key: cord-0021335-kh01d1mp authors: Cannata, Giulia; Mariotti Zani, Elena; Argentiero, Alberto; Caminiti, Caterina; Perrone, Serafina; Esposito, Susanna title: TEG(®) and ROTEM(®) Traces: Clinical Applications of Viscoelastic Coagulation Monitoring in Neonatal Intensive Care Unit date: 2021-09-07 journal: Diagnostics (Basel) DOI: 10.3390/diagnostics11091642 sha: ee052b4ee93fd31d7463de3bb59a985261d45685 doc_id: 21335 cord_uid: kh01d1mp The concentration of the majority of hemostatic proteins differs considerably in early life, especially in neonates compared to adulthood. Knowledge of the concept of developmental hemostasis is an essential prerequisite for the proper interpretation of conventional coagulation tests (CCT) and is critical to ensure the optimal diagnosis and treatment of hemorrhagic and thrombotic diseases in neonatal age. Viscoelastic tests (VETs) provide a point-of-care, real-time, global, and dynamic assessment of the mechanical properties of the coagulation system with the examination of both cellular and plasma protein contributions to the initiation, formation, and lysis of clots. In this work, we provide a narrative review of the basic principles of VETs and summarize current evidence regarding the two most studied point-of-care VETs, thromboelastography (TEG(®)) and rotational thromboelastometry (ROTEM(®)), in the field of neonatal care. A literature analysis shows that viscoelastic hemostatic monitoring appears to be a useful additive technique to CCT, allowing targeted therapy to be delivered quickly. These tools may allow researchers to determine the neonatal coagulation profile and detect neonatal patients at risk for postoperative bleeding, coagulation abnormalities in neonatal sepsis, and other bleeding events in a timely manner, guiding transfusion therapies using the goal-oriented transfusion algorithm. However, diagnosis and treatment algorithms incorporating VETs for neonatal patients in a variety of clinical situations should be developed and applied to improve clinical outcomes. Further studies should be performed to make routinary diagnostic and therapeutic application possible for the neonatal population. Thromboelastography (TEG ® ) and rotational thromboelastometry (ROTEM ® ) are viscoelastic coagulation tests that quantify the process of clot formation and degradation. They both rapidly provide a graphical representation of the hemostatic process, including the interactions of platelets and fibrinolytic agents, leading to prompt and accurate therapeutic management [1] [2] [3] [4] [5] . The conventional coagulation test (CTT) performed in plasma including prothrombin time (PT), activated partial thromboplastin time (aPTT), thrombin time (TT), and fibrinogen partially reflects the interactions of all these elements, which contribute to clot formation and disruption; for this reason, the test does not seem to provide reliable information to guide transfusion or coagulation factor therapies [6] [7] [8] [9] [10] . A Neonatal Intensive Care Unit (NICU) patient often requires challenging management due to the underlying disease and needs swift hemostatic monitoring for the pre-vention of hemorrhagic and thrombotic complications. TEG ® and ROTEM ® constitute a simple method to evaluate the hemostatic profile in vivo with a very small sample of blood volume, representing a promising alternative to monitor coagulation in the neonatal population, although available TEG ® and ROTEM ® neonatal reference values are still very limited [11] [12] [13] . In this narrative review of the literature, we analyze the basic principles of viscoelastic hemostatic assays (VHAs) and current evidence-based clinical applications of TEG ® and ROTEM ® in the field of NICU. This review was carried out by the Department of Pediatrics in Parma. Systematic searches were performed in PubMed, Embase, Cochrane Library, Scopus, Google Scholar, and ClinicalTrials.gov up to 10 May 2021. Language was restricted to English. Search terms included viscoelastic testing (VET), thromboelastography (TEG ® ), and rotational thromboelastometry (ROTEM ® ) in combination with neonate, newborn, infant, and preterm infant. Case reports, case series, original research studies, review articles, letters to the editor, randomized controlled trials (RCTs), non-RCTs, and cohort studies (prospective or retrospective) were included. TEG ® is performed with a 360 µL whole blood sample (TEG ® 5000 Thrombelastograph ® Hemostasis Analyzer system). The blood sample is placed in a heated (37 • C) cylindrical cup that rotates with an angle of 4 • 45 and a cycle duration of 10 s, simulating venous flow. A pin is immersed into the blood sample and suspended by a torsion wire connected to a mechanical-electrical transducer and subsequent data-processing unit ( Figure 1 ) [14] [15] [16] [17] . Diagnostics 2021, 11, x FOR PEER REVIEW 2 of 18 A Neonatal Intensive Care Unit (NICU) patient often requires challenging management due to the underlying disease and needs swift hemostatic monitoring for the prevention of hemorrhagic and thrombotic complications. TEG ® and ROTEM ® constitute a simple method to evaluate the hemostatic profile in vivo with a very small sample of blood volume, representing a promising alternative to monitor coagulation in the neonatal population, although available TEG ® and ROTEM ® neonatal reference values are still very limited [11] [12] [13] . In this narrative review of the literature, we analyze the basic principles of viscoelastic hemostatic assays (VHAs) and current evidence-based clinical applications of TEG ® and ROTEM ® in the field of NICU. This review was carried out by the Department of Pediatrics in Parma. Systematic searches were performed in PubMed, Embase, Cochrane Library, Scopus, Google Scholar, and ClinicalTrials.gov up to 10 May 2021. Language was restricted to English. Search terms included viscoelastic testing (VET), thromboelastography (TEG ® ), and rotational thromboelastometry (ROTEM ® ) in combination with neonate, newborn, infant, and preterm infant. Case reports, case series, original research studies, review articles, letters to the editor, randomized controlled trials (RCTs), non-RCTs, and cohort studies (prospective or retrospective) were included. TEG ® is performed with a 360 μL whole blood sample (TEG ® 5000 Thrombelastograph ® Hemostasis Analyzer system). The blood sample is placed in a heated (37 °C) cylindrical cup that rotates with an angle of 4°45′ and a cycle duration of 10 s, simulating venous flow. A pin is immersed into the blood sample and suspended by a torsion wire connected to a mechanical-electrical transducer and subsequent data-processing unit ( Figure 1 ) [14] [15] [16] [17] . At the onset of each measurement, whole blood behaves as a non-Newtonian fluid, and the viscous shear between the rotating cup and stationary pin results in no torque and absent deflection in the viscoelastic trace. Once a clot starts to form and grows in strength, the whole blood within the rotating cup begins its transition from a viscous to elastic state; the resultant rotational force exerted on the immersed pin causes rotation on its axis, which is tracked and plotted as a deflection from the baseline to maximum amplitude (maximum cloth strength). Once clot lysis begins, the deflection returns back toward the baseline [18, 19] . Results are graphically represented, with the amplitude of motion detected via torsion wire and an electromagnetic transducer plotted against time. A typical TEG ® trace is shown (Figure 2 ). At the onset of each measurement, whole blood behaves as a non-Newtonian fluid, and the viscous shear between the rotating cup and stationary pin results in no torque and absent deflection in the viscoelastic trace. Once a clot starts to form and grows in strength, the whole blood within the rotating cup begins its transition from a viscous to elastic state; the resultant rotational force exerted on the immersed pin causes rotation on its axis, which is tracked and plotted as a deflection from the baseline to maximum amplitude (maximum cloth strength). Once clot lysis begins, the deflection returns back toward the baseline [18, 19] . Results are graphically represented, with the amplitude of motion detected via torsion wire and an electromagnetic transducer plotted against time. A typical TEG ® trace is shown ( Figure 2 ). TEG ® parameters depict a coagulation profile representing the cell-based model of hemostasis, including the stages of initiation, amplification, propagation, and termination through fibrinolysis (Table 1 ) [8] . Different available reagent kits provide additional information on hemostatic pathways (Table 2) [14] [15] [16] [17] 20, 21] . TEG ® parameters depict a coagulation profile representing the cell-based model of hemostasis, including the stages of initiation, amplification, propagation, and termination through fibrinolysis (Table 1 ) [8] . Different available reagent kits provide additional information on hemostatic pathways ( ROTEM ® is performed with a 300 µL citrated whole blood sample (ROTEM ® delta Thromboelastometry systems). The blood sample is placed in a fixed and heated (37 • C) cup; contrary to the TEG ® , the immersed pin makes a 4 • 75 rotation. As the clot builds up, the rotation of the pin is restricted in proportion to the clot strength; pin movement is detected by an optical sensor ( Figure 3 ). The graphical tracing of cloth strength is plotted as movement amplitude relative to the time elapsed from the initiation of the coagulation cascade to fibrinolysis ( Figure 4 ) [16, 17, 22] . VET results are graphically represented, with the amplitude of movement originating from the rotating pin detected via an optical detector plotted against time. A typical ROTEM ® trace is illustrated in Figure 4 . ROTEM ® parameters provide global information about the kinetics of clot formation and dissolution, including the stages of initiation, amplification, propagation, and termination through fibrinolysis (Table 1) [6, 23] . A wide range of assays is available for the functional investigation of the hemostatic and fibrinolytic systems (Table 3) [16, 17, [20] [21] [22] . The graphical tracing of cloth strength is plotted as movement amplitude relative to the time elapsed from the initiation of the coagulation cascade to fibrinolysis ( Figure 4 ) [16, 17, 22] . VET results are graphically represented, with the amplitude of movement originating from the rotating pin detected via an optical detector plotted against time. A typical ROTEM ® trace is illustrated in Figure 4 . ROTEM ® parameters provide global information about the kinetics of clot formation and dissolution, including the stages of initiation, amplification, propagation, and termination through fibrinolysis (Table 1) [6, 23] . A wide range of assays is available for the functional investigation of the hemostatic and fibrinolytic systems (Table 3) [16, 17, [20] [21] [22] . The hemostatic system is a dynamic, evolving system that changes and matures over time from fetal to adult life. Physiological, age-dependent changes of the coagulation system are referred to as developmental hemostasis [24] [25] [26] [27] [28] . Hemostasis is an active, interconnected network of processes balancing the simultaneous and opposing forces of coagulation and anti-coagulation. The holistic cell-based coagulation model involves both cellular and plasma protein components interacting together in the four overlapping phases of initiation, amplification, propagation and termination. Multiple enzymes, cofactors, adhesion molecules, receptors, platelets, and cellular elements including red blood cells, white cells, the endothelium, and extravascular fibroblasts play a key role in this process [6, 29] . Maternal coagulation factors are unable to cross the placental barrier because of their size [30] [31] [32] ; fetal synthesis of coagulation proteins begins at 11 weeks of gestation. Physiological concentrations of coagulation proteins gradually increase with time; levels are lower in premature infants compared to full-term babies and healthy children [25] [26] [27] [33] [34] [35] [36] [37] . Adult plasma levels of most coagulation proteins are reached between a few months of age and 16 years old. Neonatal plasma concentrations of vitamin K-dependent coagulation factors (II, VII, IX, X) and contact factors (XII, XI, high-molecular-weight kininogen and prekallikrein) are about 50% of adult values [24, 25, 37] . However, in contrast, the major inhibitors of the coagulation system antithrombin (AT), protein C (PC), and protein S (PS) show low levels at birth. Thrombin generation and fibrinolytic capacity are reduced ( The neonatal platelet count and mean platelet volume are usually normal or high, reaching adult values within 1 year. However, platelet hyporeactivity has been reported in newborns; potential reasons reported include impaired signal transduction, the deficiency of thromboxane synthesis, and a decreased density of platelet receptors [42] [43] [44] [45] . Overall, a shorter primary hemostasis bleeding time, reflecting an interaction between platelets and damaged endothelial cells to form a platelet plug, was found in healthy neonates compared with adults, which was normalized before the first month of life [46] [47] [48] [49] . Reasons reported are an increased concentration and activity of von Willebrand factor (vWf; i.e., a high-molecular-weight adhesive glycoprotein that plays an essential role in primary hemostasis) potentiating vWF-mediated platelet adhesion and high hematocrit with the presence of large nucleated red cells [24, 37, 50, 51] . Published data about VET in a healthy neonatal population are limited and mainly based on cord blood samples. Recalcified citrate cord blood TEG ® reference intervals were established by Edwards et al. The hemostatic profile of healthy neonates displayed by TEG ® showed a significantly shorter reaction time (R-time) (p-value < 0.001) as opposed to children and adults and higher values for alpha angle (α angle) than for children (p-value < 0.001), with a moderately higher maximum amplitude (MA) as opposed to children (p-value < 0.05). The coagulation index (CI) (CI = −0.2454R + 0.0184K + 0.1655MA − 0.0241a − 5.0220) and the G parameter (G) (shear elastic modulus strength or clot strength) were higher compared to children (p-value < 0.001), while G values were significantly lower compared to adult controls (p-value < 0.001) [52] . NATEM ROTEM ® analysis with an increasing concentration of tissue plasminogen activator (tPA) performed on neonatal cord blood samples by Sidlik and CFT were shorter compared to adults (p-value ≤ 0.002), while alpha angle (α angle) values were higher than those for adults (p-value = 0.002). Increasing concentrations of tissue plasminogen activator (tPA) correlated with a significant decrease in maximum clot firmness (MCF) (p-value = 0.0001) [53] . ROTEM ® was used for the first time to assess clot formation in preterm infants by Strauss et al. [54] . Modified EXTEM ROTEM ® analysis performed on full-term neonatal cord blood samples showed accelerated clot formation kinetics characterized by significantly shorter CT and CFT compared to adult controls (p-value = 0.001). Modified EXTEM ROTEM ® analysis performed on preterm neonatal cord blood samples showed shorter CT (p-value = 0.001) and CFT (p-value = 0.002) compared to adults. Maximum clot firmness (MCF) values were lower than those for full term neonates (p-value = 0.004) and adults (p-value = 0.001). They also showed a significant correlation between CT and maximum clot firmness (MCF) values and gestational age (GA) in preterm newborns (Pearson correlation r = 0.132, p-value = 0.045 and r = 0.259, p-value < 0.001, respectively) [54] . Cvirn et al., using thromboelastometry with an extrinsic activator assay, recorded no difference in coagulation times between cord blood samples and adult controls. Maximum clot firmness (MCF) values were significantly lower in neonates compared to adults, which was probably due to an impaired polymerization of neonatal fibrin [55] . Wiegele et al. have recently published reference ranges in preterm (30 + 0 to 36 + 6 weeks/days) and term neonates (37 + 0 to 39 + 6 weeks/days) delivered by cesarian section for all types of commercially available thrombelastometry tests [56] . Reference values for cord blood TEG ® in full-term healthy newborns were established by Mirabella et al. [57] . Finally, no statistical differences (p < 0.01) were observed by Schott et al. in TEG measurements (reaction time, kinetics time, α angle, maximum amplitude) between healthy neonates following vaginal and cesarean section deliveries [58] . Table 5 summarizes the main TEG®/ROTEM®results obtained in the neonatal population. Scarce data exist about the hemostatic profiles assessed for peripheral blood samples with VETs in healthy neonates. Caution must be taken when interpreting experimental data due to the lack of homogeneity in commercially available viscoelastic hemostatic assays and the reagents used, the small size samples, and the variability of age ranges included in studies. Motta et al. established reference values for citrated-native whole blood TEG ® assay in 65 healthy preterm neonates (gestational age at birth < 37 weeks). Blood samples were collected within 36 h after birth; exclusion criteria were the following: death within 7 days of life, early-onset sepsis confirmed by a positive blood culture, suspected sepsis with raised C-reactive protein defined as values >10 mg/L, presence of bleeding (intracranial, gastrointestinal, cutaneous, and pulmonary hemorrhage), platelet count < 100 × 109 /L. Reference intervals were established in early-preterm neonates (gestational age at birth < 32 weeks; 32 neonates included in the group) and moderate/late-preterm neonates (gestational age at birth from 32 to 37 weeks; 33 neonates included in the group). Similar TEG ® parameters values were found comparing early-preterm neonates with moderate/late-preterm neonates: reaction time (R-time) (p-value = 0.840), kinetics time (K-time) (p-value = 0.681), alpha angle (α angle) (p-value = 0.830), and maximum amplitude (MA) (p-value = 0.984) parameters did not statistically differ among the two groups. The hemostatic profile of healthy moderate/late-preterm neonates displayed by TEG ® showed a lower coagulation index (CI) (p-value = 0.847) and higher values of the G parameter (G) (shear elastic modulus strength, or clot strength) (p-value = 0.984). TEG ® analysis performed on early-preterm neonatal blood samples showed statistically significant higher values of lysis at 30 min (LY30) compared to moderate/late-preterm neonates (p-value = 0.013) [59] . Native-blood TEG ® analysis was used to assess clot formation in 237 healthy pediatric patients of less than 2 years of age undergoing elective non-cardiac surgery by Miller et al.; a total of 37 term infants were enrolled. Exclusion criteria were the use of drugs known to interfere with blood coagulation, which is a previously diagnosed congenital heart disease and ongoing acute systemic disease. Blood samples were drawn after the induction of general anesthesia. Reaction time (R-time), kinetics time (K-time), alpha angle (α angle), and amplitude of TEG ® tracing 60 min after maximum amplitude (A-60) parameters were measured from the TEG ® tracing. Neonatal TEG ® values did not statistically differ from those of older patients (aged from 1 to 24 months). When compared to adult control subjects, native-blood TEG ® analysis performed on pediatric patients of less than 12 months of age described an accelerated initiation (R-time) and propagation of coagulation (K-time, α angle) along with increased clot strength (MA) [60] . Sewell et al. established reference ranges for citrated-modified and heparinasemodified TEG ® assay after kaolin activation in 30 term newborns admitted to the NICU and diagnosed with known or suspected congenital surgical anomalies, neurosurgical diseases, urological abnormities, and non-critical congenital heart diseases. Exclusion criteria were a personal or family history of bleeding disorders and known or suspected major chromosomal anomaly [61] . Sokou et al. established reference values for peripheral arterial whole blood EXTEM ROTEM ® assay in 198 full-term neonates (gestational age at birth ≥ 37 weeks) and 84 preterm newborns (gestational age at birth < 37 weeks). Exclusion criteria were a personal history of perinatal blood loss, birth asphyxia or perinatal stress, a known or suspected major chromosomal anomaly, a family history of bleeding disorders, and ongoing septicemia. Comparable EXTEM ROTEM ® parameters values were found comparing full-term neonates with preterm newborns: CT (p-value = 0.19), CFT (p-value = 0.06), alpha angle (α angle) (p-value = 0.45), amplitude 10 min after CT (A10) (p-value = 0.90), amplitude 20 min after CT (A20) (p-value = 0.15), and MCF (p-value = 0.17) parameters did not statistically differ among the two groups. An enhanced fibrinolytic activity in preterm neonates was observed: lysis at 60 min (LY60) values were significantly lower than those for full-term neonates (p-value = 0.006). The authors also demonstrated a significant correlation between lysis at 60 min (LY60) and GA and birth weight in preterm newborns (Spearman correlation coefficient ρ =−0.3763, p-value = 0.002 and Spearman correlation coefficient ρ =−0.2988, p-value = 0.016, respectively) [62] . EXTEM ROTEM ® analysis was used for the first time by Sokou et al. to assess clot formation in small for gestational age (SGA) neonates in comparison to appropriate for gestational age weight (AGA) neonates. SGA newborns are defined as having a birth weight of less than the 10th percentile for gestational age. A total of 93 neonates were enrolled in the study: the SGA group included 23 full-term newborns and 22 preterm newborns, while the AGA group included 23 full-term newborns and 25 preterm newborns. Exclusion criteria included a personal history of significant bleeding events, transfusion of blood, platelets, fresh frozen plasma or cryoprecipitate, a family history of bleeding disorders, birth asphyxia, moderate or severe acidemia, ongoing septicemia, and a known or suspected major chromosomal anomaly. Comparable EXTEM ROTEM ® parameters values were found comparing small for gestational age (SGA) neonates with appropriate for gestational age weight (AGA) neonates: CT (p-value = 0.36), CFT (p-value = 0.29), amplitude 5 min after CT (A5) (p-value = 0.45), amplitude 10 min after CT (A10) (p-value = 0.43), alpha angle (α angle) (p-value = 0.18), MCF (p-value = 0.54), and lysis at 60 min (LY60) (p-value = 0.21) values did not statistically differ among the two groups. Moreover, no significant statistical differences were noticed in any of the EXTEM ROTEM ® parameters with regard to delivery time when comparing SGA full-term neonates to AGA full-term neonates and SGA preterm neonates to AGA preterm neonates [63] . TEG ® and ROTEM ® have been extensively studied and applied in adult patients, especially in major bleeding related to severe trauma and surgical fields, such as liver transplantation as well as cardiac and obstetric procedures. Their use has been largely considered in the management of trauma-induced coagulopathy because of their great precision to identify the lacking element involved in the hemostasis reaction [66] [67] [68] [69] [70] [71] [72] [73] [74] [75] [76] [77] [78] . Neonatal data are very limited because of the scarcity of viscoelastic reference values, the extreme variability of the NICU patients, and the differences in coagulation status, which is influenced by gestational age, weight, and maturation of hepatic functions [24] [25] [26] [27] [28] [33] [34] [35] [36] [37] . TEG ® and ROTEM ® clinical applications in children have been almost entirely anesthesiology-driven, especially in newborn candidates for surgery, including cardiac procedures and ExtraCorporeal Membrane Oxygenation (ECMO) [79, 80] . With regard to neonatology, the main application areas include neonatal sepsis, intraventricular hemorrhage (IVH), hypoxic-ischemic encephalopathy (HIE), and newborn candidates for therapeutic hypothermia [11, 12] . Thromboelastometry is an efficient tool to assess the hemostatic efficacy of fibrinogen infusions in patients with quantitative fibrinogen disorders undergoing major surgical procedures [81] . Most patients with hypofibrinogenemia are asymptomatic, while others show signs of severe bleeding events (e.g., umbilical cord bleeding, intracranial hemorrhage, splenic rupture, or bleeding episodes induced by trauma or surgery [82] [83] [84] ). A recent study conducted by Sirmuda et al. observed a reduced level of fibrinogen in a 6-year-old boy with delayed bleeding after 6 h of cleft lip surgery and a 34-year-old woman diagnosed with mild hypofibrinogenemia detected in the first trimester of her second pregnancy with further complications due to retroplacental hematoma. Laboratory analysis confirmed mild hyperfibrinolysis, and in both cases, a new nonsense mutation in the FGB gene was detected, leading to mild hypofibrinogenemia [23] . Thromboelastometric evaluation of coagulation in children undergoing cardiac surgery is associated with a reduction in postoperative blood loss and transfusion requirements and shorter hospitalization [85, 86] . Neonates undergoing CPB are at an increased risk of bleeding events due to their immature coagulation system. The prospective observational study conducted by Scott et al. on 44 NICU patients undergoing cardiopulmonary bypass (CBP) concluded that ROTEM ® can rapidly detect thrombocytopenia and hypofibrinogenemia and consequently implement measures that may improve peri-CPB hemostasis and minimize transfusion-related complications such as venous thromboembolism and transfusion-related acute lung injury. EXTEM A10 can predict thrombocytopenia and FIBTEM A10 can predict hypofibrinogenemia in neonates undergoing CPB. The diminished clot strength in neonatal age is probably the result of the impaired polymerization properties of the neonatal fibrin and the increased fibrinolytic activity due to the increased tissue-plasminogen activator (tPA) levels and the reduced levels of the fibrinolysis inhibitors plasminogen activator inhibitor (PAI) and α2-antiplasmin [87] . Moreover, in surgical settings, the administration of fresh frozen plasma (FFP) is a realistic alternative to normalize hemostasis [88] , and fibrinogen replacement could be guided by MCF values using rotational thromboelastometry [89] . TEG ® and ROTEM ® may detect coagulation status abnormalities in newborns with sepsis. The study conducted by Sakou et al. on 91 septic neonates concluded that hypercoagulation, but especially hypocoagulation, represents an early finding of sepsis and seems to be the main coagulopathy associated with neonatal sepsis [90] . The explanations of this phenomenon lie in the evidence that vitamin K-dependent factors are reduced to 50% of the normal value reached in adults. Moreover, neonatal platelets have a decreased granule secretion and expression of fibrinogen-binding sites [47] . All these factors play a role in the promotion of a pro-hemorrhagic status, especially in sick neonates. The degree of the disruption of the hemostatic balance might be correlated with the gravity of sepsis [90] . Nevertheless, limited data are available about the role of viscoelastic tests as an indicator of early neonatal sepsis, and further studies should be performed. Perinatal hypoxia and asphyxia represent a serious issue that neonatologists are confronted with in clinical practice; one of the dysfunctions caused by the interruption of fetal blood flow and impairment of gas exchange is the derangement of the hemostatic balance, predisposing to hemorrhage and/or thrombosis [91] . A recent study conducted by Konstantinidi et al. evaluated the hemostatic profiles of 164 neonates with perinatal hypoxia and compared them to healthy controls using viscoelastic whole blood tests (EXTEM ROTEM ® analysis). The authors demonstrated a hypocoagulable profile in neonates with perinatal asphyxia or fetal distress compared to healthy neonates. EXTEM ROTEM ® analysis performed on hypoxic neonates showed prolonged CT and CFT compared to healthy neonates (p-value < 0.001); alpha angle (α angle) and maximum clot firmness (MCF) values were lower than those of healthy newborns (p-value < 0.001). EXTEM ROTEM ® analysis performed on asphyxiated neonates showed a significantly prolonged CT (p-value = 0.005) and CFT (p-value = 0.003) and reduced alpha angle (α angle) (p-value = 0.028) compared with newborns with fetal distress. These findings indicate that this test potentially may be used to explore coagulation in hypoxic neonates further [92] . In addition, a hypoxic event and the subsequent reperfusion injury can result in a thromboxane and thrombopoietin release, affecting platelet function [91] . A study conducted by Pakvasa et al. on 98 neonates with hypoxic-ischemic encephalopathy (HIE) measured PT, platelet count, and fibrinogen concentration. Initial hemostatic dysfunction was defined by the authors as one or more of the following: PT ≥ 18 s, platelet count <100 × 103 /µL, or fibrinogen concentration < 150 mg/dL. The prevalence of initial hemostatic dysfunction was 69% (95% CI 59% to 78%); 27 neonates (28%; 95% CI 19% to 38%) had abnormal bleeding events, 56 (57%) received at least one blood product transfusion, and three neonates died from bleeding complications [93] . These findings suggest that hemostatic dysfunction is prevalent and associated with an increased risk of bleeding and high transfusion burden, and the assessment of initial hemostatic profiles should be routinely performed in all newborns with moderate-to-severe HIE. In conclusion, in the light of all the above, viscoelastic tests may be useful tools to determine the neonatal coagulation profile and detect neonatal patients at risk of developing postoperative bleeding, coagulation abnormalities in neonatal sepsis and other bleeding events in a timely manner, guiding transfusion therapies using the goal-oriented transfusion algorithm. Table 6 shows the main TEG®/ROTEM®results obtained in the sick neonatal population. Hemostasis is a dynamic process that changes and matures over time from fetal to adult life. The concept of developmental hemostasis is critical to ensure the optimal diagnosis and treatment of hemorrhagic and thrombotic diseases in pediatric patients, especially in neonates [24] [25] [26] [27] [28] . Healthy neonates are born with an apparent combined deficiency in plasma coagulation factors, natural inhibitors of hemostasis, and components of the fibrinolytic system directly dependent on gestational age, birth weight, and maturation of hepatic function. Neonatal plasma concentrations of vitamin K-dependent coagulation factors (II, VII, IX, X) and contact factors (XII, XI, high-molecular-weight kininogen and prekallikrein) are about 50% of adult values and reach adult plasma levels between a few months of age and 16 years old [24, 25, 37] . Nevertheless, the hemostatic system is functionally balanced with no tendency toward coagulopathy or prothrombosis [26, [38] [39] [40] [41] [46] [47] [48] [49] [50] [51] . This delicate balance is overturned in sick neonates, with CCT appearing not to be completely suitable to ensure the optimal diagnosis and treatment of the main neonatal hemorrhagic and thrombotic diseases. The approach to coagulation disorders in newborns in current clinical practice includes standard coagulation tests, such as PT, aPTT, and TT as well as fibrinogen and platelet count. TEG ® and ROTEM ® both provide real-time whole-blood measurements of hemostatic process kinetics, providing a global rapid assessment of clotting, platelet function, and fibrinolysis. Compared to CCT, viscoelastic tests more closely reflect the in vivo hemostatic conditions. The availability of different reagents and simultaneous sample investigation allow a more accurate and specific evaluation of the hemostatic cascade and permit researchers to investigate clot initiation, propagation, stabilization, and dissolution processes separately. Viscoelastic hemostatic monitoring has been rapidly expanding and appears to be a useful additive technique to CCT, allowing targeted therapy to be delivered quickly. These may be useful tools to determine the neonatal coagulation profile and detect neonatal patients at risk of developing postoperative bleeding, coagulation abnormalities in neonatal sepsis, and other bleeding events in a timely manner, guiding transfusion therapies using the goal-oriented transfusion algorithm. However, TEG ® and ROTEM ® reference data are scarce in the neonatal population. Diagnosis and treatment algorithms incorporating VETs for neonatal patients in a variety of clinical situations should be developed and applied to improve clinical outcome. However, most available data come from small-sample-size studies, differing in study designs both in terms of patient inclusion criteria and the whole blood viscoelastic assays performed. Further studies on homogeneous and larger study populations should be performed to make routinary diagnostic and therapeutic application possible for the neonatal population. Author Contributions: G.C. and E.M.Z. co-wrote the manuscript. A.A. and C.C. performed the literature review. S.P. made a substantial scientific contribution. S.E. supervised the project, revised the first draft, and made a substantial scientific contribution. All authors have read and agreed to the published version of the manuscript. Viscoelastic Hemostatic Assays: Moving from the Laboratory to the Site of Care-A Review of Established and Emerging Technologies A new era of thromboelastometry Diagnostics 2021 Rotational thromboelastometry (ROTEM ® ) TEG and ROTEM: Technology and clinical applications Global Hemostasis Testing Thromboelastography: Old Technology A cell-based model of hemostasis Usefulness of standard plasma coagulation tests in the management of perioperative coagulopathic bleeding: Is there any evidence? Viscoelastic testing inside and beyond the operating room Modern Management of Bleeding, Clotting, and Coagulopathy in Trauma Patients: What Is the Role of Viscoelastic Assays? Thromboelastography as a Better Indicator of Hypercoagulable State After Injury Than Prothrombin Time or Activated Partial Thromboplastin Time Clinical Application of Thromboelastography/Thromboelastometry (TEG/TEM) in the Neonatal Population: A Narrative Review Might work in neonatology too? Comparison of clinical phenotype with genetic and laboratory results in 31 patients with congenital dysfibrinogenemia in northern Slovakia Blutgerinnungsstudien mit der Thrombelastographie, einem neuen Untersuchungsverfahren Managing and Monitoring Haemostasis: Viscoelastometric Point-of-Care Testing (ROTEM, TEG and Sonoclot Systems) Coagulation Monitoring: Current Techniques and Clinical Use of Viscoelastic Point-of-Care Coagulation Devices Bark, N. Fibrinopeptides and fibrin gel structure Rheometry for blood coagulation studies Quality Assurance and Quality Control of Thrombelastography and Rotational Thromboelastometry: The UK NEQAS for Blood Coagulation Experience Interchangeability of rotational elastographic instruments and reagents A Novel Nonsense Mutation in FGB (c.1421G>A; p.Trp474Ter) in the Beta Chain of Fibrinogen Causing Hypofibrinogenemia with Bleeding Phenotype Development of the human coagulation system in the full-term infant Development of the human coagulation system in the healthy premature infant Maturation of the hemostatic system during childhood Development of the Hemostatic System in the Neonate and Young Infant Developmental Hemostasis: Relevance to Hemostatic Problems During Childhood How well do we really understand coagulation? Issues Hemost Manag Placental Barrier to Coagulation Factors: Its Relevance to the Coagulation Defect at Birth and to Haemorrhage in the Newborn Maternal and neonatal primary hemostasis The neonatal coagulation system and the vitamin K deficiency bleeding-A mini review Hemostasis in neonates and children: Pitfalls and dilemmas Developmental hemostasis: Age-specific differences in the levels of hemostatic proteins Differences in the mechanism of blood clot formation and nanostructure in infants and children compared with adults Antithrombin III (ATIII) and Heparin Cofactor II (HCII) in Normal Human Fetuses (21st-27th Week) Evolution of blood coagulation activators and inhibitors in the healthy human fetus Comparison of Human Normal, Full-Term, Fetal and Adult Plasminogen by Physical and Chemical Analyses Thrombin Generation in Newborn Plasma Is Critically Dependent on the Concentration of Prothrombin Effects of Antithrombin and Protein C on Thrombin Generation in Newborn and Adult Plasma Platelet function in term and preterm neonates Platelet Function of Newborns as Tested by Cone and Plate(let) Analyzer Correlates with Gestational Age Platelet Hyporeactivity in Very Low Birth Weight Neonates Deficient Thromboxane Synthesis and Response in Platelets from Premature Infants Template bleeding times of 240 neonates born at 24 to 41 weeks gestation Shorter PFA-100 closure times in neonates than in adults: Role of red cells, white cells, platelets and von Willebrand factor The platelet hyporeactivity of extremely low birth weight neonates is age-dependent Platelets in the neonatal period: Developmental differences in platelet production, function, and hemostasis and the potential impact of therapies Nucleated red blood cells in the fetus and newborn Relationship between human development and disappearance of unusually large von Willebrand factor multimers from plasma Parameters of Thromboelastography in Healthy Newborns Assessment of Functional Fibrinolysis in Cord Blood Using Modified Thromboelastography Clot formation of neonates tested by thromboelastography correlates with gestational age Clot Strength: A Comparison Between Cord and Adult Blood by Means of Thrombelastometry Establishing reference ranges of cord blood: Point-of-care hemostatic function assessment in preterm and term neonates Reference values for coagulation assessment in full-term newborns Thromboelastography in term neonates Reference intervals of citrated-native whole blood thromboelastography in premature neonates Functional Maturity of the Coagulation System in Children Thromboelastography in term neonates: An alternative approach to evaluating coagulopathy Reference ranges of thromboelastometry in healthy full-term and pre-term neonates Studying hemostatic profile in small for gestational age neonates-a pilot observational study Does whole blood coagulation analysis reflect developmental haemostasis? Blood Coagul Reference Values of Thrombolastometry Parameters in Healthy Term Neonates Intraoperative Changes in Blood Coagulation and Thrombelastographic Monitoring in Liver Transplantation Utility of viscoelastic coagulation testing in liver surgery: A systematic review Clinical Utility of Viscoelastic Tests of Coagulation (TEG/ROTEM) in Patients with Liver Disease and during Liver Transplantation Viscoelastic testing in cardiac surgery Tracing the Lines: A Review of Viscoelastography for Emergency Medicine Clinicians Thromboelastography (TEG) or thromboelastometry (ROTEM) to monitor haemostatic treatment versus usual care in adults or children with bleeding Systematic review of viscoelastic testing (TEG/ROTEM) in obstetrics and recommendations from the women's SSC of the ISTH Management of Trauma-Induced Coagulopathy with Thrombelastography Thromboelastography and rotational thromboelastometry for the surgical intensivist: A narrative review Thromboelastography (TEG) and rotational thromboelastometry (ROTEM) for trauma-induced coagulopathy in adult trauma patients with bleeding Effect of thromboelastography (TEG ® ) and rotational thromboelastometry (ROTEM ® ) on diagnosis of coagulopathy, transfusion guidance and mortality in trauma: Descriptive systematic review Utility of thromboelastography and/or thromboelastometry in adults with sepsis: A systematic review Viscoelastometric Testing to Assess Hemostasis of COVID-19: A Systematic Review Thromboelastography-guided management of coagulopathy in neonates with congenital diaphragmatic hernia supported by extracorporeal membrane oxygenation Use of Thromboelastography to Predict Thrombotic Complications in Pediatric and Neonatal Extracorporeal Membranous Oxygenation Efficacy and safety of a new human fibrinogen concentrate in patients with congenital fibrinogen deficiency: An interim analysis of a Phase III trial Perioperative management of a severe congenital hypofibrinogenemia with thrombotic phenotype Congenital hypofibrinogenemia in pregnancy Incidence of bleeding symptoms in 100 patients with inherited afibrinogenemia or hypofibrinogenemia Thromboelastometry-guided intraoperative haemostatic management reduces bleeding and red cell transfusion after paediatric cardiac surgery Development of a specific algorithm to guide haemostatic therapy in children undergoing cardiac surgery Rotational Thromboelastometry Rapidly Predicts Thrombocytopenia and Hypofibrinogenemia During Neonatal Cardiopulmonary Bypass Clinical Features and Management of Congenital Fibrinogen Deficiencies Thromboelastometryguided administration of fibrinogen concentrate for the treatment of excessive intraoperative bleeding in thoracoabdominal aortic aneurysm surgery Thromboelastometry for diagnosis of neonatal sepsis-associated coagulopathy: An observational study Hemostasis and Platelet Dysfunction in Asphyxiated Neonates Thromboelastometry Variables in Neonates with Perinatal Hypoxia Observational study of haemostatic dysfunction and bleeding in neonates with hypoxic-ischaemic encephalopathy Is placental blood a reliable source for the evaluation of neonatal hemostasis at birth? Transfusion 2020 Thromboelastographic profiles of healthy very low birthweight infants serially during their first month Establishing a reference range for thromboelastograph parameters in the neonatal period Thromboelastometry (ROTEM ® ) in children: Age-related reference ranges and correlations with standard coagulation tests Heparinase-Modified Thrombelastography in Term and Preterm Neonates Data Availability Statement: Not applicable. The authors declare no conflict of interest.