key: cord-0025378-96vxix0x authors: Ren, Huihui; Xu, Tengyan; Liang, Kun; Li, Jiye; Fang, Yu; Li, Fanfan; Chen, Yitong; Zhang, Hongyue; Li, Dingwei; Tang, Yingjie; Wang, Yan; Song, Chunyan; Wang, Huaimin; Zhu, Bowen title: Self-assembled peptides-modified flexible field-effect transistors for tyrosinase detection date: 2021-12-22 journal: iScience DOI: 10.1016/j.isci.2021.103673 sha: 91bd37b2185772590b2dc3aba31c8c0521129652 doc_id: 25378 cord_uid: 96vxix0x Flexible biosensors have received intensive attention for real-time, non-invasive monitoring of cancer biomarkers. Highly sensitive tyrosinase biosensors, which are important for melanoma screening, remained a hurdle. Herein, high-performance tyrosinase-sensing field-effect transistor-based biosensors (bio-FETs) have been successfully achieved by self-assembling nanostructured tetrapeptide tryptophan–valine–phenylalanine–tyrosine (WVFY) on n-type metal oxide transistors. In the presence of target tyrosinase, the phenolic hydroxyl groups in WVFY are rapidly converted to benzoquinone with the consumption of protons, which could be detected potentiometrically by bio-FETs. As a result, the WVFY-modified bio-FETs exhibited an ultra-low detection limit of 1.9 fM and an optimal detection range of 10 fM to 1 nM toward tyrosinase sensing. Furthermore, flexible devices fabricated on ∼2.9-μm-thick polyimide (PI) substrates illustrated robust mechanical flexibility, which could be attached to human skin conformally. These achievements hold promise for wearable melanoma screening and provide designing guidelines for detecting other important cancer biomarkers with bio-FETs. As a type of malignant skin cancer along with high metastasis, recurrence, and mortality, melanoma has become a severe health-threatening problem worldwide (Huang et al., 2013; Sauer et al., 2017; Jang and Atkins, 2013) . The incidence of cutaneous melanoma has risen rapidly over past several decades (Schadendorf et al., 2018; Leiter et al., 2020) . Once tumors have spread to other organs, such skin cancer will quickly become lethal, owing to few treatment options and unsuccessful clinical trials (Schadendorf et al., 2018) . Therefore, it is imperative to diagnose cancer in its early stages considering the rapid metastasis and evolution of melanoma. To detect reliable melanoma biomarkers is one of the most effective strategies for the early diagnosis of cancer. Tyrosinase (TYR) has been proposed as a potential melanoma biomarker, because it overexpresses melanoma cells and accumulates in skin cells, closely related to the degree of malignancy (Welsh and Corrie, 2015; Robert et al., 2015) . For example, high tyrosinase rate was observed in many patients with untreated melanoma at the late stage, while intervention treatment could decrease tyrosinase concentration, demonstrating that monitoring the dynamic change of tyrosinase concentration could provide a method of therapy prognosis to patients and a predictive factor for cancer evolution (Osella-Abate et al., 2003) . The detection of tyrosinase is, therefore, crucial for practical clinical diagnosis and prognosis. Several laboratory-based methodologies have been developed for detecting tyrosinase activity, including fluorometry, colorimetry, spectrophotometry, electrochemistry (Zhan et al., 2018; Liu et al., 2013 Liu et al., , 2017 Ciui et al., 2018; Fan et al., 2021) . Among them, field-effect transistor-based biosensor (bio-FET) has risen as a promising biosensing technology due to its point-of-care testing capabilities with high sensitivity and accuracy, intrinsic amplification, rapid and real-time feedback, and easy integration with complementary metal-oxide-semiconductor (CMOS) processors (Liu et al., 2017; Allen et al., 2007; Kim et al., 2020b) . The specific interactions between target molecules and receptors modified on bio-FETs could alternate the potentials at the liquid/solid interface, resulting in variations in channel conductance. Also, the channel conductance can be modulated by conformational changes after receptors recognition or depended on alterations in surrounding environmental charges when exposure to target molecules (Bay et al., 2019) . In addition, flexible and biocompatible bio-FETs could provide a new platform for continuously monitoring individuals' health in a wearable and continuous way, attracting widespread attention in recent years (Gao et al., 2016; Bariya et al., 2020; Tai et al., 2018; Li et al., 2020 Li et al., , 2021 Kim et al., 2020a) . However, wearable bio-FETs capable of detecting early-stage melanoma remained a challenge (Emaminejad et al., 2017; Koh et al., 2016; Lee et al., 2016) . Considering high mortality in the cancer stage and the colossal challenge of direct detection, it is significant to screen them in the early stage by monitoring biomarkers. Here, we present a wearable tyrosinase-sensing bio-FET based on biocompatible self-assembled nanostructured peptides and n-type metal oxide FETs. The bio-FETs were achieved by solution-processed In 2 O 3 /Al 2 O 3 semiconductor/dielectric interfaces, where In 2 O 3 worked as the active channel and Al 2 O 3 served as a passivation layer for preventing ion and water diffusion . By modifying a biocompatible tetrapeptide of tryptophan-valine-phenylalanine-tyrosine (WVFY) onto the In 2 O 3 /Al 2 O 3 interfaces via self-assembly, the peptide-modified bio-FETs could successfully detect tyrosinase with high sensitivity, exhibiting an ultra-low detection limit of 1.9 fM and an optimal detection range of 10 fM to 1 nM. Furthermore, by fabricating Al 2 O 3 /In 2 O 3 bio-FETs on ultrathin polyimide (PI) substrate ($2.9 mm thick), the devices could be attached to the skin conformally for wearable sensing applications. The results provide a flexible and biocompatible tyrosinase-sensing platform, promising for future melanoma screening applications. The flexible Al 2 O 3 /In 2 O 3 -based bio-FETs with self-assembled nanostructured peptides were fabricated on ultrathin PI substrates ( Figure S1 ). Figure 1A depicts the schematic illustration of the fabrication process (see Method details section). Typically, the ultrathin PI substrates were prepared via spin-coating PI solution on a carrier glass and subsequent annealing. Next, a thin layer of Al 2 O 3 was spin-coated on PI film to serve as the barrier layer. Then, In 2 O 3 precursor solution was spin-coated and annealed in air. Ni/Au electrodes were thermally evaporated to form source/drain (S/D) electrodes. The interdigitated S/D electrodes were utilized to generate strong electric fields and improve the active sensing region, which depicted low current crowding effects (Rim et al., 2015) . Subsequently, another Al 2 O 3 layer was spin-coated to serve as an effective barrier preventing ion and water diffusion to the underneath In 2 O 3 channel. The as-fabricated bio-FETs were then mechanically delaminated from the carrier glass and could be transferred to other soft substrates, for example, polydimethylsiloxane (PDMS), to facilitate operation. Also, the ultrathin devices could be conformably attached to human skin at both flat and stretched states via van der Waals forces ( Figure 1B ). Figures 1C and 1D depict the electrical performance of Al 2 O 3 /In 2 O 3 bio-FETs before and after delamination from carrier glass and being transferred to a PDMS film, which was characterized in 0.13 phosphatebuffered saline (PBS) solution with a standard Ag/AgCl reference electrode as the gate electrode. Compared to the rigid substrates, the transfer characteristics (I ds -V gs , solid lines, Figure 1C ) of the devices exhibited only tiny degradation when transferred to soft PDMS substrates. The corresponding threshold voltage (V th ) shift, current on/off (I on/off ) ratio, mobility (m), and subthreshold swing (SS) change before and after transfer to PDMS are extracted in Table S1 . Furthermore, the liquid-gated Al 2 O 3 /In 2 O 3 bio-FETs demonstrated fast saturation behaviors and stable output currents with distinctive pinch-off characteristics, as shown in Figure 1D . The low operation voltage (<0.5 V) with small leakage current (I gs , open circles, Figure 1C ) benefits to avoid the redox reactions between the electrodes and biomolecules in the liquid environment, improving stability in long-term biomonitoring applications. To characterize the mechanical flexibility and durability of flexible Al 2 O 3 /In 2 O 3 bio-FETs, electrical bending tests under different bending radii and bending strain cycles were carried out. The device structure of an individual flexible bio-FET is shown in Figure 2A . Figure 2B shows a digital photo of the Al 2 O 3 /In 2 O 3 bio-FET under the bending state, where a flexible polyethylene terephthalate (PET, $50 mm thick) substrate was used to support the device to facilitate the bending test. The transfer curves of flexible Al 2 O 3 /In 2 O 3 bio-FETs at the initial state and bending states with radii of $2 and 1 cm are illustrated in Figure 2C , where almost no perceptible changes in electrical performance could be observed. To investigate mechanical ll OPEN ACCESS 2 iScience 25, 103673, January 21, 2022 iScience Article robustness, the devices were bent/released up to 1000 cycles with a bending radius of 1 cm, and no significant performance degradation was observed ( Figure 2D ). The corresponding V th and SS values at different cycles are extracted in Figure 2E . Both SS and V th showed minor variations from 70.9 G 0.22 to 72.8 G 1.4 mV dec À1 and from 0.38 G 0.004 to 0.39 G 0.005 V, respectively. Also, the corresponding variation of m and I on/off ratio were illustrated in Figure S2 , showing minor performance degradation from 0.66 G 0.02 to 0.69 G 0.02 cm 2 V À1 s À1 and from (6.5 G 0.28) 3 10 1 to (7.4 G 0.33) 3 10 1 , respectively. Importantly, the devices exhibited high environmental stability. Figure 2F illustrates the electrical characteristics of flexible Al 2 O 3 /In 2 O 3 bio-FETs measured at different times. The devices showed no obvious performance degradation up to 7 days. Hence, the robust mechanical flexibility and stable electrical performance render the Al 2 O 3 /In 2 O 3 bio-FETs suitable for wearable and portable biosensing applications. To enable the bio-FETs to detect tyrosinase, a significant biomarker for melanoma screening applications, we designed a tetrapeptide of WVFY with excellent self-assembly properties through p-p stacking (Adler-Abramovich et al., 2009; ( Figure 3A ). The WVFY was synthesized by standard Fmoc solidphase peptide synthesis (SPPS) and purified by reverse-phase high-performance liquid chromatography S4 , and S5). Typically, the WVFY film was prepared by dropping peptide solution in 1,1,1,3,3,3-hexafluoro-2-propanol (HFIP) on the bio-FET surface in air. When the peptide film was exposed to water vapor at 60 C for 12 h, the assembled nanostructures spontaneously grew on the film under the interaction of hydrogen bonds (peptide bonds and water molecules) and p-p stacking (between aromatic rings) ( Figure S6 ). Figure 3B shows a top-view scanning electron microscopy (SEM) image of the assembled peptide film, indicating twisted fiber bundles. Next, we used circular dichroism spectra (CD) to characterize the structure of peptide film before and after water vapor treatment. As shown in Figure 1C , no obvious signal peak was found in the reference WVFY film, indicating no assembly occurred. In comparison, with water vapor treatment, the WVFY film showed a significant positive peak at around 202 nm and two negative peaks at around 222 and 235 nm, indicating the b-sheet-like arrangement ( Figure 3C ). Because the phenolic hydroxyl groups on WVFY could be oxidized by tyrosinase into melanin analogs, the WVFY film modified on bio-FETs could serve as the sensitive layer for tyrosinase detection (Lampel et al., 2017) ( Figure S7 ). Figures 4A and S8 illustrate the tyrosinase sensing setup and mechanism. The WVFY peptide film was selfassembled atop the Al 2 O 3 /In 2 O 3 bio-FETs, forming a nanostructured film of $6.6 mm thick ( Figure S9 ). When the WVFY-modified bio-FETs were exposed to tyrosinase, the phenolic hydroxyl groups in WVFY would be oxidized to catechol and finally transferred to benzoquinone with the consumption of protons. The deprotonation of terminal groups on the bio-FETs would cause surface potential changes and result in channel conductance decrease due to electrostatic repulsion, which could be reflected by V th shifts of bio-FETs. In this way, the concentration of tyrosinase could be quantified potentiometrically. To verify the feasibility, we first quantified tyrosinase sensing performance of WVFY-modified Al 2 O 3 /In 2 O 3 bio-FETs fabricated on rigid silicon substrates. As shown in Figure S10A , the transfer curves illustrate iScience Article characteristic positive shifts with an increase in the concentrations of tyrosinase (C TYR ) from 10 fM to 1 nM. As a result, the device showed a linear relationship (n = 3) between the DV th and C TYR on the logarithmic scale ( Figure S10B ). In comparison, the bio-FETs without self-assembled WVFY peptides were also investigated using the same sensing procedure, which showed no obvious V th shifts but only with minor random variations ( Figures S11A and S11B ). Figure S12 illustrates the output characteristics of WVFY-modified bio-FETs when exposed to various C TYR , where the saturation current decreases with the improvement of C TYR corresponding to the depletion of protons. According to the equation of LOD (limit of detection) = 3.3 3 SD of intercept/slop (Wang et al., 2017) , where SD is the standard deviation of intercept, an outstanding LOD of 1.9 fM was extracted (Figures S13A and S13B). A comparison of the sensing performance of tyrosinase activity between our bio-FETs and documented methods (Ciui et al., 2018; Ding et al., 2021; Li et al., 2012 Li et al., , 2018 Liu et al., 2017; Wang et al., 2019; Yan et al., 2019; Yarman, 2018; Zhou et al., 2016) is shown in Table S2 . Compared to previously reported methods for tyrosinase sensing, our bio-FETs are more advantageous in LOD with an acceptable linear range, which could be used as an accurate screening approach to monitor the tyrosinase level. To explore the tyrosinase sensing performance of flexible devices, we further modified nanostructured WVFY peptides on bio-FETs fabricated on PI substrates with the same process. The typical transfer characteristics (I ds -V gs ) of flexible devices at various C TYR from 10 fM to 1 nM at a low drain bias (V ds ) of 0.05 V were plotted in Figure 4B . Still, the transfer curves shifted positively with the increase of C TYR . In addition, the control devices whose surfaces were not functionalized with WVFY peptides were also studied with the same procedure for comparison. In the absence of WVFY peptides, tyrosinase itself was not puissant enough to influence electrostatic interaction on the surface of bio-FETs. Therefore, only minute and random fluctuations in the family of transfer curves were observed ( Figure 4C ). The linear relationships between DV th and C TYR of the flexible bio-FETs with/without WVFY peptides are depicted in Figure 4D . Furthermore, the flexible WVFY-modified bio-FETs exhibited stable and fast saturation with distinctive pinch-off behaviors as shown in Figure 4E . Also, the saturation current decreased gradually with the improvement of C TYR from 10 fM to 1 nM. The result was consistent with the transfer characteristics (Figure 4B ), owing to the depletion of electrons and decreased channel conductivity. Reproducibility is one of the most significant parameters of sensor response, which is strongly related to sensor reliability and consistency (Najafi et al., 2019) . The average reproducibility of about 82% was obtained with the WVFY-modified bio-FET for monitoring various concentrations of tyrosinase ( Figure S14 ). In addition, we also carried out additional experiments to evaluate the durability. The sensitivity of the WVFY-modified bio-FET was tested after the device was stored in air for one week. As shown in Figure S15 , the bio-FET showed the same detection range of 10 fM to 1 nM, when compared to the original device (Figures 4B and 4D) , which demonstrated high durability and environmental stability. iScience Article To further investigate the electrical performance of flexible devices, real-time sensing response was characterized. As shown in Figure 4F , the drain current (I ds ) decreases with the accumulation of tyrosinase from 10 fM to 1 nM. The I ds -time curve was also measured for referenced bio-FETs without WVFY, showing no obvious response ( Figure S16 ). The relationship of the sensing response (1 -I/I 0 ) as a function of C TYR of flexible devices with/without WVFY was illustrated in Figure 4G . The WVFY-modified bio-FETs exhibited a much higher response than the control devices, illustrating the importance of WVFY in tyrosinase detection. In contrast, the increased sensing response in control devices may originate from the current drift. It was reported that the sensitivity is closed related to the transconductance when the V gs was fixed (Kang et al., 2021; Rim, 2020; Chen et al., 2017) . Figure S17A shows the I ds and g m increased with the improvement of V gs , where the maximum g m of 70.0 mS was achieved at V gs = 0.3 V and V ds = 0.05 V. Subsequently, timeresolved response (time-I/I 0 ) of the bio-FETs with or without WVFY modification was characterized in PBS at V gs = 0.3 V and V ds = 0.05 V. As depicted in Figure S17B , an increased I/I 0 response of WVFY-modified bio-FETs was obtained upon 10 fM to 1 nM tyrosinase. In comparison, the bare devices (without WVFY) only exhibited a small current vibration without obvious response ( Figure S17C ). Figure S17D shows the relationship between the sensing response (1 -I/I 0 ) and the tyrosinase concentration on the logarithm scale. It is noted that the devices with WVFY have a maximum sensing response of 0.82 approximately, much higher than the control group (0.09), when the V ds and V gs were fixed at 0.05 and 0.3 V, respectively. As shown in Figure S18 ( Figure S18 is the time extension of Figure S17B ), after the addition of 1 3 10 À9 M tyrosinase, certain amount of PBS solution was injected (at $660 s) to dilute the tyrosinase concentration from 1 3 10 À9 to 1 3 10 À10 M again. The normalized current (I/I 0 ) did not return to the previous value with 1 3 10 À10 M. On the contrary, after dilution from 1 3 10 À9 to 1 3 10 À10 M with PBS, the sensing response didn't change much (saturated), owing to the non-reversible reaction between the phenolic hydroxyl groups and catechol in WVFY. We further demonstrate the application of WVFY peptide-modified bio-FET for living cells detection. Two different kinds of cell lines, B16F10 and HeLa cells were chosen to regard melanoma cells with overexpressed tyrosinase and non-melanoma cancer cells with low tyrosinase expression, respectively (Suzuki et al., 1994; White and Hu, 1977) . Therefore, B16F10 and HeLa cells were selected as the model cell line and the negative control group, respectively. As shown in Figures S19A and S19B, the normalized current (I/I 0 ) decreases with the density of B16F10 cells accumulated from 0.5 3 10 4 to 2 3 10 5 cells mL À1 , owing to the increase of tyrosinase concentration. The real-time sensing response was also measured in HeLa cell lysates, demonstrating no significant sensing signals. The sensing response (1 -I/I 0 ) as a function of cell densities was depicted in Figure S19C , illustrating the feasibility of the bio-FETs in distinguishing these two kinds of cells. We have described flexible tyrosinase-sensing bio-FETs based on biocompatible self-assembled WVFY peptide nanostructures. The devices exhibited outstanding mechanical flexibility and could be conformally attached to the skin for future wearable sensing applications. The sensing response of the devices originated from the specific reactions between tyrosinase and phenolic hydroxyl groups in WVFY, which consume protons and cause the surface potential change on the bio-FET. As a result, the channel conductivity decreased, and V th shifted positively with elevated tyrosinase concentration. Notably, the devices showed an ultralow detection limit of 1.9 fM with a detection range of 10 fM to 1 nM by virtue of the intrinsic amplification of bio-FETs. Taking full advantages of high sensitivity, excellent mechanical flexibility, and low-cost solution-based processibility, the self-assembled peptide-modified bio-FETs offer considerable promise for tyrosinase detection and melanoma screening. In addition, the flexible and biocompatible bio-FETs open new opportunities for future health monitoring in a wearable manner. Still, our strategy of applying peptide-modified bio-FETs could be extended to the detection of other important cancer biomarkers by the exquisite design of functional peptides and FETs, promising for obviating invasive biopsy and delays in related diseases. Our method provides a FET-based sensor platform capable of detecting tyrosinase levels. However, largescale quantitative investigations aimed at tyrosinase detection at different stages of cancer are required to utilize the bio-FETs for the future real-world practice of melanoma detection, as tyrosinase shows different ll OPEN ACCESS iScience 25, 103673, January 21, 2022 7 iScience Article activity levels when patients are at different cancer stages. In addition, the accuracy of this approach after extensive melanoma screening needed to be further verified. Detailed methods are provided in the online version of this paper and include the following: Bariya, M., Li, L., Ghattamaneni, R., Ahn, C.H., Nyein, H.Y.Y., Tai, L.-C., and Javey, A. (2020) . Glove-based sensors for multimodal monitoring of natural sweat. Sci. Adv. 6. eabb8308. https:// doi.org/10.1126/sciadv.abb8308. Bay, H.H., Vo, R., Dai, X., Hsu, H.H., Mo, Z., Cao, S., Li, W., Omenetto, F.G., and Jiang, X. (2019) . Hydrogel gate graphene field-effect transistors as multiplexed biosensors. Nano Lett. 19, 2620-2626. https://doi.org/10.1021/acs.nanolett. 9b00431. Chen, H., Rim, Y.S., Wang, I.C., Li, C., Zhu, B., Sun, M., Goorsky, M.S., He, X., and Yang, Y. (2017) . iScience Article Zhou, J., Shi, W., Li, L., Gong, Q., Wu, X., Li, X., and Ma, H. (2016) . Detection of misdistribution of tyrosinase from melanosomes to lysosomes and its upregulation under psoralen/ultraviolet A with a melanosome-targeting tyrosinase fluorescent probe. WVFY was synthesized by solid phase peptide synthesis (SPPS) using 2-chlorotrityl chloride resin, and the side chains of the corresponding N-Fmoc protected amino acids are appropriately protected by different groups. Firstly, the C-terminal of the first amino acid was conjugated on the resin. Anhydrous N,N'-dimethyl formamide (DMF) containing 20% piperidine was used to remove Fmoc protected group. To couple the next amino acid to the free amino group, O-(Benzotriazol-1-yl)-N,N,N',N'-tetramethyluronium hexafluorophosphate (HBTU) was used as coupling reagent. The growth of the peptide chain was according to the established Fmoc SPPS protocol. After the last coupling step, the sample was washed with DMF for 1 minute (5 mL per gram of resin) for 3 times to remove excess reagents, then washed with DCM for 1 minute (5 mL per gram of resin) for another 3 times. The peptide was cleaved using 95% of trifluoroacetic acid (TFA) with 2.5% of triisopropylsilane (TIS) and 2.5% of H 2 O for 60 minutes. 20 mL per gram of resin of ice-cold diethyl ether was then added to the concentrated cleavage reagent. The resulting precipitate was centrifuged for 10 min at 4 C at 5000 rpm. Afterward the supernatant was decanted, and the resulting solid was dissolved in H 2 O/CH 3 CN (1:1) for HPLC separation using CH 3 CN and H 2 O containing 0.1% of TFA as eluents. WVFY was first dissolved in hexafluoroisopropanol (HFIP) at a concentration of 160 mM at 25 C. After vortex for 1.5 min, the mixtures were allowed to stand for 30 minutes. Then, 20 mL of the solutions were dropped on the surface of FETs. The samples were placed in a vacuum desiccator and vacuumed for 2 minutes to form the peptide film. Finally, the FET was stuck on the top of a chromatographic bottle, then put the bottle in a screw-cap vial involving 3 ml H 2 O. Then, the vial was put into an oven installed with designed temperature (60 C) during the growth of the assemblies. Cells (B16F10 and HeLa) were seeded in cell culture dish with MEM (Minimum Essential Medium)/RPMI (Roswell Park Memorial Institute) 1640 culture medium containing 10% fetal bovine serum and 1% penicillin-streptomycin. After incubation for 24 h, the upper solution was removed and then washed three times with PBS buffer. The remaining cells were ultrasonically cleaved in PBS buffer at 0 C, then the lysates were collected and centrifuged at 14,000 rpm for 15 min at 4 C. The supernatants were collected to detect tyrosinase activities inside of cells. The synthesized compounds were characterized using 1 H NMR (Bruker AVANCE NEO 500 MHz). HPLC was conducted at Agilent 1260 Infinity II Manual Preparative Liquid Chromatography system using a C18 RP ll OPEN ACCESS iScience 25, 103673, January 21, 2022 iScience Article column with CH 3 CN (0.1% of TFA) and water (0.1% of TFA) as the eluents. LC-MS was conducted at the Agilent Infinity Lab LC/MSD system. Circular dichroism (CD) spectra were measured by an Applied Photophysics Ltd (Chirascan V100) system. All the samples were placed into quartz spectrophotometer cell and the wavelength range was from 185 to 300 nm and the step was 1 nm. The resultant CD spectrum was acquired after subtracting the solvent background. The samples for SEM analysis were coated with platinum on a High Vacuum Coating instrument (Leica, EM ACE600), and then characterized by a Regulus 8230 field emission SEM (Hitachi High-tech Corp., Japan) at the acceleration voltage of 3.0 kV. The thickness of PI film was obtained by a surface profile measuring system (Bruker, Dektak XT). Electrical characteristics were performed in dry and liquid environments by a semiconductor parameter analyzer (4200A-SCS, Keithley) and/or source meter (Agilent B2912A). The mobility ( , where C EDL is the electrical double layer capacitance in 0.1 M ionic strength solution (25.52 mF$cm À2 ) (Park et al., 2015) . V ds , V gs , and I ds stand for drain voltage, gate voltage, and drain current, respectively. W and L are channel width and length of the transistors. Origin software were used to compile and analyze data. Error bars stood for means G SD. p < 0.05 was considered to be statistically significant. Selfassembled arrays of peptide nanotubes by vapour deposition Carbon nanotube field-effect-transistor-based biosensors Wearable wireless tyrosinase bandage and microneedle sensors: toward melanoma screening Spectrophotometric assays for sensing tyrosinase activity and their applications Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis Highly recurrent TERT promoter mutations in human melanoma Which drug, and when, for patients with BRAF-mutant melanoma? Ultrasensitive detection of SARS-CoV-2 antibody by graphene field-effect transistors Wireless, continuous monitoring of daily stress and management practice via soft bioelectronics Low-temperature fabrication of high-performance metal oxide thin-film electronics via combustion processing A soft, wearable microfluidic device for the capture, storage, and colorimetric sensing of sweat Polymeric peptide pigments with sequenceencoded properties A graphene-based electrochemical device with thermoresponsive microneedles for diabetes monitoring and therapy Epidemiology of Skin Cancer: Update 2019. Sunlight, Vitamin D and Skin Cancer Flexible and air-stable near-infrared sensors based on solution-processed inorganic-organic hybrid phototransistors Flexible low-power sourcegated transistors with solution-processed metaloxide semiconductors Highly selective fluorescent probe based on hydroxylation of phenylboronic acid pinacol ester for detection of tyrosinase in cells Spectrophotometric detection of tyrosinase activity based on boronic acidfunctionalized gold nanoparticles Colorimetric detection of tyrosinase during the synthesis of kojic acid/silver nanoparticles under illumination Probing biocatalytic transformations with luminescent DNA/silver nanoclusters Remarkable reproducibility and significant sensitivity of ZnO nanoparticles covered by chromium (III) oxide as a hydrogen sulfide gas sensor Tyrosinase expression in the peripheral blood of stage III melanoma patients is associated with a poor prognosis: a clinical follow-up study of 110 patients Sub-0.5 V highly stable aqueous salt gated metal oxide electronics Interface engineering of metal-oxide field-effect transistors for low-drift pH sensing Review of metal oxide semiconductors-based thin-film transistors for point-of-care sensor applications Printable ultrathin metal oxide semiconductor-based conformal biosensors Pembrolizumab versus ipilimumab in advanced melanoma Updated review of prevalence of major risk factors and use of screening tests for cancer in the United States Activation of the tyrosinase gene promoter by neurofibromin Methylxanthine drug monitoring with wearable sweat sensors Electrochemistry-regulated recyclable SERS sensor for sensitive and selective detection of tyrosinase activity Management of BRAF and MEK inhibitor toxicities in patients with metastatic melanoma Characteristics of tyrosinase in B16 melanoma Photofuel cell coupling with redox cycling as a highly sensitive and selective self-powered sensing platform for the detection of tyrosinase activity Development of a molecularly imprinted polymer-based electrochemical sensor for tyrosinase A fluorescent probe for early detection of melanoma and its metastasis by specifically imaging tyrosinase activity in a mouse model NO 3 ) 3 $9H 2 O, 99.997%) and poly(pyromellitic dianhydride-co-4,4 0 -oxydianiline), amic acid solution (PI, 12 wt.%) were purchased from Sigma-Aldrich. 2-methoxyethanol (2-ME, 99.3%, Alfa Aesar), acetylacetone (AcAc, 99%, Alfa Aesar), and ammonium hydroxide (NH 3 $H 2 O, 28%, Alfa Aesar) were used as received. Fmoc-amino acids were obtained from GL Biochem (Shanghai). 2-Cl-trityl chloride resin was obtained from Nankai Resin Co Chemicals, peptides, and recombinant proteins Indium (III) nitrate hydrate (99.999%) Sigma-Aldrich CAS Aluminum (III) nitrate nonahydrate (99.997%) Sigma-Aldrich CAS amic acid solution (12 wt.%) LTD Experimental models: Cell lines HeLa cells Cellcook Biotech Co We acknowledge financial support from the National Natural Science Foundation of China (Grant No. 62174138), and the Westlake Multidisciplinary Research Initiative Center (MRIC) (Grant No. MRIC20200101). We thank Westlake Center for Micro/Nano Fabrication, and the Instrumentation and Service Center for Physical Sciences (ISCPS), and the Instrumentation and Service Center for Molecular Sciences (ISCMS) at Westlake University for the facility support and technical assistance. Data were collected by H.R., T.X., and K.L., and analyzed by H.R., T.X., F.L., H.Z., D.L., Y.T., Y.W., H.W., and B.Z. Figures were prepared by H.R., T.X., J.L., Y.C., and C.S. The manuscript was written by H.R., T.X., H.W, and B.Z. with assistance from all authors. All authors have given approval to the final version of the manuscript. The authors declare no competing interests. Further information and requests for resources and reagents should be directed to and will be fulfilled by the lead contact, Dr. Bowen Zhu (zhubowen@westlake.edu.cn). All materials generated in this study are available from the lead contact without restriction. Data reported in this paper will be shared by the lead contact upon request. This study did not generate/ analyze code. Any additional information required to reanalyze the data reported in this paper is available from the lead contact upon request.