key: cord-0027558-co16l735 authors: Işın, Deniz; Eksin, Ece; Erdem, Arzum title: Graphene-Oxide and Ionic Liquid Modified Electrodes for Electrochemical Sensing of Breast Cancer 1 Gene date: 2022-02-04 journal: Biosensors (Basel) DOI: 10.3390/bios12020095 sha: e34c92104ac5628a4f824f3816ad8c308b302f12 doc_id: 27558 cord_uid: co16l735 Graphene-oxide and ionic liquid composite-modified pencil graphite electrodes (GO-IL-PGEs) were developed and used as a sensing platform for breast cancer 1 (BRCA1) gene detection. The characterization of GO-IL modified electrodes was executed by scanning electron microscopy (SEM), cyclic voltammetry (CV), and electrochemical impedance spectroscopy (EIS). The nucleic-acid hybridization was monitored by a differential pulse voltammetry (DPV) technique by directly measuring the guanine oxidation signal without using any indicator. The effects of the IL concentration, the probe concentration, and the hybridization time were optimized to the biosensor response. The limit of detection (LOD) was calculated in the concentration range of 2–10 μg/mL for the BRCA1 gene and found to be 1.48 µg/mL. The sensitivity of the sensor was calculated as 1.49 µA mL/µg cm(2). The developed biosensor can effectively discriminate the complementary target sequence in comparison to a three-base-mismatched sequence or the non-complementary one. Breast cancer is one of the most serious types of cancer affecting human health worldwide. The region of human chromosome 17q12-21, which contains the breast-cancer susceptibility gene known as BRCA1, is responsible for approximately 80% of breast-cancer families and ovarian cancer and is also associated with other types of cancer, such as pancreatic and colon cancers. BRCA1 encoded a tumor-suppressor protein that controls the cell cycle, repairs damaged DNA, and maintains genomic stability. Mutations in BRCA1 affect the proteins produced, resulting in an increased risk of breast cancer. Besides BRCA1 mutation, the alteration of the gene-expression level can also affect biological processes and contribute to both sporadic and hereditary breast-tumor progression [1] . Conventional methods such as capillarity electrophoresis [2] , sequencing [3, 4] , and PCR-mediated techniques [5, 6] are available for analyzing cancer-related genes. Even though these methods are sensitive, there are some limitations, such as high costs, time consuming, and complex experimental steps with harmful chemical use and the requirement of highly trained staff. Thus, it is urgent to develop a low-cost, fast, and sensitive detection method for the analysis of cancer-related gene sequences. In this context, nucleic acid biosensors provide a simplified method to detect cancer related nucleic-acid sequences [1, [7] [8] [9] [10] [11] [12] . Recently, the fluorescent method has become popular in the diagnosis of cancer or tumor tissues due to its advantages of sensitivity and specificity. For instance, Liu et al. developed a fluorescent method based on hairpin DNA-templated copper nanoclusters (CuNCs) for the detection of the BRCA1 gene [10] . Although the fluorescent sensor for the BRCA1 gene assay showed good assay performance, fluorescent-based assays also have some limitations, such as limited applications, being labor-intensive (i.e., labelling molecules), and validating tools often being required. On the other hand, surface plasmon resonance biosensors can provide an alternative platform for detection of point mutations. For example, Li et al. presented a method that combines primer extension analysis with SPR to discriminate single-base mutations related to the BRCA1 gene [11] . One potential limitation of the SPR method is that the ligand may not maintain its native configuration upon immobilization on the sensor chip surface, or its orientation may sterically hinder analyte binding. Non-specific binding effects at the sensor surface must also be carefully controlled, and this requires meticulous experimental design for SPR [12] . Electrochemical nucleic acid biosensors are based on the electrochemical behavior of the redox active nucleobases (e.g., guanine or adenine oxidation) or changes in the interfacial properties of the recognition layer or a signal change in the electroactive indicator agent after hybridization, which are the main approaches. Electrochemical biosensors do not suffer the drawback of high sensor setup complexity and cost. Other essential advantages of electrochemical biosensors are their robustness, easy miniaturization, sufficient detection limits, and requirement of low analyte volumes [13] . Graphene oxide (GO) has gained great attention in the field of biosensing [14] . Plenty of works based on incorporation of GO to the sensor surface in the absence or presence of other (nano)materials in a composite form have been reported so far [15] [16] [17] [18] . For instance, Arfin and Rangari developed a ZnO-functionalized graphene oxide (GO)-modified glassy carbon electrode (GO-ZnO/GCE) used for the electrochemical sensory detection of phenol [15] . The cost-effective carbon monoxide sensor was developed on the interdigitated copper electrode based on PPy-GO composite materials by Farea et al. [16] . TiO 2 -reduced graphene oxide composite-modified glassy carbon electrodes (TiO 2 -ErGO-GCE) for the sensitive detection of tartrazine were reported by He et al. [17] . Ionic liquids (ILs) are considered as important alternative "Green" solvents with unique properties. Thus, ILs are attractive for a range of applications [19] . There are many biosensor studies that involve the use of various ILs as a supporting medium or an electrode modifier [20] [21] [22] [23] [24] [25] [26] [27] [28] [29] [30] [31] [32] [33] . To date, numerous biosensors developed by using graphene-IL composites have been applied for detection of different analytes (i.e., small molecules). For example, Kim et al. [34] developed an electrochemical detection method for capsaicin using an IL-doped graphene-titania-nafion composite-modified electrode. Li et al. [35] designed a simple and sensitive electrochemical sensor based on an IL-functionalized graphene-oxide supported gold nanoparticles (GO-IL-AuNPs) coated glassy carbon electrode (GCE) for detection of dopamine. Ma et al. [36] developed a strategy for efficient determination of the sulfadiazine residues in the animal feed by applying the composite-modified electrode fabricated with the rGO and N-octyl-pyridinium-hexafluorophosphate (OPPF6) IL. On the other hand, Sukumaran et al. [37] developed a label-free platform based on an IL-functionalized nitrogen doped graphene (NrGO)-modified GCE for the discrimination of mutations in DNA. A hydrophilic IL was used to trap the DNA targets onto it by electrostatic interaction, and thereby it retained them close to the electrode surface suitable for the electroanalysis. The change in the peak current was explained based on the extent of the hindrance experienced by the redox species [Fe(CN) 6 ] 3− for the electron transfer at various DNA modified electrodes. To the best of our knowledge, a GO-IL composite modified single-use graphite electrode was developed for the first time herein as a sensing platform and applied for detection of breast cancer 1 (BRCA1) gene through the direct measurement of the guanine oxidation signal. In this work, we aimed to develop a sensitive nucleic-acid biosensor related to the BRCA1 gene sequence. Under this goal, graphene-oxide and ionic liquid composite modified pencil graphite electrodes (GO-IL-PGEs) were developed for the first time herein as a single-use biosensing platform. The characterization of GO-IL modified electrodes was carried out with scanning electron microscopy (SEM), cyclic voltammetry (CV), and electrochemical impedance spectroscopy (EIS). Synthetic DNA sequences were used to elevate the experimental conditions as well as to test the selectivity of the biosensor. The Table 1 . The base sequences of all oligonucleotides. Probe (P 1 ) 5 -(CH 2 ) 6 -GGGGGGGGGGGG 3 SH-link probe (P 2 ) 5 -SH-(CH 2 ) 6 -GGGGGGGGGGGG-3 Note that the BRCA1-specific DNA probe has an inosine (I) base instead of a guanine base since it is a synthetic analogue of a guanine base (Table 1 ). Based on this phenomenon, indicator-free detection of BRCA1 can be performed by directly measuring the guanine oxidation signal after full-match hybridization of the probe and the BRCA1 target [27, 38] . The detection protocol is given below and represented in Figure 1 . GO was prepared in dimethyl sulfoxide (DMSO) as 50 µg/mL by following the sonication step during 30 min. The GO concentration was used in our study as 50 µg/mL, which was optimized in our previous work [39] . After sonication, IL was pipetted into the GO solution, and we obtained 7.5% IL in GO-IL solution and sonicated it for 30 min. PGEs were activated by applying +1.40 V for 30 s in ABS (pH 4.80) via DPV. These electrochemically activated PGEs were dipped into 100 µL of GO-IL and kept for 15 min to achieve the surface modification [39] [40] [41] [42] [43] . There was no precipitation in the solution of GO-IL observed during its immobilization onto the PGE surface. Then, GO-IL-PGEs were air-dried for 5 min. The BRCA1-specific DNA probe and the BRCA1 target were mixed and allowed for solution-phase hybridization during 5 min in PBS. Then, GO-IL-modified PGEs were dipped in vials containing BRCA1 probe-target hybrids and allowed for immobilization onto GO-IL-PGE surface during 30 min. Immobilization of hybrids arose by the formation of peptide bonds between 5 -end amino groups of BRCA1 probe-target hybrids and carboxyl groups of GO as reported in our previous study [39] . Due to this bond formation, no or negligible precipitation was expected during the sensing process. After 30 min of the immobilization step, the electrodes were washed with PBS to eliminate non-specific binding. The electrochemical characterization of electrodes was studied by using the CV technique. CVs were carried out by scanning between −0.8 V and +1.4 V with a 25 mV/s step potential and a 100 mV/s scan rate in the redox probe solution of 2.5 mM Fe(CN) 6 3−/4− in 0.1 M KCl. The guanine oxidation signal was measured by the DPV technique using the same conditions in our earlier studies [27, 28, 38] . DPVs were carried out by scanning with a 50 mV pulse amplitude and a 50 mV/s scan rate in ABS (pH, 4.80). The redox probe solution of 2.5 mM Fe(CN) 6 3−/4− in 0.1 M KCl was used for EIS measurements. More information about impedimetric measurements is given in the supplementary information. The microscopic characterizations of PGE, GO-PGE, IL-PGE, and GO-IL-PGEs were recorded by a Quanta 400 FEI field emission scanning electron microscope (Tokyo, Japan) and are given in Figure 2 . There are similarities between PGE (Figure 2a The effect of IL concentration was examined by measuring I a values after modification of GO and IL at different percentages (2.5 % to 10 %) by the CV technique ( Figure 3 ). Additionally, the electrochemical behavior of GO-IL-PGEs was investigated in comparison to PGE or GO-PGE. The corresponding anodic current (I a ) and cathodic current (I c ) with the relative charge Q (mC) and peak potential separation value (∆E p ) and the electroactive surface areas of PGE, GO-PGE, and GOL-IL-PGE are listed in Table 2 . An increase was obtained at I a when IL % was increased till 7.5% (Figure 3e ), after then it decreased. As seen from Figure 3 , the anodic and cathodic peak currents of GO-IL-PGE were higher than the ones measured by PGE or GO-PGE. This confirmed that the GO-IL expedited the electron convey between the electrode interface and the electrolyte. Additionally, the electroactive surface areas of the electrodes were calculated according to the Randles-Sevcik equation [44] , and accordingly, the highest surface area (i.e., 0.314 cm 2 ) was obtained in the case of GO-IL modification. The enhanced sensor response can be attributed to the acceleration of ion transportation from the electrolyte to the electrode surface in the presence of GO-IL in comparison to the one measured with PGE [27] [28] [29] [39] [40] [41] [42] [43] . ∆E p was found to be 151 mV by GO-IL-GE, while it was 173 mV with PGE. Thus, it was a quasi-reversible electrochemical response and convincing electrocatalytic effect of GO-IL modification toward the redox reaction of [Fe(CN) 6 ] 3−/4− . The effect of IL concentration was examined by measuring Ia values after modification of GO and IL at different percentages (2.5 % to 10 %) by the CV technique ( Figure 3 ). Additionally, the electrochemical behavior of GO-IL-PGEs was investigated in comparison to PGE or GO-PGE. The corresponding anodic current (Ia) and cathodic current (Ic) with the relative charge Q (mC) and peak potential separation value (ΔEp) and the electroactive surface areas of PGE, GO-PGE, and GOL-IL-PGE are listed in Table 2 . An in- attributed to the acceleration of ion transportation from the electrolyte to the electrode surface in the presence of GO-IL in comparison to the one measured with PGE [27] [28] [29] [39] [40] [41] [42] [43] . ΔEp was found to be 151 mV by GO-IL-GE, while it was 173 mV with PGE. Thus, it was a quasi-reversible electrochemical response and convincing electrocatalytic effect of GO-IL modification toward the redox reaction of [Fe(CN)6] 3−/4− . Ia ( The electrochemical-impedance spectroscopy technique was used for stepwise characterization of GO-IL-PGEs before and after DNA immobilization. Figure 4 shows the impedimetric data obtained by PGE and GO-PGE and the results by using GO-IL-PGE before/after DNA immobilization. A standard Randles-equivalent circuit (Figure 4 -inset) was used to fit EIS data. The Randles-equivalent circuit considers the resistance from the electrolyte solution (Rs) and the double-layer capacitance (C) due to the accumulation of ions at the electrode surface as well as the charge-transfer resistance of the electrode surface (Rct) and the Warburg component (W) for diffusive proves [45] . Thus, the effect on the impedimetric response was elucidated by means of Rs, Rct, C, and Warburg by fitting The electrochemical-impedance spectroscopy technique was used for stepwise characterization of GO-IL-PGEs before and after DNA immobilization. Figure 4 shows the impedimetric data obtained by PGE and GO-PGE and the results by using GO-IL-PGE before/after DNA immobilization. A standard Randles-equivalent circuit (Figure 4 -inset) was used to fit EIS data. The Randles-equivalent circuit considers the resistance from the electrolyte solution (R s ) and the double-layer capacitance (C) due to the accumulation of ions at the electrode surface as well as the charge-transfer resistance of the electrode surface (R ct ) and the Warburg component (W) for diffusive proves [45] . Thus, the effect on the impedimetric response was elucidated by means of R s , R ct , C, and Warburg by fitting the experimental curves with the Randles-equivalent circuit. Further analysis of circuit components was extracted from the Randles-circuit model and is listed in Table 3 . The average R ct value of PGE was 73.30 ± 10.50 Ohm (RSD %, 14.30 %, n = 3). In the presence of GO modification, a decrease in the R ct value was observed since the GO-modified surface was becoming more conductive in contrast to the unmodified PGE, similarly as reported in the literature [38] [39] [40] 43] . After GO-IL modification onto the electrode surface, there was a 43 % decrease observed in the R ct value as GO-IL composites improved conductivity at the electrode surface by facilitating more electron transfer between the solution and the electrode interface by graphene oxides, similarly as reported in earlier work [30] . the data recorded by the CV and the EIS techniques in our study. Table 3 . EIS response parameters obtained after curve fit of the Nyquist diagrams presented in Figure 4 using Randles circuit. There was an increase obtained in the R ct value due to a repulsive interaction between the anionic redox probe and the negatively charged DNA-immobilized electrode [42, 43] . Additionally, a decrease was observed in the C value after DNA immobilization, which is coherent with the theory that as a biomolecule binds it reduces the surface area by resulting with a 47 % decrease in the capacitance. DNA immobilization onto the electrode surface also creates a shielding effect and lowers the columbic attraction between the electrode surface and the electrolyte [45] . Hence, a lower capacitance was recorded by DNA-immobilized GO-IL-PGE in contrast to bare GO-IL-PGE. By means of IL, an increase in the conductivity of the PGE surface can be obtained because of the increase in the electron transfer between the electrolyte and the electrode [29, 41] . Regarding the results obtained by the electrode (PGE, GO-PGE, and GO-IL-PGE), there was a good agreement between the data recorded by the CV and the EIS techniques in our study. In the presence of fsDNA immobilization on the PGE, the GO-PGE, or the GO-IL-PGE, the R ct values were increased to 619 ± 130 Ohm (RSD %, 21%, n = 3), 570.50 ± 33.20 Ohm (RSD %, 5.80%, n = 3), and 1559.50 ± 193 Ohm (RSD %, 12.40%, n = 3), respectively ( Figure 5 ). The increase in R ct in case of fsDNA immobilization was due to the charge repulsion between the phosphate backbone of DNA and [Fe(CN) 6 ] 3−/4− , which resulted in a reduced ability for electron transfer at the electrode surface, as similarly reported in earlier works [42, 43, 46] . The highest increase in R ct was obtained when fsDNA was immobilized onto the GO-IL-PGE relatively to the PGE or the GO-PGE. Furthermore, the apparent fractional coverage values (θ R IS ) for each of the electrode types were calculated by Janek et al.'s equation [47] . The θ R IS values for the fsDNA-immobilized PGE, GO-PGE, and GO-IL-PGE were found to be 0.881, 0.900, and 0.973, respectively. The highest θ R IS indicated successful sensor surface modification with GO-IL. In the presence of fsDNA immobilization on the PGE, the GO-PGE, or the GO-IL-PGE, the Rct values were increased to 619 ± 130 Ohm (RSD %, 21 %, n = 3), 570.50 ± 33.20 Ohm (RSD %, 5.80 %, n = 3), and 1559.50 ± 193 Ohm (RSD %, 12.40 %, n = 3), respectively ( Figure 5 ). The increase in Rct in case of fsDNA immobilization was due to the charge repulsion between the phosphate backbone of DNA and [Fe(CN)6] 3−/4− , which resulted in a reduced ability for electron transfer at the electrode surface, as similarly reported in earlier works [42, 43, 46] . The highest increase in Rct was obtained when fsDNA was immobilized onto the GO-IL-PGE relatively to the PGE or the GO-PGE. Furthermore, the apparent fractional coverage values ( ) for each of the electrode types were calculated by Janek et al.'s equation [47] . The values for the fsDNA-immobilized PGE, GO-PGE, and GO-IL-PGE were found to be 0.881, 0.900, and 0.973, respectively. The highest indicated successful sensor surface modification with GO-IL. For better attachment of the DNA probe onto the sensor surface, the effect of 5′-end modification of the DNA probe was investigated. DNA probes with different 5′-end modifiers as amino, thiol, or phosphate were used and compared with the unmodified version of the probe ( Figure 6 ). The guanine oxidation signals and RSD % values recorded by the PGE or the GO-IL-PGE are given in Table 4 . The most reproducible and highest guanine signal with the relatively high increment % was obtained in the case of the amino-linked probe (P4) immobilization. Therefore, NH2 linkage was approved for effective probe immobilization onto the GO-IL-PGE surface. For better attachment of the DNA probe onto the sensor surface, the effect of 5 -end modification of the DNA probe was investigated. DNA probes with different 5 -end modifiers as amino, thiol, or phosphate were used and compared with the unmodified version of the probe ( Figure 6 ). The guanine oxidation signals and RSD % values recorded by the PGE or the GO-IL-PGE are given in Table 4 . The most reproducible and highest guanine signal with the relatively high increment % was obtained in the case of the aminolinked probe (P 4 ) immobilization. Therefore, NH 2 linkage was approved for effective probe immobilization onto the GO-IL-PGE surface. The BRCA1 probe concentration was further optimized by testing five probe concentrations (from 2 to 10 µg/mL) (Figure 7) . For this purpose, the solution-phase hybridization between the BRCA1 probe and the BRCA1 target was performed. After hybridization, the highest guanine oxidation signal was measured as 5.88 ± 0.73 µA (RSD %, 12.41 %, n = 3), in the case of the 6 µg/mL BRCA1 probe (Figure 7 ). The hybridization between the probe and the BRCA1 target was performed during 1, 5, 10, and 20 min. The guanine signal was recorded as 5.58 ± 0.39 µ A (RSD %, 7.05 %, n = 3), which was the highest signal when hybridization was performed during 5 min (Figure 8) . Therefore, the optimum hybridization time was chosen as 5 min. The hybridization between the probe and the BRCA1 target was performed during 1, 5, 10, and 20 min. The guanine signal was recorded as 5.58 ± 0.39 µA (RSD %, 7.05%, n = 3), which was the highest signal when hybridization was performed during 5 min (Figure 8) . Therefore, the optimum hybridization time was chosen as 5 min. The effect of the BRCA1 target concentration at the hybridization process was then investigated. Firstly, the hybridization between the 6 µ g/mL BRCA1 probe and different concentrations of the BRCA1 target from 2 to 12 µ g/mL was performed. The signal was increased up to 10 µ g/mL; then, it was fixed (shown in Figure S2 ). Therefore, the 10 µ g/mL BRCA1 target concentration was chosen as the optimum. The limit of detection (LOD) was calculated in the concentration range of 2-10 μg/mL with the equation y = 0.47x + 0.91 and R 2 = 0.98 (shown in Figure 9B ), and the LOD was found to be 1.48 µ g/mL [48] . Additionally, the sensor sensitivity was estimated from the slope of the calibration curve divided by the surface area of the GO-IL-PGE and was found to be 1.49 µ A.mL/µ g.cm 2 . The effect of the BRCA1 target concentration at the hybridization process was then investigated. Firstly, the hybridization between the 6 µg/mL BRCA1 probe and different concentrations of the BRCA1 target from 2 to 12 µg/mL was performed. The signal was increased up to 10 µg/mL; then, it was fixed (shown in Figure S2 ). Therefore, the 10 µg/mL BRCA1 target concentration was chosen as the optimum. The limit of detection (LOD) was calculated in the concentration range of 2-10 µg/mL with the equation y = 0.47x + 0.91 and R 2 = 0.98 (shown in Figure 9B ), and the LOD was found to be 1.48 µg/mL [48] . Additionally, the sensor sensitivity was estimated from the slope of the calibration curve divided by the surface area of the GO-IL-PGE and was found to be 1.49 µA.mL/µg.cm 2 . The selectivity of the assay was tested in case of hybridization between the probe and the BRCA1 target or 3-base mismatched (MM) or the mixture samples containing target:MM. The highest guanine signal was obtained with the full-match hybridization between the BRCA1 probe and the target ( Figure 10 ) and was measured as 5.08 ± 0.69 µ A (RSD %, 13.74 %, n = 3). The negligible guanine oxidation signal was 0.62 ± 0.10 µ A (RSD %, 17.32 %, n = 3) in case of hybridization between the BRCA1 probe and the MM sequence. Otherwise, in the case of hybridization between the probe and the BRCA1 target in the presence of MM sequence, almost the same guanine signal was measured with the one obtained by the full-match hybridization signal. Therefore, it can be concluded that our assay presented a selective behavior on BRCA1-detection. The selectivity of the assay was tested in case of hybridization between the probe and the BRCA1 target or 3-base mismatched (MM) or the mixture samples containing target:MM. The highest guanine signal was obtained with the full-match hybridization between the BRCA1 probe and the target ( Figure 10 ) and was measured as 5.08 ± 0.69 µA (RSD %, 13.74%, n = 3). The negligible guanine oxidation signal was 0.62 ± 0.10 µA (RSD %, 17.32%, n = 3) in case of hybridization between the BRCA1 probe and the MM sequence. Otherwise, in the case of hybridization between the probe and the BRCA1 target in the presence of MM sequence, almost the same guanine signal was measured with the one obtained by the full-match hybridization signal. Therefore, it can be concluded that our assay presented a selective behavior on BRCA1-detection. The intra-day repeatability of the assay by GO-IL-PGEs was examined by measuring the guanine oxidation signal in the case of hybridization between the probe and the 10 µ g/mL BRCA target consecutively during three days (Table S1 ). The values of RSD % varied from 0.8 % and 8.8 %. In order to examine the inter-day reproducibility, these results were combined, and the average signal was calculated and found to be 5.01 ± 0.33 µ A with a RSD % of 6.62 % (n = 7) (shown in Table S2 ). The results revealed that the GO-IL-PGEs exhibited a satisfactory reproducibility with a mean change in the response of 0.33 µ A and a relative standard deviation of 6.62 % (n = 7). In this study, graphene-oxide (GO) and ionic-liquid (IL) composite-modified pencilgraphite electrodes (GO-IL-PGEs) were developed for the first time and then applied as a biosensing platform for electrochemical detection of breast cancer 1 (BRCA1) gene. The detection protocol consists of GO-IL modification onto the PGE surface, hybridization of a BRCA1-specific probe with its target, and finally the immobilization of DNA-DNA hybrid onto the surface of the GO-IL-PGE. The GO-IL incorporated the advantages of GO and IL, which provide a thin film at the electrode surface for immobilization of the BRCA1 probe-target hybrid. Additionally, the GO-IL composite could offer enhanced electrocatalytic properties and a larger electroactive surface area than the bare electrode or only the GO-modified electrode. The hybridization was monitored by directly measuring the guanine oxidation signal at +1.0 V using the differential pulse voltammetry (DPV) technique. In our previous work [41] , chemically activated and graphene oxide (GO) modified graphite electrodes (CA-GO-PGEs) were used for impedimetric detection of miRNA-34a, which was a biomarker of Alzheimer's disease. In that assay, the PGE surface was firstly activated by using covalent agents (CA), and then GO modification was done by itself. In the meantime, hybridization of the miRNA-34a-specific DNA probe with the miRNA-34a RNA target was performed in the solution phase. After all, the sample containing DNA-miRNA hybrids were immobilized onto the CA-GO-PGE surface. The impedimetric detection of miRNA-34a was then performed by measuring the fractional change at the The intra-day repeatability of the assay by GO-IL-PGEs was examined by measuring the guanine oxidation signal in the case of hybridization between the probe and the 10 µg/mL BRCA target consecutively during three days (Table S1 ). The values of RSD % varied from 0.8% and 8.8%. In order to examine the inter-day reproducibility, these results were combined, and the average signal was calculated and found to be 5.01 ± 0.33 µA with a RSD % of 6.62% (n = 7) (shown in Table S2 ). The results revealed that the GO-IL-PGEs exhibited a satisfactory reproducibility with a mean change in the response of 0.33 µA and a relative standard deviation of 6.62% (n = 7). In this study, graphene-oxide (GO) and ionic-liquid (IL) composite-modified pencilgraphite electrodes (GO-IL-PGEs) were developed for the first time and then applied as a biosensing platform for electrochemical detection of breast cancer 1 (BRCA1) gene. The detection protocol consists of GO-IL modification onto the PGE surface, hybridization of a BRCA1-specific probe with its target, and finally the immobilization of DNA-DNA hybrid onto the surface of the GO-IL-PGE. The GO-IL incorporated the advantages of GO and IL, which provide a thin film at the electrode surface for immobilization of the BRCA1 probetarget hybrid. Additionally, the GO-IL composite could offer enhanced electrocatalytic properties and a larger electroactive surface area than the bare electrode or only the GOmodified electrode. The hybridization was monitored by directly measuring the guanine oxidation signal at +1.0 V using the differential pulse voltammetry (DPV) technique. In our previous work [41] , chemically activated and graphene oxide (GO) modified graphite electrodes (CA-GO-PGEs) were used for impedimetric detection of miRNA-34a, which was a biomarker of Alzheimer's disease. In that assay, the PGE surface was firstly activated by using covalent agents (CA), and then GO modification was done by itself. In the meantime, hybridization of the miRNA-34a-specific DNA probe with the miRNA-34a RNA target was performed in the solution phase. After all, the sample containing DNA-miRNA hybrids were immobilized onto the CA-GO-PGE surface. The impedimetric detection of miRNA-34a was then performed by measuring the fractional change at the charge transfer resistance (R ct ). The aim of the earlier work was to develop an impedimetric assay for the detection of microRNA by GO-modified electrodes. On the other hand, the aim of our present work was to develop a voltammetric assay in combination with GO-IL-composite-modified electrodes to detect breast cancer 1 (BRCA1) gene while incorporating the advantages of these two materials while increasing the electrocatalytic properties of the electrode. The proposed assay was able to identify the complementary target sequence even in the mixture with mismatched sequences. The preparation of GO-IL-PGEs was performed in 20 min, and BRCA1 gene detection could be managed within 35 min with a LOD of 1.48 µg/mL (equal to 251 nM or 25 pmol in a 100 µL sample). Some of the earlier studies related to the electrochemical detection of BRCA1 gene were summarized in Table 5 [49] [50] [51] [52] [53] [54] [55] [56] [57] . Although lower LOD values have been reported in previous studies [50] [51] [52] [53] [54] 56, 57] , the LOD value obtained herein was in the acceptable level and was even lower than the ones reported by some of the earlier studies [49, 55] . The single-use GO-IL-modified sensors brought herein the following important advantages: they are easy to use and cost effective per measurement, resulting with reproducible results in a short time (35 min) in comparison to earlier works. Therefore, in comparison to the conventional nucleic acid detection technologies, our assay provides a rapid, practical, low-cost, and user-friendly nucleic acid analysis with a satisfactory detection limit [2] [3] [4] [5] [6] . With any of the conventional methods (e.g., capillarity electrophoresis, sequencing, and PCR-mediated techniques) BRCA1 gene detection could be achieved in a short time; however, the determination was carried out in 35 min in our electrochemical DNA biosensor-based assay. This simple modification and detection protocol can be extended for the monitoring of different type of nucleic acids or protein detection further, which can prompt new opportunities to design novel nucleic-acid-sensing strategies and applications. Supplementary Materials: The following supporting information can be downloaded at: https: //www.mdpi.com/article/10.3390/bios12020095/s1, Figure S1 : The electrochemical cell with threeelectrode setup including PGE as a working electrode, silver chloride electrode as reference and platinum wire as counter electrode during measurement. Left image represents the enlarged view of the electrochemical cell with three-electrode system, Figure S2 : Line graph with the average guanine signal of hybridization between 6 µg/mL BRCA1 probe and 2 µg/mL, 4 µg/mL, 6 µg/mL, 8 µg/mL, 10 µg/mL, 12 µg/mL BRCA 1 target on the surface of GO-IL-PGE (n = 3), Table S1: The guanine oxidation signals measured in the presence of the hybridization between probe and 10 µg/mL BRCA1 target by GO-IL-PGEs for three different days with the values of the average guanine oxidation signals (n = 2 or n = 3) and the standard deviation with the RSD % for presenting the intra-day reproducibility, Table S2 : The guanine oxidation signals measured in the presence of the hybridization between probe and 10 µg/mL BRCA1 target by GO-IL-PGEs for three different days with the average guanine oxidation signals (n = 7) and the standard deviation with the RSD % for presenting the inter-day reproducibility. The data presented in this study are available within the article and its Supplementary Materials. Other data that support the findings of this study are available upon request from the corresponding author. Ultrasensitive detection of cancer biomarkers using conducting polymer/electrochemically reduced graphene oxide-based biosensor: Application toward BRCA1 sensing Capillary electrophoresis as alternative method to detect tumor genetic mutations: The model built on the founder BRCA1 c.4964_4982del19 variant BRCA sequencing of tumors: Understanding its implications in the oncology community A Reference System for BRCA Mutation Detection Based on Next-Generation Sequencing in the Chinese Population A reliable method for the detection of BRCA1 and BRCA2 mutations in fixed tumour tissue utilising multiplex PCR-based targeted next generation sequencing Targeted capture-based NGS is superior tomultiplex PCR-based NGS for hereditary BRCA1 and BRCA2 gene analysis in FFPE tumor samples A novel biosensor for the ultrasensitive detection of the lncRNA biomarker MALAT1 in non-small cell lung cancer Magnetic-Bead based Label-Free Electrochemical Detection of DNA Hybridization Genomagnetic Electrochemical Assays of DNA Hybridization Fluorometric determination of the breast cancer 1 gene based on the target-induced conformational change of a DNA template for copper nanoclusters Surface plasmon resonance biosensor for label-free and highly sensitive detection of point mutation using polymerization extension reaction Real-time and Label-free Bio-sensing of Molecular Interactions by Surface Plasmon Resonance: A Laboratory Medicine Perspective Nanomaterial based electrochemical DNA sensing strategies Biosensors based on graphene oxide and its biomedical application Graphene oxide-ZnO nanocomposite modified electrode for the detection of phenol Carbon monoxide sensor based on polypyrrole-graphene oxide composite: A cost-effective approach Sensitive and Selective Detection of Tartrazine Based on TiO 2 -Electrochemically Reduced Graphene Oxide Composite-Modified Electrodes Graphene and Graphene-Based Nanomaterials for DNA Detection: A Review Biocompatible ionic liquids: Fundamental behaviours and applications Rapid diagnosis of SARS-CoV-2 using potential point-of-care electrochemical immunosensor: Toward the future prospects Ionic liquid functionalized 3D graphene-carbon nanotubes-AuPd nanoparticlesmolecularly imprinted copolymer based paracetamol electrochemical sensor: Preparation, characterization and application A novel ratiometric electrochemical sensor for the selective detection of citrinin based on molecularly imprinted poly (thionine) on ionic liquid decorated boron and nitrogen co-doped hierarchical porous carbon A sensitive and fast approach for voltammetric analysis of bisphenol a as a toxic compound in food products using a Pt-SWCNTs/ionic liquid modified sensor An electrochemical chiral sensor based on the synergy of chiral ionic liquid and 3D-NGMWCNT for tryptophan enantioselective recognition Electrochemical sensor based on ionic liquid and carbon black for voltammetric determination of Allura red colorant at nanomolar levels in soft drink powders Chitosan/Ionic liquid composite electrode for electrochemical monitoring of the surface-confined interaction between mitomycin C and DNA Development of Ionic Liquid Modified Disposable Graphite Electrodes for Label-Free Electrochemical Detection of DNA Hybridization Related to Microcystis spp Ionic Liquid Modified Single-use Electrode Developed for Voltammetric Detection of miRNA-34a and its Application to Real Samples An Impedimetric Biosensor Based on Ionic Liquid-Modified Graphite Electrodes Developed for microRNA-34a Detection An ultrasensitive label-free electrochemical immunosensor based on 3D porous chitosan-graphene-ionic liquid-ferrocene nanocomposite cryogel decorated with gold nanoparticles for prostate-specific antigen Electrochemical genosensor based on carbon nanotube/amine-ionic liquid functionalized reduced graphene oxide nanoplatform for detection of human papillomavirus (HPV16)-related head and neck cancer Label-Free Electrochemical Immunosensor Based on a Functionalized Ionic Liquid and Helical Carbon Nanotubes for the Determination of Cardiac Troponin I Simultaneous determination of ascorbic acid, dopamine, and uric acid using graphene quantum dots/ionic liquid modified screen-printed carbon electrode Electrochemical Determination of Capsaicin by Ionic Liquid Composite-Modified Electrode A novel ionic liquid functionalized graphene oxide supported gold nanoparticle composite film for sensitive electrochemical detection of dopamine Simple Fabrication of Reduced Graphene Oxide-Ionic Liquid Composite Modified Electrode for Sensitive Detection of Sulfadiazine Ionic liquid modified N-doped graphene as a potential platform for the electrochemical discrimination of DNA sequences Electrochemical detection of microRNAs by graphene oxide modified disposable graphite electrodes Impedimetric detection of microRNA at graphene oxide modified sensors Graphene oxide modified single-use electrodes and their application for voltammetric miRNA analysis Graphene Oxide Modified Chemically Activated Graphite Electrodes for Detection of microRNA Carboxylated-Graphene Decorated Pencil Graphite Electrode as a Platform for Voltammetric Detection of DNA Electrochemical monitoring of biointeraction by graphene-based material modified pencil graphite electrode Determination of the electrochemically effective electrode area An insight into impedimetric immunosensor and its electrical equivalent circuit Ultrasensitive DNA hybridization biosensor based on polyaniline Impedance spectroscopy of self-assembled monolayers on Au (111) Statistics and Chemometrics for Analytical Chemistry Electrochemical sensing of label free DNA hybridization related to breast cancer 1 gene at disposable sensor platforms modified with single walled carbon nanotubes Graphene-DNA electrochemical sensor for the sensitive detection of BRCA1 gene An ultrasensitive electrochemical impedance sensor for a special BRCA1 breast cancer gene sequence based on lambda exonuclease assisted target recycling amplification Detection of BRCA1 gene on partially reduced graphene oxide biosensors Carbon-nanotube-modified glassy carbon electrodes for amplified label-free electrochemical detection of DNA hybridization A label-free electrochemical DNA biosensor for breast cancer marker BRCA1 based on self-assembled antifouling peptide monolayer Detection of Breast Cancer 1 (BRCA1) Gene Using an Electrochemical DNA Biosensor Based on Immobilized ZnO Nanowires A Novel Biosensor Nanomaterial for Ultraselective and Ultrasensitive Electrochemical Diagnosis of Breast Cancer-Related BRCA1 Gene Electrochemical DNA Sensor for Sensitive BRCA1 Detection Based on DNA Tetrahedral-Structured Probe and Poly-Adenine Mediated Gold Nanoparticles A.E. would like to express her gratitude to the Turkish Academy of Sciences (TUBA) as a Principal member for its partial support. The authors declare no conflict of interest.