key: cord-0022523-23j9627y authors: Pflüger, Lisa Sophie; Nörz, Dominik; Volz, Tassilo; Giersch, Katja; Giese, Annika; Goldmann, Nora; Glebe, Dieter; Bockmann, Jan-Hendrik; Pfefferle, Susanne; Dandri, Maura; Schulze zur Wiesch, Julian; Lütgehetmann, Marc title: Clinical establishment of a laboratory developed quantitative HDV PCR assay on the cobas6800 high-throughput system date: 2021-08-28 journal: JHEP Rep DOI: 10.1016/j.jhepr.2021.100356 sha: 37b1737a106bc79a0be42215c0a80aecdacd11d0 doc_id: 22523 cord_uid: 23j9627y BACKGROUND & AIMS: Currently available HDV PCR assays are characterized by considerable run-to-run and inter-laboratory variability. Hence, we established a quantitative reverse transcription real-time PCR (RT-qPCR) assay on the open channel of a fully automated PCR platform (cobas6800, Roche) offering improved consistency and reliability. METHODS: A primer/probe-set targeting a highly conserved region upstream of the HDV antigen was adapted for use on the cobas6800. The lower limit of detection (LLOD) was determined using a dilution panel of the HDV WHO standard (n = 21/dilution). Linearity and inclusivity were tested by preparing 10-fold dilution series of cell culture-derived virus (genotype [GT]1-8; n = 5/dilution). Patient samples containing a variety of bloodborne viral pathogens were tested to confirm exclusivity (n = 60). RESULTS: The LLOD of the HDV utility-channel (HDV_UTC) assay was determined as 3.86 IU/ml (95% CI 2.95–5.05 IU/ml) with a linear range from 10–10ˆ8 IU/ml (GT1). Linear relationships were observed for all HDV GTs with slopes ranging from -3.481 to -4.134 cycles/log and R(2) from 0.918 to 0.994. Inter-run and intra-run variability were 0.3 and 0.6 Ct (3xLLOD), respectively. No false-positive results were observed. To evaluate clinical performance, 110 serum samples of anti-HDV-Ab+ patients were analyzed using the HDV_UTC and CE-IVD RoboGene assays. 58/110 and 49/110 samples were concordant positive or negative, respectively (overall agreement 97.3%). Quantitative comparison demonstrated a strong correlation (R(2) 0.8733; 95% CI 0.8914–0.9609; p value <0.0001). CONCLUSION: The use of highly automated, sample-to-result solutions for molecular diagnostics holds many inherent benefits over manual workflows, including improved reliability, reproducibility and dynamic scaling of testing capacity. The assay we established showed excellent analytical and clinical performance, with inclusivity for all HDV GTs and a limit of quantification of 10 IU/ml, making it a sensitive new tool for HDV screening and viral load monitoring. LAY SUMMARY: The hepatitis delta virus (HDV) causes a severe form of inflammation in the liver. We developed a tool for molecular diagnostics, a polymerase chain reaction HDV assay that showed great performance. It can be used to improve diagnosis of HDV, as well as for monitoring treatment responses. The assay allows for quantification of the virus in the tested samples and is performed on a fully automated platform (cobas6800), which provides various benefits including less hands-on time and excellent comparability of test results. RT-qPCR assay with excellent relaibility and reproducibility Master mix (Roche) 1 2 Highlights Establishment and validation of new quantitative HDV PCR assay on a fully automated platform (cobas6800). New assay allows for dynamic scaling of testing capacity and drastically reduces hands-on time as wells as manual steps. Improved reliability and reproducibility of test results. Lower limit of quantification of 10 IU/ml and a linear range from 10 1 -10 8 IU/ml. Inclusivity for all 8 known HDV genotypes. The hepatitis delta virus (HDV) causes a severe form of inflammation in the liver. We developed a tool for molecular diagnostics, a polymerase chain reaction HDV assay that showed great performance. It can be used to improve diagnosis of HDV, as well as for monitoring treatment responses. The assay allows for quantification of the virus in the tested samples and is performed on a fully automated platform (cobas6800), which provides various benefits including less handson time and excellent comparability of test results. Chronic HDV infection (cHDV) is considered the most severe form of viral hepatitis. Studies have shown that high HDV viral loads (>600,000 copies/ml) in non-cirrhotic patients are associated with a higher risk of developing cirrhosis, 1 which combined with the subsequently elevated risk of developing hepatocellular carcinoma and liver decompensation, 2,3 contributes to the high morbidity and mortality associated with cHDV. 4 Moreover, therapeutic options for cHDV are very limited and mainly based on PEGylated interferon alpha. Further, treatment is associated with low response rates (17-47%) 5 and relapse after therapy. 6 However, only in 2020, a synthetic polypeptide derived from the envelope protein of HBV (bulevirtide) was approved as a new treatment option by the European Medicines Agency. The drug inhibits the HBV/HDV entry receptor and was shown to induce a decrease of HDV RNA and normalization of alanine aminotransferase levels 7 opening new therapeutic avenues, but also increasing the clinical need for tools to monitor HDV viral loads and hence treatment responses. On the molecular level, HDV is a negative strand, circular RNA virus with strong self-base-pairing and high genomic diversity between the different genotypes (GT). 8, 9 Both of these features pose a considerable challenge for the diagnostic workflow. Further, many available reverse transcription real-time PCR (RT-qPCR) assays are characterized by considerable run-to-run and inter-laboratory variability. Herein, we describe the establishment of a quantitative RT-qPCR assay on the open channel of a fully automated PCR-platform (cobas6800, Roche). A primer/probe-set targeting a highly conserved region upstream of the HDV antigen coding region 10 was selected and adapted for use on the cobas6800. Primers were modified with 2'-O-methyl bases in their penultimate base to prevent formation of primer dimers and the probe was conjugated to a minor groove binder at the 3'end in order to increase melting temperature and binding stability. Primers and probes were ordered from IDT DNA Technologies (Coralville, USA) and biomers.net GmbH (Ulm, Germany), respectively. Forward primer (5'-CTCCCTTWGCCATCCmGAG -3'; 208 ll; 1000 nM), reverse primer (5'-CTCTTCGGGTCGGCATmGG-3'; 208 ll; 1000 nM), probe (5'-ATGCCCAGGTCGGACCRC5-3'; 10.4 ll; 50 nM) and Utility Channel Master Mix Reagent 2 (UC MMX-R2, 5573.6 ll; Roche) were combined and transferred to a reagent cassette (96 reactions/ cassette). Sequence-specific forward and reverse primers allow for selective amplification of the internal control (included in the UC MMX-R2). The temperature profile and Utility Channel software settings included a predefined uracil-DNA N-glycosylase incubation step, a pre-PCR step (1 cycle: 55 C for 120 s, 60 C for 360 s, 65 C for 240 s), 5 cycles of 95 C for 5 s and 55 C for 30 s (1 st measurement at the end of each cycle) followed by 45 cycles of 91 C for 5 s and 58 C for 25 s (2 nd measurement at the end of each cycle) and finished with a predefined cooling step. The minimum of the relative fluorescence increase was set at 1.25. The lower limit of detection (LLOD) was determined using a 2-fold dilution panel of the HDV world health organization (WHO) standard (PEI-code: 7657/12; n = 21/dilution, 11 ), which included 10 different concentrations, with the highest and lowest concentrations being 1,000 IU/ml and 1.95 IU/ml, respectively, and a set of 21 negative samples. Linearity and inclusivity were tested by preparing 10-fold dilution series of cell culture-derived virus (GT1-8; n = 5/dilution). To determine interand intra-run variability, the dilution panel used to assess LOD Evaluation of the performance of the HDV utility-channel (HDV_UCT) assay included 20 samples of an external quality assessment (EQA) panel for HDV (INSTAND, Düsseldorf, Germany) and serum samples from HDV RNA-positive patients (n = 4) collected at different timepoints. To further assess assay performance and whether HDV PCR-positive patients have been missed due to a lack of sensitivity of the testing method currently in use, the clinical records of admitted patients at our center from 2008 until July of 2020 were screened for anti-HDVantibody status. In total, 170 HDV patients were identified of whom 60 had to be excluded due to a lack of sample material. Of the remaining 110 patients, 1 serum sample each was included in this study. According to the records, 52.7% (58/110) of patients were HDV PCR positive at the time of blood sampling. Serum samples were stored at or below -18 Celsius until use. Samples containing less than 2,000 ll were diluted using negative human plasma (quantitative HBsAg dilution reagent, Abbott) to a total volume of 2,000 ll. All 110 samples were subjected to the new HDV_UCT assay and an already established CE-marked in vitro diagnostics (CE-IVD) assay (Analytik Jena, Jena, Germany). For the HDV_UCT assays, 500 ll of the sample was used. For the CE-IVD assay, 400 ll were utilized using the INSTANT Virus RNA/ DNA Kit (Analytik Jena, Jena, Germany) to extract RNA before administering 5 ll of the eluate following the manufacturer's recommendations. The nucleic acid extraction and RT-qPCR were carried out according to the manufacturer's instructions and the latter was performed on a LightCycler 480 II (Roche, Rotkreuz, Switzerland). The study was approved by the local ethics committee (PV5626). GraphPad Prism version 8 (San Diego, California, USA) was used for statistical analysis. To determine the LLOD a Probit analysis was computed. 12 The graphical abstract was created with BioRender.com. The LLOD of the HDV_UCT assay was determined as 3.86 IU/ml (95% CI 2.95-5.05 IU/ml) at a 95% detection probability with a linear range from 10 1 -10 8 IU/ml (GT1; Fig. 1A,B and Fig. S1 ). A linear relationship was observed for all 8 HDV GTs, with slopes ranging from -3.481 to -4.134 cycles/log and an R 2 range from 0.918-0.994 (Fig. 2) . Inter-run and intra-run variability were 0.3 and 0.6 Ct (3xLLOD), respectively, demonstrating a high assay reproducibility. No cross reactivity in patient serum samples containing a variety of bloodborne viral pathogens (n = 60) was observed. All samples of the qualitative EQA panel were identified correctly (16/16 positive, 4/4 negative). Viral loads in quantitative EQA panel samples available from 2020 (n = 4) were within the expected range (Fig. S1A) . Comparison between results of the HDV_UCT and the CE-IVD assays showed that 58/110 and 49/110 samples were concordant positive or negative, respectively (Fig.1E) . Three samples (all Ct: >35.6) only tested positive on the HDV_UCT assay (overall-agreement: 97.3%). Quantitative comparison demonstrated a strong correlation between assays (R 2 0.873; 95% CI 0.891-0.961; p value <0.0001; Fig. 1C ). The Bland-Altman analysis revealed a mean of the bias of -0.2997 (95% CI 0.6914-1.291). Three datapoints appeared outside of the margins of ± 1.96 standard deviations of the mean (Fig. 1D) . For HDV RNA-positive patient serum samples collected at different time points, both assays demonstrated similar viral load kinetics (Fig. S1B-D) . To improve the options for clinical HDV RNA monitoring, we established a new quantitative HDV RT-qPCR assay with excellent analytical performance and built-in full process control on the cobas6800 system. The use of highly automated, sample-toresult solutions for molecular diagnostics holds many inherent benefits over manual workflows, including improved reliability and reproducibility, as well as dynamic scaling of testing capacity. Moreover, the interpretation of test results is less errorprone since internal controls are included in every run and the quantitative test results are calculated using the Ct value by the laboratory information software based on the corresponding standard curve. Studies on assay performance on the cobas6800 already demonstrated high reliability for other viruses, including HBV, HCV and HIV 13 and showed excellent comparability between testing sites. 14 In line with previous studies, 15 considerably less hands-on time (n = 24; HDV_UCT 20 min; CE-IVD 180 min) and manual steps (HDV_UCT 2; CE-IVD 36, without preparing reagents and data management) are required as nucleic acid extraction, purification, amplification, detection and data transfer to the laboratory information system are fully automated, hence minimizing the potential for errors. The HDV_UCT assay showed excellent analytical and clinical performance. However, these promising results need to be further validated by examining a larger patient cohort. The new assay demonstrated an inclusivity for all 8 currently known GTs and a low limit of quantification of 10 IU/ml. Combined with the excellent comparability of the PCR results, the new assay will be advantageous for HDV screening and viral load monitoring during HDV treatment or clinical therapeutic trials. Abbreviations CE-IVD, CE-marked in vitro diagnostics; cHDV, chronic HDV infection; EQA, external quality assessment; GT, genotypes; HDV_UCT, HDV utilitychannel; LLOD, lower limit of detection; RT-qPCR, reverse transcription quantitative real-time PCR; WHO, world health organization. Financial support High serum levels of HDV RNA are predictors of cirrhosis and liver cancer in patients with chronic hepatitis delta Estimating the global prevalence, disease progression, and clinical outcome of hepatitis delta virus infection The global prevalence of hepatitis D virus infection: systematic review and meta-analysis Influence of hepatitis delta virus infection on morbidity and mortality in compensated cirrhosis type B EASL clinical practice guidelines: management of chronic hepatitis B virus infection Late HDV RNA relapse after peginterferon alpha-based therapy of chronic hepatitis delta Treatment of chronic hepatitis D with the entry inhibitor myrcludex B: first results of a phase Ib/IIa study Structure, sequence and expression of the hepatitis delta (delta) viral genome The hepatitis delta (delta) virus possesses a circular RNA Development and performance of prototype serologic and molecular tests for hepatitis delta infection World Health Organization International Standard for Hepatitis D Virus RNA for Nucleic Acid Amplification Techniques (NAT)-Based Assays, PEI code 7657/12 22.5 × 14 cm. xiv + 318 pp Verification of the Roche cobas PCR 200 ll and 500 ll protocols for the quantification of HIV-1 RNA, HBV DNA and HCV RNA and evaluation with COBAS( ® ) Ampliprep/COBAS( ® ) TaqMan( ® ) assays Pushing beyond specifications: evaluation of linearity and clinical performance of the cobas 6800/8800 SARS-CoV-2 RT-PCR assay for reliable quantification in blood and other materials outside recommendations Clinical evaluation of multiplex RT-PCR assays for the detection of influenza A/B and respiratory syncytial virus using a high throughput system The study was supported by Roche Diagnostics. Marc Lütgehetmann, Maura Dandri, Dieter Glebe, Annika Giese, Nora Goldmann, Jan-Hendrick Bockmann and Julian Schulze zur Wiesch are funded by the German Center for Infection Research (DZIF). Additional funding was provided by the German Research Foundation (DFG), SFB841 to Julian Schulze zur Wiesch, Marc Lütgehetmann, Maura Dandri and Lisa Sophie Pflüger; B08/ SFB1021 and GL595/9-1 to Dieter Glebe. The National Reference Centre for Hepatitis B Viruses and Hepatitis D Viruses is supported by the German Ministry of Health via the Robert Koch Institute. The study was supported by Roche Diagnostics. Marc Lütgehetmann, Maura Dandri, Dieter Glebe, Annika Giese, Nora Goldmann, Jan-Hendrick Bockmann and Julian Schulze zur Wiesch are funded by the German Supplementary data to this article can be found online at https://doi.org/1 0.1016/j.jhepr.2021.100356. Marc Lütgehetmann has received travel expenses and speakers' honoraria (Roche Diagnostics). The other authors declare that they have no conflict of interest.Please refer to the accompanying ICMJE disclosure forms for further details. The data shown in this article are available from the corresponding authors upon a reasonable request. All reagents, antibodies and resources used in this research can be found in the CTAT table.