key: cord-0034015-2cv9tr7t authors: Rutala, William A.; Weber, David J. title: Disinfection, Sterilization, and Control of Hospital Waste date: 2014-10-31 journal: Mandell, Douglas, and Bennett's Principles and Practice of Infectious Diseases DOI: 10.1016/b978-1-4557-4801-3.00301-5 sha: 11affb5d07de4970d928a698779fa119d128a350 doc_id: 34015 cord_uid: 2cv9tr7t nan William A. Rutala Each year in the United States there are approximately 53 million outpatient surgical procedures and 46 million inpatient surgical procedures. 1 For example, there are at least 10 million gastrointestinal endoscopies per year. 2 Each of these procedures involves contact by a medical device or surgical instrument with a patient's sterile tissue or mucous membranes. A major risk of all such procedures is the introduction of infection. Failure to properly disinfect or sterilize equipment carries not only the risk associated with breach of the host barriers but also the additional risk for person-to-person transmission (e.g., hepatitis B virus) and transmission of environmental pathogens (e.g., Clostridium difficile). Achieving disinfection and sterilization through the use of disinfectants and sterilization practices is essential for ensuring that medical and surgical instruments do not transmit infectious pathogens to patients. Because it is unnecessary to sterilize all patient care items, health care policies must identify whether cleaning, disinfection, or sterilization is indicated based primarily on the item's intended use. Multiple studies in many countries have documented lack of compliance with established guidelines for disinfection and sterilization. 3, 4 Failure to comply with scientifically based guidelines has led to numerous outbreaks of infectious diseases. 2, [4] [5] [6] [7] [8] In this chapter, which is an update of previous chapters, [9] [10] [11] [12] [13] a pragmatic approach to the judicious selection and proper use of disinfection and sterilization processes is presented, based on well-designed studies assessing the efficacy (via laboratory investigations) and effectiveness (via clinical studies) of disinfection and sterilization procedures. In addition, we briefly review the management of medical waste in health care facilities. Sterilization is the complete elimination or destruction of all forms of microbial life and is accomplished in health care facilities by either physical or chemical processes. Steam under pressure, dry heat, ethylene oxide (ETO) gas, hydrogen peroxide gas plasma, vaporized hydrogen peroxide, and liquid chemicals are the principal sterilizing agents used in health care facilities. Sterilization is intended to convey an absolute meaning, not a relative one. Unfortunately, some health care professionals as well as the technical and commercial literature refer to "disinfection" as "sterilization" and items as "partially sterile. " When chemicals are used for the purposes of destroying all forms of microbiologic life, including fungal and bacterial spores, they may be called chemical sterilants. These same germicides used for shorter exposure periods may also be part of the disinfection process (i.e., high-level disinfection). Disinfection describes a process that eliminates many or all pathogenic microorganisms on inanimate objects, with the exception of bacterial spores. Disinfection is usually accomplished by the use of liquid chemicals or wet pasteurization in health care settings. The efficacy of disinfection is affected by a number of factors, each of which may nullify or limit the efficacy of the process. Some of the factors that affect both disinfection and sterilization efficacy are the prior cleaning of the object; the organic and inorganic load present; the type and level of microbial contamination; the concentration of and exposure time to the germicide; the nature of the object (e.g., crevices, hinges, and lumens); the presence of biofilms; the temperature and pH of the disinfection process; and, in some cases, the relative humidity of the sterilization process (e.g., with ETO). By definition then, disinfection differs from sterilization by its lack of sporicidal property, but this is an oversimplification. A few disinfectants will kill spores with prolonged exposure times (e.g., 3 to 12 hours) and are called chemical sterilants. At similar concentrations but with shorter exposure periods (e.g., 12 minutes for 0.55% orthophthalaldehyde) these same disinfectants will kill all microorganisms with the exception of large numbers of bacterial spores and are called high-level disinfectants. Low-level disinfectants may kill most vegetative bacteria, some fungi, and some viruses in a practical period of time (≤10 minutes), whereas intermediate-level disinfectants may be cidal for mycobacteria, vegetative bacteria, most viruses, and most fungi but do not necessarily kill bacterial spores. The germicides differ markedly among themselves primarily in their antimicrobial spectrum and rapidity of action. Cleaning, on the other hand, is the removal of visible soil (e.g., organic and inorganic material) from objects and surfaces, and it normally is accomplished by manual or mechanical means using water with detergents or enzymatic products. Thorough cleaning is essential before high-level disinfection and sterilization because inorganic and organic materials that remain on the surfaces of instruments interfere with the effectiveness of these processes. Also, if the soiled materials become dried or baked onto the instruments, the removal process becomes more difficult and the disinfection or sterilization process less effective or ineffective. Surgical instruments should be presoaked or rinsed to prevent drying of blood and to soften or remove blood from the instruments. Decontamination is a procedure that removes pathogenic microorganisms from objects so they are safe to handle, use, or discard. Terms with a suffix "-cide" or "-cidal" for killing action also are commonly used. For example, a germicide is an agent that can kill microorganisms, particularly pathogenic organisms ("germs"). The term germicide includes both antiseptics and disinfectants. Antiseptics are germicides applied to living tissue and skin, whereas disinfectants are antimicrobial agents applied only to inanimate objects. Preservatives are agents that inhibit the growth of microorganisms capable of causing biologic deterioration of substances/materials. In general, antiseptics are only used on the skin and not for surface disinfection and disinfectants are rarely used for skin antisepsis because they may cause injury to skin and other tissues. Other words with the suffix "-cide" (e.g., virucide, fungicide, bactericide, sporicide, and tuberculocide) can kill the type of microorganism identified by the prefix. For example, a bactericide is an agent that kills bacteria. [14] [15] [16] [17] [18] [19] RATIONAL APPROACH TO DISINFECTION AND STERILIZATION About 45 years ago, Earle H. Spaulding 15 devised a rational approach to disinfection and sterilization of patient care items or equipment. This classification scheme is so clear and logical that it has been retained, refined, and successfully used by infection control professionals and others when planning methods for disinfection or sterilization.* Spaulding believed that the nature of disinfection could be understood more readily if instruments and items for patient care were divided into three categories based on the degree of risk for infection involved in the use of the items. Although the scheme remains valid, some examples of disinfection studies with viruses, mycobacteria, and protozoa challenge the current definitions and expectations of highand low-level disinfection. 22 The three categories Spaulding described were critical, semicritical, and noncritical. Critical items are so called because of the high risk for infection if microorganisms, although small numbers of bacterial spores may be present. Intact mucous membranes, such as those of the lungs or the gastrointestinal tract, generally are resistant to infection by common bacterial spores but susceptible to other organisms such as bacteria, mycobacteria, and viruses. Semicritical items minimally require highlevel disinfection using chemical disinfectants. Glutaraldehyde, hydrogen peroxide, ortho-phthalaldehyde, peracetic acid, and peracetic acid with hydrogen peroxide are cleared by the U.S. Food and Drug Administration (FDA) and are dependable high-level disinfectants provided the factors influencing germicidal procedures are met (see Tables 301-1 and . When a disinfectant is selected for use with certain patient care items, the chemical compatibility after extended use with the items to be disinfected also must be considered. The complete elimination of all microorganisms in or on an instrument, with the exception of small numbers of bacterial spores, is the traditional definition of high-level disinfection. The FDA's definition of high-level disinfection is a sterilant used for a shorter contact time to achieve at least a 6-log 10 kill of an appropriate Mycobacterium species. Cleaning followed by high-level disinfection should eliminate sufficient pathogens to prevent transmission of infection. 30, 31 Semicritical items should be rinsed with sterile water after highlevel disinfection to prevent their contamination with organisms that may be present in tap water, such as nontuberculous mycobacteria, 8, 32 Legionella, 33, 34 or gram-negative bacilli such as Pseudomonas. 18, 20, [35] [36] [37] In circumstances where rinsing with sterile water rinse is not feasible, a tap water or filtered water (0.2-µm filter) rinse should be followed by an alcohol rinse and forced air drying. 9, [37] [38] [39] Forced-air drying markedly reduces bacterial contamination of stored endoscopes, most likely by removing the wet environment favorable for bacterial growth. 38 After rinsing, items should be dried and stored (e.g., packaged) in a manner that protects them from recontamination. bacterial spores. Thus, it is critical that objects that enter sterile tissue or the vascular system be sterile because any microbial contamination could result in disease transmission. This category includes surgical instruments, cardiac and urinary catheters, implants, arthroscopes, laparoscopes, and ultrasound probes used in sterile body cavities . Most of the items in this category should be purchased in sterile form or be sterilized by steam sterilization if possible. If heat sensitive, the object may be treated with ETO, hydrogen peroxide gas plasma, hydrogen peroxide vapor, or liquid chemical sterilants if other methods are unsuitable. Tables 301-1 and 301-2 list several germicides categorized as chemical sterilants and high-level disinfectants. These include 2.4% or greater glutaraldehyde-based formulations, hypochlorous acid/ hypochlorite 650 to 675 ppm free chlorine, 1.12% glutaraldehyde with 1.93% phenol/phenate, 3.4% glutaraldehyde with 26% isopropanol, 23 7.5% stabilized hydrogen peroxide, 2.0% hydrogen peroxide, 7.35% hydrogen peroxide with 0.23% peracetic acid, 8.3% hydrogen peroxide with 7.0% peracetic acid, 0.2% peracetic acid, 0.55% or greater orthophthalaldehyde, and 0.08% peracetic acid with 1.0% hydrogen peroxide. 24 Liquid chemical sterilants can be relied on to produce sterility only if cleaning (to eliminate organic and inorganic material) precedes treatment and if proper use as to concentration, contact time, temperature, and pH is met. 25 Semicritical items are those that come in contact with mucous membranes or nonintact skin. Respiratory therapy and anesthesia equipment, some endoscopes, laryngoscope blades and handles, 26 esophageal manometry probes, endocavitary probes, 26 nasopharyngoscopes, prostate biopsy probes, 27 infrared coagulation device, 28 anorectal manometry catheters, cystoscopes, 29 and diaphragm fitting rings are included in this category. 26 These medical devices should be free of all B. Ethylene oxide gas (see manufacturer's recommendations, generally 2 to 6 hours processing time plus aeration time of 8 to 12 hours at 50° to 60° C). C. Hydrogen peroxide gas plasma (see manufacturer's recommendations for internal diameter and length restrictions, processing time between 24 to 47 minutes) and vaporized hydrogen peroxide (see manufacturer's recommendations for internal diameter and length restrictions). D. Glutaraldehyde-based formulations: ≥2% glutaraldehyde (caution should be exercised with all glutaraldehyde formulations when further in-use dilution is anticipated); glutaraldehyde (1.12%) with 1.93% phenol/phenate; and glutaraldehyde (3.4%) with isopropanol (26%). One glutaraldehyde-based product has a high-level disinfection claim of 5 minutes at 35° C. E. Ortho-phthalaldehyde (OPA) 0.55%. F. Hydrogen peroxide, standard 7.5% (will corrode copper, zinc, and brass). G. Peracetic acid, concentration variable but ≥0.2% is sporicidal. A 0.2% peracetic acid immersion reprocessor operates at 50° to 56° C. Per guidance from the FDA, most hospitals use the 0.2% peracetic acid reprocessor for reprocessing semicritical items that require high-level disinfection. Thus, as a general rule, the reprocessor will not be used to reprocess critical items because critical items should be sterile and with the reprocessor using 0.2% peracetic acid the final processed device cannot be assured to be sterile. Thus, heat-sensitive critical devices should be sterilized by other validated, FDA-cleared, sterilization processes such as hydrogen peroxide gas plasma, ethylene oxide, and vaporized hydrogen peroxide. If a heat-sensitive critical device truly cannot be processed by any other modality than the reprocessor using 0.2% peracetic acid, then the decision is between not using the device at all or reprocessing it in the 0.2% peracetic acid reprocessor (at 50° to 56° C). The decision to use the 0.2% peracetic acid reprocessor at 50° to 56° C for a heat-sensitive critical item that cannot be processed by an alternative sterilization process should be made on a case-by-case basis. H. Hydrogen peroxide (7.35%) with 0.23% peracetic acid; hydrogen peroxide 1% with peracetic acid 0.08%; 8.3% hydrogen peroxide with 7.0% peracetic acid (will corrode metal instruments). I. Wet pasteurization at 70°C for 30 minutes with detergent cleaning. J. Hypochlorite, single-use chlorine generated on site by electrolyzing saline containing >400 to 675 active free chlorine (will corrode metal instruments). K. Improved hydrogen peroxide ≥2%. L. Sodium hypochlorite (5.25% to 6.15% household bleach diluted 1 : 500 provides >100 ppm available chlorine). M. Phenolic germicidal detergent solution (follow product label for use-dilution). N. Iodophor germicidal detergent solution (follow product label for use-dilution). O. Quaternary ammonium germicidal detergent solution (follow product label for use-dilution). P. Ethyl and isopropyl alcohol 60% to 95%. Q. Improved hydrogen peroxide 0.5% and 1.4%. 1 See text for discussion of hydrotherapy. 2 The longer the exposure to a disinfectant, the more likely it is that all microorganisms will be eliminated. Twenty-minute exposure at 20° C is the minimum time needed to reliably kill Mycobacterium tuberculosis and nontuberculous mycobacteria with 2% glutaraldehyde. With the exception of >2% glutaraldehyde (see text), follow the FDA-cleared high-level disinfection claim. Some high-level disinfectants have a reduced exposure time (e.g., OPA at 12 minutes at 20° C) because of their rapid activity against mycobacteria or reduced exposure time due to increased mycobactericidal activity at elevated temperature (e.g., 2.5% glutaraldehyde at 5 minutes at 35° C, 0.55% OPA at 5 minutes at 25° C in automated endoscope reprocessor). 3 Tubing must be completely filled for high-level disinfection and liquid chemical sterilization; care must be taken to avoid entrapment of air bubbles during immersion. 4 Material compatibility should be investigated when appropriate. 5 A concentration of 1000 ppm available chlorine should be considered where cultures or concentrated preparations of microorganisms have spilled (5.25% to 6.15% household bleach diluted 1 : 50 provides >1000 ppm available chlorine). This solution may corrode some surfaces. 6 Pasteurization (washer-disinfector) of respiratory therapy or anesthesia equipment is a recognized alternative to high-level disinfection. Some data challenge the efficacy of some pasteurization units. 7 Thermostability should be investigated when appropriate. 8 Do not mix rectal and oral thermometers at any stage of handling or processing. 9 By law, all applicable label instructions on EPA-registered products must be followed. If the user selects exposure conditions that differ from those on the EPA-registered products label, the user assumes liability from any injuries resulting from off-label use and is potentially subject to enforcement action under the Federal Insecticide, Fungicide, and Rodenticide Act. EPA, U.S. Environmental Protection Agency; FDA, U.S. Food and Drug Administration; MR, manufacturer's recommendations; NA, not applicable. Note: The selection and use of disinfectants in the health care field is dynamic, and products may become available that are not in existence when this chapter was written. As newer disinfectants become available, persons or committees responsible for selecting disinfectants and sterilization processes should be guided by products cleared by the FDA and the EPA as well as by information in the scientific literature and manufacturer recommendations. Modified from the works of Rutala and Simmons and their colleagues. 9 36 when they are used as noncritical items and do not contact nonintact skin or mucous membranes. However, these items (e.g., bedside tables, bed rails) could potentially contribute to secondary transmission by contaminating hands of health care workers or by contact with medical equipment that will subsequently come in contact with patients. 14, [42] [43] [44] [45] [42] [43] [44] [45] [46] [47] [48] [49] [50] [51] [52] [53] [54] [55] [56] [57] [58] Thus, it is acceptable to disinfect noncritical medical equipment (e.g., blood pressure cuff) and noncritical surfaces (e.g., bedside table) with an EPA-registered disinfectant or disinfectant/detergent at the proper use-dilution and a contact time of at least 1 minute. 9,59 Because the typical drying time for a germicide on a surface is 1 to 3 minutes (unless the product contains Some items that may come in contact with nonintact skin for a brief period of time (i.e., hydrotherapy tanks, bed side rails) are usually considered noncritical surfaces and are disinfected with low-or intermediate-level disinfectants (i.e., phenolic, iodophor, alcohol, chlorine). 40 Because hydrotherapy tanks have been associated with spread of infection, some facilities have chosen to disinfect them with recommended levels of chlorine. 40 Noncritical items are those that come in contact with intact skin but not mucous membranes. Intact skin acts as an effective barrier to most microorganisms; therefore, the sterility of items that come in contact with intact skin is "not critical. " Examples of noncritical items are bedpans, blood pressure cuffs, crutches, bed rails, bedside tables, patient furniture, and floors. The five most commonly touched noncritical items in the patient environment have been quantitatively shown to be bed rails, bed surface, supply cart, overbed table, and intravenous-line pump. 41 In contrast to critical and some semicritical items, most noncritical reusable items may be decontaminated where they are used and do not need to be transported to a central processing area. There is virtually no documented risk of transmitting infectious Note: All products effective in presence of organic soil, relatively easy to use, and have a broad spectrum of antimicrobial activity (bacteria, fungi, viruses, spores, and mycobacteria). The above characteristics are documented in the literature; contact the manufacturer of the instrument and sterilant for additional information. Modified from references 13, 93, 278, and 304. underrated: ethyl alcohol and isopropyl alcohol. 70 These alcohols are rapidly bactericidal rather than bacteriostatic against vegetative forms of bacteria; they also are tuberculocidal, fungicidal, and virucidal but do not destroy bacterial spores. Their cidal activity drops sharply when diluted below 50% concentration, and the optimal bactericidal concentration is in the range of 60% to 90% solutions in water (volume/ volume). 71, 72 Alcohols are not recommended for sterilizing medical and surgical materials, principally because of their lack of sporicidal action and their inability to penetrate protein-rich materials. Fatal postoperative wound infections with Clostridium have occurred when alcohols were used to sterilize surgical instruments contaminated with bacterial spores. 73 Alcohols have been used effectively to disinfect oral and rectal thermometers, computers, 60 hospital pagers, scissors, cardiopulmonary resuscitation (CPR) manikins, applanation tonometers, 74 external surfaces of equipment (e.g., ventilators), and stethoscopes. 75 Alcohol towelettes have been used for years to disinfect small surfaces such as rubber stoppers of multiple-dose medication vials or vaccine bottles. Alcohols are flammable and consequently must be stored in a cool, well-ventilated area. They also evaporate rapidly, and this makes extended exposure time difficult to achieve unless the items are immersed. Hypochlorites are the most widely used of the chlorine disinfectants and are available in liquid (e.g., sodium hypochlorite) or solid (e.g., calcium hypochlorite) forms. The most prevalent chlorine products in the United States are aqueous solutions of 5.25% to 6.15% sodium hypochlorite, which usually are called household bleach. They have a broad spectrum of antimicrobial activity (i.e., bactericidal, virucidal, fungicidal, mycobactericidal, sporicidal), do not leave toxic residues, are unaffected by water hardness, are inexpensive and fast acting, 74, 76 remove dried or fixed organisms and biofilms from surfaces, 77 and have a low incidence of serious toxicity. 78, 79 Sodium hypochlorite at the concentration used in domestic bleach (5.25% to 6.15%) may produce ocular irritation or oropharyngeal, esophageal, and gastric burns. 69, 80, 81 Other disadvantages of hypochlorites include corrosiveness to metals in high concentrations (>500 ppm), inactivation by organic matter, discoloring or "bleaching" of fabrics, release of toxic chlorine gas when mixed with ammonia or acid (e.g., household cleaning agents), 82 and relative stability. 83 Reports have examined the microbicidal activity of a new disinfectant, "superoxidized water. " The concept of electrolyzing saline to create a disinfectant or antiseptic is appealing because the basic materials of saline and electricity are inexpensive and the end product (i.e., water) is not damaging to the environment. The main products of this "water" are hypochlorous acid (e.g., at a concentration of about 144 mg/L) and chlorine. This is also known as electrolyzed water; and, as with any germicide, the antimicrobial activity of superoxidized water is strongly affected by the concentration of the active ingredient (available free chlorine). 84 The free available chlorine concentrations of different superoxidized solutions reported in the literature range from 7 to 180 ppm. 84 Data have shown that freshly generated superoxidized water is rapidly effective (<2 minutes) in achieving a 5-log 10 reduction of pathogenic microorganisms (i.e., M. tuberculosis, Mycobacterium chelonae, poliovirus, human immunodeficiency virus (HIV), MRSA, E. coli, Candida albicans, Enterococcus faecalis, Pseudomonas aeruginosa) in the absence of organic loading. However, the biocidal activity of this disinfectant was substantially reduced in the presence of organic material (5% horse serum). 85, 86 Hypochlorites are widely used in health care facilities in a variety of settings. 76 Inorganic chlorine solution is used to disinfect tonometer heads 87 and for disinfection of noncritical surfaces and equipment. A 1 : 10 to 1 : 100 dilution of 5.25% to 6.15% sodium hypochlorite (i.e., household bleach) [88] [89] [90] [91] or an EPA-registered tuberculocidal disinfectant 18 has been recommended for decontaminating blood spills. For small spills of blood (i.e., drops of blood) on noncritical surfaces, the area can be disinfected with a 1 : 100 dilution of 5.25% to 6.15% sodium hypochlorite or an EPA-registered tuberculocidal disinfectant. Because hypochlorites and other germicides are substantially inactivated in the alcohol [e.g., a 60% to 70% alcohol will dry in about 30 seconds]) (N. Omidbakhsh, written communication), one application of the germicide on all hand contact surfaces to be disinfected is recommended. Mops (microfiber and cotton string), reusable cleaning cloths, and disposable wipes are regularly used to achieve low-level disinfection. 60, 61 Microfiber mops have demonstrated superior microbial removal compared with cotton string mops when used with detergent cleaner (95% vs. 68%, respectively). Use of a disinfectant did significantly improve microbial removal when a cotton string mop was used. 61 Mops (especially cotton-string mops) are commonly not kept adequately cleaned and disinfected, and if the water-disinfectant mixture is not changed regularly (e.g., after every 3 to 4 rooms, no longer than 60-minute intervals), the mopping procedure may actually spread heavy microbial contamination throughout the health care facility. 62 In one study, standard laundering provided acceptable decontamination of heavily contaminated mop heads but chemical disinfection with a phenolic was less effective. 62 The frequent laundering of cotton-string mops (e.g., daily) is, therefore, recommended. Hospital cleanliness continues to attract patient attention and in the United States it is still primarily assessed via visual appearance, which is not a reliable indicator of surface cleanliness. 63 Three other methods have been offered for monitoring patient room hygiene and they include adenosine triphosphate (ATP) bioluminescence, 64,65 fluorescent markers, 66, 67 and microbiologic sampling. 65 Studies have demonstrated suboptimal cleaning by aerobic colony counts as well as the use of the ATP bioluminescence and fluorescent markers. 64,66 ATP bioluminescence and fluorescent markers are preferred to aerobic plate counts because they provide an immediate assessment of cleaning effectiveness. A great number of disinfectants are used alone or in combinations (e.g., hydrogen peroxide and peracetic acid) in the health care setting. These include alcohols, chlorine and chlorine compounds, formaldehyde, glutaraldehyde, ortho-phthalaldehyde, standard and improved hydrogen peroxide, iodophors, peracetic acid, phenolics, and quaternary ammonium compounds. With some exceptions (e.g., ethanol or bleach), commercial formulations based on these chemicals are considered unique products and must be registered with the EPA or cleared by the FDA. In most instances, a given product is designed for a specific purpose and is to be used in a certain manner. Therefore, the label should be read carefully to ensure that the right product is selected for the intended use and applied in an efficient manner. Additionally, caution must be exercised to avoid hazards with the use of cleaners and disinfectants on electronic medical equipment. Problems associated with the inappropriate use of liquids on electronic medical equipment have included equipment fires, equipment malfunctions, and health care worker burns. 68 Disinfectants are not interchangeable and an overview of the performance characteristics of each is provided in the next section so the user has sufficient information to select an appropriate disinfectant for any item and use it in the most efficient way. It should be recognized that excessive costs may be attributed to incorrect concentrations and inappropriate disinfectants. Finally, occupational diseases among cleaning personnel have been associated with the use of several disinfectants, such as formaldehyde, glutaraldehyde, and chlorine, and precautions (e.g., gloves, proper ventilation) should be used to minimize exposure. 69 Asthma and reactive airway disease may occur in sensitized individuals exposed to any airborne chemical, including germicides. Clinically important asthma may occur at levels below ceiling levels regulated by the Occupational and Safety Health Administration (OSHA) or recommended by the National Institute for Occupational Safety and Health. The preferred method of control is to eliminate the chemical (via engineering controls, or substitution) or relocate the worker. In the health care setting, "alcohol" refers to two water-soluble chemical compounds, the germicidal characteristics of which are generally Bacillus, 105, 106 and this includes the hypervirulent binary toxin strains of C. difficile spores (W.A. Rutala, unpublished data, December 2012). There have been reports of microorganisms with relative resistance to glutaraldehyde, including some mycobacteria (M. chelonae, Mycobacterium avium-intracellulare, Mycobacterium xenopi), [107] [108] [109] Methylobacterium mesophilicum, 110 Trichosporon, fungal ascospores (e.g., Microascus cinereus, Chaetomium globosum), and Cryptosporidium. 111 M. chelonae persisted in a 0.2% glutaraldehyde solution used to store porcine prosthetic heart valves, 112 and a large outbreak of Mycobacterium massiliense infections in Brazil after videolaparoscopy equipment used for different elective cosmetic procedures (e.g., liposuction) was highly tolerant to 2% glutaraldehyde. 113 Porins may have a role in the resistance of mycobacteria to glutaraldehyde and ortho-phthalaldehyde. 114 Dilution of glutaraldehyde during use commonly occurs, and studies show a glutaraldehyde concentration decline after a few days of use in an automatic endoscope washer. 115 This decline occurs because instruments are not thoroughly dried and water is carried in with the instrument, which increases the solution's volume and dilutes its effective concentration. This emphasizes the need to ensure that semicritical equipment is disinfected with an acceptable concentration of glutaraldehyde. Data suggest that 1.0% to 1.5% glutaraldehyde is the minimal effective concentration (MEC) for 2% or greater glutaraldehyde solutions when used as a high-level disinfectant. [115] [116] [117] Chemical test strips or liquid chemical monitors are available for determining whether an effective concentration of glutaraldehyde is present despite repeated use and dilution. The frequency of testing should be based on how frequently the solutions are used (e.g., used daily, test daily; used weekly, test before use), but the strips should not be used to extend the use life beyond the expiration date. Data suggest the chemicals in the test strip deteriorate with time, 118 and a manufacturer's expiration date should be placed on the bottles. The bottle of test strips should be dated when opened and used for the period of time indicated on the bottle (e.g., 120 days). The results of test strip monitoring should be documented. The glutaraldehyde test kits have been preliminarily evaluated for accuracy and range, 118 but their reliability has been questioned. 119 The concentration should be considered unacceptable or unsafe when the test indicates a dilution below the product's MEC (generally to 1.0% to 1.5% glutaraldehyde or lower) by the indicator not changing color. Glutaraldehyde is used most commonly as a high-level disinfectant for medical equipment such as endoscopes, 93 endocavitary probes, spirometry tubing, dialyzers, transducers, anesthesia and respiratory therapy equipment, hemodialysis proportioning and dialysate delivery systems, and reuse of laparoscopic disposable plastic trocars. 75 Glutaraldehyde is noncorrosive to metal and does not damage lensed instruments, rubber, or plastics. The FDA-cleared labels for high-level disinfection with 2% or greater glutaraldehyde at 25° C range from 20 to 90 minutes depending on the product. However, multiple scientific studies and professional organizations support the efficacy of 2% or greater glutaraldehyde for 20 minutes at 20° C. 9, 18, 37 Minimally, this latter recommendation should be followed. Glutaraldehyde should not be used for cleaning noncritical surfaces because it is too toxic and expensive. Colitis believed to be due to glutaraldehyde exposure from residual disinfecting solution in the endoscope solution channels has been reported and is preventable by careful endoscope rinsing. 69 One study found that residual glutaraldehyde levels were higher and more variable after manual disinfection (<0.2 to 159.5 mg/L) than after automatic disinfection (0.2 to 6.3 mg/L). 120 Similarly, keratopathy and corneal damage were caused by ophthalmic instruments that were inadequately rinsed after soaking in 2% glutaraldehyde. 121 Glutaraldehyde exposure should be monitored to ensure a safe work environment. In the absence of an OSHA permissible exposure limit, if the glutaraldehyde level is higher than the American Conference of Industrial Hygienists ceiling limit of 0.05 ppm, it would be prudent to take corrective action and repeat monitoring. 122 The literature contains several accounts of the properties, germicidal effectiveness, and potential uses for stabilized hydrogen peroxide in the health care setting. Published reports ascribe good germicidal activity presence of blood, 54,92 large spills of blood require that the surface be cleaned before an EPA-registered disinfectant or a 1 : 10 (final concentration) solution of household bleach is applied. If there is a possibility of a sharps injury, there should be an initial decontamination, 69, 93 followed by cleaning and terminal disinfection (1 : 10 final concentration). 54 Extreme care should always be used to prevent percutaneous injury. At least 500 ppm available chlorine for 10 minutes is recommended for decontamination of CPR training manikins. Other uses in health care include as an irrigating agent in endodontic treatment and to disinfect laundry, dental appliances, hydrotherapy tanks, 40 regulated medical waste before disposal, 76 applanation tonometers, 74 and the water distribution system in hemodialysis centers and hemodialysis machines. 9,75 Disinfection with a 1 : 10 dilution of concentrated sodium hypochlorite (i.e., bleach) has been shown to be effective in reducing environmental contamination in patient rooms and in reducing C. difficile infection rates in hospital units where there is a high endemic C. difficile infection rates or in an outbreak setting. 9, [94] [95] [96] 97, 98 At our institution, we use a sporicidal solution (5000 ppm chlorine) in all C. difficile-infected patient rooms for routine daily and terminal cleaning. This is done by one application of the sporicide covering all hand contact surfaces to allow sufficient wetness for a greater than 1-minute contact time. Chlorine has long been favored as the preferred disinfectant in water treatment. Hyperchlorination of a Legionella-contaminated hospital water system 40 resulted in a dramatic decrease (30% to 1.5%) in the isolation of Legionella pneumophila from water outlets and a cessation of health care-associated legionnaires' disease in the affected unit. 99, 100 Chloramine T and hypochlorites have been used in disinfecting hydrotherapy equipment. 75 Hypochlorite solutions in tap water at a pH greater than 8 stored at room temperature (23° C) in closed, opaque plastic containers may lose up to 40% to 50% of their free available chlorine level over a period of 1 month. Thus, if a user wished to have a solution containing 500 ppm of available chlorine at day 30, a solution containing 1000 ppm of chlorine should be prepared at time 0. There is no decomposition of sodium hypochlorite solution after 30 days when stored in a closed brown bottle. 83 Glutaraldehyde is a saturated dialdehyde that has gained wide acceptance as a high-level disinfectant and chemical sterilant. 101 Aqueous solutions of glutaraldehyde are acidic and generally in this state are not sporicidal. Only when the solution is "activated" (made alkaline) by use of alkalinizing agents to pH 7.5 to 8.5 does the solution become sporicidal. Once "activated" these solutions have a shelf life of minimally 14 days because of the polymerization of the glutaraldehyde molecules at alkaline pH levels. This polymerization blocks the active sites (aldehyde groups) of the glutaraldehyde molecules that are responsible for its biocidal activity. Novel glutaraldehyde formulations (e.g., glutaraldehyde-phenolsodium phenate, potentiated acid glutaraldehyde, stabilized alkaline glutaraldehyde) produced in the past 40 years have overcome the problem of rapid loss of activity (e.g., now use life of 28 to 30 days) while generally maintaining excellent microbicidal activity. 74, 75, 102, 103 However, it should be recognized that antimicrobial activity is dependent not only on age but also on use conditions such as dilution and organic stress. The use of glutaraldehyde-based solutions in health care facilities is common because of their advantages, which include excellent biocidal properties; activity in the presence of organic matter (20% bovine serum); and noncorrosive action to endoscopic equipment, thermometers, rubber, or plastic equipment. The advantages, disadvantages, and characteristics of glutaraldehyde are listed in Table 301 -2. The in vitro inactivation of microorganisms by glutaraldehydes has been extensively investigated and reviewed. 104 Several investigators showed that 2% or greater aqueous solutions of glutaraldehyde, buffered to pH 7.5 to 8.5 with sodium bicarbonate, were effective in killing vegetative bacteria in less than 2 minutes; M. tuberculosis, fungi, and viruses in less than 10 minutes; and spores of Bacillus and Clostridium species in 3 hours. 104 Spores of C. difficile are more rapidly killed by 2% glutaraldehyde than are spores of other species of Clostridium and is a combination of iodine and a solubilizing agent or carrier; the resulting complex provides a sustained-release reservoir of iodine and releases small amounts of free iodine in aqueous solution. The best known and most widely used iodophor is povidone-iodine, a compound of polyvinylpyrrolidone with iodine. This product and other iodophors retain the germicidal efficacy of iodine but, unlike iodine, are generally nonstaining and are relatively free of toxicity and irritancy. 137 There are several reports that documented intrinsic microbial contamination of antiseptic formulations of povidone-iodine and poloxamer-iodine. [138] [139] [140] It was found that "free" iodine (I 2 ) contributes to the bactericidal activity of iodophors and dilutions of iodophors demonstrate more rapid bactericidal action than does a full-strength povidone-iodine solution. Therefore, iodophors must be diluted according to the manufacturers' directions to achieve antimicrobial activity. Published reports on the in vitro antimicrobial efficacy of iodophors demonstrate that iodophors are bactericidal, mycobactericidal, and virucidal but may require prolonged contact times to kill certain fungi and bacterial spores. 15, [141] [142] [143] [144] Besides their use as an antiseptic, iodophors have been used for the disinfection of blood culture bottles and medical equipment such as hydrotherapy tanks and thermometers. Antiseptic iodophors are not suitable for use as hard-surface disinfectants because of concentration differences. Iodophors formulated as antiseptics contain less free iodine than those formulated as disinfectants. 145 Iodine or iodinebased antiseptics should not be used on silicone catheters because the silicone tubing may be adversely affected. 146 Ortho-phthalaldehyde (OPA) is a high-level disinfectant that received FDA clearance in October 1999. It contains at least 0.55% 1,2benzenedicarboxaldehyde or OPA, and it has supplanted glutaraldehyde as the most commonly used "aldehyde" for high-level disinfection in the United States. OPA solution is a clear, pale-blue liquid with a pH of 7.5. The advantages, disadvantages, and characteristics of OPA are listed in Table 301-2. Studies have demonstrated excellent microbicidal activity in in vitro studies, 74, 75, 93, 111, [147] [148] [149] [150] [151] [152] including superior mycobactericidal activity (5-log 10 reduction in 5 minutes) compared with glutaraldehyde. Walsh and colleagues also found OPA effective (>5-log 10 reduction) against a wide range of microorganisms, including glutaraldehyde-resistant mycobacteria and Bacillus atrophaeus spores. 150 OPA has several potential advantages compared with glutaraldehyde. It has excellent stability over a wide pH range (pH 3 to 9), is not a known irritant to the eyes and nasal passages, does not require exposure monitoring, has a barely perceptible odor, and requires no activation. OPA, like glutaraldehyde, has excellent material compatibility. A potential disadvantage of OPA is that it stains proteins gray (including unprotected skin) and thus must be handled with caution. 93 However, skin staining would indicate improper handling that requires additional training and/or personal protective equipment (gloves, eye and mouth protection, fluid-resistant gowns). OPA residues remaining on inadequately water-rinsed transesophageal echocardiographic probes may leave stains on the patient's mouth. Meticulous cleaning, use of the correct OPA exposure time (e.g., 12 minutes), and copious rinsing of the probe with water should eliminate this problem. Because OPA has been associated with several episodes of anaphylaxis after cystoscopy, 153 the manufacturer has modified its instructions for use of OPA and contraindicates the use of OPA as a disinfectant for reprocessing all urologic instrumentation for patients with a history of bladder cancer. Personal protective equipment should be worn when handling contaminated instruments, equipment, and chemicals. 148 In addition, equipment must be thoroughly rinsed to prevent discoloration of a patient's skin or mucous membrane. The MEC of OPA is 0.3%, and that concentration is monitored by test strips designed specifically for the OPA solution. OPA exposure level monitoring found that the concentration during the disinfection process was significantly higher in the manual group (median, 1.43 ppb) than in the automatic group (median, 0.35 ppb). These findings corroborate other findings that show it is desirable to introduce automatic endoscope reprocessors to to hydrogen peroxide and attest to its bactericidal, virucidal, sporicidal, and fungicidal properties. [123] [124] [125] [126] [127] Some other studies have shown limited bactericidal and virucidal activity of standard 3% hydrogen peroxide. 58, 74 The advantages, disadvantages, and characteristics of hydrogen peroxide are listed in Table 301 -2. As with other chemical sterilants, dilution of the hydrogen peroxide must be monitored by regularly testing the MEC (i.e., 7.5 to 6.0%). Compatibility testing by Olympus America of the 7.5% hydrogen peroxide found both cosmetic changes (e.g., discoloration of black anodized metal finishes) 93 and functional changes with the tested endoscopes (Olympus, October 15, 1999, written communication) . Commercially available 3% hydrogen peroxide is a stable and effective disinfectant when used on inanimate surfaces. It has been used in concentrations from 3% to 6% for the disinfection of soft contact lenses (e.g., 3% for 2 to 3 hours), 123, 128 tonometer biprisms, ventilators, fabrics, 129 and endoscopes. 130 Hydrogen peroxide was effective in spotdisinfecting fabrics in patients' rooms. 129 Corneal damage from a hydrogen peroxide-soaked tonometer tip that was not properly rinsed has been reported. 131 An improved hydrogen peroxide-based technology has been introduced into health care for disinfection of noncritical environmental surfaces and patient equipment 132 and high-level disinfection of semicritical equipment such as endoscopes. [133] [134] [135] Improved hydrogen peroxide contains very low levels of anionic or nonionic surfactants or both in an acidic product that act with hydrogen peroxide to produce microbicidal activity. This combination of ingredients speeds the antimicrobial activity of hydrogen peroxide and cleaning efficiency. 134, 135 Improved hydrogen peroxide is considered safe for humans and equipment and benign for the environment. In fact, improved hydrogen peroxide has the lowest EPA toxicity category (i.e., category IV) based on its oral, inhalation, and dermal toxicity, which means it is practically nontoxic and not an irritant. 132, 134, 136 It is prepared and marketed by several companies in various concentrations (e.g., 0.5% to 7%), and different products may use different terminology for these products, such as "accelerated" or "activated. " Lower concentrations (i.e., 0.5%,1.4%) are designed for the low-level disinfection of noncritical environmental surfaces and patient care objects, whereas the higher concentrations can be used as high-level disinfectants for semicritical medical devices (e.g., endoscopes). A recent study compared the bactericidal activity of a quaternary ammonium compound with two new improved hydrogen peroxide products. The improved hydrogen peroxide products were superior or similar to the quaternary ammonium compound tested. When the two improved hydrogen peroxide products were compared with standard 0.5%, 1.4%, and 3% hydrogen peroxide formulations, the improved hydrogen peroxide-based environmental surface disinfectants proved to be more effective (>6-log 10 reduction) and fast-acting (30-60 seconds) microbicides in the presence of a soil load (to simulate the presence of body fluids) than commercially available hydrogen peroxide. Only 30-to 60-second contact time was studied because longer contact times (e.g., 10 minutes) are not achievable in clinical practice. Additionally, the improved hydrogen peroxide products have an EPAregistered contact time that is substantially less (e.g., 30 seconds, 1 minute for bacteria) than most EPA-registered low-level disinfectants. 58 We have also recently shown that the 1.4% activated hydrogen peroxide is very effective in reducing microbial contamination of hospital privacy curtains. In our study, the activated hydrogen peroxide completely eliminated contamination with MRSA and VRE and resulted in a 98.5% reduction in microbes (only Bacillus spp. recoverable). Thus, at our institution, privacy curtains are being disinfected at the grab area by spraying the grab area of the curtain three times with activated hydrogen peroxide at discharge cleaning. Iodine solutions or tinctures have long been used by health care professionals, primarily as antiseptics on skin or tissue. The FDA has not cleared any liquid chemical sterilant/high level disinfectants with iodophors as the main active ingredient. However, iodophors have been used both as antiseptics and disinfectants. An iodophor functional group (e.g., alkyl, phenyl, benzyl, halogen) replaces one of the hydrogen atoms on the aromatic ring. Two phenol derivatives commonly found as constituents of hospital disinfectants are orthophenylphenol and ortho-benzyl-para-chlorophenol. Published reports on the antimicrobial efficacy of commonly used phenolics showed that they were bactericidal, fungicidal, virucidal, and tuberculocidal. 15, 53, 75, 141, [174] [175] [176] [177] [178] Many phenolic germicides are EPA registered as disinfectants for use on environmental surfaces (e.g., bedside tables, bedrails, laboratory surfaces) and noncritical medical devices. Phenolics are not FDA cleared as high-level disinfectants for use with semicritical items but could be used to preclean or decontaminate critical and semicritical devices before terminal sterilization or high-level disinfection. The use of phenolics in nurseries has been questioned because of the occurrence of hyperbilirubinemia in infants placed in bassinets in which phenolic detergents were used. 179 In addition, Doan and co-workers demonstrated bilirubin level increases in phenolic-exposed infants compared with nonphenolic-exposed infants when the phenolic was prepared according to the manufacturers' recommended dilution. 180 If phenolics (or other disinfectants) are used to clean nursery floors, they must be diluted according to the recommendation on the product label. Phenolics (and other disinfectants) should not be used to clean infant bassinets and incubators while occupied. If phenolics are used to terminally clean infant bassinets and incubators, the surfaces should be rinsed thoroughly with water and dried before the infant bassinets and incubators are reused. 18 The quaternary ammonium compounds are widely used as surface disinfectants. There have been some reports of health care-associated infections associated with contaminated quaternary ammonium compounds used to disinfect patient care supplies or equipment such as cystoscopes or cardiac catheters. 181, 182 As with several other disinfectants (e.g., phenolics, iodophors), gram-negative bacteria have been found to survive or grow in them. 140 Results from manufacturers' data sheets and from published scientific literature indicate that the quaternaries sold as hospital disinfectants are generally fungicidal, bactericidal, and virucidal against lipophilic (enveloped) viruses; they are not sporicidal and generally not tuberculocidal or virucidal against hydrophilic (nonenveloped) viruses. † Poor mycobactericidal activities of quaternary ammonium compounds have been reported. 49, 141 The quaternaries are commonly used in ordinary environmental sanitation of noncritical surfaces such as floors, furniture, and walls. EPA-registered quaternary ammonium compounds are appropriate to use when disinfecting medical equipment that comes into contact with intact skin (e.g., blood pressure cuffs). Pasteurization is not a sterilization process; its purpose is to destroy all pathogenic microorganisms with the exception of bacterial spores. The time-temperature relation for hot-water pasteurization is generally greater than 70° C (158° F) for 30 minutes. The water temperature and time should be monitored as part of a quality assurance program. 186 Pasteurization of respiratory therapy 187, 188 and anesthesia equipment 189 is a recognized alternative to chemical disinfection. Ultraviolet (UV) light has been recognized as an effective method for killing microorganisms. It has been suggested for use in health care for several purposes, including air disinfection, room decontamination (see "Room Decontamination, " later), surface disinfection, biofilm disinfection, 190 and ultrasound probe disinfection. 191 Contaminated ultrasound probes can potentially transmit pathogens. When the probe is only in contact with the patient's skin there is a low risk for infection and low-level disinfection is recommended; however, a higher level of disinfection is recommended when the probe contacts mucous membranes or nonintact skin. An evaluation of a new disinfection procedure for ultrasound probes using UV light demonstrated the median decrease disinfectant exposure levels among scope reprocessing technicians. 154 Peracetic, or peroxyacetic acid, is characterized by a very rapid action against all microorganisms. A special advantage of peracetic acid is its lack of harmful decomposition products (i.e., acetic acid, water, oxygen, hydrogen peroxide); it enhances removal of organic material 155 and leaves no residue. It remains effective in the presence of organic matter and is sporicidal even at low temperatures. Peracetic acid can corrode copper, brass, bronze, plain steel, and galvanized iron, but these effects can be reduced by additives and pH modifications. The advantages, disadvantages, and characteristics of peracetic acid are listed in Table 301-2. Peracetic acid will inactivate gram-positive and gram-negative bacteria, fungi, and yeasts in less than 5 minutes at less than 100 ppm. In the presence of organic matter, 200 to 500 ppm is required. For viruses the dosage range is wide (12 to 2250 ppm), with poliovirus inactivated in yeast extract in 15 minutes with 1500 to 2250 ppm. A processing system using peracetic acid at a temperature of 50° C to 56° C can be used for processing heat-sensitive semicritical and critical devices that are compatible with the peracetic acid and processing system and cannot be sterilized by other legally marketed traditional sterilization methods validated for that type of device (e.g., steam, hydrogen peroxide gas plasma, vaporized hydrogen peroxide). After processing, the devices should be used immediately or stored in a manner similar to that of a high-level disinfected endoscope. [156] [157] [158] The sterilant, 35% peracetic acid, is diluted to 0.2% with tap water that has been filtered and exposed to ultraviolet light. Simulated-use trials with the earlier version of this processing system have demonstrated excellent microbicidal activity, 74, [158] [159] [160] [161] 162 and three clinical trials have demonstrated both excellent microbial killing and no clinical failures leading to infection. [163] [164] [165] Three clusters of infection using the earlier version of the peracetic acid automated endoscope reprocessor were linked to inadequately processed bronchoscopes when inappropriate channel connectors were used with the system. 166, 167 These clusters highlight the importance of training, proper model-specific endoscope connector systems, and quality control procedures to ensure compliance with endoscope manufacturer's recommendations and professional organization guidelines. An alternative high-level disinfectant available in the United Kingdom contains 0.35% peracetic acid. Although this product is rapidly effective against a broad range of microorganisms, 168, 169 it tarnishes the metal of endoscopes and is unstable, resulting in only a 24-hour use life. 169 Three chemical sterilants are FDA-cleared that contain peracetic acid plus hydrogen peroxide (0.08% peracetic acid plus 1.0% hydrogen peroxide, 0.23% peracetic acid plus 7.35% hydrogen peroxide, and 8.3% hydrogen peroxide plus 7.0% peracetic acid). The advantages, disadvantages, and characteristics of peracetic acid with hydrogen peroxide are listed in Table 301 The bactericidal properties of peracetic acid plus hydrogen peroxide have been demonstrated. 170 Manufacturer's data demonstrated that this combination of peracetic acid plus hydrogen peroxide inactivated all microorganisms with the exception of bacterial spores within 20 minutes. The 0.08% peracetic acid plus 1.0% hydrogen peroxide product was effective in inactivating a glutaraldehyde-resistant mycobacteria. 171 The combination of peracetic acid and hydrogen peroxide has been used for disinfecting hemodialyzers. 172 The percentage of dialysis centers using a peracetic acid with hydrogen peroxide-based disinfectant for reprocessing dialyzers increased from 5% in 1983 to 72% in 1997. 173 Phenol has occupied a prominent place in the field of hospital disinfection since its initial use as a germicide by Lister in his pioneering work on antiseptic surgery. In the past 40 years, however, work has been concentrated on the numerous phenol derivatives or phenolics and their antimicrobial properties. Phenol derivatives originate when a microbial reduction for UV light was 100%, 87.5% for antiseptic wiping, and 88% for dry wiping. 191 Surface disinfection with UV light (100-280 nm) has been evaluated with three hospital-related surfaces, namely, aluminum (bed railings), stainless steel (operating tables), and scrubs (laboratory coats). Acinetobacter baumannii were inoculated on small coupons (10 3 or 10 5 /coupon) and exposed to 90 J/m 2 . This exposure was effective in the inactivation of Acinetobacter from the metal coupon surfaces but ineffective in the decontamination of scrubs. 192 A hand-held room decontamination technology that utilizes far-UV radiation (185 to 230 nm) to kill pathogens was evaluated and found that it rapidly kills C. difficile spores and other health care-associated pathogens on surfaces. However, the presence of organic matter reduces the efficacy of far-UV radiation, possibly explaining the more modest results observed on surfaces in hospital rooms that were not precleaned. 193 Most medical and surgical devices used in health care facilities are made of materials that are heat stable and thus are sterilized by heat, primarily steam sterilization. However, since 1950, there has been an increase in medical devices and instruments made of materials (e.g., plastics) that require low-temperature sterilization. ETO has been used since the 1950s for heat-and moisture-sensitive medical devices. Within the past 15 years, a number of new, low-temperature sterilization systems (e.g., hydrogen peroxide gas plasma, vaporized hydrogen peroxide) have been developed and are being used to sterilize medical devices. This section reviews sterilization technologies used in health care and makes recommendations for their optimum performance in the processing of medical devices. 9, 194 Sterilization destroys all microorganisms on the surface of an object or in a fluid to prevent disease transmission associated with the use of that item. Although the use of inadequately sterilized critical items represents a high risk for transmitting pathogens, documented transmission of pathogens associated with an inadequately sterilized critical item is exceedingly rare. [195] [196] [197] This is likely due to the wide margin of safety associated with the sterilization processes used in health care facilities. The concept of what constitutes "sterile" is measured as a probability of sterility for each item to be sterilized. This probability is commonly referred to as the sterility assurance level (SAL) of the product and is defined as the probability of a single viable microorganism occurring on a product after sterilization. SAL is normally expressed as 10 −n . For example, if the probability of a spore surviving were one in 1 million, the SAL would be 10 −6 . 198, 199 Dual SALs (e.g., 10 −3 SAL for blood culture tubes, drainage bags; 10 −6 SAL for scalpels, implants) have been used in the United States for many years, and the choice of a 10 −6 SAL was strictly arbitrary and not associated with any adverse outcomes (e.g., patient infections). 198 Medical devices that have contact with sterile body tissues or fluids are considered critical items. These items should be sterile when used because any microbial contamination could result in disease transmission. Such items include surgical instruments, biopsy forceps, and implanted medical devices. If these items are heat resistant, the recommended sterilization process is steam sterilization, because it has the largest margin of safety due to its reliability, consistency, lethality, and least effect from organic/inorganic soils. However, reprocessing heatand moisture-sensitive items requires use of a low-temperature sterilization technology (e.g., ETO, hydrogen peroxide gas plasma, vaporized hydrogen peroxide). 200 A summary of the advantages and disadvantages for commonly used sterilization technologies is presented in Table 301 -3. Of all the methods available for sterilization, moist heat in the form of saturated steam under pressure is the most widely used and the most dependable. Steam sterilization is nontoxic, inexpensive, 201 rapidly microbicidal, and sporicidal and rapidly heats and penetrates fabrics (see Table 301 -3). 202 Like all sterilization processes, steam sterilization has some deleterious effects on some materials, including corrosion is defined as the shortest possible time between a sterilized item's removal from the sterilizer and its aseptic transfer to the sterile field. This implies that the sterilized item is used during the procedure for which it was sterilized and in a manner that minimizes its exposure to air and other environmental contaminants. The same critical reprocessing steps (e.g., cleaning, decontamination, rinsing, and aseptic transfer from the sterilizer to the point of use) must be followed. Immediate-use steam sterilization should not be used for convenience, as an alternative to purchasing sufficient instrument sets, or as a time saver. 209, 210 Ethylene Oxide "Gas" Sterilization ETO is a colorless gas that is flammable and explosive. The four essential parameters (operational ranges) are gas concentration (450 to 1200 mg/L); temperature (37° C to 63° C); relative humidity (40% to 80%; water molecules carry ETO to reactive sites); and exposure time (1 to 6 hours). These parameters influence the effectiveness of ETO sterilization. [211] [212] [213] [214] Within certain limitations, an increase in gas concentration and temperature may shorten the time necessary for achieving sterilization. The main disadvantages associated with ETO are the lengthy cycle time and its potential hazards to patients and staff; the main advantages are that it is highly penetrating and can sterilize occluded locations in medical items and can sterilize heat-or moisture-sensitive medical equipment without deleterious effects on the material used in the medical devices (see Table 301 -3). 212 Acute exposure to ETO may result in irritation (e.g., to skin, eyes, or gastrointestinal or respiratory tracts) and central nervous system depression. 69 Chronic inhalation has been linked to the formation of cataracts, cognitive impairment, neurologic dysfunction, and disabling polyneuropathies. 69 Occupational exposure in health care facilities has been linked to hematologic changes and an increased risk for spontaneous abortions and various cancers. 69 ETO should be considered a known human carcinogen. 215 The use of ETO evolved when few alternatives existed for sterilizing heat-and moisture-sensitive medical devices; however, favorable properties (see Table 301 -3) account for its continued widespread use. 216 Two ETO gas mixtures are available to replace ETO-chlorofluorocarbon (CFC) mixtures for large capacity, tank-supplied sterilizers. The ETOcarbon dioxide (CO 2 ) mixture consists of 8.5% ETO and 91.5% CO 2 . This mixture has limited use in U.S. health care facilities but is sometimes used in hospitals in India and China. It is less expensive than ETO-hydrochlorofluorocarbons (HCFC), but a disadvantage is the need for pressure vessels rated for steam sterilization, because higher pressures (28-psi gauge) are required. The other mixture, which is a drop-in CFC replacement, is ETO mixed with HCFC. HCFCs are approximately 50-fold less damaging to the earth's ozone layer than are CFCs. The EPA will begin regulation of HCFC in the year 2015 and will terminate production in the year 2030. The ETO-HCFC mixtures have been provided by companies as a drop-in replacement for CFC-12 (one mixture consists of 8.6% ETO and 91.4% HCFC, and the other mixture is composed of 10% ETO and 90% HCFC) but will be phased out by the end of 2013. 216 An alternative to the pressurized mixed-gas ETO systems is 100% ETO. Partly because of the events just described, the 100% ETO sterilizers that use unit-dose cartridges will become the systems for ETO use in U.S. health care facilities. The excellent microbicidal activity of ETO has been demonstrated in several studies 25, 161, 162, [217] [218] [219] and summarized in published reports. 220 ETO inactivates all microorganisms, although bacterial spores (especially B. atrophaeus) are more resistant than other microorganisms. For this reason, B. atrophaeus is the recommended biologic indicator organism. Like all sterilization processes, the effectiveness of ETO sterilization can be altered by lumen length, lumen diameter, inorganic salts, and organic materials. ‡ For example, although ETO is not used commonly for reprocessing endoscopes, 39 several studies have shown failure of ETO in inactivating contaminating spores in endoscope channels 221 or lumen test units. 25, 161, 218 Residual ETO levels averaging 66.2 ppm have been found even after the standard degassing time. 130 Failure of ETO also has been observed when dental handpieces were contaminated and combustion of lubricants associated with dental handpieces, 203 reduction in ability to transmit light associated with laryngoscopes, 204 and increased hardening time (fivefold to sixfold) with plaster cast. 205 The basic principle of steam sterilization, as accomplished in an autoclave, is to expose each item to direct steam contact at the required temperature and pressure for the specified time. Thus, there are four parameters of steam sterilization: steam, pressure, temperature, and time. The ideal steam for sterilization is dry saturated steam and entrained water (dryness fraction ≥97%). 194 Pressure serves as a means to obtain the high temperatures necessary to quickly kill microorganisms. Specific temperatures must be obtained to ensure the microbicidal activity. The two common steam sterilizing temperatures are 121° C (250° F) and 132° C (270° F). These temperatures (and other high temperatures) must be maintained for a minimal time to kill microorganisms. Recognized minimum exposure periods for sterilization of wrapped health care supplies are 30 minutes at 121° C in a gravity displacement sterilizer or 4 minutes at 132° C in a prevacuum sterilizer. At constant temperatures, sterilization times vary depending on the type of item (e.g., metal versus rubber, plastic, items with lumens), whether the item is wrapped or unwrapped, and the sterilizer type. The two basic types of steam sterilizers (autoclaves) are the gravity displacement autoclave and the high-speed prevacuum sterilizer. In the former, steam is admitted at the top or the sides of the sterilizing chamber and, because the steam is lighter than air, forces air out the bottom of the chamber through the drain vent. The gravity displacement autoclaves are primarily used to process laboratory media, water, pharmaceutical products, regulated medical waste, and nonporous articles whose surfaces have direct steam contact. With gravity displacement sterilizers the penetration time into porous items is prolonged because of incomplete air elimination. The high-speed prevacuum sterilizers are similar to the gravity displacement sterilizers except they are fitted with a vacuum pump (or ejector) to ensure air removal from the sterilizing chamber and load before the steam is admitted. The advantage of a vacuum pump is that there is nearly instantaneous steam penetration even into porous loads. Like other sterilization systems, the steam cycle is monitored by physical, chemical, and biological monitors. Steam sterilizers usually are monitored using a printout (or graphically) by measuring temperature, the time at the temperature, and pressure. Typically, chemical indicators are affixed to the outside and incorporated into the pack to monitor the temperature or time and temperature. The effectiveness of steam sterilization is monitored with a biological indicator containing spores of Geobacillus stearothermophilus (formerly Bacillus stearothermophilus). Positive spore test results are a relatively rare event and can be attributed to operator error, inadequate steam delivery, 206 or equipment malfunction. Portable steam sterilizers are used in outpatient, dental, and rural clinics. These sterilizers are designed for small instruments, such as hypodermic syringes and needles and dental instruments. The ability of the sterilizer to reach physical parameters necessary to achieve sterilization should be monitored by physical, chemical, and biologic indicators. Steam sterilization should be used whenever possible on all critical and semicritical items that are heat and moisture resistant (e.g., steam sterilizable respiratory therapy and anesthesia equipment), even when not essential to prevent pathogen transmission. Steam sterilizers also are used in health care facilities to decontaminate microbiologic waste and sharps containers, 207 but additional exposure time is required in the gravity displacement sterilizer for these items. "Flash" steam sterilization was originally defined by Underwood and Perkins as sterilization of an unwrapped object at 132° C for 3 minutes at 27 to 28 pounds of pressure in a gravity displacement sterilizer. 208 It was intended for instruments (e.g., dropped instruments) when there is insufficient time to sterilize an item by the preferred package method. The term "flash" arose out of the abbreviated time of exposure of the unwrapped instrument. Flash sterilization is an antiquated term that does not fully describe the various steam sterilization cycles now used to process items not intended to be stored for later use. Immediate use guidelines on reprocessing, [236] [237] [238] [239] and outbreaks of infection continue to occur. 240, 241 To ensure that reprocessing personnel are properly trained, there should be initial and annual competency testing for each individual who is involved in reprocessing endoscopic instruments. 9, 38, 167, 234 In general, endoscope disinfection or sterilization with a liquid chemical sterilant or high-level disinfectant involves five steps after leak testing: 1. Clean-mechanically clean internal and external surfaces, including brushing internal channels and flushing each internal channel with water and a enzymatic cleaner. 2. Disinfect-immerse endoscope in high-level disinfectant (or chemical sterilant) and perfuse (eliminates air pockets and ensures contact of the germicide with the internal channels) disinfectant into all accessible channels such as the suction/biopsy channel and air/ water channel and expose for a time recommended for specific products. 3. Rinse-rinse the endoscope and all channels with sterile water, filtered water (commonly used with automated endoscope reprocessors), or tap water. 4. Dry-rinse the insertion tube and inner channels with alcohol and dry with forced air after disinfection and before storage. 5. Store-store the endoscope in a way that prevents recontamination and promotes drying (e.g., hung vertically). Unfortunately, there is poor compliance with the recommendations for reprocessing endoscopes, which may result in patient exposure to bloodborne pathogens. 242 In addition, there are rare instances in which the scientific literature and recommendations from professional organizations regarding the use of disinfectants and sterilants may differ from the manufacturer's label claim. One example is the contact time used to achieve high-level disinfection with 2% glutaraldehyde. Based on FDA requirements (FDA regulates liquid sterilants and high-level disinfectants used on critical and semicritical medical devices), manufacturers test the efficacy of their germicide formulations under worstcase conditions (i.e., minimum recommended concentration of the active ingredient) and in the presence of organic soil (typically 5% serum). The soil is used to represent the organic loading to which the device is exposed during actual use and that would remain on the device in the absence of cleaning. These stringent test conditions are designed to provide a margin of safety by ensuring that the contact conditions for the germicide provide complete elimination of the test bacteria (e.g., 10 5 to 10 6 M. tuberculosis in organic soil and dried on a scope) if inoculated into the most difficult areas for the disinfectant to penetrate and in the absence of cleaning. However, the scientific data demonstrate that M. tuberculosis levels can be reduced by at least 8 log 10 with cleaning (4 log 10 ) followed by chemical disinfection for 20 minutes at 20° C (4 to 6 log 10 ). 9, 37, 243 Because of these data, professional organizations (at least 14 professional organizations worldwide) that have endorsed an endoscope reprocessing guideline recommend contact conditions of 20 minutes at 20° C (or <20 minutes outside the United States) with 2% glutaraldehyde to achieve high-level disinfection that differs from that of the manufacturer's label. 37, [244] [245] [246] It is important to emphasize that the FDA tests do not include cleaning, a critical component of the disinfection process. Therefore, when cleaning has been included in the test methodology, 2% glutaraldehyde for 20 minutes has been demonstrated to be effective in eliminating all vegetative bacteria. In December 1991 OSHA promulgated a standard entitled "Occupational Exposure to Bloodborne Pathogens" to eliminate or minimize occupational exposure to bloodborne pathogens. 247 One component of this requirement is that all equipment and environmental and working surfaces be cleaned and decontaminated with an appropriate disinfectant after contact with blood or other potentially infectious materials. Although the OSHA standard does not specify the type of disinfectant or procedure, the OSHA original compliance document 248 suggested that a germicide must be tuberculocidal to kill hepatitis B virus (e.g., phenolic, chlorine). However, in February 1997, OSHA amended its policy and stated that EPA-registered disinfectants that are labeled as with Streptococcus mutans and exposed to ETO. 222 It is recommended that dental handpieces be steam sterilized. ETO is used in health care facilities to sterilize critical items (and sometimes semicritical items) that are moisture or heat sensitive and cannot be sterilized by steam sterilization. New sterilization technology based on hydrogen peroxide and plasma was patented in 1987 and marketed in the United States in 1993. Gas plasmas have been referred to as the fourth state of matter (i.e., liquids, solids, gases, and gas plasmas). Gas plasmas are generated in an enclosed chamber under deep vacuum using radiofrequency or microwave energy to excite the gas (i.e., hydrogen peroxide) molecules and produce charged particles, many of which are in the form of free radicals (e.g., hydroxyl and hydroperoxyl). This system works by diffusing hydrogen peroxide into the chamber and then "exciting" the hydrogen peroxide into a plasma state. The combined use of hydrogen peroxide vapor and plasma safely and rapidly sterilizes instruments without leaving toxic residues. The biologic indicator used with this system is Geobacillus stearothermophilus spores. This process has the ability to inactivate a broad range of microorganisms, including resistant bacterial spores. Studies have been conducted against vegetative bacteria (including mycobacteria), yeasts, fungi, viruses, and bacterial spores. 25, 161, 219, [223] [224] [225] [226] [227] [228] [229] Like all sterilization processes, the effectiveness can be altered by lumen length, lumen diameter, inorganic salts, and organic materials. § Materials and devices that cannot tolerate high temperatures and humidity, such as some plastics, electrical devices, and corrosionsusceptible metal alloys, can be sterilized by hydrogen peroxide gas plasma. This method has been compatible with most (>95%) medical devices and materials tested. 230, 231 Vaporized Hydrogen Peroxide A new low temperature sterilization system uses vaporized hydrogen peroxide to sterilize reusable metal and nonmetal devices used in health care facilities. The system is compatible with a wide range of medical instruments and materials (e.g., polypropylene, brass, polyethylene). There are no toxic by-products because only water vapor and oxygen are produced. The system is not intended to process liquids, linens, powders, or any cellulose materials. The system can sterilize instruments with diffusion-restricted spaces (e.g., scissors) and medical devices with a single stainless steel lumen based on lumen internal diameter and length (e.g., an inside diameter of 1 mm or larger and a length of 125 mm or shorter; see manufacturer's recommendations). Thus, gastrointestinal endoscopes and bronchoscopes cannot be sterilized in this system at the current time. Although this system has not been comparatively evaluated with other sterilization processes, vaporized hydrogen peroxide has been shown to be effective in killing spores, viruses, mycobacteria, fungi, and bacteria. Table 301 -3 lists the advantages and disadvantages of this and other processes. Physicians use endoscopes to diagnose and treat numerous medical disorders. Although endoscopes represent a valuable diagnostic and therapeutic tool in modern medicine and the incidence of infection associated with use has been reported as very low (about 1 in 1.8 million procedures), 232 more health care-associated outbreaks have been linked to contaminated endoscopes than to any other medical device. [4] [5] [6] 233 To prevent the spread of health care-associated infections, all heat-sensitive endoscopes (e.g., gastrointestinal endoscopes, bronchoscopes, nasopharyngoscopes) must be properly cleaned and at a minimum subjected to high-level disinfection after each use. Highlevel disinfection can be expected to destroy all microorganisms; although when high numbers of bacterial spores are present, a few spores may survive. Recommendations for the cleaning and disinfection of endoscopic equipment have been published and should be strictly followed. 9 The current recommendations consider inactivation data but also use epidemiologic studies of prion transmission, infectivity of human tissues, and the efficacy of removing proteins by cleaning. 257, 259, 260 On the basis of scientific data, only critical (e.g., surgical instruments) and semicritical devices contaminated with high-risk tissue (i.e., brain, spinal cord, and eye tissue) from high-risk patients (e.g., known or suspected infection with CJD or other prion disease) require special prion reprocessing. A moist environment after contamination reduces the attachment of both protein and prion amyloid to the stainless steel surface so moist conditions should be maintained. 261 After the device is clean, it should be sterilized by either autoclaving (i.e., steam sterilization) or using a combination of sodium hydroxide and autoclaving 262 using one of the options below 257 : Option 1-autoclave at 134° C for 18 minutes in a prevacuum sterilizer Option 2-autoclave at 132° C for 1 hour in a gravity displacement sterilizer 9,257,263 Option 3-immerse in 1N sodium hydroxide for 1 hour; remove and rinse in water, then transfer to an open pan and autoclave (121° C gravity displacement or 134°C porous or prevacuum sterilizer) for 1 hour Option 4-immerse in 1N sodium hydroxide for 1 hour and heat in a gravity displacement at 121° C for 30 minutes, then clean and subject to routine sterilization. Some data suggest the temperature should not exceed 134° C because the effectiveness of autoclaving may decline as the temperature is increased (e.g., 136° C, 138° C). 264 Prion-contaminated medical devices that are impossible or difficult to clean should be discarded. To minimize environmental contamination, noncritical environmental surfaces should be covered with plastic-backed paper; and when contaminated with high-risk tissues, the paper should be properly discarded. Noncritical environmental surfaces (e.g., laboratory surfaces) contaminated with high-risk tissues should be cleaned and then spot decontaminated with a 1 : 10 dilution of hypochlorite solutions. 257 There is excellent evidence in the scientific literature that environmental contamination plays an important role in the transmission of several key health care-associated pathogens, including MRSA, VRE, Acinetobacter, norovirus, and C. difficile. [265] [266] [267] [268] All these pathogens have been demonstrated to persist in the environment for days (in some cases months), frequently contaminate the environmental surfaces in rooms of colonized or infected patients, transiently colonize the hands of health care personnel, be transmitted by health care personnel, and cause outbreaks in which environmental transmission was deemed to play a role. Importantly, a recent study by Steifel and associates demonstrated that contact with the environment was just as likely to contaminate the hands of health care workers as was direct contact with the patient. 269 Further, admission to a room in which the previous patient was colonized or infected with MRSA, VRE, Acinetobacter, or C. difficile has been shown to be a risk factor for the newly admitted patient to develop colonization or infection. 270, 271 It has long been recommended in the United States that environmental surfaces in patient rooms be cleaned and disinfected on a regular basis (e.g., daily or three times per week), when surfaces are visibly soiled, and after patient discharge (terminal cleaning). 9 Disinfection is generally performed using an EPA-registered hospital disinfectant such as a quaternary ammonium compound. Recent studies have demonstrated that adequate environment cleaning is frequently lacking. For example, Carling and co-workers assessed the thoroughness of terminal cleaning in the patient's immediate environment in 23 acute care hospitals (1119 patient rooms) by using a transparent, easily cleaned, stable solution that fluoresces when exposed to hand-held UV light. 272 The overall thoroughness of cleaning, expressed as a percent of surfaces evaluated, was 49% (range for all hospitals, 35% to 81%). Using a similar design, Carling and co-workers assessed the environmental cleaning in intensive care unit rooms in 16 hospitals (2320 objects) and demonstrated effective against HIV and hepatitis B virus would be considered as appropriate disinfectants "provided such surfaces have not become contaminated with agent(s) or volumes of or concentrations of agent(s) for which higher level disinfection is recommended. " When bloodborne pathogens other than hepatitis B virus or HIV are of concern, OSHA continues to require the use of EPA-registered tuberculocidal disinfectants or hypochlorite solution (diluted 1 : 10 or 1 : 100 with water). 89, 249 Recent studies demonstrate that, in the presence of large blood spills, a 1 : 10 final dilution of EPA-registered hypochlorite solution initially should be used to inactivate bloodborne viruses 54, 250 to minimize risk for disease to the health care worker from percutaneous injury during the clean-up process. Emerging Pathogens, Antibiotic-Resistant Bacteria, and Bioterrorism Agents Emerging pathogens are of growing concern to the general public and infection control professionals. Relevant pathogens include Cryptosporidium parvum, C. difficile, severe acute respiratory syndrome (SARS)coronavirus, Helicobacter pylori, E. coli O157:H7, HIV, hepatitis C virus (HCV), rotavirus, multidrug-resistant M. tuberculosis, human papillomavirus, norovirus, and nontuberculous mycobacteria (e.g., M. chelonae). Similarly, publications have highlighted the concern about the potential for bioterrorism. 251 The Centers for Disease Control and Prevention (CDC) has categorized several agents as "high priority" because they can be easily disseminated or transmitted person to person, can cause high mortality, and are likely to cause public panic and social disruption. 252 These agents include Bacillus anthracis (anthrax), Yersinia pestis (plague), variola major (smallpox), Francisella tularensis (tularemia), filoviruses (Ebola hemorrhagic fever, Marburg hemorrhagic fever); and arenaviruses (Lassa [Lassa fever], Junin [Argentine hemorrhagic fever]), and related viruses. 252 With rare exceptions, the susceptibility of each of these pathogens to chemical disinfectants/sterilants has been studied and all of these pathogens (or surrogate microbes such as feline-calicivirus for norovirus, vaccinia for variola, 142 and B. atrophaeus [formerly Bacillus subtilis] for B. anthracis), are susceptible to currently available chemical disinfectants/sterilants. 9,253,254 Standard sterilization and high-level disinfection procedures for patient care equipment (as recommended in this chapter) are adequate to sterilize or disinfect instruments or devices contaminated with blood or other body fluids from persons infected with bloodborne pathogens, emerging pathogens, and bioterrorism agents, with the exception of prions (see later). No changes in procedures for cleaning, disinfecting, or sterilizing need to be made. 9 In addition, there are no data to show that antibiotic-resistant bacteria (methicillin-resistant Staphylococcus aureus [MRSA], vancomycin-resistant Enterococcus [VRE], multidrug-resistant M. tuberculosis) are less sensitive to the liquid chemical germicides that antibiotic-sensitive bacteria at currently used germicide contact conditions and concentrations. 255, 256 Creutzfeldt-Jakob disease (CJD) is a degenerative neurologic disorder of humans with an incidence in the United States of approximately 1 case/million population/year. 257, 258 CJD is believed to be caused by a proteinaceous infectious agent or prion. CJD is related to other human transmissible spongiform encephalopathies (TSEs) that include kuru (0 incidence, now eradicated), Gertsmann-Straussler-Sheinker (GSS) syndrome (1/40 million), and fatal insomnia syndrome (FFI) (<1/40 million). The agents of CJD and other TSEs exhibit an unusual resistance to conventional chemical and physical decontamination methods. Because the CJD agent is not readily inactivated by conventional disinfection and sterilization procedures and because of the invariably fatal outcome of CJD, the procedures for disinfection and sterilization of the CJD prion have been both conservative and controversial for many years. units (CFU)/cm 2 and of VRE by more than 3-to 4-log 10 CFU/cm 2 . 282 Thus, there are now three studies that have demonstrated that a UV-C system is capable of reducing vegetative bacteria inoculated on a carrier by more than 3-to 4-log 10 in 15 to 20 minutes and C. difficile by more than 1.7-to 4-log 10 in 35 to 100 minutes. The studies also demonstrate reduced effectiveness when surfaces were not in direct line of sight. [280] [281] [282] Several systems that produce hydrogen peroxide (e.g., vapor, aerosolized dry mist) have been studied for their ability to decontaminate environmental surfaces and objects in hospital rooms. Hydrogen peroxide vapor (HPV) has been used increasingly for the decontamination of rooms in health care facilities. [283] [284] [285] [286] [287] [288] [289] [290] [291] [292] Investigators found that hydrogen peroxide systems are a highly effective method for eradicating various pathogens (e.g., MRSA, M. tuberculosis, Serratia, C. difficile spores, Clostridium botulinum spores) from rooms, furniture, and equipment. Importantly, using a before-after study design, Boyce and co-workers have shown that use of hydrogen peroxide vapor was associated with a significant reduction in the incidence of C. difficile infection on five high-incidence wards. 283 The UV-C system studied and the systems that use hydrogen peroxide have their own advantages and disadvantages 268 and there is now ample evidence that these "no-touch" systems can reduce environmental contamination with health care-associated pathogens. However, each specific system should be studied and its efficacy demonstrated before being introduced into health care facilities. The main advantage of both units is their ability to achieve substantial reductions in vegetative bacteria. As noted earlier, manual cleaning has been demonstrated to be suboptimal because many environmental surfaces are not cleaned. Another advantage is their ability to substantially reduce C. difficile because low-level disinfectants (e.g., quaternary ammonium compounds) have limited or no measurable activity against spore-forming bacteria. 278 Both systems are residual free and they decontaminate all exposed surfaces and equipment in the room. The major disadvantages of both decontamination systems are the substantial capital equipment costs, the need to remove personnel and patients from the room, thus limiting their use to terminal room disinfection (must prevent/minimize exposure to UV and hydrogen peroxide), the staff time needed to transport the system to rooms to be decontaminated and monitor its use, the need to physically clean the room of dust and debris, and the sensitivity to use parameters. There are several important differences between the two systems. The UV-C system offers faster decontamination that reduces the "down" time of the room before another patient can be admitted. The hydrogen peroxide systems have been demonstrated to be more effective in eliminating spore-forming organisms. Whether this improved sporicidal activity is clinically important is unclear because studies have demonstrated that although environmental contamination is common in the rooms of patients with C. difficile infection, the level of contamination is relatively low (also true for MRSA and VRE). Finally, the hydrogen peroxide system was demonstrated to reduce C. difficile incidence in a clinical study, whereas similar studies with the UV-C system have not been published. If additional studies continue to demonstrate a benefit, then widespread adoption of these technologies should be considered for terminal room disinfection of certain patient rooms (e.g., contact precautions) in health care facilities. Health care facilities that generate medical, chemical, or radiologic waste have a moral and legal obligation to dispose of these wastes in a manner that poses minimal potential hazard to the environment or public health. The proper disposal of these wastes requires a dynamic waste management plan that conforms to federal, state, and local regulations and provides adequate personnel and financial resources to ensure implementation. that only 57.1% of sites were cleaned after discharge of the room's occupant. 273 A recent study using ATP bioluminescence assays and aerobic cultures demonstrated that medical equipment frequently had not been disinfected as per protocol. 274 Investigators have reported that intervention programs aimed at environmental services workers resulted in significant improvement in cleaning practices. 65, 66 Such interventions have generally included multiple activities: improved education, monitoring the thoroughness of cleaning (e.g., by use of ATP assays or fluorescent dyes) with feedback of performance to the environmental service workers or use of cleaning checklists or both. We have found that assignment of cleaning responsibility (e.g., medical equipment to be cleaned by nursing; environmental surfaces to be cleaned by environmental service) is also important to ensure all objects and surfaces are decontaminated, especially the surfaces of medical equipment (e.g., cardiac monitors). Improved environmental cleaning has been demonstrated to reduce the environmental contamination with VRE, 275, 276 MRSA, 276 and C. difficile. 277 Importantly, no study has reported in the postintervention period proper cleaning of more than 85% of objects. Further, all studies have only focused improvement on a limited number of "high risk" objects. Thus, a concern of published studies is that they have only demonstrated improved cleaning of a limited number of "high risk" objects (or "targeted" objects), not an improvement in the overall thoroughness of room decontamination. As noted earlier, multiple studies have demonstrated that environmental surfaces and objects in rooms are frequently not properly cleaned and these surfaces may be important in transmission of health careassociated pathogens. Further, although interventions aimed at improving cleaning thoroughness have demonstrated effectiveness, many surfaces remain inadequately cleaned and therefore potentially contaminated. For this reason, several manufacturers have developed room disinfection units that can decontaminate environmental surfaces and objects. These systems use one of two methods-either ultraviolet light or hydrogen peroxide. 268 These technologies supplement, but do not replace, standard cleaning and disinfection because surfaces must be physically cleaned of dirt and debris. UV irradiation has been used for the control of pathogenic microorganisms in a variety of applications, such as control of legionellosis, as well as disinfection of air, surfaces, and instruments. 278, 279 At certain wavelengths, UV light will break the molecular bonds in DNA, thereby destroying the organism. UV-C has a characteristic wavelength of 200 to 270 nm (e.g., 254 nm), which lies in the germicidal active portion of the electromagnetic spectrum of 200 to 320 nm. The efficacy of UV irradiation is a function of many different parameters such as intensity, exposure time, lamp placement, and air movement patterns. An automated mobile UV-C unit has been shown to eliminate more than 3-log 10 vegetative bacteria (MRSA, VRE, Acinetobacter baumannii) and more than 2.4-log 10 C. difficile seeded onto Formica surfaces in patients' rooms experimentally contaminated. 280 Boyce and colleagues report the results of assessing the effectiveness of the same UV-C unit to reduce environmental contamination with vegetative bacteria (measured using aerobic colony counts) and C. difficile inoculated onto stainless steel carrier disks. 281 Room decontamination with the UV system resulted in significant reductions in aerobic bacteria on five high-touch surfaces. Mean C. difficile log 10 reductions ranged from 1.8 to 2.9 using cycle times of 34.2 to 100.1 minutes. Surfaces in direct line-of-sight were significantly more likely to yield negative cultures after UV decontamination than before decontamination. Nerandzic and colleagues showed that UV-C at a reflected dose of 22,000 mWs/ cm 2 for approximately 45 minutes consistently reduced recovery of C. difficile spores and MRSA by more than 2-to 3-log 10 colony-forming (e.g., the CDC, EPA, or states) define waste as infectious when it is suspected to contain pathogens in sufficient number to cause disease. Not only does this subjective definition result in conflicting opinions from the CDC, EPA, and state agencies on what constitutes infectious waste and how it should be treated, but it also gives undue emphasis to the mere presence of pathogens. Guidelines produced by the CDC have designated five types of hospital waste as regulated medical waste (i.e., microbiology laboratory waste, pathology and anatomy waste, contaminated animal carcasses, blood, and sharps). 40 The EPA guidelines consider the same types of waste as infectious or regulated medical waste but also designate communicable disease isolation waste. 296 In the MWTA, the EPA modified its position on "communicable disease isolation waste" by including only certain "highly" communicable disease waste such as Class 4 (e.g., Marburg, Ebola, and Lassa viruses) as regulated medical waste 298 (Table 301-4 ). In a systematic random survey of all U.S. hospitals conducted in July 1987 and January 1988, the overall compliance rates with the CDC and EPA recommendations were 82% and 75%, respectively. Not only were the majority of hospitals in compliance, but the hospitals frequently treated other hospital waste as infectious, including contaminated laboratory waste (87%), surgery waste (78%), dialysis waste (69%), items contacting secretions (63%), intensive care (37%), and emergency department waste (41%). 293 A key component in evaluating the impact of a medical waste management program is the quantity of waste produced per patient. Hospitalized patients generate about 15 pounds of hospital waste per day. The amount of hospital waste generated by U.S. hospitals is approximately 6700 tons per day. U.S. hospitals designate approximately 15% of the total hospital waste by weight as infectious (about 1000 tons of infectious waste per day). 293 Not surprisingly, the percentage of medical waste treated as infectious increases with the number and types of medical waste classified as infectious. For example, about 6% of hospital waste would be treated as infectious waste if the CDC guidelines are followed but 45% of hospital waste could be considered infectious waste under the MWTA. 293, 299 The vast majority of U.S. hospitals designate and treat microbiologic, pathologic, isolation, blood, and sharp waste as infectious. 293 In the late 1980s, treatment of infectious waste by U.S. hospitals was most Medical waste disposal has been as a major problem in the United States for the past 40 years. The problem has developed as a result of medical waste washing ashore in some coastal states in 1987 and 1988 and the perceived threat of acquiring HIV infection via this waste. This has led to restrictive rules governing the disposal of medical waste in many states and an increase in the volume of waste defined as regulated medical waste. Coincidentally, with an increase in volume of regulated medical waste (formerly called "infectious waste"), the options for medical waste treatment and disposal are diminishing because of space and environmental concerns. This section will review some of the principles associated with medical waste management, but a more detailed description of collection, storage, processing, transporting, treatment, and public health implications of medical waste may be found elsewhere. [293] [294] [295] [296] [297] Despite the attention given to medical waste by the public, the media, and all levels of government, the terms hospital waste, medical waste, regulated medical waste, and infectious waste are often used synonymously. Hospital waste refers to all waste, biologic or nonbiologic, that is discarded and not intended for further use. Medical waste refers to materials generated as a result of patient diagnosis, immunization, or treatment, such as soiled dressings or intravenous tubing. Infectious waste refers to that portion of medical waste that could potentially transmit an infectious disease. Congress and the EPA used the term regulated medical waste rather than infectious waste in the Medical Waste Tracking Act (MWTA) of 1988 in deference to the remote possibility of disease transmission associated with this waste. Thus, medical waste is a subset of hospital waste, and regulated medical waste (which is synonymous with infectious waste from a regulatory perspective) is a subset of medical waste. 293 As stated, regulated medical waste (or infectious waste) is capable of producing an infectious disease. This definition requires a consideration of the factors necessary for disease induction that include dose, host susceptibility, presence of a pathogen, virulence of a pathogen, and the most commonly absent factor, a portal of entry. For a waste to be infectious, therefore, it must contain pathogens with sufficient virulence and quantity so that exposure to the waste by a susceptible host could result in an infectious disease. Because there are no tests that allow infectious waste to be identified objectively, responsible agencies The CDC guidelines specify "microbiology laboratory waste" as infectious waste. This term includes stocks and cultures of etiologic agents and microbiology laboratory waste contaminated with etiologic agents (e.g., centrifuge tubes, pipettes, tissue culture bottles). † The Act went into effect on June 22, 1989 , and expired June 22, 1991 . It affected only four states (New Jersey, New York, Connecticut, and Rhode Island). The Act required both treatment (any method, technique, or process designed to change the biologic character or composition of medical waste so as to eliminate or reduce its potential for causing disease) and destruction (waste is ruined, torn apart, or mutilated so that it is no longer generally recognizable as medical waste). ‡ MWTA specified used and unused sharps. The Act regulated wastes from persons with highly communicable diseases such as Class 4 etiologic agents (e.g., Marburg, Ebola, Lassa viruses). § Optional infectious waste: EPA states that the decision to handle these wastes as infectious should be made by a responsible, authorized person or committee at the individual facility. CDC, Centers for Disease Control and Prevention 40 ; EPA, U.S. Environmental Protection Agency 296 ; MWTA, Medical Waste Tracking Act. 298 Disposal/Treatment abbreviations: C, chemical disinfection for liquids only; CB, cremation or burial by mortician; I, incineration; IW, infectious waste; S, steam sterilization; Sew, sanitary sewer (EPA requires secondary treatment); SW, steam sterilization with incineration or grinding; TI, thermal inactivation. Note: The Joint Commission requires that there be a hazardous waste system designed and operated in accordance with applicable law and regulations. acknowledge that much of the medical waste that washed ashore in the summer of 1988 was syringe related (65%) and came from home health care and illegal drug use. Fourth, studies have shown that most U.S. hospitals are in compliance with the CDC infectious waste guidelines. Fifth, although the principal purpose of the MWTA was to reduce medical waste on beaches, it has not demonstrated its intended benefit. The relative number of syringes on the beaches in the MWTA states was significantly greater during implementation of the Act (17.23%) than before the Act went into effect (3.2%). 299 If regulatory control were based on epidemiologic, microbiologic, and environmental data, only two types of medical waste would require special handling and treatment-sharps and microbiologic waste. Federal medical waste regulations have been promulgated by the U.S. Department of Transportation and OSHA. The Department of Transportation regulation involves the transport of infectious substances and medical waste and went into effect January 1996. 302 The OSHA Bloodborne Pathogen Standard requires labeling to designate waste that poses a health threat in the workplace. The OSHA definition of regulated waste is not intended to designate waste that must be treated. In fact, generators who apply the OSHA definition of regulated waste (rather than state regulations) to designate infectious waste for treatment by incineration or other means may unintentionally incur additional expenses. 247 When properly used, disinfection and sterilization can ensure the safe use of invasive and noninvasive medical devices. However, current disinfection and sterilization guidelines must be strictly followed. commonly accomplished by incineration (range, 64% to 93%, depending on the type of waste), but emission regulations that limit air pollutants has reduced the number and use of incineration for medical waste. For example, in September 1997 there were an estimated 2373 medical waste incinerators in the United States, but based on the EPA's 2010 inventory there are 54 infectious waste incinerators. 300 Autoclaves or steam sterilizers have become the primary nonincineration technology used by hospitals to process their regulated medical waste (except pathology waste) (E. Krisiunas, written communication, 2008). Several other nonincineration alternatives have been proposed for treating regulated medical waste (e.g., mechanical/chemical disinfection, microwave decontamination, steam disinfection, and compacting). 297 Nonregulated medical waste is generally discarded in a properly sited and operated sanitary landfill because this is a safe and inexpensive disposal method (e.g., landfill disposal costs $0.02 to 0.05 per pound for nonregulated medical waste versus a contract incinerator cost of $0.20 to 0.60 per pound for regulated medical waste). The conflicting opinions of state and federal regulations are related to the paucity of microbiologic and epidemiologic evidence that medical waste represents a threat to the public health. First, with the exception of "sharps" such as needles, which have caused disease only in an occupational setting, there is no scientific evidence that medical waste has caused disease in the hospital or the community. Second, data demonstrate that household waste contains on average 100 times as many microorganisms with pathogenic potential for humans than medical waste. 301 Third, detailed reports of the beach washups found that the vast majority of waste on beaches was debris (about 99%) such as wood, plastic, and paper, not medical waste. EPA documents The complete reference list is available online at Expert Consult. Healthcare Infection Control Practices Advisory Committee. Guideline for Disinfection and Sterilization in Healthcare Facilities Selection and use of disinfectants in healthcare An overview of disinfection and sterilization Disinfection and sterilization in health-care facilities FDA-Cleared Sterilant and High-Level Disinfectants with General Claims for Processing Reusable Medical and Dental Devices Comparison of ion plasma, vaporized hydrogen peroxide and 100% ethylene oxide sterilizers to the 12/88 ethylene oxide gas sterilizer New developments in reprocessing semicritical devices Multisociety guideline on reprocessing flexible GI endoscopes Guidelines for environmental infection control in health-care facilities Antimicrobial activity of home disinfectants and natural products against potential human pathogens Efficacies of selected disinfectants against Mycobacterium tuberculosis The effect of blood on the antiviral activity of sodium hypochlorite, a phenolic, and a quaternary ammonium compound Efficacy of improved hydrogen peroxide against important healthcare-associated pathogens Improving cleaning of the environment surrounding patients in 36 acute care hospitals Efficacy of hospital germicides against adenovirus 8, a common cause of epidemic keratoconjunctivitis in health care facilities. Antimicrob Agents Chemother Uses of inorganic hypochlorite (bleach) in health-care facilities Stability and bactericidal activity of chlorine solutions Activity of selected oxidizing microbicides against spores of Clostridium difficile: relevance to environmental control A targeted strategy to wipe out Clostridium difficile Inactivation of Clostridium difficile spores by disinfectants Bactericidal, virucidal, and mycobactericidal activities of reused alkaline glutaraldehyde in an endoscopy unit Outbreaks associated with contaminated antiseptics and disinfectants Inactivation of Mycobacterium tuberculosis and Mycobacterium bovis by 14 hospital disinfectants The inactivation of viruses by germicides Virucidal activity of ortho-phthalaldehyde solutions against hepatitis B and C viruses Steris System 1E Liquid Chemical Sterilant New low temperature sterilization technologies: microbicidal activity and clinical efficacy Ineffectiveness of hospital disinfectants against bacteria: a collaborative study Sterility assurance: concepts for patient safety Immediate use steam sterilization: moving beyond current policy Association for the Advancement of Medical Instrumentation, Accreditation Association for Ambulatory Health Care, Association of Perioperative Registered Nurses, et al. Immediate-Use Steam Sterilization Association for the Advancement of Medical Instrumentation. Ethylene Oxide Sterilization in Health Care Facilities: Safety and Effectiveness How to assess risk of disease transmission when there is a failure to follow recommended disinfection and sterilization principles FDA labeling requirements for disinfection of endoscopes: a counterpoint Occupational Safety and Health Administration. Occupational exposure to bloodborne pathogens; final rule Susceptibility of antibiotic-susceptible and antibiotic-resistant hospital bacteria to disinfectants Use of germicides in the home and health care setting: is there a relationship between germicide use and antimicrobial resistance? Guideline for disinfection and sterilization of prion-contaminated medical instruments Disinfection and sterilization of prion-contaminated medical instruments Disinfection and sterilization of prion-contaminated medical instruments: reply to Belay World Health Organization. WHO Infection Control Guidelines for Transmissible Spongiform Encephalopathies The role played by contaminated surfaces in the transmission of nosocomial pathogens Environmental contamination makes an important contribution to hospital infection Role of hospital surfaces in the transmission of emerging health careassociated pathogens: norovirus, Clostridium difficile, and Acinetobacter species Are room decontamination units needed to prevent transmission of environmental pathogens? Risk of acquiring antibioticresistant bacteria from prior room occupants Evaluation of hospital room assignment and acquisition of Clostridium difficile infection Healthcare Environmental Hygiene Study Group. Identifying opportunities to enhance environmental cleaning in 23 acute care hospitals Room decontamination by ultraviolet radiation Impact of hydrogen peroxide vapor room decontamination on Clostridium difficile environmental contamination and transmission in a healthcare setting Management of infectious waste by United States hospitals Management of infectious waste from hospitals Infectious waste: mismatch between science and policy Society for Hospital Epidemiology of America. SHEA position paper: Medical waste Outpatient Surgery Passes Inpatient, to 53 Million a Year Managing and preventing exposure events from inappropriately reprocessed endoscopes Audit of bronchoscope disinfection: a survey of procedures in England and Wales and incidents of mycobacterial contamination Transmission of infection by gastrointestinal endoscopy and bronchoscopy Lessons from outbreaks associated with bronchoscopy The prevention of infection following gastrointestinal endoscopy: the importance of prophylaxis and reprocessing An outbreak of Mycobacterium chelonae infection following liposuction Mycobacterium chelonae causing otitis media in an ear-nose-and-throat practice Healthcare Infection Control Practices Advisory Committee. Guideline for Disinfection and Sterilization in Healthcare Facilities Selection and use of disinfectants in healthcare Disinfection, sterilization, and control of hospital waste An overview of disinfection and sterilization Disinfection and sterilization in health-care facilities Chemical disinfection of medical and surgical materials Chemical disinfection of medical and surgical materials CDC guidelines for the prevention and control of nosocomial infections: guideline for hospital environmental control APIC Guidelines Committee. APIC guideline for selection and use of disinfectants. Association for Professionals in Infection Control and Epidemiology Prevention and Control of Nosocomial Infections CDC guideline for handwashing and hospital environmental control APIC guideline for selection and use of disinfectants Disinfection: is it time to reconsider Spaulding? Aldahol high-level disinfectant FDA-Cleared Sterilant and High-Level Disinfectants with General Claims for Processing Reusable Medical and Dental Devices Comparison of ion plasma, vaporized hydrogen peroxide and 100% ethylene oxide sterilizers to the 12/88 ethylene oxide gas sterilizer New developments in reprocessing semicritical devices Disinfection of a probe used in ultrasound-guided prostate biopsy Disinfection of an infrared coagulation device used to treat hemorrhoids Outbreak of cystoscopy related infections with Pseudomonas aeruginosa: New Mexico Efficacy of high-level disinfectants for reprocessing gastrointestinal endoscopes in simulated-use testing High-level disinfection of gastrointestinal endoscopes: are current guidelines adequate? Nosocomial outbreaks/pseudo-outbreaks caused by nontuberculous mycobacteria Waterrelated nosocomial pneumonia caused by Legionella pneumophila serogroups 1 and 10 Legionella: from environmental habitats to disease pathology, detection and control Water as a reservoir of nosocomial pathogens Prevention and Control of Nosocomial Infections Multisociety guideline on reprocessing flexible GI endoscopes Cleaning and disinfection of fiberoptic endoscopes: evaluation of glutaraldehyde exposure time and forced-air drying Disinfection practices for endoscopes and other semicritical items Guidelines for environmental infection control in health-care facilities A quantitative approach to defining high-touch surfaces in hospitals Institutional outbreaks of rotavirus diarrhoea: potential role of fomites and environmental surfaces as vehicles for virus transmission Role of environmental contamination in the transmission of vancomycin-resistant enterococci Prevention of surface-to-human transmission of rotaviruses by treatment with disinfectant spray Chemical disinfection to interrupt transfer of rhinovirus type 14 from environmental surfaces to hands Transmission of experimental rhinovirus infection by contaminated surfaces Interruption of rotavirus spread through chemical disinfection Antimicrobial activity of home disinfectants and natural products against potential human pathogens Efficacies of selected disinfectants against Mycobacterium tuberculosis Efficacy of a variety of disinfectants against Listeria spp Feasibility of a combined carrier test for disinfectants: studies with a mixture of five types of microorganisms Chemical disinfection of human rotaviruses: efficacy of commercially-available products in suspension tests Survival and disinfectant inactivation of the human immunodeficiency virus: a critical review The effect of blood on the antiviral activity of sodium hypochlorite, a phenolic, and a quaternary ammonium compound How stable is the hepatitis C virus (HCV)? Environmental stability of HCV and its susceptibility to chemical biocides The antiviral action of common household disinfectants and antiseptics against murine hepatitis virus, a potential surrogate for SARS coronavirus Inactivation of influenza A virus H1N1 by disinfection process Efficacy of improved hydrogen peroxide against important healthcare-associated pathogens Are disinfectants accurately prepared for use in hospital patient care areas? Bacterial contamination of keyboards: efficacy and functional impact of disinfectants Microbiologic evaluation of microfiber mops for surface disinfection Hospital sanitation: the massive bacterial contamination of the wet mop Monitoring the effectiveness of cleaning in four British hospitals Monitoring the effectiveness of hospital cleaning practices by use of an adenosine triphosphate bioluminescence assay A modified ATP benchmark for evaluating the cleaning of some hospital environmental surfaces Improving cleaning of the environment surrounding patients in 36 acute care hospitals Use of a fluorescent chemical as a quality indicator for a hospital cleaning program Public Health Notification from FDA, CDC, EPA and OSHA: Avoiding Hazards with Using Cleaners and Disinfectants on Electronic Occupational risks associated with the use of selected disinfectants and sterilants Alcohol as a surgical disinfectant The relationship of concentration and germicidal efficiency of ethyl alcohol A note on the fallacy of using alcohol for the sterilization of surgical instruments Efficacy of hospital germicides against adenovirus 8, a common cause of epidemic keratoconjunctivitis in health care facilities. Antimicrob Agents Chemother Selection and use of disinfectants in healthcare Uses of inorganic hypochlorite (bleach) in health-care facilities Safety and cleaning of medical materials and devices Poisoning with sodium hypochlorite solution: report of a fatal case, supplemented with an experimental and clinico-epidemiological study Treatment of acute hypoxemic respiratory failure caused by chlorine exposure Esophageal stenosis produced by ingestion of bleach: report of two cases Response of the human eye to accidental exposure to sodium hypochlorite Home exposures to chlorine/chloramine gas: review of 216 cases Stability and bactericidal activity of chlorine solutions Not all super-oxidized waters are the same Evaluation of the antimicrobial activity of a new super-oxidized water, Sterilox®, for the disinfection of endoscopes Evaluation of microbicidal activity of a new disinfectant: Sterilox® 2500 against Clostridium difficile spores, Helicobacter pylori, vancomycin resistant Enterococcus species, Candida albicans and several Mycobacterium species Tonometer disinfection and viruses Acquired immune deficiency syndrome (AIDS): precautions for clinical and laboratory staffs Recommendations for prevention of HIV transmission in health-care settings Guidelines for prevention of transmission of human immunodeficiency virus and hepatitis B virus to health-care and public-safety workers Guideline for isolation precautions in hospitals A comparison of hypochlorite and phenolic disinfectants for disinfection of clean and soiled surfaces and blood spillages Disinfection of endoscopes: review of new chemical sterilants used for high-level disinfection Measures to control and prevent Clostridium difficile infection Activity of selected oxidizing microbicides against spores of Clostridium difficile: relevance to environmental control A targeted strategy to wipe out Clostridium difficile Significant impact of terminal room cleaning with bleach on reducing nosocomial Clostridium difficile Environmental control to reduce transmission of Clostridium difficile Legionnaires' disease associated with a hospital water system: a cluster of 24 nosocomial cases Legionnaires' disease associated with a hospital water system. A five-year progress report on continuous hyperchlorination GI endoscopic reprocessing practices in the United States Antimicrobial and other properties of a new stabilized alkaline glutaraldehyde disinfectant/sterilizer Comparison of the activities and stabilities of alkaline glutaraldehyde sterilizing solutions Inactivation of Clostridium difficile spores by disinfectants The activity of glutaraldehyde against Clostridium difficile Glutaraldehyde resistant mycobacteria from endoscope washers Glutaraldehyderesistant Mycobacterium chelonae from endoscope washer disinfectors Aldehyde-resistant mycobacteria bacteria associated with the use of endoscope reprocessing systems Microbial contamination of activated 2% glutaraldehyde used in high-level disinfection of endoscopes Inactivation of Cryptosporidium parvum oocyst infectivity by disinfection and sterilization processes Fastidious mycobacteria grown from porcine prosthetic-heart-valve cultures Rapidly growing mycobacteria infection after videosurgical procedures-the glutaraldehyde hypothesis Role of porins in the susceptibility of Mycobacterium smegmatis and Mycobacterium chelonae to aldehyde-based disinfectants and drugs Bactericidal, virucidal, and mycobactericidal activities of reused alkaline glutaraldehyde in an endoscopy unit Effect of methodology, dilution, and exposure time on the tuberculocidal activity of glutaraldehyde-based disinfectants Mycobactericidal activity of glutaraldehyde solutions Glutaraldehyde test kits: evaluation for accuracy and range Monitoring glutaraldehyde dilution in automated washer/disinfectors Residual glutaraldehyde levels in fiberoptic endoscopes: measurement and implications for patient toxicity Glutaraldehyde keratopathy Glutaraldehyde: a potential health risk to nurses Hydrogen peroxide and other oxidant disinfectants Disinfection, Sterilization, and Preservation A product based on accelerated and stabilized hydrogen peroxide: evidence for broad-spectrum germicidal activity Sporicidal activity of chemical sterilants used in hospitals Innate resistance to sporicides and potential failure to decontaminate The effect of currently available contact lens disinfection systems on Acanthamoeba castellanii and Acanthamoeba polyphaga The 1999 Lindberg award. 3% hydrogen peroxide for the gram-positive disinfection of fabrics Significant factors in the disinfection and sterilization of flexible endoscopes Corneal damage from improperly cleaned tonometer tips Broad-spectrum microbicidal activity, toxicologic assessment, and materials compatibility of a new generation of accelerated hydrogen peroxidebased environmental surface disinfectant Combined application of simulated reuse and quantitative carrier test to assess high-level disinfection: experiments with an accelerated hydrogen peroxide-based formulation A new peroxide-based flexible endoscopecompatible high-level disinfectant An accelerated hydrogen peroxide (AHP)-based fast-acting and reusable microbicide for manual disinfection of heat sensitive semi-critical medical devices Products based on accelerated and stabilized hydrogen peroxide: evidence for cleaning and sanitizing efficiency, environmental and human safety and non-corrosiveness Iodine and iodine compounds Pseudobacteremia caused by povidone-iodine solution contaminated with Pseudomonas cepacia Pseudobacteremia attributed to contamination of povidone-iodine with Pseudomonas cepacia Outbreaks associated with contaminated antiseptics and disinfectants Inactivation of Mycobacterium tuberculosis and Mycobacterium bovis by 14 hospital disinfectants The inactivation of viruses by germicides Increased bactericidal activity of dilute preparations of povidone-iodine solutions Wescodyne: lack of activity against poliovirus in the presence of organic matter Chemical disinfection of medical and surgical materials Medcom Frequently Asked Questions. Available at www.medcompnet.com/faq/faq/html Enhancement of mycobactericidal activity of glutaraldehyde with α,βunsaturated and aromatic aldehydes New disinfection and sterilization methods The mycobactericidal efficacy of ortho-phthalaldehyde and the comparative resistances of Mycobacterium bovis, Mycobacterium terrae, and Mycobacterium chelonae Ortho-phthalaldehyde: a possible alternative to glutaraldehyde for high level disinfection In-hospital evaluation of ortho-phthalaldehyde as a high level disinfectant for flexible endoscopes Virucidal activity of ortho-phthalaldehyde solutions against hepatitis B and C viruses Nine episodes of anaphylaxis following cystoscopy caused by Cidex OPA (ortho-phthalaldehyde) highlevel disinfectant in 4 patients after cystoscopy Ortho-phthalaldehyde exposure levels among endoscope disinfection workers Surface analysis of clinically used expanded PTFE endoscopic tubing treated by the STERIS PROCESS Steris System 1E Liquid Chemical Sterilant Medical applications of peracetic acid The use of peracetic acid for the reprocessing of flexible endoscopes and rigid cystoscopes and laparoscopes Evaluation of the Steris System 1 Peracetic Acid Endoscope Processor Evaluation of the microbicidal efficacy of Steris System 1 for digestive endoscopes using GERMANDE and ASTM validation protocols New low temperature sterilization technologies: microbicidal activity and clinical efficacy Comparison of liquid chemical sterilization with peracetic acid and ethylene oxide sterilization for long narrow lumens Liquid sterilization versus high level disinfection in the urologic office Clinical evaluation of a liquid chemical sterilization system for flexible bronchoscopes Liquid chemical sterilization using peracetic acid: an alternative approach to endoscope processing Centers for Disease Control and Prevention. Bronchoscopy-related infections and pseudoinfections Centers for Disease Control and Prevention. FDA and CDC Public Health Advisory: Infections from Endoscopes Inadequately Reprocessed by an Automated Endoscope Reprocessing System Disinfection of bronchoscopes, contaminated in vitro with Mycobacterium tuberculosis, Mycobacterium avium-intracellulare and Mycobacterium chelonae in sputum, using stabilized, buffered peracetic acid solution In-use stability of Nu-Cidex Bactericidal properties of peracetic acid and hydrogen peroxide, alone and in combination, and chlorine and formaldehyde against bacterial water strains Destruction of a glutaraldehyde-resistant mycobacterium by a per-oxygen disinfectant Dialyser reprocessing with Renalin National surveillance of dialysis-associated diseases in the United States Ineffectiveness of hospital disinfectants against bacteria: a collaborative study Disinfection, Sterilization, and Preservation Disinfectant testing using a modified use-dilution method: collaborative study Phenolic compounds Comparative sensitivity of 13 species of pathogenic bacteria to seven chemical germicides Epidemic neonatal hyperbilirubinemia and use of a phenolic disinfectant detergent Phenol and neonatal jaundice Bacteremia following cardiac catheterization Antibioticsensitive Serratia marcescens infections complicating cardiopulmonary operations: contaminated disinfectant as a reservoir Chemical disinfection of hepatitis A virus on environmental surfaces Surface active agents: quaternary ammonium compounds Inactivation of feline calicivirus, a Norwalk virus surrogate An outbreak of respiratory tract infection resulting from incomplete disinfection of ventilatory equipment Disinfection of respirator tubing: a comparison of chemical versus hot water machine-assisted processing Efficacy of a washer-pasteurizer for disinfection of respiratory-care equipment Evaluation of two hot water washer disinfectors for medical instruments Dose requirements for UVC disinfection of catheter biofilms Evaluation of a new disinfection procedure for ultrasound probes using ultraviolet light Disinfection of Acinetobacter baumannii-contaminated surfaces relevant to medical treatment facilities with ultraviolet C light Evaluation of a hand-held far-ultraviolet radiation device for decontamination of Clostridium difficile and other healthcareassociated pathogens Association for the Advancement of Medical Instrumentation, American National Standards Institute. Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities Outbreak of viral hepatitis B in a rural community in India linked to inadequately sterilized needles and syringes An outbreak of surgical wound infections due to Clostridium perfringens Outbreak of Pseudomonas aeruginosa surgical site infections after arthroscopic procedures: Texas Sterility assurance: concepts for patient safety Sterility testing-validation of sterilization processes, and sporicide testing Clinical effectiveness of low-temperature sterilization technologies Costs of low-temperature plasma sterilization compared with other sterilization methods Sterilization by heat Evaluation of sterilization of dental handpieces by heating in synthetic compressor lubricant The effect of steam sterilization at 134 degrees C on light intensity provided by fibrelight Macintoch laryngoscopes The effect of steam sterilization on plaster casting material When the biological indicator is positive: investigating autoclave failures Decontamination of laboratory microbiological waste by steam sterilization Disinfection and flash sterilization in the operating room Immediate use steam sterilization: moving beyond current policy Association for the Advancement of Medical Instrumentation, Accreditation Association for Ambulatory Health Care, Association of Perioperative Registered Nurses, et al. Immediate-Use Steam Sterilization Sterilization with gaseous ethylene oxide: a review of chemical and physical factors Gaseous chemical sterilization Association for the Advancement of Medical Instrumentation. Ethylene Oxide Sterilization in Health Care Facilities: Safety and Effectiveness Association for the Advancement of Medical Instrumentation. Ethylene Oxide Sterilization in Health Care Facilities: Safety and Effectiveness Department of Health and Human Services Ethylene oxide sterilization: how hospitals can adapt to the changes Safety and efficacy of ethylene oxide sterilized polyethylene in total knee arthroplasty Bacterial killing ability of 10% ethylene oxide plus 90% hydrochlorofluorocarbon sterilizing gas Comparative evaluation of the sporicidal activity of new low-temperature sterilization technologies: ethylene oxide, 2 plasma sterilization systems, and liquid peracetic acid Sterilization with ethylene oxide and other gases The efficacy of low temperature plasma (LTP) sterilization, a new sterilization technique Effectiveness of ethylene oxide for sterilization of dental handpieces Efficacy testing of low-temperature plasma sterilization (LTP) with test object models simulating practice conditions Validation of low-temperature-plasma (LPT) sterilization systems: comparison of two technical versions, the Sterrad 100, 1.8 and the 100S Sterilization processes utilizing lowtemperature plasma Sporicidal activity of a new low-temperature sterilization technology: the Sterrad 50 sterilizer Inactivation of human immunodeficiency virus type 1, hepatitis A virus, respiratory syncytial virus, vaccinia virus, herpes simplex virus type 1, and poliovirus type 2 by hydrogen peroxide gas plasma sterilization Assessment of the efficacy of a low temperature hydrogen peroxide gas plasma sterilization system Evaluation of low temperature hydrogen peroxide plasma sterilization Compatibility of medical devices and materials with low-temperature hydrogen peroxide gas plasma The new Sterrad 100S sterilization system: features and advantages Infectious complications associated with gastrointestinal endoscopy Lessons learned from outbreaks and pseudo-outbreaks associated with bronchoscopy Multi-society guideline for reprocessing flexible gastrointestinal endoscopes Society of Gastroenterology Nurses and Associates. SGNA Standards of infection control in reprocessing of flexible gastrointestinal endoscopes Endoscope reprocessing methods: a prospective study on the impact of human factors and automation Correction of deficiencies in flexible fiberoptic sigmoidoscope cleaning and disinfection technique in family practice and internal medicine offices Italian National Survey of Digestive Endoscopy Disinfection Procedures An audit of bronchoscopy practice in the United Kingdom: a survey of adherence to national guidelines An outbreak of Pseudomonas aeruginosa infections associated with flexible bronchoscopes Outbreak of infection associated with bronchoscopes How to assess risk of disease transmission when there is a failure to follow recommended disinfection and sterilization principles FDA labeling requirements for disinfection of endoscopes: a counterpoint Guidelines on cleaning and disinfection in GI endoscopy: update 1999. The European Society of Gastrointestinal Endoscopy Report of a working party of the British Society of Gastroenterology Endoscope Committee British Thoracic Society guidelines on diagnostic flexible bronchoscopy Occupational Safety and Health Administration. Occupational exposure to bloodborne pathogens; final rule OSHA Instruction CPL 2-2.44C. Office of Health Compliance Assistance OSHA Memorandum from Stephen Mallinger. EPA-Registered Disinfectants for HIV/HBV Inactivation of hepatitis B virus in plasma by hospital in-use chemical disinfectants assessed by a modified HepG2 cell culture The looming threat of bioterrorism Biological and chemical terrorism: strategic plan for preparedness and response Infection control: the role of disinfection and sterilization Research to support the decontamination of surfaces and buildings contaminated with biothreat agents Susceptibility of antibiotic-susceptible and antibiotic-resistant hospital bacteria to disinfectants Use of germicides in the home and health care setting: is there a relationship between germicide use and antimicrobial resistance? Guideline for disinfection and sterilization of prion-contaminated medical instruments Surveillance for Creutzfeldt-Jakob disease-United States Disinfection and sterilization of prion-contaminated medical instruments Disinfection and sterilization of prion-contaminated medical instruments: reply to Belay Adsorption of prion and tissue proteins to surgical stainless steel surfaces and the efficacy of decontamination following dry and wet storage conditions World Health Organization. WHO Infection Control Guidelines for Transmissible Spongiform Encephalopathies Creutzfeldt-Jakob disease: recommendations for disinfection and sterilization Inactivation of prions by physical and chemical means The role played by contaminated surfaces in the transmission of nosocomial pathogens Environmental contamination makes an important contribution to hospital infection Role of hospital surfaces in the transmission of emerging health care-associated pathogens: norovirus, Clostridium difficile, and Acinetobacter species Are room decontamination units needed to prevent transmission of environmental pathogens? Contamination of hands with methicillin-resistant Staphylococcus aureus after contact with environmental surfaces and after contact with the skin of colonized patients Risk of acquiring antibioticresistant bacteria from prior room occupants Evaluation of hospital room assignment and acquisition of Clostridium difficile infection Healthcare Environmental Hygiene Study Group. Identifying opportunities to enhance environmental cleaning in 23 acute care hospitals Intensive care unit environmental cleaning: an evaluation in sixteen hospitals using a novel assessment tool Cleanliness of portable medical equipment disinfected by nursing staff Intervention evaluation of environmental contamination by vancomycin-resistant enterococci: failure of personnel, product, or procedure? Impact of an environmental cleaning intervention on the presence of methicillinresistant Staphylococcus aureus and vancomycin-resistant enterococci on surfaces in intensive care unit rooms Reduction of Clostridium difficile and vancomycin-resistant Enterococcus contamination of environmental surfaces after an intervention to improve cleaning methods Sterilization, high-level disinfection, and environmental cleaning Applications of ultraviolet germicidal irradiation disinfection in health care facilities: effective adjunct, but not stand-alone technology Room decontamination by ultraviolet radiation Terminal decontamination of patient rooms using an automated mobile UV light unit Evaluation of an automated ultraviolet radiation device for decontamination of Clostridium difficile and other healthcareassociated pathogens in hospital rooms Impact of hydrogen peroxide vapor room decontamination on Clostridium difficile environmental contamination and transmission in a healthcare setting Tackling contamination of the hospital environment by methicillin-resistant Staphylococcus aureus (MRSA): a comparison between conventional terminal cleaning and hydrogen peroxide vapour decontamination Environmental methicillin-resistant Staphylococcus aureus (MRSA) disinfection using dry-mist-generated hydrogen peroxide Use of hydrogen peroxide vapor for deactivation of Mycobacterium tuberculosis in a biological safety cabinet and a room Rapid recontamination with MRSA of the environment of an intensive care unit after decontamination with hydrogen peroxide vapour Evaluation of hydrogen peroxide vapour as a method for the decontamination of surfaces contaminated with Clostridium botulinum spores Efficacy of vaporized hydrogen peroxide against exotic animal viruses Vapor-phase hydrogen peroxide as a surface decontaminant and sterilant Use of hydrogen peroxide vapour for environmental control during a Serratia outbreak in a neonatal intensive care unit Activity of a dry mist hydrogen peroxide system against environmental Clostridium difficile contamination in elderly care wards Management of infectious waste by United States hospitals Management of infectious waste from hospitals Agency for Toxic Substances and Disease Registry. The Public Health Implications of Medical Waste: A Report to Congress EPA Guide for Infectious Waste Management Office of Technology Assessment (OTA-0-459). Finding the Rx for Managing Medical Wastes Environmental Protection Agency. Standards for the tracking and management of medical waste; interim final rule and request for comments Infectious waste: mismatch between science and policy Environmental Protection Agency. Hospital/Medical/ Infectious Waste Incinerators: Summary of requirements for revised or new section 111(d)/129 State plans following amendments to the emission guidelines Society for Hospital Epidemiology of America. SHEA position paper: Medical waste Department of Transportation. Infectious substances: final rule Selection and use of disinfectants in healthcare Sterilization and disinfection