key: cord-0032546-l5jzukp9 authors: Fontecha, Gustavo; Escobar, Denis; Ortiz, Bryan; Pinto, Alejandra; Serrano, Delmy; Valdivia, Hugo O. title: Field Evaluation of a Hemozoin-Based Malaria Diagnostic Device in Puerto Lempira, Honduras date: 2022-05-11 journal: Diagnostics (Basel) DOI: 10.3390/diagnostics12051206 sha: 9476cdcaff78211bcdb5b277f4ed5a8792c0f07e doc_id: 32546 cord_uid: l5jzukp9 The diagnosis of malaria in Honduras is based mainly on microscopic observation of the parasite in thick smears or the detection of parasite antigens through rapid diagnostic tests when microscopy is not available. The specific treatment of the disease depends exclusively on the positive result of one of these tests. Given the low sensitivity of conventional methods, new diagnostic approaches are needed. This study evaluates the in-field performance of a device (Gazelle™) based on the detection of hemozoin. This was a double-blind study evaluating symptomatic individuals with suspected malaria in the department of Gracias a Dios, Honduras, using blood samples collected from 2021 to 2022. The diagnostic performance of Gazelle™ was compared with microscopy and nested 18ssr PCR as references. The sensitivity and specificity of Gazelle™ were 59.7% and 98.6%, respectively, while microscopy had a sensitivity of 64.9% and a specificity of 100%. The kappa index between microscopy and Gazelle™ was 0.9216 using microscopy as a reference. Both methods show similar effectiveness and predictive values. No statistical differences were observed between the results of the Gazelle™ compared to light microscopy (p = 0.6831). The turnaround time was shorter for Gazelle™ than for microscopy, but the cost per sample was slightly higher for Gazelle™. Gazelle™ showed more false-negative cases when infections were caused by Plasmodium falciparum compared to P. vivax. Conclusions: The sensitivity and specificity of Gazelle™ are comparable to microscopy. The simplicity and ease of use of the Gazelle™, the ability to run on batteries, and the immediacy of its results make it a valuable tool for malaria detection in the field. However, further development is required to differentiate Plasmodium species, especially in those regions requiring differentiated treatment. Malaria in Honduras increased by more than 46% between 2020 and 2021 when 1671 cases were reported in the country, with 98% of them concentrated in the department of Gracias a Dios (Personal communication, National Office of the Pan American Health Organization). This increase in malaria cases represents a setback in efforts to achieve elimination [1] , which had been commendable before the COVID-19 pandemic [2] . Another recent change in the epidemiology of malaria in Honduras is the increase in the proportion of cases caused by P. falciparum versus P. vivax, from 29.3% in 2020 to 43 .5% in 2021 (Personal communication by the National Malaria Surveillance Laboratory, Health Ministry, Honduras). Diagnosis and timely treatment are mainstays in malaria control and elimination, preventing local transmission. The diagnosis of malaria in Honduras is mainly based on the detection of the parasite by light microscopy (LM) carried out by expert technicians or by rapid diagnostic tests (RDTs). LM remains the gold standard in many endemic countries worldwide [3] due to its low cost and because it allows the identification of the species and stages of the parasite, in addition to being useful for the assessment of the severity of malaria through the quantification of parasitaemia. However, even under optimal conditions, the detection limit of LM is low (0.001% parasitemia) [4, 5] . Furthermore, microscopy is a time-consuming method and relies heavily on skilled technicians. The loss of diagnostic capacity based on microscopy is a common problem in countries where the number of malaria cases has decreased [6, 7] . The declining skills of microscopists have negative consequences for diagnosis, such as an underestimation of cases due to low parasite burden among symptomatic and asymptomatic persons [8] , misidentification of parasite species [9] , or inability to detect mixed infections [10] . On the other hand, RDTs are valuable alternatives to microscopy, especially in remote areas lacking qualified personnel or continuous electrical flow. RDTs detect parasite-specific antigens, are inexpensive and easy to use in the field, and allow immediate treatment after diagnosis. However, the sensitivity of common RDTs is usually lower than that of microscopy, with a threshold of 100-200 parasites per µL of blood [11, 12] . Furthermore, RDTs do not indicate parasite burden and may face detection problems when the parasite does not express the target antigen [13, 14] . Due to the intrinsic limitations of conventional malaria diagnostic methods and the challenges faced by countries approaching elimination, a large number of new diagnostic methods have been proposed [15] [16] [17] [18] [19] [20] [21] [22] . The most sensitive methods (around five parasites/µL of blood) are those based on DNA amplification. However, they require specialized infrastructure and highly qualified technicians, are expensive, labor-intensive, and cannot be implemented at the point of care [23] . Therefore, molecular methods are restricted to research or reference laboratories. Thus, there is a real need to develop new diagnostic approaches that are easy to implement at the local level, inexpensive, and with a sensitivity at least comparable to the gold standard, LM (below 50-100 parasites/µL of blood), if we want to achieve the United Nations Sustainable Development Goal number 3 for the elimination of malaria [1] . One of the most studied biomarkers at present is hemozoin or malarial pigment. Hemozoin is an inert crystal by-product of hemoglobin degradation, produced by all species of Plasmodium spp. within the digestive vacuole. The purpose of hemozoin production is to avoid the toxicity that heme groups would have on the parasite through oxidative stress [24, 25] . Hemozoin is a good biomarker because it is generally absent in uninfected individuals, and its blood concentration increases with the progression of malaria [26] . Due to its chemical structure, hemozoin is paramagnetic and optically birefringent [27, 28] . These properties have allowed the development of micro and miniaturized devices, extensively reviewed by Baptista et al. [29] , to detect malaria using simple and portable instruments. This study evaluates the performance of Gazelle™ (Hemex Health, Portland, OR, USA), a late-stage device to diagnose malaria by detecting hemozoin. The study was conducted in a setting of moderate endemicity for malaria in Central America, where two species of Plasmodium circulate. Gazelle™ was compared with LM and nested PCR. This was a double-blind study evaluating symptomatic individuals suspected of malaria comparing three diagnostic assays: thick blood smear, Gazelle™, and nested PCR. The study was carried out in the department of Gracias a Dios in the Moskitia region between 2021 and the first trimester of 2022. La Moskitia is an isolated territory located in eastern Honduras ( Figure 1 ) that contributed to 98% of national malaria cases last year. Two species of human Plasmodia coexist in the region, P. falciparum and P. vivax [30] . The most common vector species in the country and in La Moskitia is Anopheles albimanus [31] . The geographic isolation of the Honduran Moskitia is caused by multiple lagoons and a thick strip of forest that limits land access. Around this system of lagoons of 3700 square kilometers live more than 24,000 people far from each other. The remoteness and the difficulties of access limit the development of the communities with respect to the rest of the country, which is reflected in the low health and education indicators, with a human poverty rate of 53%. The mangrove forest is the dominant coastal ecosystem with more than 23,000 hectares, which privileges fishing as the main means of subsistence and favors the development of malaria vector mosquitoes [32] . Diagnostics 2022, 12, x FOR PEER REVIEW 3 thick strip of forest that limits land access. Around this system of lagoons of 3700 sq kilometers live more than 24,000 people far from each other. The remoteness and the ficulties of access limit the development of the communities with respect to the rest o country, which is reflected in the low health and education indicators, with a human erty rate of 53%. The mangrove forest is the dominant coastal ecosystem with more 23,000 hectares, which privileges fishing as the main means of subsistence and favor development of malaria vector mosquitoes [32]. Blood samples were collected from febrile patients who attended routine me care at the public hospital in the municipality of Puerto Lempira, in the departme Gracias a Dios, located within the region of La Moskitia. Participants resided in one o following municipalities: Puerto Lempira (91.8%), Ramón Villeda Morales (4%), A (2.7%), Brus Laguna (0.9%), and Wampusirpi (0.5%) ( Figure 1 ). Patients of both sexes of all ages were recruited. The patients or their legal guardians were informed of th jectives of the study and signed an informed consent form before collecting the b samples. There were no exclusion criteria for this study. The study was conducted i cordance with the guiding principles of the Declaration of Helsinki and was approve the Institutional Ethics Committee of the National Autonomous University of Hond (CEI-MEIZ/UNAH). The protocol for this study (03-2020/NAMRU6.2018.0002) wa viewed and approved by the Research Administration Program of the Naval Medica search Unit-6 (NAMRU-6). No previous studies have tested the sensitivity of Gazelle TM in a low transmis setting similar to this study. Therefore, the sample size was calculated as describe Buderer [33] , assuming a sensitivity of at least 55%, a specificity of at least 85%, a disease prevalence of 1/3 (33.3%) on collected samples, 12% relative precision, and power. This yielded a minimum required sample size of 199 subjects. Further, 10% o Blood samples were collected from febrile patients who attended routine medical care at the public hospital in the municipality of Puerto Lempira, in the department of Gracias a Dios, located within the region of La Moskitia. Participants resided in one of the following municipalities: Puerto Lempira (91.8%), Ramón Villeda Morales (4%), Ahuas (2.7%), Brus Laguna (0.9%), and Wampusirpi (0.5%) ( Figure 1 ). Patients of both sexes and of all ages were recruited. The patients or their legal guardians were informed of the objectives of the study and signed an informed consent form before collecting the blood samples. There were no exclusion criteria for this study. The study was conducted in accordance with the guiding principles of the Declaration of Helsinki and was approved by the Institutional Ethics Committee of the National Autonomous University of Honduras (CEI-MEIZ/UNAH). The protocol for this study (03-2020/NAMRU6.2018.0002) was reviewed and approved by the Research Administration Program of the Naval Medical Research Unit-6 (NAMRU-6). No previous studies have tested the sensitivity of Gazelle™ in a low transmission setting similar to this study. Therefore, the sample size was calculated as described by Buderer [33] , assuming a sensitivity of at least 55%, a specificity of at least 85%, and a disease prevalence of 1/3 (33.3%) on collected samples, 12% relative precision, and 80% power. This yielded a minimum required sample size of 199 subjects. Further, 10% of the samples were added to cover possible sample losses for any unavoidable reasons leading to a final sample size of 219 participants. After the physician evaluated the feverish patients suspected of malaria, trained laboratory technicians collected the blood samples in EDTA tubes. According to the Honduran national malaria guidelines, thin and thick blood smears were prepared for microscopic examination [34] . The slides were examined within the first 24 h after the sample was collected. A total of 500 microscopic fields were observed with a 100X objective before reporting negative samples. Positive samples were reported estimating the parasite density through a quantitative method, reporting both asexual and sexual stages of the parasite per 200 leukocytes. The density of parasitemia was classified as described by Alger et al. [35] as high, moderate, or low, depending on the parasite species. The Gazelle™ device was handled by a technician in the clinical laboratory of Puerto Lempira hospital. Blood samples were processed at the end of each working day by blinded study microscopists to avoid bias. The Gazelle™ device uses magneto-optical technology to detect hemozoin produced by Plasmodium as a by-product of hemoglobin digestion. By exposing the iron contained in the hemozoin to an intense magnetic field, the opacity of the sample changes, and the increase in opacity is indicative of malaria. The Gazelle™ system consists of a portable reading device and single-use disposable cartridges that operate on either battery power or electricity. The cartridges consist of a chamber, where 65 microliters of the diluent provided by the manufacturer were deposited, followed by 30 microliters of patient blood into the diluent. The upper part of the cartridge was assembled in the lower chamber, and once assembled, it was inserted into the reader, and the analysis was started. The results were recorded after at least one minute to determine the presence or absence of malaria. The Gazelle™ system is unable to differentiate between Plasmodium species responsible for the infection. The Gazelle™ device can be seen in Figure 2 . After the physician evaluated the feverish patients suspected of malaria, t boratory technicians collected the blood samples in EDTA tubes. According to duran national malaria guidelines, thin and thick blood smears were prepared f scopic examination [34] . The slides were examined within the first 24 h after th was collected. A total of 500 microscopic fields were observed with a 100X object reporting negative samples. Positive samples were reported estimating the par sity through a quantitative method, reporting both asexual and sexual stages of site per 200 leukocytes. The density of parasitemia was classified as described b al. [35] as high, moderate, or low, depending on the parasite species. The Gazelle TM device was handled by a technician in the clinical laboratory Lempira hospital. Blood samples were processed at the end of each workin blinded study microscopists to avoid bias. The Gazelle TM device uses magne technology to detect hemozoin produced by Plasmodium as a by-product of he digestion. By exposing the iron contained in the hemozoin to an intense magn the opacity of the sample changes, and the increase in opacity is indicative of ma Gazelle TM system consists of a portable reading device and single-use dispo tridges that operate on either battery power or electricity. The cartridges consist o ber, where 65 microliters of the diluent provided by the manufacturer were d followed by 30 microliters of patient blood into the diluent. The upper part of the was assembled in the lower chamber, and once assembled, it was inserted into th and the analysis was started. The results were recorded after at least one minute mine the presence or absence of malaria. The Gazelle TM system is unable to dif between Plasmodium species responsible for the infection. The Gazelle TM devi seen in Figure 2 . A volume of 50 μL of blood was deposited on Whatman filter paper Nº 3 f DNA extraction. The samples were stored at room temperature in individual se tic bags with desiccant for up to five months before being transported to Tegucig DNA was extracted using the AutoMate TM system and the PrepFiler Express DNA Extraction Kit TM (Applied Biosystems, Waltham, MA, USA) according to t facturer's instructions. A volume of 50 µL of blood was deposited on Whatman filter paper Nº 3 for further DNA extraction. The samples were stored at room temperature in individual sealed plastic bags with desiccant for up to five months before being transported to Tegucigalpa. The DNA was extracted using the AutoMate™ system and the PrepFiler Express Foren- A nested PCR (nPCR) was performed to amplify a segment of the ribosomal 18S gene using the primers and conditions described by Singh et al. [36, 37] with modifications. Briefly, both reactions were performed in 50 µL total volume containing 2X master mix Taq polymerase (Promega Corp. Madison, WI, USA) and 2 µL of each primer 10 µM (Table 1 ). The first reaction included 11 µL of nuclease-free water and 10 µL of DNA. The second reaction included 20 µL of nuclease-free water and 1 µL of the product of the first reaction. Negative samples were considered negative after only one result, while positive samples were amplified a second time to confirm the result. If a discordant result was obtained between two PCR reactions or between a PCR reaction and microscopy/Gazelle™, the PCR was repeated a third time, and the result was determined by two concordant tests. The third PCR reaction was performed with newly extracted DNA from the filter paper to exclude contamination. Samples with two positive results were analyzed to determine the parasite species through two independent reactions that were carried out in a final volume of 25 µL containing 12.5 µL of 2X master mix Taq polymerase, 1 µL of each primer 10 µM, 9.5 µL of nuclease-free water and 1 µL of the product of the first PCR ( Table 1 ). The analysts of the nPCR were blinded to the malaria status of the samples before performing the technique. Once the result of the nPCR was obtained, the result of the microscopy and Gazelle™ were revealed to decide if it was necessary to repeat the PCR. The reactions were carried out by an initial denaturation at 94 • C for 4 min, 35 cycles of 94 • C for 30 sec, annealing temperature for 60 sec (Table 1) , and 72 • C for 60 s, with a final extension at 72 • C for 4 min. Amplicons were visualized by 2% agarose gel electrophoresis with ethidium bromide. The positive and negative controls were included in each set of reactions. The turnaround time and cost per test were calculated for the thick blood smear, Gazelle™, and nPCR. The cost calculation included consumables and reagents. The cost of collecting the blood sample and preparing the thick smear for microscopy was included, as well as the cost of sampling, cartridges, and buffers for Gazelle™. Filter paper, semiautomatic DNA extraction, amplification reactions, and electrophoresis for PCR were considered. The fixed costs for labor, the laboratory, and the equipment were not included. The sensitivity (Sen), specificity (Spe), positive and negative predictive values (PPV, NPV) were calculated for microscopy and Gazelle™ using the PCR as the reference assay, where Sen = true positives/total positives * 100; Spe = true negatives/total negatives * 100; PPV = true positives/(true positives + false positives) * 100; NPV = true negatives/(true negatives + false negatives) * 100. The diagnostic accuracy (or effectiveness) was calculated as follows: (true positives + true negatives)/(true positives + true negatives + false positives + false negatives) *100 [38] . The total number of samples were those on which all three assays were successfully performed. Then, 95% confidence intervals were calculated for the sensitivity and specificity of microscopy and Gazelle™. Receiver operating characteristic (ROC) analysis and areas under the curve (AUC) were carried out to assess diagnostic accuracy and to compare the diagnostic performance of microscopy, Gazelle™, and nested PCR. AUC was interpreted as follows: 0.9-1.0, excellent; 0.8-0.9, very good; 0.7-0.8, good; 0.6-0.7, sufficient; 0.5-0.6, bad; <0.5, test not useful [39] . The Cohen s kappa coefficient of agreement of microscopy and Gazelle™ using the nPCR as reference was also calculated as follows: k = p o − p e /1 − pe, where p o is the relative observed agreement among assays, and p e is the hypothetical probability of chance agreement. In addition, the kappa index between microscopy and Gazelle™ was calculated using microscopy as a reference. The kappa index result was interpreted as follows: values ≤ 0 indicate no agreement and 0.01-0.20 as none to slight, 0.21-0.40 as fair, 0.41-0.60 as moderate, 0.61-0.80 as substantial, and 0.81-1.00 as almost perfect agreement [40] . The McNemar chi-square was calculated to assess statistical differences between the results of Gazelle™ compared to LM. We enrolled 220 febrile patients who requested medical care at the Puerto Lempira Hospital. Out of those, 145 (66%) were female, and 73 (33.2%) were minors under the age of 18 years. The average age of the participants was 24.6 years, with an age range of one month to 74 years. Microscopy diagnosed 50 (22.73%) samples with malaria, whereas Gazelle™ detected 48 (21.82%) positives, and nPCR detected 77 (35%) positives (Table 2) . Microscopy detected 30 samples infected by P. vivax, 19 by P. falciparum, and one sample with mixed infection. Nested PCR revealed 50 samples infected with P. vivax, 19 with P. falciparum, and eight with mixed infections (Table 2) . When comparing the concordance in the identification of Plasmodium species between microscopy and nPCR, microscopy misidentified three P. vivax infections as P. falciparum and missed seven cases with mixed infection (Table 3) . According to microscopy, the median parasite density (parasites/100 leukocytes) was 57, with a range of 0.4 to 223. Parasitaemia was considered high, moderate, or low in 28 (56%), 14 (28%), and eight (16%) patients, respectively. Only one patient showed a parasitemia under 200 parasites/µL and was shown to be negative by Gazelle™. The WHO uses this value in the RDT quality assurance program as a cut-off point to assess the sensitivity of rapid diagnostic tests [41] . In four positive cases of P. falciparum detected by microscopy, Gazelle™ did not detect hemozoin (one case with high parasitemia, two moderates, and one low). All four cases were confirmed by nPCR, and accordingly, they were considered false negatives by Gazelle™. In contrast, two positive cases by Gazelle™ were not detected by microscopy. In none of them were the infections were confirmed by nPCR, and therefore they were classified as false positives by Gazelle™. The kappa index between microscopy and Gazelle™ was 0.9216 using microscopy as reference. Using nPCR as the reference test, microscopy had a sensitivity of 64.9% (95%CI: 62.8-67.0%) and a specificity of 100% (95%CI: 98.6-101.4%), while Gazelle™ had a sensitivity of 59.7% (95%CI: 57.9-61.6%) and a specificity of 98.6% (95%CI: 98.2-99.0%) ( Table 4 ). The ROC analysis showed that the area under the curve (AUC) was 0.79 for Gazelle™ and 0.82 for microscopy ( Figure 3 ). The McNemar chi-square test did not show statistical differences between the results of Gazelle compared to LM (p = 0.6831). (56%), 14 (28%), and eight (16%) patients, respectiv sitemia under 200 parasites/μL and was shown to uses this value in the RDT quality assurance progr sitivity of rapid diagnostic tests [41] . In four positive cases of P. falciparum detected tect hemozoin (one case with high parasitemia, tw cases were confirmed by nPCR, and accordingly, th Gazelle TM . In contrast, two positive cases by Gazel In none of them were the infections were confirm classified as false positives by Gazelle TM . The kapp zelle TM was 0.9216 using microscopy as reference. Using nPCR as the reference test, microscopy h 67.0%) and a specificity of 100% (95%CI: 98.6-101. of 59.7% (95%CI: 57.9-61.6%) and a specificity of 98 ROC analysis showed that the area under the curve for microscopy ( Figure 3 ). The McNemar chi-squa ences between the results of Gazelle compared to L The predictive values, effectiveness, and kappa index for microscopy and Gazelle™ are also shown in Table 4. According to the McHugh table, the agreement between microscopy and Gazelle™ with respect to nPCR was moderate, and the agreement between microscopy and Gazelle™ was almost perfect. When comparing the percentage of false negatives for Gazelle™ vs. nPCR by parasite species, Gazelle™ missed 20 of 50 (40%) P. vivax infections and 10 of 19 (52.6%) P. falciparum infections. If mixed infections are included in the analysis, Gazelle™ missed 21 of 58 P. vivax/mixed infections (36.2%) and 11 of 27 P. falciparum/mixed infections (40.7%). Due to inconsistencies between two assays, 56 of 220 (25.5%) samples were analyzed more than once by nPCR. Out of those, 43 of 56 (76.8%) samples were positive in two or three consecutive nPCR assays, while 10 of 56 (17.9%) had two positives and one negative result, and 3 of 56 (5.4%) had two negatives and one positive result. As shown in Table 5 , the turnaround time was shorter for Gazelle™ than for microscopy, but the cost per sample was slightly higher for Gazelle™, not considering fixed costs, human resources, and equipment. The turnaround time and costs per sample are considerably higher for nPCR. In this study, we evaluated the in-field performance of a novel diagnostic device (Gazelle™) based on the detection of hemozoin, comparing it with microscopy and nested PCR. The sensitivity of Gazelle™ was lower but comparable to microscopy (60% versus 65%, respectively), with similar effectiveness and predictive values and a kappa index showing almost perfect agreement between both methods. Other field studies have evaluated the performance of Gazelle™. A study conducted in the Peruvian Amazon that included samples infected by P. vivax reported a sensitivity of 88.2% for Gazelle™, slightly lower than the sensitivity of LM, although statistically similar, and using a nested PCR as the gold standard [42] . In the Brazilian Amazon, Gazelle™ and LM were compared using TaqMan qPCR as a reference. Only malaria vivax samples were evaluated, with a sensitivity of 72.1% for Gazelle™ and 75% for microscopy. Gazelle™ sensitivity was 96.2% compared to LM as a reference [43] . In India, Gazelle™ was evaluated in a region where four Plasmodium species are endemic. The sensitivity of Gazelle™ was 82.1% compared to nPCR, while the sensitivity of LM was 75%. The sensitivity of Gazelle™ was 97.6% using LM as the reference test [44] . In Papua New Guinea, the sensitivities of Gazelle™ and LM were 78% and 82%, respectively, compared to nested PCR, in a region where three parasite species coexist [45] . Regarding the specificity, our results are similar to those of other studies, approaching 100% in all cases, which shows that Gazelle™ is a reliable tool to differentiate between positive and negative cases. In general, the results obtained here are consistent with the literature, and the relatively low sensitivity for both Gazelle™ and LM could be attributed to unknown biological characteristics of the parasites circulating in the region or low parasitic burden [43] . Low parasitaemia and submicroscopic infections have been described as the main causes of the low sensitivity of common malaria diagnostic tools [46, 47] , particularly among asymptomatic individuals and in low malaria-endemic settings [7, 48] . In this study, parasitaemia was only calculated for positive samples by microscopy; therefore, it was not possible to determine the parasite burden in samples positive by PCR but negative by microscopy or Gazelle™. However, in 13 of 56 cases (23.2%) in which PCR was performed in triplicate, there were discordant results between the PCRs. This could suggest that the parasitemia of these individuals was around the detection limit of the most sensitive technique (nPCR) and could be responsible for the low sensitivity of Gazelle™. On the other hand, and at the direct suggestion of the manufacturer, in this study, the method of preparing the blood sample deposited in the Gazelle™ cartridge was modified. Previous studies have used 15 µL of blood in 80 µL of the diluent [42] [43] [44] ; instead, in this study, we used 30 µL of blood in 65 µL of diluent. Although doubling the sample volume should have increased sensitivity, this variable could be responsible for an inverse effect. However, further studies will be necessary to evaluate this possibility. In this study, as well as in previous reports, Gazelle™ detected a few false-positive cases. Arndt et al. showed that past malaria infections are associated with residual levels of hemozoin, especially in individuals infected within the last two weeks [45] . However, this hypothesis cannot be verified here due to the lack of access to the clinical history of the participants. On the other hand, Gazelle™ missed 40% of P. vivax infections but 52.6% of P. falciparum infections, which could be due to a difference in sensitivity for both species. Some authors have shown that P. falciparum infections tend to go unnoticed by magnetooptical detection of hemozoin [45] . Since the concentration of hemozoin in the peripheral bloodstream depends in part on the circulating stages of the parasite, and because ring immature trophozoites predominate in P. falciparum infections in the absence of other asexual stages, the amount of hemozoin is lower compared to P. vivax infection [49, 50] . About the implementation of the technique, Gazelle™ has the advantage of a very fast response time (less than 5 min), which would positively impact timely diagnosis and treatment. In addition, the cost per test is similar to that of microscopy and RDTs, and although an initial investment of USD 700 is required to acquire the device, the amount is justified by the high average life, with a capacity of up to 20,000 tests [42] . Another advantage of Gazelle™ is its portability and the ability to run on battery power and in high temperatures, making it suitable for remote locations in the tropics with limited infrastructure. One drawback is that Gazelle™ requires phlebotomy to collect the venous blood sample, as well as the correct use of a micropipette to dispense the proper volume of blood and diluent into the cartridge. These skills are greater than those required by an RDT but much less than those of an expert microscopist. The subjectivity of microscopy is another limitation overcome with Gazelle™. The relatively little training required to operate Gazelle™ would allow its use to be expanded in remote communities with a trained technician available. This makes Gazelle™ a possible solution to the lack of skilled microscopists in countries approaching elimination. One of the most attractive features of the device is the ability to store data and transmit it electronically over a wireless connection, enabling real-time responses from decision-makers. The most notable disadvantage of Gazelle™ is its inability to distinguish between the species of the parasite. Plasmodium vivax and P. falciparum circulate in Honduras, and the national guidelines establish a differentiated treatment with primaquine 0.25 mg/kg for 14 days against P. vivax to ensure the elimination of liver stages and 0.75 mg/kg in singledose against P. falciparum on the first day of treatment. In case of mixed infection, the scheme against P. vivax is followed [34] . Consequently, the diagnosis of malaria requires mandatory identification of the species of the parasite and the determination of the parasitemia. Including a method such as Gazelle™ in the national malaria program would require that all positive samples be subsequently analyzed by microscopy before the administration of treatment, which would be impractical and increase costs. A new version of Gazelle should detect all parasite species with sensitivity at least equal to that of LM or RDTs. This study evaluated the in-field performance of a device based on the magneto-optical detection of hemozoin (Gazelle™), demonstrating sensitivity and specificity comparable to LM. Its short turnaround time, low cost, and ease of use make it a valuable tool for point-of-care malaria detection and an alternative to the lack of trained microscopists. However, further development is required to differentiate the species of Plasmodium if the device is to be introduced in places whose regulations establish a differentiated treatment. World Health Organization. Global Technical Strategy for Malaria Malaria diagnostic update: From conventional to advanced method Laboratory diagnosis of malaria Rapid diagnostic tests for malaria parasites Low quality of routine microscopy for malaria at different levels of the health system in Dar es Salaam Diagnostic performance of rapid diagnostic test, light microscopy and polymerase chain reaction during mass survey conducted in low and high malaria-endemic areas from two North-Eastern states of India Malaria rapid diagnostic tests in elimination settings-Can they find the last parasite? Dependence of malaria detection and species diagnosis by microscopy on parasite density Identification of the four human malaria parasite species in field samples by the polymerase chain reaction and detection of a high prevalence of mixed infections A review of the WHO malaria rapid diagnostic test product testing programme Sensitivity and specificity of malaria rapid diagnostic test (mRDT CareStatTM) compared with microscopy amongst under five children attending a primary care clinic in southern Nigeria False-negative rapid diagnostic tests for malaria and deletion of the histidine-rich repeat region of the hrp2 gene Deletions of pfhrp2 and pfhrp3 genes of Plasmodium falciparum from Honduras, Guatemala and Nicaragua A direct, sensitive and high-throughput genus and species-specific molecular assay for large-scale malaria screening Ultraviolet/Visible and Near-Infrared Dual Spectroscopic Method for Detection and Quantification of Low-Level Malaria Parasitemia in Whole Blood An Optimized Real-Time qPCR Method for the Effective Detection of Human Malaria Infections Sensitive detection of Plasmodium vivax malaria by the rotating-crystal magneto-optical method in Thailand A novel computation method for detection of Malaria in RBC using Photonic biosensor The Role of Ultrasensitive Molecular Methods for Detecting Malaria-The Broader Perspective An Automated Microscopic Malaria Parasite Detection System Using Digital Image Analysis Malaria Screening Using Front-Line Loop-Mediated Isothermal Amplification Novel molecular diagnostic tools for malaria elimination: A review of options from the point of view of high-throughput and applicability in resource limited settings The peculiarities and paradoxes of Plasmodium heme metabolism Hemoglobin metabolism in the malaria parasite Plasmodium falciparum Full blood count and haemozoincontaining leukocytes in children with malaria: Diagnostic value and association with disease severity Birefringent hemozoin identifies malaria The Future in Sensing Technologies for Malaria Surveillance: A Review of Hemozoin-Based Diagnosis Evidence of a Recent Bottleneck in Plasmodium falciparum Populations on the Honduran-Nicaraguan Border Distribution and phylogenetic diversity of Anopheles species in malaria endemic areas of Honduras in an elimination setting Statistical methodology: I. Incorporating the prevalence of disease into the sample size calculation for sensitivity and specificity Secretaría de Salud de Honduras. Norma Nacional de Malaria en Honduras Densidad parasitaria en malaria: Métodos de determinación y su interpretación. Rev. Médica Hondureña A genus-and species-specific nested polymerase chain reaction malaria detection assay for epidemiologic studies Comparison of molecular tests for the diagnosis of malaria in Honduras Measures of Diagnostic Accuracy: Basic Definitions Receiver Operating Characteristic (ROC) Curve Analysis for Medical Diagnostic Test Evaluation Interrater reliability: The kappa statistic World Health Organization. Malaria Rapid Diagnostic Test. Performance: Results of WHO Product Testing of Malaria RDTs: Round Field validation of a magnetooptical detection device (Gazelle) for portable point-of-care Plasmodium vivax diagnosis Performance of a sensitive haemozoin-based malaria diagnostic test validated for vivax malaria diagnosis in Brazilian Amazon First successful field evaluation of new, one-minute haemozoin-based malaria diagnostic device Magneto-optical diagnosis of symptomatic malaria in Papua New Guinea High-resolution melting analysis reveals low Plasmodium parasitaemia infections among microscopically negative febrile patients in western Kenya Low parasitemia in submicroscopic infections significantly impacts malaria diagnostic sensitivity in the highlands of Western Kenya Evaluation of the loop mediated isothermal DNA amplification (LAMP) kit for malaria diagnosis in P. vivax endemic settings of Colombia Limitations of haemozoin-based diagnosis of Plasmodium falciparum using dark-field microscopy Haemozoin detection in infected erythrocytes for Plasmodium falciparum malaria diagnosis-prospects and limitations