of an adequate proximal or distal landing zone. We perform extra anatomic bypass grafting of the arch or mesenteric vessels to facilitate stenting in such cases. Methods and Materials: Thirteen patients were treated with bypass prior to thoracic stent grafting. Aortic pathologies included aneurysms affecting the arch and descending aorta, aneurysmal type B dissections, thoraco-abdominal aneurysm and one aneurysm of an aberrant right subclavian artery. Patients underwent a combination of grafting from the ascending aorta, carotid-carotid artery bypass, carotid-subclavian artery bypass and mesenteric artery bypass. Four patients were treated with Medtronic Talent stents, one patient with a Gore TAG stent and eight patients with the new Medtronic Valiant stent graft. Results: Initial technical success was achieved in all patients. One procedure was complicated by rupture of the iliac artery. One patient died in the perioperative period giving a 30 day mortality of 7.7%. There were two small endoleaks (15%) on the immediate post operative CT. However there were no endoleaks present at three months (0%). At a mean follow-up of eight months there have been no further deaths, ruptures or complications related to the bypass grafts. Conclusion: Extra anatomic bypass is a valid and safe technique to expand the proportion of patients suitable for endovascular stenting of the thoracic aorta. Descending thoracic aorta dissections were experimentally created with a catheter-based technique under X-ray in 8 domestic pigs weighing 63-98 kg. All subsequent MR imaging was performed on a 1.5-T wholebody MR imaging scanner (Avanto, Siemens, Erlangen, Germany). Pre-interventional dissection evaluation was commenced with ECG-gated TrueFISP imaging and with contrast enhanced 3D FLASH MRA. A self-expandable, nitinol thoracic aortic stent-graft device (GoreTAG, Gore Inc., Flagstaff, AZ) was chosen for the experiments. For implantation, the delivery system with mounted stent-graft was advanced from the iliac artery to the thoracic aorta under real-time MR imaging fluoroscopy featuring an interactive radial TrueFISP sequence providing 7fps. Images were projected onto an in-room 60" high-contrast projection screen (MR-Iscreen, MR-Innovation GmbH, Essen, Germany). Following stent-graft placement, the pre-interventional MR imaging protocol was repeated to evaluate procedure success. Results: Descending aortic dissections were successfully created in all 8 pigs. Real-time MR imaging permitted: 1) successful and safe device navigation within the true lumen from the iliac arteries to the thoracic aorta, 2) precise positioning of the stent-graft, 3) successful stent-graft deployment, and 4) safe catheter withdrawal. Post-interventional MR imaging demonstrated correct position of the stentgraft and complete obliteration of the false lumen which was confirmed by necropsy. Conclusion: Endovascular stent-graft placement for experimental aortic dissection using commercially available catheter devices is feasible and safe under real-time MR imaging-guidance. Beyond detailed pre-interventional evaluation and procedure guidance, MR imaging permits immediate post-interventional anatomic and functional evaluation of procedure success. Midterm results of transcaval approach in treatment of type II endoleak: One patients with type II endoleaks after EVAR. Via a venous access, a curved guiding cannula was advanced into the caval vein to the site of the aneurysm sac. A flexible needle in the cannula was oriented towards the sac and embolization inside the sac was performed with transcaval puncture. Systemic and intrasac pressures were recorded before and after the embolization. CT scans were performed at 24 hours, 30 days and 1 year after the procedure: Both the presence of leaks and the changes in sac diameter were evaluated. Technical success was identified by feasibility of the procedure and clinical success with no evidence of leaks at follow-up. Results: Transcaval puncture was feasible in 10/12 procedures (technical success 83%). The mean systemic pressure was 150/86 mmHg (range 117/69-176/ 100). The mean intrasac pressure before embolization was 79 mmHg (range 140-40). After embolization, the mean intrasac pressure was 10 mmHg (range 5-39) in 10 patients, while in one it increased. At the 24 hours CT scan, gas bubbles and stable contrast medium with coils inside the sac were seen in 10 Patients. 10/11 patients were successfully treated; at 1 year CT scan, a reduction in diameter of the sac and no leak recurrence were observed (clinical success 91%). Conclusion: Transcaval approach is feasible for type II endoleak treatment. High clinical success rate makes it an alternative to direct translumbar puncture. Purpose: To evaluate whether endoleaks, simulated in an in-vitro model of an abdominal aortic aneurysm (AAA), can be detected by a percutaneously implantable telemetric pressure sensor (TPS). An elastic silicone AAA model including a deployed endograft was connected to a pulsatile circulation with physiologic flow and pressure rates. Different types (I-III) and sizes (3-11 Fr) of endoleaks were created. Before and after generating endoleaks aneurysm sac pressure (ASP) was telemetrically obtained with our newly designed TPS and simultaneously measured at several positions with 3 Fr. wire mounted pressure sensors (WPS) as a reference. Measurements were performed in fluid filled and in partly/fully thrombosed aneurysm sacs. ASP readings detected by TPS and WPS were compared to each other and to the intra-aortic pressure (IAP). C D E F B A G renal infarcts, 3 acute cortical necrosis (ACN), 6 renal abscesses, 15 acute pyelonephritis, 6 atheroembolic renal disease, 5 not perfused areas after blunt abdominal trauma, 2 segmental renal artery stenosis. Cysts, renal tumors, renal cortical scars and scarred kidneys were excluded. Results: An excellent depiction of normal renal vasculature was obtained even in elderly, obese and nephropathic patients. Renal infarcts presented as wedgeshaped segmental perfusion defects. In ACN only small cortical areas were perfused, while the hilar renal vessels enhanced normally. Abscesses presented with rounded perfusion defects. Pyelonephritis presented with wedge-shaped areas of decreased enhancement (4/15), slightly hypoperfused areas (7/15), or barely visible perfusion defects (4/15). In patients with atheroembolic renal disease bilateral hypoperfused areas can be identified during the early arterial phase. Segmental renal artery stenosis presented as segmental areas with delayed enhancement. Conclusion: CEUS allows excellent depiction of renal lesions presenting as perfusion abnormalities. Day case renal biopsies -desirable or dangerous? A.J. Sebastian, S. Nair, R. Panigrahi, A.M.B. Bowker; York/UK (arunsebastian@hotmail.com) Purpose: To determine the rate and timing of complications after percutaneous ultrasound guided renal biopsy (PRB) and to assess if this biopsy can be performed as a day case procedure. Methods and Materials: 141 consecutive renal biopsies performed during the 4 year study period between January 2001 and December 2004 were retrospectively assessed to identify post biopsy complications and the adequacy of the biopsy samples. The patients were all monitored for a minimum duration of 24 hours after the procedure. A median of two 18G samples (range 1-5) were taken from each patient. Diagnostic samples were obtained in 138/141 (98%) patients. Results: One patient (0.7%) developed post procedure renal haemorrhage within 4 hours after the biopsy requiring selective renal arterial coil embolization and blood transfusion. There were no other major complications and there was no procedure related mortality. Literature suggests that PRB performed using 14G needles had a major complication rate of 6.4%, with 33% of these complications being detected more than 8 hours after the procedure. We found no additional benefit in monitoring patients for 24 hours rather than for a period of 8 hours. Performing PRB as day case procedures will also be cost effective as admitting each patient for overnight observation can cost more than £300. Conclusion: Day case PRB is a safe, cost effective procedure that will continue to provide renal biopsy samples of diagnostic quality. We aim to start performing PRB as day case procedures and re-audit our results. Arterial spin labeling for kidney transplant monitoring C. de Bazelaire 1 , P. Martirosian 2 , O. Mathieu 1 , E. Pillebout 1 , P. Bourrier 1 , E. de Kerviler 1 , J. Frija 1 ; 1 Paris/FR, 2 Tübingen/DE (cedric.de-bazelaire@sls.aphp.fr) Purpose: Kidney transplants are monitored by MR imaging using a morphological approach based on the analysis of arteries, nephrogram, urinary cavities and the surrounding structures. This study aimed to assess whether cortex perfusion determined by Arterial Spin Labeling (ASL) could help to evaluate the transplant kidney. Methods and Materials: Twenty patients with kidney transplants were included in the study. The median delay between transplantation and ASL was 19 days. Reasons for MR imaging evaluation were anuria, unsatisfactory improvement in serum creatine or rise of serum creatine. MR imaging was performed on a Siemens-Symphony 1.5 Tesla scanner using the body coil for transmission and a surface coil array for reception. For ASL, a single slice FAIR method with a True-FISP imaging acquisition was used. Perfusion was measured on a coronal view of the kidney in free breathing conditions. Imaging protocol included an MRA sequence (FLASH 3D) with contrast injection (Dotarem, Guerbet, France) and a morphological sequence (VIBE). Results: Perfusion was fully evaluated by ASL in all patients. The pelvic position of kidney transplants did not cause reduced motion artefact. Renal cortex blood flow was found to correlate with serum creatine (R=0.80). No correlation between perfusion and transplant artery stenosis was found. Perfusion maps were unrelated to enhancement of the renal parenchyma. Conclusion: Our results suggest that physiological mechanisms could be highlighted by ASL: 1) renal cortex perfusion and glomerular function are related, 2) renalcortex perfusion could be maintained for a long time despite significant renal artery stenosis and 3) renal enhancement does not reflect perfusion. Diffusion-weighted MR imaging of renal allografts: A comparative study to healthy volunteers H.C. Thoeny 1 , D. Zumstein 1 , S. Simon-Zoula 1 , U. Eisenberger 1 , F. De Keyzer 2 , C. Boesch 1 , F. Frey 1 , P. Vock 1 , P. Vermathen 1 ; 1 Berne/CH, 2 Leuven/BE (Harriet.Thoeny@insel.ch) Purpose: To prospectively evaluate the feasibility and reproducibility of diffusionweighted magnetic resonance imaging (DW-MRI) for the assessment of kidney function in patients with transplanted kidneys and to correlate the results with age and sex matched native kidneys. Methods and Materials: Fifteen patients with transplanted kidneys and stable kidney function and 15 age and sex matched healthy volunteers underwent coronal DW-MRI with respiratory triggering on a 1.5 T MR system using a large set of 10 b-values (b=0-900sec/mm 2 ). The following diffusion parameters were calculated: Apparent diffusion coefficient (ADC tot ) from all b-values, a diffusion coefficient (D) and the perfusion fraction F p . Cortex and medulla were analyzed separately. Serum creatinine concentration was obtained from each subject. Seven of the 15 patients with transplanted kidneys underwent a repeat study using the identical protocol to determine reproducibility. Results: In transplanted kidneys almost identical values were obtained in the cortex and medulla for ADC tot , D, and for F p . In contrast, in native kidneys ADC tot , D and F p were significantly higher in cortex than in medulla (p < 0.0001, p < 0.002 and p < 0.01, respectively). The reproducibility study showed a very low CV w (coefficient of variation within subjects) in cortex and medulla especially for ADC tot and D of less than 3.2%. Diffusion and perfusion indices correlated with serum creatinine concentrations. Conclusion: DW-MRI in patients with transplanted kidneys is feasible with highly reproducible results and is able to detect differences in diffusion-based corticomedullary differentiation as compared to native kidneys. It might therefore be able to noninvasively assess acute and chronic graft rejection after kidney transplantation. Implication of dilated renal pericapsular vein in renal cell carcinoma on CT or MR imaging J. Lee, S. Kim, S. Lee; Pusan/KR (junwlee@pusan.ac.kr) Purpose: To evaluate the factors affecting dilatation of renal pericapsular veins in renal cell carcinoma (RCC), and the relationship between dilated renal pericapsular veins (DRPV) and hematogenous metastases to other organs. Methods and Materials: CT and MR images of 127 patients with pathologically proven RCC were retrospectively reviewed. CT or MR imaging was obtained with excretory phase only, or both corticomedullary and excretory phase after administration of contrast material. The parameters analysed on imaging were: (1) the diameter and number of DRPV; (2) the tumor size and degree of contour bulging; (3) the presence of renal vein thrombosis; and (4) the degree of tumor enhancement. In addition the degree of DRPV was compared to the nuclear grade and histological subtypes of RCC by using the Spearman rank correlation coefficient. Results: DRPV were seen in 90 (70.9%) patients. Hematogenous metastases of RCC to other organs were found in 15 (12%) patients. Metastases were found in the lung (n = 9), liver (n = 3), brain (n = 1), vertebra (n = 1) and pleura (n = 1). Of the 15 patients with metastases, 14 patients showed a variable degree of DRPV. DRPV were identified only in patients with an exophytic mass larger than 2 cm in diameter and patients with renal vein thrombosis. Renal vein thrombosis was present in 6 patients. There was a weak correlation between the degree of DRPV and mean tumor size and degree of contour bulging (P <.05). Conclusion: Tumor size and exophytic growth of renal cell carcinoma were related to the development of DRPV. Hematogenous metastasis of RCC to other organs was high in patients with DRPV. High resolution ASL perfusion imaging of the kidneys at 3 Tesla A. Boss, P. Martirosian, C.D. Claussen, H.-P. Schlemmer, F. Schick; Tübingen/DE (andreas.boss@med.uni-tuebingen.de) Purpose: Arterial spin labeling (ASL) provides a non-invasive modality for quantitative perfusion measurements. The feasibility of high resolution ASL imaging of the kidneys was tested and proven at 3 Tesla using a flow-sensitive alternating inversion recovery (FAIR) true fast imaging in steady precession (TrueFISP) technique. Methods and Materials: An ASL sequence with FAIR spin-preparation and True-FISP signal detection strategy was adapted for high-resolution perfusion imag- ing of the kidneys in a 3 Tesla whole-body scanner. Quantitative perfusion maps were acquired in six healthy volunteers and two patients. To avoid banding artefacts, which are prominent at 3 Tesla, a TrueFISP frequency scout was implemented. Perfusion maps with an in-plane resolution of 1.5 mm were recorded in transverse and coronal planes. For fast mapping of the whole kidney, an in-plane resolution of 2 mm was applied. Results: High-resolution perfusion images with excellent image quality could be obtained during a measuring time of 10 minutes in all subjects. The whole kidney could be mapped with diagnostic image quality in less than 10 minutes. For all slices, a frequency offset of less than 100 Hz provided the kidneys free of True-FISP artifacts. Perfusion values of the renal cortex ranged from 250 up to 400 ml/ 100 g/min (mean cortical perfusion right kidney 316 ± 43, left 336 ± 40). Conclusion: Perfusion images of the kidney could be obtained with excellent image quality by means of a 3 Tesla MR setting within a clinically applicable measuring time offering an alternative to conventional perfusion imaging with potentially nephrotoxic contrast media. Purpose: There is sparse data on the MR appearance of the normal development of the structures of the fetal posterior fossa, and how prenatal MR imaging data compares to US investigations at the same gestational age (GA). In this study we present preliminary MR imaging data and compare that to published, GA matched ultrasound (US) data. Methods and Materials: Image sets originally obtained by using (T2 weighted) single-shot fast spin-echo, (EXPRESS) of 91 normal fetuses between GA's of 40 and 20 weeks were retrospectively analysed using proprietary measuring software. Measurements were taken of the trans-cerebellar diameter (TCD) the crainiocaudal (CCD), anteroposterior (AP), midline sagittal surface area (CSA) dimensions of the cerebellar vermis, the AP diameter of the fourth ventricle (AP4), anterior and posterior axial cerebrospinal fluid (CSF), and the midline sagittal surface area of an area defined as the 'bony posterior fossa' (PF). Results: We present data from 63 normal fetuses. We have shown that there is significant correlation between MR imaging and US measurements of fetal PF structures and that there is a constant discrepancy between vermis CCD and AP diameters as measured by MR imaging. Conclusion: An in vivo MR template of normal fetal cerebellar (posterior fossa) development is essential for conclusive and accurate interpretation of the foetal brain, to this end our data underlines previous work by Griffiths, Whitby and other groups corroborating previously published MR depictions of the developing cerebellum and posterior fossa and is consistent with published US data of similar age matched measurements. Comparison of BOLD signal changes in the fetal brain at 3 T with local pO2 measurements U. Wedegärtner, N. Beindorff, A.N. Priest, H. Kooijman, H. Schröder, G. Adam; Hamburg/DE (wedegaer@uke.uni-hamburg.de) Purpose: To compare BOLD signal changes (SI) with local pO 2 measurements in the fetal brain during hypoxia. Methods and Materials: A T2*-weighted single-shot gradient-echo sequence (TE 30 ms) was used to measure BOLD SI changes in the fetal sheep brain during hypoxia. 9 pregnant ewes were examined in a 3 T scanner (Intera, Philips Medical Systems). Measurements were performed during a control period with normal oxygen saturation, then during hypoxia and recovery induced by reduction and subsequent increase of oxygen in the ventilated gas mixture. Regions of interest were placed in the fetal cerebrum to measure SI. The changes in the relaxation rate (Delta_R2*) were estimated for each plateau by ln[(SI (t)/SI control )/ TE]. Fetuses were chronically instrumented with a probe in the left hemisphere to continuously measure the local pO 2 . The difference of the control and hypoxic/ recovery pO 2 values was calculated (Delta_pO 2 ). Linear regression analysis was used and a correlation coefficient (r) was calculated to indicate the correlation between Delta_R2* and Delta_pO 2 . Results: BOLD measurements were obtained during control (n = 15), hypoxia (n = 22) and recovery (n = 12). Mean values of relative SI% (SI (t)/SI control ) and pO 2 were reduced from 100% and 14 mmHg pO 2 [95% confidence interval 10-17] to 81% SI% [76-86] and 3 mmHg [2-4] and increased to 99% [93-105] and 15 mmHg [11] [12] [13] [14] [15] [16] [17] [18] [19] [20] during recovery. Delta_R2* and Delta_pO 2 were correlated significantly (r = 0.64; p = 0.00005). Conclusion: BOLD SI changes reflect changes of local brain pO 2 . R2* measurements in the brain correlate with local brain pO 2 values. Therefore R2* provides information about fetal brain tissue oxygenation. Purpose: It has been respectively shown that MR spectroscopy (MRS) and diffusion tensor imaging (DTI) can depict abnormalities in temporal lobe epilepsy (TLE). Our purpose was to use both techniques to assess TLE and compare their discriminative ability. Single-voxel proton MRS, DTI and routine MR imaging were performed in 42 patients with TLE and in 29 age-matched controls using a 3.0 T MR scanner. The voxel of MRS was placed in the hippocampal region bilaterally in each subject. The N-acetylaspartate (NAA)/creatine and choline/creatine values were calculated automatically. DTI was performed with a b factor of 1,000s/mm2 along 25 directions. Fractional anisotropy (FA) values were obtained from heads of hippocampus and selected regions in temporal lobe white matter through Functool. The metabolic parameters and FA values were compared between the patient and control groups. Receiver operating characteristic (ROC) analysis was utilized to compare the discriminative efficiency of the two techniques. Purpose: To determine if there are significant differences in concentrations and ratios of metabolites in different gray and white matter areas in the brain between fibromyalgia (FM) patients and normal controls using MR spectroscopy. Methods and Materials: 26 FM patients and 22 age-matched healthy control subjects underwent conventional pre and post-contrast enhanced MR imaging as well as 2D-CSI spectroscopy (TE/TR=144/1500 ms). Large volumes of interest were placed at the level of basal ganglia and also in the periventricular white matter at the level of the centrum semiovale. Within the larger voxel, eighteen 1 cm x 1 cm x 1 cm smaller voxels were placed in gray and white matter locations that have been implicated in pain processing. Post-processing of the spectra was done with vendor-provided software (Functool, 2000 GE) . NAA/Cr, Cho/ Cr and Cho/NAA ratios were calculated for each voxel. Student's t-test was used to determine significant differences in the metabolite ratios between the two groups. Results: Mean Cho/Cr ratios were significantly higher in FM patients compared to normal controls in the right prefrontal subcortical white matter (p = 0.02) and the left parietal white matter (p = 0.04). Non-significant similar trends were seen in the left thalamus (p = 0.07) and the left internal capsule (p = 0.09). Conclusion: MR spectroscopy may be of use in evaluating patients with fibromyalgia. Our data suggests that there are regional metabolite differences between FM patients and healthy controls in areas that have been implicated in pain processing. Larger studies are required to further validate this potential tool and correlate MR spectroscopy with clinical symptoms and response to therapies. In a prospective randomised study 100 conventional localization clips (Micromark™ II, Ethicon, USA) and 100 gelatin pledget/Metallic Marker clips (Gel Mark™ ultra, SenoRX Inc, USA) were deployed using a stereotactic system (Mammotome™, Fisher Imaging, USA) following 11-g VABB. Preand post-interventional two-view mammograms were analyzed by two radiologists, independently. The position of the markers were assessed. In case of a dislocation, defined as a clip migration > 5 mm from the target, the distance and direction of dislocation was recorded. The dislocation was scored on a threepoint scale [1=no dislocation (≤ 5 mm); 2=minor dislocation (6-20 mm); 3=major dislocation (> 20 mm)]. Results: Post-interventional mammograms revealed clip displacement in 43/200 (21.5%) cases. Clip displacement was seen significantly more often in calcifications (39/43; 90.7%) than in masses (4/43; 9.3%; p < 0.05). Displacement was seen significantly more often in conventional localizing clips (30/100; 30%) than in gelatin pledget/metallic clips (13/100; 13%; p0.05). Major dislocation was seen significantly more often using conventional clips only (7% vs. 0%; p < 0.05). Conclusion: Following 11 g VABB clip displacement more than 20 mm was seen significantly more often using conventional localization clips in comparison to the gelatine pledget/metallic clips. Conventional localization clips should no longer be used following stereotactic 11 g VABB. Purpose: The purpose of our study was to evaluate the accuracy of galactography-guided vacuum-assisted biopsy (VAB) and its feasibility as potential therapeutic procedure for intraductal lesions of the breast. Methods and Materials: 16 patients (mean age 58.2 years) with nipple discharge and galactographically verified intraductal lesions, were included in this prospective trial. Galactography-guided stereotactic 11-gauge VAB was performed in 16 lesions followed by open biopsy. VAB and surgical histology were correlated. Results: After VAB, histopathologic findings revealed ductal cancer in situ in five (31.3%) cases, invasive ductal cancer in three (18.7%), atypical in two (12.5%), and a benign result in six (37.5%) lesions. In ten cases (62.5%) histopathology after open biopsy confirmed the previous established diagnosis. Five (31.3%) carcinomas were missed or underestimated at VAB, showing one DCIS, two atypical papillomas and two benign results at VAB. One (6.2%) atypical ductal hyperplasia was underestimated as papilloma at VAB. At surgery tumor remnants were found in all cases. In all patients nipple discharge ceased after open surgery (follow-up at least 18 months). Conclusion: Our results suggest that galactography-guided VAB is a feasible method to evaluate intraductal lesions presenting with nipple discharge. As underestimation or false negative results occurred in six of 16 patients (37.5%) and complete removal cannot be assured due to the nature of lesions, 11-gauge VAB cannot be recommended as alternative to open biopsy. Stereotactic large-core needle breast biopsy (SLCNBB): Analysis of pain and discomfort related to the biopsy procedure J.M. Hemmer, J.C. Kelder, J.P.M. van Heesewijk; Nieuwegein/NL (judith@reijnen-hemmer.nl) Purpose: This prospective study was designed to evaluate the significance of variables (duration of the procedure, type of breast tissue, number of biopsies, depth of the biopsies and the operator performing the procedure) and their effect on women's perception of pain and discomfort during SLCNBB. Methods and Materials: Between June 2003 and October 2005, 150 patients were included with a non-palpable suspicious mammographic lesions, in the analysis. 14 Gauge biopsies were taken using a prone stereotactic table (Lorad stereoguide). Following their biopsy procedure the patients were asked to complete a questionnaire. Results: 150 biopsy procedures were analyzed. There was no discomfort in lying on the prone table. There is no relation between type of breast lesion and pain, underlying pathology and pain and performing operator and pain. The type of breast tissue is correlated with the pain experienced from biopsy (p = 0.0001). The depth of the biopsy correlates with pain from biopsy (p = 0.0028). There is a correlation between the amount of biopsies and pain in the neck (p = 0.0188) and shoulder (p = 0.0366). The duration of the procedure is correlated with pain experienced in neck (p = 0.0116) but not with pain experienced from biopsy. Conclusion: SLCNBB is not experienced as painful. No correlation was found between the performing operator and pain. There is a correlation between duration of the procedure and pain in neck. Pain from biopsy is correlated with depth and type of breast tissue. Computer-navigated MR-guided vacuum-assisted breast biopsy: A feasibility study U. Mastalier, H. Fauster, A. Ruppert-Kohlmayr, M. Uggowitzer; Leoben/AT (ursula.mastalier@lkh-leoben.at) Purpose: To evaluate a recently available computer-navigated targeting system for MR-guided biopsy and localization of non-palpable breast lesions which are not detectable with mammography or sonography. Methods and Materials: 14 patients underwent vacuum-assisted breast biopsy (VABB) because of BIRADS-IV and -V lesions, exclusively detected on prior MR exams. 14 lesions with a diameter between 6 and 36 mm underwent biopsy using a computer-navigated target system including a dedicated workstation (Dy-naCAD ® ). Sequences for intervention consisted of a T1W FS-markersequence and CE-dynamic sequences using a dedicated breast coil. 20-24 specimen were taken of each lesion and underwent histological evaluation. Immediately after biopsy a MR-compatible clip with bioabsorbable attachment was placed via a coaxial needle into the biopsy site. Histologically proven malignant lesions underwent surgery guided by the clip within 2 weeks. Results: VABB could be performed in all lesions well tolerated by all patients (100%). 8/14 lesions (57%) were malignant (6 invasive carcinomas, 2 DCIS) which could be surgically confirmed. 6/14 lesions were benign (43%). Placement of a clip into the biopsy hole was possible in all cases. Surgery confirmed the presence of the clip within or adjacent to the tumor in all surgically proven malignant lesions. No clip displacement occurred. Mean duration of the procedure was 50 minutes. Conclusion: Computer-navigated MR-guided breast biopsy with DynaCAD® is a robust and highly accurate and feasible method for histological evaluation and localization of breast lesions which are neither detectable with mammography nor with sonography. Purpose: Vacuum assisted breast biopsy (VABB) under MR imaging guidance is increasingly used for the histological verification of MR imaging-detected breast lesions that lack a correlation at mammography and ultrasound. The aim of this study was to compare three commercially available VABB devices for their yield of breast tissue. Methods and Materials: Three VABB devices [Mammotome (Ethicon USA), Vacora (Bard, USA), and ATEC (Suros Surgical System, USA)] were included in this study. Each device was used to obtain 50 specimens from fresh turkey breasts. Turkey breasts were immobilized using a stereotactic device (Noras, Germany) and positioned on the table of a 1.5 T magnetic resonance imager (Vision, Siemens, Germany) 60 cm from the gantry. Obtained specimens were measured for weight, length and fragmentation. Results: Biopsy could be successfully performed with each of the devices inside the MR imaging suite. The ATEC device (mean sample length 25.5 mm, mean sample weight 188.6) performed best, followed by the Vacora (mean sample length 18 mm, mean sample weight 155.1) and the Mammotome (mean sample length 19.8 mm, mean sample weight 77.6). Differences in mean sample weight were statistically significant (p < 0.05). 15 samples obtained with the ATEC device yielded a total of 3004.8 mg of breast tissue. To retrieve the same amount of breast tissue, 21 passes were necessary with the Vacora, and 39 passes with the Mammotome. Conclusion: Biopsies inside the MR imaging suite could be successfully performed with each of the devices. The ATEC device performed significantly better than the Vacora and the Mammotome. Breast abscess treatment with ultrasound-guided percutaneous aspiration G. Jahr, P. Skaane, O. Westerheim; Oslo/NO (jgun@uus.no) To study the effect of needle aspiration of breast abscesses as alternative to percutaneous pig-tail drainage or surgical incision. Methods and Materials: Ultrasonography (US) was performed on 63 patients with suspected breast abscesses. Patients with phlegmonous mastitis were treated with antibiotics only. US-guided needle aspiration was performed with a 14-18 G needle, if an abscess was diagnosed. The patients were scheduled for shortterm follow-up and the procedure was repeated in case of relapse. Patients with progression or no therapeutic response were referred for surgical treatment. Results: 9 of the 63 patients were excluded because of complex etiological factors. US showed phlegmonous mastitis in 4 patients. Of the remaining 50 patients having an abscess formation, 43 were successfully treated with needle aspiration and antibiotics (median abscess diameter 26 mm, range 7-105 mm). 24 abscesses had a diameter of more than 25 mm, of which only five had to undergo surgical drainage. Mean number of aspirations for each patient was 1.6 (range 1-5). Median number of days from first aspiration until treatment was considered successfully finished, was 13 days (range 4 to 70 days). Seven patients had to be referred for surgical incision. Treatment success was not influenced by abscess size or duration of symptoms before therapy. Conclusion: US-guided abscess aspiration in combination with antibiotics as a minimally invasive procedure is an alternative to percutaneous pig-tail drainage and surgical incision, even in patients with larger abscess formations. Purpose: To evaluate the mechanisms responsible for recurrence after varicose vein surgery with color duplex imaging (CDI). Methods and Materials: One hundred and thirteen limbs (93 patients, female: 69, mean age: 48 years), were preoperatively examined using CDI in order to detect the presence and the extent of venous reflux and to design the appropriate surgical procedure. All limbs were followed-up clinically and ultrasonographically one month and 5-years after surgery. Results: Forty-two sources of reflux were identified among 28 (25%) of the 113 operated limbs that had presented with recurrence during the follow-up period. The one-month postoperative examination revealed the preservation of 1) an incompetent greater saphenous vein (GSV) distal remnant in 6 limbs that were subjected to a limited stripping (6/42, 14.3%) and 2) a saphenous stump with incompetent tributaries in 2 limbs (2/42, 4.8%). The 5 year follow-up examination showed: 1) the presence of refluxing serpentine tributaries arising from the ligated SFJ as a result of neovascularization in 13 limbs (31%), 2) the expansion of reflux along the preoperatively competent GSV remnant in 4 limbs with limited stripping, due to chronic venous disease progression (CVDP) (9.5%), 3) the development of reflux in the preoperatively competent superficial veins (in non-operated areas) due to CVDP in 17 limbs (40.8%). The one-month postoperative ultrasonographic examination validated the sufficiency and the appropriateness of the surgery performed, and the 5-year examination, including both the operated and non-operated areas of the limb, revealed unavoidable causes of recurrences such as neovascularization and CVDP. Diagnostic accuracy and immediate discharge rate of a DVT pathway I. Britton 1 , P. Vlachou 2 , A. Morley-Davies 1 ; 1 Stoke-upon-Trent/UK, 2 Boston/UK (ingrid.morley-davies@virgin.net) Purpose: Implementation of a combined clinical, haematological, venometer and Doppler DVT diagnostic pathway, was assessed for diagnostic accuracy and immediate discharge rate. Methods and Materials: Over a five-month period, 200 patients with suspected DVT, had a Wells pre-test clinical probability score calculated. Exclusionary testing was performed for "low" and "moderate risk" groups with d-dimer and venometer. Positive results were clarified by Doppler. Patients were tracked through the pathway, and for subsequent events over a three-month period. Results: 57 had DVT (28.5%). 3 of 41 (7%) "low risk" patients, 18 of 86 (21%) "moderate risk" patients, and 36 of 73 (49%) of the "high-risk" group had DVT. 31 (15.5%) patients were discharged following exclusionary testing. Negative predictive value of venometer and d-dimer were 94% and 96% respectively, but PPV was poor at 38% and 25% respectively. One patient re-presented following a Purpose: Early detection and proper management of deep vein thrombosis (DVT) after major orthopedic surgery is very important to the prognosis. The aim of this prospective study was to evaluate the ability of indirect multidetector row CT venography (CTV) in detecting DVT after hip or knee arthroplasty in comparison with Doppler sonography. Methods and Materials: Sixty-two patients had undergone hip (n = 28), knee (n = 32) or both (n = 2) arthroplasty. For the detection of DVT, CTV and Doppler sonography was performed 8 -40 days after the operation. CTV (collimation 16 x 1.5 mm, table feed 24 mm) from diaphragm to feet was done. Doppler sonography was performed within 30 days (average 2 days) following CTV. In the detection of DVT, the ability of CTV was evaluated in comparison with Doppler sonography. CAM-180 dual-detector gamma camera. Qualitative assessment using visual grading (SLS) was performed. Quantitative parameters assessed were percentage ilio-inguinal node tracer uptake at 120 min, and tracer clearance from injection sites. Statistical analysis was performed to assess the data. Results: Lymphoedema was established in 14 oedematous extremities (primary = 9, secondary = 5) and excluded in 7 oedematous extremities for other reasons (Prader-Willi syndrome = 2, venous insufficiency = 2, surgery = 1, lipoedema = 2). Lymphatic function of 5 non-oedematous extremities was also analysed. Early lymphatic filling was observed in 7, and dermal flow in 4 extremities. For SLS, sensitivity = 71%, specificity = 40%, PPV = 83.3%, NPV = 25%; and for tracer clearance from injection site, sensitivity = 55.5%, specificity = 80%, PPV = 90.9%, NPV = 33.3% were calculated. Conclusion: Our findings suggest that visual grading of lymphoscintigrams may be standardised using summed lymphatic score and combined with a measure of tracer clearance from injection sites for establishing the diagnosis of lower extremity lymphoedema. High-resolution magnetic resonance lymphangiography in patients with primary and secondary lymphedema C. Purpose: To evaluate the feasibility of high-resolution magnetic resonance lymphangiography with intracutaneous injection of gadodiamide in patients with primary and secondary lymphedema. Methods and Materials: 15 patients with lymphedema of the lower extremities underwent high-resolution magnetic resonance lymphangiography. Gadodiamide was injected into the dorsal aspect of both feet. First, the extent and distribution of the lymphedema was evaluated using a heavily T2-weighted 3D-TSE sequence. For high-resolution magnetic resonance lymphangiography a 3D spoiled gradient-echo sequence (VIBE) was used. Results: No complications were observed during or after the examination. In all patients the lymphedema demonstrated an epifascial distribution. In 14 patients, lymphatic vessels extending from the injection site were reliably detected 15 minutes after injection. After 5 minutes, concomitant venous enhancement was detected in all patients. In 13 patients, the lymphatic vessels in the upper leg could be seen. In 14 patients, the inguinal lymph nodes with external iliac lymphatic pathways were reliably depicted at 35 minutes. The external iliac lymph nodes were observed merely in 6 patients. Collateral vessels with dermal back-flow were seen in 12 patients. A lymphocutaneous fistula and a lymphocele with the afferent lymphatic vessel were detected in one patient. Conclusion: High-resolution magnetic resonance lymphangiography is feasible in non-invasively visualizing the lymphatic vessels and accompanying complications in patients with primary and secondary lymphedema. The method is not aimed at the depiction of lymph node morphology, but provides complementary information about the lymphatic vessels when lymph nodes are examined with super-paramagnetic iron oxide particles. Methods and Materials: 59 consecutive males with FI were recruited. Symptoms assessed with Wexner Score. All patients followed standard investigation pathway starting with manometry, rectal sensation studies then anal endosonography (EAUS). Co-morbidities and anorectal histories were recorded 12 had anal surgery (8 haemorrhoidectomy, 2 anal fissure, 2 fistula-in-ano). 5 had rectal surgery. No significant difference in sphincter function with surgery vs no surgery. No significant difference in function was found when co-morbidities were assessed. Age > 60 yrs was associated with reduced Maximum Resting Pressure (MRP) EAUS found entirely normal sphincters in 38/59 (64.4%). 16 had thinning of IAS or EAS or both, but no defects. 5/59 (8.5%) defects were recorded, 4/5 post surgery. All rectal surgery patients had abnormal rectal sensation. Conclusion: Male incontinence is an existing problem that deserves more attention. It is infrequently attributed to sphincter defects, even post-surgery, and other causes should be investigated Posterior pelvic floor disorders: A prospective comparison with colpocistodefecography and dynamic vaginal endosonography S Rome/IT Biopsy and localisation techniques Moderators: W. Buchberger; Innsbruck/AT P. Viehweg Stainless-steel versus titanium needles for MR imaging-guided large core breast biopsy: Comparison for yield of breast tissue D. Flöry, M. Duenkelmeyer Purpose: To investigate the role of MR imaging in patients with peripheral neuropathies of the upper extremity in whom clinical evaluation (including electrodiagnostic studies) revealed unclear or ambiguous findings and to demonstrate the impact of MR imaging on therapy. Methods and Materials: Fifty patients with peripheral median, ulnar and/or radial neuropathies were referred to MR imaging at 1.5 T. In these patients, either the etiology (32/50) or the site and localization (18/50) of the lesion was unclear. A total of 50 MR imaging examinations were performed (1 shoulder/axilla, 2 upper arms, 26 elbows including the forearm, 21 wrists including the hand). Two radiologists reviewed all MR images with regard to nerve, muscle, and other soft-tissues abnormalities. A panel of a radiologist, a hand surgeon and a neurologist reviewed all clinical records. The correlation of clinical and radiological findings, and the impact of MR imaging on therapy were assessed. Cohen's kappa and Spearman's (p) rank correlation test were calculated. Results: A total of 17 neurogenic tumors, 20 compressive nerve lesions, and 10 cases of focal or multifocal neuritis were detected. In 13 of the 18 patients (72%), MR imaging demonstrated exactly the site and location of the neuropathy. MR imaging revealed the etiology in 27 of the 32 patients (84%). Cohen's kappa was 0.73 to 1.0 and p was 0.78 to 0.94. The impact on therapy was "major" in 29, "remarkable" in 18, and "no impact" in 3 patients. Purpose: To evaluate the accuracy of dynamic vaginal endosonography in detecting diseases of posterior pelvic floor as a possible alternative technique for early diagnosis, particularly in young patients. A comparison of ultrasound and radiological findings is presented. Methods and Materials: Thirty-seven consecutive females (age, range) with functionality impairment of posterior pelvic floor were enrolled after clinical evaluation. Exclusion criteria were age under 18 and previous history of vulvectomy. They underwent colpocistodefecography (DF) and dynamic endovaginal US examination (Siemens Antares). Recto-anal angle was measured during "squeeze" maneuver to evaluate the strength of voluntary musculature contraction and during straining to assess pelvic floor descent. Images were blindly evaluated by two radiologists experienced in pelvic disorders and the interobserver agreement was also calculated. The same measuring reference lines were obtained for each group of examinations. Results: No statistical differences were observed between colpocistodefecography and dynamic ultrasound measurement during the "squeeze" maneuver (p = 0.09) and during straining (p = 0.59). Presence of intussusception was blindly evaluated with a 94.5% agreement between ultrasound and defecography. Moreover US study allowed measurements of internal and external sphincters: sphincter thickness and degree of clinically evaluated faecal incontinence showed good correlation. Conclusion: We propose dynamic endovaginal ultrasound as a simple, non-invasive way to assess recto-anal angle functionality, allowing in the same time the evaluation of the thickness and integrity of the sphinterial complex whereas DF can provide only functional information. together with colpocistodefecography, anatomic evaluation of the perineal body in diagnostic evaluation of pelvic floor diseases. Purpose: To assess the feasibility of half-Fourier single-shot turbo spin-echo (HASTE) of the lung at 3 T using parallel imaging with 32-receiver channel technology. In an ongoing feasibility study 5 healthy volunteers were examined with HASTE on a new 32-channel 3 T MR imaging system (Magnetom TIM Trio, Siemens Medical Solutions) using a prototype 32-channel torso array coil (RAPID Biomedical). HASTE MR imaging (TR=800 ms, FOV=420x420 mm, matrix=256x256, slice thickness = 5 mm) was acquired at both end-inspiratory and end-expiratory breathhold using parallel acquisition techniques (GRAPPA, 64 reference lines) with acceleration factors ranging between 2 and 4. Depending on the acceleration factor, a minimal TE between 26 ms (PAT=2) and 16 ms (PAT=4) was achieved. The image quality of pulmonary and extrapulmonary anatomical structures and noise level was evaluated. Results: All images were of diagnostic quality. Using an acceleration factor of 4 image blurring was substantially reduced compared to lower acceleration factors resulting in sharp delineation of intrapulmonary and extrapulmonary structures. Expiratory scans provided a higher intrapulmonary signal intensity than inspiratory scans. As visually assessed, loss of signal to noise did not impair image quality even at acceleration factors of 4. Conclusion: HASTE MR imaging of the lung is feasible at 3 T. The use of high parallel imaging acceleration factors allows shorter echo times and reduces image blurring. Due to optimised multi-element coil geometry, sufficient signal-tonoise can be achieved despite high acceleration factors. Expiratory scanning may be favourable to compensate for susceptibility associated signal loss of the lung at 3 T. Detection of small pulmonary nodules in high-field MR at 3 T: Experimental evaluation of different pulse sequences using porcine lung explants Purpose: To compare and evaluate different MR imaging sequences for the detection of injected pulmonary nodules in porcine lungs. Methods and Materials: 84 agarose nodules of different diameter (range 2 mm-20 mm) were injected into 11 porcine lungs within a chest phantom. Three-dimensional (3D) gradient echo (GRE) and HASTE sequences were performed in axial and coronal orientations, followed by multislice CT (MSCT) as gold-standard in the diagnosis of pulmonary nodules. Images were read by 3 blinded observers with different diagnostic experience. Sizes and locations of the detected nodules were assessed. Sensitivity, specificity, positive and negative predictivevalue (PPV, NPV) for all lesions were calculated for each sequence. Results: The 3D-GRE sequence showed a sensitivity greater than 90% for all readers in nodules greater than 5 mm in diameter which was comparable to CT. In lesions smaller than 5 mm the sensitivity and PPV declined below 70%. The HASTE sequence showed a sensitivity greater than 86% in nodules over 5 mm. A specificity, sensitivity, PPV and NPV of nearly 100% was recorded for all sequences in nodules greater than 10 mm. The rate of false positive findings depended on the reader's experience, but was lower for the 3D-GRE than the HASTE sequence. The 3D-GRE sequence showed a higher overall accuracy than the HASTE sequence. The study confirms the feasibility of 3 T-MR for the detection of small solid lung lesions. In lesions greater than 5 mm, the sensitivity of the 3D-GRE All patients had a history of nephrectomy of contralateral kidney. RFA was indicated due to inoperability or due to high probability of complete renal failure after surgical enucleation of the tumor. If tumor-diameter was larger than 3 cm arterial embolization was performed prior to RFA to reduce heat sink effect and risk of bleeding. RFA was performed under CT-fluoroscopy using analgosedation and local anaesthesia. Follow-up examinations were after 6 and 12 weeks, then every three months using CT or MR imaging. Definition of success was lack of contrast enhancement and shrinking of the treated area. Results: Targeting of RCC using CT fluoroscopy was feasible in all patients. Within the follow-up period no major complication or tumor relapse occurred. In one patient a second RCC after was detected and treated with second RF ablation. None of the patients developed renal failure needing hemodialysis. In one of the patients bleeding into pararenal tissue was noticed, which stopped spontaneously. Conclusion: RFA for inoperable RCC in a solitary kidney is a very effective therapeutic option due to the small complication rate, no deterioration of the renal function and excellent tumor control. Purpose: Safety margins are a prerequisite to avoid local tumor recurrence after RF ablation of liver tumors. The aim was to determine the prognostic value of a volumetric necrosis/tumor quotient. Methods and Materials: Twenty-nine colorectal cancer (CRC) metastases and 24 hepatocellular cancer (HCC) nodules were treated by radiofrequency ablation. Volumetric measurements of the tumor volumes were performed prior to ablation and the necrosis volume was measured after ablation based on contrast-enhanced CT data sets. The quotient of necrosis and tumor volumes was calculated for all lesions. Follow-up exams were performed up to 1 year after ablation. The CRC metastases and HCC nodules showed a mean tumor volume of 12.8 ml and 11.5 ml. The mean ablation volumes were 37.6 ml and 29.5 ml, respectively. This resulted in post-/pre-ablation quotient of 5.08 ± 3.8 for mets and 4.16 ± 2.7 for HCC. The follow-up revealed local tumor recurrence in 16 of 29 (55%) metastases and 11 of 24 HCC (46%). The necrosis/tumor quotient was significantly smaller in local recurrence with 3.64 compared to 5.71 in lesions free of tumor (p < 0.002). CRC metastases, but not HCC nodules showed a correlation between initial tumor volume and local outcome.Conclusion: Although small lesions were treated and the area of necrosis was more than four times larger than the tumor volumes, local tumor control could be achieved only in the half of patients. However, the volumetric necrosis/tumor quotient can identify patients at risk for local tumor recurrence. Percutaneous CT-guided radiofrequency ablation of osteoid osteoma with multitined expandable electrode R. Cioni, L. Crocetti, R. Lencioni, V. Zampa, C. Cappelli, I. Bargellini, C. Della Pina, V. Consoli, C. Bartolozzi; Pisa/IT (lcrocetti@libero.it)Purpose: To determine the feasibility, safety and effectiveness of percutaneous CT-guided radiofrequency (RF) ablation of osteoid osteoma (OO) with a multitined expandable electrode. Methods and Materials: Eleven patients (male, age 14-58 years, median age 22 years) with single OO were enrolled in a prospective, single center, single-arm clinical trial. Diagnosis was obtained by clinical and imaging findings in 10 patients and by CT-guided biopsy in 1 patient. CT-guided RF ablation was performed under conscious sedation by using a 150-200-W generator and expandable multitined electrode (RITA Medical Systems, Mountain View, CA). Follow-up period ranged 3-22 months (mean 12 months ± 7) and included MR examinations and clinical visits performed 6 months after the procedure and at 12-month intervals thereafter. Results: RF ablation was technically feasible in all 11 patients (technical success 100%) without need of surgical access. No major complications occurredin particular, no skin burns were observed. Prompt pain relief and complete ablation of the nidus, as shown by 6-month MR, was achieved in 10 of 11 osteomas (primary effectiveness rate 91%). In one patient, recurrence of pain and persistent contrast uptake of the nidus at MR were present. The OO was effectively retreated by means of RF ablation. No recurrences were observed on follow-up. Conclusion: Percutaneous CT-guided RF ablation of OO performed with a multitined expandable electrode yields high effectiveness rates and has a favourable safety profile.Purpose: The aim of our study was to determine the efficacy of RFA in the treatment of osteoid osteoma regarding complication rate and duration of pain relief. Methods and Materials: Within 42 months, 22 patients (12 male, 10 female, age 9 to 45 years) suffering from osteoid osteoma (OO) were treated. In 8 of 22 patients the OO were localized in the lumbar (5 of 8) or thoracic spine (3 of 8), whereas the other OO were located in the femur (5 of 14), the acetabulum (2 of 14) or other long bones (radius, tibia). In children RFA was performed using general anaesthesia, in adults analgo-sedation and local anaesthesia was preferred. Ablation periods of 5 to 8 minutes at a temperature of 90 o C were used. Cooling of the skin was performed in osteoid osteoma next to the skin (tibia, radius) and an additional needle was inserted to do saline flushing if the OO was next to neural structures (spine). Primary success rate, complications, disease-free interval and follow-up time were evaluated. Results: Within the observation period (3-42 months) all of our patients were successfully treated and had no more complaints. After the first ablation 19 of 22 patients were free of pain, while in 3 patients the ablation was repeated to obtain complete response. No major complications occurred. In 2 patients minor complications (1 hematoma, 1 first degree skin burn) were observed. Conclusion: Radiofrequency ablation is a highly effective, efficient, minimally invasive and safe method of treating osteoid osteoma. Purpose: The aim of this paper is to review our treatment protocols for hepatomas and to highlight the efficacy of combined trans-arterial chemo-embolization (TACE) followed by RFA in the treatment of large lesions beyond 10 cm in diameter. Over 250 sessions of RF ablation have been performed for hepatomas in our centre since 2001. Our protocol is as follows: < 2 cm -RFA alone; 2-3 cm -RFA with overlapping ablation; 3-10 cm -TACE followed by RFA (± repeat sessions); > 10 cm: -multiple sessions of TACE followed by RFA; impaired liver function -bland embolization with lipiodol ± gelfoam followed by RFA. For lesions above 10 cm, treatment protocols in many centres advocate TACE alone. However, we are finding that the combination of TACE followed immediately by RFA for lesions above 10 cm has favorable outcomes. Repeat sessions of combined TACE and RFA are usually performed for these lesions at 3-6 week intervals, depending on the appearance of the lesion at follow-up, correlated with tumour markers. Results: Overall, results are extremely favorable. For large lesions, a regime of multiple sessions of chemoembolization and RFA yields good results with complete tumour eradication in many lesions, even those > 13 cm in diameter. Complication rates are low and post-procedure recovery is usually uneventful. Conclusion: Multiple sessions of TACE followed by RFA yield encouraging results for large hepatomas. Further studies are strongly recommended to establish the limits of this regime and its survival rates. Assessment of liver tissue following radiofrequency ablation: An experimental study with different imaging procedures G. Antoch, F.M. Vogt, P. Veit, N. Blechschmid, A. Bockisch, H. Kühl; Essen/DE (gerald.antoch@uni-essen.de)Purpose: Rim-like contrast enhancement on morphologic imaging and increased tracer uptake on FDG-PET may hamper differentiation of residual tumor from regenerative tissue following radiofrequency ablation (RF-ablation). The purpose of this study is to assess the typical appearance of normal liver tissue following RF-ablation when imaged with FDG-PET/CT and contrast-enhanced CT and MR imaging. Methods and Materials: Two lesions of 3 cm each were created by RF-ablation in normal liver parenchyma of 21 fully anesthesized mini-pigs. Post-interventional follow-up was performed by FDG-PET/CT as well as contrast-enhanced (CE) MR imaging and CE CT. Imaging was performed 1 day, 3 days, 10 days, 1, 2, 3 and 6 months post-intervention. All imaging procedures were evaluated qualitatively and quantitatively for areas of increased contrast enhancement and re-gions of elevated tracer uptake. Imaging findings were compared with histopathology. Results: One day following RF-ablation no increase in FDG uptake was visible while elevated contrast enhancement was noticed in the periphery of the necrosis with all morphologic imaging procedures. However, on further follow-up, an area of rim-like increase of FDG-uptake was detected surrounding the necrosis with PET/CT. Both, increased glucose metabolism as well as elevated contrast enhancement were detectable qualitatively and quantitatively for 6 months postintervention. Histology showed an hemorrhagic border converting into a regeneration capsule. Conclusion: FDG-PET and PET/CT may be of benefit over morphologic imaging procedures when assessing patients following RF-ablation for residual tumor, if performed within one day of intervention. Follow-up at a later time will be limited by visualization of peripheral hyperperfusion and tissue regeneration on functional and morphologic data sets. Comparative study of wet electrode-mediated and hypertonic salineenhanced radiofrequency ablation in in vivo porcine liver J. Lee, J. Han, J. Lee, S. Kim, K. Sohn, J. Chang, B. Choi; Seoul/KR (leejm@radcom.snu.ac.kr)Purpose: The aim of this study was to compare the in-vivo efficiency of wet radiofrequency ablation (RFA) with that of hypertonic saline (HS)-enhanced RFA for creating larger coagulation necrosis. Methods and Materials: Twenty six ablation zones were created in seven pigs with five different regimens: Group A, RFA with 6 mL of 36% HS preinjection; group B, wet RFA with infusion of 14.6% HS at 1 mL/min; group C, dry RFA; group D, 6 mL of 36% HS injection; group E, 12 mL of 14.6% HS infusion. RF energy was applied to the liver for 12 minutes using a 200 W generator and a cooled-wet electrode. The dimensions of the ablation zones and delivered energy during RFA were then compared between the groups. Results: The mean delivered energy was higher in group B (104.6 kJ) than in group A (97.2 kJ), which was higher than in group C (49.5 kJ) (p < 0.05). Wet RFA (55.0 ± 18.1 cm 3 ) created greater coagulation necrosis than HS-enhanced RFA (31.7 ± 8.8 cm 3 ) or dry RFA (11.1 ± 3.5 cm 3 )(p < 0.05). An infusion of 36% or 14.6% HS into the liver induced coagulation necrosis measuring 2.3 ± 2.6 cm 3 and 0.9 ± 0.9 cm 3 , respectively. Conclusion: Wet RFA is more efficient in creating a large coagulation necrosis than either the dry RFA or HS-enhanced RFA. MR-guided galvano-therapy for the treatment of localized prostate cancer: Results of a phase-III trial T.J. Vogl 1 , F. Mayer 2 , H. Mayer 2 , S. Zangos 1 ; 1 Frankfurt a. Main/DE, 2 Regensburg/DE (t.vogl@em.uni-frankfurt.de)Purpose: To evaluate a novel technique of MR-guided galvano-therapy (MR-GG) (electro-chemical therapy (ECT)) for the treatment of prostate cancer in a phase-III trial. Methods and Materials: Thirty three patients with histologically proven prostate cancer were prospectively treated with MR-guided galvano-therapy. After transgluteal puncturing of the prostate, under local anesthesia two MR compatible electrodes were positioned under MR guidance in the periphery of the right and left lobes of the prostate. The patients were treated three times at a weekly time interval. A total of 350 Coulomb was applied to the localized cancer and the prostate gland. Follow-up was performed after three, six and twelve months, respectively, using laboratory data (PSA) and endorectal MR imaging. Results: All patients tolerated the MR-GG well without any major side effects. Twelve of the thirty three patients had some reversible urinating problems. In the follow-up MR evaluation 12 months post therapy twenty four (72.7%) of the patients were classified as "stable" with a constant lesion size. In five patients (15.2%) MR imaging showed a homogenization of the neoplastic area, in two patients (6%) we found a size reduction of 30%, and two patients (6%) presented an increase in size of 20% and 30%, respectively. The PSA values showed a decrease in the twelve-month control from an average PSA of 13.5 ng/ml to 5.3 ng/ ml, which is equivalent to a reduction of 60.8% and statistically significant (p > 0.01). Conclusion: MR-guided galvano-therapy possibly results in local control of limited prostatic carcinoma with a concomitant reduction in the PSA value.Results: Both telemetrically and wire based obtained ASP measurements showed a significant increase (p < 0.05) after generating a type I endoleak. For a fluid filled aneurysm sac the exemplary results are averaged over all investigated sizes and positions. With WPS systolic ASP shown as the percentage of IAP increased from 10.1 ± 1.1 to 72.4 ± 1.5 (TPS: 14.6 ± 4.2 to 77.0 ± 2.2), diastolic ASP from 13.6 ± 2.2 to 111.7 ± 3.0 (TPS: 17.3 ± 7.5 to 113.8 ± 4.6), mean ASP from 11.6 ± 1.5 to 93.6 ± 0.7 (TPS: 16.2 ± 5.8 to 98.8 ± 4.5) respectively. Conclusion: Type I endoleaks of all investigated sizes cause an increase in ASP, high enough to be telemetrically detected at all investigated positions even in a thrombosed aneurysm sac. Therefore a telemetric pressure sensor might be integrated into the EVAR procedure and become a routinely used device for follow-up. Eight Purpose: To compare the impact of suprarenal and infrarenal fixation of endografts on renal artery complications and late postoperative renal function. A retrospective analysis of the data of 102 patients with unruptured infrarenal abdominal aortic aneurysms who underwent endovascular repair (EVAR) was performed: 60 patients were treated with a suprarenal fixation endograft (Cook Zenith, Medtronic Talent) (group A) whereas in 42 patients an infrarenal fixation device (Gore Excluder, Medtronic AneuRX) (Group B) was implanted. Baseline patient demographic data, pre-and post-procedural renal function (CrCl: creatinine clearance), incidence of renal artery (steno-obstructive disease) and procedure-related (proximal type-I endoleak and migration rates) complications, detected on post-operative surveillance performed with CT-angiography (1, 3, 6 and 12 months) were collected and compared. Results: No significant differences were found between the two groups in terms of demographic data and pre-operative risk factors. Proximal neck length was significantly shorter in group A (2.14 ± 0.84) than in group B (3.41 ± 1.21). CrCl decreased during the follow-up period in both groups (Group A: 64.24 ml/min to 60.66 ml/min; Group B: 72.41 ml/min to 68.73 ml/min) without statistically significant differences in terms of rate of CrCl decrease during the follow-up period as well as renal artery and procedure-related complications. Conclusion: In our experience, suprarenal fixation, with proximal extent of the proximal anchoring zone, allowed us to expand the inclusion criteria for EVAR without increasing renal artery and procedure-related complication rates as well as the likelihood of postoperative renal impairment. Endovascular treatment of thoracic aorta aneurysms and dissections after transposition of supra-aortic trunks R. Rostagno, R. Llorens, T. Zander, I. Zerolo, S. Baldi, E. Sanabria, M. Maynar; Santa Cruz de Tenerife/ES (romandrostagno@yahoo.com.ar)Purpose: To report our experience with a combined repair of aortic arch aneurysms or dissections by sequential transposition of the supra-aortic trunks and endovascular stent-graft placement in cases without a proximal neck. Methods and Materials: Nine patients were selected for transposition of supraaortic trunks previous to endovascular treatment. The indications were: thoracic aneurysms in 6 cases and dissections in 3. TST was performed using minithoracotomy without extracorporeal circulation. Endovascular procedures were performed in a second operative time in 8 cases. Long term follow-up was planned at 1, 3, 6 months and yearly. Results: Technical success was achieved in all but one cases, who died after surgery due to major stroke. Aneurysms were excluded and proximal entry occluded in all cases. Early complications were: 1 acute limb ischemia, 1 minor stroke and 1 Purpose: To report the long term follow-up results of endovascular aneurysm repair (EVAR) of inflammatory aortic aneurysms (IAA). In a retrospective cohort study we identified 8 patients suffering from IAA based on findings of the pre-interventional computed tomography angiography (CTA) who underwent EVAR. Primary and follow-up images were reviewed by two observers for changes of the aneurysm sac diameter, of the periaortic fibrosis (PAF), of potential renal impairment and the procedure related and long term complications. Results: EVAR was successfully completed in all patients. 5 patients had a minimum follow-up period of 36 months (median 65 months). The maximum diameter of their aneurysm sac decreased from median 57 mm (47-95) to median 45 mm (36-60), resulting in a median relative regression of 21.1% (4.8-63.1). PAF regressed in all of these patients, prior hydronephrosis was present in two patients, remaining unchanged in one of them and disappearing in the other one. In two patients a type 3 endoleak was detected and overstented during the follow-up period. Conclusion: EVAR of IAA excludes the aneurysm and seems to reduce the size of the aneurysm sac and the extent of PAF with an acceptable peri-interventional and long term morbidity. Purpose: Open surgery is associated with considerable morbidity and mortality in patients with complicated type B aortic dissection. We report our 5 year experience for the use of thoracic aortic stent grafts in these patients. Thoracic stent grafts were deployed in 35 patients (28 men) with acute complicated type B aortic dissections (rupture, end organ ischaemia or continued pain) and chronic type B dissections which had undergone aneurysmal dilatation. The patients were followed up with computed tomography at 1.3 and 6 months post stenting and then yearly. Results: Initial technical success was 97%. One patient required a further procedure on the initial admission to seal other entry sites. Three patients died in the post operative period giving a 30 day mortality of 8.5%. In two patients (6%) proximal (type A) extension of the dissection occurred subsequent to stenting. No patient developed post operative paraplegia. Follow-up averaged 14 months (range 1-57 months). In that time the false lumen failed to thrombosis or was reperfused in six patients (17%). There were two late deaths which were unrelated to the procedure (6%). The use of stent grafts in selected patients with type B dissection provides an attractive alternative treatment option. The procedure can be accomplished with acceptable initial morbidity and mortality when compared to surgery. However long term follow-up is essential to monitor the durability of these devices. Purpose: To evaluate the optimal technique for first pass renal perfusion imaging at 1.5 T. Methods and Materials: 15 healthy male volunteers (25-38 years) underwent a prospective intra-individual comparison study of renal perfusion measurement on a 1.5 T MR-scanner (Avanto, Siemens). Each volunteer underwent three exams on three different days with three different sequences after administration of 7 ml Gd-BOPTA (MultiHance®, Bracco-Altana Pharma): Saturation-recovery (SR) turboFLASH without parallel imaging (TFl), SR-trueFISP (TFi) and SR-turbo-FLASH with parallel imaging (TFl-PI). The temporal resolution of the sequences was 4 slices/s (TFl, TFi) and 6 slices/s (TFl-PI). Field of view (450x373 mm²) and matrix (192x134) were kept constant for all sequences. Evaluation comprised the semiquantitative assessment of first pass perfusion parameters mean transit time (MTT), time to peak signal intensity (TTP), maximal upslope (MUS) and maximal signal intensity (MSI). In addition the baseline and peak signal-to-noise ratio (SNR) were assessed as well as the image quality which was rated by two radiologists in consensus. Purpose: Magnetic resonance diffusion weighted imaging (MR-DWI) is useful to assess proton motion by computation of an apparent diffusion coefficient (ADC). This property could be used to assess renal damage, with special regard to unilateral dysfunction. The aim of this study is to estimate the correlation between ADC and stage of chronic renal failure. Methods and Materials: Twenty patients (mean age 51.6 years, range 38-66 years) underwent MR examination on a 1.5 T system; in 10 cases there was a history of hypertension, in 3 diabetes, in 3 obstructive uropathy, Takayasu's disease and nephrovascular hypertension respectively. Four subjects without known renal pathology were used as a normal reference. Glomerular filtration rate (GFR) assessed by Cockroft-Gault's equation was used as a functional marker. Imaging protocol consisted of the following sequences: Fast-field echo T1-weighted, TSE T2-weighted and Single-Shot DWI echo-planar sequence (b-value 600 sec/mm2), with application of SENSitivity encoding factor of 2, by using a four channel phasedarray body coil. In five patients SS-DWI sequence was repeated after iv administration of 1 mg of furosemide. The ADC was correlated with the GFR by Pearson's test. Results: ADC was 2.37 ± 0.25*10-3 mm2/sec in subjects with normal GFR and 2.05 ± 0.33*10-3 mm2/sec (p < 0.05) in subjects with altered GFR; a significant difference was found between stage III and IV (p < 0.01). A good correlation was found between GFR and ADC (r = 0.76; p < 0.01), with no significant modification after furosemide administration. Purpose: Evaluation of a new standardized ultrasound (US)technique for diagnosis of acute rejection of kidney grafts. Methods and Materials: 22 kidney recipients underwent US examination following administration of 1.6 ml US contrast medium (USCM, SonoVue™) 6 days after transplantation. The examinations were performed on an Aplio US system (Toshiba™). The difference in time from the first increase in signal intensity between the renal artery and renal cortex was determined. Subsequently the temporal course of contrast enhancement in the area of the renal artery following the first peak was recorded over 10 sec and depicted in color using a Windowsbased software. The resulting color-time images were evaluated by three readers who rated the images on an analogue scale from 1 (normal) to 5 (abnormal). Results: Twelve of the 22 patients had an uneventful clinical course. US demonstrated rapid inflow of the USCM into the renal cortex. The calculated time difference was 1.0 ± 0.4 s. The score assigned to the parametric images was 1.7 ± 0.8. Eight of the 22 patients received a biopsy and showed histologically proven rejection. The time difference in this group was twice as high as in the normal group (2.2 ± 0.7 s, p < 0.05). The scores were in the abnormal range (3.7 ± 1.6, p < 0.05). Two patients with perirenal hematoma also showed high scores, without rejection. Conclusion: Acute rejection and perirenal hematoma are associated with a delayed signal increase in the renal cortex. This information can be provided with a single image with standardized color display of the temporal course of USCM inflow. Results: Both methods demonstrated the same small arterial branches of the circle of Willis. Besides the first order branches, e. g. frontopolar a., callosomarginal a. and pericallosal a., also second and third order branches, paracentral a. and medial frontal branches, were seen. The PCMRAs exhibited a better background subtraction compared to the TOF sequence. In comparison to clinical 1.5 T/3 T images, unenhanced MRA of the brain at 7T provides superior vessel contrast and conspicuity. Ultrahigh field MRAs demonstrate higher order branches with smaller calibers than conventional MRAs. Conclusion: This study demonstrates the feasibility of both TOF and PCMRA of the brain in a whole body 7T MR scanner with improved arterial contrast and vessel conspicuity compared to lower clinical field strengths. Optimized image processing in susceptibility weighted MR-imaging (SWI) A. Deistung 1 , A. Rauscher 1 , J. Sedlacik 1 , S. Witoszynskyj 2 , M. Barth 3 , J.R. Reichenbach 1 ; 1 Jena/DE, 2 Vienna/AT, 3 Nijmegen/NL (andreas.deistung@web.de) Purpose: SWI visualizes small cerebral veins and susceptibility differences by combining magnitude and phase information. Prior to using phase images in SWI, phase-wraps have to be eliminated. This study investigated the effects of different phase-unwrapping algorithms and phase-masks (weighting functions). Additionally, a technique for visualizing the whole SWI data over maximum intensity projection (MIP) was tested. Methods and Materials: High-resolution 3D scans were acquired with a velocity-compensated gradient-echo sequence (TR/TE/FA = 67 ms/40-50 ms/25°) on a 1.5 T scanner (Vision, Siemens, Germany). Different filter sizes for homodyne filtering (2D and 3D) were analyzed in 6 healthy volunteers. The results were compared with venograms created using a region-growing phase-unwrapping algorithm in image-space. Corrected phase information was weighted with different functions and multiplied with the magnitude. For the MIP data were masked and inverted. Results: The conspicuity of vessels depended critically on the quality of phaseunwrapping. Good results were obtained with filter sizes of 25% for 2D and 50% of each k-space dimension for 3D. With homodyne filtering there exists a tradeoff between CNR and phase-wrap suppression in regions with field inhomogeneities. For transverse slices the optimum weighting function maps negative phases between zero and one and sets positive phases to one. The MIP visualizes venous vessels well, whereas tissue is suppressed. Conclusion: Optimized homodyne filters provide a robust method for SWI data processing, whereas in regions with field inhomogeneities phase-unwrapping by region-growing yields superior results. Performing a MIP of SWI data allows visualization of the vessel network similar to other angiographic methods. A technique for assessing changes of connectivity of motor pathways in patients with post-traumatic head injury and mostly recovered motor functions with 3 Tesla fMRI and DTI G. Luccichenti, A. Cherubini, P. Péran, C. Barba, R. Formisano, U. Sabatini; Rome/IT (g.luccichenti@email.it) Purpose: To develop a method for assessing functional plasticity of white matter in patients with post-traumatic head injury. Methods and Materials: Six right-handed patients with post traumatic head injury involving motor pathways with hemiparesis mostly recovered after six months from trauma, underwent the following MR protocol using a 3 T scanner: 1) fMRI right and left hand finger tapping using a standard block acquisition design; 2) DTI with the following parameters: 52 images with 1x1x2 mm 3 voxel size in axial plane, NEX=16, b=0 and b=1000 s/mm 2 in 12 directions; 3) 3D FLASH (1x1x1 mm 3 voxel) and PD-T2 3 mm-thick images. The same protocol was performed in six age-matched right-handed healthy volunteers. Diffusion and fractional anisotropy maps were generated from DTI acquisition. Conventional fiber tracking reconstructions of pyramidal motor pathways were generated. Probabilistic connectivity maps of pyramidal motor pathways were generated using cortical activation obtained from finger tapping fMRI as seed point. Results: In all patients, conventional fiber tracking algorithms showed structural integrity of pyramidal motor pathways, that were in all cases close but not involved by post-traumatic lesions. Finger tapping of the hand contralateral to the side of the lesion showed bilateral activation, while connectivity maps generated from these active cortical areas showed increased probability of connection with pyramidal pathway contralateral to the side of the lesion (omolateral to finger tapping) through the corpus callosum. Conclusion: Probabilistic quantification of connectivity of brain areas may potentially allow the assessment of white matter plasticity, representing a powerful tool for understanding mechanisms of recovery and response to therapy. Traumatic diffuse axonal injury (DAI): Diffusion-weighted MR imaging findings Y. Li, X. Feng, W. Tang; Shanghai/CN (liyuan5714@163.com) Purpose: Diffuse axonal injury (DAI) accounts for a significant portion of primary intra-axial lesions in cases of traumatic brain injury. In this study we want to use diffusion-weighted MR imaging to characterize DAI in the setting of acute and subacute traumatic brain injury. Methods and Materials: Nineteen patients (13 men and 6 women) ranging in age from 26 to 63 years were examined with conventional MR imaging (including fast T2-weighted, FLAIR) as well as diffusion-weighted MR imaging at 1 to 18 days after closed head trauma. According to the Glasgow Coma Scale (GCS), 14 patients suffered from severe head injury (GCS < 8) and 5 patients had moderate head injury (GCS 9-12). Lesions were characterized as DAI on the basis of their location and their appearance on conventional MR images. Trace apparent diffusion coefficient (ADC) maps were computed off-line with the diffusion-weighted and base-line images. Areas of increased signal were identified on the diffusionweighted images, and regions of interests were used to obtain trace ADC values. Results: In the nineteen patients studied, lesions were located in the cerebral hemispheric gray-white matter interface and subcortical white matter (14 cases), body and splenium of corpus callosum (3 cases), basal ganglia (1 csaes), dorsolateral aspect of brainstem (1 case) and isotropic diffusion-weighted images showed areas of increased signal with correspondingly decreased ADC. In one case, decreased ADC was seen 18 days after the initial event. Conclusion: Decreased ADC can be demonstrated in patients with DAI in the acute setting and may persist into the subacute period, beyond that described for cytotoxic edema in ischemia. Purpose: To use diffusion tensor imaging (DTI) to semiquantitatively analyze the severity of diffuse axonal injury (DAI) in cases of traumatic brain injury (TBI). Twelve closed head trauma patients were prospectively studied with conventional MR imaging and DTI with a Siemens 3-Tesla MR system within fourteen days of injury. The fractional anisotropy (FA) values were measured at multiple regions of interest (ROI) by two radiologists in consensus.Results were compared with those of thirteen healthy control subjects. The FA values were correlated with the Glasgow Coma Scale (GCS) score in the trauma group.Results: In comparison with the control group, the mean FA values of patients were significantly reduced within the posterior limb of the internal capsule (P < 0.001) and corpus callosum (P < 0.05), no significant reduction within the thalami (P > 0.05). The mean FA value of the posterior limb of internal capsule was significantly correlated with GCS score in trauma group (r = 0.823, P < 0.001).No correlation was found between the FA value of thalami and GCS scores in the trauma group (P > 0.5). Purpose: Both stainless-steel and titanium needles are commercially available for MR imaging guided large-core breast biopsy (LCBB). The purpose of this study was to compare these needles for their yield of breast tissue. Methods and Materials: Six 14G LCBB needles consistent of stainless steel and three consistent of titanium alloys were included. Each needle was used to obtain 50 specimens from fresh turkey breasts. Turkey breasts were immobilized using a stereotactic device (Noras, Germany) and positioned on the Six had ductal carcinoma in situ, and 6 had invasive ductal carcinoma. 60 and 180 seconds after administration of 370 mgI/ml non-ionic contrast media (CM), CT scanning was started. Radiologists identified abnormal density spots on 3 mm slice transverse images and marked foci in a felt-tip pen. Surgeons enucleated tumors with a 1 cm margin from marking. We considered the margin as positive when foci existed within 5 mm from margin. We evaluated a rate of positive margin, distance between foci and margin, and block number and axis of positive foci. Results: Of the 12 cases, 5 (42%) had a positive margin. Mean + SD of distance between margin and foci were 19 + 13 mm. The number of positive block was within two in all margin-positive cases. Of 9 foci in margin-positive cases, mean + SD of distance between margin and foci was 4.4 + 1.1 mm. Six of 9 positivemargin foci (67%) existed on an axis from nipple to focus center. Conclusion: Since more than half of positive foci existed on the nipple-focus axis and mean distance between margin and foci was 4.4 mm, improvement of the results would be expected if transverse images are examined in more detail and the surgical margin is slightly elongated. Purpose: To assess a potential clip displacement of two biopsy marker clips following stereotactic 11-gauge vacuum assisted breast biopsy (VABB). Conclusion: Doppler sonography is time consuming and operator-dependent in the detection of DVT. In this study, the diagnostic ability of CTV was comparable to that of Doppler sonography and beam-hardening artifact due to arthroplasty did not become a serious problem. Therefore, CTV is a promising alternative modality in the evaluation of post-arthroplasty patients. Methods and Materials: Forty-seven consecutive patients (18 men, 29 women; mean age, 50.8 years; range, 13-76 years) confirmed as having acute DVT on CT underwent follow-up CT at least 2 months later. Mean duration of follow-up CT was 10. 6 months (range 2-42 months). Localization and extent of primary DVT on initial CT, intraluminal residual thrombi, decreased caliber, luminal obliteration, intravenous fibrotic bands, ipsilateral muscular enlargement, and ipsilateral subcutaneous edema on follow-up CT were assessed retrospectively by two radiologists in consensus. Frequencies of residual thrombi, decreased caliber, luminal obliteration, and fibrotic bands were calculated on each affected venous segment; iliac, femoral, popliteal, and calf vein. Results: Nine patients (19.1%) were completely normal on follow-up CT. Thirtyeight patients (80.9%) showed abnormal findings on follow-up CT; intraluminal residual thrombi (n = 7, 18.4%), fibrotic bands (n = 23, 60.5%), decreased caliber (n = 16, 42.1%), luminal obliteration (n = 16, 42.1%), ipsilateral muscular enlargement (n = 28, 73.7%), and ipsilateral subcutaneous edema (n = 7, 18.4%). Most commonly affected site was iliac vein (n = 29/37, 78.4%), followed by femoral vein (n = 29/45, 64.4%), popliteal vein (n = 22/43, 51.2%), and calf vein (n = 15/34, 44.1%), in order of frequency. Conclusion: Fibrotic bands, decreased caliber, luminal obliteration, and ipsilateral muscular enlargement are common CT findings of chronic DVT in the lower extremity. Indirect upper extremity (UE) multidetector-row computed tomographic (MDCT) venography: Initial technical experience J.C. Hellinger, D. Fleischmann, S. Cheung, G.D. Rubin; Stanford, CA/US (jchellinger@stanford.edu) Purpose: To assess the UE central and peripheral venous enhancement patterns when performing indirect UE-MDCTV. Methods and Materials: 17 UE 16-channel MDCT venograms were performed for either vascular mapping (N=6), occlusive disease (N=5), failing hemodialysis access (N=3), trauma (N=2), or vasculitis (N=1). Slice thickness (1.25 mm), reconstruction interval (0.8), and table speed (55 mm/sec) were identical. Images were acquired 60 seconds following the arrival of ionic contrast medium (concentration 350 mgI/ml; volume 90-140 ml; rate 3-5 cc/sec) to the aortic arch. For each UE, enhancement values were recorded in 8 regions (SVC, brachiocephalic, subclavian, axillary, brachial, upper cephalic, median cubital, and forearm deep veins) resulting in a total of 136 analyzed venous segments. All studies were reviewed by two radiologists for qualitative enhancement (0-4 scale). Results: 96% (N=130) of the vascular segments were well depicted and of diagnostic image quality (mean 3.25). Segments deemed to have poor enhancement (N=6) had attenuation values of < 90 HU (mean 1.2). The mean venous attenuation value across all segments was 137 HU ± 23 SD (92-182, 95% CI). Central veins (N=51, mean 147 HU) yielded 13% greater enhancement (p = 0.005) than peripheral veins (N=85, mean 130 HU). Within the central venous segments, 100% had values above 95 HU (range 103-195 HU) and 90%(N=46) achieved values ≥ 115 HU. Peripheral veins revealed no significant difference between venous attenuation in the upper arm (mean 131 HU) or elbow/forearm (mean 128 HU). 93% of peripheral venous segments reached values above 95 HU (range 95-214 HU) while 59% (N=48) achieved values ≥ 115 HU. Conclusion: Indirect MDCT venography can be used to image the UE central and peripheral veins. A delay of 60 seconds after aortic arch enhancement is necessary to achieve robust UE venous enhancement. Impact of gadolinium MR venography on the management of patients with haemodialysis-related thoracic venous disease P.S. Goh, S.C. Wang, C. Au; Singapore/SG (dnrgohps@nus.edu.sg) Purpose: To assess the impact of gadolinium MR venography on the management of patients with haemodialysis-related thoracic venous disease. Methods and Materials: Retrospective study of 44 consecutive patients with suspected haemodialysis related thoracic venous disease presenting with head or arm swelling, failure of AV fistula or dialysis line placement, or history of multiple central lines was performed over 20 consecutive months. MR imaging findings, results of subsequent interventional procedures, and medical records were reviewed. Scans were performed on a GE or Siemens 1.5 T MR system, with fast SPGR or FLASH sequences, an injection system for triggered gadolinium delivery, and multiphasic image acquisition. Post-processing with image subtraction, and both maximum intensity and 3D volumetric projections from different angles of view was used. Results: 21 female and 23 male patients with end stage renal disease, between 17 to 80 years old were evaluated. Multiple stenotic and occlusive lesions were demonstrated, most commonly involving the right internal jugular vein (23). 18 patients had subsequent interventional procedures -9 had venography confirming mediastinal venous stenosis with 5 successful venoplasty and stenting, and the other 9 had dialysis catheter venous access successfully via the non affected intenal jugular vein (7) or subclavian vein (2). MR findings were also used in 23 patients to plan AV fistula creation or revision, and in 9 to refer for peritoneal dialysis. Conclusion: Gadolinium MR venography is a useful non-invasive investigation which has significant impact for optimising the management of patients with thoracic venous disease. Purpose: To evaluate anatomic structures of patients with idiopathic subclavian vein thrombosis (SVT) compared to healthy controls and to evaluate a potential correlation between clinical provocative tests and the results of MR imaging (MRI). Methods and Materials: Both thoracic outlets of 29 consecutive patients with idiopathic SVT and 29 healthy controls were investigated by MR imaging with the arms alongside the body, patients were additionally imaged after a postural maneuver. We performed analyses of the costoclavicular and the interscalene spaces and evaluated the thoracic outlets for the presence of aberrant muscles. Results: Asymmetric interscalene spaces were not related to the presence of SVT. We detected an aberrant muscle, the musculus subclavius posticus, in four shoulders of patients, three of them symptomatic. In the remaining patients and in all controls no anatomical variant was detected. The minimum costoclavicular distance (MCD) of patients was significantly smaller than in controls (P < 0.001). The thickness of the subclavius muscle did not differ between patients and controls. The fat surrounding the subclavian vessels had a significant influence on the MCD, because patients with less fat exhibited a smaller MCD. The results of clinical provocative tests were not significantly related to any measurement or any anatomical structure determined by MR imaging and exhibited a sensitivity of about 60% for detection of arm sides with SVT. Conclusion: A smaller MCD was the only anatomical finding in patients with spontaneous SVT. Clinical evaluation including provocative tests had a low sensitivity and did not correlate with anatomical conditions evaluated by MR imaging. Lymphoscintigraphy in lower extremity lymphoedema: Relevance of visual grading using summed lymphatic score A. Singh, W.E. Svensson, D.J. Towey, L. Rahman, J. Murrell, K.S. Nijran, A. Al-Nahhas; London/UK (dr_asin@yahoo.co.uk) Purpose: Lymphoscintigraphy is used in the evaluation of lower extremity lymphoedema. We evaluated the role of visual grading using a Summed Lymphatic Score (SLS) and measure of tracer clearance from injection sites in the interpretation of lymphoscintigrams. Methods and Materials: 99m Tc-albumin nanocolloid lymphoscintigraphy was performed in 13 patients (7 males, 6 females; age range 18-74) for 26 lower extremities (21 oedematous and 5 non-oedematous). Clinical records were reviewed for all patients. Images were acquired at 30 and 120 minutes using a Siemens E.