key: cord-0017881-slmyrylv authors: Chihab, Samir; Grommersch, Bryan M.; Erens, Greg A.; Guild, George N. title: Trunnion Failure in Revision Total Knee Arthroplasty date: 2021-02-26 journal: Arthroplast Today DOI: 10.1016/j.artd.2021.01.003 sha: ece89310a83051e04c0a63b2ab15365feea357e6 doc_id: 17881 cord_uid: slmyrylv In revision total knee arthroplasty, joint kinematics must be maintained amid bone and ligamentous insufficiency. Current modular designs address defects while allowing for intraoperative prosthesis customization through a variety of stem extensions and constraints. Additional constraint improves knee stability while increasing stress at the implant-host interface and modular junction of the implant. This renders the prosthetic stem-condyle junction more prone to fatigue failure. We report 2 cases of prosthetic stem-condyle junction failure in in a varus-valgus constrained revision total knee arthroplasty. Primary and revision total knee arthroplasty (rTKA) cases are expected to increase six-fold between 2005 and 2030 [1] . rTKA Presents many challenges. Bone loss and ligamentous insufficiency must be addressed while maintaining joint kinematics. The use of stems in rTKA is often necessary to obtain fixation in deficient bone and allow for long-term fixation with increasing constraint [2] [3] [4] . Constrained implants improve the stability of the knee but increase stress at the implant-host interface and modular junction of the implant. As a result, the prosthetic stem-condyle junction becomes a potential area for fatigue failure [3, 5] . Therefore, the benefit of constrained implants must be balanced against the risk of added construct stress. Several types of constraint are currently available to compensate for ligamentous insufficiencies [6] . In general, using the minimum necessary constraint when performing rTKA is recommended [7] . Failure of TKA implants at the prosthetic stem-condylar junction has been reported in the literature [8] [9] [10] . We report 2 cases of modular junction failure in posterior stabilized constrained rTKA. An 80-year-old male patient with a BMI of 25 underwent primary right TKA in 2007 for symptomatic osteoarthritis at an outside facility. Unfortunately, his postoperative course was complicated by infection requiring two-stage revision. Over 10 years later, the patient developed worsening right knee pain and instability. His radiographs revealed gross mechanical failure at the junction of the femoral component and stem (Fig. 1) . The infection workup was negative. He underwent rTKA because of mechanical failure of a posterior stabilized legacy constrained condylar knee (LCCK) prosthesis (ZimmerBiomet, Warsaw, IN). Informed consent for the procedure and publication was obtained and documented. Upon entry into the knee, it was noted that the patient had thickened black-stained synovium (Figs. 2 and 3). The joint fluid appeared relatively normal. The LPS liner did not show any signs of excessive or focal wear. As revealed on the preoperative radiographs, the femoral component was grossly broken at the femoral stem-condyle junction. Because the broken femoral stem remained lodged within the femoral canal, osteotomes followed by a trephine were used. The femoral component was easily extracted, and the tibial plate was noted to be solidly fixed. The decision was made to keep the tibial base plate to avoid more proximal bone loss. The femoral component was revised to a posterior stabilized LCCK with a small cone while the tibial component remained in place (Fig. 4) . Unfortunately, the initial revision was performed at an outside institution in 2008, and documentation of the implant sizes was not available. The implants used for the revision reported herein were femur size D with 13 Â 100 cemented femoral stem and a porous tantalum cone size small 30 mm in height, as well as a 20-mm LCCK polyethylene liner. At 21-month follow-up, the patient had experienced no postoperative complications and demonstrated an active range of motion of 0-110 with excellent strength and stability. Informed consent for the procedure and publication was obtained and documented. Upon entry into the joint, the fluid appeared normal with no signs of infection or metallosis. The polyethylene liner did not show any signs of excessive wear. As expected, the tibial base plate was grossly loose within the proximal tibia and was extracted (Fig. 6a) . This allowed visualization of the stem, which remained within the canal. The stem had broken at the trunnion (Fig. 6b ). After clearing out cement and debris around the stem, it was easily removed from the canal. A Persona Revision Total Knee System (ZimmerBiomet, Warsaw, IN) was used with cemented stem extensions and a tibial cone (Fig. 7) . The implants revised were femur size F with 20 Â 130 sharp fluted stem extension and 5-mm augments medially and laterally, wedged tibial plate size 5 with offset 13 Â 155 stem extension, and a 12-mm liner. The implants used in the revision were femur size 9 with 20 Â 135 smooth cemented stem, tibia size D with 12 Â 135 smooth cemented stem extension, and a 22-mm LCCK polyethylene liner. At the 2-week postoperative visit, radiographs demonstrated a revision right total knee replacement without complication. His surgical staples were removed. The patient was able to ambulate short distances with a walker and was recovering without issue. Unfortunately, the patient has not returned for additional follow-up amid the COVID-19 pandemic. Primary and rTKAs are increasingly prevalent in the United States [1] . As TKAs are performed in younger patients, surgeons will be faced with rising amounts of severe bone loss and ligamentous deficiencies in revision surgeries. In such cases, the surgeon must combat the many challenges associated with restoring original anatomy and function while maintaining stability. Suboptimal distal femoral or proximal tibial fixation and ligamentous insufficiency dictate the use of constrained stemmed components [9, 11, 12] . Stem-condyle junction failure remains a rare clinical event, with approximately a dozen cases previously reported in the literature. years after most recent revision [15] . Howell and Rorabeck disclosed an incident of intraoperative femoral intramedullary stem disengagement, likely from an assembly error, that lead to night pain but no significant implant loosening at 3-year follow-up [16] . Boe et al. reported a similar dissociation of the femoral condylar component in a Triathlon TS prosthesis 7 years after most recent revision (Stryker, Kalamazoo, MI) [17] . Most recently, Kahan and Estes detailed failure at the femoral stem-extension condylar interface in a NexGen Rotating Hinge Knee 7 years after revision (ZimmerBiomet, Warsaw, IN) [18] . To our knowledge, the cases presented in this report are the first to identify stem-condyle junction failure in a posterior stabilized LCCK prosthesis. Meta-diaphyseal stem fixation is required to share articular loads, augment suboptimal condylar fixation, and protect ligament deficiencies [19, 20] . Although articular constraint provides knee stability, it increases stress at the bone-component interface and at the condylar-stem junction in case of a modular prosthesis [5, 21] . As a general principle, the minimum necessary amount of constraint should be used when performing rTKA [7] . Modular design helps with fixation and stability by allowing for intraoperative customization. In the case of modular prostheses, the area of maximal stress concentration is at the stem-prosthesis junction and tip of the stem [2, 5] . The stem extensions, however, are unable to resist the entire load when limited juxta-articular bone support is available. As the condylar portion of the femoral or tibial implant loosens, stresses at the stem junction increase, especially if the part of the stem further from the joint remains stable. The prosthetic stem-condyle junction may not be able to withstand such concentrated forces and becomes an area susceptible to fatigue failure [5] . In our patients described in case 1 and case 2, there was radiographic evidence of osteolysis and bone loss at the distal femur and proximal tibia, respectively. This may have contributed to aseptic component loosening. In these cases, bone loss at the distal femur or proximal tibia may have been caused by stress-shielding of the respective cemented stemmed component [22, 23] . It is likely that this component aseptic loosening increased stresses throughout the stem and the taper connection. The benefits of the current modular knee systems are intraoperative customization and improved fixation and kinematics. However, there is a potential for failure at the modular junction. When increased constraint is required, the minimum necessary amount of constraint should be used, and special consideration should be given to the amount of stress that will be placed on the condyle-stem junction. Adequate distal bone support should be optimized with a variety of techniques (bone graft, augments, cone, sleeve) to avoid excessive stress on the femoral and tibial component-stem junctions. The authors declare there are no conflicts of interest. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030 Stemmed revision arthroplasty for aseptic loosening of total knee replacement Failure of the stemecondyle junction of a modular femoral stem in revision total knee arthroplasty: a report of five cases Revision total knee arthroplasty with use of a constrained condylar knee prosthesis Detection of locking bolt loosening in the stem-condyle junction of a modular femoral stem in revision total knee arthroplasty Revision total knee arthroplasty: how much constraint is enough? Revision total knee arthroplasty with use of modular components with stems inserted without cement Cemented long-stem revision total knee arthroplasty Three cases of femoral stem failure in rotating hinge revision total knee arthroplasty: causes and surgical considerations Disengagement of a locking screw from a modular stem in revision total knee arthroplasty. A report of three cases Total condylar III knee prosthesis: long-term follow-up study Revision total knee arthroplasty with modular components inserted with metaphyseal cement and stems without cement Coupling failure between stem and femoral component in a constrained revision total knee arthroplasty Failure at the taper lock of a modular stemmed femoral implant in revision knee arthroplasty: a report of two cases and a retrieval analysis Fracture at the stemecondylar junction of a modular femoral prosthesis in a varusevalgus constrained total knee arthroplasty Femoral stem disengagement in modular total knee revision arthroplasty Failure of the stem-condyle junction of a modular femoral stem in revision total knee arthroplasty Failure at the femoral stem extension-condylar interface in a rotating hinge knee without radiographic evidence of loosening Structural allografting in revision total knee arthroplasty Cancellous bone strains indicate efficacy of stem augmentation in constrained condylar knees Revision total knee arthroplasty with modular cemented stems: long-term follow-up Experimental evaluation of strain shielding in distal femur in revision TKA The role of femoral stem extension in total knee arthroplasty Postoperative anteroposterior (a) and lateral (b) radiographs