key: cord-0015753-m2hm8vfe authors: Dünnwald, Tobias; Kienast, Roland; Niederseer, David; Burtscher, Martin title: The Use of Pulse Oximetry in the Assessment of Acclimatization to High Altitude date: 2021-02-10 journal: Sensors (Basel) DOI: 10.3390/s21041263 sha: 0599a4b8bd7638904839baeb29c52c6267404a93 doc_id: 15753 cord_uid: m2hm8vfe Background: Finger pulse oximeters are widely used to monitor physiological responses to high-altitude exposure, the progress of acclimatization, and/or the potential development of high-altitude related diseases. Although there is increasing evidence for its invaluable support at high altitude, some controversy remains, largely due to differences in individual preconditions, evaluation purposes, measurement methods, the use of different devices, and the lacking ability to interpret data correctly. Therefore, this review is aimed at providing information on the functioning of pulse oximeters, appropriate measurement methods and published time courses of pulse oximetry data (peripheral oxygen saturation, (SpO(2)) and heart rate (HR), recorded at rest and submaximal exercise during exposure to various altitudes. Results: The presented findings from the literature review confirm rather large variations of pulse oximetry measures (SpO(2) and HR) during acute exposure and acclimatization to high altitude, related to the varying conditions between studies mentioned above. It turned out that particularly SpO(2) levels decrease with acute altitude/hypoxia exposure and partly recover during acclimatization, with an opposite trend of HR. Moreover, the development of acute mountain sickness (AMS) was consistently associated with lower SpO(2) values compared to individuals free from AMS. Conclusions: The use of finger pulse oximetry at high altitude is considered as a valuable tool in the evaluation of individual acclimatization to high altitude but also to monitor AMS progression and treatment efficacy. Wearable sensors can provide athletes, coaches, patients and physicians with useful physiological data, e.g., on the actual cardiovascular and respiratory stress of the individual [1] [2] [3] . Such information may be gathered by continuous or spot measurements depending on specific objectives. For example, the monitoring of bio-vital markers like heart rate and peripheral oxygen saturation has become a standard of patient care [4] but is also frequently applied by people visiting high altitudes for sight-seeing, trekking, skiing or climbing [5, 6] . Simple and inexpensive devices, finger pulse oximeters, are widely used to monitor physiological responses to high-altitude exposure, the progress of acclimatization, and/or the potential development of high-altitude related diseases [7] [8] [9] . Although there is increasing evidence for the usefulness of pulse oximetry at high altitude some controversy remains [10, 11] . This is largely due to differences in individual preconditions, different evaluation purposes, different measurement methods, not considering limitations of devices in certain conditions, the use of different devices, and the lacking ability to interpret data correctly [12] . Therefore, this review is aimed at providing information on the functioning of pulse oximeters, appropriate measurement methods and published time courses of pulse oximetry data (peripheral oxygen saturation, SpO 2 ; and heart rate, HR) recorded at rest and submaximal exercise during exposure to various altitudes. Moreover, where available, alterations in acute mountain sickness (AMS) scores for the particular exposure will be illustrated. Interpretation notes are intended to enable the reader to differentiate between normal and potentially pathologic values recorded by pulse oximetry in consideration of different individual conditions. Part 1: In the first part of this review, we present basic principles of functioning and, based on a selected literature review, most relevant pitfalls and possible countermeasures for pulse oximetry particularly concerning healthy people going to high altitudes. Part 2: In order to show the various changes in SpO 2 and HR observed following acute exposures to altitude, the second part of this review presents the results of a literature search performed in the PubMed database including studies published prior to June 2020, using the search terms "oxygen saturation", "heart rate", "high altitude", "acclimatization", "exercise", "performance" and "acute mountain sickness". No restriction was done regarding to the type of studies. The search was complemented by articles known to the authors and by screening reference lists of selected review articles. Studies reporting SpO 2 data from subjects acutely exposed to high altitude with subsequent active or passive stays were included in this review. Exclusion criteria included the absence of reporting baseline data on SpO 2 , the presentation of data obtained at sea level and once at altitude only, stays at higher altitude levels during the exposure, administration of medications as well as stays lasting less than 2 days and more than 28 days. Core findings of the presented literature overview are summarized. Figure 1 shows a flow chart of the selection process. Studies predominantly involved an observational study design. Part 3: In the third part, physiologic and pathophysiologic mechanisms explaining pulse oximetric measures when acutely exposed to high altitude and during acclimatization are discussed, including practical aspects for the mountaineer going to high altitude. The introduction of pulse oximetry represents one of the most important technological advances in medicine, permitting to continuously, non-invasively and simultaneously monitor SpO 2 of hemoglobin in the arterial blood and HR. These vital parameters provide exceptionally important information on well-being of the individual, e.g., patients in the hospital setting, in emergency situations at home or in the field, but also in people going to high altitudes. Hemoglobin (Hb) is a prominent protein complex in erythrocytes. Based on its ability to bind oxygen (O 2 ) it is essential for the transport of O 2 from the alveoli to tissues. Hb exists in two forms: (1) the deoxyhemoglobin (HHb) without attached O 2 and (2) the oxyhemoglobin (O 2 Hb) with bound O 2 molecules. O 2 molecules change the light absorption of Hb at specific wave lengths [6, 13] . This effect can be observed even under normal light conditions since well oxygenated (arterial) blood with high O 2 Hb concentration exhibits a bright red staining, while venous (deoxygenated) blood appears dark red for the eye. Pulse oximetry makes advantage of this optical effect of diverging light absorption of Hb and utilizing red light at wavelength of 660nm and near-infrared (IR) light at 940 nm to estimate SpO 2 . The distinct feature of these two wavelengths is that red light is more strongly absorbed by HHb than by O 2 Hb, whereas infrared light exhibits the opposite characteristic (see Figure 2 ). For both wavelengths, the absorption during systole (A AC ) and diastole (A DC ) is measured and the modulation ratio R is calculated, where R is expressed as: Based on a comparison of R to an empirically generated calibration curve, the SpO 2 value is estimated [13] [14] [15] . The data for this calibration curve was acquired from adult healthy volunteers and includes saturation values from 70% to 100% [12, 13, [15] [16] [17] . By utilizing multiple wavelengths, selected instruments are also capable of determining the most important dyshemoglobins-carboxyhemoglobin (SpCO) and methemoglobin (SpMet)-in addition to oxy-and deoxyhemoglobin [14, 18, 19] . Sensors commonly used in pulse oximetry can be categorized into two types, according to the measurement method: (1) Transmission sensors, where the emitter and receiver are placed opposite to each other and the light passes through the tissue (e.g., finger-or earlobe sensors), and (2) reflection sensors, where the emitter and receiver are placed next to each other and the backscattered light is analyzed (e.g., forehead-or wrist sensors) [14] . Because of the easy-to-use and wide clinical acceptance, pulse oximetry has become a well-established indirect method for continuous and noninvasive monitoring of blood oxygenation [21] . By default, medical pulse oximeters comply with the international standard for pulse oximeter manufacture ISO 80601-2-61 and must maintain high accuracy (average root mean square error A rms ≤ 4%) in SpO 2 reading within the range of 70-100% when compared to arterial oxygen saturation (SaO 2 ) obtained from arterial blood gas (ABG) analysis. The FDA recommends A rms values of ≤3.0% for transmission sensors and ≤3.5% for ear clip and reflectance sensors [22] . This also implies that not each oximeter would provide exactly the same reading if theoretically measured in the same individual, at the same time and at the same location. However, the variance should be within the limits specified. Assuming faultless handling of the instrument, these variations are attributable to technical variations, e.g., different signal averaging times, incorrect calibration or differences in the number and precision of wavelengths used [23] . Devices for non-medical use not conforming to this ISO standard may have larger deviations in SpO 2 readings. There exist a limited number of studies comparing reasonably priced commercially available handheld devices to medical-standard-devices. These studies indicate that "low-cost" handheld devices provide sufficiently accurate SpO 2 values in the range of about 90-100% compared to medical devices, however, below 90%, non-medical devices decrease in accuracy [16, [24] [25] [26] . This property might be a drawback for measurements at high altitudes. Although, no explicit statement for the use of non-medical devices at high altitudes can be made based on these few studies. In general, there is a lack of data on the measurement accuracy of pulse oximeters at high altitudes comparing SpO 2 to SaO 2 obtained from ABG analysis. This complicates to identify devices that are appropriate for their usage at high altitudes without restrictions. However, certain parameters may require increased attention when the decision for a specific pulse oximeter has to be made: (1) accuracy, precision, and bias of the device [12] ; (2) environmental conditions such as maximum operating altitude respectively minimum air pressure or the minimum operating temperature; (3) the availability of advanced algorithms to reduce motion artifacts or the detection of low perfusion; (4) the selection of the sensor location, typically using the finger [12, 15, 27, 28] -however, other common positions such as the forehead may be considered, in particular if the measurement is conducted during motion [28] [29] [30] ; (5) if required: the opportunity to sense carboxyhemoglobin and methemoglobin. In addition to the technical aspects, however, the measurement process itself becomes more error-prone with increasing altitude [6, 12] . Table 1 summarizes the most significant pitfalls that may lead to inaccurate SpO 2 readings and possible countermeasures particularly for healthy people visiting high altitudes. The application of standardized procedure by trained users can prevent incorrect measurements and data interpretation [6, 12] . Unfortunately, there exists no uniform standard for measuring protocols at high altitudes but there are some specific recommendations available to minimize measurement uncertainties which largely coincides with the countermeasures listed in Table 1 : (1) The test person should remain in a sitting position for about 5 min, (2) the measuring site (normally the finger) should be kept as warm as possible, e.g., by wearing gloves, (3) motions of the sensor should be prevented, (4) the sensor should be shielded from ambient light, (5) a trained and experienced examiner should perform the measurements, and (6) SpO 2 values should be monitored and averaged over a period of 2-3 min [6, 12, 31] . Additionally, if the device offers the ability to display the pulse wave graphically, it should be ensured that it remains as stable as possible [31] . In the following section, studies implementing a pulse oximeter to document the progress of acclimatization are illustrated. [13, 32, 33] , however modern devices implement advanced algorithm to reduce motion artefacts [14, 15, 17, 33, 34] . These devices may be identified by indications such as "motion tolerant" or "motion resistant" [34] [35] [36] . With increasing altitude, the temperature drops. This may result in cold extremities and an increased shivering and affects the sensor position and the sensor signal. During measurement, keep the measuring position steady and avoid too much trembling. Poor probe positioning The red or infrared light of the sensor may bypass the tissue or too high levels of ambient light hit the light-detector of the sensor [13, 15] . This results in a wrong SpO 2 reading. An imperfectly fitting of the sensor should be avoided, and the sensor should always be used in the appropriate position. If these countermeasures do not achieve the desired results, measurement at a different site (e.g., earlobe or forehead) may be considered. Excessive ambient light can lead to erroneous SpO 2 readings. However, modern devices are capable of handling strong ambient light more effectively [13, 37] . Especially snow-covered areas with high solar radiation can lead to increased ambient light intensity at the sensor. Protect the sensor from sunlight (e.g., by covering the measuring site). A decrease in PaO 2 (<60 mmHg) results in a significant change in oxygen saturation where small variations of the pressure have a strong effect on the saturation [12, 38, 39] . Increasing altitude results in a decreasing PaO 2 . Especially at altitudes above 3000 m a PaO 2 below 60% can be expected [12, 40, 41] . To reduce fluctuations in PaO 2 , SpO 2 measurements should be conducted after the person keeps silent and gentle breathing for several minutes. The measurement duration should be extended over a few minutes and the most frequent occurring value should be used [27] . Devices complying with the international standard ISO 80601-2-61 (medical electrical equipment. Part 2-61: Specific requirements for basic safety and essential performance of pulse oximeter equipment) must measure accurate oxygen saturations (A rms ≤ 4%) between 70-100%. Below 70%, they are less reliable [11, 12, 16, 23, 25, 42] . the oxygen saturation is estimated based on human calibration data measured from 100% to 70%. Saturation values below 70% are only based on an extrapolation of this determined curve [13, 25] . At high altitudes, however, the occurrence of low saturation values is not abnormal. The possibility of a slight deviation of the measured value should be considered if oxygen saturation is below 70%. Especially when comparing data with devices of several manufacturers. High Altitude Cold-induced vasoconstriction (poor perfusion) Cold skin temperatures reduce SpO 2 reading accuracy [43] [44] [45] . This effect is based on a reduced blood flow due to cold-induced vasoconstriction [12, 31] . However, modern devices can handle this condition and/or report it to the user. These devices may be identified by indications such as "oximeter with perfusion index" or "sensitivity to low perfusion signals". With increasing altitude, the temperature drops. Warming the measuring site before and preferably during the measurement (e.g., using heating pads). Pulse oximeters are possibly less accurate during hypoxia in dark-skinned individuals at lower saturation (<80%) resulting in overestimations [46, 47] . Feiner et al. mentioned [48] : "further study is needed to confirm these observations in the relevant populations." However, an actual study is consistent with Bickler et al. and Feiner et al. [49] . At high altitude, the occurrence of saturation values below 80% is not abnormal [12, 50] . Until the scientific data is more definite, the possibility of a slight overestimation (about +2% [51] ) of the measured value should be kept in mind when interpreting the data for a person with oxygen saturation below approx. 90% combined with a dark pigmentation of the skin. Some fingernail polish can lower the SpO 2 readings [48] . Previous studies, however, have shown that the variance is not clinically relevant using actual devices [51] [52] [53] [54] . Especially with older devices, the nail polish should be removed to avoid variations in the measurement accuracy. However, the deviation in the SpO 2 readings is less than 2% [53, 54] . A lack of knowledge regarding device application and interpretation of the measurement data can lead to incorrect conclusions [14, 23] . Conditions at high altitudes complicate the use of the device and the accurate interpretation of the data [12, 27, 50] . As Tannheimer et al. [50] concluded it "requires an experienced examiner who can include altitude anamnesis, clinical examination and mountaineering aspects in the overall assessment" to avoid possible pitfalls during SpO 2 measurement and interpretation on high altitudes. High Altitude Based on an absorption of the red and infrared light, methemoglobin (MetHb) and carboxyhemoglobin (HbCO) cause SpO 2 overestimation and mask serious hypoxia [14, 23, 48] . However, as already mentioned, certain multiple-wavelengths devices are capable of detecting dyshemoglobins. Unlikely at high altitude, however, it can be a danger using a cooking stove in small, enclosed areas like tents. In the worst case, this can lead to carbon monoxide poisoning [12, 55] . When using devices that are not capable of analyzing dyshemoglobins, possible carbon monoxide poisoning should be considered if the person has remained in a small, enclosed space for an extended period while a combustion process (e.g., a stove) has taken place. Symptoms of carbon monoxide poisoning may include headache, nausea and drowsiness. However, these symptoms are similar to those associated to altitude sickness [56] . The findings from 18 studies reporting resting SpO 2 data are presented in Table 2 . Two studies were performed at an altitude of 2000-2200 m [57, 58] . In 10 studies, subjects were exposed to altitudes ranging from 3400 m to 4350 m [59] [60] [61] [62] [63] [64] [65] [66] [67] [68] . The remaining six studies were performed at altitude levels ranging from 5050 m to 5700 m [69] [70] [71] [72] [73] [74] . Altitude stays lasted from five to 22 days. In four studies, an ascent phase of four to eight days preceded the sojourn at altitude [64, [71] [72] [73] . In the majority of the studies, subjects were adults, one study included children [64] whereas one study was conducted in the elderly [58] . Subjects remained physically active by daily hiking [58] , trekking [64] or few ascents to higher altitudes [66] in four studies. After an initial fall in SpO 2 in the beginning of the stay at altitude ≤ 2200 m (−3% to −6%), the subsequent increase in SpO 2 during acclimatization ranged from 0% to 3% [57, 75] . In studies performed at altitudes varying from 3400 m to 4350 m, the initial decline in SpO 2 was reported to range from −5% to −19%. During acclimatization, SpO 2 rose by 3% [62, 63] in studies with short duration of exposure (5 days) and from 5%-8% in studies with longer duration (i.e., 6-22 days) [59, 61, 65, 67, 68] . When exercise was repeatedly performed during the stay at these altitudes, the increase in SpO 2 over time ranged from 1%-3% [64, 66] . One study that evaluated SpO 2 data immediately before the start of an exercise test did not observe an acclimatization effect on SpO 2 [60] . At altitudes > 5000 m, SpO 2 initially decreased by −8% to −20%. SpO 2 changes during acclimatization ranged from 6-8% [71] [72] [73] . In one study, SpO 2 increased by 4% after 5 days at 5050 m but decreased again by 3% after 14 days [70] . Baillie et al., assessing AMS in 23 out of 35 subjects (66%), observed no acclimatization effect on SpO 2 [74] . Changes in resting HR during the stay at altitude are mentioned in 10 out of the 20 studies. An initial increase in HR of 8 bpm at moderate altitude was observed in the study implementing daily hiking activities, with only small changes during acclimatization (−1 bpm) [58] . In the three studies performed at 3400 m to 3500 m, HR increased by 14-21 bpm, remained elevated when trekking exercise was performed during the stay [64] and decreased during acclimatization by −6 bpm after 5 days [62] and by −20 bpm after 9 days [60] . At altitudes >5000 m, the initial increases in HR ranged from 9 bpm to 25 bpm, whereas changes during acclimatization phase ranged from −1 to −9 bpm [69-73]. Effects of acclimatization in SpO 2 and HR during submaximal exercise are shown in Table 3 . Studies were performed at altitudes ranging from 2800 m to 4300 m, except one study implementing exposures at 2000 m [75] . Altitude stays lasted from three to 22 days. In the studies with shorter duration of exposure (i.e., <7 days), initial fall in SpO 2 after exercise ranged from −15% to −22%. In the end of the stay, SpO 2 following exercise remained unchanged or increased by 3% [63, 76, 77] . In the studies lasting from seven to 22 days, SpO 2 declined by −15% to −23% during exercise tests at acute hypoxia [60, 66, 68, 78] . In these studies, the rise in SpO 2 during acclimatization ranged from 2% to 10%. Exercising HR at acute altitude exposure increased by 11 bpm to 38 bpm in six out of seven studies when compared to sea level. In the remaining two studies, reporting either 55% slower time trial durations [78] or significant reductions in mean power [77] , exercising HR during acute altitude exposure declined in a range of −1 bpm to −7 bpm. In those studies, showing a primary increase in HR during the first exercise test, the reduction in exercising HR ranged from −4 bpm to −8 bpm [60, 63, 66, 76] , still being elevated compared to sea level. In the study performed at 2000 m, exercising HR returned to baseline after 7 days of active stay at altitude (−14 bpm) [58] . The findings of 12 studies evaluating symptoms of acute mountain sickness during acclimatization using the Lake Louise Scoring System (LLS) are presented in Table 4 . Reductions in SpO 2 in these studies ranged from −5% to −29%. Altitudes ranged from 3180 m to 5400 m, and duration of exposures lasted between two to 11 days. In the studies with resting SpO 2 levels below 80%, LLS ranged from 2.7 to 6.0 [72, 74, [79] [80] [81] [82] [83] . In all other studies, resting SpO 2 remained above 80% with LLS ranging from 1.4 to 3.9, except in one study where all subjects developed AMS showing a LLS of 5.3 [84] . In most studies, LLS decreased over time when SpO 2 was improved. However, in one study, LLS increased from 0.6 to 0.9 in a group of non-AMS subjects [85] . The only study that reported AMS-C scores observed similar time-courses as in the studies using LLS to assess AMS [86] . The presented findings from the literature review demonstrate rather large variations of pulse oximetry measures (SpO 2 and HR) during acute exposure and acclimatization to high altitude. This is not surprising as conditions (levels of altitude/hypoxia, type of ascent, extent of pre-acclimatization, physical activity levels, characteristics of study populations, etc.) are also considerably different between studies. Main conclusions derived from this review are, that particularly SpO 2 levels decrease with acute altitude/hypoxia exposure and partly recover during acclimatization, with an opposite trend of HR, i.e., initial increase and slight decrease during acclimatization. In addition, AMS development is consistently associated with lower SpO 2 values compared to individuals free from AMS. When acutely ascending to high altitude, the human organism is exposed to the reduced availability of oxygen (hypoxia) because of the decreasing barometric pressure and related partial pressure of oxygen (pO 2 ). As a consequence, several physiological responses are initiated to counteract hypoxia and the associated risk to get sick [88] [89] [90] . Such responses are targeted to improve oxygen delivery to tissues including hyperventilation (hypoxic ventilatory response, HVR), hemoconcentration due to diuresis, and elevated cardiac output due to sympathetic activation. During the first few days at high altitude there is, compared to acute exposure, a progressive increase in resting ventilation (ventilatory acclimatization) which is accompanied by an increase in the arterial oxygen partial pressure (PaO 2 ) and a decrease in the alveolar partial pressure of carbon dioxide (pACO 2 ) [88, 91] . Ventilatory acclimatization is characterized by an initial decrease and a subsequent increase in resting SpO 2 values, reaching a maximum between 4 to 8 days, at least at an altitude up to 4300 m [65, 68] . Provided altitude levels are comparable, increase in SpO 2 seems to be higher when the time for acclimatization is prolonged (e.g., >5 days). In contrast to passive stays at altitude, frequent trekking activities or other exercises at altitude may delay the recovery of SpO 2 . The initial SpO 2 decline is more pronounced and acclimatization may take some more days at higher compared to lower altitudes. An example for changes of SpO 2 values during acclimatization at two different altitudes is depicted in Figure 3 . Moreover, larger SpO 2 variation during the first days become narrower with progressing acclimatization indicating slightly different individual time courses. Resting HRs are typically elevated during the first few days after acute ascent to high altitude showing a subsequent decrease with acclimatization. However, the acclimatization effect on HR may be counteracted if the stay involves physical exercises [64] . An example for changes of HR values during acclimatization at 3600 m is shown in Figure 4 . In the healthy general population, individual HR variation is much more pronounced than that of SpO 2 values, e.g., depending on age, sex, and fitness, requiring careful consideration of the individual baseline values. Therefore, including HR values to assess the progress of acclimatization is a valuable complement to the SpO 2 measurements. It is important to mention that the presented pulse oximetry data reflect the time course of acclimatization predominantly after rapid ascent to a certain level of altitude. Various pre-acclimatization strategies including the use of hypoxia chambers, staging or mountaineering activities at moderate altitudes may accelerate the acclimatization process at a given high altitude [65, 92, 93] . On the other hand, higher than normal susceptibility to hypoxia exposure or pre-existing illnesses may delay this process [94, 95] . . Example for changes of heart rate (HR) when acutely ascending from low (LA) to high altitude (HA1) and during the subsequent day acclimatization period at 3600 m based on a study with young soccer players (16 ± 0.4 years) [51] . HR data were collected in the morning after awakening with a Polar Team system (Polar Electro Oy, Kempele, Finland). Changes of SpO 2 and HR values when acutely ascending to high altitude are more pronounced during submaximal exercise when compared to resting conditions [96] . From rest until the anaerobic threshold, there is an almost linear increase in heart rate and minute ventilation with increasing workload, the slope being steeper at high than low altitude [97] . The SpO 2 decline during exercise depends, at least partly, on the individual ventilatory response and will become less steep with ventilatory acclimatization. This is based on the fact that the PaO 2 at altitude is on the steep portion of the oxyhemoglobin dissociation curve, and an only slight fall in PaO 2 , e.g., because of hypoventilation, results in a marked SpO 2 reduction [98] . The SpO 2 during submaximal exercise is lowest during the first days at altitude and improves with acclimatization. Compared to resting conditions (4 to 8 days), the increase of exercising SpO 2 values plateaus after 2 to 3 weeks during acclimatization to high altitude [68, 99] . Thus, as observed in the studies presented in this review and similarly to resting SpO 2 , the magnitude of improvement of exercising SpO 2 may be less pronounced in studies with shorter duration. An example for changes of (submaximal) exercising SpO 2 during acclimatization to 4300 m is shown in Figure 5 . Exercising heart rates decrease with increasing SpO 2 during acclimatization exhibiting a similar time course [93] . In contrast to the aerobic capacity (VO 2 max), the initially reduced submaximal endurance performance at acute altitude improves during 2 to 3 weeks of acclimatization, accompanied by characteristic changes in cardiorespiratory responses [99] [100] [101] [102] . For instance, the study by Horstman et al. reported a 31% and 59% improvement on day 9 and 15 without considerable changes until day 22 [99] . It has to be mentioned, that relevant acclimatization effects, i.e., hemoconcentration, hyperventilation and associated improved oxygenation and submaximal exercise performance, even occur during the first days at high altitude [77] . Figure 5 . Example for changes of peripheral oxygen saturation (SpO 2 ) during submaximal exercise when acutely ascending from low (LA) to high altitude (HA) and during the subsequent 22-day acclimatization period based on a study performed at 4300 m [68] . Resting SpO 2 was measured in a sitting (upright) position for a 4-min period after relaxing for 20 min, by ear oximetry (Hewlett-Packard 47201A ear oximeter, Palo Alto, CA, USA). Although the usefulness of pulse oximetry for the prediction and diagnosis of AMS is still debated [103] [104] [105] , there is increasing evidence confirming that subjects developing AMS are more hypoxic (lower SpO 2 values) than those who stay free from AMS, provided conditions (e.g., level of pre-acclimatization, health status, etc.) are comparable and measurement methods are appropriate [6, 8, 12, 106, 107] . Various studies reported certain SpO 2 cut-off values differentiating between AMS+ and AMS-at a certain level of altitude. For instance, Mandolesi [108] . However, different conditions between individuals can mislead prediction, e.g., subjects taking aspirin before highaltitude exposure tolerated lower SpO 2 values (with regard to suffering from headache at high altitude, 3480 m) than those who were pretreated with placebo (83% vs. 88%) [109] . Thus, one might conclude that, when individual pre-conditions are carefully considered, the use of non-invasive pulse oximetry provides a simple and specific indicator of inadequate acclimatization to high altitudes associated with the risk for developing AMS [108] . In this context, the studies incorporated in the present review clearly show that when SpO 2 levels are improved during the stay at altitude, AMS scores decline. However, a limiting factor regarding these studies is, that a more detailed information about AMS scores (i.e., separate scores of those who got AMS and those who did not) is lacking in most studies. The presented findings indicate that despite the existing large variability, the use of finger pulse oximetry at high altitude is an invaluable tool in the evaluation of the individual course of acclimatization to high altitude but also for the monitoring of AMS progression and treatment efficacy. However, there is a lack of data on the measurement accuracy of pulse oximeters at high altitudes. Therefore, in order to allow for optimal preconditions to obtain reliable data, complying with some specific technical and usage-oriented recommendations may enhance the accuracy and precision and reduce the bias of the measurement results. Especially when going to high altitudes, besides accuracy and precision, other technical characteristics of the device such as maximum operating altitude, minimum operating temperature, or the capability to detect low perfusion should be considered. There are also indications that medical pulse oximeters have an increased accuracy and lower deviations in SpO 2 readings below 90% compared to "low-cost" handheld devices. In parallel to the technical aspects, the measurement process itself becomes more error-prone with increasing height. Some specific pitfalls in pulse oximetry as illustrated in this review affect the measurement results, in particular at high altitudes (e.g., SpO 2 saturation below 70% or considerably decreased PaO 2 ). Therefore, and as no uniform standard for measuring protocols at high altitude exists, an examiner with sufficient experiences in the data interpretation should perform the measurements. Importantly, when interpreting SpO 2 data regarding to acclimatization, one should keep in mind that varying conditions such as pre-acclimatization phases, ascent rates, accomplished altitude levels, extent of physical activity or prevalence of AMS during the altitude stay may have a considerable impact on these data. Wearable Performance Devices in Sports Medicine Balance Improvement Effects of Biofeedback Systems with State-of-the-Art Wearable Sensors: A Systematic Review Cardiorespiratory fitness estimation using wearable sensors: Laboratory and free-living analysis of context-specific submaximal heart rates Pulse oximeters to self monitor oxygen saturation levels as part of a personalised asthma action plan for people with asthma Evaluating Health Impact at High Altitude in Antarctica and Effectiveness of Monitoring Oxygen Saturation The correct measurement of oxygen saturation at high altitude heart rate variability, and the diagnosis of mild-to-moderate acute mountain sickness Prediction of susceptibility to acute mountain sickness by SaO2 values during short-term exposure to hypoxia Oxygen saturation course and altitude symptomatology during an expedition to broad peak (8047 m) Prevalence and prediction of hypoxemia in children with respiratory infections in the Peruvian Andes Arterial Oxygen Saturation, Pulse Oximetry, and Cerebral and Tissue Oximetry in Hypobaric Hypoxia Pulse oximetry at high altitude Pulse oximetry: Understanding its basic principles facilitates appreciation of its limitations Pulse oximetry Current progress of photoplethysmography and SPO2 for health monitoring The Accuracy of 6 Inexpensive Pulse Oximeters Not Cleared by the Food and Drug Administration: The Possible Global Public Health Implications The effect of motion on pulse oximetry and its clinical significance The measurement of carboxyhemoglobin and methemoglobin using a non-invasive pulse CO-oximeter Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia Tabulated Molar Extinction Coefficient for Hemoglobin in Water Pulse oximetry: Fundamentals and technology update Pulse Oximeters-Premarket Notification Submissions [510 (k)s] Guidance for Industry and Food and Drug Administration Staff ; U.S. Department of Health and Human Services Clinical use of pulse oximetry: Official guidelines from the Thoracic Society of Australia and New Zealand Clinical Interpretation of Peripheral Pulse Oximeters Labeled "Not for Medical Use Pulse Oximetry for Monitoring Patients with COVID-19 at Home. Potential Pitfalls and Practical Guidance Perioperative comparison of the agreement between a portable fingertip pulse oximeter v. a conventional bedside pulse oximeter in adult patients (COMFORT trial) Measuring arterial oxygenation in a high altitude field environment: Comparing portable pulse oximetry with blood gas analysis Investigation of Photoplethysmography Behind the Ear for Pulse Oximetry in Hypoxic Conditions with a Novel Device (SPYDR) A Comparison of Reflective Photoplethysmography for Detection of Heart Rate, Blood Oxygen Saturation, and Respiration Rate at Various Anatomical Locations Validity of pulse oximetry during maximal exercise in normoxia, hypoxia, and hyperoxia Estimation of arterial oxygen saturation in relation to altitude Motion-resistant" pulse oximetry: A comparison of new and old models Effects of motion artifact on the blood oxygen saturation estimate in pulse oximetry New-generation pulse oximetry in the care of critically ill patients Four Types of Pulse Oximeters Accurately Detect Hypoxia during Low Perfusion and Motion Recent advances in pulse oximetry Does ambient light affect the accuracy of pulse oximetry? World Health Organization. Pulse Oximetry Training Manual In StatPearls; StatPearls Publishing: Treasure Island, FL, USA, 2020. 40. Peacock, A.J. ABC of oxygen: Oxygen at high altitude High-altitude medicine Errors in 14 pulse oximeters during profound hypoxia Effects of skin surface temperature on photoplethysmograph Investigating the Effects of Temperature on Photoplethysmography Analysing the effects of cold, normal, and warm digits on transmittance pulse oximetry Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: The effects of oximeter probe type and gender Effects of skin pigmentation on pulse oximeter accuracy at low saturation. Anesthesiology Potential errors in pulse oximetry. III: Effects of interferences, dyes, dyshaemoglobins and other pigments Racial Bias in Pulse Oximetry Measurement The Use of Pulse Oximetry at High Altitude Does fingernail polish affect pulse oximeter readings? Intensive Crit. Care Nurs What is the effect of fingernail polish on pulse oximetry? Chest Effect of Glittered Nail Polish on Pulse Oximetry Measurements in Healthy Subjects Influence of nail polish on pulse oximeter readings of oxygen saturation: A systematic review Carbon monoxide poisoning at high altitudes Wellness, fatigue and physical performance acclimatisation to a 2-week soccer camp at 3600 m (ISA3600) Effect of six days of staging on physiologic adjustments and acute mountain sickness during ascent to 4300 meters Cardiopulmonary and metabolic responses in healthy elderly humans during a 1-week hiking programme at high altitude Intermittent normobaric hypoxia facilitates high altitude acclimatization by curtailing hypoxia-induced inflammation and dyslipidemia Prediction of physiological responses and performance at altitude using the 6-minute walk test in normoxia and hypoxia Factors associated with B-lines after exposure to hypobaric hypoxia Maximum exercise responses of men and women mountaineering trainees on induction to high altitude (4350 m) by trekking Effect of repeated normobaric hypoxia exposures during sleep on acute mountain sickness, exercise performance, and sleep during exposure to terrestrial altitude The Young Everest Study: Effects of hypoxia at high altitude on cardiorespiratory function and general well-being in healthy children Time course of augmentation and depression of hypoxic ventilatory responses at altitude Physiological changes induced by pre-adaptation to high altitude Sea-level PCO2 relates to ventilatory acclimatization at 4300 m Increased exercise SaO2 independent of ventilatory acclimatization at 4300 m The effect of sleep quality in Sherpani Col High Camp Everest Chemoreceptor Responsiveness at Sea Level Does Not Predict the Pulmonary Pressure Response to High Altitude Regional cerebral blood flow in humans at high altitude: Gradual ascent and 2 wk at 5050 m Index measured at an intermediate altitude to predict impending acute mountain sickness High-altitude exposure of three weeks duration increases lung diffusing capacity in humans Oral antioxidant supplementation does not prevent acute mountain sickness: Double blind, randomized placebo-controlled trial Effects of aspirin during exercise on the incidence of high-altitude headache: A randomized, double-blind, placebo-controlled trial Acetazolamide pre-treatment before ascending to high altitudes: When to start? Effects of short-term acclimatization to altitude (3200 m) on aerobic and anaerobic exercise performance Effects of carbohydrate supplementation on aerobic exercise performance during acute high altitude exposure and after 22 days of acclimatization and energy deficit High-Altitude Hypoxia Decreases Plasma Erythropoietin Soluble Receptor Concentration in Lowlanders Decreased contrast sensitivity at high altitude Disturbed eating at high altitude: Influence of food preferences, acute mountain sickness and satiation hormones Effect of short-term acclimatization to high altitude on sleep and nocturnal breathing Quantification of optic disc edema during exposure to high altitude shows no correlation to acute mountain sickness Increased Cytokines at High Altitude: Lack of Effect of Ibuprofen on Acute Mountain Sickness, Physiological Variables, or Cytokine Levels Effect of rapid ascent to high altitude on autonomic cardiovascular modulation Efficacy of residence at moderate versus low altitude on reducing acute mountain sickness in men following rapid ascent to 4300 m Endurance Athletes Are at Increased Risk for Early Acute Mountain Sickness at 3450 m Adaptation to high altitude High-altitude illness Hypoxia-related altitude illnesses Man's respiratory response during and after acclimatization to high altitude Exercise responses after altitude acclimatization are retained during reintroduction to altitude Altitude preexposure recommendations for inducing acclimatization Symptom progression in acute mountain sickness during a 12-hour exposure to normobaric hypoxia equivalent to 4500 m Awareness, prevalence, medication use, and risk factors of acute mountain sickness in tourists trekking around the Annapurnas in Nepal: A 12-year follow-up Submaximal exercise testing at low altitude for prediction of exercise tolerance at high altitude Minute ventilation and heart rate relationship for estimation of the ventilatory compensation point at high altitude: A pilot study Ventilatory acclimatisation is beneficial for high-intensity exercise at altitude in elite cyclists Work capacity during 3-wk sojourn at 4300 m: Effects of relative polycythemia Effects of high-altitude exposure on submaximal endurance capacity of men Maximal performance at altitude and on return from altitude in conditioned runners Preparation for Endurance Competitions at Altitude: Physiological, Psychological, Dietary and Coaching Aspects. A Narrative Review Pulse oximetry in the diagnosis of acute mountain sickness Change in oxygen saturation does not predict acute mountain sickness on Jade Mountain Diagnosis and prediction of the occurrence of acute mountain sickness measuring oxygen saturation-independent of absolute altitude? Sleep Breath Prediction of acute mountain sickness by monitoring arterial oxygen saturation during ascent Arterial oxygen saturation for prediction of acute mountain sickness Long-term monitoring of oxygen saturation at altitude can be useful in predicting the subsequent development of moderate-to-severe acute mountain sickness Aspirin for prophylaxis against headache at high altitudes: Randomised, double blind, placebo controlled trial Data Availability Statement: All available data are provided. The authors declare no conflict of interest.