key: cord-327948-stwmxpbv authors: Dolai, Subhashish; Tabib‐Azar, Massood title: Whole virus detection using aptamers and paper‐based sensor potentiometry date: 2020-08-01 journal: Med Devices Sens DOI: 10.1002/mds3.10112 sha: doc_id: 327948 cord_uid: stwmxpbv Paper‐based sensors, microfluidic platforms, and electronics have attracted attention in the past couple of decades because they are flexible, can be recycled easily, environmentally friendly, and inexpensive. Here we report a paper‐based potentiometric sensor to detect the whole Zika virus with a minimum sensitivity of 0.26 nV/Zika and a minimum detectable signal (MDS) of 2.4x10(7) Zika. Our paper sensor works very similar to a P‐N junction where a junction is formed between two different regions with different electrochemical potentials on the paper. These two regions with slightly different ionic contents, ionic species and concentrations, produce a potential difference given by the Nernst equation. Our paper sensor consists of 2‐3 mm x 10 mm segments of paper with conducting silver paint contact patches on two ends. The paper is dipped in a buffer solution containing aptamers designed to bind to the capsid proteins on Zika. We then added the Zika (in its own buffer) to the region close to one of the silver‐paint contacts. The Zika virus (40 nm diameter with 43 kDa or 7.1x10(‐20) gm weight) became immobilized in the paper’s pores and bonded with the resident aptamers creating a concentration gradient. Atomic force microscopy and Raman spectroscopy were carried out to verify that both the aptamer and Zika become immobilized in the paper. The potential measured between the two silver paint contacts reproducibly became more negative upon adding the Zika. We also showed that a Liquid Crystalline Display (LCD) powered by the sensor can be used to read the sensor output. Sensors built with paper as a substrate have short response time, are low-cost [1] and they are flexible [2] , [3] . Moreover, they are biodegradable and suitable for mass deployment in resource-limited areas and can be easily used by unskilled operators. Paper is also a great medium for immobilization and trapping and, in some cases, for binding with biomolecules. Its porous structure with large connected pores composed of cellulose fibers allows transporting liquid by means of capillary forces that result in short response time. The porous structure of paper also allows any nano and microparticles to remain immobilized in the paper structure. Paper can be functionalized with certain materials such as nitrocellulose paper used for immobilizing nucleic acids for selective sensing [4] . Paper-based potentiometric sensors are reported for detecting many ions and proteins [4] - [7] . Potentiometric paper-based sensors [1] , [8] - [16] utilize gradients of ion distributions that generate open-circuit voltages (V oc ) that are measured to transduce analytes. Paper-based pathogen and virus sensors are also easy to incinerate. They can be used one-time use front end in sensor systems that can be pealed-off and replaced. There are many Zika sensors and detection methods reported in the literature. These include serum analysis using the antibody detection assays [17] , [18] , and detection of viral RNA using molecular-based techniques like conventional polymerase chain reaction (PCR). Real-time reverse transcription-polymerase chain reaction (RT-PCR) [19] - [25] Zika has also been reported. These techniques involve using viral antibody or DNA/RNA extraction, followed by labeled detection using fluorescent probes. They provide high sensitivity and specificity but require specialized equipment, expensive procedures, and are time-consuming. The paper-based sensors are fast, inexpensive, and can be used in regions with limited resources. We show that a simple Liquid Crystalline Display (LCD) can be used to measure the sensor output with the power generated from the sensor itself. The paper sensor reported here can be modified by replacing its aptamer with other virus aptamers, such as COVID-19. Our work shows for the first time that a potentiometric paper sensor can reliably detect a whole virus using standard printer papers functionalized with antigens or aptamers. Aptamers are single strand DNA structures that are artificially synthesized using a combinatorial selection process called systematic evolution of ligands by exponential enrichment (SELEX) [26] to detect biomolecules and pathogens with high specificity [27] , [28] . In our sensors, when Zika was added to one side of the paper, it resulted in a concentration gradient with the associated electrochemical potential difference. Electrical contacts to the paper sensor can be made with Graphene [15] , [16] conducting glues and epoxies and silver paint. We also This article is protected by copyright. All rights reserved developed a proof-of-concept device ( Fig. 2(a) ) that can be printed with silver paint contacts and show that the paper-based Zika sensor can be connected to an LCD for electronic readout. The charge distribution and ionic transport of differently charged species in the paper device and the resulting open-circuit voltage can be explained using the phase boundary model formed by two different ionic species (aptamer in the background and Zika added on the anode side). The total electrochemical potential difference (V oc ) (Fig. 1a) is the sum of the phase boundary potentials [29] between the aptamer-electrode (V oc1 ), the liquid junction potential [30] , [31] between the Zika/aptamer-aptamer (V oc2 ) and between the Zika/aptamer-electrode (V oc3 ): V oc =V oc1 + V oc2 +V oc3 . Fig. 1 .a shows the schematic of the phase boundary between electrodes and the paper and liquid junction potential at the center. The phase boundary potential between the aptamer-electrode (V oc1 ) and Zika/aptamer-electrode (V oc3 ) are obtained from the Nernst equation: where, is the standard potential, R is the gas constant, T is temperature, F is the Faraday constant 0 (Coul./mol), z 1 =1 is the valence of the aptamer, and z 3 =1 is the valence of the Zika. Zika is a complex extended pathogen with surface proteins, DNA, RNA, etc. and has many oxidation/reduction potentials corresponding to more than one charge transfer that we have indicated in Eq. 1. However, under the small voltages and chemically "mild" conditions of our paper sensors, it is reasonable to assume the single valency indicated in Eq. 1. The liquid junction potential between (Zika-aptamer) and aptamer in the middle of the device (Fig.1) can be written as [31, 32] : where, is the valence of the particular sample species (in our case z's for aptamers and the Zika are equal to 1) , denotes the activity of the ions/sample species and is the transference number and This article is protected by copyright. All rights reserved signifies the fractional conductance of the i th ion/sample species. The transference number defined by = , where u is the mobility of the i th ion, and M j is the molar concentration, with j ranging over all the ions. We note that there are 3 space charge regions that form (Fig.1a) between the anode/cathode and the paper and at the boundary of the Zika/aptamer-aptamer regions in the middle. In our experiments, the paper was uniformly primed with the buffer solution and aptamers. We assume that the anode-paper and cathode-paper Nernst potentials (V oc1 and V oc3 in Fig. 1a ) are approximately equal and are determined by the electrochemical potential differences between these electrodes and the primed paper. Thus, V oc1 =V oc3 and the total V oc between the cathode and the anode is approximately equal to V oc2 (Fig. 1a) . The space-charge region at the Zika boundary ( Fig. 1a and 1b) is composed of charges that are induced by the Zika-aptamer-buffer on one side of this junction and aptamer-buffer on the other side ( Fig. 1b) . The boundary is controlled by the amount of Zika added and the flow of the buffer that accompanies the Zika. We note that aptamers and Zika viruses become immobilized in the pores of the paper. According to our experiments, the aptamer and Zika had negative residual charges. Each of these analytes is charge-neutral in their respective medium, but when they are added together, because of their different polarizability, they exhibit different residual charges. In the case of Zika + aptamer, there is an additional charge re-distribution because of the hydrogen binding. The binding process does not happen instantly and proceeds as a function of time. In our experiments, the Zika-aptamer complex provided higher V oc change than Zika alone. The selectivity of the sensor to detect Zika is primarily due to its preferential binding with the aptamer. When we added the Zika to the anode region, V oc became negative, suggesting that the space charge region has a charge, electric field, and potential distributions schematically shown in Figs. 1c, 1d, and 1e. This article is protected by copyright. All rights reserved c)-e) Space charge, resulting in the electric field and potential at the boundary of Zika in the center of the paper device. The devices used in our work ( Fig. 2 ) consisted of a sample holder with two clips and electrodes mounted on a glass slide. The devices were connected to a data acquisition system (NI-USB 6341 DAQ) and This article is protected by copyright. All rights reserved a computer with a custom LabVIEW program (Fig. 2b) . We measured the open-circuit voltage (V oc ) of the paper device as a function of time as buffer, aptamer, and Zika were added to the sensor. The glass slide was located inside a grounded copper enclosure to shield and improve the signal-to-noise ratio. Paper strips (0.2-0.3 cm x 1 cm) used in our devices were manually cut from a standard printer paper and were contacted with silver paint (from Ted-Pella). These paper devices were then coated (impregnated) with appropriate analytes such as de-ionized (DI) water/buffer/aptamer, and the buffer/Zika solutions were added from the anode side. This article is protected by copyright. All rights reserved The aptamer used in our experiments had a thiol end group and consisted of 32 nucleotide-chain that folded to bind with the Zika SF9 envelope protein [33] . Fig. 2c shows the schematic of the structure of the aptamer. The inactivated Zika virus (0810521CFHI) was obtained from the Zeptometrix corporation. 40 µl of TCID_50 Zika stock solution was diluted with 160 µl buffer. The number of Zika viruses in the diluted 2 µl volume was ~7x10 7 , estimated in the same manner as in [34] . The open-circuit voltage (V oc ) of the sensor containing DI water with the buffer solution introduced on the anode is shown in Fig. 3a . The V oc , in this case, was measured immediately after the buffer was added to the anode region. V oc diminishes as a function of time (Fig. 3b) as the buffer diffuses in the paper, and its concentration gradient diminishes. The Zika virus has a small residual negative charge (Fig. 4a ) that reduces the V oc as soon as it is added to the sensor. Fig. 4 shows the response of the sensor when Zika was added. The initial rise in Fig. 4a is due to the pipette tip approaching the sensor to deliver the Zika. Fig. 4b shows another run with the Zika added in the beginning and later at 3.1 min. In both cases, the sensor output reached a steady-state value. This indicates that the concentration gradient introduced by the addition of Zika did not change as a function of time, and it appears that the Zika became immobilized in the paper. In the second case (Fig. 4b) , the paper was not primed with the aptamer, and the change in V oc was due to the residual charge of the Zika and buffer ions that accompany the Zika as part of its solution. It is also possible that the printer paper we are using has its own impurities, which in the presence of Zika + buffer became ionized. Zika is spherical with 40 nm diameter, and it is expected to get entangled in the paper pores. If Zika was mobile, the This article is protected by copyright. All rights reserved resulting V oc would diminish as a function of time as the Zika diffuses and reduces its concentration gradient. In separate experiments with buffer solutions alone (Fig. 3b) , the sensor response diminished after the initial introduction of the buffer at the anode, indicating that the buffer ions could diffuse in the paper, decreasing their initial concentration gradients. in the device charge content and its gradient, are introduced by the additional Zika alone that has negative residual charges. Successive additions of Zika (used to increase its concentration) introduced larger changes in V oc followed by smaller changes once all the resident aptamers were saturated (Fig. 5b) , and ∆V oc became smaller at the end. The average "point" sensitivity in these devices was 0.26 nV/Zika calculated at ∆V oc of ~ 18 mV for 2 µl Zika (~7x10 7 number of Zika per 2 µl). The sensitivity was nearly the same for 4 l Zika in device P1 but not in device P2 (Fig. 5) . The paper-based sensor's output response had a rms noise voltage of 2 mV, and the minimum detectable voltage was 2 mV that resulted in the MDS of 2.4x10 7 Zika. This article is protected by copyright. All rights reserved To verify the presence of aptamers and Zika in the paper-based devices, we carried out two additional experiments. The first series of experiments used atomic force microscopy (AFM) to image Zika-coated regions of the paper-based device, as shown in Fig. 6a . Although one can see large Zika complexes, individual Zika viruses could not be resolved. The paper has large surface roughness in excess of 2 m rms that prevented performing high spatial resolution scans. We then performed AFM scans on aptamers and Zika deposited on gold-coated glass samples [35] . As can be seen in Fig. 6b one could resolve individual Zika viruses on gold. Next, AFM studies using AFM probes with an aptamer-functionalized bead (Fig. 6c) were carried out. The stiction force [36] obtained from Force-displacement curve using aptamer-functionalized AFM probe (Fig. 6c) . The AFM probe has been prepared by gluing a 150 µm diameter polyester microbead to the AFM tip and subsequently coating the microbead with the aptamer (1 µM) for 30 min. It showed larger stiction forces with papers coated with aptamers than Zika papers with aptamers or Zika alone, as shown in Fig. 6d . The stiction force of Zika/aptamer/paper was higher due to aptamer-specific binding with Zika on the aptamer/paper. This article is protected by copyright. All rights reserved The second set of experiments used Raman spectroscopies on paper-based devices coated with different analytes. The Raman spectra shown in Fig. 7 are from different regions of the paper coated with aptamers, Zika, and Zika/aptamers. A DeltaNu spectrometer with Examiner 785 Raman unit was used in these experiments. 10 l, 1 M aptamer solution was used to prime the paper, and the Zika concentration was 1.7x10 8 /l, with the amount used in each application was 1 l. The buffer solution was 1 l of 1x PBS with 1 mM MgCl 2 . The Raman peak due to Zika was around 1590 cm -1 [37] . This article is protected by copyright. All rights reserved Buffer/paper Aptamer/paper Zika/paper Zika+aptamer/paper cm -1 Raman Signal Figure 7 . Raman spectra of buffer/paper, aptamer/paper, Zika/paper, and Zika/aptamer/paper. The spectra are normalized with dry paper Raman spectrum. These spectra clearly show the difference between aptamers, Zika, and buffer deposited on the paper. We then used printed patterns on paper, followed by adding silver paint contacts, as shown in Fig. 2a and Fig. 8 insets. The purpose of the printed device is to clearly show the location of the Zika solution (or fluids that may contain Zika) on the device. Fig. 8a shows the V oc -time response for paper-based devices (D1-D3). The ∆V oc in these devices are shown in Fig. 8b . We verified that the aptamer stayed and was resident in the paper even after rinsing the paper with DI water. Fig. 8c shows the response of adding Zika on three different aptamer-primed sensors (D4-D6). The V oc levels in Fig. 8c are similar and consistent with that of the aptamer-primed device in Fig. 8b . The actual interactions between different components in the paper matrix is quite complex, and the sensor response depends on the history of the paper, temperature, air current on the sensor, and other environmental factors [15] . The results shown here had a standard deviation of ~ 2 mV from run to run. Uniformity of the electrical contacts can be a concern that can be addressed using 3-D printing. Another very important source of error was in making sure that the paper sensor was not saturated with the analytes introduced at each step. When the paper was saturated, the addition of Zika resulted in a thin layer of liquid above the paper sensor, and Zika did not incorporate in the paper resulting in unreliable sensor output. The sensor was also very sensitive to spurious environmental electric fields generated by us and other objects nearby that could be addressed by proper grounding. We next explored the possibility of detecting the virus from the paper-based sensor in the form of direct electronic readout through an LCD instead of measuring the open-circuit voltage using a digital voltmeter. The LCD was configured to display a letter such as "A" once the paper sensor voltage reached a threshold value generated by the presence of sufficient number of Zika viruses. In order to demonstrate the feasibility of such application, we used an LCD screen from a digital watch and experimentally found the turn-on voltage for a single display segment to be ~100 mV. From the previous results, we observed that the addition of Zika resulted in V oc ~ 50 mV. We used 6 paper sensors connected in series to generate ~ 300 mV (Fig. 9a) that was sufficient to power many readout devices, including LCDs. The external LCD load did not draw more than 1 A current to cause internal sensor resistor to become important. As can be seen in Fig. 9b , the LCD displayed "A" upon adding Zika. The LCD can be replaced by an electronic-ink display that requires much lower voltages/power eliminating the necessity of using many devices in series. (a) (b) The paper-based sensors we constructed and reported here were very reproducible as long as they were not saturated (with buffer or water) and as long as they were grounded properly. Their "point" sensitivity of 0.26 nV/Zika and their minimum detectable signal of 2.4x10 7 Zika are quite good given their simple structures and ease of operation. Selectivity of this sensor as well as other aptamer-based sensors are limited by the selectivity of their respective aptamers that can be modified to achieve better selectivity if needed. In the case of paper-based sensors discussed here, the selectivity may also be affected by viruses becoming immobilized in the paper even if they are do not bind with the resident aptamer. These other viruses can be removed by mildly washing them away that will not affect the Zika-aptamer complexes appreciably. The real-world application of Zika paper sensors will involve the presence of bodily fluids such as urine or sweat or saliva. These biofluids are very complex, and proper filtering on the sensor will be used to remove their components that may affect the paper sensor operation adversely. As with any sensor development, the research and development efforts have many different stages. It usually starts with developing a transduction mechanism using modified "clean" samples and progresses toward more realistic samples in the real world. The present work describes the transduction phenomena of "sensing pure Zika" virus using a paper-based potentiometric device that, to the best of our knowledge, was not done before. The innovation has two parts: a) using paper-based sensors to detect the whole virus and b) using resident aptamers in the paper to bind with Zika. The sensor's selectivity comes from its Accepted Article resident aptamers. We will measure and report our sensors' selectivity and verify their operations in the presence of bodily fluid in the near future. The paper can also be used as an active sensing medium located over electrodes connected to the measurement apparatus. In this approach, the paper can be peeled off and replaced with a fresh paper. Paper itself can be used as a filter preventing larger particles from reaching the sensitive layers in intimate contact with the electrodes. This work demonstrates, for the first time a paper-based sensor for detecting Zika virus using potentiometry. The sensor had a minimum sensitivity of 0.26 nV/Zika and a minimum detectable signal of 2.4x10 7 Zika. These sensors can be generalized to detect any other viruses, pathogens and bacteria by simply using appropriate aptamers. The most important and unique aspect of the potentiometric paper-based Zika virus is its structural simplicity and ease of use. This study demonstrates the feasibility of detecting whole Zika viruses and the use of simple electronic readout of paper-based sensors. Although aptamers are known for their high specificity to capture target pathogens and viruses, more study is required with different viruses to verify and measure the cross-sensitivity of aptamers and our sensors. A low-cost paper-based inkjet-printed platform for electrochemical analyses Foldable printed circuit boards on paper substrates Paper electronics A paper-based potentiometric cell for decentralized monitoring of Li levels in whole blood Potentiometric sensing utilizing paper-based microfluidic sampling Towards Protein Assays on Paper Platforms with Potentiometric Detection A Disposable Planar Paper-Based Potentiometric Ion-Sensing Platform Paper-based potentiometric ion sensors constructed on ink-jet printed gold electrodes Integrated, paper-based potentiometric electronic tongue for the analysis of beer and wine A fully integrated paper-microfluidic electrochemical device for simultaneous analysis of physiologic blood ions A paper based, all organic, reference-electrode-free ion sensing platform Paper-based microfluidic sampling and separation of analytes for potentiometric ion sensing Paper-based enzymatic electrode with enhanced potentiometric response for monitoring glucose in biological fluids Paper-based microfluidic sampling for potentiometric Accepted Article This article is protected by copyright. All rights reserved determination of ions Paper-based potentiometric pH sensor using carbon electrode drawn by pencil High-performance flexible potentiometric sensing devices using free-standing graphene paper Zika virus: A new pandemic threat The zika challenge Sensing viruses by mechanical tension of DNA in responsive hydrogels Prolonged detection of zika virus in vaginal secretions and whole blood Detection and sequencing of Zika virus from amniotic fluid of fetuses with microcephaly in Brazil: a case study One-step RT-PCR for detection of Zika virus Detection of zika virus in urine Dengue virus detection using whole blood for reverse transcriptase PCR and virus isolation Detection of Zika virus in saliva Systematic evolution of ligands by exponential enrichment: RNA ligands to bacteriophage T4 DNA polymerase Aptamer-Based ELISA Assay for Highly Specific and Sensitive Detection of Zika NS1 Protein High-resolution molecular discrimination by RNA The phase-boundary potential model The liquid junction potential -the free diffusion junction Calculation of liquid junction potentials Base Pair Biotechnologies -Aptamer Discovery Company 433 MHz Lithium Niobate Microbalance Aptamer-Coated Whole Zika Virus Sensor with 370 Hz/ng Sensitivity Microfabricated Nano-Gap Tunneling Current Zika Virus Sensors with Single Virus Detection Capabilities Recommendations for the use of an atomic force microscope as an in-fab stiction monitor Surface-Enhanced Raman Spectroscopy-Based Sandwich Immunoassays for Multiplexed Detection of Zika and Dengue Viral Biomarkers Accepted Article Esterowitz. This article is protected by copyright. All rights reserved Accepted Article