key: cord-0030126-uqflzi8b authors: Kaushik, Saurabh; Selvanathan, Prabhakaran; Soni, Gautam Vivek title: Customized low-cost high-throughput amplifier for electro-fluidic detection of cell volume changes in point-of-care applications date: 2022-04-20 journal: PLoS One DOI: 10.1371/journal.pone.0267207 sha: 997c788de22c5ec1ea1159a18ca155b15192c759 doc_id: 30126 cord_uid: uqflzi8b Physical parameters of the pathogenic cells, like its volume, shape, and stiffness, are important biomarkers for diseases, chemical changes within the cell, and overall cell health. The response of pathogenic bacteria and viruses to different chemical disinfectants is studied widely. Some of the routinely employed techniques to measure these changes require elaborate and expensive equipment which limits any study to a non-mobile research lab facility. Recently, we showed a micropore-based electro-fluidic technique to have great promise in measuring subtle changes in cell volumes at high throughput and resolution. This method, however, requires commercial amplifiers, which makes this technique expensive and incompatible for in-field use. In this paper, we develop a home-built amplifier to make this technique in-field compatible and apply it to measure changes in bacterial volumes upon exposure to alcohol. First, we introduce our low-cost and portable transimpedance amplifier and characterize the maximum range, absolute error percentage, and RMS noise of the amplifier in the measured current signal, along with the amplifier’s bandwidth, and compared these characteristics with the commercial amplifiers. Using our home-built amplifier, we demonstrate a high throughput detection of ~1300 cells/second and resolve cell diameter changes down to 1 μm. Finally, we demonstrate measurement of cell volume changes in E. coli bacteria when exposed to ethanol (5% v/v), which is otherwise difficult to measure via imaging techniques. Our low-cost amplifier (~100-fold lower than commercial alternatives) is battery-run, completely portable for point-of-care applications, and the electro-fluidic devices are currently being tested for in-field applications. Diagnosis of a disease is the first critical step for finding a cure for a patient [1] [2] [3] [4] . Lack of infield diagnostic facilities or diagnosis tools has caused numerous casualties in past [5] [6] [7] . The diagnostic industry includes X-ray, flow cytometry, ultrasound, coagulation analyzers, CTscan, MRI, cell counters, micro sedimentation centrifuges, platelet aggregometer, enzyme assay kits, and many associated consumables [8] [9] [10] [11] [12] [13] [14] [15] [16] [17] [18] [19] [20] [21] [22] . The urban population has access to these facilities via multi-specialist hospitals, door-to-door ambulance services, private clinics, and even online medical assistance, whereas in many rural areas around the globe, the nearest medical help or essential medical equipment are miles away [23] [24] [25] [26] . Diseases like Tuberculosis, Tetanus, Cholera, Anthrax, Pneumonia, etc., are caused by bacterial infection, which has been fatal for centuries [27] [28] [29] [30] [31] [32] [33] [34] [35] [36] . Certain alcohol-based disinfectants are used to kill bacterial cells and can help prevent infections [37, 38] . Disinfectants cause denaturation, and the bacterial cells lose their structural integrity by the breakdown of membrane proteins [39] [40] [41] [42] . There are pathogens that resist certain disinfectants [43] [44] [45] [46] , and hence a quantitative study of alcohol-based physiological changes can lead to a better understanding of how these cells evolve to develop such resistance. In this work, we present our electro-fluidic device to measure alcohol-dependent changes in bacterial cells. This device makes a high resolution and high throughput electrical measurement which directly corresponds to changes in cell volume. We start with introducing a lowcost portable amplifier (referred to as "lab-amplifier") customized with a microfluidic platform which is easy to build, plug and play in use, portable to be used in the field, and about a 100-fold lower in cost than the existing commercial amplifiers. Accurate electrical readout of cell volume changes is recorded with the lab amplifier with high resolution as the cells translocate through the microfluidic device (micropore). The calibration of the micropore device to estimate the volume is based on resistive pulse technique [47] [48] [49] [50] [51] [52] [53] [54] [55] [56] . The lab amplifier is characterized for the maximum range, absolute error, and RMS noise in the measured current signals. We have two series of lab amplifiers (L1 and L10 series), and both are characterized to have different bandwidths with different current gains. Lab amplifier with higher bandwidth (L1 series) works better for higher throughput but at the cost of slightly higher noise than the L10 series. L1 series amplifiers are more suitable for larger cells with large resistive pulse signals [48] , whereas the L10 series amplifiers are more suitable to measure subtle changes in cell volumes with higher resolution. All the electrical characterization of the lab amplifier is compared with two commercial amplifiers under identical experimental conditions. Using model cells, we demonstrate resolution and high throughput detection of up to~1300 model cells/sec using our lab amplifier. Finally, we apply our device to quantitatively measure volumetric changes in cells (E. Coli bacteria) caused due to mild ethanol exposure in the suspension buffer. These subtle changes in cell physiology were not detectable using typical fluorescence or bright field microscopy imaging but were resolved successfully using our electro-fluidic device with lab-amplifier-based detection. There have been previous reports of making customized amplifiers for resistive pulse sensing, however, those designs are not suitable for measurement in cellular changes as they are aimed at molecular detection insteads [57] [58] [59] [60] [61] [62] [63] . Given the custom design of our lab amplifiers, we have made our portable measurement system lowcost, high throughput, and demonstratively aimed towards measuring changes in cell sizes. We foresee its multiple applications in hospitals and in-field rural settings. Our custom lab amplifier is designed with an inverting mode operational amplifier (Op-amp) (IC AD820) which is powered by two 9V DC batteries (+V cc and -V cc ) at IC pins 7 and 4, respectively. The feedback resistor (R F ) and capacitor (C F ) are connected across pins 2 and 6 of the Op-amp in parallel. We use a data acquisition card (National Instruments, NI myDAQ) to apply a constant DC input voltage (V in ) across the load resistor (R P ) in a virtual ground configuration. It is important to note that the input voltage (V in ) can also be applied from a third DC battery with a voltage regulator circuit. The analog signal across the input (V in and ground) and output (pin 6 and ground) terminals were recorded by the DAQ system using a custom-written LabVIEW code. All ground cables were connected to the Aluminum box to keep the electrical noise low. The schematic of the above-mentioned electrical circuit is shown in Fig 1A. Two BNCs for connecting the load resistor (R P ), a DB9 connector for sending and receiving the signal from DAQ, an On/Off DPST toggle switch connected to the batteries to power the amplifier, the circuit board, and a power LED are shown in the interior of the lab amplifier in Fig 1B. Simulations for these amplifiers were done using Tina-TI simulation tool. During cell experiments, the load resistor R P is replaced by the micropore device. In Fig 1C, we show two lab amplifiers (L10 and L1 series) enclosed in an aluminum box, with the feedback resistor, bandwidth, and maximum measurable current labeled. A microfluidic device and a one rupee coin are shown in the image for size references. The description and cost of all the electronics parts used in making lab amplifiers are mentioned in S1 Table in S1 File. Borosilicate glass capillaries (Part # B100-50-10, Sutter Instruments) were pulled using a micropipette puller (P-2000, Shutter Instrument) and then polished using a flame polisher (MF-900, Micro Forge, Narishige) to fabricate the desired micropores (electro-fluidic device). The detailed information on the fabrication of the micropores (electro-fluidic device), sample preparation, and electro-fluidic measurements are presented in our previous work [48] . The schematic and brief details on how the micropore devices are prepared is included in the supplementary information file (see S2 Fig in S1 File). The two commercial amplifiers used in this work as a standard comparison are Dagan Chem Clamp and AM Systems Model 2400 (See S2 Table in S1 File for detail). Latex beads of different diameters (see text) were used as model cells for testing the signal-to-noise, throughput, and resolution. Events in the conductance traces are detected using atleast 1.5σ (1.5 times the standard deviation of baseline noise) thresholding from the baseline. DH5α strain of E. coli bacteria were grown overnight in Luria broth at 37˚C, and 180 rpm overnight. Cells were washed with 1X-PBS (137 mM NaCl, 4.3 mM Na 2 HPO 4, 2.7 mM KCl , and 1.47 mM KH 2 PO 4 at pH 7.4) buffer before its cell volume measurements. Bacterial cells were stained with FM4-64 (Cat # T13320 Invitrogen) dye to a final concentration of 10 μM and imaged using Andor iXon DU-885K-CS0 camera (Oxford Instruments) and fluorescent lamp illumination on an IX-73 Olympus microscope with 100X objective. For testing the effect of alcohol on the cell volumes, 1000 μl of E. coli bacterial culture was incubated with 50 μl of ethanol (EtOH) (5% v/v) for 30 minutes before measurement. Current range. To measure the current range of our L1 series (gain = 0.001 mV/pA, R F = 1 MΩ, C F = 4.7 pF) and L10 series (gain = 0.01 mV/pA, R F = 10 MΩ, C F = 1.5 pF) amplifiers, we measure their I-V characteristics for different R P load resistors. The I-V characteristics of L1A, L1B and D001-1 amplifiers (all with R F = 1 MΩ) for R P values of 50 kΩ, 100 kΩ, 500 kΩ, and 1 MΩ electrical resistors are shown in the Fig 1D. The inset shows the I-V characteristics of L10A and L10B compared with commercial amplifiers D01-10 and AM10 (R F = 10 MΩ) for R P values of 0.5, 1, 5, and 10 MΩ electrical resistors. The current across R P = 50 kΩ resistor with L1A, L1B, and D001-1 amplifiers saturates at ±10 μA, and for R P = 500 kΩ the L10A, L10B, D01-10, and AM10 amplifiers saturate at ± 1000 nA. The current range is limited by the saturation voltage of the Op-amps in the amplifiers. The values of the current ranges for all the amplifiers are listed in S2 Table in S1 File. Absolute Error and Root Mean Square (RMS) Noise in Current: Current was measured across load resistors of values 100 kΩ, 500 kΩ, 1 MΩ, 5 MΩ, 10 MΩ, 50 MΩ and 500 MΩ at voltages ± 200 mV, ± 400 mV, ± 600 mV, ± 800 mV, and ± 1000 mV. The absolute error percentages, as defined by Eq 1, in the current was calculated at all the measured voltages (Note: the current values in the saturation region (beyond measurable current range) were neglected for all the amplifiers): Here, I Th is the theoretically expected current, and I Meas is the experimentally measured current for a known load resistor R P . Fig 2A shows a semi-log plot of the absolute percentage error vs. the theoretically expected current values for different lab and commercial amplifiers. The dotted regions in the plot show the electrical resistor used as the load. We note that the current measured by our lab amplifiers has an error of less than 6% and does as good as (or better than) the commercial amplifiers. The load resistance of our microfluidic devices used for cell volume measurements is in the range of~1 MΩ, where we find the absolute percentage errors to be less than 1%. We next recorded the current at ± 300 mV for electrical resistors 500 kΩ and 1 MΩ (our experimental range) and measured the RMS noise in current at 1 kHz filter frequency. As seen in Fig 2B and 2C , in the experimental range of load resistances (R P ), our lab amplifiers show RMS noise of < 200 pA (L1 series) and < 80 pA (L10 series). The RMS noise values at other filter frequencies are shown in S3 and S4 Tables in S1 File. We note that the RMS noise values of our lab-amplifiers (comparable to the commercial amplifiers) and the large signal-to-noise (see later in Fig 4A) in the cell measurements demonstrates suitability of our devices for such measurements. Bandwidth of lab amplifier. We next calibrated the frequency response of our lab amplifiers. The low-pass cutoff frequency decides the time response of the amplifier, which in turn decides the maximum throughput of cells that can be measured per second. The Gain (dB)-Frequency (Hz) response curves of different lab amplifiers (with R P = 500 kΩ) is shown in Fig 3, where the black horizontal dotted line shows the -3dB decrease in the gain value. For this measurement, a data acquisition card (DAQ) with a maximum sampling rate of 2 MHz was used to apply a clean sine signal of 150 mV amplitude (V in ) of different frequencies and then recorded the output sine signal (V out ). The Gain (dB) was measured as 20 , for each frequency. The blue (circle-L1A, square-L1B) and green (circle-L10A, square-L10B) markers are experimental data, whereas the solid blue (L1) and green (L10) lines are simulation results of the lab amplifier frequency response. The y-axis is scaled so that maximum gain appears at 0 dB. The cutoff frequencies were estimated from the graph at the intersection point of the -3dB line (horizontal black dotted line) and the response curve, and the values are listed in S2 (1), is plotted against the expected current for different amplifiers. Note that the current was measured at both positive and negative voltages to get an absolute mean current value and errorbars. Blue and green markers are for lab-amplifiers and red, yellow and black are for commercial amplifiers. b and c show RMS noise and the error bar values (at 1 kHz) in the current, measured at ± 300 mV for R P = 500 kΩ (blue) and 1 MΩ (grey), by different amplifiers with feedback resistor (R F ) values 1 MΩ (b) and 10 MΩ (c). A 600 ms current series was recorded at 100k sample/sec, 10 such sets at ± 300 mV were used for estimation of RMS noise and respective errorbars. Online version in colour. https://doi.org/10.1371/journal.pone.0267207.g002 Table in S1 File. We show that L1-series amplifiers have a relatively higher bandwidth of~30 kHz, whereas the L10 series amplifiers are of~10 kHz bandwidth. Detection of model cells using lab-amplifiers and electro-fluidic devices. We next show cell volume detection capabilities using model cells measured with our electro-fluidic device and lab-amplifiers. Our devices make electrical measurements of cell volumes, with single-cell resolution, as individual cells translocate through the micropore under an applied flow and electrical potential [48] . Fig 4A shows 1 .5 sec long time series of translocation events caused by 4.98 μm beads translocating through a 6.8 μm micropore device as measured using the L1A (blue), L10A (green) and D001-1 (red) amplifiers. Schematic of a typical micropore (inset (right)) and a measurement device (inset (left)) is shown in Fig 4B. Translocation of the model cells was maintained by a 500 nL/min constant fluid flow and an applied potential of 300 mV. Ions in the 1X-PBS buffer move across the unobstructed micropore resulting in an open pore conductance. As the model cells translocate through the micropore, they block the pore conductance (ΔG (nS)) for the duration of translocation (dwell time, Δt (ms)). An electrical conductance drop (ΔG) signals the translocation of a single cell and is directly proportional to the cell volume [48, 51, 52, 54] . In each dataset, we collect electrical conductance blockage events for 500 cells or more that translocate through the pore to measure population average cell volumes and changes in it, if any. Fig 4B shows the population average of ΔG histograms. We show high signal-to-noise measurements on the same device and sample using the L1 and L10 lab amplifiers and compare it with the D001-1 commercial amplifier. The identical and overlapping histograms in Fig 4B show numerically identical ΔG values for the model cell population when measured using any amplifier. We demonstrate the effect of the bandwidth of the amplifier on the measured signal by translocating the 4.98 μm latex beads through the same 6.8 μm micropore at different fluid flow rates. At larger flow rates, the event rate of cells translocating through the device increases dramatically (see Fig 4C inset) . The ΔG values of model cells measured at different flow rates are shown in Fig 4C. We show that the amplifiers maintain the correct ΔG values up to certain throughput, after which, although the cells are detected successfully, the measured ΔG values drops. This is due to the low pass filtering of the amplifiers. We show in Fig 4C that the L1 amplifier (30 kHz bandwidth) maintains the measured ΔG values up to much higher flow rates (750 nL/min of fluid flow), whereas for other amplifiers (L10 and D00-1, both with 10 kHz bandwidth) ΔG values drop after 500 nL/min of fluid flow. The choice of amplifier bandwidth is important to accurately quantify the translocation dynamics. For the translocation of particles and cells used in our study, 10 kHz bandwidth was found sufficiently high so as not to affect the ΔG conductance drop values (see S3 Fig in S1 File) . Hence, the amplifier and the measurements were optimized to be recorded by 10 kHz bandwidth so that the signal doesn't experience any distortion at the fluid flow rates used for experiments in this paper. Our amplifiers can be used in two different modes. The sensitivity mode (see Fig 4B and later in Fig 5) where the translocation speeds are more controlled and the current drop amplitudes correspond to cell volumes and the high-throughput mode where the cells are detected at a very high speed. We demonstrate the high-throughput detection of model cells in Fig 4D. We use the 6.8 μm micropore device to translocate 4.06 μm model cells at a fluid flow of 50 μL/min and demonstrate a detection event rate of~1300 particles per second. In Fig 4D, we show a conductance trace of~3.5 seconds (using an L10A amplifier), showing such a high detection throughput. The inset of Fig 4D shows a 10 Resolving and quantifying mixed sample population. We next demonstrate the resolution of our measurement electronics by resolving model cells that differ in diameter by 1 μm. In Fig 5A, we show the conductance blockade events corresponding to model cells of diameters 4.06 μm (blue), 4.98 μm (red), and a mixture of both 4.06 and 4.98 μm (black) translocating through a 7.6 μm micropore device. A comparison of the translocation time (Δt) and change in conductance (ΔG) along with the respective lognormal (inset) and Gaussian fits for the same samples are shown in Fig 5B respectively . Although the two different model cells are not resolvable in the Δt histogram, they are very well resolved in the ΔG histograms (red and blue histograms). In the mixed sample experiment, the ΔG histogram shows two clear wellseparated populations (Fig 5B, black histogram) corresponding to individual populations of 4.06 μm and 4.98 μm model cells. In Fig 5C, we also show the ΔG-Δt scatter plot that also clearly shows the presence of two distinct cell size populations. We further quantify the mixed sample data by comparing the number-ratio of events in each population histogram to the number-ratio in which the mixed sample was prepared. Since there is a random chance of any type of cell to translocate through the pore, the two ratios must be equal. The two populations were isolated by fitting Gaussian peaks (with no yoffset) to the histograms (Fig 5B) and estimating number of events in each peak. The quantitative data for the mixed sample is shown in Table 1 (average of three independent experiments). The first column of each set is the ΔG value and the spread in the histogram, and the second column is the number of events corresponding to the respective latex beads. The last column shows the sample concentration used in the experiment from the bead manufacturer's numbers. The 4.06 and 4.98 μm beads were mixed in a known ratio of 2.5:1, the last row of the Table 1 shows the event ratio measured from the translocation experiment. The average of the event ratio from three sets gives us (2.49 ± 0.08): 1, which is remarkably close to the ratio in which the beads were mixed. Thus, our devices, along with our custom amplifiers, may be used in the field for testing multiple types of cells that differ in size and concentrations. In Fig 5D, we demonstrate the resolution capabilities of our system by detecting changes in particle volumes as low as 0.6 femtoliter. The bar plot in this figure shows comparison of measured ΔG values of 24.1 ± 0.7, 26.7 ± 0.3 and 33.9 ± 0.9 nS respectively, for beads of diameter 1.90 ± 0.04, 2.00 ± 0.04, and 2.1 ± 0.3 μm translocating through a 3.0 μm micropore device. In our previous work, we have shown similar contrast between particles using commercial amplifiers [48] , Fig 5D demonstrates that in terms of resolutions, our custom made lab-amplifier compares very well to the commercial amplifiers. Effect of alcohol on bacterial cell volume. Finally, we apply our measurement system with the lab amplifier (L10 series) to quantitatively understand the effect of ethanol on bacterial cell physiology. The FM4-64 stained fluorescent images of the cells suspended in 1X PBS buffer, exposed with and without ethanol, are shown in Fig 6A and 6B , respectively. Particle analyzer plugin of ImageJ software was used post thresholding to estimate the projected area for the bacteria in the fluorescent images. The histogram in Fig 6C shows the projected area estimated from~400 bacteria cells each with and without alcohol exposure. The Gaussian fit to the histograms estimates the values as 5.09 ± 0.14 μm 2 and 4.87 ± 0.12 μm 2 for native and ethanol exposed state of bacteria, respectively. The changes in the cell's projected area are hard to quantify when measured using imaging techniques. The electrical measurement data, on the other hand, shows excellent contrast in the cell volumes of these two cell populations. The ΔG histogram in Fig 6D shows the change in signal for bacterial translocation through a 2.1 (see S1 Fig in S1 File for image) μm micropore device. The ΔG value from the Gaussian fits are 11.4 ± 4.3 (Blue-Native) and 8.7 ± 3.2 nS (Red-5% EtOH exposed). Relative change in the ΔG values directly corresponds to the relative change in cell volume [48] using the following equation: Here, the ΔG Control is the measured value for the native cells, and ΔG Sample is the measured value for cells exposed to 5% ethanol. The bar plot in the inset shows the relative change in volume when the cells are treated with ethanol for 30 minutes, with respect to their native state. We note a 12.6% reduction (average of 4 experiments) in bacterial cell volume upon exposure to just 5% ethanol solution. It is important to note that all the pulse measurements are made relative to the baseline and the reduction in bacterial size is solely due to the shrinkage of cells in the presence of ethanol and not due to the decrease in absolute conductivity of the solution as shown in S4A Fig in S1 File. In order to demonstrate this, we translocate 4.0 μm beads through a 6.1 μm diameter micropore (See S4B Fig in S1 File) and, 1.9 and 2.1 μm beads through a micropore of 3.0 μm diameter (See S4C and S4D Fig in S1 File) in suspension In this work, we present an in-field point-of-care compatible system which is cost-effective, portable, and has simple-to-build electronics. This system consists of electrical measurements using a microfluidic device and a lab-made amplifier (L1 and L10 series). The lab amplifier is characterized for the maximum current range, absolute error, RMS noise in the measured current signals, and bandwidth. All the electrical characterization demonstrate that the lab amplifiers perform comparable to the commercially available amplifiers. Using our lab amplifiers, we showed high throughput detection of up to 1300 cells/sec, demonstrating its potential as a cell counter device. We demonstrate the high resolution of our system by measuring cell size differences down to 100 nm. Estimation of absolute volumes of the translocating particles from the measured ΔG values, require detailed modeling of particle and pore's internal geometry. However, the relative change in cell volume, upon exposure to ethanol, is straight forwardly estimated provided the particle and pore shape remains the same. We finally presented a real-world measurement example by quantitatively measuring small physiological changes in the bacterial cells upon mild alcohol exposure. We report a 12.6% decrease in bacterial volume upon ethanol exposure of 5% v/v. The quantitative knowledge of cell shrinkage upon mild alcohol exposure is a vital step in understanding the adaptive behaviour of bacterial cells which they use to maintain their cellular integrity against certain alcohol-based disinfectants. The measurement of changes in bacterial volume is presented here as proof of concept for sensitive detection of volume-change in cells, using our lab-amplifiers. The simple design of our lab amplifier, its low cost, portability, high throughput, and resolution makes it a promising device for large-scale population screening applications in hospital and in-field rural areas. We envisage possible application of this system for red blood cell-based population screening for disease that directly affect cell size, such as, malaria and sickle cell anemia. REASSURED diagnostics to inform disease control strategies, strengthen health systems and improve patient outcomes The importance of diagnostic test parameters in the interpretation of clinical test findings: The Prone Hip Extension Test as an example The science of clinical practice: Disease diagnosis or patient prognosis? Evidence about "what is likely to happen" should shape clinical practice The Value of Diagnostic Information in Personalised Healthcare: A Comprehensive Concept to Facilitate Bringing This Technology into Healthcare Systems Factors associated with COVID-19-related death using OpenSAFELY Mortality due to low-quality health systems in the universal health coverage era: a systematic analysis of amenable deaths in 137 countries Rate of diagnostic errors and serious misdiagnosis-related harms for major vascular events, infections, and cancers: toward a national incidence estimate using the "Big Three Multiparameter flow cytometry quantification of bone marrow plasma cells at diagnosis provides more prognostic information than morphological assessment in myeloma patients Optical and Mechanical Clot Detection Methodologies: A Comparison Study for Routine Coagulation Testing Platelet Aggregation Testing on a Routine Coagulation Analyzer: A Method Comparison Study Platelet aggregometry testing: Molecular mechanisms, techniques and clinical implications Separation of blood microsamples by exploiting sedimentation at the microscale Technology Advancements in Blood Coagulation Measurements for Point-of-Care Diagnostic Testing Trends in Use of Medical Imaging in US Health Care Systems Use of Flow Cytometry in Clinical Practice A case of partial dedicator of cytokinesis 8 deficiency with altered effector phenotype and impaired CD8+ and natural killer cell cytotoxicity Automated cell count in body fluids: a review Resonance Imaging in the Differentiation of Tumors of the Liver Beyond ultrasound: CT and MRI of ectopic pregnancy Multicolor Flow Cytometry for the Diagnosis of Primary Immunodeficiency Diseases Comparison of commercial ELISA assays for quantification of corticosterone in serum Access to specialty healthcare in urban versus rural US populations: A systematic literature review Rural and urban differences in treatment status among children with surgical conditions in Uganda A comparison of evidence-based medicine practices between primary care physicians in rural and urban primary care settings in Malaysia: a qualitative study A century of tuberculosis Anthrax: A disease of biowarfare and public health importance Cholera epidemics of the past offer new insights into an old enemy Deaths from bacterial pneumonia during 1918-19 influenza pandemic Historical Epidemiology of the Second Cholera Pandemic: Relevance to Present Day Disease Dynamics Introduction: Anthrax History Mortality from tetanus between 1990 and 2015: findings from the global burden of disease study 2015 Natural history of tuberculosis: Duration and fatality of untreated pulmonary tuberculosis in HIV negative patients: A systematic review Tetanus: A review of the literature Proceedings of the Royal Society of Medicine! ection of tbe lbtitorp of fIebictne The History of the 1832 Cholera Epidemic in Yorkshire Efficacy and effectiveness of alcohol in the disinfection of semi-critical materials: A systematic review Alcohols as surface disinfectants in healthcare settings Role of alcohols in growth, lipid composition, and membrane fluidity of yeasts, bacteria, and archaea Mechanism of lysis of Escherichia coli by ethanol and other chaotropic agents Effect of Ethanol on Differential Protein Production and Expression of Potential Virulence Functions in the Opportunistic Pathogen Acinetobacter baumannii Characterization of the E. coli proteome and its modifications during growth and ethanol stres Increasing tolerance of hospital Enterococcus faecium to handwash alcohols Anti-bacterial efficacy of alcoholic hand rubs in the kenyan market Antibiotics and bacterial resistance in the 21st century Né rome S. Bacteria which are highly resistant to antibiotics are not resistant hydro-alcoholic products Detection of nucleosomal substructures using solid-state nanopores Measurement of Alcohol-Dependent Physiological Changes in Red Blood Cells Using Resistive Pulse Sensing Synchronous optical and electrical detection of biomolecules traversing through solid-state nanopores Fabrication of Low Noise Borosilicate Glass Nanopores for Single Molecule Sensing Fingerprinting branches on supercoiled plasmid DNA using quartz nanocapillaries Counting and sizing of submicron particles by the resistive pulse technique Controlling protein translocation through nanopores with bio-inspired fluid walls Electrical Sizing of Particles in Suspensions: III. Rigid Spheroids and Red Blood Cells Detection of local protein structures along DNA using solid-state nanopores Detecting DNA folding with nanocapillaries A Course of Hands-On Nanopore Experiments for Undergraduates: Single-Molecule Detection with Portable Electrochemical Instruments A cartridge based Point-of-Care device for complete blood count Construction, calibration, and validation of a simple patchclamp amplifier for physiology education A low noise amplifier system for nanopore-based single molecule analysis Ultra-low noise measurements of nanopore-based single molecular detection We thank Prof. V. A. Raghunathan's lab for the microforge facility and RRI machining workshop for fluid cell fabrication. We would like to acknowledge initial discussions with Mr. Koushik Sreenivasa and other constructive discussions with Mr. R. Somashekhar and Mr. H. N. Nagraja from RRI's EEG group. We acknowledge Mrs. Divya Shet for her help in bacterial work. Data curation: Saurabh Kaushik, Prabhakaran Selvanathan.