key: cord-0021629-iefeezj7 authors: Doshi, Siddharth; Banavar, Samhita P.; Flaum, Eliott; Kulkarni, Surendra; Vaidya, Ulhas; Kumar, Shailabh; Chen, Tyler; Bhattacharya, Arnab; Prakash, Manu title: Applying heat and humidity using stove boiled water for decontamination of N95 respirators in low resource settings date: 2021-09-30 journal: PLoS One DOI: 10.1371/journal.pone.0255338 sha: 8453ea0e3c03ece8ad9fd5a734cba9c2de9bde56 doc_id: 21629 cord_uid: iefeezj7 Global shortages of N95 respirators have led to an urgent need of N95 decontamination and reuse methods that are scientifically validated and available world-wide. Although several large scale decontamination methods have been proposed (hydrogen peroxide vapor, UV-C); many of them are not applicable in remote and low-resource settings. Heat with humidity has been demonstrated as a promising decontamination approach, but care must be taken when implementing this method at a grassroots level. Here we present a simple, scalable method to provide controlled humidity and temperature for individual N95 respirators which is easily applicable in low-resource settings. N95 respirators were subjected to moist heat (>50% relative humidity, 65–80°C temperature) for over 30 minutes by placing them in a sealed container immersed in water that had been brought to a rolling boil and removed from heat, and then allowing the containers to sit for over 45 minutes. Filtration efficiency of 0.3–4.99 μm incense particles remained above 97% after 5 treatment cycles across all particle size sub-ranges. This method was then repeated at a higher ambient temperature and humidity in Mumbai, using standard utensils commonly found in South Asia. Similar temperature and humidity profiles were achieved with no degradation in filtration efficiencies after 6 cycles. Higher temperatures (>70°C) and longer treatment times (>40 minutes) were obtained by insulating the outer vessel. We also showed that the same method can be applied for the decontamination of surgical masks. This simple yet reliable method can be performed even without electricity access using any heat source to boil water, from open-flame stoves to solar heating, and provides a low-cost route for N95 decontamination globally applicable in resource-constrained settings. The ongoing COVID-19 pandemic has caused a worldwide shortage of N95 filtering facepiece respirators (FFRs). Many health facilities are rationing N95 FFRs and reusing them with various decontamination protocols. Nations with emerging economies and high population densities (e.g. India, Pakistan, Bangladesh, Brazil, Peru, Ecuador), are expected to face significant public health challenges due to these shortages. Therefore, it is important to develop accessible procedures for decontamination and re-use of N95 FFRs without damaging their fit or filtration efficiency. At a minimum, decontamination requires inactivation of the SARS-CoV-2 virus and maintenance of the fit and filtration efficiency of the N95 respirator. Temperature treatments, which can inactivate viruses through protein denaturation, represent a potential disinfection method [1] . While there is recent data suggesting that SARS-CoV-2 in liquid media can be inactivated by exposure to 70˚C for 5 minutes [2] , it appears that the virus requires a much longer heat-treatment for inactivation on N95 FFR surfaces. A recent, non-peer-reviewed report indicates that application of 70˚C dry heat for 30 min only led to a 1.9-log reduction in viral load of SARS-CoV-2 in an unknown media on N95 fabric [3] , which is below the minimum 3-log inactivation level suggested by the FDA for emergency use authorization for decontamination [4] . Additionally, the media used for inoculation may not be representative of human saliva or mucin, which may falsely increase inactivation rates compared to a realworld scenario [5] [6] [7] [8] . While direct studies involving the effects of heat and humidity on the SARS-CoV-2 virus on N95 FFRs are limited, increased humidity was found to increase inactivation rates of the H1N1 influenza virus at 65˚C on steel surfaces [9] [10] [11] . A non-peerreviewed report found that increasing relative humidity from 13% to 48% increased inactivation of MS2 and Phi6 bacteriophages deposited with PBS on N95 FFRs by >4-log during a 72˚C heat treatment for 30 min [7] . After this 30 min treatment at 72˚C and 48% relative humidity, both viruses were inactivated by over 6-log. MS2 belongs to a class of non-enveloped viruses, which are traditionally considered to be more difficult to inactivate than enveloped viruses such as SARS-CoV-2 [12] . Given this emerging data, it is likely, though unproven, that temperatures of 65˚C and relative humidities of >50% applied for at least 30 minutes will lead to at least 3-log inactivation of SARS-CoV-2 on an N95 FFR. While increasing temperature and humidity may improve viral inactivation, it also carries the risk of damaging N95 fit and filtration. Several N95 FFR models have been shown to fail fit tests after greater than one 121˚C autoclave treatment cycle, indicating that temperature, humidity, and duration of decontamination must be carefully chosen to strike a balance between viral inactivation and N95 performance [13] . Fortunately, many models of N95 FFR have been shown to undergo at least three cycles of elevated temperature (65-85˚C) at greater than 50% relative humidity for 20-30 min while maintaining filtration efficacy and fit [14] [15] [16] , indicating that conditions of moist heat at 65-80˚C, with 50% relative humidity for 30 minutes may be a promising target for decontamination of SARS-CoV-2 on N95 FFRs. However, reliably achieving these heat and humidity conditions in low resource environments is challenging, particularly those without equipment for heating or stable access to electricity. Equipment typically used for surgical sterilisation (e.g. autoclaves) run at temperatures which, as outlined earlier, may result in a loss of fit and filtration efficiency [14, 15] . There is a need to develop methods applicable to low resource contexts that can achieve 65-80˚C temperatures at high humidity (>50%) for over 30 minutes. A method of heating N95 FFRs with moisture to achieve temperatures of 85˚C at 65-80% humidity was recently implemented by placing plastic containers containing N95 FFRs and 500 μL of water into an oven [16] . Implementation of a similar method involving equipment available in low resource settings could have wide applicability. Open flame stoves are widely used for cooking in a range of remote and low resource contexts where access to electricity is limited or intermittent. Using these stoves to achieve heat and humidity conditions could lead to a widely accessible method of decontamination, if validated properly. Boiling water and cooking utensils are commonly used across India, Bangladesh, Pakistan and many other countries to disinfect surgical equipment in clinical settings lacking traditional autoclaves [17] [18] [19] . Regulation of the temperatures inside a cooker or closed cooking vessel could be achieved through the use of high thermal mass elements such as boiling water. Heating of the water to its boiling point at atmospheric pressure provides a reliable way of reaching a set temperature. Due to the high heat capacity of water, heat transfer may occur at slow enough rates to maintain 65-80˚C temperatures and >50% humidity for over 30 minutes after the vessel is removed from the stove. Here we implement a method in which N95 respirators are placed inside containers, which are placed inside larger cooking vessels with water that has been heated to a rolling boil and then removed from heat. We demonstrate that this method can maintain a consistent elevated temperature and humidity as required for decontamination of N95 FFRs using materials available in low-resource settings. This method could be implemented using a wide variety of conditions based on practical considerations and the materials available in a local context. In this work we demonstrate two specific implementations of this protocol. The first, carried out in an air conditioned building (Stanford, CA, USA) uses a 1.65L Pyrex glass container as the FFR holding container, a saturated 5x5cm paper towel with water from a tap as a moisture source to achieve humidity, and a standard 6 quart (5.7L) cooking vessel containing 2L of water as the larger vessel. The second uses a 2L steel container contained within a 5L aluminium vessel and was carried out in a non air-conditioned kitchen environment (Mumbai, MH, India). Specific procedural details are described for each implementation in the sections below. Additionally, we also demonstrate that a similar decontamination process can be used for surgical masks as well (See Supporting Information S1.4 in S1 File for details). 1. A Kimberly Clark N95 respirator was placed in a 1.75 quart (1.65L) Pyrex container (Fig 1a) . 5. The large vessel was covered with a non-airtight lid and allowed to sit with the closed respirator container inside for at least 45 minutes (Fig 1d) . 6. The container was first removed from the large vessel and then opened. The ambient temperature was~22˚C and measured ambient humidity was~37%. Temperature and humidity inside the container were measured with a BME280 sensor (Sparkfun Inc) and logged with an Arduino Uno. They are reported in Fig 2. The minimum temperature target of 65˚C was typically reached after approximately 4-7 minutes and max temperatures of between 75-80˚C were observed. The initial spike in humidity is due to saturation of air with water vapour at low temperatures, followed by a rapid increase in vapour pressure (and hence decrease in relative humidity) upon heating, followed by a gradual increase in humidity as the water in the sealed container evaporates. The rapid drop in humidity occurs when the container is opened at the end of treatment. The minimum humidity target was reached between 10-13 minutes after testing began and consistently reached a maximum of~60%. The variance in data is partly caused by slight differences in time taken to move the container into the large vessel after removing from boil, water volumes, average water temperature at the end of the boil, precise amount of liquid soaked into the towel and in data logging start time. These are likely to occur during normal implementation of this method in any low resource setting. Despite these operating variations, across all trials the humidity and temperature remained simultaneously above the minimum targets (after reaching both the 50% humidity and 65˚C temperature targets) for over 30 minutes. A total treatment time of 45 mins allowed the system to ramp up to the targets and then maintain them for 30 minutes. A simple experimental test rig (Fig 3) and method was used to test the particle filtration efficiency of various materials including N95-grade masks. The setup includes a LightHouse handheld particle counter (Model 3016 IAQ), Incense: Satya Sai Baba Nag Champa 100 gram, and connectors (universal cuff adaptor, teleflex multi-adaptor). The detector measures particles at an airflow rate of 2.83 L/min, and reports particle counts for 4 different size bins: 0.3-0.49 μm, 0.5-0.99 μm, 1.0-2.49 μm, and 2.5-4.99 μm. The incense produces particles of various sizes, including those in the range picked up by the detector (0.3 μm-10 μm). We first measure the number of particles produced by the incense in each of 4 particle size bins. Then, we place the filter on the setup and run the particle counter to measure the number of unfiltered particles (note the incense is moved back the length of the filter mount away from the original position to maintain similar incense emission). To calculate the filtration efficiency for a given particle size range, we calculate the ratio of unfiltered particles in that range to the number of particles produced by the incense in that range, and then subtract from one. The filter efficiencies tested for Kimberly Clark N95 masks after repeated heating cycles, for particles sized 0.3-04.99 μm, are reported below in Fig 4. Filtration efficiency after 5 treatment cycles for particle sizes after 5 treatment cycles are reported in Table 1 . Note the filter efficiency testing is done using a simple in house experimental setup as opposed to the standard testing which typically uses the TSI Automated filter tester 8130A. Future work may involve carrying out a long term study to measure changes in filtration efficiency after treatment cycles following intermittent exposures. It has been shown that intermittent, low-level sodium chloride aerosol loading of N-95 FFR's, over periods of greater than 100 days, resulted in a degradation of filter efficiency, with the efficiency of some manufacturer models dropping below 95% [20] . However, this is not expected to be an issue within the limited 5 treatment cycles investigated in this study. Furthermore, we note from the referenced study that the long term usage was not was not accompanied by a significant increase in breathing resistance, indicating that long term usage does not result in a reduction in airflow [20] . It may be desirable to achieve time and temperature parameters greater than 65˚C for 30 minutes. A change in the dimensions of containers or water volumes used in the system will result in a change in the maximum temperature, and the time spent at that temperature. In particular, the cooling rate of the system is dependent upon the large vessel surface area to water volume ratio where the higher the surface area to water volume ratio, the higher the cooling rate. The effect of increasing boiled water volume to 3L while maintaining vessel sizing is shown in Fig 5a. This test was carried out without FFRs. The maximum achieved temperature increases along with time spent at a given temperature. However, as the vapour pressure is higher at those higher temperatures, it takes longer to reach the required relative humidity, and the humidity reached is lower (10 minutes to achieve 50% for 2L vs 17 minutes for 3L). This could be adjusted by increasing the size or number of soaked folded paper towels included in the FFR container. The time spent simultaneously above a temperature threshold (65˚C or 70˚C) and the humidity threshold of 50% is plotted for 2L and 3L boiling volumes in Fig 5b. Use of 3L boiled water would allow for 70˚C moist heat to be applied for over 35 minutes, or 65˚C moist heat to be applied for almost 60 minutes. It is desirable to consider how this method could be implemented using other materials based on local availability. A variant of this method was implemented using a smaller vessel (a 2.5L Table 1 . Filtering efficiency of particles for different particle sizes after the 5th treatment cycle. As previously discussed, a 5x5cm folded wet paper towel was placed inside the Ziploc bag along with the respirator halves (Fig 6a) and the sensor. Water in the open pressure cooker was brought to a boil and removed from heat. The Ziploc bag, resting in a steel saucepan, was immediately placed in the pressure cooker (Fig 6b) , which was then sealed and allowed to sit for 45 minutes. It was noted during implementation of this procedure that it was challenging to reliably keep the moist towel and associated water droplets separate from the respirator, representing a risk of filtration piece getting wet, which could affect filtration efficiency. While temperature and humidity targets were reached, filtration efficiency of one of the respirator halves was measured to be below 95% after cycle 3 (Fig 7) . This indicates that holding N95 respirators in Ziploc bags while implementing this method may risk reducing filtration efficiency below 95%, though further study is required to gain confidence in this assessment. Most kitchens in India have steel or aluminium vessels typically with fitting metal lids. To verify that this method can be implemented in a variety of global settings, we implement the humid heat decontamination method using a 5L aluminum outer vessel with a 2L stainless steel inner container. We additionally wrap the outer vessel with cloth to retain heat longer. A Venus 4400 N95 respirator, one of the most common brands used in India, was chosen for the experiments. Five cycles were carried out using the configuration shown in Fig 8. Further experimental details are outlined in Supporting Information S1.1 in S1 File. For the 6th cycle, to illustrate the scalability of the method, we used a 4-plate "idli cooker" where 4 respirators could be loaded, one at each level, as shown in Fig 9(a) and 9(b) . As in the previous experiments the 4-plate vessel was kept in a closed container, hence not exposing the masks directly to steam. The use of such steamers may represent a route to greater scalability of the method, as large idli steamers, or other large dumpling steamers commonly found in South East Asia could be configured to hold up to 40-50 respirators (see Supporting Information S1.3 in S1 File). However, as the heat transfer characteristics of a larger system would be different, further work would be required to validate the method specifically for a larger system. The temperature and humidity profiles in the different cycles are shown in Fig 10. In this implementation, measured temperature and humidity exceeded targets for a longer period than the initial implementation described in Figs 1 and 2. Temperature and humidity reached 65˚C within 5-10 minutes and remained above 65˚C for over 40 minutes even in the case without insulation. In most cases the temperature remained above the higher threshold of 70˚C for~30 minutes in the un-insulated case, and~45 minutes in the insulated case. This is expected to be partly due to the higher ambient temperature in Mumbai, and the effect of the insulation. A simple home-built experimental setup for measuring filtration efficiency (see Supporting Information S1 in S1 File. 2 for further details) was used to evaluate the particulate filtration efficiency at 0.3 μm at a flow rate of 10 litres per minute for the India dataset. The data are shown in Fig 11. Filtration efficiency consistently remained above~94% across all trials, showing no degradation in efficiency due to the heat treatment. A similar experiment was carried out to explore the decontamination of surgical masks, the data are presented in the Supporting Information S1.4 in S1 File. Here we demonstrate a highly accessible heat-based protocol for N95 respirator decontamination that subjects them to >65˚C and 50% humidity for over 30 minutes without requiring any advanced instrumentation or electricity. In this protocol, a 2L pot of water is brought to a rolling boil and removed from heat. Then, a Kimberly Clark N95 respirator and a damp paper towel are placed within a sealed container which is immersed in the boiled water for the treatment duration. The minimum and average filtration efficiencies measured after 5 treatment cycles of this method remained above 97% for particles sized 0.3-4.99 μm at a flow rate of 2.83 L/min, indicating filtration efficiency was not affected by this decontamination method. This outcome is consistent with previously reported literature, where many N95 FFR models have been shown to survive treatments of similar heat and humidity conditions without loss of filtration efficiency. Increasing the boiled water volume while maintaining vessel size allows for a longer treatment duration at high temperatures. A second implementation of this protocol was carried out on a Venus 4400 N95 FFR using standard sized utensils commonly found in South Asia. It achieved similar temperatures and humidities (>70˚C at 50-90% humidity for over 30 minutes) and also maintained filtration efficiencies consistently above 94% for 6 treatment cycles. Higher treatment temperatures, such as 70˚C for 30 minutes, or longer treatment durations can be achieved by increasing the boiled water volume while maintaining vessel size, or by insulation of the vessel. We also show that the same method can be applied for the decontamination of surgical masks. We reiterate that although these protocols were not tested on FFRs inoculated with SARS-CoV-2, they either use disinfection precedents set by either SARS-CoV or recent data based on inactivation of SARS-CoV-2 on FFR surfaces. While further work is still required to ensure viral inactivation of SARS-CoV-2 on FFRs under the given conditions, we believe this simple method can be easily modified to address any changes in guidelines for required heat and humidity. We believe this simple method is robust to variations that may be encountered in individual settings, and provides a reliable, low-cost route for N95 decontamination applicable in resource-constrained settings anywhere in the world. Supporting information S1 File. (PDF) Guideline for disinfection and sterilization in healthcare facilities Stability of SARS-CoV-2 in different environmental conditions Assessment of N95 respirator decontamination and re-use for SARS-CoV-2. medRxiv. Advance online publication Enforcement Policy for Face Masks and Respirators During the Coronavirus Disease (COVID-19) Public Health Emergency (Revised) Inactivation of the coronavirus that induces severe acute respiratory syndrome, SARS-CoV Evaluation of inactivation methods for severe acute respiratory syndrome coronavirus in noncellular blood product Humidity and deposition solution play a critical role in virus inactivation by heat treatment on N95 respirators. medRxiv. Advance online publication Validation of N95 filtering facepiece respirator decontamination methods available at a large university hospital. medRxiv. Advance online publication A pandemic influenza preparedness study: use of energetic methods to decontaminate filtering facepiece respirators contaminated with H1N1 aerosols and droplets Effectiveness of Three Decontamination Treatments against Influenza Virus Applied to Filtering Facepiece Respirators Role of Absolute Humidity in the Inactivation of Influenza Viruses on Stainless Steel Surfaces at Elevated Temperatures Recommendations for Sponsors Requesting EUAs for Decontamination and Bioburden Reduction Systems for Surgical Masks and Respirators During the Coronavirus Disease 2019 (COVID19) Public Health Emergency N95 Mask Decontamination using Standard Hospital Sterilization Technologies. medRxiv. Advance online publication Can N95 Respirators Be Reused after Disinfection? How Many Times Is the fit of N95 facial masks effected by disinfection? A study of heat and UV disinfection methods using the OSHA protocol fit test A scalable method of applying heat and humidity for decontamination of N95 respirators during the COVID-19 crisis Sterilization of Instruments in Isolated Hospitals Sterilization of surgical instruments Limited sterile processing capabilities for safe surgery in low-income and middleincome countries: experience in the Republic of Congo, Madagascar and Benin Electrostatic N-95 respirator filter media efficiency degradation resulting from intermittent sodium chloride aerosol exposure The methods used in this study were informed by insights from summary reports on heatbased decontamination of N95 masks put out by the N95 decon consortium (www. N95DECON.org). SK, UV and AB thank the Soft Matter Lab at TIFR and Emroj Hossain for help with the development of the PFE test setup.