key: cord-0026010-lz8uto7k authors: Welden, Melanie; Poghossian, Arshak; Vahidpour, Farnoosh; Wendlandt, Tim; Keusgen, Michael; Wege, Christina; Schöning, Michael J. title: Towards Multi-Analyte Detection with Field-Effect Capacitors Modified with Tobacco Mosaic Virus Bioparticles as Enzyme Nanocarriers date: 2022-01-14 journal: Biosensors (Basel) DOI: 10.3390/bios12010043 sha: 0cdd1777e536f0a50b90c10b264d05c435db7b9d doc_id: 26010 cord_uid: lz8uto7k Utilizing an appropriate enzyme immobilization strategy is crucial for designing enzyme-based biosensors. Plant virus-like particles represent ideal nanoscaffolds for an extremely dense and precise immobilization of enzymes, due to their regular shape, high surface-to-volume ratio and high density of surface binding sites. In the present work, tobacco mosaic virus (TMV) particles were applied for the co-immobilization of penicillinase and urease onto the gate surface of a field-effect electrolyte-insulator-semiconductor capacitor (EISCAP) with a p-Si-SiO(2)-Ta(2)O(5) layer structure for the sequential detection of penicillin and urea. The TMV-assisted bi-enzyme EISCAP biosensor exhibited a high urea and penicillin sensitivity of 54 and 85 mV/dec, respectively, in the concentration range of 0.1–3 mM. For comparison, the characteristics of single-enzyme EISCAP biosensors modified with TMV particles immobilized with either penicillinase or urease were also investigated. The surface morphology of the TMV-modified Ta(2)O(5)-gate was analyzed by scanning electron microscopy. Additionally, the bi-enzyme EISCAP was applied to mimic an XOR (Exclusive OR) enzyme logic gate. The field effect in an electrolyte-insulator-semiconductor (EIS) system offers a universal transducer principle for designing many kinds of chemical sensors and biosensors (see e.g., recent reviews [1] [2] [3] [4] [5] [6] [7] [8] [9] ). EIS capacitors (EISCAP) are considered as the simplest type of such field-effect sensors [10] . They have been implemented for detecting various biochemical species such as ions [11] [12] [13] [14] , charged molecules (DNA (deoxyribonucleic acid) [15] [16] [17] , protein biomarkers [18] [19] [20] [21] [22] , polyelectrolytes [23, 24] ), virus-like particles [25, 26] , ligandstabilized nanoparticles [27, 28] , etc. In addition, numerous enzyme-modified EISCAP biosensors were constructed for the detection of various analytes such as glucose [29] [30] [31] , urea [30] [31] [32] , creatinine [33] , penicillin [31, 34] , formaldehyde [35] , triglycerides [36] , and acetoin [37] . The operation mechanism of these biosensors is based on the detection of local pH changes resulting from the catalytic reaction of the immobilized enzyme on the sensor surface with its specific substrate [10] . Moreover, the ability of EISCAP sensors for multi-analyte detection using a single EISCAP chip or an array of EISCAPs has been demonstrated [31, [38] [39] [40] [41] [42] [43] [44] . Generally, the analytical characteristics of enzyme-based biosensors are strongly affected by the enzyme immobilization method [45, 46] . Therefore, the choice of an appropriate enzyme immobilization strategy is a key factor for designing the biorecognition part of biosensors, including EISCAPs. The immobilization method must provide a high enzyme load on the biosensor surface, a good accessibility of the active sites for the target analytes, as well as to retain the structure, function, and catalytic activity of the enzyme [45, 46] . Intensive efforts have been made during the last decade to develop novel immobilization techniques to improve the biosensor performance: the use of nanomaterials (e.g., metal, oxide or organic nanoparticles, nanowires, carbon nanotubes, magnetic beads) as nanoscale scaffolds for the immobilization of receptor molecules [47] [48] [49] [50] as well as the incorporation of enzymes within nanoscale structures (e.g., alginate gels [51] , polyelectrolyte/enzyme [52] or carbon nanotube/enzyme multilayers [53] ). More recently, due to their regular shape, high surface-to-volume ratio, and extremely high density of surface docking sites, biological nanoscaffolds such as plant virus-like particles have increasingly been used for the precisely positioned immobilization of receptors on different transducers for biosensing purposes [54] [55] [56] . In contrast to various chemically synthesized nanoparticles, which are typically polydisperse with a more randomized and unreproducible distribution of particle size and density of shell molecules, many types of virus particles are monodisperse with regard to uniform morphologies and specific dimensions. The tobacco mosaic virus (TMV) is one of the most comprehensively investigated plant viruses. Native full-length TMV particles are 300 nm long, nanotube-like nucleoprotein complexes with inner and outer diameters of 4 and 18 nm, respectively [56, 57] . TMV particles exhibit excellent chemical and physical stability: in solutions, they can withstand 90 • C [58] and pH values between pH 3 and pH 9 [57] . The outer surface of each TMV nanotube holds thousands of docking sites accessible on about 2130 identical, helically arranged coat protein (CP) subunits, capable in the coupling of functional molecules. Therefore, TMV particles have often been utilized as biological nanoscaffolds for an extremely dense and precisely controlled immobilization of biorecognition molecules (receptors), including enzymes [55, 56, 59] . Moreover, TMV particles functionalized with bioreceptors can be simply combined with different transducer structures for chemical and biological sensing. For example, TMV particles were applied to assist the detection of the explosive agent trinitrotoluene [60] , volatile organic compounds [61] , and antigen-antibody binding [62, 63] . The authors introduced amperometric glucose [64] as well as colorimetric and field-effect penicillin biosensors [65, 66] by using TMV particles as nanoscaffolds for the immobilization of glucose oxidase and penicillinase, respectively. TMV particles as receptor nanocarriers usually enable increased receptor densities per sensor area, an enhanced mass transport of the target molecules to the TMV surface in comparison to planar surfaces, and a favorable orientation of the receptor molecules for an increased receptor-substrate interaction; all these factors enhance the overall biosensor performance [56, 64] . In this work, the ability of TMV-assisted enzyme-based EISCAPs for multi-analyte detection was demonstrated. Penicillinase and urease were co-immobilized onto TMV surfaces for the sequential detection of penicillin and urea using the same p-Si-SiO 2 -Ta 2 O 5 field-effect sensor. These two enzyme/substrate systems (i.e., penicillinase/penicillin and urease/urea) represent typical model experiments inducing counter-rotating pH shifts on the sensor surface: a) generating hydrogen ions (penicillinase/penicillin → pH decrease) or b) consuming hydrogen ions (urease/urea → pH increase). For comparison, TMV-modified EISCAPs immobilized with either penicillinase or urease (no co-immobilization) were also studied. Finally, in a proof-of-concept experiment, the newly developed TMV-assisted bienzyme EISCAP biosensor was applied to mimic an XOR (Exclusive OR) enzyme logic gate. A TMV variant (TMV Cys ) containing a S3C mutation close to the N-terminus of every CP was employed as a viral enzyme nanocarrier with a spacing of 2.5 to 3.5 nm of coupling sites on their outer protein coat [67] . This mutant exposed 2130 cysteine residues on each TMV particle presenting surficial sulfhydryl groups. TMV Cys extracted from infected Nicotiana tabacum leaves were equipped with PEG 11 -biotin moieties via maleimidesulfhydryl-coupling (EZ-Link®Maleimide-PEG 11 -Biotin, Thermo Scientific, Rockford, IL, USA) according to [56] . Briefly, a maleimide-PEG 11 -biotin linker was incubated with TMV Cys particles in a molar ratio of 3:1 (linker/CP) for 3 h at 26 • C under agitation. Unbound linker molecules were removed by centrifugal ultrafiltration (Amicon Ultra, 30 kDa molecular weight cut-off, Merck-Millipore, Darmstadt, Germany) in five consecutive washing steps with 10 mM of sodium-potassium-phosphate (SPP) buffer (pH 7.0). The resulting biotinylated TMV Cys (TMV Cys / Bio ) particles were resuspended and stored in this buffer at 4 • C. The particles were analyzed by sodium dodecyl sulphate-polyacrylamide gel electrophoresis and colloidal Coomassie Brilliant Blue G-250 staining [68] . Densitometric comparison of the signals corresponding to the biotinylated and the non-modified form of the CP confirmed biotinylation of >90 % of the CPs. Penicillinase from Bacillus cereus (1500-3000 Units/mg protein, Sigma Aldrich, Darmstadt, Germany) and urease from Canavalia ensiformis (75,265 Units/g solid, from Jack bean, Sigma Aldrich, Germany) were applied as the model enzymes. In order to immobilize them on the biotinylated TMV particles, they were conjugated with streptavidin (SA), allowing strong attachment via SA-biotin high affinity binding. The streptavidin conjugation was performed by utilizing a commercial conjugation kit (LYNX Streptavidin rapid conjugation kit, Bio-Rad, Great Britain). The conjugation of penicillinase was performed as described in [65] , using a molar ratio between penicillinase and SA of 1:25 in the reaction mixture. For the urease conjugation, the same protocol was applied with a molar ratio of 1:10. The streptavidin-conjugated enzymes (SA-enzymes) were stored in stock solutions (10 mM phosphate buffered saline (PBS)) with a concentration of 600 Units/mL (SA-penicillinase) and of 2000 Units/mL (SA-urease) at 4 • C. For the fabrication of the bi-enzyme biosensors, both enzyme solutions were mixed in a ratio of 1:1, resulting in an enzyme cocktail containing 300 Units/mL SA-penicillinase and 1000 Units/mL SA-urease. In this study, EISCAPs with an Al/p-Si/SiO 2 (30 nm)/Ta 2 O 5 (60 nm) layered structure were used as the sensor platform, as schematically represented in Figure 1a . Details of the fabrication process were described in [25] . Prior to their modification with the TMV particles and enzymes, the sensors were cleaned in an ultrasonic bath for 5 min each with acetone, isopropanol, ethanol, and deionized water, and then installed into a homemade measurement cell. In the measurement cell, the sensor chip was sealed by an O-ring, with 0.5 cm 2 of the sensor surface in contact with the electrolyte (see Figure 1a) . A 50 µL TMV solution (0.1 mg/mL) was incubated for one hour at room temperature (RT) on this exposed Ta 2 O 5 surface to allow TMV adsorption. Subsequently, not-adsorbed TMV particles were washed away with 10 mM PBS buffer. Afterwards, 50 µL of the particular SA-enzyme solution (single enzyme EISCAP: 50 µL SA-penicillinase or 50 µL SA-urease solution; bi-enzyme EISCAP: 50 µL SA-penicillinase + SA-urease solution) were incubated for two hours at RT. The sensor surface was then flushed three times with 0.33 mM PBS and conditioned in 0.33 mM PBS buffer solution for at least one hour, before the electrochemical measurements were started. When the EISCAPs were not in use, they were stored in 0.33 mM PBS buffer at 4 • C. For electrochemical characterization of the TMV-assisted EISCAPs, an Ag/AgCl reference electrode (filled with 3 M KCl, Metrohm, Filderstadt, Germany) was immersed in the buffer solution and connected to an impedance analyzer (Zahner Zennium, Zahner Elektrik, Kronach, Germany). Furthermore, the Al rear-side contact was electrically connected to the impedance analyzer. All measurements were performed in a measurement buffer (0.33 mM PBS buffer) with varying concentrations of penicillin G (Sigma Aldrich, Darmstadt, Germany) and/or urea (GE Healthcare Bio-Sciences AB, Uppsala, Sweden) at RT. In order to avoid signal interferences, the set-up (except of the impedance analyzer) was integrated in a dark Faraday cage. As a first step, leakage-current measurements were carried out in a measurement solution without penicillin and urea. Therefore, gate voltages from -3 V to +3 V were applied with 100 mV steps between the reference electrode and the Al rear-side contact. These measurements served as a quality control of the insulator layer, where only sensors exhibiting a leakage current < 10 nA were selected for further electrochemical measurements. In the next step, capacitance-voltage (C-V) measurements were performed in the measurement buffer (by applying a gate voltage between -2 V and +2 V with 100 mV steps) to check the correct functioning of the field-effect EISCAPs. In order to measure the capacitance of the EISCAP, a small AC (alternating current) voltage of 20 mV with a frequency of 120 Hz was superimposed. For the following constant-capacitance (ConCap) measurements, a working point was set in the depletion For electrochemical characterization of the TMV-assisted EISCAPs, an Ag/AgCl reference electrode (filled with 3 M KCl, Metrohm, Filderstadt, Germany) was immersed in the buffer solution and connected to an impedance analyzer (Zahner Zennium, Zahner Elektrik, Kronach, Germany). Furthermore, the Al rear-side contact was electrically connected to the impedance analyzer. All measurements were performed in a measurement buffer (0.33 mM PBS buffer) with varying concentrations of penicillin G (Sigma Aldrich, Darmstadt, Germany) and/or urea (GE Healthcare Bio-Sciences AB, Uppsala, Sweden) at RT. In order to avoid signal interferences, the set-up (except of the impedance analyzer) was integrated in a dark Faraday cage. As a first step, leakage-current measurements were carried out in a measurement solution without penicillin and urea. Therefore, gate voltages from -3 V to +3 V were applied with 100 mV steps between the reference electrode and the Al rear-side contact. These measurements served as a quality control of the insulator layer, where only sensors exhibiting a leakage current < 10 nA were selected for further electrochemical measurements. In the next step, capacitance-voltage (C-V) measurements were performed in the measurement buffer (by applying a gate voltage between -2 V and +2 V with 100 mV steps) to check the correct functioning of the field-effect EISCAPs. In order to measure the capacitance of the EISCAP, a small AC (alternating current) voltage of 20 mV with a frequency of 120 Hz was superimposed. For the following constant-capacitance (ConCap) measurements, a working point was set in the depletion region of the C-V curve at about 60% of the maximum capacitance. The ConCap mode offered the time-dependent detection of surface-charge (potential) changes induced by, e.g., local pH changes at the sensor surface. During the ConCap mode, the capacitance of the sensor structure in the working point was kept constant by a control loop: Changes of the surface potential, e.g., induced by the enzymatic conversion of penicillin or urea, were compensated by applying an opposed voltage at the reference electrode. These voltage changes were recorded over time, allowing for the dynamic detection of changes in the penicillin and urea concentrations. A detailed description of the C-V and ConCap-measurement mode was provided in [23] . The ConCap measurements were conducted in a measurement buffer as well as in penicillin and urea solutions with concentrations between 0.1 and 5 mM for single-enzyme sensors and between 0.1 mM and 3 mM for bi-enzyme sensors. The penicillin and urea were purchased from Sigma Aldrich (Darmstadt, Germany) and GE Healthcare Bio-Sciences AB (Uppsala, Sweden), respectively. To control and characterize the TMV adsorption on the Ta 2 O 5 -gate surface, SEM images of the TMV-modified sensor chip were taken using a JEOL JSM-7800F Schottky fieldemission microscope (JEOL GmbH, Freising, Germany). For this purpose, the sensors were mounted out of the measurement chamber, rinsed with deionized water, and dried with nitrogen to remove the salt residues of the buffer solution. Subsequently, an approximately 5 nm thick platinum-palladium layer was sputtered onto the sensor surface to provide conductivity and prevent the additional charging of the sensor surface. To ensure that the TMV particles had adsorbed to the Ta 2 O 5 -gate surface within the O-ring and thus could act as nanocarriers for enzyme attachment, SEM images of the sensor surface were obtained after TMV loading. Figure 1b-d shows exemplary SEM images of the chip surface with a clear boundary between the TMV-modified region and the area sealed by the O-ring (see Figure 1c ). Within the O-ring, TMV particles were homogeneously distributed on the Ta 2 O 5 surface, with some virus particles present as lateral or head-to-tail aggregates (Figure 1d ), as is typical for TMV particles [25, 26, 66] . The TMV particles were adsorbed in a high density of about 6.3 × 10 9 particles/cm 2 on the Ta 2 O 5 surface, which was slightly higher than was reported in previous works [25, 26, 66] and revealed that the TMVs had been successfully attached to the Ta 2 O 5 surface where they were available for enzyme immobilization. In order to study the sensor performance of TMV-assisted single enzyme EISCAPs, penicillin and urea biosensors with penicillinase and urease, respectively, were fabricated separately. Penicillinase catalyzes the conversion of penicillin to penicilloic acid, whereby H + ions are produced, leading to a local pH decrease [69] . In contrast, during the hydrolysis reaction of urea catalyzed by the urease, OHions are produced (or H + ions are consumed), resulting in a local pH increase [70, 71] . Changes in pH at the EISCAP surface cause the surface to become more positively charged (in the case of a pH decrease) or more negatively charged (in the case of a pH increase). This surface-charge change influences the width of the space-charge region in the semiconductor layer and, thus, the total capacitance of the sensor structure. Choosing these two enzymes enabled investigating whether EISCAPs modified with TMV particles as enzyme nanocarriers were suitable for enzymatic reactions involving both acidification and alkalization. The direction of the EISCAP-signal changes would directly correlate with these two kinds of enzymatic reactions, generating (penicillinase/penicillin) or consuming (urease/urea) hydrogen ions. The measurements were carried out in 0.33 mM PBS at pH 8.0, which corresponds to the pH optimum of the penicillinase [72] . For each penicillin concentration, the ConCap signal was recorded for approximately 5 min. Figure 2a shows a ConCap curve of the TMV/SA-penicillinase modified EISCAP biosensor in the loop of penicillin concentrations of 0.1, 0.5, 1, 3, 5, 3, 1, 0.5, and 0.1 mM. The measurements were carried out in 0.33 mM PBS at pH 8.0, which corresponds to the pH optimum of the penicillinase [72] . For each penicillin concentration, the ConCap signal was recorded for approximately 5 min. As expected, with increasing penicillin concentration, the recorded signal shifted towards less positive voltages. By increasing the penicillin concentration, more H + ions were generated and the local pH value at the sensor surface became lower: The produced H + ions protonate the hydroxyl groups on the Ta2O5 surface making it more positively charged. Consequently, the total capacitance of the EISCAP was decreased. To keep the total capacitance of the sensor constant, the applied voltage at the reference electrode should become more negative (or less positive), which is visible as a shift in the ConCap signal. Conversely, with decreasing penicillin concentration, the ConCap signal shifted in As expected, with increasing penicillin concentration, the recorded signal shifted towards less positive voltages. By increasing the penicillin concentration, more H + ions were generated and the local pH value at the sensor surface became lower: The produced H + ions protonate the hydroxyl groups on the Ta 2 O 5 surface making it more positively charged. Consequently, the total capacitance of the EISCAP was decreased. To keep the total capacitance of the sensor constant, the applied voltage at the reference electrode should become more negative (or less positive), which is visible as a shift in the ConCap signal. Conversely, with decreasing penicillin concentration, the ConCap signal shifted in the opposite direction. The clear steps at different penicillin concentrations also underlined a fast response time of the TMV-based penicillin biosensor. The measurement curve revealed that the sensor had a low hysteresis of 1 and 3 mV in the buffer without penicillin and in the 3 mM penicillin solution, respectively. However, it was observed that hysteresis is somewhat higher at lower penicillin concentrations with a maximum value of 8 mV at 0.5 mM. Generally, hysteresis of pH-sensitive field-effect devices is interpreted as slow response due to the slow buried sites underneath the gateinsulator surface (see e.g., [73, 74] ). It was reported that the hysteresis increases with an increasing pH-loop time, and in acid solutions it is smaller than in alkaline solutions [73, 74] In addition, the hysteresis width may be affected by the background long-term drift, slow states at the Si-SiO 2 interface, as well as by the possible alteration of the enzyme activity due to local pH changes. The experiments performed in this study did not allow us to provide a clear explanation for the analyte-concentration dependence of the hysteresis width. Therefore, additional in-depth studies are needed to quantify this phenomenon. The inset figure indicated the calibration plots for upward and downward penicillinconcentration loops evaluated from the ConCap response. The TMV-assisted EISCAP biosensor exhibited a high penicillin sensitivity of about 98 and 95 mV/dec for the increasing and decreasing concentration series of measurements, respectively. The small difference in penicillin sensitivities of 3 mV/dec observed for upward and downward concentration loops could be attributed to the hysteresis effect. The obtained sensitivity values were comparable to our previous work with TMV-based penicillin biosensors [66] and were higher than sensitivity values obtained for EISCAPs with adsorptively immobilized penicillinase (68.7 mV/dec) [31] . One of the important operation characteristics of biosensors is the reproducibility of the sensor response. To demonstrate the reproducibility of the TMV-assisted EISCAP penicillin biosensor, the ConCap signal was repeatedly measured in buffer (six times) and in a 0.5 mM penicillin solution (five times) in alternating order. The results of these experiments are shown in Figure 2b . The mean signal was 67 ± 1 mV, which underlines the high reproducibility of the developed penicillin biosensor. Figure 3a depicts a ConCap curve of an EISCAP modified with TMV/SA-urease, which was recorded in 0.33 mM PBS at pH 7.4, specified by the supplier as the optimum pH for urease [75] . Here, the ConCap signal behaved in the opposite way to that of the penicillin biosensor: with an increasing urea concentration from 0.1 to 5 mM, the measurement signal rose in the direction of the more positive (or less negative) voltages. The observed signal behavior could be explained as follows: an increase of the urea concentration resulted in a rise of the local pH value and deprotonation of the hydroxyl groups on the Ta 2 O 5 surface, making it more negatively charged, whereby the total capacitance of the EISCAP increased. To keep the overall capacitance of the EISCAP sensor constant, the applied voltage at the reference electrode must be more positive (or less negative), which appeared as a signal shift in the ConCap curve. Again, as with the penicillin measurements, clearly delineated signal levels could be seen at different urea concentrations with a low hysteresis of 2 mV for 3 mM urea. However, the hysteresis width increased with a decreasing urea concentration and amounts of 11 mV at 0.1 mM. In contrast to the penicillin measurements, a relatively large hysteresis of 8 mV was also recorded in the buffer solution. Additionally, at low urea concentrations, it took longer before a stable sensor signal was achieved. From the ConCap response at different urea concentrations, the calibration curves for the upward and downward concentration loops were evaluated, which are depicted in the inset figure. The TMV-assisted urea biosensor reveals a high urea sensitivity of 55 and 49 mV/dec for the increasing and decreasing concentration series of measurements. The difference in urea sensitivities for the upward and downward concentration loops was larger (6 mV/dec) than that of the TMV-assisted penicillin biosensor. In other studies with urea sensors based on EISCAPs, urea sensitivities of 16 mV/dec (1 to 100 mM) by using a layer-by-layer nanofilm of ZnO nanocrystals and carbon nanotubes [76] , 32 mV/dec by the immobilization of urease on magnetic particles [71] , and 40.5 mV/dec (1 to 25 mM) in the case of nano-spotted urease [31] were achieved. Thus, the EISCAP modified with TMV/SA-urease offers a higher sensitivity and the ability of detecting low urea concentrations. the upward and downward concentration loops was larger (6 mV/dec) than that of the TMV-assisted penicillin biosensor. In other studies with urea sensors based on EISCAPs, urea sensitivities of 16 mV/dec (1 to 100 mM) by using a layer-by-layer nanofilm of ZnO nanocrystals and carbon nanotubes [76] , 32 mV/dec by the immobilization of urease on magnetic particles [71] , and 40.5 mV/dec (1 to 25 mM) in the case of nano-spotted urease [31] were achieved. Thus, the EISCAP modified with TMV/SA-urease offers a higher sensitivity and the ability of detecting low urea concentrations. The results of the reproducibility experiments are shown in Figure 3b . Like the penicillin biosensor, the ConCap signal of the TMV-assisted EISCAP urea biosensor was repeatedly measured in buffer (six times) and in a 0.5 mM urea solution (five times) in alternating order. The urea biosensor signal was reproducible with a mean signal of (88 ± 5) mV. The TMV-assisted bi-enzyme EISCAP, where both enzymes had been co-immobilized on the same sensor chip, was applied for the serial detection of urea and penicillin The results of the reproducibility experiments are shown in Figure 3b . Like the penicillin biosensor, the ConCap signal of the TMV-assisted EISCAP urea biosensor was repeatedly measured in buffer (six times) and in a 0.5 mM urea solution (five times) in alternating order. The urea biosensor signal was reproducible with a mean signal of (88 ± 5) mV. The TMV-assisted bi-enzyme EISCAP, where both enzymes had been co-immobilized on the same sensor chip, was applied for the serial detection of urea and penicillin for the first time. Figure The measurements were performed in the following order: First, the ConCap response was recorded in the buffer solution (0.33 mM PBS, pH 7.4) to obtain the baseline signal for the sensor, followed by measurements in urea solutions with different concentrations. Second, the sensor signal was recorded again in the buffer solution to obtain the baseline signal for the subsequent penicillin measurements with varying penicillin concentrations. Finally, the ConCap signal was recorded again in the buffer and urea solution. Such sequential arrangements of experiments are useful to examine the recoverability of the sensor signal as well as to identify possible hysteresis and drift effects. As expected, the bi-enzyme sensor showed distinct signal steps towards more positive voltages with an increasing urea concentration and towards more negative voltages with an increasing penicillin concentration. For example, at a urea and penicillin concentration of 3 mM, the signal shift reached approximately 114 and 146 mV, respectively. Taking into account a pH sensitivity of 56 mV/pH for the Ta 2 O 5 -gate EISCAPs studied in this work, the local pH value on the TMV/SA-enzyme modified EISCAP surface was estimated to be pH = 9.4 or pH = 4.8 at a urea or penicillin concentration of 3 mM, respectively. This indicated the stability of the TMV/SA-enzyme system on the EISCAP surface over a wide pH range. In Figure 4b , the evaluated calibration curves of the TMV-assisted bi-enzyme EIS-CAP for urea and penicillin are illustrated. The biosensor exhibited a urea sensitivity of 54 mV/dec, which was nearly similar to that of the single-enzyme biosensor. The penicillin sensitivity of the bi-enzyme EISCAP amounted to 85 mV/dec, which was slightly lower than that achieved with the single-enzyme EISCAP (see Section 3.2.1). The results indicated that both enzymes maintained their activity after the co-immobilization. In general, a possible non-specific adsorption of analyte molecules onto the gate surface of the field-effect device may induce an unwanted background signal, and, thus, may reduce the signal-to-noise ratio of the sensor signal [1, 4, 10] . In our case, such nonspecific adsorption may occur either on the TMV particles or on TMV-free areas of the gate surface of the EISCAP. Therefore, we studied the signal behavior of the TMV-modified (but enzyme-free) EISCAP in the urea and penicillin solutions. Figure 4c depicts the ConCap signal of the EISCAP recorded in 0.33 mM PBS (pH 7.4) buffer before (unmodified sensor) and after the loading of TMV particles, followed by measurements in buffer containing 0.5 mM urea or penicillin. As TMV particles are negatively charged [26] , the sensor signal shifted to more positive voltages after their loading onto the Ta 2 O 5 surface. However, upon subsequent measurements in analyte solutions, the signal remained almost constant, indicating that the existence of urea or penicillin molecules in the solution had practically no impact on the EISCAP sensor signal. In the past, intensive research was performed in the field of enzyme-based logic gates, which mimic the operation of electronic logic gates (see e.g., [77, 78] ). In enzyme logic gates, Boolean logic operations are activated by specific molecular inputs via enzymatic reactions. An interfacing of enzyme logic gates with electronic transducers was considered a very promising approach for designing digital biosensors that could provide qualitative evidence (in a YES/NO format) concerning the presence or absence of a specific analyte in the sample [79] [80] [81] . In previous works, we demonstrated the successful integration of enzyme logic gates with pH-sensitive EISCAPs as AND-Reset, OR-Reset, CNOT (controlled NOT) and XOR gates [82] [83] [84] [85] . In these logic gate devices, enzymes were immobilized onto the biosensor surface by means of physical entrapment within a membrane or through physical adsorption. In this work, the EISCAP modified with TMV particles (as bi-enzyme nanocarriers) was applied to mimic an XOR enzyme logic gate. The XOR gate is one of the important elements of biocomputing systems: it provides a true output (1) if only one of the inputs is true. If both input signals are false (0) or both are true (1), it must remain inactive (output signal 0). As was noted in [86] , the enzyme-based XOR gate is difficult to realize, because it requires enzymatic reactions that are able to produce approximately equal signals of opposite direction (net output signal should be 0), when both input signals simultaneously appear. The enzyme/substrate systems used in this work (i.e., urease/urea and penicillinase/penicillin) represented typical examples of such enzymatic reactions producing pH changes and consequently EISCAP signals in opposite direction (see Figure 4) . Figure 5a shows the schematic symbol and truth table for the enzyme-based XOR gate with urea as input A and penicillin as input B. The expected ConCap signal of the TMV-assisted bi-enzyme EISCAP biosensor for different input combinations is illustrated in Figure 5b . Biosensors 2022, 12, 43 11 of 16 to produce approximately equal signals of opposite direction (net output signal should be 0), when both input signals simultaneously appear. The enzyme/substrate systems used in this work (i.e., urease/urea and penicillinase/penicillin) represented typical examples of such enzymatic reactions producing pH changes and consequently EISCAP signals in opposite direction (see Figure 4 ). Figure 5a shows the schematic symbol and truth table for the enzyme-based XOR gate with urea as input A and penicillin as input B. The expected ConCap signal of the TMV-assisted bi-enzyme EISCAP biosensor for different input combinations is illustrated in Figure 5b . The presence of the respective analyte in the solution corresponded to the input signal 1, while the absence of an analyte was considered as the input signal 0. When no analyte was present in the solution (input combination (0,0)), the resulting output signal became 0, because no local pH change occurred. If one of the analytes (urea or penicillin) was present (input combination (1,0) or (0,1)), the output signal became 1 due to the enzymatically induced pH change. The presence of both analytes led to an output signal 0, as both enzymatic reactions occurred simultaneously with opposing pH shifts which compensated each other. Figure 5c represents an experimental ConCap logic signal of the TMV-assisted bienzyme EISCAP biosensor corresponding to various input combinations: The measurement was started in the buffer solution in the absence of the analyte (input combination (0,0)), followed by measurement in a 0.1 mM urea (input combination (1,0)) and 0.1 mM penicillin solution (input combination (0,1)). To return the local pH to its initial value, after each measurement in the particular analyte solution, the biosensor was exposed to the buffer solution and the baseline signal was recorded again. Finally, the ConCap signal was recorded in the solution containing both 0.1 mM urea and 0.1 mM penicillin (input combination (1,1)). Relatively large signal shifts of opposite direction were observed when only one of the analytes was present in the solution (24 mV at 0.1 mM urea; 12 mV at 0.1 mM penicillin). In contrast, a very small signal shift of 3 mV was detected if both analytes were present in the solution: two simultaneously occurring enzymatic reactions produced opposing pH changes, resulting in a nearly negligible sensor output signal. The described proof-of-principle experiment demonstrated the successful application of the TMV-assisted bi-enzyme EISCAPs for the development of XOR enzyme logic gates. Future work will be directed to minimize the output signal (should be ideally 0) of the EIS-CAP in the presence of both analytes via the optimization of the ratio of urease/penicillinase activities and/or urea/penicillin concentrations in the solution. The appropriate enzyme immobilization strategy is a key factor for the design of enzyme-based biosensors. Plant virus-like particles offer the possibility of extremely dense and precise immobilization of enzymes, due to uniform 3D structures and a high density of surface binding sites. In this study, TMV particles were utilized for the coimmobilization of penicillinase and urease onto the Ta 2 O 5 -gate surface of an EISCAP sensor for the serial detection of penicillin and urea. These two enzyme/substrate systems (i.e., penicillinase/penicillin and urease/urea) allowed the performance of typical model experiments with opposing pH shifts on the sensor surface: (a) generating hydrogen ions (penicillinase/penicillin with associated pH decrease) or (b) consuming hydrogen ions (urease/urea with associated pH increase). The sensitive characteristics of TMV-assisted bi-enzyme EISCAPs were characterized in the ConCap mode and compared with those of single-enzyme EISCAPs. The single-enzyme biosensors offered a high penicillin (95-98 mV/dec) and urea (49-55 mV/dec) sensitivity and remarkable reproducibility. These notable sensor properties were retained even when both enzymes were co-immobilized on the TMV particles, resulting in a penicillin and urea sensitivity of 85 and 54 mV/dec, respectively. The successful application of the TMV-assisted biosensor for designing an XOR enzymatic gate further highlighted the potential of the presented sensor arrangement. The results achieved in this study demonstrated the great prospects of TMV particles as enzyme nanocarriers in constructing EISCAP biosensors for multi-analyte detection. In future studies, TMV-assisted multi-enzyme EISCAPs could be extended to other enzymes. Furthermore, TMV-assisted EISCAPs could be employed for the characterization and application of two-or multi-step enzymatic cascades. Due to the uniform 3D structure of the TMV particles, the enzymes are geometrically close to each other, whereby enzymatic product/substrate diffusion can optimally take place, and the influence of interfering substances could be reduced. This may include not only arrangements with blends of distinct enzymes immobilized on the TMV nanoscaffolds but could also be tailored to enzyme systems collaborating between separate particles. If installed on TMV carriers with length-defined, selectively addressable longitudinal subdomains, such sensors might even provide fundamental insights into spacing-dependent interactions of enzyme groups and, thus, ease the design of high-efficiency artificial biocatalytic systems [87] . The data presented in this study are available on request from the corresponding author. Recent progress in silicon-based biologically sensitive field-effect devices Silicon nanowire field-effect-transistor-based biosensor for biomedical applications Field-effect transistor biosensing: Devices and clinical applications Field-effect sensors for virus detection: From Ebola to SARS-CoV-2 and plant viral enhancers. Front EGFET-based sensors for bioanalytical applications: A review Recent trends in field-effect transistors-based immunosensors Biologically coupled gate field-effect transistors meet in vitro diagnostics Light-addressable potentiometric sensors for quantitative spatial imaging of chemical species Light-addressable potentiometric sensors for cell monitoring and biosensing Capacitive field-effect chemical sensors and biosensors: A status report Electrochemical impedance spectroscopy study of Ta 2 O 5 based EIOS pH sensors in acid environment Miniaturized pHsensitive field-effect capacitors with ultrathin Ta 2 O 5 films prepared by atomic layer deposition Anion sensing and interfering behaviors of electrolyte-insulator-semiconductor sensors with nitrogen plasma-treated samarium oxide LaF 3 electrolyte-insulator-semiconductor sensor for detecting fluoride ions Label-free detection of DNA using high-k Lu 2 Ti 2 O 7 electrolyteinsulator-semiconductors Detection of PCR-amplified tuberculosis DNA fragments with polyelectrolyte-modified field-effect sensors Real-time monitoring of PCR amplification of proto-oncogene c-MYC using a Ta 2 O 5 electrolyte-insulator-semiconductor sensor Investigation of mechanisms involved in the enhanced label free detection of prostate cancer biomarkers using field effect devices Detection of protein kinase using an aptamer on a microchip integrated electrolyteinsulator-semiconductor sensor Label-free detection of rheumatoid factor using YbY x O y electrolyte-insulator-semiconductor devices Highly sensitive aptamer-based method for the detection of cardiac biomolecules on silicon dioxide surfaces Capacitance electrochemical biosensor based on silicon nitride transducer for TNF-α cytokine detection in artificial human saliva: Heart failure (HF) Electrical monitoring of polyelectrolyte multilayer formation by means of capacitive field-effect devices The effect of the type of doping on the electrical characteristics of electrolyteoxide-silicon sensors: pH sensing and polyelectrolyte adsorption Capacitive field-effect biosensor studying adsorption of tobacco mosaic virus particles Detection of plant virus particles with a capacitive field-effect sensor Oxygen plasma-treated gold nanoparticle-based field-effect devices as transducer structures for bio-chemical sensing Gold nanoparticle-modified capacitive field-effect sensors: Studying the surface density of nanoparticles and coupling of charged polyelectrolyte macromolecules Microfluidic chip integrated with an electrolyte-insulator-semiconductor sensor for pH and glucose level measurement NH 3 plasma-treated magnesium doped zinc oxide in biomedical sensors with electrolyte-insulator-semiconductor (EIS) structure for urea and glucose applications Towards a multi-enzyme capacitive field-effect biosensor by comparative study of drop-coating and nano-spotting technique High-k Dy 2 TiO 5 electrolyte-insulator-semiconductor urea biosensors High-k Tm 2 Ti 2 O 7 electrolyte-insulator semiconductor creatinine biosensor Studying the spatially resolved immobilization of enzymes on a capacitive field-effect structure by means of nano-spotting Formaldehyde-sensitive sensor based on recombinant formaldehyde dehydrogenase using capacitance versus voltage measurements Novel applications of silicon and porous silicon based EISCAP biosensors Capacitive field-effect biosensor applied for the detection of acetoin in alcoholic beverages and fermentation broths Fabrication techniques for an arrayed EIS biosensor An array of field-effect nanoplate SOI capacitors for (bio-)chemical sensing An array of on-chip integrated, individually addressable capacitive field-effect sensors with control gate: Design and modelling Fabrication of multianalyte CeO 2 nanograin electrolyte-insulator-semiconductor biosensors by using CF 4 plasma treatment Integrating solid-state sensor and microfluidic devices for glucose, urea and creatinine detection based on enzyme-carrying alginate microbeads Multianalyte biosensor based on pH-sensitive ZnO electrolyte-insulator-semiconductor structures Solid-state sensor incorporated in microfluidic chip and magnetic-bead enzyme immobilization approach for creatinine and glucose detection in serum Potential applications of enzymes immobilized on/in nano materials: A review Immobilization strategies to develop enzymatic biosensors Gating capacitive field-effect sensors by the charge of nanoparticle/molecule hybrids Field-effect nanoparticle-based glucose sensor on a chip: Amplification effect of coimmobilized redox species Urease capacitive biosensors using functionalized magnetic nanoparticles for atrazine pesticide detection in environmental samples Alcohol determination via covalent enzyme immobilization on magnetic beads Extended-gate field-effect transistor packed in micro channel for glucose, urea and protein biomarker detection Capacitive electrolyte-insulatorsemiconductor structures functionalized with a polyelectrolyte/enzyme multilayer: New strategy for enhanced field-effect biosensing Penicillin biosensor based on a capacitive field-effect structure functionalized with a dendrimer/carbon nanotube multilayer Controlled immobilisation of active enzymes on the cowpea mosaic virus capsid Virus scaffolds as enzyme nano-carriers Modified TMV particles as beneficial scaffolds to present sensor enzymes The physics of tobacco mosaic virus and virus-based devices in biotechnology From stars to stripes: RNA-directed shaping of plant viral protein templates-structural synthetic virology for smart biohybrid nanostructures Nanoscale device architectures derived from biological assemblies: The case of tobacco mosaic virus and (apo)ferritin An electrochemical sensor for selective TNT sensing based on tobacco mosaic virus-like particle binding agents Tobacco mosaic virus based thin film sensor for detection of volatile organic compounds Integration of genetically modified virus-like-particles with an optical resonator for selective bio-detection Capillary microfluidics-assembled virus-like particle bionanoreceptor interfaces for label-free biosensing Tobacco mosaic virus as enzyme nanocarrier for electrochemical biosensors Penicillin detection by tobacco mosaic virus-assisted colorimetric biosensors Field-effect biosensor using virus particles as scaffolds for enzyme immobilization TMV nanorods with programmed longitudinal domains of differently addressable coat proteins Fast and sensitive colloidal coomassie G-250 staining for proteins in polyacrylamide gels Light-addressable actuator-aensor platform for monitoring and manipulation of pH gradients in microfluidics: A case study with the enzyme penicillinase The emerging role of urease as a general microbial virulence factor Tailoring of analytical performances of urea biosensors using nanomaterials Inducible oxacillin-hydrolyzing penicillinase in aeromonas hydrophila isolated from fish Comparison of the hysteresis of Ta 2 O 5 and Si 3 N 4 pH-sensing insulators Preparation and study on the drift and hysteresis properties of the tin oxide gate ISFET by the sol-gel method Product Information Urease, Type IX. Available online Nanofilm of ZnO nanocrystals/carbon nanotubes as biocompatible layer for enzymatic biosensors in capacitive field-effect devices Enzyme-based logic systems interfaced with signal-responsive materials and electrodes Enzyme-based logic gates and circuits-analytical applications and interfacing with electronics Digital biosensors with built-in logic for biomedical applications-Biosensors based on a biocomputing concept Biosensors with built-in biomolecular logic gates for practical applications Stimulus-responsive nanomaterials containing logic gates for biomedical applications Integration of biomolecular logic gates with field-effect transducers Enzyme logic AND-Reset and OR-Reset gates based on a field-effect electronic transducer modified with multi-enzyme membrane An enzyme-based reversible controlled NOT (CNOT) logic gate operating on a semiconductor transducer Enzyme-based XOR logic gate with electronic transduction of the output signal Realization and properties of biochemical-computing biocatalytic XOR gate based on signal change Dynamic DNA-controlled "stop-and-go" assembly of well-defined protein domains on RNA-scaffolded TMV-like nanotubes The authors gratefully thank H. Iken, D. Rolka, T. Karschuck, and D. Rashid for technical support. The authors declare no conflict of interest. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.