key: cord-0029226-osc85gd8 authors: Sprowls, Mark; Victor, Shaun; Mora, Sabrina Jimena; Osorio, Oscar; Pyznar, Gabriel; Destaillats, Hugo; Wheatley-Guy, Courtney; Johnson, Bruce; Kulick, Doina; Forzani, Erica title: A Smart System for the Contactless Measurement of Energy Expenditure date: 2022-02-10 journal: Sensors (Basel) DOI: 10.3390/s22041355 sha: e8b20d289fddd93572beff6471336030c1e48466 doc_id: 29226 cord_uid: osc85gd8 Energy Expenditure (EE) (kcal/day), a key element to guide obesity treatment, is measured from CO(2) production, VCO(2) (mL/min), and/or O(2) consumption, VO(2) (mL/min). Current technologies are limited due to the requirement of wearable facial accessories. A novel system, the Smart Pad, which measures EE via VCO(2) from a room’s ambient CO(2) concentration transients was evaluated. Resting EE (REE) and exercise VCO(2) measurements were recorded using Smart Pad and a reference instrument to study measurement duration’s influence on accuracy. The Smart Pad displayed 90% accuracy (±1 SD) for 14–19 min of REE measurement and for 4.8–7.0 min of exercise, using known room’s air exchange rate. Additionally, the Smart Pad was validated measuring subjects with a wide range of body mass indexes (BMI = 18.8 to 31.4 kg/m(2)), successfully validating the system accuracy across REE’s measures of ~1200 to ~3000 kcal/day. Furthermore, high correlation between subjects’ VCO(2) and λ for CO(2) accumulation was observed (p < 0.00001, R = 0.785) in a 14.0 m(3) sized room. This finding led to development of a new model for REE measurement from ambient CO(2) without λ calibration using a reference instrument. The model correlated in nearly 100% agreement with reference instrument measures (y = 1.06x, R = 0.937) using an independent dataset (N = 56). Obesity is an epidemic affecting 39.8% of the United States adult population as per latest data [1] . With the decreasing life expectancy by an estimated 5-20 years as a result of obesity-associated co-morbidities [2] [3] [4] , obesity is clearly a major burden on society. The primary clinical intervention in obesity treatment is referred to as a deficit diet [5] , where a medical professional outlines a dietary plan with the objective of reducing daily caloric intake from food to less than the total daily energy expenditure (TEE) of that person. Energy expenditure (EE) is the rate of energy consumption of a person at given time [6] , with resting energy expenditure (REE) being a common conceptual form of EE [7] with great focus in obesity treatment. REE can be measured accurately using a principle referred to as indirect calorimetry [8, 9] , a technique based on the measurement of rates of O 2 consumption and CO 2 production by the well-established Weir formula [10] . Indirect calorimetry is used as the basis for all current U.S. Food and Drug Administration (FDA)-cleared/approved medical devices for REE measurement. These devices are widely available, yet measurement of EE is not commonly adopted by clinical practices. Instead, predictive equations for REE calculation are common practice, which is a potential cause for limited success in treatment of clinical obesity. These predictive equations have low accuracy [11] . For instance, for a given individual, REE differences of 900 kcal/day can be found between the calculated REE from Mifflin St. Jeor equation [12] and measured REE [13] . Many studies have observed similar findings [14] [15] [16] [17] . Besides application to obesity medicine, personalized metabolic assessment has invigorated a passion for personal health monitoring and "calorie tracking" across the world, evidenced by worldwide Smartwatch adoption, although this technology has notably low accuracy [18] . As such, a convenient and accurate EE measurement technology could substantially benefit both obesity patients and individuals interested in their own personal health and physical performance. Given known inaccuracies of predictive equations for REE, it is reasonable to question why EE measurement tools are not used extensively in clinical practice. An argument for this could be that the EE measurement using current technologies is a technique affected by widely recognized issues [19] that are a result of a fundamental design flaw of these devices: they typically require a subject to breathe into a mouthpiece, mask, or some object worn on the subject's face. Wearing a facial accessory results in errors, including elevated VCO 2 and VO 2 from hyperventilation while wearing an object for breath gas collection on the face, or, alternatively, mistakenly breathing into an EE measurement medical device that is not fitted in an airtight manner to the subject's face. These tools have the added operational issues related to wearing an object on the subject's face, including discouraged repeated measurement due to discomfort, which is important given high day-to-day variability in REE, 11.8% [20] . Other considerations are related to sterilization of facial accessories, an aspect which increases both device cost (purchase of additional single-use accessories) and procedural time requirements (sterilization of facial accessories), and inherently increases the risk of transmission of various pathogens. In addition to usability concerns, many current FDA-cleared/approved technologies for REE measurement have reproducibility errors worse than ±10% (68% CI) by multiple definitions of reproducibility [21] . Alternatively, there is a technique for EE measurement from ambient sensors, in what is referred to as an "indirect calorimeter room" [22] [23] [24] . Current methodologies for development of an indirect calorimeter room are remiss in the sense that they are not portable and require extensive installation time; they are therefore unsuitable for widespread clinical adoption. There are only seventeen of these calorimeter rooms in the entire U.S., and an additional twenty-five in the rest of the world [24] . In this work, we present a novel EE measurement technique offering clear advantages over modern indirect calorimeters. The system presented here, the Smart Pad, resolves many of these issues, since the Smart Pad is not wearable and therefore fundamentally reduces analytical accuracy errors from wearable facial accessories and eliminates the operational issues mentioned above. In comparison with indirect calorimetry rooms, the Smart Pad, utilizing a mathematical model for VCO 2 from a single sensor in the bulk of the measurement environment [25] , can be transported from room to room. The Smart Pad's room conditioning involves the connection of the inlet and/or outlet ventilation fan/s to the Smart Pad actuator (see details below). A previous work [26] reports on the analytical accuracy of the Smart Pad system in one simple configuration within a clinical setting and performing measurements on 20 subjects using a reference instrument for EE measurement in parallel with the Smart Pad. This manuscript builds upon that work by optimizing the operating parameter of EE measurement resolution while also maximizing measurement accuracy. Eventually, the system could be implemented synergistically with other recent advancements in the field of medical device engineering, where there is a strong trend towards non-invasive [27] [28] [29] [30] or totally contactless [31] [32] [33] [34] [35] [36] [37] measurement. The assessment of a room's air exchange rate, referred to as "λ" in this work, is a critical challenge for measurement of REE from ambient CO 2 accumulation patterns. λ is a common consideration in hospitals striving to minimize pathogen transmission [38] , schools [39] where ventilation can influence learning performance [40] , and office buildings [41] to prevent incidences of sick building syndrome [42] . It can be approximated via quantitative assessment of CO 2 concentration decay in environments with no CO 2 sources (i.e., no occupants), and a well-established CO 2 model has been studied extensively [43] [44] [45] . Further, the CO 2 decay modelling procedure has shown good agreement with reference methods that use a tracer gas [46, 47] , suggesting that the technique is accurate in unoccupied environments. Recently, λ values obtained from decay modelling have been applied to CO 2 accumulation data for determination of human CO 2 production rate (VCO 2 ) in cognitively stressed subjects [48] . Our previous study [26] accounted for λ by calibration with a reference instrument's VCO 2 measurements during CO 2 accumulation in an 8 m 3 sized room and the technique was shown to be accurate for assessment of an individual's VCO 2 . Here, we implemented this EE assessment technique by utilizing a fully developed and automated Smart Pad system while investigating two methods of assessing λ values: Method A, named "λ Acc ", with λ values assessed from CO 2 accumulation data collected in presence of a subject of known VCO 2 ; and Method B, named "λ o ", with λ values assessed from CO 2 decay data collected in the absence of the subject. Both methods were carried out in a medium-sized test room. Additionally, a strong correlation between λ Acc and VCO 2 was observed, which led to the development of a novel model for VCO 2 and REE assessment within an occupied room that does not rely directly on λ calibration using a reference instrument [26] or λ assessment from CO 2 decay rate [46] . The model shows good agreement for REE assessment when compared to independent measurements from a reference instrument. The following equations were used, derived in a previous work [25] . Equation (1) describes a physical environment where CO 2 (concentration units expressed in part-permillion, ppm) accumulates due to a human's occupancy; air exchange to the surrounding environment occurs and is characterized by λ (h −1 ), specifically referred to as λ Acc when it is applied to CO 2 accumulation data. k gen is the CO 2 generation rate in the environment (ppm h −1 ) due to occupant CO 2 production from breath: Previous findings [26] suggested that k gen was correlated to the actual k gen via an environmental correction factor (CF env ) of 1.143, an empirical constant determined to correct for mean error bias originating from any source not accounted for in the "ideal model" used in Equation (1), e.g., imperfect mixing, CO 2 sensor response time, etc., and is used in the model as follows: k gen = k gen CF env (2) Once λ Acc is estimated, k gen is then assessed for the subject and used to calculate k gen from Equation (2) . k gen ' is then used in the following equation to determine the value of VCO 2 (mL/min): where VCO 2 is the subject's volumetric production of CO 2 (mL/min), V Room is the room's volume (mL), and CF STPD (dimensionless) is a correction factor used to convert measured ambient VCO 2 to standard temperature, pressure, and dry conditions (STPD). The correction factor was calculated as follows: where P bar (mmHg) is barometric pressure, P H20 is partial pressure of H 2 O (mmHg), and T represents temperature within the environment ( • C). EE (kcal/day) of the subject was calculated using a simplified version of the Weir formula [10] (Equation (12) of the original In addition, the λ values assessed from CO 2 decay data collected in absence of the subject are referred to as λ 0 (hour −1 ) and were estimated from a standard first-order decay equation as follows: where λ o is the air exchange rate (hour −1 ) in the absence of a subject, and distinctively different from λ Acc, the air exchange rate (hour −1 ) in the presence of a subject (Equation (1)). The Smart Pad system is comprised of three components. (A) The measurement system, a wireless device with an embedded sensor array located in the rear of a seat pad ( Figure 1A ). This array includes sensors for carbon dioxide (CO 2 ) and environmental conditions (humidity, temperature, and barometric pressure) necessary for REE measurement from Equations (1)- (7) . (B) The actuator system, which controls the room's ventilation ( Figure 1C ) using logic defined by user-selected maximum and minimum threshold levels set for CO 2 using an iPhone (iOS) application (see Figure 2A ). (C) The mobile (iOS) application, which hosts algorithms with defined functions of sensing, actuation, information processing, and data analytics ( Figure 1B ). The mobile application is under development and currently has functional wireless and automated sensing and actuation abilities. where Pbar (mmHg) is barometric pressure, PH20 is partial pressure of H2O (mmHg), and T represents temperature within the environment (°C). EE (kcal/day) of the subject was calculated using a simplified version of the Weir formula [10] (Equation (12) of the original publication) that assumes a constant respiratory quotient (RQ) of 0.85 (see more details in Supplementary Materials, Section S1): (kcal/day) = 3.941 * 2 + 1.106 * 2 (5) In addition, the λ values assessed from CO2 decay data collected in absence of the subject are referred to as λ0 (hour −1 ) and were estimated from a standard first-order decay equation as follows: where λo is the air exchange rate (hour −1 ) in the absence of a subject, and distinctively different from λAcc, the air exchange rate (hour −1 ) in the presence of a subject (Equation (1)). The Smart Pad system is comprised of three components. (A) The measurement system, a wireless device with an embedded sensor array located in the rear of a seat pad ( Figure 1A ). This array includes sensors for carbon dioxide (CO2) and environmental conditions (humidity, temperature, and barometric pressure) necessary for REE measurement from Equations (1)- (7) . (B) The actuator system, which controls the room's ventilation ( Figure 1C ) using logic defined by user-selected maximum and minimum threshold levels set for CO2 using an iPhone (iOS) application (see Figure 2A ). (C) The mobile (iOS) application, which hosts algorithms with defined functions of sensing, actuation, information processing, and data analytics ( Figure 1B ). The mobile application is under development and currently has functional wireless and automated sensing and actuation abilities. accumulation data, using known λAcc from first set of CO2 accumulation data. In the first CO2 accumulation cycle, the CO2 accumulation fitting is performed using Equation (1) (Eq (1)) and the known kgen from the reference instrument. In the second CO2 accumulation cycle, λAcc from the first accumulation cycle is used to determine kgen using Equation (1) . Kgen is then used to calculate EE via a combination of Equations (3)-(6). The measurement system's CO2 measurements are based on non-dispersive infrared (NDIR) absorption by a commercial CO2 sensor (Telaire TM 7001D CO2 monitor, Test Equipment Depot, MA). A brand-new CO2 measurement device was used for the entire study and factory calibration was validated using a CO2 reference gas and outdoor air, which is well known to contain approximately 415 ppm CO2 in most areas of the world [49] , to the rated accuracy of the CO2 measurement device (±50 ppm). An inlet tube with a diameter of approximately 0.3 cm was connected to the sample inlet port of the CO2 sensor, sealed with parafilm, and then allowed to rest approximately 5 cm out of the seatpad. The CO2 sensor was connected to the sensing module, which also includes embedded temperature, barometric pressure, and relative humidity sensors. The iOS application was programmed to collect data with a resolution of 1 measurement/5 s for all measurements. It is worth noting that the Smart Pad CO2 sensor does not collect real-time respiration CO2 patterns. The sensor is meant to collect changes in environment CO2, which are reflective of the metabolic rate of the individual present in the environment. The Smart Pad was installed on a chair in a rectangular room with a measured volume of 14.0 m 3 , which accounts for the size of objects, and approximate dimensions of 2.7 m × 1.4 m × 3.7 m (see layout and object locations in Figure 2B ). The room was well sealed using tape to cover interstitial space between ceiling tiles and cardboard panels to block the room's built-in HVAC (heating, ventilation, and air conditioning) system. A special Retrotec TM (https://retrotec.com/, accessed 23 December 2021) blower door was installed on the doorway fixed with an outlet fan and also a transparent plastic window allowing for observation of test subjects. The λo values for this room ranged from 1.5 to 3 h −1 , which is in agreement with similar findings for administrative offices [50] , suggesting that sig- (C) Subject during test procedure for parallel REE measurement with Smart Pad and reference instrument. (D) Graphical demonstration of Smart Pad's k gen assessment with two sets of CO 2 accumulation data, using known λ Acc from first set of CO 2 accumulation data. In the first CO 2 accumulation cycle, the CO 2 accumulation fitting is performed using Equation (1) and the known k gen from the reference instrument. In the second CO 2 accumulation cycle, λ Acc from the first accumulation cycle is used to determine k gen using Equation (1). K gen is then used to calculate EE via a combination of Equations (3)-(6). The measurement system's CO 2 measurements are based on non-dispersive infrared (NDIR) absorption by a commercial CO 2 sensor (Telaire TM 7001D CO 2 monitor, Test Equipment Depot, Woburn, MA, USA). A brand-new CO 2 measurement device was used for the entire study and factory calibration was validated using a CO 2 reference gas and outdoor air, which is well known to contain approximately 415 ppm CO 2 in most areas of the world [49] , to the rated accuracy of the CO 2 measurement device (±50 ppm). An inlet tube with a diameter of approximately 0.3 cm was connected to the sample inlet port of the CO 2 sensor, sealed with parafilm, and then allowed to rest approximately 5 cm out of the seatpad. The CO 2 sensor was connected to the sensing module, which also includes embedded temperature, barometric pressure, and relative humidity sensors. The iOS application was programmed to collect data with a resolution of 1 measurement/5 s for all measurements. It is worth noting that the Smart Pad CO 2 sensor does not collect real-time respiration CO 2 patterns. The sensor is meant to collect changes in environment CO 2 , which are reflective of the metabolic rate of the individual present in the environment. The Smart Pad was installed on a chair in a rectangular room with a measured volume of 14.0 m 3 , which accounts for the size of objects, and approximate dimensions of 2.7 m × 1.4 m × 3.7 m (see layout and object locations in Figure 2B ). The room was well sealed using tape to cover interstitial space between ceiling tiles and cardboard panels to block the room's built-in HVAC (heating, ventilation, and air conditioning) system. A special Retrotec TM (https://retrotec.com/, accessed 23 December 2021) blower door was installed on the doorway fixed with an outlet fan and also a transparent plastic window allowing for observation of test subjects. The λ o values for this room ranged from 1.5 to 3 h −1 , which is in agreement with similar findings for administrative offices [50] , suggesting that significant air exchange will still occur even with the room mostly sealed from the surrounding environment (Retrotec TM door's opening for an outlet fan was likely the primary air exchange medium). Additionally, two small (approximately 40 cm diameter) mixing fans were pointed towards each other at opposite corners of the measurement environment to support mixing of CO 2 in the room. The fans were left on continuously during both CO 2 accumulation and decay periods. Supplementary Materials, Section S2 provides more details on system design, standard operation, and testing environment. The system prevents CO 2 exposure to levels above a maximum threshold CO 2 through ventilation actuation, given that exposure to a high concentration of CO 2 can compromise a person's cognitive functions [51, 52] . Figure 2A illustrates the concept: when the CO 2 levels reach a level of 700 ppm, the system actuates the ventilation, which causes the CO 2 level to decrease. Once the minimum threshold CO 2 is reached, the SmartPad turns off the room air ventilation. This enables the accumulation of CO 2 in the room and causes significant changes in CO 2 in the room environment, which allows for the assessment of VCO 2 through data analysis. This cycle repeats until the subject departs from the environment and many sequential CO 2 accumulation cycles are often recorded. This design makes the system uniquely innovative: collecting CO 2 data with each cycle, which can be analyzed and provides access to multiple EE data points for high-temporal resolution monitoring while also preventing exposure to unhealthy CO 2 . Experiments were performed using three distinct methodologies. (1) Sequential REE assessments (see Sections 2.3 and 2.5): subject remains seated in room while REE measurement cycles are recorded every 10-30 min. (2) Sequential Exercise VCO 2 assessments (see Section 2.4): subject remains in room while exercise VCO 2 measurements are recorded every 5-10 min. (3) Sequential CO 2 accumulation and decay assessments (see Section 2.6): subject remains seated in room for a singular REE assessment and then immediately departs upon CO 2 accumulation to a particular threshold limit. Subject 1 performed the first phase of this research work. The physical characteristics of the subject are shown in Table 1 . Subject 1 performed a total of 113 measurements with both systems: Smart Pad and a reference instrument for VCO 2 and REE measurement, the MGC Ultima CPX TM Indirect Calorimeter, an FDA 510(k) cleared system [53] , considered one of the best breath-by-breath indirect calorimeters. The basic configuration of the experiment is shown in Figure 2C , with Subject 1 performing parallel reference instrument and Smart Pad measures by wearing the MGC Ultima CPX TM mask (connected to the metabolic cart) while occupying a room with the Smart Pad (sensing module) installed on a chair. The MGC Ultima CPX TM system measures both VCO 2 and VO 2 by direct physical measurement of flow rate (via pitot tube), O 2 concentration (via galvanic cell potential), and CO 2 concentration (via NDIR absorption) [53] . The 113 REE measurements were split between 5 CO 2 threshold ranges of 500-600 ppm, 500-625 ppm, 500-650 ppm, 500-675 ppm, and 500-700 ppm. Multiple CO 2 accumulation cycles were measured sequentially. Λ Acc was assessed by recording a first CO 2 accumulation curve and fitting the data with reference k gen value assessed from the reference instrument's VCO 2 as shown in Figure 2D . In parallel, the subject's VCO 2 measures were assessed from the Smart Pad system (shown in Figures 1 and 2C) . The subject fasted for 8 h prior to each REE assessment to minimize the wellstudied thermic effect of food, which is known to increase EE by a significant margin [54] . The subject remained seated at a desk while measurement cycles of CO 2 accumulation data were recorded for REE assessment. See Supplementary Materials, Section S3 for more details on the experimental protocol. Subject 1 performed N = 46 simultaneous Smart Pad and MGC Ultima CPX TM exercise VCO 2 assessments using a stationary exercise bike. The subject often biked at a high intensity; however, no formal measure of exertion was recorded. The bike was facing the Smart Pad unit and was approximately 1 m away from the seat pad (see Figure 2B ). See Supplementary Materials, Section S4 for more details on the experimental protocol. Test subjects were recruited with approval from Arizona State University's institutional review board (STUDY00006547) after giving consent to participate in the study and were asked to perform six sequential MGC Ultima CPX TM and Smart Pad measurements at the CO 2 threshold range of 500-650 ppm (one for λ Acc assessment from reference instrument VCO 2 and five comparative measurements). This was done to determine whether subject BMI, VCO 2 , or EE had any significant effect on Smart Pad accuracy or λ Acc . The physical characteristics of each subject are shown in Supplementary Materials (Table S1 ). In the scientific literature [43] [44] [45] 48] , there is a general assumption that λ 0 from CO 2 decay following subject departure is in agreement with λ Acc . In this study, a reference instrument for VCO 2 measurement was used to assess λ Acc with high accuracy, as shown in Figure 2D , and λ Acc was compared with λ 0 assessed from CO 2 decay data, collected following subject departure from the room. A total of twenty-six (N = 26) CO 2 accumulation periods and subsequent CO 2 decay periods were collected with the actuator system turned off. A minimum of five (N = 5) assessments were performed for each of the following CO 2 concentration ranges (ppm): 500-600, 500-625, 500-650, 500-675, and 500-700. For the 500-650 ppm data set, CO 2 accumulation and subsequent CO 2 decay was collected using N = 1 measurement from each subject #2-5 and N = 2 measurements from Subject 1. Fitting of CO 2 accumulation data or CO 2 decay data was performed with Equation (1) and Equation (7), respectively, using the least square error method. In general, errors between the Smart Pad and reference method were assessed as: Error (%) = [(REE Smart Pad -REE MGC CPX ) / REE MGC CPX ] × 100%. For biking assessments, VCO 2 was used to calculate errors in place of REE given the reasoning discussed below. To study the effect of the Smart Pad's upper and lower CO 2 threshold range settings on REE measurement accuracy, a single subject performed a total of 113 measurements as follows: 24 measurements for a range between 500-600 ppm, 22 measurements for a range between 500-625 ppm, 32 measurements for a range between 500-650 ppm, 19 measurements for a range between 500-675 ppm, and 16 measurements for a range between 500-700 ppm. Comparative measurements between the REE values obtained from Smart Pad and the reference instrument are summarized in Figure 3 . Figure 3A shows relatively good correlation (y = 1.02x) between parallel REE measurements recorded by the Smart Pad and the reference instrument, with an averaged mean bias of 3.3% across all measurement conditions suggesting that Equation (1) is robust overall across all five CO2 threshold ranges examined within this work. However, Figure 3A shows relatively good correlation (y = 1.02x) between parallel REE measurements recorded by the Smart Pad and the reference instrument, with an averaged mean bias of 3.3% across all measurement conditions suggesting that Equation (1) is robust overall across all five CO 2 threshold ranges examined within this work. However, the relatively low Pearson's R of 0.559 indicates that further optimization of the measurement error is required. Figure 3B details the effect of Smart Pad threshold range on system precision and accuracy, showing clear differences in measurement characteristics across differing threshold ranges. The 500-650 ppm CO 2 concentration range was concluded to be the most accurate and precise operating range, with performance characteristics of −1.0% ± 10.5% (SD) for the subject across N = 32 total REE measurements. Figure 3C shows the effect of differing CO 2 threshold ranges on measurement duration. Clearly, measurement duration increased consistently with increasing CO 2 threshold range. From Figure 3B ,C, one may reasonably conclude that the 500-650 ppm CO 2 threshold range provides optimal precision and accuracy, with REE measurement accuracy of −1.0% ± 10.5% for a 14-19 min measurement duration. Repeated measures can improve the Smart Pad accuracy significantly (see Supplementary Materials, Section S7). In addition, the Smart Pad's maximum error after λ Acc calibration using a reference instrument (11.5% for ±1 SD) is comparable to errors reported for well-respected FDA-cleared/approved medical devices for REE measurement [21] ; for example: Korr ReeVue TM (11.1% for ±1 SD) and MGC CPX Ultima TM (12.2% error for ±1 SD). Figure 4A-C shows Smart Pad VCO 2 measurement performance for exercise (biking on a fixed bike) in a similar analysis format presented in Figure 3 for REE data. Figure 4A shows high correlation between parallel exercise VCO2 measurements from the Smart Pad and reference instrument, with an average mean bias of −0.8% across all measurement conditions, suggesting that Equation (1) is robust across all measurement conditions. Figure 4B shows how the Smart Pad app's CO2 threshold range affects accuracy for VCO2 measurement, with the 500-675 ppm threshold range showing optimal measurement with errors of −0.8% ± 9.3% (SD). Further, Figure 4C shows the effect of CO2 threshold range on measurement duration. A less pronounced increase in measurement duration with increasing size of CO2 threshold range was observed compared to the set of Figure 4A shows high correlation between parallel exercise VCO 2 measurements from the Smart Pad and reference instrument, with an average mean bias of −0.8% across all measurement conditions, suggesting that Equation (1) is robust across all measurement con- ditions. Figure 4B shows how the Smart Pad app's CO 2 threshold range affects accuracy for VCO 2 measurement, with the 500-675 ppm threshold range showing optimal measurement with errors of −0.8% ± 9.3% (SD). Further, Figure 4C shows the effect of CO 2 threshold range on measurement duration. A less pronounced increase in measurement duration with increasing size of CO 2 threshold range was observed compared to the set of experiments for REE. One possible explanation is that the effort level on the bike was not a control variable of the study. Measurement duration for the lowest error range (500-675 ppm) corresponded to only 4.8-7 min of contactless VCO 2 measurement. EE calculation via the Weir formula in Equation (5) assumes an RQ value of 0.85. However, the RQ value may vary significantly during the course of exercise [55, 56] , and may potentially be a source of error for Smart Pad under exercising conditions. For that reason, exercise EE accuracy is not reported. Five subjects were studied across a wide range of BMIs and REEs using the reference instrument and Smart Pad at the optimized CO 2 threshold range of 500-650 ppm. The procedure shown in Figure 2D was followed, which uses a reference instrument to calibrate for λ Acc . Each subject performed a minimum of five independent measurements, and a total of 52 REE measurements were performed on all the subjects. Figure 4A -C summarizes the effects of BMI on measurement accuracy, the effects of subject REE on measurement accuracy, and the effects of subject REE on Smart Pad measurement duration. Figure 5A shows Smart Pad's REE measurement mean error as a function of BMI. Mean error biases across all five subjects were within ±10%, an analytically accepted error for FDA-cleared devices in the U.S. [21] . Figure 5A 's slope was determined to be statistically insignificant (p = 0.267), suggesting that the Smart Pad is accurate over a wide range of BMI values. In addition, Figure 5B shows a Bland-Altman plot of Smart Pad error % for REE measurement, and confirms low mean bias (µ = −2.6%) between the MGC Ultima CPX TM and Smart Pad and an insignificant (p = 0.088) effect of REE on system accuracy across all N = 52 measurements performed on the N = 5 subjects. These findings suggest that the Smart Pad is accurate for a wide range of metabolic rates and body types. of BMI values. In addition, Figure 5B shows a Bland-Altman plot of Smart Pad error % for REE measurement, and confirms low mean bias (µ = −2.6%) between the MGC Ultima CPX TM and Smart Pad and an insignificant (p = 0.088) effect of REE on system accuracy across all N = 52 measurements performed on the N = 5 subjects. These findings suggest that the Smart Pad is accurate for a wide range of metabolic rates and body types. Figure 5C shows a plot between subject REE and Smart Pad's measurement duration. The general trend in the plot indicates decreasing measurement duration with increasing REE. This is an intuitive result based on system design, as, while holding upper CO2 level constant, elevated CO2 generation rates will reach the upper CO2 level more quickly than would a reduced CO2 generation rate. In general, from regressions shown in Figure 5A ,B, we conclude that neither subject BMI nor subject REE have statistically significant effects on Smart Pad accuracy when using a reference instrument to calibrate for λAcc. A room's air exchange rate (λ) can be determined safely within indoor environments with the estimation of subject VCO2 during CO2 accumulation periods (λAcc) [43, 57] , or recording CO2 decay following exit of human occupants from an environment (λo) [43- Figure 5 . Smart Pad system performance for REE assessment at optimal CO 2 range in 5 subjects: (A) Smart Pad mean error % ± SD for REE measurement categorically grouped using BMI and ranging from just above the underweight cutoff (18.8 kg/m 2 ) to obese (31.4 kg/m 2 ); (B) Bland-Altman plot for REE measurement accuracy across various subject REEs; (C) Effect of subject REE on Smart Pad measurement duration at 500-650 ppm CO 2 threshold range. Figure 5C shows a plot between subject REE and Smart Pad's measurement duration. The general trend in the plot indicates decreasing measurement duration with increasing REE. This is an intuitive result based on system design, as, while holding upper CO 2 level constant, elevated CO 2 generation rates will reach the upper CO 2 level more quickly than would a reduced CO 2 generation rate. In general, from regressions shown in Figure 5A ,B, we conclude that neither subject BMI nor subject REE have statistically significant effects on Smart Pad accuracy when using a reference instrument to calibrate for λ Acc . A room's air exchange rate (λ) can be determined safely within indoor environments with the estimation of subject VCO 2 during CO 2 accumulation periods (λ Acc ) [43, 57] , or recording CO 2 decay following exit of human occupants from an environment (λ o ) [43] [44] [45] 48] . Previous reports of λ assessed from CO 2 accumulation data (λ Acc ) [43, 57] were based on predictive equations for REE (e.g., [12, 58] ). However, it is widely reported that assessment of REE based on REE predictive equations is often inaccurate [11, 13] . In this work we focus on assessing λ Acc with high accuracy so that we can formally study the correlation of λ Acc vs. λ o and λ Acc vs. VCO 2 (k gen ). In order to do this, we build upon the previous work [26] in which λ Acc was assessed using a reference instrument. Here, we build upon that work by using a reference instrument to provide a high-accuracy value of λ Acc (using k gen derived from an FDA 510(k)-cleared device's VCO 2 measure) and λ 0 from CO 2 decay in the exact same environment following the subject's departure. It is worthy to mention that in general, there is an assumption within the scientific community that these parameters (i.e., λ Acc and λ 0 ) are in agreement. Figure 6A shows the results of λ Acc vs. λ 0 from this study. Λ Acc values were obtained with a procedure described in Figure 2D and λ 0 values were assessed from decay data as shown in Figure 6B , right branch of CO 2 data. The data shown in Figure 6A were obtained from Subjects 1-5's data for a total of N = 26 measurements, including sequential CO 2 accumulation and decay at the five threshold ranges (500-600, 500-625, 500-650, 500-675, and 500-700 ppm) for a minimum of N = 5 sequential measurements for each CO 2 threshold range. It can be observed that in our study there was no agreement between λ Acc vs. λ 0 (R = −0.228). Additionally, there was relatively low variance in λ 0 assessed from CO 2 decay, with sample means of 1.9 ± 0.2 h −1 (CV = 11%), 1.9 ± 0.2 h −1 (CV = 11%), and 1.8 ± 0.2 h −1 (CV = 11%) for 15, 30, and 45 min of CO 2 decay, respectively. Figure 6C shows λ Acc assessed from Equation (1) using Smart Pad in comparison with VCO 2 measured from the reference instrument. A high correlation can be observed between those parameters (λ Acc and VCO 2 ) for resting state (R = 0.785, p < 0.00001) and a moderate correlation for exercise state (R = 0.565, p < 0.01). A possible explanation for these correlations is that the λ Acc is strongly influenced by changes in mass transport factor due to the subject's presence in the room, leading to a rise in air exchange across openings connecting the room with adjacent spaces. Such changes may be due to increased convective transport associated with body movements and air density gradients due to human heat dissipation [59] [60] [61] . These effects are maximized during biking tests, where the subject's heat flux is elevated from exercise and there is additional air convection due to pedaling. In order to accurately measure contactless EE in free-living conditions, it is important to assess k gen from the CO 2 concentration profiles described by Equation (1) . To do so, it is essential to assess the environment's air exchange rate during occupancy, λ Acc . Figure 6A demonstrates that λ 0 is not an adequate predictor of λ Acc . Using the strong observed correlation between λ Acc and VCO 2 as shown in Figure 6C , this issue may be resolved by simultaneous assessment of λ Acc and VCO 2 without immediate prior calibration using a reference instrument. This is only possible given that VCO 2 correlated so strongly with λ Acc in the test environment. As such, λ Acc can be ascertained via regression utilizing the relationship found in this work ( Figure 6C ), shown as follows: In this study, α = 0.0107 (h −1 min −1 mL), and one might expect the term to vary in differing measurement environments. By introducing Equation (8) in Equation (1), we obtain Equation (9) as follows: where variable meanings and dimensions are the same as for Equation (1), except for a new term, β (ppm −1 ), which is as follows: β considers unit conversion (mL min −1 to ppm h −1 ) and the factor from Equation (8) that represents "the change in λ Acc resulting from a 1 unit increase in "k gen ". (9)) with Smart Pad system; (D) Sample data analysis using Equation (9) , € Correlation between REE measured using Equation (9) (Eq (9)) and REE measured using MGC Ultima CPX TM for N = 56 total measurements on N = 5 subjects; (E) Correlation for the same dataset € (F) but presenting mean REE ± SD for each subject's measurements. Figure 6C shows λAcc assessed from Equation (1) using Smart Pad in comparison with VCO2 measured from the reference instrument. A high correlation can be observed between those parameters (λAcc and VCO2) for resting state (R = 0.785, p < 0.00001) and a moderate correlation for exercise state (R = 0.565, p < 0.01). A possible explanation for these correlations is that the λAcc is strongly influenced by changes in mass transport factor due to the subject's presence in the room, leading to a rise in air exchange across openings Figure 6 . Results of λ Acc correlative study. (A) λ Acc vs. λ 0 scatter plot showing little to no correlation (R = −0.228) for N = 26 sequential CO 2 accumulation/decay analyses; (B) Sample data analysis for sequential CO 2 decay/accumulation experiments, (C) Correlation between VCO 2 measured from the MGC Ultima CPX TM and λ Acc as assessed from Equation (1) with Smart Pad system; (D) Sample data analysis using Equation (9) , € Correlation between REE measured using Equation (9) and REE measured using MGC Ultima CPX TM for N = 56 total measurements on N = 5 subjects; (E) Correlation for the same dataset € (F) but presenting mean REE ± SD for each subject's measurements. An independent dataset of N = 56 measurements from N = 5 subjects was used to validate accuracy for the determination of energy expenditure from these new equations. This dataset contained no overlap in individual measurements with regards to the model development dataset. The results of the application of Equation (9) to independent data are shown in Figure 6E ,F. Figure 6D shows an example of fitting Equation (9) to a CO 2 accumulation profile for a subject of our study. Figure 6E shows the correlation plot of the Smart Pad's REE measurement from Equation (9) with the REE measurement from the MGC Ultima CPX TM . The results show a low mean bias relative to the reference instrument: slope = 1.02 versus 1.00 for ideal measurement with R~0.8. Taking the mean of multiple repeated REE measures ( Figure 6F ) increases correlation with the reference instrument and a Pearson's R = 0.937, which is a logical consequence of standard error [62] and the general ability of measurement devices to mitigate imprecision using repeated measures. This new model represents an important step forward from the progress made in [26] towards development of a contactless IoT device for REE measurement, given that the promising application of Equation (9) requires no direct calibration for λ Acc using a VCO 2 measurement reference instrument. The data presented in Figure 6E is displayed in comparison with current FDA 510(k)-cleared (authorized for prescription use) technologies for metabolic assessment in Table 1 below. A new contactless system named Smart Pad was evaluated. The system displayed promising accuracy characteristics for contactless resting energy expenditure (REE) and exercise CO 2 production rate (VCO 2 ) measurements in a medium size room after calibrating for air exchange rate with a VCO 2 measurement reference instrument method once. The Smart Pad is capable of performing accurate REE measurements in 14-19 min and exercise VCO 2 measurements in 5-7 min after calibrating with the reference instrument once. In this configuration, measurement characteristics were comparable to multiple wearable (i.e., with facial accessories) FDA-cleared devices as reported by another study. Additionally, via nonlinear regression of sequential CO 2 accumulation and decay data in the same environment using a high accuracy FDA-cleared medical device, observed differences in air exchange rate could be rationalized statistically in terms of a strong correlation with VCO 2 . This led to development of a new model for REE assessment from ambient CO 2 , which does not require a direct air exchange rate calibration using a reference instrument method or CO 2 decay rate. The model shows good agreement for REE assessment (y = 1.06x, R = 0.937) when evaluated on a dataset independent from the one used to develop the model and measurements for the same subjects were averaged. The model has been demonstrated in a few subjects as a proof-of-concept, and it is currently being tested in a larger study. So far, the new model was concluded to meet the FDA 510(k) standard for authorization of a medical device for prescription use by comparison with five contact-based indirect calorimeters currently used today in clinical practice, with these comparisons shown in Table 1 . Future work will focus on validation of Equation (9) in a new environment and the development of a contactless physical fitness test. Arizona State University owns intellectual property related to the measurement technology: US Patent Application No. 20210048206. The US Patent Application No. 20210048206 has been licensed to TF Health Co. Dr. Erica Forzani is co-founder of TF Health Corp., Tempe, AZ, USA. Supplementary Materials: The following supporting information can be downloaded at: https: //www.mdpi.com/article/10.3390/s22041355/s1: S1. Smart Pad System Operating Equations and Measurement Technique; Figure S1 : Sample regressions for Smart Pad system; S2. Smart Pad: Physical Characteristics, Design, and Testing Environment; Figure S2 : Graphic illustrating Smart Pad assembly; Figure S3 : Graphic demonstrating how the Retrotec TM door was sealed for every experiment; Figure S4 : The Smart Pad actuator system; Figure S5 : The Smart Pad room's carbon dioxide level distribution; S3. CO 2 Accumulation Range Optimization for REE assessment; S4. CO 2 Accumulation Range Optimization for Exercise assessment; S5. CO 2 Accumulation Range Optimization for REE assessment: Results; S6. Additional information, including Table S1 . Physical Characteristics and Body Type Classification, Figure S6 : Sample raw data for many sequential REE assessments using a CO 2 threshold range of 500-650; Figure S7 : Sample baseline adjusted CO 2 fitting; Table S2 : Figure Funding: This work was supported by the A. J. and Sigismunda Palumbo Charitable Trust and Arizona State University. Neither institution was actively involved in study execution or the planning or publication of this article (other than IRB approval). This work was also supported by NIH (R03 EB027336). The study was conducted in accordance with the Declaration of Helsinki and approved by the Institutional Review of Arizona State University (STUDY00006547) on 8 April 2017. The protocol was annually renewed. Informed Consent Statement: Informed consent was obtained from all subjects involved in the study. Data Availability Statement: Some raw sample data presented in this study are available in Supplemental Information. For more inquiries concerning data, please contact the corresponding author. Prevalence of Obesity among Adults and Youth: United States Body-Mass Index and Mortality among 1.46 Million White Adults Years of Life Lost Due to Obesity Body-mass index and cause-specific mortality in 900 000 adults: Collaborative analyses of 57 prospective studies Update on treatment strategies for obesity Human Energy Expenditure Measurement of energy expenditure The theoretical bases of indirect calorimetry: A review Indirect Calorimetry: A Practical Guide for Clinicians New methods for calculating metabolic rate with special reference to protein metabolism Bias and accuracy of resting metabolic rate equations in non-obese and obese adults A new predictive equation for resting energy expenditure in healthy individuals Comparison of Resting Metabolic Rates: Calculated using predictive equation and measured using Portable Indirect Calorimeter. Glob Analysis of Predictive Equations for Estimating Resting Energy Expenditure in a Large Cohort of Morbidly Obese Patients Accuracy of four resting metabolic rate prediction equations: Effects of sex, body mass index, age, and race/ethnicity Accuracy of Resting Energy Expenditure Predictive Equations in Patients with Cancer Accuracy of predictive equations for the measurement of resting energy expenditure in older subjects Assessment of laboratory and daily energy expenditure estimates from consumer multi-sensor physical activity monitors Indirect calorimetry: Methodological and interpretative problems Within-and between-subject variation in energy expenditure measured by the doubly-labelled water technique: Implications for validating reported dietary energy intake Assessing Validity and Reliability of Resting Metabolic Rate in Six Gas Analysis Systems Comparison of Methods for Achieving 24-Hour Energy Balance in a Whole-Room Indirect Calorimeter Evaluation of a new whole room indirect calorimeter specific for measurement of resting metabolic rate Room Indirect Calorimetry Operating and Reporting Standards (RICORS 1.0): A Guide to Conducting and Reporting Human Whole-Room Calorimeter Studies Assessing metabolic rate and indoor air quality with passive environmental sensors A system for contact free energy expenditure assessment under free-living conditions: Monitoring metabolism for weight loss using carbon dioxide emission A Flexible and Pervasive IoT-Based Healthcare Platform for Physiological and Environmental Parameters Monitoring Wearable sensors: Modalities, challenges, and prospects Soft Wearable Pressure Sensors for Beat-to-Beat Blood Pressure Monitoring Colorimetric Optoelectronic Dynamics Analyzer for Measuring Total Ammonia of Biological Samples Contactless cardiac arrest detection using smart devices An Unobstructive Sensing Method for Indoor Air Quality Optimization and Metabolic Assessment within Vehicles Accurate measurement of the pulse wave delay with imaging photoplethysmography An integrated system of air sampling and simultaneous enrichment for rapid biosensing of airborne coronavirus and influenza virus Vital-sign monitoring and spatial tracking of multiple people using a contactless radar-based sensor Noncontact Monitoring of Blood Oxygen Saturation Using Camera and Dual-Wavelength Imaging System Noncontact Monitoring Breathing Pattern, Exhalation Flow Rate and Pulse Transit Time Air quality in hospital operating rooms Indoor air quality investigation according to age of the school buildings in Korea Ventilation rates in schools and pupils' performance European Indoor Air Quality Audit Project in 56 Office Buildings Sick-building syndrome Review and Extension of CO 2 -Based Methods to Determine Ventilation Rates with Application to School Classrooms Air Stuffiness and Air Exchange Rate in French Schools and Day-Care Centres Lazović, I. Indoor CO 2 measurements in Serbian schools and ventilation rate calculation Simple and Cheap Air Change Rate Measurement Using CO2Concentration Decays CO 2 tracer gas concentration decay method for measuring air change rate Impact of Cognitive Tasks on CO 2 and Isoprene Emissions from Humans Trends in Atmospheric Carbon Dioxide: NOAA ESRL Global Monitoring Division Development and application of an indoor air quality audit to an air-conditioned building in Singapore Direct human health risks of increased atmospheric carbon dioxide Is CO 2 an Indoor Pollutant? Direct Effects of Low-to-Moderate CO 2 Concentrations on Human Decision-Making Performance Ultima System 510(k) Premarket Notification: U.S. Food and Drug Administration The Thermic Effect of Food: A Review Decrease in Respiratory Quotient During Exercise Following L-Carnitine Supplementation Respiratory quotient during exercise Association between substandard classroom ventilation rates and students' academic achievement Standard Guide for Using Indoor Carbon Dioxide Concentrations to Evaluate Indoor Air Quality and Ventilation: American Society for Testing and Materials (ASTM) Experimental Investigation of the Air Exchange Effectiveness of Push-Pull Ventilation Devices Estimating Energy Expenditure Using Heat Flux Measured at a Single Body Site Comparison of Air Exchange Efficiency and Contaminant Removal Effectiveness as IAQ Indices ASHRAE Transactions Standard deviations and standard errors CPXTM 510(k) Premarket Notification (K072323). U.S. Food and Drug Administration Pro 510(k) Premarket Notification (K992214): U.S. Food and Drug Administration SentrySuite Product Line 510(k) Premarket Notification (K133925): U.S. Food and Drug Administration Mobile 510(k) Premarket Notification (K023120): U.S. Food and Drug Administration Indirect Calorimeter, Model#8100 510(k) Premarket Notification The Antoine Equation for Vapor-pressure Data Indirect calorimetry: Technical aspects Fasting Respiratory Quotient as a Predictor of Long-Term Weight Changes in Non-Obese Women Clinical use of the respiratory quotient obtained from indirect calorimetry The authors would like to thank the A. J. and Sigismunda Palumbo Charitable Trust, Arizona State University, ASU's Ira A Fulton School of Engineering, NIH (R03 EB027336) and the test participants. Hugo Destaillats participated in this study as an external consultant, and the Lawrence Berkeley National Laboratory was not involved in this research. has been licensed to TF Health Corp. Erica Forzani is co-founder of TF Health Corp. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.