key: cord-0014149-ttrnuxn4 authors: Datta, A.; Zhong, Z.; Motakef, S. title: A new generation of direct X-ray detectors for medical and synchrotron imaging applications date: 2020-11-18 journal: Sci Rep DOI: 10.1038/s41598-020-76647-5 sha: 1705bd3a36ac80a572b6ba6c8080b4d352da5937 doc_id: 14149 cord_uid: ttrnuxn4 Large-area X-ray imaging is one of the most widely used imaging modalities that spans several scientific and technological fields. Currently, the direct X-ray conversion materials that are being commercially used for large-area (> 8 cm × 4 cm without tiling) flat panel applications, such as amorphous selenium (a-Se), have usable sensitivities of up to only 30 keV. Although there have been many promising candidates (such as polycrystalline HgI(2) and CdTe), none of the semiconductors were able to assuage the requirement for high energy (> 40 keV) large-area X-ray imaging applications due to inadequate cost, manufacturability, and long-term performance metrics. In this study, we successfully demonstrate the potential of the hybrid Methylammonium lead iodide (MAPbI(3)) perovskite-based semiconductor detectors in satisfying all the requirements for its successful commercialization in synchrotron and medical imaging. This new generation of hybrid detectors demonstrates low dark current under electric fields needed for high sensitivity X-ray imaging applications. The detectors have a linear response to X-ray energy and applied bias, no polarization effects at a moderate bias, and signal stability over long usage durations. Also, these detectors have demonstrated a stable detection response under BNL’s National Synchrotron Light Source II (NSLS-II) 70 keV monochromatic synchrotron beamline. than CsI. Beyond 40 keV, however, the attenuation coefficient of a-Se is lower than that of CsI by close to one order of magnitude, rendering a-Se direct conversion FPXIs impractical for chest and torso imaging as well as in any type of tomography applications that operate at energies > 40 keV. Mammography is the biggest market for current direct FPXIs, yet these FPXIs lose their effectiveness for dense breast tissues. Figure 2 shows a plot of MTF versus DQE for direct and indirect commercial FPXIs. Clearly, the indirect FPXIs have a much smaller MTF than direct FPXIs based on a-Se. The highlighted region denotes our best estimate on the expected performance of next-generation X-ray detectors that combine the best properties of both indirect and direct FPXIs. Materials engineering and discovery have significantly benefitted from synchrotron-based imaging techniques in understanding the interplay between atomic-scale structures and macroscopic properties. Modern X-ray imaging techniques, such as coherent X-ray diffraction imaging and phase-contrast imaging, require high beam coherence and highly penetrating X-rays. In the current third-generation synchrotrons, such coherence is only attainable for the lower end of the hard X-ray spectrum. The new generation of synchrotrons, however, achieves enhanced coherence at beam energies higher than 40 keV. This increase in coherence benefits a number of scientific research areas: (1) intensely focused and coherent X-ray beams will dramatically advance nanoscale imaging, nano-spectroscopy, and nano-diffraction; (2) coherent diffractive imaging and ptychography will reach spatial resolution approaching atomic length-scales and permit time-resolved studies; and (3) the statistical information derived from photon correlation spectroscopy will open up previously inaccessible timescales and length-scales 26 . These new prospects for non-destructive imaging require high spatial resolution and efficient detectors, which again can only be successfully fulfilled using high sensitivity FPXIs. www.nature.com/scientificreports/ A typical three-dimensional lead halide perovskite structure consists of eight octahedra of lead halide with Pb at the center. When the halide ions are changed, the displacement of the octahedra directly adjusts the bandgap of this class of material. The size of the cation (such as CH 3 NH 3 + ) is an important factor for the formation of the cubo-octahedral perovskite structure by coordination with the anions 27 . MAPbI 3 has emerged as a new generation of photovoltaic materials that achieved high power conversion efficiencies of around 23% within four years of their development. It is also an excellent candidate for radiation detection due to the presence of high-Z elements Pb. MAPbI 3 has a very high attenuation coefficient, especially at high energies, similar to HgI 2 . MAPbI 3 can, however, be produced at much higher thicknesses than HgI 2 . The strong spin-orbital coupling of MAPbI 3 perovskites due to the presence of heavy elements and inversion asymmetry in the crystal structure results in Rashba effects 28, 29 . Due to this, MAPbI 3 single crystals demonstrate high diffusion lengths of up to 175 µm 30 and carrier lifetimes of 15 µs 31 . Moreover, the recombination of charge carriers also minimizes in perovskite single crystals and polycrystalline thin films. Studies of Rashba-type effects in perovskites are still in their early stages, and further understanding of the phenomena and impact on charge-carrier recombination and transport is needed 28 . In our study, the mobility-lifetime (µτ) product of holes in 200 µm-thick polycrystalline MAPbI 3 measured using the Hecht relation was found to be of the order of 10 -4 cm 2 /V, similar to 250 µm thick polycrystalline HgI 2 sensor and three orders of magnitude better than polycrystalline CZT. The µτ product of a single-crystalline MAPbI 3 device is of the order of 10 -2 cm 2 /V 32 , comparable to the state-of-the-art CZT device. Also, perovskites have a very low density of defects and traps within the bandgap 33 . Although there has been a lot of progress in recent years on the development of MAPbI 3 based X-ray detectors [e.g. 34 ], none of these studies have demonstrated the repeatability and reliability of these devices. In addition, most of these studies are performed using thin sensor layers that significantly limits the X-ray stopping power 35 or have used detector configurations that are incompatible with the widely used large-area high spatial resolution flat panel imaging 36 . There also has not been any stability data, which shows that these detectors can work for a considerable duration of time under X-ray irradiation. Reliability and repeatability have been the main problem with other direct halide semiconductor materials (such as HgI 2 ). These detectors only demonstrate its excellent properties (such as the low dark current) for a few days under electric bias and then becomes unstable. As a result of this, even after a few decades of research, these are not commercially available for FPXI applications. In the following sections, we will discuss the fabrication techniques and properties of the large-area MAPbI 3 -based X-ray detectors that are prime candidates for a new generation of direct FPXIs. The main target of this study was to develop X-ray detectors that can be reliably and reproducibly used for the applications mentioned above. Along with the efficiency and sensitivity of the X-ray detectors, the dark current density is a fundamentally important factor for the proper functioning of the readout matrix on which the halide sensor layer is being deposited. Nearly all detectors based on MAPbI 3 have been reported to have a very high leakage current due to the relatively lower MAPbI 3 bandgap. In order to tackle this problem, we focused on minimizing the dark current of the MAPbI 3 -based detectors with repeatable results while maintaining high X-ray sensitivity. Table 1 lists the performance of the MAPbI 3 -based detector configurations. Table S1 schematically shows the detector configurations. Figure S1 shows the legend for the various layers that are included in each configuration. Two types of polymers (polymers A and B in Figure S1 ) were used to fabricate the charge transporting layers. MAPbI 3 detectors with different sensor thicknesses were fabricated in this study. The thickness range varied from 200 to 1400 µm. An SEM image of a typical MAPbI 3 layer is shown in Supplementary Figure S2 . Figures S3 and S4 show the X-ray characterization setups used in this study. More details on these setups are given in "Methods" section. The characterization experiments were started using a MAPbI 3 -based sensor with no additional charge manipulating layers, i.e., the MAPbI 3 layer was biased directly from both sides. In subsequent steps, the charge controlling layers were added between the MAPbI 3 sensor and the electrical contacts. A total of 16 configurations were tested, and the most promising six configurations are shown in Table 1 . The typical dark currents at an electric field of 0.08 V/µm for each configuration is also shown in the table. Figure 3 shows the current-voltage characteristics of the detectors for each configuration. As expected, a very high leakage current is seen in configuration 1 due to the lower MAPbI 3 bandgap. The dark current baseline for configurations 1-3 was unstable and hence provided inferior detector stability. The lowest dark current was obtained using a single layer of polymer B between the MAPbI 3 semiconductor layer and the contact (configuration 6). At 0.083 V/µm bias, the dark current was measured to be 1.29 × 10 -6 mA/cm 2 . Five of these detectors were encapsulated using an optical epoxy and put under bias for 240 days. The dark current baseline of the detectors was stable under constant bias voltage, and in fact, the dark current decreased to ~ 2.5 × 10 -7 mA/cm 2 over 240 days. The variations in the sensitivity of the MAPbI 3 detector was less than ± 2%. X-ray response data for one of these detectors is shown in Supplementary Figure S5 . On the other hand, after prolonged bias (~ 2 days), the baseline of all the other devices with configurations 4 and 5 started to show a significant amount of noise with up to a two-times increase in dark current. The encapsulation used for all these detectors were not fully optimized. Interaction with moisture and oxygen results in emanation of the organic species from the MAPbI 3 matrix, leaving it Pb-rich, thereby deteriorating the photo-response of these detectors 34, 37 . Hermetic encapsulation is essential for the long-term functioning of these detectors and is still a crucial challenge for perovskite materials that are currently under development for different applications 34, 38 . Several cation and anion doping schemes have been suggested to alleviate this stability problem and can be used to stabilize these X-ray detectors for long-term applications under ambient atmospheres [39] [40] [41] . Future studies will be performed on optimizing such detectors for long-term X-ray detection applications. The X-ray sensitivity of the detectors was characterized using the set up shown in Figure S3 . The detectorto-source distance was kept constant at around 20 cm. The detectors fabricated using configurations 1 through www.nature.com/scientificreports/ 5 demonstrated high sensitivities up to 17 µC mGy −1 cm −2 at an electric field of 0.08 V/µm. However, due to the lack of stability of the detectors and poor reproducibility, we focused our studies on configuration 6. The sensitivity values for the 1200 µm-thick detector were 1.9-7.5 µCmGy −1 cm −2 for 0.041-0.16 V/µm applied bias. For the 200 µm-thick detector, the sensitivity values were 7.5-13.5 µCmGy −1 cm −2 for 0.25-0.5 V/µm applied bias. Both these detectors were tested at 90 kV with 85 µA tube current. A comparison of the sensitivity values for detectors with different thicknesses are given in Fig. 4 . Here we can see that the sensitivity increases as the applied bias is increased. Also, for similar electric fields, thinner detectors show higher sensitivity, showing the effects of charge trapping inside the thicker detectors. The plot also shows the lower efficiency of the 200 µm detector in comparison to the 600 and 1200 µm detectors. Figures 5 and 6 show the linearity with respect to the incoming X-ray energy and X-ray exposure rate of the 1200 µm-and 200 µm-thick detectors, respectively. As can be seen, the X-ray response of both these detectors are linear and hence shows the feasibility of linear X-ray response of MAPbI 3 -based detectors. Figure 7 shows the response of the 1200 µm-and 200 µm-thick detectors at the same wattage. This again shows the highly linear and uniform response from two separate detectors. As expected, higher polarization effects were seen in the thicker detector. Figure 8 shows the difference in polarization of the detectors biased at different bias voltages. It is clear that higher bias voltages result in polarization effects, visible as a decay of the signal after the initial signal rise. The prolonged polarization effect turns off around 0.5 V/µm applied bias, whereas the fast polarization is eliminated at 0.2 V/µm. Figure 9 shows the falling edge of the 200 µm-thick MAPbI 3 detector after turning off the X-ray tube. The charge trapping in bulk contributes to the decay lag in these detectors. The lower decay lag in these detectors as compared to other polycrystalline semiconductors such as a-Se is due to the presence of shallow defects in MAPbI 3 crystallites 42 . Deeper traps result in longer decay times, thereby increasing the decay lag. Another factor that adds to this time lag is the delay in charge injection through the increased electric field generated due to the X-ray illumination at the contact electrodes and the barrier layer (such as the polymer B layer in configuration 6). We also estimated the mobility-lifetime characteristics of MAPbI 3 detectors using the classical Hecht-equation and obtained mobility-lifetime values in the order of ~ 10 -4 cm 2 /V. Figure 10 shows the data for one of these detectors. It is clear that these detectors possess excellent charge carrier properties and could potentially become the best performing sensor material for direct detection of higher energy X-rays. Table 1 for the details of the X-ray detector configurations. Table 1 . This table lists all the MAPbI 3 -based detector configurations that were tested in this study. The thickness of the electrical contacts was ~ 2 nm-thick, and the charge transport layers were ~ 2 µm-thick. Device structure Typical dark current at 0.08 V/µm (mA/ cm 2 ) X-ray sensitivity at 0.08 V/µm (µC mGy −1 cm −2 ) www.nature.com/scientificreports/ www.nature.com/scientificreports/ The performance of 1200 µm and 200 µm-thick detectors was also tested at the NSLS II XPD beamline at Brookhaven National Laboratory (BNL), Figure S4 . The BNL beamline safety regulations restricted the allowed voltage for any new equipment (such as the X-ray detector) being operated in the beamline room, which in turn limited the maximum electric field to 0.25 V/µm. In addition, these detectors were not hermetically encapsulated and were exposed to an ambient atmosphere for three days before they were tested at the BNL NSLS-II www.nature.com/scientificreports/ beamline. Figure 11 shows the response of these detectors under a monochromatic 70 keV synchrotron X-ray for more than 30 min. Although polycrystalline MAPbI 3 detectors yielded measurable responses under high energy monochromatic synchrotron radiation under a low applied bias, these responses merely represent the feasibility of these detectors for synchrotron detection and are not fully optimized. With further optimization, these detectors hold the potential to demonstrate much higher SNR, as demonstrated with the microfocus X-ray source. The future direction of this study will be focused on the development of large-area high spatial resolution FPXIs. This will include sensor material and detector structure optimization, design and fabrication of an appropriate pixelated backplane optimized for MAPbI 3, and testing of these detectors for synchrotron and medical imaging applications. In conclusion, we established and validated the path forward for a new generation of polycrystalline X-ray detectors that have applications in numerous fields that require large area FPXIs, specifically in medical and synchrotron imaging. Not only MAPbI 3 -based semiconducting layers are highly efficient and highly sensitive to X-rays with sensitivity values as high as 13.5 µCmGy −1 cm −2 as measured using a microfocus X-ray source, but they are also easily manufacturable and highly reliable for long-term applications when optimally encapsulated. These multi-layered detectors exhibit extremely low dark currents in the range of ~ 1 nA/cm 2 under a high bias voltage of 1 V/µm and ~ 150 pA/cm 2 under a lower bias voltage of 0.25 V/µm, suitable for the fabrication of low noise FPXIs on active pixel array backplanes such as a-Si TFTs. We demonstrated a variety of performance and stability tests with these detectors, including detector response testing with monochromatic 70 keV synchrotron radiation at BNL. The epoxy-encapsulated detectors also showed stable dark current and X-ray detection sensitivity for over eight months in an ambient atmosphere. Detector fabrication. All the physical MAPbI 3 devices comprised of a 5 cm × 5 cm ITO-coated glass substrate (1850 Å thick ITO with a resistivity of 10-15 Ω per sq. in.) with Ag bus lines for electrical contact. We fabricated 2 cm × 2 cm MAPbI 3 sensor layers that were 600 µm-thick and deposited 1 cm × 1 cm electrical contacts on the top. The polycrystalline MAPbI 3 sensor layers were synthesized using a solvent-based process and deposited using doctor blading technique. Specific molar ratios of CH 3 NH 3 I and lead iodide (99.999% trace metal basis) was added to a polar solvent such as N-Methyl-2-Pyrrolidone (NMP) or γ-butyrolactone typically at a ratio of 60:40 wt% (solvent to solute). It was observed that the grain characteristics of the MAPbI 3 sensor are highly dependent on the starting stoichiometry of the MAI and PbI 2 precursors. Supplementary Figure S6 shows the differences in the grain characteristics. www.nature.com/scientificreports/ This mixture was then placed in a convection oven at 100 °C and was vigorously mixed for 24 h. As observed, the reaction initiates within five minutes, and the colorless solution with white precursors turns dark grey. Once the solution appeared homogeneous with a smooth texture, a co-solvent (e.g. α-terpineol) was added to the mixture for thicker sensors. The solution was again placed inside an oven at 95 °C and was vigorously mixed for the next 24 h. The solution was decanted, and the viscous MAPbI 3 was used to fabricate the X-ray detectors using the doctor blading technique. The hole-transporting and hole-blocking layers were fabricated by mixing MAPbI 3 and MAPbBr 3 with polymer A respectively. These layers were deposited using spin coating. A similar strategy was used by Kim et al. using polyimide based polymers 43 . The top and bottom electrodes were chosen to be chemically non-reactive oxide contacts (e.g., SnO 2 , ITO) for halide semiconductors, as shown in our previous study 44 . Flat panel detectors for digital radiography, medical imaging, physics and psychophysics Digital mammography: A review of technical development and clinical applications Digital detectors for mammography: The technical challenges Status and prospects of digital detector technology for CR and DR Flat-panel detectors: How much better are they? Review of medical imaging with emphasis on X-ray detectors Flat-detector computed tomography (FD-CT) Advances in digital radiography: Physical principles and system overview Recent developments in digital radiography detectors On the development of digital radiography detectors: A review Flat detectors and their clinical applications Performance of a high fill factor, indirect detection prototype flat-panel imager for mammography Status of direct conversion detectors for medical imaging with X-rays Amorphous selenium and its alloys from early xeroradiography to high resolution x-ray image detectors and ultrasensitive imaging tubes Recent advances in X-ray photoconductors for direct conversion X-ray image detectors Direct-conversion flat-panel x-ray image detectors Direct conversion flat panel X-ray image sensors for digital radiography Amorphous semiconductors usher-in-digital X-ray imaging Optimizing sawing of boards for furniture production using CT log scanning Chapter 2 X-ray and neutron digital radiography and computed tomography for cultural heritage Frequency and distribution of chest radiographic findings in COVID-19 positive patients Ultrafast 3D reconstruction for X-ray real-time tomography (RTT) Health impacts of low-dose ionizing radiation: Current scientific debates and regulatory issues Organic-inorganic perovskites: Structural versatility for functional materials design Rashba effect and carrier mobility in hybrid organic-inorganic perovskites The influence of the Rashba effect Electron-hole diffusion lengths > 175 μm in solution-grown CH 3 NH 3 PbI 3 single crystals Charge carrier lifetimes exceeding 15 μs in methylammonium lead iodide single crystals Recent progress in single-crystalline perovskite research including crystal preparation Unique properties of halide perovskites as possible origins of the superior solar cell performance Towards enabling stable lead halide perovskite solar cells; interplay between structural, environmental, and thermal stability Flexible inkjet-printed triple cation perovskite X-ray detectors Perovskite-filled membranes for flexible and large-area direct-conversion X-ray detector arrays Investigation of CH 3 NH 3 PbI 3 degradation rates and mechanisms in controlled humidity environments using in situ techniques Enhanced stability of MAPbI 3 perovskite solar cells using poly(p-chloro-xylylene) encapsulation Chemical management for colorful, efficient, and stable inorganicorganic hybrid nanostructured solar cells Pseudohalide-induced moisture tolerance in perovskite CH 3 NH 3 Pb(SCN)2I thin films Stabilizing perovskite structures by tuning tolerance factor: Formation of formamidinium and cesium lead iodide solid-state alloys Understanding of the formation of shallow level defects from the intrinsic defects of lead tri-halide perovskites Printable organometallic perovskite enables large-area, low-dose X-ray imaging Novel electrodes and engineered interfaces for halide-semiconductor radiation detectors The work discussed in this document has been partially supported by the US Department of Energy and US National Institute of Health. The authors would also like to extend their thanks to the NSLS II XPD beamline scientists and Dr. Yonggang Cui from BNL Nonproliferation and National Security Department for their help during the synchrotron measurements. A.D. and S.M. conceived the experiments and the research direction. A.D. synthesized and fabricated the X-ray detectors. A.D. characterized the detectors, including current-voltage and X-ray sensitivity properties. Z.Z. Performed the synchrotron beamline measurements. A.D. and S.M. wrote the manuscript. All authors discussed the results and commented on the manuscript. The authors declare no competing interests. Supplementary information is available for this paper at https ://doi.org/10.1038/s4159 8-020-76647 -5.Correspondence and requests for materials should be addressed to A.D.Reprints and permissions information is available at www.nature.com/reprints.Publisher's note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations. License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creat iveco mmons .org/licen ses/by/4.0/.