key: cord-0032345-0vnmnqts authors: Sharma, Sandhya; Thomas, Emmanuel; Caputi, Massimo; Asghar, Waseem title: RT-LAMP-Based Molecular Diagnostic Set-Up for Rapid Hepatitis C Virus Testing date: 2022-05-05 journal: Biosensors (Basel) DOI: 10.3390/bios12050298 sha: e9570c3a77975d584268fcc06a0e6c4415610dd0 doc_id: 32345 cord_uid: 0vnmnqts Hepatitis C virus (HCV) infections occur in approximately 3% of the world population. The development of an enhanced and extensive-scale screening is required to accomplish the World Health Organization’s (WHO) goal of eliminating HCV as a public health problem by 2030. However, standard testing methods are time-consuming, expensive, and challenging to deploy in remote and underdeveloped areas. Therefore, a cost-effective, rapid, and accurate point-of-care (POC) diagnostic test is needed to properly manage the disease and reduce the economic burden caused by high case numbers. Herein, we present a fully automated reverse-transcription loop-mediated isothermal amplification (RT-LAMP)-based molecular diagnostic set-up for rapid HCV detection. The set-up consists of an automated disposable microfluidic chip, a small surface heater, and a reusable magnetic actuation platform. The microfluidic chip contains multiple chambers in which the plasma sample is processed. The system utilizes SYBR green dye to detect the amplification product with the naked eye. The efficiency of the microfluidic chip was tested with human plasma samples spiked with HCV virions, and the limit of detection observed was 500 virions/mL within 45 min. The entire virus detection process was executed inside a uniquely designed, inexpensive, disposable, and self-driven microfluidic chip with high sensitivity and specificity. According to the World Health Organization (WHO), more than 354 million people worldwide are infected with the Hepatitis C virus (HCV), of which 70 million are chronically infected. Each year, an estimated 1 million people die from this disease [1, 2] . According to the WHO, in the year 2019, 58 million people were reported with HCV infection worldwide [1] . During the same time, in the United States, 4136 were reported with acute infection; however, the estimated number of cases was 57,500 [3] . The Centers for Disease Control and Prevention (CDC) estimates that, in the United States, about 50% of the infected individuals are unaware of their infection [4] . HCV is primarily transmitted by parenteral drug administration, blood/plasma transfusion, the reuse of medical equipment, sexual practices resulting in blood transfer, and sharing the same needle while injecting drugs [4] [5] [6] . Patients are diagnosed with HCV after they exhibit symptoms arising from the chronic infection. However, patients are often chip exhibited a LOD of 500 virion copies/mL within 45 min. The device is cost-effective, user-friendly, portable, and provides a visual confirmation while utilizing a small amount of sample and few reagents in a time-efficient manner. This device is, therefore, ideal for application at POC and in underdeveloped countries. Figure 1 summarizes the experimental workflow and the schematics of the microfluidic chip utilized in the assays. 12, x FOR PEER REVIEW 3 of 13 orange to green in the presence of double-stranded DNA, resulting in easy analysis without fluorescent imaging. To test our chip, we utilized plasma spiked with HCV. The RT-LAMP-based microfluidic chip exhibited a LOD of 500 virion copies/mL within 45 min. The device is cost-effective, user-friendly, portable, and provides a visual confirmation while utilizing a small amount of sample and few reagents in a time-efficient manner. This device is, therefore, ideal for application at POC and in underdeveloped countries. Figure 1 summarizes the experimental workflow and the schematics of the microfluidic chip utilized in the assays. The 5' untranslated region (5' UTR) of the JFH-1 isolate RNA genomic sequence (Gen-Bank: AB047639.1) [52] was used as a target. The RT-LAMP primers were designed with the freely available online software Primer Explorer V5 from Eiken Chemical Co. Ltd (Tokyo, Japan). A BLAST search was carried out in the GenBank nucleotide of the primers. The conserved target DNA sequence was synthesized (Integrated DNA Technologies) for the initial sensitivity and specificity testing of the designed primer set. The LavaLAMP RNA Master Mix kit from Lucigen (Middleton, WI USA) was used for all the RT-LAMP reactions (both benchtop and on-chip testing). The LAMP reactions set-up included Master Mix (12.5 L), green fluorescent dye (1 L), HCV RT-LAMP primers (2.5 L), HCV, ZIKA, SARS-CoV-2, HIV or RNase/DNase-free water (1 L), and RNase/DNase-free water (8 L). The AriaMx Real-Time PCR system (Agilent) was utilized to maintain 70 °C for 40 min for the isothermal amplification. The LAMP-amplified products were further confirmed with 1.5% agarose gel electrophoresis. The gel was subjected to 90 volts for 90 min. The HCV cultured samples were obtained from the University of Miami. The HCV strain was propagated in Huh 7.5.1 human hepatoma cell line. The virus titer was quantified by qPCR. Human blood samples from deidentified healthy individuals were obtained from Continental Services Group, Inc. (Fort Lauderdale, FL, USA). Under the institutional The 5' untranslated region (5' UTR) of the JFH-1 isolate RNA genomic sequence (Gen-Bank: AB047639.1) [52] [M1] was used as a target. The RT-LAMP primers were designed with the freely available online software Primer Explorer V5 from Eiken Chemical Co. Ltd (Tokyo, Japan). A BLAST search was carried out in the GenBank nucleotide of the primers. The conserved target DNA sequence was synthesized (Integrated DNA Technologies) for the initial sensitivity and specificity testing of the designed primer set. The LavaLAMP RNA Master Mix kit from Lucigen (Middleton, WI, USA) was used for all the RT-LAMP reactions (both benchtop and on-chip testing). The LAMP reactions set-up included Master Mix (12.5 µL), green fluorescent dye (1 µL), HCV RT-LAMP primers (2.5 µL), HCV, ZIKA, SARS-CoV-2, HIV or RNase/DNase-free water (1 µL), and RNase/DNase-free water (8 µL). The AriaMx Real-Time PCR system (Agilent) was utilized to maintain 70 • C for 40 min for the isothermal amplification. The LAMP-amplified products were further confirmed with 1.5% agarose gel electrophoresis. The gel was subjected to 90 volts for 90 min. The HCV cultured samples were obtained from the University of Miami. The HCV strain was propagated in Huh 7.5.1 human hepatoma cell line. The virus titer was quantified by qPCR. Human blood samples from deidentified healthy individuals were obtained from Continental Services Group, Inc. (Fort Lauderdale, FL, USA). Under the institutional review board (IRB)-approved protocol, the plasma was separated from the blood by centrifuga-tion for 15 min at 2000× g. The plasma was spiked with HCV varying from 2.8 × 10 7 to 28 virion copies/mL. The viral RNA was isolated from the plasma samples, utilizing the Dynabeads SILANE viral Nucleic Acid (NA) kit (Invitrogen). A total of 100 µL of plasma sample was used for the RNA isolation, following the manufacturer's instruction. The RNA was then eluted in 50 µL of elution buffer and the RT-LAMP assay was carried out in a thermocycler, as described above. Fluorescence data was collected for the endpoint analysis. In addition, the amplification products of the LAMP reaction were analyzed by 1.5% gel electrophoresis. A total of 1 µL of SYBR green I nucleic A dye (Invitrogen) was added to the reaction solution after the isothermal amplification for the colorimetric visualization. Microfluidic chips were fabricated from three-layered poly(methyl methacrylate) (PMMA) sheets. The layers of the chip were attached together using double-sided adhesive (DSA) tape. Table S2 and Figure S1 show the thickness and dimension of each layer and the chambers of the microfluidic chip. Figure 2a ,b show the complete chip (top and vertical view). The chip layout was designed in AutoCAD, and a CO 2 laser cutter was utilized to obtain the chambers, as previously reported [22, 40, [52] [53] [54] . Each microfluidic chip consists of 4 independent diamond-shaped aqueous chambers: one sample inlet chamber, two washing buffer chambers, and one reaction chamber separated by three elliptical-shaped valving chambers containing mineral oil (viscosity-15 cSt). An unconnected oval-shaped sensor chamber adjacent to the reaction chamber is imprinted for sensor attachment. The assembled microfluidic chip was subjected to UV for 30 min. After 30 min, all the inlets were blocked with scotch tape. The scotch tape was removed from the microfluidic chip for the filling of reagents. was then eluted in 50 l of elution buffer and the RT-LAMP assay was carried out in a thermocycler, as described above. Fluorescence data was collected for the endpoint analysis. In addition, the amplification products of the LAMP reaction were analyzed by 1.5% gel electrophoresis. A total of 1 L of SYBR green I nucleic A dye (Invitrogen) was added to the reaction solution after the isothermal amplification for the colorimetric visualization. Microfluidic chips were fabricated from three-layered poly(methyl methacrylate) (PMMA) sheets. The layers of the chip were attached together using double-sided adhesive (DSA) tape. Table S2 and Figure S1 show the thickness and dimension of each layer and the chambers of the microfluidic chip. Figures 2a,b show the complete chip (top and vertical view). The chip layout was designed in AutoCAD, and a CO2 laser cutter was utilized to obtain the chambers, as previously reported [22, 40, [52] [53] [54] . Each microfluidic chip consists of 4 independent diamond-shaped aqueous chambers: one sample inlet chamber, two washing buffer chambers, and one reaction chamber separated by three elliptical-shaped valving chambers containing mineral oil (viscosity-15 cSt). An unconnected oval-shaped sensor chamber adjacent to the reaction chamber is imprinted for sensor attachment. The assembled microfluidic chip was subjected to UV for 30 min. After 30 min, all the inlets were blocked with scotch tape. The scotch tape was removed from the microfluidic chip for the filling of reagents. The automated diagnostic platform designed and optimized by our lab is depicted in Figure 2c [55] . The diagnostic platform moves the magnetic beads automatically. The bead's movement is directed by two small magnets (5 mm-diameter neodymium) fitted in a 3-Dprinted inclusion. This 3-D-printed inclusion can move bidirectionally, and its movement is coordinated by a stepper motor. The stepper motor is guided by a printed circuit board (PCB) Arduino. The movement of the magnetic beads from one chamber to another and the incubation time were controlled by a G-code scripted in Python. We utilized an Arduino temperature controller to maintain the 70 °C temperature required for on-chip amplification [22] . The sensor is placed in the sensor chamber. The automated diagnostic platform designed and optimized by our lab is depicted in Figure 2c [55] . The diagnostic platform moves the magnetic beads automatically. The bead's movement is directed by two small magnets (5 mm-diameter neodymium) fitted in a 3-D-printed inclusion. This 3-D-printed inclusion can move bidirectionally, and its movement is coordinated by a stepper motor. The stepper motor is guided by a printed circuit board (PCB) Arduino. The movement of the magnetic beads from one chamber to another and the incubation time were controlled by a G-code scripted in Python. We utilized an Arduino temperature controller to maintain the 70 • C temperature required for on-chip amplification [22] . The sensor is placed in the sensor chamber. The microfluidic chip was loaded with reagents and buffers, as described in Table S3 . The inlet chamber (a in Figure 1 ) contained the lysis/binding buffer + proteinase K + magnetic beads + isopropanol. Washing buffers I and II were added to the washing chambers (b and c in Figure 1 ) and their viscosity was increased by adding RNase/DNase-free water in a 1:1 ratio. The reaction chamber and sensor chamber (d and e in Figure 1 ) solution contained LavaLAMP RNA Master Mix, HCV-specific primers, and elution buffer. The MgSO 4 concentration of LavaLAMP RNA Master Mix was increased from 5 mM to 9.8 mM for on-chip testing. A total of 100 µL of plasma sample was loaded in the inlet chamber of the pre-filled chip and placed on the diagnostic set-up. Next, the heater and magnetic actuation were started concurrently, and the automated set-up was incubated for 45 min. SYBR green 1 dye was added to the reaction chamber after the isothermal incubation and mixed using the oscillatory movement of the magnetic beads to detect the reaction amplification products. All the on-chip experiments were repeated at least three times. HCV is comprised of seven distinct genotypes, which can differ by as much as 30% at the nucleotide level [56] [57] [58] . Though sequence variations are evenly distributed throughout the genome, the 5' UTR is conserved [59] [60] [61] . The HCV RT-LAMP primers we designed (Table 1 ) are mapped on the highly conserved 5' UTR region. The designed primer set was BLAST, against all the genomes present in the National Center of Biotechnology Information (NCBI). The results showed high diversity against other genomes and identify to the HCV genome 5' UTR region. We performed an initial RT-LAMP assay to test the sensitivity and specificity of the primer set we designed ( Figure 3A) , utilizing a synthetic target HCV DNA sequence with different target concentrations (10 9 to 0 copies per reaction). The lowest LOD observed was 5 HCV copies/reaction within 40 min. A no-template control (NTC) confirmed the absence of primer dimer formation. The R 2 value (0.967) of the time to amplification relative to the concentration of the target substrate indicates that the amplification time of the target is inversely proportional to the target concentration in the reaction. These results were confirmed by analyzing the LAMP products by agarose gel electrophoresis. Amplification products can be observed in the reactions that amplify the target HCV sequences, forming long sharp bands ( Figure 3B ). The HCV primer set was tested by amplifying the cDNA obtained from the RNA genome of ZIKA virus, [22] SARS-CoV-2, [62] and HIV [63] (Figure 4 and Figure S2 ). The HCV primer sets did not amplify viral sequences from these viruses, confirming the specificity of the RT-LAMP primers. Amplification products can be observed in the reactions that amplify the target HCV sequences, forming long sharp bands ( Figure 3B ). The HCV primer set was tested by amplifying the cDNA obtained from the RNA genome of ZIKA virus, [22] SARS-CoV-2, [62] (a) (b) As a proof-of-principle, human plasma was spiked with a cultured HCV strain to replicate clinical samples. HCV preparations were added to the plasma samples with 10fold dilutions, obtaining final concentrations ranging between 2.8 × 10 7 and 28 virions/mL. The viral RNA was extracted from the spiked plasma samples utilizing magnetic Amplification products can be observed in the reactions that amplify the target HCV sequences, forming long sharp bands ( Figure 3B ). The HCV primer set was tested by amplifying the cDNA obtained from the RNA genome of ZIKA virus, [22] SARS-CoV-2, [62] (a) (b) As a proof-of-principle, human plasma was spiked with a cultured HCV strain to replicate clinical samples. HCV preparations were added to the plasma samples with 10fold dilutions, obtaining final concentrations ranging between 2.8 × 10 7 and 28 virions/mL. The viral RNA was extracted from the spiked plasma samples utilizing magnetic As a proof-of-principle, human plasma was spiked with a cultured HCV strain to replicate clinical samples. HCV preparations were added to the plasma samples with 10-fold dilutions, obtaining final concentrations ranging between 2.8 × 10 7 and 28 virions/mL. The viral RNA was extracted from the spiked plasma samples utilizing magnetic Dynabeads. The RT-LAMP of the samples yielded a LOD of 28 virions/mL within 60 min (Figure 5a ). From this assay, a fitted linear trendline was observed with R 2 = 0.865, indicating the high efficiency of the RNA extraction/RT-LAMP analysis with limited amounts of virus. These results validate the efficiency of the magnetic beads' RNA extraction and the fact that the RNA maintains its integrity throughout the purification steps. Figure S3 represents the gel results. A colorimetric test was performed utilizing SYBR green I dye, which is initially orange but turns to green in the presence of DNA, to enable a fast and simpler readout [64] . The SYBR green I was added to the reaction tubes after the isothermal incubation and confirmed the LOD for the assay for the samples containing 28 virion copies/mL, which showed an attenuated shift to the green coloration when compared with the control reactions that retained the original orange color (Figure 5b ). Dynabeads. The RT-LAMP of the samples yielded a LOD of 28 virions/mL within 60 min (Figure 5a ). From this assay, a fitted linear trendline was observed with R 2 = 0.865, indicating the high efficiency of the RNA extraction/RT-LAMP analysis with limited amounts of virus. These results validate the efficiency of the magnetic beads' RNA extraction and the fact that the RNA maintains its integrity throughout the purification steps. Figure S3 represents the gel results. A colorimetric test was performed utilizing SYBR green I dye, which is initially orange but turns to green in the presence of DNA, to enable a fast and simpler readout [64] . The SYBR green I was added to the reaction tubes after the isothermal incubation and confirmed the LOD for the assay for the samples containing 28 virion copies/mL, which showed an attenuated shift to the green coloration when compared with the control reactions that retained the original orange color (Figure 5b ). Color change from orange to green was observed due to the amplification in the reactions (2.8 × 10 7 to 282, and slightly in 28 HCV virion copies/mL) using SYBR green I dye. In recent years, molecular detection microfluidic platforms have transformed the healthcare landscape, as they offer the rapid detection of viruses and other pathogens [65] [66] [67] [68] [69] . Herein, we have designed a microfluidic platform that incorporates different steps that are usually performed by trained personnel with sophisticated lab settings on a single platform. The microchip is designed with distinct shapes of chambers so that the solutions can be retained during the entire execution process. The diamond-shape chambers contain different buffers performing different tasks for optimal RNA purification. The inlet and reaction chambers are dual-purpose chambers, making the plasma processing steps less complex. These diamond chambers are separated by elliptical chambers which contain mineral oil. The oil reduces the interfacial energy barrier that facilitates the smooth passage of the beads from one buffer to another. The geometry of the chambers is designed in such a way that the magnetic beads align with magnets situated on the diagnostic platform. This provides an approachable magnetic field to the beads for flexible RNA isolation processing. Using a 750 µm PMMA sheet at the bottom also contributed to the maximum magnetic force, as the distance between the magnetic beads and the magnet was Color change from orange to green was observed due to the amplification in the reactions (2.8 × 10 7 to 282, and slightly in 28 HCV virion copies/mL) using SYBR green I dye. In recent years, molecular detection microfluidic platforms have transformed the healthcare landscape, as they offer the rapid detection of viruses and other pathogens [65] [66] [67] [68] [69] . Herein, we have designed a microfluidic platform that incorporates different steps that are usually performed by trained personnel with sophisticated lab settings on a single platform. The microchip is designed with distinct shapes of chambers so that the solutions can be retained during the entire execution process. The diamond-shape chambers contain different buffers performing different tasks for optimal RNA purification. The inlet and reaction chambers are dual-purpose chambers, making the plasma processing steps less complex. These diamond chambers are separated by elliptical chambers which contain mineral oil. The oil reduces the interfacial energy barrier that facilitates the smooth passage of the beads from one buffer to another. The geometry of the chambers is designed in such a way that the magnetic beads align with magnets situated on the diagnostic platform. This provides an approachable magnetic field to the beads for flexible RNA isolation processing. Using a 750 µm PMMA sheet at the bottom also contributed to the maximum magnetic force, as the distance between the magnetic beads and the magnet was minimized. The magnetic beads used for the isolation process are composed of cross-linked polystyrene and magnetic material. Silica-like surface chemistry offers excellent binding of RNA, and the magnetic property of the beads aids with fast mobility under the magnetic field. Furthermore, given the beads' low sediment rate, no adhesion was observed at the bottom. The incubation of the beads in each chamber is facilitated by oscillatory movement directed by a stepper motor on the platform, which is controlled by a G-code. Table S4 represents the beads' incubation time in each chamber, directed by the automated magnets located on the platform. For on-chip isothermal temperature, an automated Arduino-based temperature controller is used, which consists of a K-type thermocouple sensor and an ultrathin nano-carbon flexible heater. A K-type thermocouple sensor reads the real-time temperature. The heater heats the reagents in the sensor chamber and reaction chamber. To maintain the set isothermal temperature, the sensor reads the temperature every 2 s. Once the temperature is higher than the set temperature, the sensor turns "off" the heater, and if the temperature is lower than the set temperature, it turns "on" the heater. To avoid direct contact between the sensor and the reaction chamber solution, the sensor was inserted into the sensor chamber. The sensor chamber contains the same reagents as the reaction chamber, and one surface heater was attached on the top of both chambers. RNA degradation could lead to false-negative results in an RNA-based detection device. To address this potential problem, we exposed the RT-LAMP microfluidic chip to UV for 30 min, which eliminated traces of RNases from solid surfaces [70, 71] . On the microfluidic chip, the key role of the inlet chamber is to lyse the viral particles and to allow binding of the viral RNA to the magnetic beads in the presence of lysis and binding buffer. Once the RNA adheres to the surface of the magnetic beads, they are directed toward the reaction chamber under the influence of a magnetic field. The two washing chambers contain two different washing buffers to ensure that reagents and other potential inhibitory compounds are not transferred from the inlet chamber to the reaction chamber. Once the beads reach the reaction chamber, which is already maintained at 70 • C with the help of the sensor chamber and surface heater, beads elute the RNA and move back to the washing chamber. The RT-LAMP HCV primers amplify the HCV RNA if present, and change the color of the chamber to green; if HCV RNA is not present initially, the color of the dye will remain unchanged. For the on-chip assay, plasma samples were spiked with different amounts of HCV concentration (from 2.8 × 10 6 to 282 virions/mL). A no-template control assay was carried with non-spiked plasma samples. Since PMMA adsorbs polar molecules, such as DNA and polymerase enzymes that inhibit amplification [72] , the addition of ions such as Mg ++ , Na + , and K + enhances the enzyme activity and provides DNA stability [73, 74] . To enhance the on-chip amplification efficiency of the target, the MgSO 4 concentration of the Master Mix was increased from 5 mM to 9.8 mM in the reaction chamber. Figure 6a shows the color comparison of the reaction chambers of the microfluidic chip subjected to a non-spiked plasma sample (orange color in the reaction chamber) and HCV-spiked sample (green color reaction chamber) after 45 min. Figure 6b demonstrated the on-chip sensitivity of the set-up: 500 virions/mL within 45 min. At the same time, no color change was observed in the negative control (non-spiked plasma) sample and 280 virions/mL. These results validate that the RT-LAMP-based microfluidic chip we developed can efficiently isolate and amplify HCV RNA. The molecular diagnostic set-up here described provides qualitative colorimetric results from plasma samples spiked with HCV. The data presented demonstrate that this compact microfluidic device could be utilized as a "sample-in-answer-out" system for HCV screening in low-to-middle-income areas. POC microfluidic platforms offer comparable results to gold-standard methods using a small quantity of reagents and the sample can be tested outside the laboratory setting. A commercial device derived from the one described here, if utilized broadly, can play an essential role in managing HCV infection worldwide, thereby controlling the spread and enabling the timely treatment and monitoring of the disease. The 3-D-printed platform used in the research consists of a microprocessor controlled by Arduino, a step-up motor, and a circuit for the power supply. The molecular diagnostic platform is fully programmed and reusable for repetitive testing. It can also be operated using batteries to direct the movement of magnetic beads present in the chambers of the disposable microfluidic chip. The operational cost of the diagnostic set-up would be minimal, as it uses relatively inexpensive and sustainable equipment (roughly USD 50) for sample processing and disease detection. This microfluidic chip also offers shorter times for a reliable diagnosis of HCV infection. The disposable microfluidic chip we describe is affordable (roughly USD 2) and can be utilized in POC settings, as the assay is performed in an automated fashion. With this molecular diagnostic set-up, the user can run multiple tests since attention is required only initially and at the end of the colorimetric analysis. Table S5 contains the full list of materials and costs for the molecular diagnostic microfluidic chip set-up and fabrication. In this study, we developed a low-cost RNA-based POC molecular diagnostic set-up for HCV with a colorimetric result readout. The developed diagnostic method will enable timely HCV screening and prevention in high-risk populations. It is an accurate method that can be implemented in low-income areas, making it accessible to people. POC microfluidic platforms offer comparable results to gold-standard methods using a small quantity of reagents and the sample can be tested outside the laboratory setting. A commercial device derived from the one described here, if utilized broadly, can play an essential role in managing HCV infection worldwide, thereby controlling the spread and enabling the timely treatment and monitoring of the disease. The 3-D-printed platform used in the research consists of a microprocessor controlled by Arduino, a step-up motor, and a circuit for the power supply. The molecular diagnostic platform is fully programmed and reusable for repetitive testing. It can also be operated using batteries to direct the movement of magnetic beads present in the chambers of the disposable microfluidic chip. The operational cost of the diagnostic set-up would be minimal, as it uses relatively inexpensive and sustainable equipment (roughly USD 50) for sample processing and disease detection. This microfluidic chip also offers shorter times for a reliable diagnosis of HCV infection. The disposable microfluidic chip we describe is affordable (roughly USD 2) and can be utilized in POC settings, as the assay is performed in an automated fashion. With this molecular diagnostic setup, the user can run multiple tests since attention is required only initially and at the end of the colorimetric analysis. Table S5 contains the full list of materials and costs for the molecular diagnostic microfluidic chip set-up and fabrication. In this study, we developed a low-cost RNA-based POC molecular diagnostic set-up for HCV with a colorimetric A lack of overt symptoms in HCV patients often results in a lack of diagnosis, which subsequently leads to increased morbidity and mortality. Therefore, it is important to implement regular screening in high-risk populations. Molecular testing is considered the most accurate test for diagnosing HCV infection. Our research describes a singlestep procedure for the molecular detection of HCV RNA from a patient's samples with clinical applicability. The microfluidic chip utilizes an automated system for the handsoff processing of plasma samples. The testing of human plasma samples spiked with HCV particles showed that this set-up offers a sensitivity of 500 viral copies/mL and high specificity, without the need for trained technicians, expensive equipment, or facilities. The hands-free microfluidic chip used for the testing is easy to assemble, low-cost, and provides a practical approach for large-scale testing outside the laboratory. The operating procedure of the chip is straightforward; once the plasma samples are introduced to the inlet chamber, the automated system will self-operate. The cost reduction, need for less equipment, and short time (45 min) required for viral RNA detection offered by this approach would substantially reduce the financial and operational burden of large-scale testing in low-tomiddle-income countries. Overall, the HCV diagnostic test approach we propose can be clinically implemented for routine HCV screening and can help in the timely execution of preventive and therapeutic steps to limit HCV spread. Supplementary Materials: The following supporting information [28, 35, [75] [76] [77] [78] [79] [80] [81] [82] [83] [84] [85] [86] [87] [88] [89] [90] [91] [92] [93] can be downloaded at: https://www.mdpi.com/article/10.3390/bios12050298/s1, Figure S1 : Design of the microfluidic chip, illustrating the dimensions of the chip and chambers; Figure. S2: The 1.5% gel electrophoresis results stained with Bromophenol blue dye (lane L contains 1 kbp size DNA ladder). Sharp bands in the wells containing the LAMP amplification product of the HCV target, and no band formation observed in the well holding the HIV, ZIKA, and SARS-CoV-2 targets; Figure. S3: The 1.5% gel electrophoresis results stained with Bromophenol blue dye (lane L contains 1 kbp size DNA ladder). Sharp bands in the wells holding the RT-LAMP amplification product of HCV (2.8 × 10 7 to 282, and slightly with 28 HCV copies/mL) clearly show the specificity of the designed primers, and also provide a sensitivity of up to 28 HCV copies/mL; Table S1 : Summary of the existing HCV detection methods illustrating the comparison of target used, detection time, and limit of detection (LOD), along with the limitations' title; Table S2 : Microfluidic chip dimensions and thickness of various layers; Table S3 : List of the reagents loaded in the microfluidic chambers along with the volume; Table S4 : Table illustrating Data Availability Statement: All data needed to evaluate the conclusions in the paper are present in the paper and/or the Supplementary Materials. Additional data related to this paper may be requested from the authors. Hepatitis Can't Wait-WHO Commemorates World Hepatitis Day. 2021. Available online Viral Hepatitis Surveillance Report/CDC. Available online Hepatitis C virus: Screening, diagnosis, and interpretation of laboratory assays Host Genetic Variants in the Pathogenesis of Hepatitis C. Viruses Hepatitis C Antibody in Patients with Chronic Liver Disease and Hepatocellular Carcinoma Era of Direct Acting Anti-Viral Agents for the Treatment of Hepatitis C Diagnostic accuracy of assay using point-of-care testing or dried blood spot samples for the determination of HCV RNA: A systematic review The Role of Resistance in HCV Treatment Effective Drugs on the Road to HCV Elimination and a Therapeutic Gap to Close HCV Testing and Treatment Initiation in an Italian Prison Setting: A Step-by-Step Model to Micro-Eliminate Hepatitis C WHO guidelines on testing for hepatitis B and C-Meeting targets for testing Report of a WHO Consultation Organized in Collaboration with the Viral Hepatitis Prevention Board Epidemiology of Hepatitis C Virus Infection Performance of HCV Antigen Testing for the Diagnosis and Monitoring of Antiviral Treatment: A Systematic Review and Meta-Analysis Improving Linkage to Care of Hepatitis C: Clinical Validation of GeneXpert®HCV Viral Load Point-of-Care Assay in Indonesia Early Detection of Hepatitis C Virus Infection by Use of a New Combined Antigen-Antibody Detection Assay: Potential Use for High-Risk Individuals High Rate of Both Spontaneous and Treatment-Induced Viral Clearance Testing for HCV Infection: An Update of Guidance for Clinicians and Laboratorians An Antibody Panel for Highly Specific Detection and Differentiation of Zika Virus Molecular Techniques of Viral Diagnosis Lab-on-a-Chip Zika Detection With Reverse Transcription Loop-Mediated Isothermal Amplification-Based Assay for Point-of-Care Settings Validation of a Strategy for HCV Antibody Testing with Two Enzyme Immunoassays in a Routine Clinical Laboratory Tools for the Diagnosis of Hepatitis C Virus Infection and Hepatic Fibrosis Staging Systematic Review of the Accuracy of Antibody Tests Used to Screen Asymptomatic Adults for Hepatitis C Infection Circulating Tumor Cell Isolation, Culture, and Downstream Molecular Analysis Diagnosis, Management, and Treatment of Hepatitis C Hepatitis C Virus (HCV) Diagnosis via Microfluidics Reducing Read Time of Point-of-Care Test Does Not Affect Detection of Hepatitis C Virus and Reduces Need for Reflex RNA The performance of hepatitis C virus (HCV) antibody point-of-care tests on oral fluid or whole blood and dried blood spot testing for HCV serology and viral load among individuals at higher risk for HCV in South Africa. Health Sci HCV virology and diagnosis Evaluation of the Elecsys®Anti-HCV II Assay for Routine Hepatitis C Virus Screening of Different Asian Pacific Populations and Detection of Early Infection Evaluation of a New Point-of-Care Oral Anti-HCV Test for Screening of Hepatitis C Virus Infection Performance Evaluation of the Hologic Aptima HCV Quant Dx Assay for Detection of HCV RNA from Dried Blood Spots Evaluation of a New, Rapid Test for Detecting HCV Infection, Suitable for Use with Blood or Oral Fluid Diagnostic Performance and Usability of the Genedrive®HCV ID Kit in Two Decentralized Settings in Cameroon and Georgia Loop-Mediated Isothermal Amplification of DNA LAMP-on-a-Chip: Revising Microfluidic Platforms for Loop-Mediated DNA Amplification A Handheld Digital Microfluidic Device for Loop-Mediated Isothermal Amplification (LAMP) A Self-Powered Rapid Loading Microfluidic Chip for Vector-Borne Viruses Detection Using RT-LAMP Selection of healthy sperm based on positive rheotaxis using a microfluidic device Microfluidic-RT-LAMP Chip for the Point-of-Care Detection of Emerging and Re-Emerging Enteric Coronaviruses in Swine Hand-Held Device for Rapid Single Tube Detection of Hepatitis-C Virus Rapid Visual Detection of Hepatitis C Virus Using a Reverse Transcription Loop-Mediated Isothermal Amplification Assay Detection of HCV Gene by Reverse Transcription Loop-Mediated Isothermal Amplification Method-Abstract-Europe PMC Detection of HCV Genotypes 1b and 2a by a Reverse Transcription Loop-Mediated Isothermal Amplification Assay A Field-Deployable Mobile Molecular Diagnostic System for Malaria at the Point of Need. Lab A Chip Sample Introduction Interface for On-Chip Nucleic Acid-Based Analysis of Helicobacter Pylori from Stool Samples. Lab A Chip Hands-free smartphonebased diagnostics for simultaneous detection of Zika, Chikungunya, and Dengue at point-of-care Instrument-Free Point-of-Care Molecular Detection of Zika Virus Loop-Mediated Isothermal Amplification (LAMP) Shield for Arduino DNA Detection Production of Infectious Hepatitis C Virus in Tissue Culture from a Cloned Viral Genome Development of a Flow-Free Automated Colorimetric Detection Assay Integrated with Smartphone for Zika NS1 Development of a Multiplex Fully Automated Assay for Rapid Quantification of CD4+ T Cells from Whole Blood Development of a Flow-Free Magnetic Actuation Platform for an Automated Microfluidic ELISA Hepatitis C Genotype 4: What We Know and What We Don't yet Know Molecular Cloning of the Human Hepatitis C Virus Genome from Japanese Patients with Non-A, Non-B Hepatitis Molecular Cloning of the Human Hepatitis C Virus Genome from Japanese Patients with Non-A, Non-B Hepatitis Clinical Significance of Hepatitis C Virus Genotypes Functionally conserved architecture of hepatitis C virus RNA genomes The Hepatitis C Virus 5 UTR Genomic Region Remains Highly Conserved Under HAART: A 4-to 8-Year Longitudinal Study from HCV/HIV Co-Infected Patients SYBR Green I: Fluorescence Properties and Interaction with DNA Assessing Molecular Point-of-Care Testing and Dried Blood Spot for Hepatitis C Virus Screening in People Who Inject Drugs. Open Forum Infect Rapid Point-of-Care (POC) Testing for Hepatitis C Antibodies in a Very High Prevalence Setting: Persons Injecting Drugs in Development of a Microfluidic Device for CD4 T Cell Isolation and Automated Enumeration from Whole Blood Estimating Maximum Plausible Conserved Synteny between Orthologous Genomes of a Species Pair Compared in an Exhaustive Search-Space Dengue Detection: Advances in Diagnostic Tools from Conventional Technology to Point of Care Bioassay and Dose Measurement in UV Disinfection The Inactivation of Enzymes by Ultraviolet Light. V. The Disruption of Specific Cystines in Ribonuclease Optimization and Troubleshooting in PCR. Cold Spring Harb Impact of Metal Ions on PCR Inhibition and RT-PCR Efficiency Simultaneous Detection and Differentiation of Clinically Relevant Relapsing Fever Borrelia with Semimultiplex Real-Time PCR Evaluation of Dried Blood Spot Samples for Screening of Hepatitis C and Human Immunodeficiency Virus in a Detection of Residual Hepatitis C Virus RNA by Transcription-Mediated Amplification in Patients with Complete Virologic Response According to Polymerase Chain Reaction-Based Assays Surface Plasmon Resonance Technology Combined with Rolling Circle Amplification for Detection of Hepatitis C Virus Current Molecular Methods for the Detection of Hepatitis C Virus in High Risk Group Population: A Systematic Review Impact of Nucleic Acid Extraction Platforms on Hepatitis Virus Genome Detection Developments in the HCV Screening Technologies Based on the Detection of Antigens and Antibodies Validation of a Solid-Phase Electrochemical Array for Genotyping Hepatitis C Virus Ag(I)-Coordinated Hairpin DNA for Homogenous Electronic Monitoring of Hepatitis C Virus Accompanying Isothermal Cycling Signal Amplification Strategy Advanced "Lab-on-a-Chip" to Detect Viruses-Current Challenges and Future Perspectives Development and Clinical Validation of the Genedrive Point-of-Care Test for Qualitative Detection of Hepatitis C Virus Electronic and electrochemical viral detection for point-of-care use: A systematic review Performance of the OraQuick HCV Rapid Antibody Test for Screening Exposed Patients in a Hepatitis C Outbreak Investigation Rapid, and Sensitive Integrated Protein Microarray for Simultaneous Detection of Multiple Antigens and Antibodies of Five Human Hepatitis Viruses Performance of Premarket Rapid Hepatitis C Virus Antibody Assays in 4 National Human Immunodeficiency Virus Behavioral Surveillance System Sites Management of Hepatitis C Virus Infection in Hemodialysis Patients Magnetic Particle-Based Immunoassay for the Detection of Hepatitis C Virus Core Antigen in Human Serum or Plasma Biosensor for Detection of Hepatitis A and C Viruses in Serum Approaches for Simplified HCV Diagnostic Algorithms We would like to thank Amy Makler for reviewing the manuscript and providing valuable comments and suggestions. We acknowledge research support from the National Institute of Health (NIH), award-R61AI154643, and the National Science Foundation (NSF) CAREER, award-1942487. The authors declare no conflict of interest.