id author title date pages extension mime words sentences flesch summary cache txt cord-320587-936cavob Ruscio, M. Analytical assessment of Beckman Coulter Access anti-SARS-CoV-2 IgG immunoassay 2020-11-07 .txt text/plain 4092 221 46 This analytical assessment encompassed the calculation of intra-assay, inter-assay and total imprecision, linearity, limit of blank (LOB), limit of detection (LOD), functional sensitivity, and comparison of anti-SARS-CoV-2 antibodies values obtained on paired serum samples using DiaSorin Liaison SARS-CoV-2 S1/S2 IgG and Roche Elecsys Anti-SARS-CoV-2 total antibodies. The assessment of this novel Beckman Coulter Access SARS-CoV-2 IgG immunoassay on UniCel DxI800 has originally encompassed the calculation of intra-assay, inter-assay and total imprecision, linearity, limit of blank (LOB), limit of detection (LOD), functional sensitivity, as well as comparison of anti-SARS-CoV-2 antibodies values obtained on paired serum samples using two other commercial anti-SARS-CoV-2 immunoassays, as comprehensively described below. As for the diagnostic performance calculated using manufacturers' cut-off, the AUCs of Beckman Coulter Access SARS-CoV-2 IgG (p<0.001) and Roche Elecsys Anti-SARS-CoV-2 (p=0.01) antibodies titers were found to be both significantly higher than that of DiaSorin Liaison SARS-CoV-2 S1/S2 IgG, whilst they were not significantly different between them (p=0.785). ./cache/cord-320587-936cavob.txt ./txt/cord-320587-936cavob.txt