key: cord-0001624-mpuovmeb authors: Bratcher, Preston E.; Gaggar, Amit title: Factors Influencing the Measurement of Plasma/Serum Surfactant Protein D Levels by ELISA date: 2014-11-03 journal: PLoS One DOI: 10.1371/journal.pone.0111466 sha: dc4332792ad6c643b1142cef1d344755a64b9625 doc_id: 1624 cord_uid: mpuovmeb BACKGROUND: Extensive variations in human surfactant protein D (SP-D) levels in circulation as measured by ELISA exist in the published literature. In order to determine the source of these variations, factors influencing the measurement by ELISA were explored. MATERIALS AND METHODS: Peripheral blood from healthy individuals was collected into various vacutainers during the same blood draw. Recombinant SP-D was diluted into different matrices and used for a standard curve. Samples were analyzed by capture ELISA using one of two distinct detection antibodies. RESULTS: The type of matrix had some effects on detection of recombinant SP-D. The type of anticoagulant used and dilution factor had very little effect, except for in plasma collected in EDTA vacutainers. The extent of variation in published values seemed to be due to the ELISA configuration employed, and, in agreement with this, we found that by switching the detection antibody, there was a 50% decrease in the extrapolated SP-D value of serum and plasma samples. Storage of samples resulted in slight changes in measured SP-D levels. CONCLUSIONS: The ELISA configuration employed to measure circulating levels of SP-D has a significant effect on the extrapolated values. In both configurations tested, the use of EDTA as a coagulant resulted in inconsistent values, and we, therefore, suggest the avoidance of this anticoagulant when assaying for SP-D by ELISA. While the demonstrated effects of several factors on measurement of SP-D may not account for all the disparities amongst the previous studies, they stress that variations in methodologies for measuring the same protein can result in very inconsistent results. Surfactant protein D (SP-D) is a pulmonary collectin involved in regulation of inflammation, innate immune defense, and surfactant homeostasis. It is expressed by Clara cells and alveolar type II cells in the lung. SP-D has a multimeric structure which gives it the ability to agglutinate pathogens, as well as aid in the clearance of apoptotic cells, cellular debris, and foreign particles in the lung [reviewed in [1] ]. Circulating levels of SP-D have been examined for their potential use as a biomarker in various diseases including dermatitis [2, 3] , acute lung injury (ALI)/acute respiratory distress syndrome (ARDS) [4] [5] [6] [7] [8] [9] [10] [11] [12] [13] , periodontitis [14] , interstitial pulmonary fibrosis (IPF) [10, 12, [15] [16] [17] [18] [19] [20] [21] [22] [23] , chronic obstructive pulmonary disease (COPD) [15, [24] [25] [26] [27] [28] [29] [30] [31] [32] [33] [34] [35] [36] , emphysema [37] , cystic fibrosis (CF) [15, 38, 39] , coronary disease [40, 41] , sclerosis [42] [43] [44] [45] [46] , cancer [47, 48] , sarcoidosis [21, 49] , allergies [28, [50] [51] [52] , rheumatoid arthritis [53, 54] , and respiratory infections [18, [55] [56] [57] [58] [59] [60] . SP-D levels have also been proposed to correlate with genetic elements [61] [62] [63] , body mass index (BMI) [64] [65] [66] [67] [68] , age [65] , circadian rhythm [69] , and with particle exposure [70, 71] and cigarette smoking habits [25, 27, 28, 31, 37, [72] [73] [74] [75] [76] [77] . In addition, there have been studies examining the levels of SP-D in subjects with Turner syndrome [78] , paraquat intoxication [79] , swimming in variably treated waters [80] , lung transplant patients [81] , patients undergoing neurosurgical operations [82] , drowning victims [83] , polymyositis/dermatomyositis [84] , dementia [85] , lupus [86] , and sleep apnea [87] . Similarly to CC-16 and KL-6, SP-D is thought to be a marker of pulmonary leak into the vasculature [88] , and therefore alveolar destruction would result in an increase in levels of these pulmonary proteins in the blood. However, protein levels in lung do not always correlate with protein levels in blood [35] , suggesting the possibility of alternative mechanisms affecting SP-D levels in circulation. Various commercially available kits and non-commercially available ELISA configurations have been used to compare SP-D levels in plasma and serum from normal, healthy controls and the various disease states described above. Interestingly, there is a very substantial discrepancy between the reported values of the healthy control populations between studies, as well as in the magnitude of the range of values in this population. While the ELISA configuration used to measure SP-D seemed to have a large impact on the values reported, there are significant variations in the healthy control SP-D levels and range amongst reports using the same configuration. The purpose of this study is to determine factors affecting the measurement of SP-D by ELISA that may, therefore, explain the variations of serum and plasma SP-D levels reported in the published literature. All human studies were approved by the University of Alabama at Birmingham Institutional Review Board for Human Use with all subjects providing written consent. Peripheral blood was collected from healthy volunteers by venipuncture into serum, heparin sulfate, K 2 EDTA, and sodium citrate vacutainers (BD Biosciences) during a single draw. Samples were kept at room temperature until blood in the serum tube was coagulated. Afterwards, samples were centrifuged at 400xg for 10 minutes to separate blood cells from serum or plasma. Samples were either directly aliquoted and stored at 280uC or given an additional round of centrifugation at 3000xg for 10 minutes to separate platelets from serum or plasma. All serum and plasma sample values depicted in figures were free of platelets. All experiments had 6 samples per group except for the experiments depicted in Mouse antihuman SP-D capture antibody, biotinylated mouse antihuman SP-D detection antibody, streptavidin conjugated to horse radish peroxidase, and recombinant human SP-D standard were purchased as a kit from R&D Systems (Catalog # DY1920) and used according to the manufacturer's protocol. For experiments testing the effects of different matrices on the detection of recombinant human SP-D, a concentrated stock was made in PBS with 1% BSA from separately purchased recombinant human SP-D expressed in NSO cells (R&D Systems, Catalog # 1920-SP-050). For studies comparing detection antibodies, polyclonal goat antihuman SP-D antibody (R&D systems, Catalog # AF1920) was used instead with a monoclonal mouse antigoat conjugated to horse radish peroxidase (Sigma) used as a secondary. All sample values were extrapolated from a second order polynomial curve fit of 9 concentrations of the standard (two-fold dilutions with a high concentration of 40 ng/mL) diluted in 1% BSA PBS. Statistical analyses were performed as described in the figure legends using Prism 5 (GraphPad Software), and all reported p values are twotailed. All error bars represent standard deviation. In order to determine the mean and range of SP-D levels in the serum/plasma of normal, healthy individuals, we performed a non-exhaustive search of the literature which revealed more than 60 publications in which these values were described. Interestingly, an unexpected amount of variation on these reported values was observed. When grouping according to the ELISA configuration used, one configuration consistently resulted in significantly higher values than the two other configurations for which multiple references were obtained (Figure 1a ). For each configuration, a substantial range of means/medians was observed in the healthy control population. In addition to the variations seen in the averages, the spectrum of values in all populations ranged from 55.9 pg/mL [22] to 3.9653 mg/mL [89] , representing a more than 70,000 fold difference. When examining the range of values seen in healthy subjects from each individual study and grouping according to the ELISA configuration used, a large variation in range was observed ( Figure 1b) . Overall, the study with the smallest range for any population (as a percentage of the average) reported a 95% confidence interval of 95.7-109.0 [54] , while the study containing the population with the largest range reported a standard deviation of 327% of the mean value [42] . Based on the above observations, we can infer that the largest difference in the measured levels of SP-D is due to the ELISA configuration employed, but that there are still significant differences between the averages of healthy individuals from studies using the same configurations. In order to compare the results of various studies in another manner, we examined the fold change from healthy populations for the four diseases/conditions for which a substantial number of publications were available (IPF, CF, cigarette smoking, and systemic sclerosis). This would allow us to control for ELISA configurations employed, technical protocol followed for the ELISA as well as for sample collection and processing, and any differences in the populations (as controls have been appropriately matched). The expectation was that the fold change from healthy would be very similar for each disease/ condition. However, large differences in the extent of this change were observed (Figure 1c ). In order to determine if the matrix used to generate the serial dilutions of recombinant human standard (rhSP-D) had any effects upon measurement, we compared values produced by dilutions in PBS, solutions of BSA (in PBS), and solutions of FBS (diluted in PBS) (Figure 2 ). At higher concentrations of protein (i.e. 5% BSA, 50% FBS, and 100% FBS), a decrease in the amount of rhSP-D detected was observed, with the amount measured in samples containing 10 ng/mL rhSP-D being the most variable of any samples measured in this assay. Using the manufacturer's recommended matrix (1% BSA) produced results very similar to a 10% FBS matrix, with both giving very consistent measurements and the greatest values. The values for PBS matrix gave similarly consistent measurements, but at a slightly lower value. This effect may be due to adsorption of rhSP-D without a carrier protein to the tubes used for serial dilutions [90] . Serum Matrix (Millipore), which had background levels of SP-D, also inhibited detection of recombinant SP-D at higher concentrations. All further standard curves were established by serial dilutions of recombinant SP-D in 1% BSA in PBS. One factor that was found to differ between studies and could, therefore, be a source of variation of reported healthy population values, was the type of anticoagulant (or lack thereof) in the collection container. When blood was simultaneously collected in various vacutainers, serum and heparin plasma gave similar measurements of SP-D, while citrate plasma gave values significantly lower than serum values (Figure 3 ). EDTA plasma gave the most inconsistent results, and values were also significantly lower than serum values. In order to determine if calcium concentration in the sample had a significant effect upon the measurement of SP-D, rhSP-D was assayed in serially diluted CaCl 2 or EDTA. The detection of rhSP-D was only slightly effected by the changes in calcium concentration in the sample (Figure 4a ). Additionally, we examined the effects of calcium concentration in the human samples by adding calcium to EDTA plasma samples and by adding EDTA to serum samples. While reconstituting the free calcium in the EDTA samples had little effect, the addition of EDTA to the serum samples had a dramatic effect that was inconsistent amongst the patient samples. While many ELISAs employ the use of PBS lacking divalent cations as a buffer during the detection of antibody-captured antigens, the ELISA configuration employed by Holmskov et al. uses a Tris-buffered saline solution containing 5 mM calcium chloride, as the monoclonal detection antibody binds to SP-D in the presence of calcium [58] . In addition, it was previously demonstrated that SP-D has the ability to bind to various immunoglobulins in a calcium-dependent manner [91] . In order to determine the effects of calcium in the context of SP-D detection by ELISA, ELISAs were performed with antibodies diluted in buffer with or without calcium. We examined these effects on both recombinant SP-D and SP-D in plasma with either the ELISA kit's detection antibody or a polyclonal goat anti-SP-D antibody produced by the same manufacturer. In all cases but one, the addition of 5 mM calcium to the dilution and wash buffer resulted in a small (,17%) but significant increase in SP-D concentration relative to detection in the absence of calcium (Figure 4b ). It is important to note that with the kit reagents, since the relative increase in detection of recombinant SP-D and SP-D in plasma in the presence of calcium is very similar, the extrapolated SP-D concentration in plasma when using a standard curve of recombinant SP-D should not significantly change. Interestingly, while the inclusion of calcium increased the recognition of native SP-D by the polyclonal antibody, it had no effect on the detection of recombinant SP-D by this antibody. Although it is beyond the scope of this study, future work will explore whether this effect is due to an increase in antibody recognition of antigen, increase in non-specific binding of antibodies by captured SP-D, or an increase through another mechanism. Given that some of the variation seen in the published literature might be explained by the use of different antibodies, we detected SP-D in serum and plasma samples using either the ELISA kit's detection antibody or the polyclonal goat anti-SP-D antibody. While there is no significant difference between detection by the [12, [16] [17] [18] [19] 21] , CF [15, 38, 39] , cigarette smoking [28, 31, [72] [73] [74] [75] , or sclerosis [42] [43] [44] 46] was different between publications. An asterisk (*) denotes p#0.001 by one-way ANOVA with Tukey's multiple comparison test. doi:10.1371/journal.pone.0111466.g001 kit antibody versus the polyclonal goat antibody with regard to the type of collection tube used, overall, there is a ,50% reduction in the extrapolated value produced by detection with the polyclonal goat antibody compared to the kit antibody ( Figure 5 ). In both cases, the same capture antibody was employed, and it is therefore possible that using a different capture antibody could have a similar effect on varying the extrapolated SP-D concentration in the sample. Another factor that fluctuated from study to study was whether the sample was assayed neat or the sample was prediluted (and the amount the sample was diluted). Given this fact and that diluting recombinant SP-D into 100% FBS had some influence in the detection of said SP-D, we compared SP-D values detected in undiluted and diluted samples. There was a very minimal difference between values produced by undiluted and 10 fold diluted serum, citrate plasma, and heparin plasma ( Figure 6 ). Ten fold diluted EDTA plasma, however, produced a significantly higher extrapolated SP-D value when compared to the same sample undiluted. This same effect was seen when the polyclonal goat antibody tested above was used for detection (data not shown). While the most common processing technique involved separating the blood cells (but not platelets) from serum and plasma and storing this sample at 280uC until assayed, some variations in processing and storage were present. To address this, we assayed serum and heparin plasma without platelets immediately after processing or after storage at 4uC, 220uC, or 280uC ( Figure 7) ; there were no significant differences between conditions. Additionally, there were only subtle differences between a sample stored at 220uC and 2 80uC, and depending on whether the sample went through a freeze/ thaw (FT) cycle, contained platelets or not, or was spun after thawing to separate any precipitates/debris (data not shown). SP-D concentrations of samples stored at 220uC and 280uC for two weeks were not different from the same samples at one week (data not shown). This study provides experimental evidence that variations in the anticoagulant used and ELISA configuration can have dramatic effects upon the measured SP-D level. Storage and processing of samples as well as diluent used for the standard have minor effects on the level of SP-D extrapolated by ELISA. Although these results may partially explain the variations seen in reported SP-D values in the blood, it is important to note the caveats associated with this review of the literature. For the comparison of healthy control groups amongst various studies, it is possible that differences in ethnic background and/or geographical location from which subjects were recruited may contribute to the variations seen when the same ELISA configuration was employed, as could the type of anticoagulant. However, we would expect the range of samples between healthy populations of different studies to be similar. In addition, assuming that an appropriately matched control population was used for comparison to the disease population, the fold change between healthy and disease should be more similar than what was observed in the literature. Measuring SP-D by ELISA presents a problem due to its multimeric structure. For each molecule (a complete dodecamer) of multimerized protein, there are at least 12 of the same epitopes present. Due to this property, it is not necessary to use discrete antibodies in a sandwich ELISA format to capture and then detect SP-D; even the same monoclonal antibody could be used for both. However, this may cause a difference in read out compared to using antibodies that detect different epitopes, due to the fact that the degree of multimerization has been shown to vary in the lungs during different disease states [92] and using the same monoclonal antibody would fail to detect a monomer. Even using two monoclonals that recognize distinct epitopes to measure SP-D in its native state would be confounded by differences in the degree of multimerization. For example, if an antibody against the carbohydrate recognition domain of SP-D was used to capture and an antibody against the neck region of SP-D was used to detect, a single, captured monomer of SP-D would be able to bind a single anti-neck antibody. However, a single, captured dodecamer would be able to bind 12 anti-neck antibodies. In order to accurately establish absolute concentrations by ELISA, it would be important that the multimeric structure of the standard was equivalent to that of the samples, as there could be differences in antigen binding in higher order multimers due to steric hinderance. Unfortunately, we did not have the necessary resources to examine the influence of the degree of multimerization on ELISA measurements. In addition to the possible effects caused by its multimeric nature, there is evidence in the literature to variable posttranslational modifications of SP-D in native samples. These modifications may, in fact, mask the epitope recognized by an ELISA antibody. This may account for one of the largest differences seen in the reported ELISA measurements of SP-D in serum/plasma, as the highest concentrations of SP-D came from the non-commercially available ELISA which used purified, native SP-D as a standard; this is opposed to almost all others, which used recombinantly expressed SP-D as a standard. Regardless of whether the antibodies used were raised against recombinant SP-D or purified native, because the post-translational modifications can vary from one patient to the next, this presents a significant obstacle to accurate measurements of natural SP-D using antibodies. The results of this study demonstrate the influence that the antibody may have on the SP-D concentration as measured by ELISA. In our experiments, it is important to note that the same recombinant standard and human serum/plasma samples were analyzed. Given the confounders mentioned above, it is possible that the antibody from the kit recognizes an epitope that is not post-translationally modified or altered in human serum/plasma SP-D, while several of the epitopes recognized by the polyclonal antibody are modified in human serum/plasma SP-D, but were unmodified in the recombinantly expressed SP-D (which was used as the immunogen). This would result in the observed relative decrease in binding of the native SP-D from the blood compared to the recombinant SP-D. Accuracy of SP-D measurements is critically important to the validation of this protein as a biomarker in pulmonary disease. Standardization of sample processing and storage, including the avoidance of EDTA as an anticoagulant, is necessary to ensure consistent results. Although absolute values may vary greatly due to the ELISA configuration employed, relative differences in SP-D concentrations amongst various disease groups should be consistent throughout the published literature. We are currently exploring alternative methods to quantify SP-D levels in the blood, as we believe this is necessary in order to establish the absolute concentration. Surfactant proteins SP-A and SP-D: structure, function and receptors Surfactant protein D (SP-D) and systemic scleroderma (SSc) Surfactant protein D in atopic dermatitis and psoriasis Plasma surfactant protein levels and clinical outcomes in patients with acute lung injury Effects of leukoreduced blood on acute lung injury after trauma: a randomized controlled trial Surfactant phospholipids, surfactant proteins, and inflammatory markers during acute lung injury in children Prognostic and pathogenetic value of combining clinical and biochemical indices in patients with acute lung injury Plasma CC16 levels are associated with development of ALI/ARDS in patients with ventilator-associated pneumonia: a retrospective observational study Plasma levels of surfactant protein D and KL-6 for evaluation of lung injury in critically ill mechanically ventilated patients Plasma biomarker profiles in acute exacerbation of idiopathic pulmonary fibrosis Biomarkers of lung epithelial injury and inflammation distinguish severe sepsis patients with acute respiratory distress syndrome Serum surfactant proteins-A and -D as biomarkers in idiopathic pulmonary fibrosis Serial changes in surfactant-associated proteins in lung and serum before and after onset of ARDS Increased plasma concentration of surfactant protein D in chronic periodontitis independent of SFTPD genotype: potential role as a biomarker Effect of single vs bilateral lung transplantation on plasma surfactant protein D levels in idiopathic pulmonary fibrosis Pulmonary surfactant protein D in sera and bronchoalveolar lavage fluids Enzymelinked immunosorbent assay using F(ab')2 fragment for the detection of human pulmonary surfactant protein D in sera Comparative study of KL-6, surfactant protein-A, surfactant protein-D, and monocyte chemoattractant protein-1 as serum markers for interstitial lung diseases Increased circulating levels of soluble Fas ligand are correlated with disease activity in patients with fibrosing lung diseases Monitoring markers of disease activity for interstitial lung diseases with serum surfactant proteins A and D Clinical importance of bronchoalveolar lavage fluid and blood cytokines, surfactant protein D, and Kerbs von Lungren 6 antigen in idiopathic pulmonary alveolar proteinosis Significance of serum vascular endothelial growth factor level in patients with idiopathic pulmonary fibrosis Pneumocyte biomarkers KL-6 and surfactant protein D reflect the distinct findings of high-resolution computed tomography in nonspecific interstitial pneumonia Integrating lung and plasma expression of pneumo-proteins in developing biomarkers in COPD: a case study of surfactant protein D Ageing and smoking contribute to plasma surfactant proteins and protease imbalance with correlations to airway obstruction Inflammatory biomarkers improve clinical prediction of mortality in chronic obstructive pulmonary disease Budesonide/formoterol enhances the expression of pro Surfactant Protein-B in lungs of COPD patients Exhaled metallic elements and serum pneumoproteins in asymptomatic smokers and patients with COPD or asthma Circulating surfactant protein D as a potential lung-specific biomarker of health outcomes in COPD: a pilot study The effects of fluticasone with or without salmeterol on systemic biomarkers of inflammation in chronic obstructive pulmonary disease Serum surfactant protein D is steroid sensitive and associated with exacerbations of COPD Serum surfactant protein D during acute exacerbations of chronic obstructive pulmonary disease Comprehensive characterisation of pulmonary and serum surfactant protein D in COPD Serum surfactant protein D: biomarker of chronic obstructive pulmonary disease Value of serum and induced sputum surfactant protein-D in chronic obstructive pulmonary disease Surfactant protein d, soluble intercellular adhesion molecule-1 and highsensitivity C-reactive protein as biomarkers of chronic obstructive pulmonary disease The role of matrix metalloproteinase-9 in cigarette smoke-induced emphysema Surfactant protein D in serum from patients with allergic bronchopulmonary aspergillosis Serum-surfactant SP-D correlates inversely to lung function in cystic fibrosis Circulating surfactant protein-D and the risk of cardiovascular morbidity and mortality The utility of lung epithelium specific biomarkers in cardiac surgery: a comparison of biomarker profiles in on-and off-pump coronary bypass surgery Can serum surfactant protein D or CC-chemokine ligand 18 predict outcome of interstitial lung disease in patients with early systemic sclerosis? Serum levels of surfactant proteins A and D are useful biomarkers for interstitial lung disease in patients with progressive systemic sclerosis Clinical significance of surfactant protein D as a serum marker for evaluating pulmonary fibrosis in patients with systemic sclerosis Comparative study of serum surfactant protein-D and KL-6 concentrations in patients with systemic sclerosis as markers for monitoring the activity of pulmonary fibrosis Surfactant protein D and KL-6 as serum biomarkers of interstitial lung disease in patients with scleroderma Clinical significance of serum pulmonary surfactant proteins a and d for the early detection of radiation pneumonitis Diagnostic significance of surfactant proteins A and D in sera from patients with radiation pneumonitis Study of Clara cell 16, KL-6, and surfactant protein-D in serum as disease markers in pulmonary sarcoidosis Serum surfactant protein D is elevated in allergic patients Involvement of eicosanoids and surfactant protein D in extrinsic allergic alveolitis Functional and biological characteristics of asthma in cleaning workers Circulating surfactant protein D is decreased in early rheumatoid arthritis: a 1-year prospective study Circulating surfactant protein -D is low and correlates negatively with systemic inflammation in early, untreated rheumatoid arthritis Elevated plasma surfactant protein D (SP-D) levels and a direct correlation with anti-severe acute respiratory syndrome coronavirus-specific IgG antibody in SARS patients Analysis of plasma surfactant protein D levels in lung transplant recipients Clinical significance of the serum surfactant protein D and KL-6 levels in patients with measles complicated by interstitial pneumonia Surfactant protein D (SP-D) serum levels in patients with community-acquired pneumonia small star, filled Serum SP-D levels as a biomarker of lung injury in respiratory syncytial virus bronchiolitis Serum KL-6 and surfactant protein D in children with 2009 pandemic H1N1 influenza infection Serum KL-6 and surfactant proteins A and D in pediatric interstitial lung disease Polymorphisms in the human surfactant protein-D (SFTPD) gene: strong evidence that serum levels of surfactant protein-D (SP-D) are genetically influenced Surfactant protein d, a marker of lung innate immunity, is positively associated with insulin sensitivity Circulating levels of Clara cell protein 16 but not surfactant protein D identify and quantify lung damage in patients with multiple injuries Blood biomarkers and measures of pulmonary function-a study from the Swedish twin registry Surfactant protein D of the innate immune defence is inversely associated with human obesity and SP-D deficiency infers increased body weight in mice Long-term stability and circadian variation in circulating levels of surfactant protein D A common polymorphism in the SFTPD gene influences assembly, function, and concentration of surfactant protein D Plasma surfactant protein D levels and the relation to body mass index in a chinese population Plasma C3d levels of young farmers correlate with respirable dust exposure levels during normal work in swine confinement buildings Respiratory effects associated with wood fuel use: a cross-sectional biomarker study among adolescents Roles of serum clara cell protein 16 and surfactant protein-D in the early diagnosis and progression of silicosis Serum surfactant protein-A, but not surfactant protein-D or KL-6, can predict preclinical lung damage induced by smoking Biomarkers of early respiratory effects in smoking adolescents Circulating markers of interstitial lung disease and subsequent risk of lung cancer Inhaled LPS challenges in smokers: a study of pulmonary and systemic effects Lung epithelium injury biomarkers in workers exposed to sulphur dioxide in a nonferrous smelter The effects of GH and hormone replacement therapy on serum concentrations of mannan-binding lectin, surfactant protein D and vitamin D binding protein in Turner syndrome Plasma surfactant D in patients following acute paraquat intoxication Respiratory function and changes in lung epithelium biomarkers after a short-training intervention in chlorinated vs. ozone indoor pools A panel of lung injury biomarkers enhances the definition of primary graft dysfunction (PGD) after lung transplantation The sitting position during neurosurgical procedures does not influence serum biomarkers of pulmonary parenchymal injury A sandwich enzyme immunoassay for pulmonary surfactant protein D and measurement of its blood levels in drowning victims Clinical significance of serum surfactant protein D (SP-D) in patients with polymyositis/ dermatomyositis: correlation with interstitial lung disease Serum surfactant protein D is correlated to development of dementia and augmented mortality Circulating surfactant protein D is decreased in systemic lupus erythematosus Comparison of biomarkers of subclinical lung injury in obstructive sleep apnea Lung epithelium-specific proteins: characteristics and potential applications as markers The association between BMI and plasma cytokine levels in patients with acute lung injury Characterization and prevention of the adsorption of surfactant protein D to polypropylene Collectin surfactant protein D binds antibodies and interlinks innate and adaptive immune systems Increased prevalence of low oligomeric state surfactant protein D with restricted lectin activity in bronchoalveolar lavage fluid from preterm infants The authors would like to thank Tarek Abdalla for his assistance with sample collection and processing. Conceived and designed the experiments: PEB. Performed the experiments: PEB. Analyzed the data: PEB. Contributed reagents/materials/ analysis tools: PEB AG. Wrote the paper: PEB AG.