id author title date pages extension mime words sentences flesch summary cache txt work_3t3ykl4vwfawdlokw7k534wcgu Jasmine A. Oliver The Mobius AIRO mobile CT for image-guided proton therapy: Characterization & commissioning 2017.0 7 .pdf application/pdf 4912 484 64 Mobile CT System, Varian EDGE radiosurgery system cone beam CT (CBCT), Philips Brilliance Big Bore 16 slice CT simulator, and Siemens SOMATOM Definition AS 20 slice CT transfer function, scaling discrepancy, geometric distortion, spatial resolution, overall uniformity, minimum uniformity, contrast, high CNR, and maximum HU deviation were Localization accuracy and CT Dose Index were measured and compared to reported values on each imaging device. Patient localization accuracy is particularly important in proton beam radiation therapy due to the sharp dose fall-off compared to conventional x-ray AIRO, images were acquired with soft, standard, and sharp reconstruction kernels with the pre-set clinical head protocol. discrepancy, geometric distortion, spatial resolution, overall uniformity, minimum uniformity, contrast, CNR, and maximum HU deviation (Table 2).14 The mean and standard deviation of five or more AIRO/Philips image quality differences were within 3% for overall uniformity, minimum uniformity, and contrast and high CNR. ./cache/work_3t3ykl4vwfawdlokw7k534wcgu.pdf ./txt/work_3t3ykl4vwfawdlokw7k534wcgu.txt