key: cord-296109-kco85lqn authors: Vanuytsel, Kim; Mithal, Aditya; Giadone, Richard M.; Yeung, Anthony K.; Matte, Taylor M.; Dowrey, Todd W.; Werder, Rhiannon B.; Miller, Gregory J.; Miller, Nancy S.; Andry, Christopher D.; Murphy, George J. title: Rapid Implementation of a SARS-CoV-2 Diagnostic qRT-PCR Test with Emergency Use Authorization at a Large Academic Safety-Net Hospital date: 2020-05-19 journal: Med DOI: 10.1016/j.medj.2020.05.001 sha: doc_id: 296109 cord_uid: kco85lqn Summary The COVID-19 pandemic is a global public health crisis. Significant delays in the rapid development and distribution of diagnostic testing for SARS-CoV-2 infection have prevented adequate public health management of the disease, impacting the timely mapping of viral spread and the conservation of personal protective equipment. Furthermore, vulnerable populations such as those served by Boston Medical Center (BMC), the largest safety net hospital in New England, represent a high-risk group across multiple dimensions, including a higher prevalence of pre-existing conditions and substance use disorders, lower health maintenance, unstable housing, and a propensity for rapid community spread, highlighting the urgent need for expedient and reliable in-house testing. Here, we report the implementation of a SARS-CoV-2 diagnostic RT-PCR assay with rapid turnaround time enabling more informed decisions with personal and public health ramifications. This work provides a blueprint for academic centers and community hospitals lacking capital-intensive automated laboratory machinery to implement in-house testing. The coronavirus disease 2019 (COVID-19) pandemic, caused by the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), is now a major global public health crisis. In the United States, significant delays in the rapid development and distribution of diagnostic testing for SARS-CoV-2 infection have prevented adequate COVID-19 patient care and public health management of the pandemic (Sharfstein et al., 2020) , impacting the timely mapping of the dynamics of viral spread in the general population, and more topically, the conservation of personal protective equipment. Furthermore, vulnerable populations such as those served by Boston Medical Center (BMC), the largest safety net hospital in New England, represent a high-risk group across multiple dimensions, including a higher prevalence of pre-existing conditions and substance use disorders, lower general health maintenance, unstable housing, and a propensity for rapid community spread, highlighting the urgent need for rapid and reliable in-house testing infrastructure. We requested and received emergency permission from the Food and Drug Administration (FDA) and the Commonwealth of Massachusetts Department of Public Health to repurpose a BMC-affiliated biosafety level 2 plus (BSL-2+) basic science research facility into a clinical diagnostic laboratory for high complexity testing. It now functions as an approved site of the BMC Department of Pathology & Laboratory Medicine, as an extension of their Clinical Laboratory Improvement Amendments (CLIA) and College of American Pathologists (CAP) accredited space. Subsequently, we implemented a quantitative, real-time reverse transcriptase polymerase chain reaction (qRT-PCR)-based assay to detect viral SARS-CoV-2 RNA from nasopharyngeal swabs, based on guidelines from the Centers for Disease Control and Prevention (CDC) and the FDA for use with in-house testing of BMC patient samples ( Figure 1 ) (CDC, 2020; Wang et al., 2020a). RNA extraction from patient swabs was performed via the QIAGEN® RNeasy Minikit (Cat #74106), to its relative abundance in the context of a limited supply of similar reagents. cDNA was prepared from patient sample RNA, followed by qRT-PCR to interrogate expression of the SARS-CoV-2 nucleocapsid protein RNA (N1, N2) as well as human Ribonuclease P (RNase P, internal housekeeping control) (IDT Cat #10006606). The assay has been submitted for approval under an FDA Emergency Use Authorization (EUA). Our medium-throughput assay enables the simultaneous testing of over 50 patient samples on a single 384-well qRT-PCR plate (including requisite controls). In order to minimize errors associated with manual interpretation and data transfer, we generated an automated data analysis and reporting pipeline capable of importing unmanipulated qRT-PCR data (including cycle threshold values for each gene per patient sample). All team members completed expedited BMC onboarding, including training protected health information (PHI) training. This automated pipeline subsequently interfaces with our institution's laboratory information system (SunQuest), allowing for direct upload into BMC's electronic medical record (Epic), further reducing overall turnaround time from sample intake to result reporting. This script is publicly available (https://github.com/TaylorMatte/Quant6-Covid_Analysis). Following FDA EUA submission requirements, our in-house SARS-CoV-2 qRT-PCR Laboratory Derived Test (LDT) was launched March 23 rd , 2020. We are currently actively testing both BMC patients and healthcare workers. Notably, the development of the described in-house testing platform greatly reduced our institutional sample turnaround-time (TAT) to under 24 hours, which had previously been 5-7 days when specimens were sent out to commercial entities and state public health agencies. As of April 6 th , 2020, we have successfully completed testing on >1000 unique patient samples, of which 44.1% were positive for SARS-CoV-2 infection, a higher ratio than other Massachusetts testing facilities per publicly available datasets ( Figure 2A ). Our assay has displayed robust accuracy and reliability in validation studies performed to fulfill FDA requirements, with a limit of detection (LOD) of 10 1 viral copies/µL (based on serial dilutions of commercially available positive control from ATCC (Cat# VR-3276T) and an average cycle threshold value (in positive samples) for the N1 probe of 24.83 ± 5.35 (n=94, mean ± SD) ( Figure 2B ). Additionally, our standard curve shows minimal run-to-run variability (n=6) ( Figure 2C ). In response to a critical testing shortage in our large urban, safety net, academic medical center, we converted our research laboratory to a CLIA certified SARS-CoV-2 diagnostic facility in seven days. Notably, our ability to extend CLIA certification from a BMC Pathology clinical facility to include basic research laboratory space was unprecedented and instrumental in our ability to test patient samples in an expedient manner, particularly at a point during the pandemic when commercial laboratory TAT for samples sent from our hospital ranged from 5-10 days. Although our hospital does possess a high throughput Roche Cobas® system for molecular testing, a nationwide shortage in the supply and distribution of SARS-CoV-2-specific kits led us to adapt and implement an in-house, qRT-PCR-based, FDA EUA diagnostic using readily sourced reagents. Our test offers a more rapid TAT in comparison to currently available commercial or state laboratory facilities, and enables clinicians and patients to make more informed decisions with significant personal and public health ramifications. Following the initial inception of this assay, and to increase the utility of our platform, bridging studies have now been performed to add additional flexibility and throughput. Notably, we have the ability to run a one-step PCR reaction (omitting a separate RT cDNA step), to multiplex samples using multiple fluorophores, and to also validate multiple swab methodologies to overcome potential bottlenecks in the supply chain. Our ability to perform several variations of our core assay ensures that if any single element of our supply chain becomes unstable, we can seamlessly pivot to an alternate approved strategy under our EUA. Furthermore, in contrast to commercial automated sample-to-answer platforms, the inherent granularity of our system is also amenable to experimental validation studies and COVID-19 related scientific inquiry. Our 44.1% rate of positivity is higher than the overall state average of 18.1%, a potential indicator that we are sampling from a higher risk population. Given that our hospital takes care of a large portion of Boston's homeless and housing unstable patients, rapid testing is vital in order to mitigate community spread in an already vulnerable population living in low resource settings in which social distancing is not possible. Based on publicly available data, 46% (roughly 4 in 10 patients) of BMC's inpatient census on April 6-7 th , 2020 (approaching the peak of the pandemic in Massachusetts) was comprised of COVID-19 patients (Wen, 2020) . This point is further highlighted by our urban location as zip codes in socioeconomically disadvantaged neighborhoods in Boston have higher than average levels of COVID-19 when compared to city-wide data (BPHC, 2020). Taken together, this data reflects emerging preliminary numbers of COVID-19 outcomes from several states and cities across the country that suggest that ethnic minorities and the underserved face higher risks of significant novel coronavirusrelated disease. Finally, we have been able to implement comprehensive testing before larger, more well-resourced institutions in our community, suggesting that the template and protocol that we have generated and validated could be useful for other smaller community hospitals lacking capital intensive automated clinical laboratory machinery to run molecular biology assays. It is important to note that LDTs such as the one described herein must be validated for reagents, hardware, and software by individual sites that implement their own in-house testing assays. Unlike sample to answer assays that are primarily run on commercially available clinical diagnostic hardware, our LDT enables customization at various stages of the assay in response to reagent supply chain availability. However, the manual nature of our assay does potentially increase the risk of human error. The authors declare no competing interests. Lead Contact: Further information and requests for resources and reagents should be directed to and will be fulfilled by the Lead Contact, George J. Murphy (gjmurphy@bu.edu) Material Availability: This study did not generate new unique reagents. The script generated during this study to analyze raw qRT-PCR data and interface with the SunQuest Laboratory Information System is available at GitHub [(https://github.com/TaylorMatte/Quant6-Covid_Analysis)]. Samples were collected from patients at Boston Medical Center who had been designated for COVID-19 testing by their attending physician. These included patients from ambulatory settings, the emergency department, as well as inpatients including those in intensive care units. Patient samples were collected at the point of care via nasopharyngeal swab, and inactivated/lysed in the clinical microbiology laboratory. 300µL of patient sample was mixed with 300uL buffer RLT (from the Qiagen RNeasy Mini Kit), and 600µL 100% molecular grade EtOH. Prior to inactivation and RNA extraction, patient samples can be stored at 4ºC for short-term or -20ºC for long-term storage. RNA was then extracted via the Qiagen RNeasy Mini Kit as per the manufacturer's protocol. RNA was then stored on ice for same day use or at -80ºC for long term storage. To generate synthetic SARS-CoV-2 RNA, commercially available stock solutions from ATCC (Cat No. VR-3276) were used as per the manufacturer's specification. Serial dilutions of the stock viral RNA were performed using 8 µL of dsDNA solution in 72 µL H 2 O for each step to generate dilutions of 10 6 , 10 5 , 10 4 , 10 3 , 10 2 , 10 1 , 10 0 genome copies/µL. cDNA was generated from patient sample-derived RNA using the High Capacity cDNA reverse transcription kit, as per the manufacturer's instructions. cDNA RT master-mix was prepared as follows: 10X RT Buffer (2.0µl/reaction), 25X dNTP Mix (0.8µl/reaction), 10X RT Random Primers (2.0µl/reaction), and RT enzyme (1.0µl/reaction). 14.2µL of clinical sample RNA was added to 5.8µL cDNA RT master-mix. The RT reaction was carried out in a thermocycler with the following incubation times: 25ºC for 5 minutes, 37ºC for 120 minutes, and 85ºC for 5 minutes. The 2019-nCoV kit from Integrated DNA Technologies (IDT: Cat No 10006606) containing qRT-PCR primers for viral genes N1 and N2, as well as human housekeeping gene RP, was thawed and aliquoted as per the manufacturer's recommended instructions. Master mixes were prepared using the following recommended dilutions: Nuclease Free Water (2.33µl/reaction), PCR Primer/Probe (1µl/reaction), and 2X TaqMan Fast Advanced Master Mix (6.66µl/reaction). 10µL of each master mix along with 5µL of sample cDNA or standard curve controls were loaded into individual wells of a 384-well PCR plate (See Additional Resources for Plate Maps and Master Mix Template). The plate was then sealed, centrifuged briefly, and loaded into the QuantStudio 6 Flex Real Time PCR System. The PCR reaction was carried out with the following incubation times: Preamplification (UNG Inactivation), 1x, 120 seconds 25ºC; Preamplification (Polymerase Activation), 1x 120 seconds 95ºC; Amplification, 45x 3 seconds 95ºC and 30 seconds 55ºC. Full instrument settings and parameters can be found in the Additional Resources Section. Data was analyzed using an automated script described above. Individual plates were checked for adequate amplification of the control standard curve, and samples were called positive or negative based on the following guidelines from the CDC (Note: C T values above 40 were considered negative): N1+, N2+, RP± indicates positive for SARS-CoV-2, either N1 or N2+, RP± indicates an inconclusive result, N1-, N2-, RP+ indicates negative for SARS-CoV-2, and N1-, N2-, RP-indicates an invalid result. Boston COVID19 Report-For the Week Ending CDC 2019-Novel Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel, D.o.V. Diseases, ed Diagnostic Testing for the Novel Coronavirus Clinical Characteristics of 138 Hospitalized Patients With 2019 Novel Coronavirus-Infected Pneumonia in Detection of SARS-CoV-2 in Different Types of Clinical Specimens Coronavirus cases surge at Boston Medical Center; nearly half of all beds taken up by such patients SARS-CoV-2 Viral Load in Upper Respiratory Specimens of Infected Patients We would like to acknowledge Anna Engler, Claire Burgess, and Andrew McCracken from the Center for Regenerative Medicine of Boston University/Boston Medical Center for their technical assistance in running this assay. In addition, we would like to thank Samar Mehta and Pardis Sabeti of the Broad Institute for their technical insights and Quantification of RT-PCR C T Values and standard curves was performed using GraphPad Prism 7. The automated analysis and reporting was performed using R Studio. Our full pre-EUA submission and detailed protocol (including a comprehensive list of all reagents and controls) are publicly available at http://www.bu.edu/dbin/stemcells/covid-19.php. In a setting of widespread community transmission of SARS-CoV-2 (COVID-19), it is vital to rapidly and reliably identify and isolate cases to mitigate and ultimately eradicate viral spread. National failures to launch widespread testing in the United States have necessitated academic and non-governmental institutions to fill this critical unmet need. Furthermore, it is particularly important for testing to be widely available to at-risk communities, which are greatly impacted by the current pandemic. This study describes the rapid implementation of a COVID-19 diagnostic test at a large safety net hospital that primarily treats an underserved population and aims to provide a blueprint for other institutions to navigate the logistical and regulatory challenges to provide equitable access to COVID-19 testing for all. • Basic research laboratories can be re-purposed into COVID-19 testing sites.• Widespread testing is essential in underserved communities to contain viral spread.• This study offers a blueprint to navigate challenges in implementing in-house testing.eTOC Paragraph:In this study, we provide a blueprint to quickly "stand-up" an in-house SARS-CoV-2 (COVID-19) RT-PCR-based diagnostic assay at a large academic safety-net hospital that predominantly serves at-risk and underserved populations, resulting in greatly improved turn-around times and conservation of PPE.