key: cord-0010390-xc65uhyu authors: Bruner, Terrance A.; Fedorko, Daniel P. title: Opportunities for rapid viral diagnosis date: 1993-05-01 journal: Clin Microbiol Newsl DOI: 10.1016/0196-4399(93)90015-f sha: 402170f5553d5261b59a806fee1417000f5d378e doc_id: 10390 cord_uid: xc65uhyu nan Rapid viral diagnostic test procedures vary in processing time from several minutes to hours depending on the method used to detect an unknown virus. Several rapid viral diagnostic assays are reviewed in this paper. The assays are all available commercially or can be performed using readily available supplies. The assay methods include enzyme immunoassay (EIA), direct immunofluorescence (DFA), and latex agglutination. In most of the assays reviewed, the mechanism of identification involves detection of viral antigen in clinical specimens. Except for DFA, the assays require minimal training or specialized equipment. Many of the procedures can be performed in laboratories during off hours and without specialized facilities. The viruses represent those that are clinically important and have rapid diagnostic tests available. Rapid identification of respiratory syncytial virus (RSV) is important for two reasons: (i) timely cohorting of patients can prevent nosocomial spread of disease and (ii) appropriate antiviral therapy can be initiated early. In winter months when the incidence of RSV in-fection is high and hospital admissions of patients increase, viral culture techniques do not allow for expedient cohorting of patients because 5 to 7 d are required before test results become available. Rapid identification techniques include DFA and several EIA systems. Methods that will be discussed include DFA, Directigen RSV (Becton Dickinson Microbiology Systems), and Test-Pack RSV (Abbott Laboratories). DFA, Directigen, and TestPack can all be processed in under 1 h. Ortho RSV, Kallestad Laboratories EIA, and Abbott Laboratories RSV EIA will not be reviewed because these "rapid" tests require several hours to process. EIA systems are more suitable during the RSV "off season," when tests can be batched, or in high-volume laboratories. DFA is considered by some to be the most sensitive technique for the identification of RSV from clinical specimens (1-3). The method requires approximately 30 min for incubation and additional time to examine the slide (4). Although DFA is rapid and sensitive, the technique requires a fluorescence microscope, high quality specific reagents, and trained, experienced personnel to prepare, examine, and interpret the stained specimen. There are two EIA test kits that provide rapid diagnosis, are self-contained, and do not require specialized training or equipment. The Directigen and Test-Pack kits are membrane enzyme immunoassay systems with assay times of 15 and 20 min, respectively. Both test kits contain internal controls. Directigen May 1,1993 RSV shows a lower sensitivity and specificity than TestPack when fresh specimens are used (2) ( Table i) . Wrenn et at. report that TestPack is more sensitive than viral culture (5). Directigen specimen samples, after a period of freezing and thawing, show increased sensitivity and specificity when compared to culture (6). The increase in sensitivity and specificity is thought to result from increased disruption of cellular material with the result that more antigenic material is available for detection (3). Current Directigen assay instructions include modifications that suggest that increased cellular disruption may be required for optimal test results (6 8.5% of specimens tested. The inability to interpret the test arises from difficulties with f'dtration of the specimen into the device. The f'dtration problem can be overcome by diluting the specimen (3). Additional processing of the specimen potentially increases the test turnaround time and may reduce the sensitivity of the test by diluting the specimen. The Directigen test kit has an improved filtration system that should help resolve the filtration problems reported previously. Influenza A virus is a pathogen associated with resph'atory tract infection. The disease spectrum ranges from mild to severe and may cause fatal pneumonia. Influenza A is associated with epidemics and can be lethal in the elderly and immunoeompromised individuals. Rapid detection of the influenza A virus is important because effective antiviral agents, amantadine and rimantidine, are available (7, 8). Commercially available identification techniques include membrane EIA and DFA. A membrane EIA kit is Directigen FLU-A (Becton Dickinson Microbiology Systems). The test can be completed in under 15 min and does not require a fluorescence microscope or specially trained personnel to be interpreted. The sensitivity and specificity of Directigen FLU-A, compared to cell culture, range from 65 to 100% and 91.6 to 96%, respectively. Positive (PPV) and negative predictive values (NPV) range from 62.6 to 85% and 89 to 100%, respectively (7, 9). The sensi-tivity, specificity, PPV, and NPV of Directigen FLU-A compared to DFA are 100, 91.6, 75, and 100%, respectively (7). In contrast to the Directigen RSV membrane EIA, there is evidence that freeze-thawing may compromise the ability of the Directigen FLU-A test to identify positive specimens. Specimens should not be frozen prior to testing with the Directigen test until the freezethaw issue has been resolved (7). Although the Directigen test is a valuable tool, it must be used appropriately in laboratory practice. Direetigen FLU-A is more expensive per test than cell culture and should be used when rapid results are likely to affect patient management. In the cooler months of the year in the United States, rotavirus infections account for approximately 50% of pediatric hospital admissions due to diarrhea and dehydration (10). Each year rotavirus epidemics follow a regional sequence from F~t to West. Outbreaks occur in November and December in the Southwestern states and spread to the New England states in April and May (11). The rapid detection of rotavirus aids in the diagnosis of gastroenteritis and in the identification of infected individuals who are sources of outbreaks in settings such as day-care centers and hospital wards. Healthcare workers must be aware that other viruses, including the fastidious adenoviruses, Norwalk and Norwalk-like viruses, ealiciviruses, aslroviruses, and possibly coronaviruses, also may be etiologic agents of gastroenteritis (10). Commercial kits for the rapid detection of rotavirus antigen in stool specimens have been available for many years. These include those with enzymelinked immunosorbent assay (ELISA), membrane EIA, and latex agglutination methodologies. Comparisons of results from evaluations of commercial rotavirus assays are quite difficult because of the variety of gold-standard reference tests used by investigators and the variability of results (12). Several monoclonal and polyclonal antibody-based ELISA kits are available for use in laboratories that receive many requests for rotavirus testing. Monoclonal antibody--based ELISA kits appear to be superior to polyclonal antibody-based kits (13, 14). Rapid rolavirus assays that have a 15 min or less performance time include latex agglutination and a membrane EIA from Abbott Laboratories Cl'estPack Rotavirus). Latex agglutination tests from Wampole Laboratories (Virogen Rotatest), Meridian Diagnostics, Inc. (Meritec Rotavirus), Murex Diagnostics, Inc. (formerly Wellcome Diagnostics, Wellcome Rotavirus), BioMerieux (Slidex Rota-kit), and Medical Technology Corporation (Rotalex) have lower sensitivity than an ELISA; however, they are good tests for laboratories that receive few requests tbr rotavirus testing or for off-season testing ( Table 2) . TestPack Rotavirus membrane EIA is also useful for small volume and offseason testing. Brooks et al. demonstrated that the TestPack performed better with fresh stool specimens (sensitivity 95%, specificity 90%, PPV 86%, NPV 97%) than with frozen specimens (16) . When receiving evaluations of TestPack, readers should notc whether fresh or frozen specimens were used. The rapid detection of herpes simplex virus (HSV) antigen is useful in cases of pregnant women with genital NOTE: No ~pomibility is almmed by the Publisher for any injury and/or damage t~ persmw or pmpeay as a matter of products liability, negligence or otherwise, or from any use or operation of any methods, predn~, imtmctinm or idcaa contained ia the matet~l hereim No militatedrut or procedure should be carried out nnl~. in the reader's judgment, its risk is justified. Beceme of ra#d advalw.m ia medical aeieae~, we rececn~md that the independag verification of diagnm~ and drug dosages should be made. Discussions. views, and recommendations as to medical precedut~es, choice of drags, and drag dosages are the respoosibility of the authors. infections and individuals who are particularly susceptible to severe infections. Several enzyme immunoassay kits are available that provide more rapid test results than the "gold standard" technique, cell culture, for detecting HSV antigen (Table 3 ). Cell culture techniques may require up to 14 d for a definitively negative result (16) . HERPCHEK (DuPont) is a microdilution plate-based enzyme immunoassay that requires approximately 50 min of hands-on time by the technologist and 4 h to complete. The assay does not require specialized laboratory equipment or highly trained personnel. Specimens must be collected in the proprietary transport media, HERPTRAN. There is no requirement for rapid transport of the specimen to the laboratory (16, 17) . Verano et al. suggest that HERPCHEK is a sensitive ELISA for symptomatic herpes lesions even if the specimen is transported in viral transport medium instead of HERPTRAN (18) . The combined labor and reagent costs for HERPCHEK are comparable to the total cost of HSV antigen detection using shell vial culture techniques. The HERPCHEK test does not differentiate between HSV types 1 and 2 (19) . The Vitek ImmunoDiagnostic Assay System (VIDAS, Vitek Systems Inc.) is a qualitative enzyme-linked fluorescent immunoassay system. The VIDAS system is fully automated and can carry out internal washes and incubation steps. Final result calculations are also performed by the system. The test results can be available in less than 3 h. The VIDAS system is limited to 13 samples per run, whereas the HERP-CHEK assay can perform up to 94 samples per run. The assay does not provide virus typing. Cell culture backup is recommended for negative HSV results obtained by VIDAS (19) . The IDEIA (NovoNordisk Ltd.) kit is an EIA that detects HSV antigen. The assay can be completed in less than 4 h. Sillis suggests that results obtained using brain tissue should be interpreted cautiously in the absence of other supporting evidence of infection (20) . SureCell (Kodak), a filtration EIA system, is performed directly with the clinical specimen. Results of the test are available within 15 rain. No instrumentation or special transport media, other than viral transport media, are required. Dorian and others suggest that all negative test results should be followed up by inoculation of cell culture (21, 22) . Rapid detection of cytomegalovirus (CMV) infection is useful in preventing overtreatment of patients who are receiving immunosuppressive therapy, to initiate antiviral therapy, and to determine if donor organs are acceptable for transplantation. The current standard method for the detection of CMV infection is cell culture. The test is slow, requiting an aver-age of 9 d (range: 7 to 2I d) to complete. The shell vial assay is more rapid than conventional tube cell cultures, requiring an average of 16 h to complete (23) . Gleaves et al. suggest that the shell vial technique is as specific and more sensitive than the conventional tube cell culture (23) . The shell vial assay uses a commercially available monoclonal antibody directed against CMV immediate early antigen (23) . An antigen detection system has been described that uses monoclonal antibodies directed against immediate early CMV antigen in cytocentrifuged blood leukocytes. The process requires 3 to 5 h to complete (24) . The technique has an overall sensitivity of 93% and specificity of 92% compared to viral isolation, IgG and IgM ELISA, and clinical symptoms (25) . van der Bij et al. suggest that the process is as specific as and more sensitive than current isolation techniques (24) . Reagents and equipment necessary to perform the assay are readily available. The clinical virology laboratory should be able to perform the procedure without diMculty. A commercially available kit (CMVvue kit: INCSTAR Corp.) contains all of the reagents required to detect CMV antigenemia. The CMV-vue kit does not require the use of a cytocentrifuge and does not require a fluorescence microscope because CMV-infected polymorphonuclear blood leukocytes are stained with horseradish peroxidase. In a recent evaluation Erice et al. (26) reported sensitivities, specificities, PPVs, and NPVs of 87, 92, 65, and 98%, respectively, for CMV-vue and 69, 96, 73, and 93% for the shell vial assay compared to cell culture isolation. Erice et al. recommend CMV antigenemia detection as the method of choice The standard method used for donor organ screening consists of antibody detection. A commercially available antibody detection kit is CMV Scan. (Becton-Dickinson) . CMV Scan is a latex agglutination test performed on donor sera and can be performed in under 20 min. In a prospective comparison to Abbott ELISA (CMV TOTAL AB ELISA), CMV Scan has a sensitivity of 98% and specificity of 97% (27). PPV and NPV are 98% and 97%, reslx~tively (27). Zbinder et al. suggest that negative results should be retested against a "highly sensitive EIA" and that sensitivity can be improved by donble-checked reading (27). Rapid nucleic acid technology is not yet available for the clinical virologist. Probes are commercially available for viral detection in situ and for staining shell vial coverslips, but these reagents are no more rapid than monoclonal antibody assays. The polymerase chain reaction (PCR) has been applied to the detection of several viruses in clinical specimens but PCR technology will be limited to reference laboratories until commercial kits are available. Clinical viral testing has evolved from an epidemiologic exercise to a practical tool providing useful diagnostic information. The more rapidly available test results allow physicians to make better patient management decisions. Clinical microbiologists need to educate the health care community with regard to availability and appropriate use of viral testing. Expectorated sputum is the specimen most commonly collected for microbiologic diagnosis of bacterial pneumonia because it is considered easy and safe to collect. Better quality specimens, such as transtracheal aspiration, transthoracic needle aspiration, and bronchoscopic aspirates are uncomfortable for the patient, uncommonly associated with significant risk, or are contraindicated in certain patients. Thus the indication for such specimens remains a clinical decision in each patient's case. Analysis of the expectorated sputum, however, is fraught with controversy. Many authorities in the field became skeptical of expectorated sputum culture over 20 yr ago when Barrett-Conner (1) reported that the yield of Streptococcus pneumoniae in patients with pneumococcal pneumonia was only 50%, which prompted her to say, "The routine sputum culture for the diagnosis of acute bacterial pneumonia may be a sacred cow." This accusation resulted in multiple attempts to improve the quality of the sputum culture by washing, quantitation, and microscopic screening. Washing the sputum with sterile saline to remove salivary contamination followed by quantitatively culturing the retained purulent portion (after straining through a tea strainer) seemed to work reasonably well (2) . Even though the method is popular in Europe, it never caught on as a routine laboratory procedure in the United States. The method most often employed in the United States' clinical laboratories is the microscopic screening of Gram-stained sputum samples viewed under low-power (10× objective) to determine the extent of salivary contamination by determining the relative numbers ofpolymorphonuclear leukocytes (PMNs) and squamous epithelial cells (SEC) (3-7). Although some authorities advocate microscopic examination as a useful gauge of the specimen's quality and to decide whether to "reject" or culture specimens, there is no consensus regarding the specific criteria needed. In addition, experience between laboratories is sometimes very different. The original report by Bartlett (7) scored sputum specimens on the basis of the gross appearance and the numbers of PMNs and SEC. Generally those specimens containing mucus, at least 10 PMNs, and <25 SEC per lowpower field (LPF) were acceptable for culture, resulting in rejection of 21% of the specimens. In the next report from the Mayo Clinic, Murray and Washington (4) used the acceptability criteria of >25 PMNs and <10 SEC per LPF and rejected 75% of the specimens. In this report, the appearance of sputum was considered an "unreliable indicator of oropharyngeal contamination." Shortly after, in another report from the Mayo Clinic by Van Scoy (6), the criteria were modified to accept any sputum specimen with >25 PMNs per LPF so that only about 25% of the sputum specimens were discarded. The most recent report was from Duke University Medical Center (8) using the crtefia of < 10 SEC per LPF and the presence of organisms on the Gram-stain to accept endotracheal suctions for cultures. In this study the "rejection" rate was 55%. The authors noted that the quantity of PMNs did not predict the quality of the endotracheal suctions. The utility of Gram stain for interpretation of expectorated sputum is also controversial. Gram stain is a subjective analysis, and there is considerable variation in expertise. In view of CLIA '88 and the lowered standards for technical personnel, it is technical expertise that must be emphasized. Critical therapeutic decisions based on these stains probably should not be entrusted to personnel who do not have special training. The main problem is the failure to recognize coccobacillary gram-negative bacilli resembling Haemophilus influenzae and Pseudomonas cepacia on Gram stains of sputum and endotracheal suctions. It is crucial not to overlook the fact that pneumonia is primarily a clinical diagnosis. Clinicians have several sources of information for evaluating a patient's pulmonary problem, including history, physical examination, and nonmicrobiologic laboratory tests such as the chest radiograph and white blood cell count. The sputum culture is used to establish etiologic diagnosis rather than to serve as a test for confirming or disproving the diagnosis of pneumonia. The sputum culture is not pathognomonic of pneumonia or completely reliable in demonstrating the etiologic agent if pneumonia exists. This is best demonstrated in a study conducted at Comparison of two rapid methods for detection of respirato~ syncytial virus (RSV) (TestPaek RSV and Ortho RSV ELISA) with direct immmaofluorescenee and virus isolation for the diagnosis of p~ialric RSV infection Reliability of . new test kits for rapid diagnosis of respkatory syncytial virus infection Comparison of Dircctigcn RSV with viral isolation and direct immunofluorescence for the identification of respiratory syncytial virus Evaluation of five methods for respiratory syncytial virus detection Detection of respiratory syncytial virus antigen in nasal washings by Abbott TestPack Enzyme Immunoassay Evaluation of the Becton Dickinson Directigen test for respiratory syneytial virus in nasopharyngeal aspirates Comparison of Directigen FLU-A with viral isolation and direct immunofluorescence for the rapid detection and identification of influenza A virus A comparison of direct immunofluorescence, shell vial culture, and conventional cell culture for the rapid detection of influenza A and B Christen.sen, M. L. 1989. Human viral gastroenteritis Annual rotavirus epidemic patterns in North America. Results of a 5-year retrospective survey of 88 centers in Canada, Mexico, and the United States Comparison of seven kits for detection of rotavims in fecal specimens with a sensitive, specific enzyme immunoassay Comparison of four latex agglutination (LA)and three enzyme-linked immunosorbent assays (ELISA) for the detection of rotavims in fecal specimens Comparison of nine commercial mununoassays for the detection of rotavirus in fecal specimens Comparison of a new rapid test (TestPack Rotavirus) with standard enzyme immunoassay and electron microscopy for the detection of rotavirus in symptomatic hospitalized children Diagnosis of Herpes simplex virus infection in a clinical setting by a direct antigen detection enzyme immunoassay kit Evaluation of the Du Pont HERP-CHEK Herpes Simplex virus antigen test with clinical specimens Herpes simplex virus antigen direct detection in standard virus transport medium by Du Pont HERPCHEX enzyme-linked immunosorbent assay Evaluation of an automated immunodiagnostic assay system for direct detection of herpes simplex virus antigen in clinical specimen Clinical evaluation of enzyme immunoassay in rapid diagnosis of herpes simplex infections Detection of herpes simplex virus by the Kodak SureCell Herpes Test Evaluation of a visual, rapid, membrane enzyme immunoassay for the detection of herpes simplex virus antigen Comparison of standard tube and shell vial cell culture techniques for the detection of cytomegalovinls in clinical specimens Comparison between viremia and antigenemia for detection of cytomegalovirus in blood Rapid immunodiagnosis of active cytomegalovirus infection by monoclonal antibody staining of blood leucocytes I