key: cord-0020762-52rz11zw authors: Bitker, Laurent; Carvalho, Nadja Cristinne; Reidt, Sascha; Schranz, Christoph; Novotni, Dominik; Orkisz, Maciej; Davila Serrano, Eduardo; Revelly, Jean-Pierre; Richard, Jean-Christophe title: Validation of a novel system to assess end-expiratory lung volume and alveolar recruitment in an ARDS model date: 2021-09-10 journal: Intensive Care Med Exp DOI: 10.1186/s40635-021-00410-x sha: 780458a40b9c4caa765f1de7af0691bb11391c1f doc_id: 20762 cord_uid: 52rz11zw BACKGROUND: Personalizing mechanical ventilation requires the development of reliable bedside monitoring techniques. The multiple-breaths nitrogen washin–washout (MBNW) technique is currently available to measure end-expiratory lung volume (EELV(MBNW)), but the precision of the technique may be poor, with percentage errors ranging from 28 to 57%. The primary aim of the study was to evaluate the reliability of a novel MBNW bedside system using fast mainstream sensors to assess EELV in an experimental acute respiratory distress syndrome (ARDS) model, using computed tomography (CT) as the gold standard. The secondary aims of the study were: (1) to evaluate trending ability of the novel system to assess EELV; (2) to evaluate the reliability of estimated alveolar recruitment induced by positive end-expiratory pressure (PEEP) changes computed from EELV(MBNW), using CT as the gold standard. RESULTS: Seven pigs were studied in 6 experimental conditions: at baseline, after experimental ARDS and during a decremental PEEP trial at PEEP 16, 12, 6 and 2 cmH(2)O. EELV was computed at each PEEP step by both the MBNW technique (EELV(MBNW)) and CT (EELV(CT)). Repeatability was assessed by performing replicate measurements. Alveolar recruitment between two consecutive PEEP levels after lung injury was measured with CT (Vrec(CT)), and computed from EELV measurements (Vrec(MBNW)) as ΔEELV minus the product of ΔPEEP by static compliance. EELV(MBNW) and EELV(CT) were significantly correlated (R(2) = 0.97). An acceptable non-constant bias between methods was identified, slightly decreasing toward more negative values as EELV increased. The conversion equation between EELV(MBNW) and EELV(CT) was: EELV(MBNW) = 0.92 × EELV(CT) + 36. The 95% prediction interval of the bias amounted to ± 86 mL and the percentage error between both methods amounted to 13.7%. The median least significant change between repeated measurements amounted to 8% [CI(95%): 4–10%]. EELV(MBNW) adequately tracked EELV(CT) changes over time (concordance rate amounting to 100% [CI(95%): 87%–100%] and angular bias amounting to − 2° ± 10°). Vrec(MBNW) and Vrec(CT) were significantly correlated (R(2) = 0.92). A non-constant bias between methods was identified, slightly increasing toward more positive values as Vrec increased. CONCLUSIONS: We report a new bedside MBNW technique that reliably assesses EELV in an experimental ARDS model with high precision and excellent trending ability. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s40635-021-00410-x. Despite improvement in therapeutic management, acute respiratory distress syndrome (ARDS) mortality remains high in observational studies [1] . Personalizing mechanical ventilation may improve ARDS prognosis [2] , but requires the development of reliable bedside monitoring techniques. End-expiratory lung volume (EELV) monitoring is appealing as it provides information on the size of the baby lung and lung strain, and may help individualize respiratory settings during mechanical ventilation of ARDS patients, by estimating alveolar recruitment related to positive end-expiratory pressure (PEEP) changes [3] . However, this clinical application has not been validated so far, using computed tomography (CT) as the reference technique. The multiple-breaths nitrogen washin-washout (MBNW) technique is currently available on some ICU ventilators, and has been shown reliable to assess EELV in both mechanically ventilated patients at PEEP 5 cmH 2 O [4] and in experimental studies at higher PEEP [5, 6] , using CT as the reference technique to measure EELV. However, the precision of the technique may be poor, with percentage errors ranging from 28 to 57%, and is accompanied with a systematic bias at high PEEP levels identified in one study [6] . Such bias at high PEEP levels was merit of attention of a previous study, where the presence of overinflated lung region and increased alveolar dead space at higher PEEP levels partially explained the substantial difference between EELV measured by CT and MBNW [7] . A new non-invasive MBNW monitor was designed, using fast mainstream sensors enabling continuous and synchronous measurements of proximal flow and O 2 / CO 2 concentrations. We hypothesized that this new system might be more reliable to assess EELV. The aim of the study was primarily to evaluate the reliability of this new non-invasive MBNW monitor to measure EELV at bedside (bias, precision, percentage error, and repeatability) in an experimental model of ARDS using CT as the gold standard. Secondly, the trending ability of the new MBNW monitor and the reliability of estimated alveolar recruitment induced by PEEP changes computed from EELV MBNW measurements were evaluated, using CT as the gold standard. The study was approved by an Institutional Review Board for the care of animal subjects (Comité d'éthique en experimentation animale n°042, APAFIS#21542-2019071909353561), and carried out on 7 female Large White pigs weighing 29 [29, 30] kg, in accordance with Good Laboratory Practice Regulation [8] . Pigs were premedicated with an intramuscular injection of xylazine 2% (0.7 mg kg −1 ), ketamine (17 mg kg −1 ), and droperidol (0.17 mg kg −1 ). Surgical preparation was performed in the supine position under continuous general anesthesia by propofol (7 mg kg h −1 ) and fentanyl (10 µg kg h −1 ). Body temperature was maintained constant at 38 ± 1 °C using a heating pad. A 6.0 tracheal tube was introduced via median surgical tracheotomy, secured hermetically after cuff inflation, and connected to a HAMILTON-C6 ventilator (Hamilton Medical AG, Bonaduz, SWITZERLAND). Mechanical ventilation was initially performed through a heated humidifier (HAMILTON-H900, Hamilton Medical AG, Bonaduz, SWITZERLAND) in volume-controlled mode, constant inspiratory flow, tidal volume (V T ) 10 mL kg −1 , inspired fraction of oxygen (FiO 2 ) 50%, PEEP 5 cm H 2 O and respiratory rate adjusted to achieve normocapnia. Muscle relaxation was obtained with cisatracurium (30 mg h −1 ). The left jugular vein was surgically cannulated with a 3 lm 8.5 Fr catheter (model Arrow # CV-12853, Teleflex, Wayne, PA, USA) for drug administration. A 7-french pulmonary artery catheter (model #131F7, Edwards, Irvine, CA, USA) was advanced through the right external jugular vein to assess central venous pressure, pulmonary artery pressure and cardiac output. The left carotid artery was surgically cannulated with an 8.5 Fr catheter for continuous monitoring of arterial blood pressure. Experimental protocol (Fig. 1) Experimental ARDS was achieved by saline lavage under FiO 2 90%, VT 10 ml.kg −1 and PEEP 5 cm H 2 O. Repeated intra-tracheal instillations of 1000 mL 0.9% sodium chloride warmed at 37 °C was performed, and lung fluid was removed by both gravity and endotracheal suctioning. The procedure aimed to achieve a PaO 2 /FiO 2 ratio < 100 mm Hg 15 min after the last lung lavage. Respiratory rate (RR) could be increased up to 30 breaths per min to maintain pH above 7.20. A recruitment maneuver was then executed by applying a continuous airway pressure of 40 cm H 2 O during 20 s, VT was set to 6 ml. kg −1 , FiO 2 was set to 80%, and a decremental PEEP trial (16, 12, 6 , and 2 cm H 2 O) was performed. PEEP could possibly be increased by 1 cm H 2 O increments at the last two PEEP steps if the peripheral oxygen saturation fell below 88% to maintain the animal alive. The following measurements were successively performed at each protocol step, i.e., at baseline, 15 min after the achievement of experimental ARDS criteria, and after a 10-min stabilization period at each PEEP step ( Fig. 1) : arterial blood gas, hemodynamic measurements, respiratory mechanics assessment, first assessment of EELV by MBNW, assessment of EELV by CT, and ultimately replicate assessment of EELV by MBNW. Arterial blood gases were analyzed with an ABL80 FLEX CO-OX blood gas analyzer (Radiometer, Copenhagen, DENMARK) for oxygen and carbon dioxide partial pressures, pH and lactate. Pressure transducers (Abbott, San Clemente, CA, USA) were positioned at the midchest level and connected to a Philips monitor (M1205A Philips Healthcare, Best, NETHERLAND). Cardiac output was measured in triplicate with the thermodilution technique, using 10 ml saline injected at room temperature. Airway pressure and flow were measured by the HAMILTON proximal flow sensor of the mechanical ventilator. Total PEEP and plateau pressure were measured at the end of 3 s end-expiratory and end-inspiratory pauses, respectively. Peripheral capillary oxygen saturation was measured continuously on the tip of the animal tong. EELV MBNW was determined using a MBNW maneuver, consisting of a change in FiO 2 by 10% and reaching a new steady-state, before resetting the FiO 2 to the original value. A novel mainstream CO 2 /O 2 gas sensor was placed between the proximal flow sensor and the tracheal tube, measuring O 2 /CO 2 partial pressures at a frequency of 100 Hz. Gas flow, airway pressure and O 2 /CO 2 partial pressures were acquired by the ventilator and relayed via a serial connection to a computer running a custom MATLAB program (MathWorks, Natick, MA, USA). The signals were processed online and automatically calculating the breath-by-breath volumetric gas exchange [9] . The cumulative volume of nitrogen washed-out by the increase in FiO 2 was determined by integrating gas flow, O 2 and CO 2 partial pressure signals and subtracting the nitrogen baseline determined during the steady-states. The cumulative volume of nitrogen retained during the wash-in phase was computed accordingly. EELV MBNW was then computed as: EELV MBNW measured during wash-in and wash-out were averaged. To allow meaningful comparisons with EELV CT , instrumental dead-space, i.e., the volume from the O 2 /CO 2 sensor to the tip of the tracheal tube was measured and subtracted from EELV MBNW . The stability of carbon dioxide production (VCO 2 ) and oxygen consumption (VO 2 ) before and after the MBNW maneuver was verified. Repeated EELV MBNW measurements (i.e., two consecutives nitrogen washin-washout) at each PEEP level were carried out in five pigs of the study to assess repeatability of the technique. An estimation of the recruited volume between two consecutive PEEP levels after lung injury was performed with the MBNW technique (Vrec MBNW ) as follows [3] : with n = rank of measurement and EELV assessed with MBNW technique. EELV CT was measured on a Biograph mCT (Siemens, Munich, GERMANY) previously calibrated with the manufacturer phantom, using the following settings: voltage 120 kV, pitch 1.2, current time-product 80 mAs, field-of-view diameter 500 mm, slice thickness 1 mm, Kernel B31f (smooth), matrix size 512 by 512, and pixel size 0.8 mm by 0.8 mm. Lung scanning was performed from apex to base during both end-expiratory and endinspiratory pauses, with clamped tracheal tubes. The lungs were semi-automatically segmented, followed by visual inspection by one author (JCR) with a previously validated in-house software [10] , excluding pleural effusions, hilar and mediastinal structures. Segmented lung volumes were analyzed using MATLAB (MathWorks, Natick, MA, USA). Gas and tissue volume in each lung voxel and end-expiratory lung volume (EELV CT ) were computed from the CT number (HU) according to the following formulae [11] : Voxel gas volume = CTnumber −1000 • voxelvolume for lung voxels with − 1000 ≤ CT number ≤ 0. Voxel gas volume = 0 for lung voxels with CT number > 0. Voxel gas volume = voxel volume for lung voxels with CT number < -1000. Voxel tissue volume = voxel volume-voxel gas volume. EELV CT = k i=1 voxelgasvolume(i) with k = total number of voxels in segmented lung volume at end-expiration. Lung parenchyma was then classified into four compartments, according to CT number [11] : non-inflated (density between + 100 and − 100 Hounsfield units (HU)), poorly inflated (density between − 101 and − 500 HU), normally inflated (density between -501 and -900 HU), and overinflated tissue (density between − 901 and − 1000 HU). The volume of each compartment was measured at end-expiration and end-inspiration. N2fraction . CT-derived recruited volume (Vrec CT in mL) between two consecutive PEEP levels after lung injury was estimated on end-expiratory CT scans as follows [12] : with n = rank of measurement, Vtis PEEPn = tissue volume of the non-aerated compartment at PEEP n at end-expiration, Vtis PEEPn+1 = tissue volume of the non-aerated compartment at PEEP n+1 at end-expiration, Fgas PEEPn = gas fraction of the aerated compartment at PEEP n at end-expiration. Statistical analyses were performed using R for Windows v4.0.2 [13] with the following packages: boot [14, 15] , lme4 [16] , lmerTest [17] , MuMIn [18] , MethComp [19] , and multcomp [20] . A p-value below 0.05 was chosen for statistical significance. Values were expressed as median [1st quartile-3rd quartile] unless otherwise stated. 95% confidence intervals (CI 95% ) of proportions were computed with the Wilson score method. Ventilatory, respiratory, hemodynamic and CT variables were compared across experimental conditions with a linear mixed model, using experimental condition as a factor with fixed effect, and pigs as a factor with random effect to consider multiple measurements. Multiple comparisons were performed with Dunnett's test using baseline measurements as a reference. Repeatability of the MBNW technique was assessed as follows. The coefficient of variation (CV) of EELV MBNW repeated measurements was computed as the standard deviation divided by the mean of the 2 replicates. The precision of replicates for EELV MBNW measurement was computed as 2 × CV. The least significant change (LSC) between repeated measurements was computed as CV × 1.96 × √ 2 [21] . The bias corrected and accelerated bootstrap method with 10,000 replicates was used to compute CI 95% for precision and LSC [22] . EELV MBNW and EELV CT were compared using a linear mixed-effect model, and Bland and Altman representation, using the first EELV MBNW replicate [23] . Bias between methods was computed as EELV MBNW minus EELV CT . Since bias was non-constant and the experimental design involved repeated measurements, limits of agreement and conversion equation between EELV MBNW and EELV CT were computed from the posterior medians of a Monte Carlo chain simulation [24] . Percentage error was computed as 2 × SD Bias /mean EELV CT [25] , with SD Bias = standard deviation of the bias. Similar analyses were performed with Vrec MBNW and Vrec CT . To explore the impact of potential confounding variables on the bias between methods, a linear mixed-effects model was built using ventilatory, respiratory, hemodynamic and CT variables as fixed effects, pigs as factor with random effect [26] , and bias between methods as the dependent variable. Variables with p values < 0.1 in univariate analyses were considered for inclusion in a multivariable model. Model simplification was performed using a backward stepwise algorithm. Ability of the MBNW technique to track directional changes in EELV between experimental conditions was assessed using four-quadrant and polar plots. Concordance rate was defined as the percentage of data points falling into one of the two quadrants of agreement on the four-quadrants plot (i.e., quadrants in which both EELV MBNW and EELV CT have the same directional changes) [27] . Since the four-quadrant plot does not quantify the distance between each data point and the line of identity, a polar plot analysis was performed [27] . Angular bias (difference in calibration between the reference and test method) was computed as the mean angle between all data points and polar axis [27] , and compared to zero using Mann-Whitney U test. Radial limits of agreement (defined as the radial sector containing 95% of the data points, after conversion of negative deflections to positive ones) were computed similarly to the limits of agreement in Bland and Altman analysis [27] . Ten measurements were missing due to various reasons, including premature death of one animal as reported in Additional file 1 and ending up in 32 available measurements. Variables repeatedly studied over time are reported in Table 1 . After experimental lung injury, PaO 2 /FiO 2 , pH and EELV decreased significantly while PaCO 2 , plateau pressure and mean pulmonary artery pressure increased significantly. EELV CT values spanned from 254 to 1200 mL, while EELV MBNW ranged from 298 to 1157 mL. 27 replicate EELV MBNW measurements were obtained. The median precision of repeated EELV MBNW measurements was 6% [CI 95% : 3-7%]. The median LSC between repeated EELV MBNW measurements was 8% [CI 95% : 4-10%]. EELV MBNW and EELV CT were significantly correlated (marginal R 2 = 0.97, p < 0.001, Fig. 2A) , and the regression equation between EELV MBNW and EELV CT , had an intercept of 36 mL (p < 0.001) and a slope of 0.92. A non-constant bias between methods was identified on Bland and Altman plot (Fig. 2B) , with a slight albeit significant trend of more negative values as EELV increased with PEEP. The conversion equation between EELV MBNW and EELV CT was computed as EELV MBNW = 0.92 × EELV CT + 36. The 95% prediction interval of the bias between methods amounted to ± 86 mL, and the percentage error between both methods was 13.7%. Multivariate analysis identified the volume of the overinflated compartment at endexpiration as the only variable independently associated with the bias between methods (Table 2) , with a linear coefficient amounting to -8.7 ± 1.8 (i.e., for each mL of overinflated volume increase, the EELV MBNW is 9 ml lower than the reference EELV value measured with CT). Ability of the MBNW technique to track directional changes in EELV EELV MBNW adequately tracked EELV CT changes over time (Fig. 3) , with a concordance rate amounting to 100% [CI 95% : 87%-100%]. Results of the polar plot analysis are reported in Fig. 4 . The angular bias amounted to -2° ± 10° and was not statistically different from 0 (p = 0.21). Radial limits of agreement amounted to ± 26°. Vrec MBNW and Vrec CT were significantly correlated (marginal R 2 = 0.92, p < 0.001, Fig. 5A ), and the regression equation between EELV MBNW and EELV CT , had an intercept of 0 mL (p < 0.001) and a slope of 1.43. A non-constant bias between methods was identified on Bland and Altman plot, slightly increasing toward more positive values as Vrec increased (Fig. 5B) . The 95% prediction interval of the bias between methods amounted to ± 147 mL. The main findings of the study are the following: (1) the new MBNW monitoring tool is reliable to assess EELV at bedside, with statistically significant trending ability; (2) the precision of the technique is good, with a percentage error against CT amounting to 13.7%, and a LSC between repeated measurements ≤ 10%; (3) estimation of alveolar recruitment from EELV MBNW and measured with CT are highly correlated. As previously reported, EELV assessed by the MBNW technique exhibited a systematic bias, as compared to EELV CT [6] , with a slight underestimation at high values of EELV. Although considered as a "gold standard" for lung volume assessment, it is important to highlight that CT scans accurately measures the "anatomical lung volume", in contrast with MBNW technique, or other gas-dilution technique, where the lung volume measured contributes to ventilation, i.e., "ventilable lung volume" [4] . Thus, in the presence of overinflated lung areas (at high PEEP levels) the "anatomical lung volume" is expected to be different from the "ventilable lung volume" [7, 28] . The current study supports previous results, since the bias observed between the tested methods was independently related to the amount of overinflation at end-expiration, it may be hypothesized that overinflated lung areas did not participate to nitrogen mixing during washin-washout, which may explain EELV underestimation at high PEEP. From the clinical standpoint, the bias observed between techniques in the current study is likely to be of minor relevance if we consider the potential clinical application of the proposed novel technology and the drawbacks of the CT technique. Nevertheless, the upper limit of the CI 95% of the LSC between repeated measurements amounted to 10%, suggesting a good ability to detect relatively small changes in EELV. This was confirmed by a 100% concordance rate and an angular bias not significantly different from 0, suggesting a good trending ability of the new technique. The percentage error of this new technique was the lowest of previously published experimental and clinical studies (ranging from 28 to 57% [4] [5] [6] ). This is presumably related to the use of the fast mainstream sensors, which ensured that the measurements of proximal gas concentrations and gas flow are precisely aligned in time [29] . Signal alignment may be more challenging when a side stream sensor is used, as in a previous study using the same ARDS animal model [6] . One might argue that despite considerable improvement on EELV MBNW accuracy provided by this novel technology, lower PEEP levels were used in the current study. However, the experimental protocol presented here is closer to the clinical setting in terms of applied PEEP levels, improving its extrinsic validity. To our knowledge, the present study is the first to provide concomitant measurements of Vrec MBNW with CT as the reference technique. As proposed by Dellamonica et al. [3] , Vrec MBNW is computed from the difference between change in EELV induced by PEEP, and minimally predicted increase in lung volume between PEEP levels (i.e., the product of respiratory system compliance assessed at lower PEEP and PEEP level change). When the change in EELV is larger than this minimum predicted volume gain, the difference is considered as an estimate of alveolar recruitment. However, this method assumes that compliance is linear over the change in pressure induced by PEEP; an assumption we are unable to confirm with our data. Furthermore, this method combines measurement errors from several devices (ventilator pneumotachograph, ventilator pressure sensors, MBNW technique), questioning the ability of this technique to detect small recruited volumes. Nevertheless, a fair correlation was identified between Vrec MBNW and alveolar recruitment computed from the pressure-volume curves in ARDS patients [3] , but mathematical coupling between both techniques (Vrec MBNW and the pressure volume curves are computed using the ventilator pneumotachograph) could question the validity of this study. Of note, 4 subjects with the highest PEEP levels (16 cm H 2 O) were excluded of Dellamonica et al. study as their EELV MBNW was underestimated [3] , a finding in line with our results. Nevertheless, we observed in the present study a strong correlation with CT measurements, although Vrec MBNW was overestimated at high recruited volume values (a consequence of the underestimation of EELV MBNW ). Whether this technique could reliably detect low and high recruiters by PEEP remains to be determined. As compared to other MBNW technologies [4] [5] [6] [7] , the current monitoring tool exhibited substantially lower percentage error (28% to 57% vs 13.7%) and hence higher precision, and narrower radial limits of agreements (± 51° [6] vs. ± 26°), i.e., higher trending ability. Furthermore, the current monitoring tool is virtually unbiased, provided the conversion equation presented above is used to account for the non-constant bias [30] . Although we did not perform a direct comparison, we hypothesize that an improved accuracy could result from fast mainstream gas concentration measurements able to better Epidemiology, patterns of care, and mortality for patients with acute respiratory distress syndrome in intensive care units in 50 countries Precision medicine in acute respiratory distress syndrome: workshop report and recommendations for future research PEEP-induced changes in lung volume in acute respiratory distress syndrome. Two methods to estimate alveolar recruitment Nitrogen washout/washin, helium dilution and computed tomography in the assessment of end expiratory lung volume Agreement between functional residual capacity estimated via automated gas dilution versus via computed tomography in a pleural effusion model Reliability of the nitrogen washin-washout technique to assess end-expiratory lung volume at variable PEEP and tidal volumes The effect of alveolar dead space on the measurement of end-expiratory lung volume by modified nitrogen wash-out/wash-in in lavage-induced lung injury Part 58-Good Laboratory Practice for Nonclinical Laboratory Studies. US Government Publishing Office; Electronic Code of Federal Regulations Estimation of functional residual capacity at the bedside using standard monitoring equipment: a modified nitrogen washout/washin technique requiring a small change of the inspired oxygen fraction Software for CT-image Analysis to Assist the Choice of Mechanical-Ventilation Settings in Acute Respiratory Distress Syndrome Relationships between lung computed tomographic density, gas exchange, and PEEP in acute respiratory failure Regional Tidal Lung Strain in Mechanically Ventilated Normal Lungs R: A language and environment for statistical computing. R Foundation for statistical Computing Bootstrap Methods and Their Applications Fitting linear mixed-effects models using lme4 lmerTest package: tests in linear mixed effects models MuMIn: Multi-Model Inference MethComp: Functions for analysis of agreement in method comparison studies Simultaneous inference in general parametric models Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies -with specific reference to the measurement of cardiac output Better bootstrap confidence intervals Agreement between methods of measurement with multiple observations per individual Comparing Clinical Measurement Methods: A Practical Guide (Statistics in Practice) A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques Statistical models for assessing agreement in method comparison studies with replicate measurements Assessment of trending ability of cardiac output monitors by polar plot methodology Bedside monitoring of lung volume available for gas exchange A mainstream monitoring system for respiratory CO2 concentration and gasflow Statistical methods for assessing agreement between two methods of clinical measurement None. Bitker et al. ICMx (2021) 9:46 accommodate the heterogeneities in the distribution of the ventilation than a device based only on end-tidal gas composition. The availability of a technology providing accurate bedside assessment and trending ability of end-expiratory lung volume allows foreseeing the regular assessment of a recruitment maneuver or prone positioning. Likewise direct assessment of the effects on EELV of a PEEP change should make PEEP optimization part of systematic daily practice. Furthermore, measurement of EELV allows computation of lung strain (the lung deformation related to its original status) and lung stress (using a presumed value for specific elastance), both being important parameters to evaluate the risk of ventilator-induced lung injury. This should help implement a personalized lung protective mechanical ventilation. Some limitations of the present study should, however, be acknowledged. First, the experimental ARDS model used is highly recruitable by PEEP, and external validity in poorly recruitable ARDS may be questionable. Second, the reliability of the technique was not tested at VT lower than 6 ml.kg −1 , at FiO 2 higher than 80% and higher respiratory rates (i.e., situations in which the MBNW technique may be less reliable [6] ). Furthermore, the present study was performed in deeply sedated animals under neuromuscular blocking agents and controlled ventilation, and the reliability of the present MBNW technique during mechanical ventilation with assisted ventilation modes remains unknown. Third, lung segmentation on CT excluded the trachea and the main bronchus (i.e., structures involved in EELV computed by the MNBW technique), which could explain a slight underestimation of EELV CT , although this would be a systematic error as these structures are cartilaginous and are not expected to increase their volumes at increasing PEEP levels. We report a novel technique that reliably assesses EELV at bedside, and might add valuable information to further personalization of mechanical ventilation. This novel system designed to assess EELV shows high measurement precision and with excellent trending ability, compared to EELV estimated using CT, as the reference technique. Finally, estimation of PEEP-induced alveolar recruitment using this novel system appears reliable in the specific setting of a highly recruitable experimental ARDS model. The online version contains supplementary material available at https:// doi. org/ 10. 1186/ s40635-021-00410-x.Additional file 1. Available and missing measurements. Authors' contributions LB made substantial contributions to study design, to data acquisition, study analysis, and interpretation of data, and drafted the manuscript, and approved the version to be published. NCC made substantial contributions to study design, to data acquisition, and interpretation of data, and revised the manuscript for important intellectual content, and approved the version to be published. SR made substantial contributions to study design, to data acquisition, and interpretation of data, and revised the manuscript for important intellectual content, AND approved the version to be published. CS made substantial contributions to study design, to data acquisition, and interpretation of data, and revised the manuscript for important intellectual content, and approved the version to be published. DN made substantial contributions to study design, to data acquisition, and interpretation of data, and revised the manuscript for important intellectual content, and approved the version to be published. MO made substantial contributions to study analysis, and interpretation of data, and revised the manuscript for important intellectual content, and approved the version to be published. EDS made substantial contributions to study analysis, and interpretation of data, and revised the manuscript for important intellectual content, and approved the version to be published. JPR made substantial contributions to study design, to data acquisition, and interpretation of data, and revised the manuscript for important intellectual content, and approved the version to be published. JCR made substantial contributions to study design, to data acquisition, study analysis, and interpretation of data, and drafted the manuscript, and approved the version to be published and agreed to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. All authors read and approved the final manuscript. The study was funded by Hamilton Medical AG. The funding body was involved in the design of the study, during data acquisition, and reviewed the manuscript, but was not involved in data analysis. LB, MO, EDS, and JCR are not employees of Hamilton Medical AG, and did not receive reimbursements, fees, or salary by the company. The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request. The study was approved by an Institutional Review Board for the care of animal subjects (Comité d'éthique en experimentation animale n°042, APAFIS#21542-2019071909353561), Not applicable.